+ All Categories
Home > Education > Autoclave

Autoclave

Date post: 13-May-2015
Category:
Upload: asanka-morawaka
View: 1,009 times
Download: 6 times
Share this document with a friend
Description:
Autoclave is a critical instrument used in laboratory,
Popular Tags:
25
Sterilization Sterilization is a term referring to any process that removes or kills all forms of microbial organisms such as fungi, bacteria, viruses, spore forms, etc. present on a surface, contained in a fluid, or in a compound such as biological culture media. Sterilization can be achieved by applying heat, chemicals, irradiation, high pressure, and filtration or combinations of them.
Transcript
Page 1: Autoclave

SterilizationSterilization is a term referring to any process that 

removes or kills all forms of microbial organisms such as  fungi, bacteria, viruses, spore forms, etc. present on a surface, contained in a fluid, or in a compound such as biological culture media.

 Sterilization can be achieved   by applying heat, chemicals,   irradiation, high pressure,  and filtration or combinations   of them.

Page 2: Autoclave

Autoclave The autoclave is a equipment used to  remove microorganisms  (Virus, Bacteria, fungus etc.) and spores using  high pressure and high temperature steam sterilization.

Page 3: Autoclave

PURPOSE OF THE AUTOCLAVE1. To prepare materials for bacteriological cell 

cultures(test tubes, pipettes, Petri dishes, etc.) without contamination.

2. Prepare elements used for taking samples. (needles, tubes, containers).

3. Sterilize contaminated material.

Page 4: Autoclave

OPERATION PRINCIPLESwhen the pressure higher than atmosphere a higher 

temperature than 1000C needs to boil the water.This steam contain higher energy than usual. Autoclaves make pressurized saturated steam in a 

chamber passing electrical current through a heating element in controlled manner.

This energized steam take place required sterilization.This method is known as steam sterilization.

Page 5: Autoclave

MAIN COMPONENTS1. Heating Elements2. Temperature Controller3. Pressure Sensor4. Chamber 5. Door gasket 6. Solenoid valve7. Water level Sensor 

Page 6: Autoclave

BLOCK DIAGRAM

Page 7: Autoclave

STERILIZATION PROCESS 

Page 8: Autoclave

Heating Elements

Page 9: Autoclave

Temperature Controller

Thermostat Pt 100

Page 10: Autoclave

Pressure Sensor

Page 11: Autoclave

Chamber 

Page 12: Autoclave

Door gasket 

Page 13: Autoclave

Solenoid valve

Page 14: Autoclave

Water level Sensor 

Page 15: Autoclave

Symbols need to remind

Page 16: Autoclave

Program selectionProgram nº Mode Temperature /

pressureSterilization

timeDrying time

1 Solid 121ºC  /  1bar 20 -

2 Solid 126ºC  /  1.33  bar 30 -

3 Solid 134C  / 2  bar 18 -

4 Solid 134ºC  /  2  bar 10 -

5 Liquid 121ºC  /  1bar 30 -

6 Solid + Dry 121ºC  /  1bar 30 35

7 Solid + Dry 126ºC  /  1.33  bar 20 35

8 Solid + Dry 126ºC  /  1.33  bar 18 35

9 Solid + Dry 134ºC  /  2  bar 10 35

0 - - - -

Page 17: Autoclave

Start 1. Main  switch2. Open  door  indicator  lamp.3.  Lack  of  water  indicator  lamp.4.  End  cycle  indicator  lamp.5.  Liquid  mode  selector.6.  Solid  mode  selector.7.  Solid+Drying  mode  selector.8.  Temperature  indicator  display.9.  Temperature select pushbutton.10. Parameter  up  pushbutton.1 1. Parameter  down  pushbutton.12. Drying  time  indicator.13. Time  indicator  display.14. Time  pushbutton.15. Start pushbutton.16. Program  indicator  display.17. Program  pushbutton.18. Stop  pushbutton.20. Alarm  indicator  lamp.

Page 18: Autoclave

Top view21. Water drainage valve22. Steam  drainage  valve.23. Internal tank filling

Page 19: Autoclave

Rear view24. Air  filter.25. Safety  valve.26. Drying  control  thermostat.27. Safety  thermostat.28. Drainage valve.29. Steam  drainage  valve30. Water  inlet.31. Condensation  outlet32. Tank  drainage.

Page 20: Autoclave

Cleaning Procedure NEVER use bleach, steel wool, a steel brush or anything abrasive to scrub or clean the Chamber. 1. remove instruments and trays from the autoclave. 2. spread the contents of a packet in chamber evenly. 3. Select and start program No. 1 (without dry). 4. At the end of the cycle drain the water from the reservoir. 5. Fill the water reservoir with distilled water. 6. Repeat a sterilization cycle without Chamber Brite ™ powder, to 

remove any excessive dirt in the pipes. 7. Select and start program No. 1 (without dry). When the cycle is 

finished it will automatically exhaust. 8. At the end of the cycle drain the water from the reservoir. 9. Now autoclave ready to use.

Page 21: Autoclave
Page 22: Autoclave

Preventive MaintenancesQuarterly1. Clean all the filters.(Water, Air )2. Fill the water level of feed tank up to maximum limit3. Check that the manometers function as expected.4. Check the accuracy of temperature reading.5. Activate the safety valves manually to verify that they are operating well.6. Conduct a general sterilization process write down pressure, temperature, 

required times for completing each phase of the cycle, put it on registration system. 

7. Verify that it is functioning within tolerances defined by manufacturer 8. Check any abnormal sound in solenoids.9. Clean the chamber and remove limescale if any, advice to use only distilled 

water.10. Check the earth leakage current, protective earth. 

Page 23: Autoclave

limescaleLimescale is the hard, off-

white, chalky deposit found in kettles, hot-water boilers

Add little amount of any acid heat the chamber.

Deposited CaCO3 will be removed.

Then rinse off with plenty of water.

Page 24: Autoclave

Quality of Process Sterilization process indicatorsThey check instrument’s control parameters like temperature, time and 

pressure.Chemical indicatorsAdhesive tapes that go inside a      component or on packages.Biological indicatorsThe spore indicator is placed in the       sterilizing load.After the process, it is incubated,      analyzed and it is determined if the      cycle meets with the sterilization      requirements.

Page 25: Autoclave

Thank you..


Recommended