ATOMIC SPECTROSCOPYATOMIC SPECTROSCOPY(Based on Flame Atomization)(Based on Flame Atomization)
LECTURE 5
Atomic SpectroscopyAtomic SpectroscopyThree techniques (methods) included in Three techniques (methods) included in atomic spectroscopy.atomic spectroscopy.
1. Atomic absorption spectroscopy2. Atomic emission spectroscopy 3. Atomic fluorescence spectroscopy
Focus on 1 and 2.Focus on 1 and 2.
2
Atomic SpectroscopyAtomic SpectroscopyIn order to perform atomic spectroscopy, atoms of the analyte must first be formed, usually in the form of an atomic vapor.
3
Atomization The process by which a sample is converted
to an atomic vapor. Atomizer A device used to convert a sample to an
atomic vapor.
Atomic SpectroscopyAtomic SpectroscopyThree types of atomizers:Three types of atomizers:
1.1. Flame atomizer2. Plasma atomizer 3. Electrothermal atomizer
Focus on 1.Focus on 1.
4
5
Atomization ProcessAtomization ProcessSolution of the analyte is evaporated
rapidly at an elevated temperature to yield a finely divided solid.
Further heating will break down into gaseous atoms.
6
FLAME FLAME ATOMIZATIONATOMIZATION
7
Processes occurring Processes occurring during atomizationduring atomization
8
NebulizationNebulization Conversion of the liquid Conversion of the liquid sample to a fine spray.sample to a fine spray.
Desolvation Desolvation Solid atoms are mixed Solid atoms are mixed with the gaseous fuel.with the gaseous fuel.
VolatilizationVolatilizationSolid atoms are converted Solid atoms are converted to a vapor in the flame to a vapor in the flame (molecules/atoms/ions.)(molecules/atoms/ions.)
9
What are the process involve to change the What are the process involve to change the analyte from one state to another?analyte from one state to another?
10
••There are three types of particles that There are three types of particles that exist in the flame:exist in the flame:
I.I. AtomsAtomsII.II. IonsIonsIII.III. MoleculesMolecules
11
Types of FlamesTypes of Flames
Selection of flame Selection of flame type type depends on depends on the volatilization the volatilization temperature of the temperature of the atom of interest.atom of interest.Most common is Most common is Acetylene/air.Acetylene/air.
Interzonal region Interzonal region is the is the hottest part of the flame hottest part of the flame and and best for atomic best for atomic absorption.absorption.
Oxidation of the atoms Oxidation of the atoms occurs in occurs in the secondary the secondary combustion zone combustion zone where where the atoms will form the atoms will form molecular oxides and are molecular oxides and are dispersed into the dispersed into the surroundings.surroundings.
12
Flame StructureFlame Structure
Temperature Profile
13
Temperature profile Temperature profile in °C for a in °C for a
natural gas-air flamenatural gas-air flame
14
Nebulizer
a) Concentric tubes.b) Cross flow.c) Fritted disk.d) Babington.
15
BurnersTwo types of burners in flame spectroscopy:
i. Turbulent flow (total consumption burner)
ii. Laminar flow (premix burner)
16
Nebulizer & burner are combined into a single unit.
Sample is drawn up the capillary & nebulized.
Sample flow rate: 1 to 3 mL/min.
17
Turbulent Flow Burner
Advantage1. introduce relatively large &
representative sample into the flame.
Disadvantages1. A relatively short path length through
flame.2. Problems with clogging of the tip.3. Burners noisy from electronic and
auditory stand point.18
Turbulent Flow Burner
Sample is nebulized Sample is nebulized by the flow of oxidant by the flow of oxidant whichwhich flow through a flow through a capillary tip.capillary tip.
ResultingResulting aerosol then aerosol then mixed with fuel & flow mixed with fuel & flow throughthrough a series of a series of baffles. baffles.
Only finest droplets Only finest droplets went through the went through the baffels.baffels.
Laminar Flow Burner
19
Bigger sample Bigger sample droplets is collected droplets is collected at the bottom of at the bottom of mixing chamber mixing chamber then drained to a then drained to a waste container.waste container.
Aerosol, oxidant & Aerosol, oxidant & fuel are burned in a fuel are burned in a slotted burner slotted burner that that provides a flame of provides a flame of 5 – 10 cm in length5 – 10 cm in length..
Laminar Flow Burner
20
21
AdvantagesAdvantagesi.i. Provide Provide quiet flamequiet flame..ii.ii. Provide longer path length Provide longer path length that enhance the that enhance the
sensitivity & reproducibility.sensitivity & reproducibility.DisadvantagesDisadvantagesi.i. Lower rate of sample introductionLower rate of sample introduction..ii.ii. Possibility of selective evaporation of mixed Possibility of selective evaporation of mixed
solvents in the mixing chamber could solvents in the mixing chamber could create create analytical uncertainties.analytical uncertainties.
iii.iii. Mixing chamber contains a potentially explosive Mixing chamber contains a potentially explosive mixturemixture that can flash back if the flow rates are that can flash back if the flow rates are too low.too low.
Laminar Flow Burner
The spectra result from the atomized sample absorbing photons of radiation of the appropriate energy (wavelength).
Energy of radiation absorbed by a vaporized atom is similar with energy needed for electron excitation transitions.
The excitation transition take place as the electrons jump from ground state to a higher energy level.
22
Atomic Absorption Spectra
23
Atomic AbsorptionAtomic AbsorptionINSTRUMENTATIONINSTRUMENTATION
24
AA SpectrophotometersSingle Beam Instrument Single Beam Instrument
The modulated power source can be replaced by a The modulated power source can be replaced by a chopper.chopper.
25
AA SpectrophotometersDouble Beam Instrument Double Beam Instrument
Radiation from HCL is split into 2 beams.1. One passes through the flame.2. The other around the flame.
A half-silvered mirror returns both beams to a single path then pass through the monochromator then detector.
Monochromator is placed between sample and detector.It used to eliminates most of the radiation
originating from the flame.
26
Double Beam Instrument Double Beam Instrument
Radiation Sources for AASRadiation source in AAS is a Radiation source in AAS is a line source line source which provide narrow emission bands. which provide narrow emission bands.
Common radiation source used in AASCommon radiation source used in AAS1.1.Hollow cathode lamp (HCL)Hollow cathode lamp (HCL)2.2.Electrodeless discharge lamp (EDL)Electrodeless discharge lamp (EDL)
27
Light from this lamp exactly light required for Light from this lamp exactly light required for the analysis, even no monochromator is used.the analysis, even no monochromator is used.
Hollow cathode lamp MUST contain the element to be determined.
28
Hollow cathode lamp (HCL)Hollow cathode lamp (HCL)
How does the HCL works?How does the HCL works?
When lamp is on, atoms are supplied with energy that causes electrons of the atoms elevate to the excited states.
Upon electrons returning to ground state, wavelength of the photon emitted are useful for the analysis.
The photon emitted will supply the exact amount of energy needed for the analyzed metal to undergoes excitation. 29
Excitation Mechanisms in HCLExcitation Mechanisms in HCL
The lamp is a sealed glass envelope filled with Argon or Neon gas.
When lamp is ON, Ar atoms are ionized with electrons at Anode (+ electrode).
The Ar ions, Ar+ bombard the surface of the Cathode (- electrode).
Metal atoms,M in the cathode are elevated to the excited state and are rejected from the surface as a result of this bombardment. 30
When atoms return to ground state, line spectrum of that specific atom emitted.
This light is directed at the flame where unexcited atoms of the same element absorb the radiation and raised to the excited state.
Absorbance is measured and related to the concentration.
31
Excitation Mechanisms in HCLExcitation Mechanisms in HCL
32
Reactions in the HCL
Ionization of filler gas: Ar + e- Ar+ + 2 e-
Excitation of metal atoms: M + Ar+ M* + ArLight emission: M* M + h
Ar + e- Ar+ + 2e-Excitation Mechanisms in HCLExcitation Mechanisms in HCL
Constructed of a metal or salt of interest Constructed of a metal or salt of interest sealed in a quartz tube filled with a noble sealed in a quartz tube filled with a noble gas (Ne or Ar) at low pressure (1 – 5 torr).gas (Ne or Ar) at low pressure (1 – 5 torr).
Electrodeless Discharge Lamp (EDL)
33
34
The noble gas is ionized and accelerated by The noble gas is ionized and accelerated by a strong radio-frequency (RF) or microwave a strong radio-frequency (RF) or microwave field field and and excite the metal or salt of interestexcite the metal or salt of interest. .
EDL can provide radiant intensities usually one to EDL can provide radiant intensities usually one to two orders of magnitude greater than HCL. two orders of magnitude greater than HCL.
EDLEDL
Source ModulationSource Modulation
35
Why source modulation is employed in AAS?1.To eliminate interference caused by
emission of the radiated flame from analyte atoms and flame gas species.
2.To distinguish between the component of radiation arising from the source and the component of radiation arising from the flame background.
Source modulator: Light chopper (circular rotating metal disk)
36
Source ModulationSource Modulation
37
The function of light chopper is to eliminate the effects of radiation from the flame.
Light is “chopped” with a rotating half-mirror so that detector could received two alternating signals. One from the radiation source and one from the flame.
At one moment (opaque), only light emitted by flame is read by the detector since the light from the radiation source is cut off. 38
Light ChopperLight Chopper
39
Next moment (transparent), light from both the flame emission and radiation source is read. Transmission from the source light is measured since the source light is allowed to pass.
Absorbance of the sample is determined by measuring the difference in radiant power between flame emission signal and signal from the radiation source.
Light ChopperLight Chopper
a) Signal provide by radiation source.
b) Sample absorbance.c) Signal provide by the
sample after absorbing part of the radiation power.
40
QuestionsQuestions1. In flame atomic absorption spectroscopy,
briefly describe the 'atomization' process which the analyte undergoes.
2. Why is source modulation employed in atomic absorption spectroscopy? Name the device used for this purpose.
3. Name the common line source used in AAS.
41
INTERFERENCESINTERFERENCES
42
Spectral Spectral interferenceinterference
Chemical Chemical interferenceinterference
Spectral interference occurs when Spectral interference occurs when the spectral line of the elements being determined are overlaps with a spectral line or band from another element present in the sample.The effect of the element will also be measured and therefore the results will be incorrect.
43
Spectral InterferenceSpectral Interference
1. Scattering by combustion or particulate products. Arise from the combustion products or particulate matters from the atomization scatters the radiation from the source. Both products reduce the power of the transmitted beam.
2. Scattering by sample matrix interference . Arise when the emission or absorption of an interfering species overlaps or lies so close to the analyte absorption so that resolution by the monochromator becomes impossible.
44
How it is occurs?How it is occurs?
1. Tune the monochromator to a different spectral line for the element of interest so that there is no overlap.
2. Use the secondary lines for the element of interest (can be found in the literature).
45
How to solve the issues?How to solve the issues?
Common types of chemical interferences that reduce the concentration of free gaseous atoms during analysis are:1.Ionization.2.Refractory Formation (formation of compound of low volatility).3.Dissociation reactions.
46
Chemical InterferenceChemical Interference
Sample either in liquids or solid is vaporized and atomized at high temperature provide by the flame source.
This high temperature environment lead to ionization of the analyte atoms.
Ionization of analyte atoms reduced the concentration of of analyte atoms to be excited.
Analysis of AAS are interested in analyte that in atomic state not the analyte ionic state.
47
1. Ionization1. Ionization
ExampleDetermination of Ba in alkaline earth
mixtures. Since the wavelength for Ba absorption appears in the center of broad absorption band of Ca.
Samples containing organics solvents where incomplete combustion of the organic matrix leaves carbonaceous particle scattered the beam.
In concentrated solutions where atomization products form refractory metal oxides with greater diameter scattered the beam.
48
By adding ionization suppressors such as Cesium, Cs and potassium, K to analyte solutions. These atoms are easily ionized & produce a high concentration of free electrons in the flame.
These atoms suppressed the analyte ionization by adding a source of electrons which shifts the equilibrium of the analyte from the ionic back to the atomic form.
49
Analyte ↔ Analyte+ + e-
How to cure ionization?How to cure ionization?
Anions that can form refractory compounds of volatility with analyte that are not atomized in flames.
Refractory compound formed reduced the fraction of analyte that is atomized.
For example, a decrease in Ca absorbance is observed with the increasing concentrations of phosphates,PO4
3- and sulphate,SO4
2- due to formation of calcium phosphate and calcium sulphate.
50
2. Refractory Formation2. Refractory Formation
Refractory compounds formation can be prevented by adding a releasing agent such as Lanthanum and Strontium ion.
Releasing agent are cations that react preferentially with the interferent and prevents its interactions with the analyte.
Lanthanum ions, La3+ is added to the sample and standard in Calcium measurement. The La3+ binds with phosphates presence as La(PO4)3 which is low volatility and not atomized.
51
How to cure Refractory How to cure Refractory Formation?Formation?
52
3. Dissociation Reactions3. Dissociation ReactionsIn the hot gaseous environment of a flame
or furnace, dissociation and associations reactions lead to conversion of the metallic constituents to the elemental state.
Dissociation reactions involve metal
oxides and hydroxide for an element and anions other than oxygen such as Cl-.
MO ↔ M+ OM(OH)2 ↔ M + 2OH
MCl ↔ M + Cl
ExampleThe emission spectra of Na is decreased
with the presence of HCl. Cl atom from the added HCl formed NaCl which later decrease the atomic concentration of Na as well as the Na emission and absorption.
Alkaline earth oxides are relatively stable but at very high temperatures, oxides and hydroxides of the alkaline metal easily dissociated. These bands are at higher intensity compared to the band of atom or ions required for analysis.
53
By the addition of Aluminium,Al and Titanium,Ti in the analysis of Vanadium,V.
Addition of Al and Ti change the concentration of Oxygen needed for the combustion of V causes an increased in Vanadium atoms concentration to be atomized and excited.
54
How to cure dissociation How to cure dissociation reactions?reactions?
Quantitative AnalysisQuantitative Analysis1. Calibration curve2. Standard Addition Method
Quantitative analysis still follows Beer’s Lambert Law and the concentration of unknown are determined by the same way using the formula( A = εbc).
Width of the flame is consider as the beam pathlength.
55
Calibration curveA general method for determining the concentration of a substance in an unknown sample by comparing the unknown to a set of standard of known concentration. Plot is a linear over a significant
concentration range (the dynamic range). Analysis should never be based on the
measurement of a single standard with assumption that Beer’s law is being followed.
56
Standard Calibration Curve
How to measure the concentration of unknown?Practically, you have measure the absorbance of your unknown. Once you know the absorbance value, you can just read the corresponding concentration from the graph .
Extensively used in AAS. This process is often called spiking the sample.
Compensate for variation caused by physical and chemical interferences in analyte solution.
Involves adding one or more increments of a standard solution to sample aliquots of the same size.
Each solution is then diluted to a fixed volume before measurement.
Standard Addition Method
If Beer’s law is obeyed,A = εbVstdCstd + εbVxCx
Vt Vt
= kVstdCstd + kVxCx
A= (kCstd)Vstd + kCxVx
k is a constant equal to εb/Vt
Standard Addition MethodsStandard Addition Methods
Plot a graph of A vs Vstd
A = (kCstd)Vstd + kCxVx
Y = mX + Cwhere the slope m and intercept C arem = kCstd
C = kCxVx
Standard Addition MethodsStandard Addition Methods
Cx can be obtained from the ratio of these two quantities m and C
C = kVxCx
m kCstd
Cx = CCstd
mVx
For single-point standard additionDividing the 2nd equation by the first & then rearrange it will give.
Cx = A1 Cs Vs (A2 – A1 ) Vx
Standard Addition MethodsStandard Addition Methods
ExampleExampleThe chromium in an aqueous sample was determined by pipetting 10.0mL of the unknown into each of five 50.0mL volumetric flasks. Various volumes of a standard containing 12.2 ppm Cr were added to the flasks, following which the solutions were diluted to the volume.
1.Plot the data.2.Calculate the concentration of Cr in the sample.
63
ADVANTAGES (Flame)1. Inexpensive in term of the equipment and
running day-to-day analysis.2. High sample throughput.3. Easy to use.4. High precision.
Flame atomization best for reproducibility (less than 1%)
64
65
1. Only solutions can be analyzed.2. Relatively large sample quantities required (1 –
2 mL).3. Less sensitivity compared to graphite furnace.
Flame AAS detection limit is >1ppm while Graphite furnace is <1ppm.
4. Interference due to refractory compound formation.
5. Only metal element can be analysed using AAS.
DISADVANTAGES
Detection Limits (ppb)
66
Applications of AASApplications of AASWater analysis. (Ca, Mg, Fe, Si, Al, Ba
content)Food analysis.Analysis of animal feedstuffs (Mn, Fe, Cu,
Cr, Se, Zn)Analysis of additives in lubricating oils and
greases. (Ba, Ca, Na, Li, Zn, Mg)Analysis of soils.Clinical analysis (Ca, Mg, Li, Na, K, Fe)
67