Date post: | 28-Dec-2015 |
Category: |
Documents |
Upload: | sophie-george |
View: | 221 times |
Download: | 3 times |
Anesthetic Equipment
Gas SuppliesBulk Supply of Anesthetic Gases
In the majority of modern hospitals, piped medical gases and vacuum (PMGV) systems have been installed
only a few cylinders are kept in reserve, attached usually to the anesthetic machine
The PMGV services comprise five sections:– bulk store– distribution pipelines in the hospital– terminal outlets, situated usually on the walls or ceilings of the
operating theatre suite and other sites– flexible hoses connecting the terminal outlets to the anesthetic
machine– connections between flexible hoses and anesthetic machinesResponsibility for the first three items lies with the engineering and pharmacy departments. Within the operating theatre, it is partly the anesthetist’s responsibility to check the correct functioning of the last two items
Bulk Store: Oxygen
In small hospitals:Oxygen cylinder manifolds consist of two groups of large cylinders (size J)
Bulk Store: Oxygen
In large hospitals: Liquid oxygen store Liquid oxygen is stored
at a temperature of approximately −165 °C at 10.5
If the pressure increases above 17 bar (1700 kPa), a safety valve opens and oxygen runs to waste
Liquid oxygen plants are housed some distance away from hospital buildings because of the risk of fire
Even when a hospital possesses a liquid oxygen plant, it is still necessary to hold reserve banks of oxygen cylinders in case of supply failure
Other Gases Nitrous Oxide
– Nitrous oxide and Entonox may be supplied from banks of cylinders connected to manifolds similar to those used for oxygen
Medical Compressed Air – Compressed air is supplied from a bank of cylinders into the
PMGV system– Air of medical quality is required, as industrial compressed
air may contain fine particles of oil Piped Medical Vacuum
– Piped medical vacuum is provided by large vacuum pumps
Terminal Outlets Six types of terminal outlet are found commonly in
the operating theatre The terminals are color-coded and also have non-
interchangeable connections specific to each gas: – Vacuum (colored yellow) – a vacuum of at least 53 kPa (400
mmHg) should be maintained at the outlet, which should be able to take a free flow of air of at least 40 L min–1
– Compressed air (colored white/black) at 4 bar – this is used for anesthetic breathing systems and ventilators
– Air (colored white/black) at 7 bar – this is to be used only for powering compressed air tools and is confined usually to the orthopedic operating theatre
– Nitrous oxide (colored blue) at 4 bar– Oxygen (colored white) at 4 bar– Scavenging designed to accept a standard 30-mm connection
Outlet of central oxygen supply system
Gas Cylinders Modern cylinders are constructed from
molybdenum steel They are checked at intervals by the manufacturer
to ensure that they can withstand hydraulic pressures considerably in excess of those to which they are subjected in normal use
The cylinders are provided in a variety of sizes (A to J), and color-coded according to the gas supplied
Cylinders attached to the anesthetic machine are usually size E
The cylinders comprise a body and a shoulder
Gas Cylinders Cylinder valves should be opened slowly to prevent sudden
surges of pressure The color codes used for medical gas cylinders in the United
Kingdom are shown in Table 15.2. Different colors are used for some gases in other countries
Cylinder sizes and capacities are shown in Table 15.3. Oxygen, air and helium are stored as gases in cylinders and
the cylinder contents can be estimated from the cylinder pressure
Nitrous oxide and carbon dioxide cylinders contain liquid and vapor
The cylinder pressure cannot be used to estimate its contents because the pressure remains relatively constant until after all the liquid has evaporated and the cylinder is almost empty
The contents of nitrous oxide and carbon dioxide cylinders can be estimated from the weight of the cylinder
The anesthetic machine comprises: A means of supplying gases either from attached
cylinders or from piped medical supplies via appropriate unions on the machine
Methods of measuring flow rate of gases Apparatus for vaporizing volatile anesthetic agents Breathing systems and a ventilator for delivery of
gases and vapors from the machine to the patient Apparatus for scavenging anesthetic gases in
order to minimize environmental pollution.
vaporizerbellow
Corrugated tube
Soda lime
Flow meter
ventilator
APL valve
Scavenging system
The Anesthesia Machine
High Intermediate Low Pressure Circuit
Oxygen Supply Failure Alarm
The machine standard specifies that whenever the oxygen supply pressure falls below a manufacturer-specified threshold (usually 30 psig) a medium priority alarm shall blow within 5 seconds
Oxygen Flush Valve (O2+)
Receives O2 from pipeline inlet or cylinder reducing
device and directs high, unmetered flow directly to
the common gas outlet (downstream of the
vaporizer)
Machine standard requires that the flow be
between 35 and 75 L/min
The ability to provide jet ventilation
Hazards
– May cause barotrauma
– Dilution of inhaled anesthetic
Low Pressure System
Consists of:– Flow meters– Vaporizer mounting device– Check valve– Common gas outlet
Flowmeter assembly When the flow control
valve is opened the gas enters at the bottom and flows up the tube elevating the indicator
The indicator floats freely at a point where the downward force on it (gravity) equals the upward force caused by gas molecules hitting the bottom of the float
Vaporizers
A vaporizer is an instrument designed to change a liquid anesthetic agent into its vapor and add a controlled amount of this vapor to the fresh gas flow
Factors That Influence Vaporizer Output Flow Rate: The output of the vaporizer
is generally less than the dial setting
at very low (< 200 ml/min) or very
high (> 15 L/min) flows
Temperature: Automatic temperature
compensating mechanisms in bypass
chambers maintain a constant
vaporizer output with varying
temperatures
Back Pressure: Intermittent back
pressure (eg positive pressure
ventilation causes a higher vaporizer
output than the dial setting)
Factors That Influence Vaporizer Output
Atmospheric Pressure: Changes in
atmospheric pressure affect
variable bypass vaporizer output
as measured by volume %
concentration, but not (or very
little) as measured by partial
pressure (lowering atmospheric
pressure increases volume %
concentration and vice versa)
Carrier Gas: Vaporizers are
calibrated for 100% oxygen
Carrier gases other than this
result in decreased vaporizer
output
The Circuit: Circle System
Arrangement is variable, but to prevent re-breathing of CO2, the following rules must be followed:– Unidirectional valves
between the patient and the reservoir bag
– Fresh-gas-flow cannot enter the circuit between the expiratory valve and the patient
– Adjustable pressure-limiting valve (APL) cannot be located between the patient and the inspiratory valve
The carbon dioxide absorberSodalime (CaOH2 + NaOH + KOH + silica) or
Baralyme (Ba[OH] 2 + Ca[OH]2) contained in the
absorber combines with carbon dioxide, forming
CaCO2 and liberating heat and moisture (H2O)
A pH-sensitive dye changes to a blue-violet
color, indicating exhaustion of the absorbing
capacity
The canister should be changed when 25% to
50% of the contents has changed color, although
it should continue to absorb satisfactorily until at
least the contents of the top canister have
changed color
Circle System
Advantages:– Relative stability of inspired concentration– Conservation of respiratory moisture and heat– Prevention of operating room pollution– PaCO2 depends only on ventilation, not fresh
gas flow– Low fresh gas flows can be used
Disadvantages:– Complex design = potential for malfunction– High resistance (multiple one-way valves) =
higher work of breathing
The reservoir bag The reservoir bag is located on the expiratory
limb The reservoir bag accumulates gas
between inspirations It is used to visualize spontaneous
ventilation and to assist ventilation manually
Adults require a 3-L bag Children a 2-L bag Most new machines have a valve used to
switch between the reservoir bag and the ventilator
Older machines may require that the bag be removed and a hose to the ventilator be connected
The Adjustable Pressure Limiting (APL) Valve
User adjustable valve that releases gases to the scavenging system and is intended to provide control of the pressure in the breathing system
Bag-mask Ventilation: Valve is usually left partially open During inspiration the bag is squeezed pushing gas into the inspiratory limb until the pressure relief is reached, opening the APL valve
Mechanical Ventilation: The APL valve is excluded from the circuit when the selector switch is changed from manual to automatic ventilation
Scavenging Systems
A scavenging system channels waste gases away
from the operating room to a location outside the
hospital building
The ambient concentration of anesthetic gases in
the operating room should not exceed 25 ppm for
nitrous oxide and 2 ppm for halogenated agents
Specific anesthetic gas-scavenging systems should
be used routinely These systems consist of a
collecting system, a transfer system, a receiving
system, and a disposal system
Scavenging Systems
The disposal system may be passive or active, although passive systems are inadequate for modern hospitals
A passive system consists of wide-bore tubing that carries gases directly to the exterior or into the exhaust ventilation ducts
Active systems can be powered by vacuum systems, fans, pumps, or Venturi systems
Scavenging Systems
The disposal system may be passive or active,
although passive systems are inadequate for
modern hospitals
A passive system consists of wide-bore tubing that
carries gases directly to the exterior or into the
exhaust ventilation ducts
Active systems can be powered by vacuum systems,
fans, pumps, or Venturi systems
Gas Analysis
Several methods are used to monitor concentrations of oxygen, carbon dioxide, and anesthetic gases in the breathing system
The oxygen analyzer is the single most important monitor for detection of a hypoxic gas mixture
Capnometry, the measurement of carbon dioxide, has many uses, including monitoring the adequacy of ventilation and detection of breathing system faults
Breath-to-breath monitoring of anesthetic concentrations provides tracking of anesthetic uptake and distribution
Most gas analyzers incorporate alarms. Among the techniques for measurement are the following: Mass spectrometry, infrared analysis and oxygen concentrations analysis
Anesthesia Ventilators Most modern anesthesia machines are fitted
with a mechanical ventilator that uses a collapsible bellows within a closed chamber
The bellows is compressed intermittently when oxygen or air is directed into the chamber, thereby pressurizing it
The ventilators are time cycled flow (as opposed to pressure) generators, controlled both mechanically and electronically, and pneumatically driven (requiring 10 to 20 L of driving gas per minute)
Ventilator controls vary among makes and models.
Some ventilators require setting of minute ventilation, rate, and inspiratory-expiratory (I:E) ratio to produce the desired tidal volume; other ventilators allow direct adjustment of tidal volume, with I : E ratio being dependent on the inspiratory flow rate, which is set independently
Anesthesia Ventilators Ventilator controls vary among makes and models Some ventilators require setting of minute ventilation, rate,
and inspiratory-expiratory (I:E) ratio to produce the desired tidal volume; other ventilators allow direct adjustment of tidal volume, with I : E ratio being dependent on the inspiratory flow rate, which is set independently
Although gas-driven ventilators can be safely driven with either oxygen or air, most often oxygen is chosen and is supplied by pipeline. Whether or not cylinder gases are used to drive the ventilator in the event of pipeline failure is usually determined by the user. If the machine is set up to drive the ventilator using cylinder oxygen, mechanical ventilation should be discontinued in the event of pipeline failure to conserve oxygen supplies
Flow Generator Ventilators Flow generators deliver a set tidal volume
regardless of changes in patients' compliance
( unlike pressure generators ) but will not
compensate for system leaks and may produce
barotrauma because high pressures can be
generated
They reliably deliver the preset tidal volume (even in
the presence of a small leak).
The risk o f barotrauma is minimal because most
patients presenting to the operating room have
healthy normally compliant lungs.
Pressure Generator Ventilators
For infants and patients with diseased lungs, the maintenance of preset tidal volumes may produce unacceptably high airway pressures and increased risk of barotrauma
Pressure generators are more appropriate in these situations, because airway pressure is controlled and barotrauma risk minimized
Checking Anesthesia Machines
Check: Emergency
ventilation equipment
High-Pressure system Low-Pressure system Scavenging system Breathing system Manual and
automatic ventilation system
Monitors
Basic Anesthetic Monitoring
ASA Standards for Basic Anesthesia Monitoring
Prepared byDr. Mahmoud Abdel-Khalek
Basic Anesthetic Monitoring
The primary goal of anesthesia is to keep the patient as safe as possible in the perioperative period
Anesthesia and surgery are serious invasions on the physiologic stability of the human body
Careful monitoring of the patient during and after surgery allows the anesthesiologist to identify problems early, when they can still be corrected
Proper monitoring of the patient can reduce the risks involved in anesthesia and surgery
Some of the physiologic disturbances that occur in the perioperative period include: apnea, respiratory depression, airway obstruction, cardiac depression, hypertension, hypotension, hypervolemia, hypovolemia, arrhythmias, blood loss, fluid shifts, weakness, bradycardia, tachycardia, hyperthermia, and hypothermia
Standards of Care
Proper monitoring standards are well-defined
The most widely accepted current anesthesia
monitoring standards are those that have been
published by the by the American Society of
Anesthesiologists (ASA)
The ASA standards were initially published in
1986, and were updated afterwards
The ASA Standards for Basic Anesthetic Monitoring Standard I states that a qualified anesthesia
provider will be present with the patient throughout the anesthetic
Standard II states that the patient's oxygenation, ventilation, circulation, and temperature will be continually monitored
Assessment of oxygenation involves two parts: measurement of inspired gas with an oxygen analyzer and assessment of hemoglobin saturation with a pulse oximeter and observation of skin color
Assessment of ventilation is by clinical assessment and preferably capnography
The ASA Standards for Basic Anesthetic Monitoring
Tracheal intubation must be verified clinically and by detection of exhaled CO2
Mechanical ventilation must be monitored with an audible disconnect monitor
Assessment of circulation involves continuous ECG monitoring, blood pressure measurement at least every five minutes, and continuous monitoring of peripheral circulation by such means as palpation, ausculation, plethysmography, or arterial pressure monitoring
The patient's temperature must be measured if changes are anticipated, intended, or suspected
Oxygen analyzers Oxygen Analyzers Oxygen analyzers are an
integral part of the newer anesthesia machines The purpose of the oxygen analyzer is to
confirm that oxygen is being delivered to the patient and that concentration of oxygen in the gas mixture is adequate
The oxygen analyzer provides one last check before the gas mixture is delivered to the patient
For the analyzer to be useful, it must be calibrated and the low-limit alarm must be working
The two main types of oxygen analyzers are galvanic (fuel cell), and the polarographic
Pulse Oximeters
Pulse oximetry have been a major advance in improving the safety of anesthesia
The pulse oximeter provides continuous monitoring of hemoglobin saturation using a two-wavelength light absorption technique
The monitor filters out the effects of ambient light, tissue, skin pigment, tissue, and venous blood It focuses on the pulsatile absorption which due to pulsatile arterial blood
Pulse oximeters were developed in the early 1980's and rapidly proved their value in anesthesia
Pulse oximetry allows rapid, beat-by-beat, noninvasive monitoring of blood oxygenation
Disadvantages of pulse oximetry are that it is motion sensitive, and that substances like carbon monoxide, methemoglobin, and some dyes like nail polish affect the readings
Capnography
The most common method of exhaled CO2 measurement is sidestream infrared (IR) capnography
Gas from the circuit is drawn into an infrared measurement chamber CO2, N2O, H20, and inhaled anesthetic agents all absorb infrared light, but at slightly different frequencies
Newer monitors have precise light sources and filters that specifically measure the individual gases These monitors provide breath-by-breath gas analysis
Problems with IR capnographs are that moisture can cause blockage of the gas path, and that they can't measure oxygen or nitrogen
Abnormalities:– Complete absence of waveform
Circuit disconnection Cardiac arrest Esophageal intubation Complete respiratory obstruction
Automatic blood pressure monitors Current automatic
noninvasive blood pressure monitors work on the oscillometric technique
The cuff inflates well above the systolic pressure and then deflates slowly The monitor first senses oscillations as the cuff drops to systolic pressure
The point at which the oscillations are the strongest is read as the mean pressure
Most of these devices calculate the diastolic pressure after they measure the systolic and mean pressures The system is normally very reliable and accurate, but motion (especially
shivering) on the part of the patient or the surgeon leaning against the cuff will cause false readings or failure to get a reading
Patient injury is possible if the tubing becomes kinked Values may be in error if the cuff is not the proper size
ECG Monitor
Lead placement is important in ischemia detection The most sensitive lead is lead V5, detecting about 75% of ischemic
episodes Lead II plus lead V5 raise the detection rate to 80%, whereas leads II, V4, and V5 together detect 98% of ischemic events
Current top-of-the line monitors do automated ST analysis which is more reliable than individual practitioner assessment as long as the measurement points are correct
The ECG monitor can provide a lot of information to the anesthesiologist
Arrhythmia detection and identification of tachycardia and bradycardia are important uses
The ECG monitor may also provide the first indication of myocardial ischemia However the absence of ST depression does not guarantee that ischemia is not present
Ventilation Monitors
Continuous measurement of exhaled tidal volume can detect circuit leaks and hypoventilation
The spirometers on the anesthesia machines may give false readings if moisture blocks the inner workings
Current anesthesia machines also have overpressure alarms and overpressure "pop-off" valves
Current anesthesia machines have ventilator disconnect alarms and built-in spirometers
The spirometers have high and low limit alarm settings
Temperature Monitors
Monitoring of skin temperature is nearly useless
Upper esophageal and nasopharyngeal temperature are affected by airway temperature
Lower esophageal temperature is normally a good reflection of core or blood temperature
Tympanic membrane temperature is also a good indication of core temperature but it is not practical in the operating room environment
Peripheral Nerve Stimulators
Peripheral nerve stimulation (PNS) monitoring is not required by the ASA standards However, it is an important safety monitor in patients who a receiving neuromuscular blocking drugs
Train-of-four monitoring assesses the level of nondepolarizer blockade and double-burst stimulation assesses return of strength at the end of the case
Clinical monitoring of neuro- muscular blockade during an anesthetic is difficult without a PNS monitor
Clinical assessment of strength is important, however, at the conclusion of an anesthetic before a final decision is made to extubate the patient
Thank You