+ All Categories
Home > Documents > André Luiz Sehnema and Antônio Martins Figueiredo Neto ...André Luiz Sehnema and Antônio Martins...

André Luiz Sehnema and Antônio Martins Figueiredo Neto ...André Luiz Sehnema and Antônio Martins...

Date post: 04-Feb-2021
Category:
Upload: others
View: 0 times
Download: 0 times
Share this document with a friend
14
arXiv:1911.11799v1 [cond-mat.soft] 26 Nov 2019 Dynamic response of a thermoelectric cell induced by ion thermodiffusion André Luiz Sehnem a and Antônio Martins Figueiredo Neto Institute of Physics, University of São Paulo, CEP 05508-090, São Paulo, Brazil (Dated: August 2018) Abstract Aqueous electrolyte solutions under the influence of a temperature gradient can generate ther- moelectric fields, which arise from different responses of the positive and negative charges. This is related to the thermodiffusion effect which is quantified by the ionic heat of transport and the thermal diffusion coefficient. When a solution is confined between two metallic electrodes with different temperatures, it is expected to measure the electrostatic potential difference due to the thermoelectric field. We performed experiments in aqueous electrolyte solutions, and our results show that a thermoelectric field appears instantly in the solution as the temperature difference between the electrodes stabilizes. This field arises from the trend to separate both ionic charges in the temperature gradient. The experimental results also show that a slow change in the potential difference is observed, which is related to thermodiffusion in the entire cell. These results are im- portant to understand how the thermoelectric response can be optimized, given the broad interest in the literature to generate thermoelectric energy from thermoelectric cells. a Corresponding author: [email protected] 1
Transcript
  • arX

    iv:1

    911.

    1179

    9v1

    [co

    nd-m

    at.s

    oft]

    26

    Nov

    201

    9

    Dynamic response of a thermoelectric cell induced by ion

    thermodiffusion

    André Luiz Sehnema and Antônio Martins Figueiredo Neto

    Institute of Physics, University of São Paulo,

    CEP 05508-090, São Paulo, Brazil

    (Dated: August 2018)

    Abstract

    Aqueous electrolyte solutions under the influence of a temperature gradient can generate ther-

    moelectric fields, which arise from different responses of the positive and negative charges. This

    is related to the thermodiffusion effect which is quantified by the ionic heat of transport and the

    thermal diffusion coefficient. When a solution is confined between two metallic electrodes with

    different temperatures, it is expected to measure the electrostatic potential difference due to the

    thermoelectric field. We performed experiments in aqueous electrolyte solutions, and our results

    show that a thermoelectric field appears instantly in the solution as the temperature difference

    between the electrodes stabilizes. This field arises from the trend to separate both ionic charges in

    the temperature gradient. The experimental results also show that a slow change in the potential

    difference is observed, which is related to thermodiffusion in the entire cell. These results are im-

    portant to understand how the thermoelectric response can be optimized, given the broad interest

    in the literature to generate thermoelectric energy from thermoelectric cells.

    a Corresponding author: [email protected]

    1

    http://arxiv.org/abs/1911.11799v1

  • I. INTRODUCTION

    The thermoelectric effect is observed when a temperature gradient exists in an aqueous

    solution with dispersed charged species. The ions diffusion response to the temperature gra-

    dient creates a concentration gradient and induces a thermoelectric field. A potential differ-

    ence develops between two points in the liquid with different temperatures [1, 2]. In aqueous

    electrolyte solutions, the difference between the thermodiffusion coefficients of cations and

    anions is one of the sources to the thermoelectric field [3–5]. A simple way to prove this effect

    is to perform experiments in a thermoelectric cell, where the electrolyte solution stays be-

    tween two metallic electrodes at different temperatures [6–9]. The ratio between the electric

    potential drop or thermovoltage △V, as measured between the electrodes, and the applied

    temperature difference △T defines the Seebeck coefficient S = −△V/△T . Recently, some

    studies have reported expressive values (as high as 10 mV/K) for S in nonaqueous electrolyte

    solutions [7, 10] after optimization of the electrode/liquid interface [11, 12].

    Some theoretical studies have been developed to predict S values in aqueous electrolytes,

    assuming the knowledge of the thermal diffusion coefficient or the heat of transport of

    different ions in solution. A noteworthy equation, which relates the Seebeck coefficient to

    ion thermodiffusion, is S = (α+ − α−) kB/e and was worked out by A. Wurger [1, 5], where

    kB is the Boltzmann constant, e the electric charge, and αi = Q∗i /2kBT are the single-

    ion Soret coefficients of ions related to the heat of transport Q∗i as obtained by Agar et al

    [3, 13, 14] for ions in water, and i represents anions and cations. This result was obtained

    considering the stationary state of ion diffusion in the liquid as due to a temperature gradient.

    Whereas this theory predicts S values to be in the range 0.05-0.3 mV/K for different aqueous

    electrolytes, the thermoelectric experiments usually give S values much higher than 0.3

    mV/K, even if common salts such as NaCl are used as electrolytes [7]. The stationary

    state is often considered in ion thermodiffusion due to the lack of a consistent description

    of charge dynamics in solution. Recently, Stout and Khair [15] analyzed theoretically a

    thermoelectric cell filled with an electrolyte and obtained time-dependent expressions for

    thermovoltage and ion distribution in the cell. Their main results show that for typical

    experimental parameters (distance between electrodes about 1 mm and ion concentrations

    around 10 mM), the time of the thermoelectric response from charge separation is below

    1 microsecond, after an instantaneous temperature gradient is established. This charge

    2

  • separation occurs near the electrode surfaces. The relationship between the characteristic

    time τφ of thermovoltage developed and Debye length κ−1, τφ ∝ 1/Diκ

    2, explains this fast

    response [15]. Di is the ion diffusion constant. This theoretical result is a consequence

    of solving the set of equations (thermophoretic-induced ion current density, charge density,

    Poisson equation, and heat equation), assuming equal diffusion coefficients Di for anions and

    cations. Some experimental studies reported a very fast induced thermovoltage associated

    with charge separation as the temperature difference develops between the electrodes [7, 16–

    18].

    In the meantime, a long timescale response is also often reported for the thermovoltage

    in thermoelectric cells, which could be associated with the classical thermodiffusion char-

    acteristic time τc given by τc ∝ l2/Di, where l is the distance between electrodes [19, 20].

    In their theoretical study, Stout and Khair [15] also demonstrated that ion thermodiffusion

    through the thermoelectric cell takes a characteristic time τ ∝ l2/Di, as reported before in

    classical studies on the thermoelectric cell geometry [19–21]. This time dependence is due

    to the slow thermodiffusion or salt Soret effect that does not create a net charge separation

    in the bulk liquid (beyond the Debye layer) inside the cell. Thus, a change in thermovoltage

    due to the slow ion thermodiffusion towards the hot or cold electrode is not theoretically

    expected.

    The physical picture changes if different diffusion constants are considered for anions

    and cations (D− 6= D+). Chikina et al [22] have considered that a local charge separation

    (between ion pairs) occurs when the diffusion coefficients are different for anions and cations,

    thus generating the Seebeck effect in the cell. As a consequence, an electric field could be

    expected to appear through the solution due to different values for Di, in the same timescale

    as the concentration gradient develops in the entire cell. As both ion species diffuse through

    the solution due to the thermodiffusion effect, one faster than the other, an electric field

    is spontaneously generated from the ambipolar diffusion effect. In a recent theory, Janssen

    and Bier [23] obtained an equation for thermovoltage containing a slow response due to

    different diffusion coefficients and a fast response related to the difference in single-ion Soret

    coefficients [23], including Stout and Khair’s result as the specific case to D+ = D− condition.

    Electric fields from ion concentration gradients have been widely investigated in the process

    of inducing migration of nano and microparticles [24–26], showing the importance to describe

    it precisely.

    3

  • In this study, the thermoelectric voltage induced in a thermoelectric cell is experimentally

    investigated in some aqueous monovalent electrolytes. The electrolytes include acids, hydrox-

    ides, and salts, which have either positive or negative values for (α+ − α−) and (D+ −D−),

    as known from the literature [3]. Our main goal is to verify their thermoelectric response in

    the different timescales. We verify, for the first time, that some aspects predicted by recent

    theories [23] appear in the experimental results, like an opposite fast thermoelectric field

    between acids and hydroxides, and a more intense field for electroytes with high difference

    of diffusion constants D+ −D− and sum of ion Soret coefficients α+ + α−.

    II. EXPERIMENTAL SECTION

    The experimental setup of the thermoelectric cell consists of a single-volume cavity con-

    taining the liquid material between two metallic electrodes. This container consists of a solid

    Teflon-made annular ring (inner diameter: 16 mm; wall thickness: 2mm; height: 8mm). Two

    O-rings are used to seal the container by pressing the metallic electrodes with the Teflon

    separator. The distance between the electrode surfaces is kept fixed (l = 11± 1 mm) in all

    experiments, as shown in Figure 1. The solution was transferred to the container using two

    needles. They connected the inner and outer parts through the O-ring and were removed

    after the chamber was filled. Photographs of the experimental setup are available in the

    Supplementary Material. The air exposure of the Teflon and O-rings outside the container

    implies a low horizontal temperature gradient to be formed at their borders, as these parts

    have a low thermal conductivity. Thus, the horizontal gradient was considered negligible

    here. The electrodes are made of stainless-steel metallic blocks (thick: about 5 mm). A

    U-shaped inner channel was built in the bottom electrode for water circulation, and the

    temperature was controlled by a Julabo thermal bath. The temperature difference between

    the electrodes was established applying an electrical current to a 4 Ohm electrical resistance

    on the top of the upper electrode. An electrically insulating and thermally conductive tape

    separated the resistance from the upper electrode surface, avoiding static charges and uni-

    formly heating up the electrode. A DC electrical current around 1 A was applied to raise its

    temperature 10 K above the temperature at the bottom electrode, which was Tcold = 293 K

    for all experiments. In both electrodes, temperature was measured using a type-K thermo-

    couple connected to a digital multimeter. Its tips were connected to the electrodes (deep:

    4

  • 5 mm) through holes (diameter: 1 mm) made in the lateral part of the electrodes. The

    electrodes were designed to be supported by a sample holder (Solartron; 12962) as shown in

    the Supplementary material. The voltage difference between the electrodes was measured

    by a multimeter (Agilent U1252A), which was connected to a computer for automatic data

    acquisition. The data acquisition program records a voltage difference value every three

    seconds.

    (a) (b)

    Figure 1: (a) Illustration (out of scale) of the thermoelectric cell in a front view showing

    the electrolyte solution isolated inside the cell. (b) Graph showing the time necessary for

    temperature (in red) stabilization in the cell and the corresponding change in the

    thermoelectric potential difference (in black) in a salt solution.

    Before applying the temperature difference, it was necessary to wait for the spontaneous

    potential difference (SPD) between electrodes to stabilize. This effect is due to a specific

    interaction between the electrodes’ interface and electrolytes, where a passivated layer is

    formed due to oxidation or reduction of the metallic layer exposed to the liquid. It is also

    related to electrons the multimeter injects in the circuit in order to measure the potential

    difference after filling of the container with the liquid, inducing a capacitive effect with time

    constant higher than 104 seconds. The stabilization process lasted a few hours to observe an

    equilibrium value for SPD when the distance between electrodes was as great as 11 mm and

    the stable SPD values varied from 0 to about 50 mV (absolute value). In Figures S3 and

    S4 of the Supplementary Material we show this stabilization, mentioned as waiting time.

    Temperature takes about 10 min (Figure 1b) to stabilize, and this is the characteristic time

    5

  • for heat diffusion τth through the cell τth = l2/πDth, where Dth is the thermal diffusivity

    of the solution (about 1.4 x 10-7 m2/s for aqueous electrolytes). Duration of experiments

    depends on the Di values for the charge species in solution, taking 3-7 h for thermovoltage

    to stabilize. When a stationary thermovoltage value is reached, the temperature difference

    is removed and the voltage difference between electrodes is recorded for at least the same

    time as in the presence of temperature difference. If separation of ion charges at the electric

    double layer and their thermodiffusion in the entire cell are reversible, one can expect the

    measured voltage difference to return to a similar value before the temperature difference was

    applied. This cycle is repeated at least three times for each of the electrolytes investigated,

    and the observed amplitudes are averaged for calculation of the Seebeck coefficient.

    The solutions were prepared by diluting acids, hydroxides, and salts in deionized water.

    The used ion concentrations cexp0 are shown in Table 1. We investigate aqueous solutions of

    NaOH (sodium hydroxide), TMAOH (tetramethylammonium hydroxide), TBAOH (tetra-

    butylammonium hydroxide), HCl (hydrochloric acid), HNO3 (nitric acid), LiCl (lithium chlo-

    ride), KCl (potassium chloride), and NH4I (ammonium iodide). The chemicals (TMAOH,

    TBAOH, NaOH, HNO3, and HCl) were purchased from Sigma-Aldrich and used without fur-

    ther purification. The NH4I solution was prepared in the Institute of Chemistry, University

    of São Paulo.

    III. RESULTS AND DISCUSSION

    Figure 2 shows experimental results obtained for the thermovoltage VT (t) = ∆Vf(t) +

    ∆Vs(t) in the thermoelectric cell. The subscripts f and s stand for the fast and slow ef-

    fects generating VT . It is possible to observe a fast change corresponding to times when the

    temperature difference stabilizes (Figure 1), followed by a slower change. The experimental

    values to the thermovoltage for the slow response is read from the moment the temperature

    difference is on until the inflection point, while the slow response is defined as the ther-

    movoltage from the inflection point until the stationary state. The equations obtained by

    Janssen and Bier [23] for the dynamics of ion charges in the thermoelectric cell were used for

    the physical interpretation of these results. They derived equations for VT (t) as a function

    of α± and D± and for ion concentration profiles ci (t) in the thermoelectric cell.

    In the presence of a temperature gradient (in the x direction of the cartesian coordinates),

    6

  • (a) (b)

    Figure 2: Potential difference ∆V between cell electrodes as a function of time for

    electrolyte solutions. (a) typical result obtained with aqueous HCl (blue curve) and LiCl

    (red curve) solutions. The time when the temperature gradient is initiated is indicated as

    “∆T on” and when it is removed is indicated as “∆T off”. (b) typical result obtained with

    aqueous TMAOH solution. The amplitudes of ∆Vf and ∆Vs are also indicated, showing

    that ∆Vf is associated with electric fields in different directions for acids and bases and

    ∆Vs is due to a similar electric field for different salts. The amplitudes for fast and slow

    responses are considered to be separated by the inflection point.

    the ion flux equations are as follows [1, 15]:

    j± = −D±n±

    (

    ∂ lnn±∂x

    −z±e

    kBTEx + 2α±

    ∂ lnT

    ∂x

    )

    , (1)

    where n± is the ion density of cations (+) and anions (-) with valence z±, and Ex is the

    thermoelectric field. Based on the experimental results, Ex has a time dynamics with one

    inflection point separating different transient times of a fast and a slow contribution, Ex =

    Ef+Es,. We will discuss our experimental results based on an equation to the thermovoltage

    recently obtained by Janssen and Bier [23]. From the set of governing equations (Equation 1,

    heat transfer, Poisson and conservation equations), they obtained the theoretical expression

    to the thermovoltage with two transient times. This leads to the understanding of the

    dynamic thermovoltage seen in the experimental results of Figure 2.

    The theoretical equation to the thermovoltage VT (t) obtained by Janssen and Bier [23],

    7

  • is given by

    VT (t)

    ∆T=

    kB(α+ − α−)

    e−

    2kB(α+ + α−)(D+ −D−)

    e(D+ +D−)

    j≥1

    1

    N j2exp[−

    2N j2D+D−t

    l2(D+ +D−)]

    −4kB(α+D+ − α−D−)

    e(D+ +D−)

    j≥1

    1

    N j2exp[−

    (D+ +D−)κ2t

    2]

    , (2)

    where Nj = (j−1/2)π. By taking any pair of values for D± the second term evolves in time

    many orders of magnitude slower than the last term, with separation between electrodes

    surface l ≈ 10mm. This allows us to recognize the faster response in our experiments ∆Vf

    in the last term of equation 2, while the slow response ∆Vs is the second term of equation

    2. We use this result in the limit t → inf, when the time dependences converge to 1, to

    compare the thermovoltage amplitudes and verify that our experiments are described by the

    theoy of equation 2.

    The thermoelectric field due to ∆Vf in the fast timescale appears in opposite directions

    for different electrolytes. ∆Vf (t) increases in acidic solutions but decreases in basic solutions

    during temperature stabilization. It means that α+D+ − α−D− should have opposite signs

    for acids and hydroxides. According to the definition αi = Q∗i /2kBT [1] and taking the Q

    ∗i

    values from Agar et al [3], we see that this is the case. In Table 1, we show our experimental

    results and those expected by the use of Agar et al [3] Q∗i -values for comparison of the

    algebraic sign(α+D+−α−D−) and we see that they are consistent. For acids and bases, the

    Q∗i and Di values for H+ and OH− ions are higher than those of counterions, thus explaining

    the opposite direction of the thermoelectric field associated with ∆Vf . The same algebraic

    sign is not observed between ∆Vf and α+D+ − α−D− for some salts (LiCl and NH4I), in

    the adopted convention. Related to the amplitudes, an agreement within the same order

    of magnitude is obtained between experiments and what is theoretically expected from the

    Seebeck component Sf amplitude. As a general trend, the highest experimental Sf values

    are observed for salts with the highest α+D+ − α−D− theoretical difference.

    Table I also presents the total Seebeck coefficient values containing the slow thermodif-

    fusion effect, Stotal = −(∆Vf +∆Vs)/∆T . The theoretical values for Stotal calculated from

    equation 2 are similar to the experimental values for acids and bases, but in different order

    of magnitude for other salts. For NaOH and TMAOH the calculations of Stotal are actually

    the same values as the experimental. The results follow the rule of highest(lowest) value

    8

  • - TBAOH TMAOH NaOH HNO3 HCl LiCl NH4I KCl

    cexp0

    (mM) 4.4 2.8 2.0 1.0 2.0 10 10 10

    Q∗+

    (kJ/mol) 20.79 10.00 3.46 13.3 13.3 0.53 1.73 2.59

    Q∗−

    (kJ/mol) 17.2 17.2 17.2 -0.63 0.53 0.53 -1.55 0.53

    D+ (10−9m2/s) 0.51 1.2 1.33 9.31 9.31 1.03 1.95 1.96

    D− (10−9m2/s) 5.32 5.32 5.32 1.9 2.03 2.03 2.04 2.03

    Stheof

    (mV/K) 0.49 0.43 0.46 -0.396 -0.378 0.006 0.057 0.035

    Stheototal

    (mV/K) -2.1 -1.3 -1.1 0.83 0.84 -0.024 0.057 0.032

    Sexpf

    (mV/K) 0.2± 0.1 0.5± 0.1 0.3± 0.1 −0.85± 0.15 −0.9± 0.1 −0.12± 0.07 −0.5± 0.1 0.3± 0.1

    Sexptotal

    (mV/K) −3.6± 0.3 −1.2± 0.2 −1.1± 0.2 −1.4± 0.15 −2.0± 0.1 −0.30± 0.11 −1.2± 0.2 0.17± 0.05

    Table I: Parameters used for the thermoelectric analysis of aqueous salt solutions.

    Concentrations (cexp0 ) are given in mM. The values for Q∗+, Q

    ∗−, D+, and D− are those from

    Agar et al. [3]. Stheof = (4kB/e)(α+D+ − α−D−)/(D+ +D−) and Stheototal is calculated from

    the complete equation 2 in the stationary state. The values for the experimental fast

    Seebeck coefficient are obtained from Sexpf = −∆Vf/∆T , and that for the total Seebeck

    coefficient as Sexptotal = −(∆Vf +∆Vs)/∆T . ∆Vs and ∆Vf are the average values from at

    least three measurement cycles, as shown in Supplementary Information.

    for the salt with highest(lowest) values of the sum α+ + α−. Regarding the Stotal signs for

    experiments and calculated values, they are the same for hydroxides and opposite for acids.

    This may come from some unknown behavior of the H3O+ and OH− ions in the thermo-

    electric cell, or from some divergence of the heat of transport values taken from ref. [3]. In

    addition, equation 2 might still not be a complete physical description of the effect.

    A special case of the investigated electrolytes is aqueous KCl, which has the condition

    D+ = D−. The slow time response of thermovoltage given by the second term of equation 2

    is expected to be null, while our experiments show a slow thermoelectric field still develops

    even for that salt. This result points to the direction that equation 2 is still not a complete

    model for the reality of the thermoelectric experiment. Some of the effects neglected in

    current descriptions are electrodes polarization from differential ionic adsorption [27, 28]

    and specific interactions between electrodes and electrolytes [29]. From the theoretical point

    of view, the temperature dependence of single-ion Soret and mass diffusion coefficients are

    necessary to be considered in Janssen and Bier’s theory [23] in order to obtain a more

    complete theory.

    Our results show that thermovoltage changes in the different timescales for ∆Vf and

    9

  • ∆Vs. This dynamic response is only observable due to a large separation between the

    electrodes (l = 11 ± 1 mm) implying in characteristic thermodiffusion times τc ∼ 104 s

    while temperature stabilizes until 1000 s. In some studies, the value for l is in the range 1

    to 2 mm to assure homogeneity of the thermoelectric field in the cell [7, 30], which means

    τc . 500 s. The temperature difference also stabilizes in the same timescale, and so does ∆Vf

    (instantaneously with ∆T , τφ ∼ 10−8 s). Then, it may usually happen for ∆Vf and ∆Vs

    to change the measured value of ∆V (t) in the same timescale, generating an experimental

    result which is a monotonous curve and suggesting a single physical effect. In this study,

    we left aside the perfect homogeneity of the generated thermoelectric field, to study both

    thermoelectric contributions separated in time.

    The physical mechanisms responsible for thermodiffusion of ions and organic molecules in

    water are the solution thermal expansion [31, 32], temperature dependences of hydration en-

    ergy [32, 33], and hydrogen bonding [34, 35]. These mechanisms give rise to thermophoretic

    forces that drive the ionic Soret effect, with charge separation due to different single-ion Soret

    coefficients of anions and cations, and an electroneutral concentration gradient through the

    bulk solution. Both effects generate an electric field accounted by equation 2.

    A. Consistency with previous experimental results

    Our results presented the steady state (t → ∞) thermoelectric fields in the same direction

    for both acidic and basic solutions. It means that a charged nanoparticle would be driven

    in opposite directions when dispersed in acids or hydroxides. In the last few years, the

    thermodiffusion response of electrically-stabilized iron oxide nanoparticles in both acidic

    and basic solutions has been investigated. In the case of nanoparticles dispersed in acidic

    solutions, a total Seebeck coefficient S ≈ - 1.5 mV/K (similar to the values obtained in

    the present study) was measured and used to predict the negative Soret coefficient for the

    positively charged nanoparticles, as observed experimentally [30]. In another publication,

    the experimentally obtained salt Soret coefficient of hydroxides [32], which is peoportional to

    α++α−, was used to predict the positive Soret coefficients of negatively charged nanoparticles

    in TMAOH and TBAOH, as also observed experimentally [36]. Thus, our present results are

    consistent with those obtained in previous studies, suggesting that the opposite Stotal sign

    predicted by equation 2 for acids and hydroxides, which does not fit the overall consistency

    10

  • from the many experimental results, may come from a still existing problem in modelling

    the Seebeck response from thermodiffusion effects.

    In the case when a negatively charged nanoparticle is dispersed only by an ionic solution

    of NaOH, three different systems in literature [4, 36, 37] show that particles move to the hot

    side while our experimental results predicts the steady-state thermoelectric field should push

    them to the cold, like in the case of particles in TMAOH and TBAOH [36]. This may be

    related to salt specific effects, which is still an open question to be solved in thermodiffusion.

    IV. CONCLUSIONS

    This study presents experimental results on the thermoelectric voltage induced by temper-

    ature difference between electrodes in a thermoelectric cell filled with aqueous electrolytes.

    The results show a fast change in thermovoltage related to temperature stabilization, fol-

    lowed by a slow change due to ion thermodiffusion in the bulk electrolyte solutions. We

    showed that the fast response is related to the difference α+D+ − α−D− and the slow re-

    sponse is related to the differences between ions diffusion coefficients and proportional to

    the Soret coefficient of the salt α+ + α−. Equation 2, from Janssen and Bier’ theory [23],

    predicts Stotal in the mV/K order of magnitude, indicating that thermodiffusive effects are

    the source of high Seebeck coefficients in aqueous electrolytes. A very good agreement has

    been found between the values predicted from their theory and our experimental results for

    some electrolytes.

    A clear conclusion of this study is that the improvement of the thermoelectric voltage

    must be a combination of two factors: a high difference in α+D+ − α−D− to improve the

    fast response and high α+ + α− and D+ − D− values to improve the slow response. It is

    also desirable that both thermoelectric fields have the same direction, as in acidic solutions.

    Many material candidates, such as organic polyelectrolytes, which promise to increase the

    performance of thermoelectric cells, can be found in the literature [7, 8]. These electrolytes

    should be composed of macroions with high positive values for Q∗i , whereas their counterions

    should have the lowest value as possible. The desired increase in the α++α− values may be

    accomplished by increasing the temperature dependence of hydration free energy [32, 38] of

    the ions and/or macroions.

    11

  • V. SUPPLEMENTARY MATERIAL

    Figures S1 and S2 show the 12962-Solartron sample holder with the thermoelectric cell

    in the open and closed conditions, respectively. In Figure S1, it is possible to see the bottom

    electrode surface, which is in contact with the liquid when the cell is filled, whereas the upper

    electrode is kept away from it. Figure S2 shows the closed cell when ready for measurements,

    when a Teflon separator and two O-rings are between the electrode surfaces.

    Figures S3 and S4 show three measurements cycles for NaOH and KCl, respectively, with

    measurement reproducibility in time for ∆V (t).

    VI. ACKNOWLEDGEMENTS

    The authors acknowledge financial support from research funding agencies CAPES (Co-

    ordenação de Aperfeiçoamento de Pessoal de Nível Superior - 88881.133118/2016-01), Cnpq

    (Conselho Nacional de Desenvolvimento Científico e Tecnológico - 465259/2014-6), FAPESP

    (Fundação de Amparo à Pesquisa do Estado de São Paulo - 2014/50983-3; 2016/24531-3),

    INCTFCx (Instituto Nacional de Ciência e Tecnologia de Fluidos Complexos).

    [1] A. Würger, Rep. Prog. Phys. 73, 126601 (2010).

    [2] M. Reichl, M. Herzog, A. Götz, and D. Braun, Phys. Rev. Lett. 112 (2014).

    [3] J. N. Agar, C. Y. Mou, and J. L. Lin, J. Phys. Chem 93, 2079 (1989).

    [4] D. Vigolo, S. Buzzaccaro, and R. Piazza, Langmuir 26, 7792 (2010).

    [5] A. Würger, Phys. Rev. Lett 101 (2008).

    [6] J. N. Agar, Thermogalvanic Cells, Advances in Electrochemistry and Electrochemical Engi-

    neering, pp. 31-121 (Interscience New York, 1963).

    [7] D. Zhao, H. Wang, Z. U. Khan, J. C. Chen, R. Gabrielsson, M. P. Jonsson, M. Berggren, and

    X. Crispin, Energy Environ. Sci. 9, 1450 (2016).

    [8] L. Zhang, T. Kim, N. Li, T. J. Kang, J. Chen, J. M. Pringle, M. Zhang, A. H. Kazim, S. Fang,

    C. Haines, et al., Advanced Materials 29, 1605652 (2017).

    [9] P. F. Salazar, S. Kumar, and B. A. Cola, Journal of Applied Electrochemistry 44, 325 (2014),

    ISSN 1572-8838, URL http://dx.doi.org/10.1007/s10800-013-0638-y.

    12

    http://dx.doi.org/10.1007/s10800-013-0638-y

  • [10] M. Bonetti, S. Nakamae, M. Roger, and P. Guenoun, J Chem Phys 134, 114513 (2011).

    [11] R. Hu, B. A. Cola, N. Haram, J. N. Barisci, S. Lee, S. Stoughton, G. Wallace,

    C. Too, M. Thomas, A. Gestos, et al., Nano Letters 10, 838 (2010), pMID: 20170193,

    https://doi.org/10.1021/nl903267n, URL https://doi.org/10.1021/nl903267n.

    [12] H. Im, T. Kim, H. Song, J. Choi, J. S. Park, R. Ovalle-Robles, H. D. Yang, K. D. Kihm, R. H.

    Baughman, H. H. Lee, et al., Nature Communications 7, 10600 (2016).

    [13] J. Agar and V. Lobo, Electrochimica Acta 20, 319 (1975), ISSN 0013-4686, URL

    http://www.sciencedirect.com/science/article/pii/0013468675900110.

    [14] V. Lobo and M. S. Teixeira, Electrochimica Acta 27, 1145 (1982), ISSN 0013-4686, URL

    http://www.sciencedirect.com/science/article/pii/0013468682801242.

    [15] R. F. Stout and A. S. Khair, Phys. Rev. E 96, 022604 (2017).

    [16] D. Tordera, D. Zhao, A. V. Volkov, X. Crispin, and M. P. Jonsson, Nano Let-

    ters 17, 3145 (2017), pMID: 28441500, https://doi.org/10.1021/acs.nanolett.7b00574, URL

    https://doi.org/10.1021/acs.nanolett.7b00574.

    [17] D. Zhao, S. Fabiano, M. Berggren, and X. Crispin, Nature Communications 8, 14214 (2017),

    URL http://dx.doi.org/10.1038/ncomms14214.

    [18] T. J. Salez, B. T. Huang, M. Rietjens, M. Bonetti, C. Wiertel-Gasquet, M. Roger, C. L.

    Filomeno, E. Dubois, R. Perzynski, and S. Nakamae, Phys. Chem. Chem. Phys. 19, 9409

    (2017).

    [19] M. Bonetti, S. Nakamae, B. T. Huang, T. J. Salez, C. Wiertel-Gasquet, and M. Roger, The

    Journal of Chemical Physics 142, 244708 (2015).

    [20] J. N. Agar and J. C. R. Turner, Proc. R. Soc. Lond. A 255, 307 (1960).

    [21] J. A. Bierlein, The Journal of Chemical Physics 23, 10 (1955).

    [22] I. Chikina, V. B. Shikin, and A. A. Varlamov, Phys. Rev. E 92, 012310 (2015), URL

    https://link.aps.org/doi/10.1103/PhysRevE.92.012310.

    [23] M. Janssen and M. Bier, Phys. Rev. E 99, 042136 (2019), URL

    https://link.aps.org/doi/10.1103/PhysRevE.99.042136.

    [24] B. Abécassis, C. Cottin-Bizonne, C. Ybert, A. Ajdari, and L. Bocquet, Nature Materials 7,

    785 (2008), URL http://dx.doi.org/10.1038/nmat2254.

    [25] D. Velegol, A. Garg, R. Guha, A. Kar, and M. Kumar, Soft Matter 12, 4686 (2016), ISSN

    1744-683X, URL http://dx.doi.org/10.1039/C6SM00052E.

    13

    https://doi.org/10.1021/nl903267nhttp://www.sciencedirect.com/science/article/pii/0013468675900110http://www.sciencedirect.com/science/article/pii/0013468682801242https://doi.org/10.1021/acs.nanolett.7b00574http://dx.doi.org/10.1038/ncomms14214https://link.aps.org/doi/10.1103/PhysRevE.92.012310https://link.aps.org/doi/10.1103/PhysRevE.99.042136http://dx.doi.org/10.1038/nmat2254http://dx.doi.org/10.1039/C6SM00052E

  • [26] R. Nery-Azevedo, A. Banerjee, and T. M. Squires, Langmuir 33, 9694

    (2017), pMID: 28853577, https://doi.org/10.1021/acs.langmuir.7b01094, URL

    https://doi.org/10.1021/acs.langmuir.7b01094.

    [27] S. Bousiadi and I. Lelidis, Physics Letters A 382, 127 (2018), ISSN 0375-9601, URL

    http://www.sciencedirect.com/science/article/pii/S0375960117310800.

    [28] G. Barbero, L. R. Evangelista, and D. Olivero, Journal of Applied Physics 87, 2646 (2000),

    https://doi.org/10.1063/1.372231, URL https://doi.org/10.1063/1.372231.

    [29] A. Ganassin, P. Sebastián, V. Climent, W. Schuhmann, A. S. Bandarenka,

    and J. Feliu, Scientific Reports 7, 1246 (2017), ISSN 2045-2322, URL

    https://doi.org/10.1038/s41598-017-01295-1.

    [30] A. L. Sehnem, A. M. Figueiredo Neto, R. Aquino, A. F. C. Campos, F. A. Tourinho, and

    J. Depeyrot, Phys. Rev. E 92, 042311 (2015).

    [31] A. Würger, Langmuir 25, 6696 (2009), URL http://pubs.acs.org/doi/abs/10.1021/la9001913.

    [32] A. L. Sehnem, D. Niether, S. Wiegand, and A. M. F. Neto, The Journal of Physical Chemistry

    B 122, 4093 (2018).

    [33] E. D. Eastman, Journal of the American Chemical Society 50, 283 (1928).

    [34] D. Niether, T. Kawaguchi, J. Hovancová, K. Eguchi, J. K. G. Dhont, R. Kita, and S. Wiegand,

    Langmuir 33, 8483 (2017).

    [35] D. Niether, S. D. Lecce, F. Bresme, and S. Wiegand, Physical Chemistry Chemical Physics

    20, 1012 (2018).

    [36] A. L. Sehnem, A. M. F. Neto, D. Niether, and S. Wiegand, Phys. Rev. E 98, 062615 (2018).

    [37] K. A. Eslahian, A. Majee, M. Maskos, and A. Würger, Soft Matter 10, 1931 (2014).

    [38] S. Duhr and D. Braun, Proc. Nat. Acad. Sci. 103, 19678 (2006).

    14

    https://doi.org/10.1021/acs.langmuir.7b01094http://www.sciencedirect.com/science/article/pii/S0375960117310800https://doi.org/10.1063/1.372231https://doi.org/10.1038/s41598-017-01295-1http://pubs.acs.org/doi/abs/10.1021/la9001913

    Dynamic response of a thermoelectric cell induced by ion thermodiffusionAbstractI IntroductionII Experimental sectionIII Results and DiscussionA Consistency with previous experimental results

    IV ConclusionsV Supplementary MaterialVI Acknowledgements References


Recommended