BIOSYNEX® TETANOS
B A C T E R I O L O G Y
Rapid test for the detection of anti-tetanus toxoid antibodies
in whole blood, serum and plasma
Efficient- 0.1 UI/mL detection limit according to WHO- Results in 10 minutes
Convenient- Storage 18 months at 2-30°C- Compatible with BIOSYNEX® Reader- Positive control available
High performances ** Clinical study performed on capillary blood, venous blood and serum from the same patients and compared to ELISA method
S
C
T
TETANUS
Positive
Kit content• 20 test cassettes• 20 disposable plastic pipettes• 1 dilution buffer vial• Instructions for use• Barcode card for test batch identification and traceability when using the BIOSYNEX® Reader
CT
CT
BIOSYNEX SA22 Boulevard Sébastien Brant
67400 ILLKIRCH-GRAFFENSTADEN - FranceTel. : +33 3 88 77 57 52 - Fax : +33 3 88 78 76 78
E-mail : [email protected] FL
Y_1030
016_
EN_V
0120
1806
R02
THE SOLUTIONThe tetanus toxin of Clostridium tetani bacteria binds to the cell membranes of the peripheral nerves and inhibits the delivery of neurotransmitters. Immunization is the best way to prevent Clostridium tetani infections in children and adults. It is sometimes desirable to know the level of anti-tetanus toxoid antibodies in a subject, in order to evaluate his/her immunitary status and determine the need for complementary vaccination. TETANOTOP® test is a rapid immunochromatographic test for the detection of anti-tetanus toxoid antibodies in human serum, plasma or whole blood, to evaluate anti-tetanus immunity.
• Automated reading of the results• Results storage• Printer, barcode reader and adhesive paper• L.I.S connectivity
Test procedure (for capillary blood) **
BIOSYNEX® Reader
References
Performances * Capillary blood Venous blood Serum
Negativeif = 0.1
Positive if = 0.1
Specificity 90% 100%
Sensitivity 83% 79%
Negative if = 0.1
Positive if = 0.1
Specificity 100% 100%
Sensitivity 93% 88%
Negative if = 0.1
Positive if = 0.1
Specificity 100% 100%
Sensitivity 91% 87%* Internal study performed on capillary blood, venous blood and serum from the same patients and compared to ELISA method (see instructions for use for more information) by AllDiag.
• Ref. 1030016 - TETANOTOP® - 20 tests
• Ref. 6030001 - Control Tetanus Positive 0.5 mL (10 controls)
• Ref. 5060001 - BIOSYNEX® Reader - 1 reader
Carefully read the instructions on the labelling and/or reagent notices.
1 2 3Results
BIOSYNEX
opTricon Gesellschaft für optische Technologien (GERMANY)
** Simplified procedure. Refer to the instructions for use for complete procedure.
Prick the fingertip using a sterile lancet.
45°
Collect the drop of blood using the pipette provided using pressing the
head of the pipette.
Transfer the blood sample into the ‘S’ well of the cassette.
Read the results at 10 minutes, either visually or using the
Biosynex® Reader.
CT
Negative
Negative result : absence of protection in an unvaccinated subject or a subject whose vaccination is not up to date.
Positive result : level of anti-tetanus antibodies in the sample > 0.1 UI/mL.
10