Blood Bank Equipment
Transcript
Page 1: Bloodbank Equipment

Blood Bank Equipment

Page 2: Bloodbank Equipment

BLOOD BANK REFRIGERATORS

Forma offers Blood Bank Refrigerators with storage capacities of 40, 84, 240,490 and 672 whole blood bags. Units feature reliable temperature performance at +4°C with digital temperature readout to verify cabinet temperature. Alarm systems are standard on all blood bank refrigerators. Inkless temperature

recorder is optional on undercounter models and standard on all other models.

PAGES 3-6

CRYOPRECIPITATE AND PLASMA THAWING BATHS

Model 2386 Cryoprecipitate Bath improves yields by 50%through controlled thawing at +4°C. Forma's Model 2032Bath is designed for plasma and red cell thawing. Bothunits feature a stainless steel pump which provides agitation

of the water for efficient thawing and a stainless steel tank, with drain, for easy clean-ing.

PAGES 18-19

PLASMA STORAGE FREEZERS

Forma's Plasma Storage Freezers are designedto provide safe storage of single donor, freshfrozen plasma and cryopre-cipitate at temperatures of -30°C and colder. Models are available with storagecapacities for 80, 300 and 350 single plasma units.

PAGES 7-11

PLATELET INCUBATORS

Maintain platelet viability safely at +22°Cwith Forma's platelet incubators. Modelsare available in single shelf and doubleshelf configurations.

PAGES 20 - 21

PLASMA/BLAST FREEZERS

Chest and Upright Plasma/Blast Freezers willfreeze up to 99 plasma units in 75 minutes.Freezers available with air or water-cooledcompressors.

PAGES 14- 17

ACCESSORIES PAGE

Blood Bank Refrigerators 21Blood Component Storage Freezers 22Blast Freezers 23Thawing Baths 23Safe Temp Gloves 24CO2 and LN2 Back-Up Systems 24General Accessories 25-27

RED CELL STORAGE FREEZERS

Forma Red Cell Storage Freezers maintain red cells safely below regulatory standards at -86°C. Upright and twochest styles are available.

PAGES 11-13

FORMA SCIENTIFIC OFFERS A COMPLETE LINE OF BLOOD BANK EQUIPMENT

For freezing, thawing, and storing ... you can depend on Forma Scientific to provide quality equipment that meets AABB, ANRC andFDA standards. Forma offers the most extensive line of blood bank equipment on the market today. These refrigerators, storage freezersand platelet incubators feature alarm systems that alert technicians to critical power failures or temperature deviations. Inkless chartrecorders track conditions for permanent records. For quality blood bank equipment, designed and engineered to meet the preciserequirements of blood banking and blood processing applications, put your trust in Forma Scientific.

TABLE OF CONTENTS

2

Page 3: Bloodbank Equipment

BLOOD BANK REFRIGERATORS

For consistent, reliable protection of blood products at a safe +4°C, youcan depend on Forma Scientific's blood storage refrigerators. Available in 240, 490 and 672 bag capacities, they main-tain temperature regardless of inventory load. Built-in over/undertem-perature alarm system and recorder, as required by AABB, ANRC andFDA for whole blood storage, are standard equipment. Forma BloodBank Refrigerators feature adjustable stainless steel drawers which rollout for easy access to inventory. High-visibility glass doors and fluo-rescent lights allow better inventory viewing. Well designed refrigera-tion systems provide quick temperature recovery for maximum productprotection after door openings.

PRODUCT FEATURES

• Select from three Blood Bank Refrigerators with inventory capacities up to 672 bags, while ensuring optimum temp-

erature uniformity throughout the interior compartment.

• Independent temperature controller is factory preset at +4ºC.

• Non-CFC, foamed-in-place, urethane insulation and triple-pane* glass door keep the cold in and reduce energy consumption.

• Six inch, seven day, inkless recorder charts product temperature for permanent records. Probe is located in a thermally-weighted product simulation bottle.

• 100% stainless steel interior for easy care and durability.

• Automatic defrost cycle keeps refrigerator free from frost build-up. Built-in condensate evaporator eliminates the need for a drain line.

• Blowers automatically cut off forced-draft air circulation when door is open.

• Key locks are standard.

• Adjustable leveling feet standard on Models 3881 (3889) and 3883 (3887). Optional customer and factory installed casters are available. Customer installed casters are standard on Forma's Model 3866/3868 (3867) Double Door

Refrigerator. Factory installed casters are optional.

• Drawer labels are provided standard to identify blood types. Optional Lexan dividers help organize inventory.

MICROPROCESSOR PERFORMANCE MONITOR/ALARM

• Microprocessor Performance Monitor/Alarm indicates temperature and alarm conditions on a two line, 40 character,alphanumeric readout in one low-voltage module.

• Monitor displays chamber temperature to a tenth of a degree C in an LCD digital readout, and also indicates high and low limit temperature conditions.

• Two temperature/alarm probes monitor chamber temperature and are conveniently located and easily removed for alarm system checks outside the cabinet. Probes are located in thermally-weighted product simulation bottles.

• Ringback alarm system, with audible and visual alarms, alerts technician to power loss and deviations beyond limits. Alarm can be temporarily silenced with the touch of a button. When silenced, the system will ring back every 30 minutes as a reminder to reset the alarm. Independent remote alarm contacts are provided for high and low alarms.

*Model 3866/3868 (3867) has double pane glass doors.

3

Model 3866/3868 (3867) Model 3881 (3889)

Page 4: Bloodbank Equipment

• Factory-set high/low alarms. Alarm setpoints are easily reprogrammed.

• Alarm silence button also silences door ajar alarm. System rings back every 30 seconds until door is closed. Warning light remains on until door is closed.

• Performance monitor tracks number of door openings in a 24 hr. period.

• Nine temperature alarm occurrences are stored in the monitor’s memory and may be reviewed at any time by the operator.

• Auto-charging nickel cadmium battery (minimum 72 hours) keeps alarm system operational during power failure.

• High and low alarm test system checks the entire alarm circuit at the touch of a button without opening the refrigerator door. Both temperature probes can be tested.

• RS-232 port permits computer interface and monitoring of temperature and alarms.

SPECIFICATIONS

Model 3881 (3889) Model 3883 (3887) Model 3866/3868 (3867)Operating Temperature +4°C (+39.2°F) +4°C (+39.2°F) +4°C (+39.2°F)

Temperature Uniformity ±0.8°C ±0.8°C ±0.7°C

Interior Dimensions 25.8"W x 55.0"H x 24.0"F-B 31.0"W x 60.0"H x 27.0"F-B 50.9"W x 59.1"H x 26.3"F-B

(65.5cm x 139.7cm x 61.0cm) (78.7cm x 152.4cm x 68.6cm) (129.3cm x 150.1cm x 66.8cm)

Drawers: 20.5"W x 3.3"H x 19.9"F-B Drawers: 29.1"W x 3.3"H x 23.8"F-B Drawers: 20.0"W x 3.3"H x 24.0"F-B

(52.1cm x 8.4cm x 50.5cm) (73.9cm x 8.4cm x 59.9cm) (50.7cm x 8.4cm x 61.0cm)

Exterior Dimensions 29.8"W x 74.3"H x 28.0"F-B* 38.0"W x 89.0"H x 34.0"F-B* 55.0"W x 79.0"H x 34.6"F-B*

(75.7cm x 188.7cm x 71.1cm) (96.5cm x 226.1cm x 86.4cm) (139.7cm x 200.7cm x 87.9cm)

*Add 2.5" (6.4cm) for condensate evaporator *Add 2.5"(6.4 cm) for condensate evaporator *Add 5.8" (14.7cm) for condensate evaporator

Capacity** 19.7 cu. ft. (588 liters) 29.1 cu. ft. (824 liters) 48 cu. ft. (1359.2 liters)

210 bags with dividers, 35 bags per drawer 392 bags with dividers, 56 bags per drawer 600 bags with dividers, 50 bags per drawer

240 bags without dividers, 40 bags per drawer 490 bags without dividers, 70 bags per drawer 672 bags without dividers, 56 bags per drawer

Drawers 6 adjustable solid stainless steel drawers 7 adjustable solid stainless steel drawers 12 adjustable solid stainless steel drawers

Interior Lights 4 fluorescent, 15 Watts each 6 fluorescent, 15 Watts each 6 fluorescent, 15 Watts each

Alarm Microprocessor Performance Monitor/Alarm System

Recorder 6", 7 day, single pen, inkless recorder

Refrigeration 1/3 HP air-cooled compressor, 1/2 HP air-cooled compressor, 1/2 HP air-cooled compressor,

Non-CFC refrigerant Non-CFC refrigerant Non-CFC refrigerant

Insulation Minimum 2" Non-CFC, foamed-in-place urethane. Glass door insulates and permits inventory viewing.

Electrical 3881: 120 VAC, 1 PH, 60 Hz, 10.5 FLA 3883: 120 VAC, 1 PH, 60 Hz, 12 FLA 3866: 120 VAC, 1 PH, 60 Hz, 10.5 FLA

3889: 220 VAC, 1 PH, 50 Hz, 6 FLA 3887: 220 VAC, 1 PH, 50 Hz, 6 FLA 3868: 220 VAC, 1 PH, 60 Hz, 16 FLA

3867: 220 VAC, 1 PH, 50 Hz, 16 FLA

Shipping Weight Ocean: 740 lbs. (336 kg) Ocean: 1200 lbs. (544 kg) Ocean: 1350 lbs. (612 kg)

Air/Container: 666 lbs. (302 kg) Air/Container: 1087 lbs. (493 kg) Air/Container: 1250 lbs. (567 kg)

Motor: 564 lbs. (256 kg) Motor: 1017 lbs. (461 kg) Motor: 1100 lbs. (499 kg)

Continuing research and improvements may result in specification changes at any time.**Unit capacity based on primary whole blood bag volume of 450ml.NOTE: All models CSA Certified; Model 3866/3868 (3867) UL Listed, CSA Certified and bears the CE Mark.

4

Control panel features an interior light switch, standard ink-less chart recorder and Forma's Microprocessor PerformanceMonitor System.

Page 5: Bloodbank Equipment

SLIM LINE BLOOD BANK REFRIGERATOR

For space-saving design in a +4°C blood storage unit, select Forma Scientific's Slim Line Blood BankRefrigerator. Three adjustable wire shelves, plus one solid stainless steel shelf, hold up to 40 wholeblood units. Standard fluorescent light illuminates the interior for easy inventory viewing.

PRODUCT FEATURES

• Two transparent inner doors protect stored product against temperature loss during door openings. Blowerautomatically cuts off when outer door is opened.

• LED digital temperature readout indicates chamber temperature at a glance.

• Alarm system activates audible and visual warnings when temperature exceeds preset low of +1.5ºC and high of +5.5ºC, or when a power failure occurs. The audible alarm can be easily silenced; total alarm system can be defeated with a switch.

• Alarm test system quickly checks the operation of the audible/visual alarms and battery back-up power supply.

• Battery back-up system keeps the alarm operational during power failure.

• Temperature/alarm probe is conveniently located for periodic checks of the alarm system. The probe, located in a thermally-weighted product simulation bottle, is easily removed.

• High/low alarm parameters are factory set (non-adjustable).

• Battery-operated, strip chart, inkless recorder is included standard and keeps monthly temperature records. Probe is located in a thermally-weighted product simulation bottle. Recorder continues to function during power outages.

• Door ajar warning light signals when door is left open.

• Triple-pane glass door permits inventory viewing.

• Built-in condensate evaporator eliminates the need for a drain line.

• Standard equipment includes adjustable leveling feet, access port with plug, key door lock and remote alarm contacts. Use contacts with #400011

Remote Alarm Monitor.

SPECIFICATIONS

Model 3644 (3646)Operating Temperature +4°C (+39.2°F) Uniformity: ±0.5°C

Interior Dimensions 12.6"W x 27.9"H x 13.8"F-B (32.0cm x 70.9cm x 35.1cm)

Exterior Dimensions 15.8"W x 59.6"H x 19.5"F-B (40.1cm x 151.4cm x 49.5cm)

Capacity* 2.8 cu.ft. (79.4 liters), 40 Bags

Shelves 3 adjustable zinc-plated wire shelves and 1 solid stainless steel shelf

Interior Light 1 fluorescent, 6 Watt

Alarm Electronic Temperature Monitor/Alarm System

Refrigeration 1/10 HP air-cooled compressor, Non-CFC refrigerant

Recorder Battery-operated, strip chart, inkless recorder. Chart paper records temperatures for one month.

Insulation Minimum 1.5" Non-CFC, foamed-in-place urethane in sides.

Electrical Model 3644: 120 VAC, 1 PH, 60 Hz, 2.5 FLA Model 3646: 220 VAC, 1 PH, 50 Hz, 1.3 FLA

Shipping Weight Ocean: 230 lbs. (104 kg) Air/Container: 200 lbs. (91 kg) Motor: 164 lbs. (74 kg)

Continuing research and improvements may result in specification changes at any time.*Unit capacity based on primary whole blood bag volume of 450ml.

5

Control panel includes an LED digital temperaturereadout, door ajar warning light, interior light switch,alarm system test switch, alarm on/off switches andindicating lights.

Model 3644 (3646)

Page 6: Bloodbank Equipment

6

CH/P® solid-state electronic temperaturemonitor/alarm is a standard feature.

Model 3626 (3629)

UNDERCOUNTER REFRIGERATOR

UL Listed and CSA Certified, this compact +4°C blood bank refrigerator is designed for undercounter installation. Ideally suited for emergency rooms, surgical suites and wher-ever space is limited, the Model 3626 (3629) features three solid stainless steel drawers thathold up to 84 whole blood units. Attractive, stainless steel cabinet houses molded, high impact ABS liner. Refrigerator is durable and easy to clean,inside and out.

PRODUCT FEATURES

• Door-mounted digital alarm/monitor continually displays temperature on an LCD read out.

• Alarm system activates audible and visual warnings when temperature exceeds pre-set low of +1.5ºC and high of +5.5ºC, or when a power failure occurs. Product

temperature can be visually verified on a right angle thermometer supplied in a thermally-weighted bottle.

• Ringback alarm system can be silenced for 60 minutes with a button, or defeated with a key switch. When conditions return to normal, the system rings back as a reminder to reset the alarm. The alarm temperature probe is located in a productsimulation bottle near the door for immediate detection of temperature change.

• Factory-set high/low alarms can be reprogrammed.

• Auto-charging nickel cadmium battery keeps alarm system operational during power failure.

• Automatic defrost cycle keeps refrigerator free from frost build-up. Condensate evaporator eliminates the need for a drain line.

• Blower automatically cuts off when door is opened, reducing cold air loss.

• Key door lock standard.

• Remote alarm contacts included for quick connection to #400011 Remote Alarm Monitor.

• Optional countertop converts the undercounter model to a freestanding unit. Optional casterdolly provides easy mobility.

• Optional inkless temperature recorder available for permanent records.

SPECIFICATIONS

Model 3626 (3629)Operating Temperature +4°C (+39.2°F)Temperature Uniformity ±0.5°CInterior Dimensions 21.0"W x 27.5"H x 19.5"F-B (53.3cm x 69.9cm x 49.5cm)Exterior Dimensions 24.0"W x 34.5"H x 25.8"F-B (61.0cm x 87.6cm x 65.5cm)Capacity* 7 cu.ft. (198.2 liters). Top and middle drawers hold 32 bags each, bottom drawer holds 20 bags.Drawers 2 perforated and 1 solid stainless steel removable drawers on Delrin rollers.Alarm Electronic Temperature Monitor/Alarm SystemRefrigeration 1/5 HP hermetically sealed, air-cooled compressor, Non-CFC refrigerant.Recorder 6", 7-day, single pen, inkless recorder (optional).Insulation Minimum 1-1/2" Non-CFC, foamed-in-place urethane.Electrical Model 3626: 120 VAC, 1 PH, 60 Hz, 4.8 FLA Model 3629: 220 VAC, 1 PH, 50 Hz, 3 FLAShipping Weight Ocean: 265 lbs. (120 kg) Air/Container: 230 lbs. (104 kg) Motor: 194 lbs. (88 kg)

Continuing research and improvements may result in specification changes at any time.*Unit capacity based on primary whole blood bag volume of 450ml. NOTE: Model 3626 UL Listed and CSA Certified.

Page 7: Bloodbank Equipment

7

UNDERCOUNTER PLASMA FREEZER

UL Listed and CSA Certified, this compact undercounter freezer provides a safe -35°C environ-ment for plasma storage. Two solid stainless steel drawers hold up to 80 plasma boxes withlabels up. Storage capacity may vary depending on configuration of plasma storage containers.Drawers roll out for easy access and quick identification. Model 8070 also features an optionalcountertop which converts it to a freestanding unit.

PRODUCT FEATURES

• Door-mounted digital alarm/monitor continually displays temperature on an LCD readout.

• Alarm system activates audible and visual warnings when temperature exceeds preset low of -40ºC and high of -20ºC, or when a power failure occurs.

• Factory set high/low alarms. Alarm setpoints can be reprogrammed.

• Alarm can be temporarily silenced with a button or defeated with a key switch. When the system returns to normal,the alarm rings back as a reminder to reset the alarm.

• Auto-charging nickel cadmium battery provides continuous alarm system operation during power failure. Audible alarm indicates when power failure occurs.

• Alarm probe is mounted in a product simulation bottle near door opening for immediate detection of temperature change.

• Built-in access port permits easy installation of probes for temperature monitoring systems or temperature recorder.

• Blower automatically shuts off when door is opened, reducing cold air loss.

• Stainless steel construction inside and out.

• Non-CFC, foamed-in-place, urethane insulation.

• Optional countertop converts the undercounter model to a freestanding unit. Optional caster dolly provides easy mobility.

• Automatic electric defrost cycle maintains frost-free operation. Condensate evaporator eliminates the need for a drain line.

• Key door lock standard. Optional inkless recorder available for permanent record keeping.

• Remote alarm contacts included for quick connection to #400011 Remote Alarm Monitor.

SPECIFICATIONS

Model 8070Operating Temperature -35°C (-31°F) Interior Dimensions 20.0"W x 19.4"H x 21.0"F-B (50.8cm x 49.3cm x 53.3cm) Exterior Dimensions 24.0"W x 34.5"H x 25.0"F-B (61.0cm x 87.6cm x 63.5cm) Capacity 4.7 cu.ft. (133.2 liters). Holds 80 single door plasma units, identification label up.Drawers 2 solid stainless steel removable drawers on Delrin rollers.Alarm Electronic Temperature Monitor/Alarm SystemRefrigeration 1/3 HP air-cooled compressor, hermetically sealed, Non-CFC refrigerant.Recorder 6", 7-day, single pen, inkless recorder (optional). Insulation Minimum 2" Non-CFC, foamed-in-place urethane on all sides, 2.5" on bottom.Electrical Model 8070: 120 VAC, 1 PH, 60 Hz, 7 FLA*Shipping Weight Ocean: 316 lbs. (143 kg) Air/Container: 264 lbs. (120 kg) Motor: 244 lbs. (111 kg)

Continuing research and improvements may result in specification changes at any time. *220V, 50 Hz model available upon request.

7

The Model 8070 featuresForma's CH/P® Temp-erature Alarm Monitor.Temperature is display-ed in LCD digital readout.Lights indicate high andlow alarms. The alarmbutton temp- orarilysilences the audible alarm.

Model 8070

DISCONTIN

UED

Page 8: Bloodbank Equipment

PLASMA FREEZER

CSA Certified, the Model 8097 (8103) Plasma Storage Freezer features a 19.7 cu. ft. storage capacity andholds up to 350 plasma boxes. Storage capacity may vary depending on configuration of plasma storagecontainer. The Non-CFC, foamed-in-place urethane insulation maintains the -30°C operating temperaturefor safe storage of single donor or fresh frozen plasma.

PRODUCT FEATURES

• Six inch, seven day, inkless recorder charts product temperature for permanent records. Product simulation bottles for recorder probe and both temperature alarm sensors are mounted near the door for immediate detection of temperature changes.

• Independent temperature controller is adjustable.

• Blower automatically cuts off when the door is opened, minimizing cold air loss.

• Solid stainless steel drawers lessen gravitational cold air loss and contain accidental spills. Optional stainless steel dividers are available for easy inventory arrangement.

• Automatic defrost cycle keeps freezer frost-free. Condensate evaporator eliminates the need for a drain line.

• Key lock and factory installed casters are standard.

MICROPROCESSOR PERFORMANCE MONITOR/ALARM

• Microprocessor Performance Monitor/Alarm indicates temperature and alarm conditions on a two line, 40 character,alphanumeric readout in one low-voltage module.

• Monitor displays chamber temperature to a tenth of a degree C in an LCD digital readout, and also indicates high and lowlimit temperature conditions.

• Two temperature/alarm probes monitor chamber temperature and are conveniently located and easily removed for alarmsystem checks outside the cabinet. Probes are located in thermally-weighted product simulation bottles.

• Ringback alarm system, with audible and visual alarms, alerts technician to power loss and deviations beyond limits. Alarm can be temporarily silenced with the touch of a button. When silenced, the system will ring back every 30 minutes as a reminder to reset the alarm. Independent remote alarm contacts are provided for high and low alarms.

• Factory-set high/low alarms. Alarm setpoints are easily reprogrammed.

• Alarm silence button also silences door ajar alarm. System rings back every 30 seconds until door is closed. Warning light remains on until door is closed.

• Performance monitor tracks number of door openings in a 24 hr. period.

• Nine temperature alarm occurrences are stored in the monitor’s memory and may be reviewed at any time by the operator.

• Auto-charging nickel cadmium battery (minimum 72 hrs.) keeps alarm system operational during power failure.

• High and low alarm test system checks the entire alarm circuit at the touch of a button without opening the

freezer door. Both temperature probes can be tested.• RS-232

port permits computer interface and monitoring of temperature and alarms.

8

Model 8097 (8103) control panel features a standard inklesschart recorder and Microprocessor Performance Monitor/Alarm.

Model 8097 (8103)

Page 9: Bloodbank Equipment

SPECIFICATIONS

Model 8097 (8103)Operating Temperature -30°C (-22°F)Interior Dimensions 25.8"W x 55.0"H x 24.0"F-B (65.5cm x 139.7cm x 61.0cm)Exterior Dimensions 29.8"W x 76.5"H x 29.0"F-B (75.7cm x 194.3cm x 73.7cm)Capacity 19.7 cu.ft. (558 liters). Holds 350 single donor plasma units identification label up.Drawers 7 solid stainless steel drawers on Delrin rollers. Dividers are optional.Alarm Microprocessor Performance Monitor/Alarm SystemRefrigeration 1/2 HP air-cooled compressorRecorder 6", 7-day single pen, inkless recorder.Insulation Minimum 2" Non-CFC, foamed-in-place urethane in sides; 3.25" in top and bottom; 2.5" in door.Electrical Model 8097: 120 VAC, 1 PH, 60 Hz, 8 FLA Model 8103: 220 VAC, 1 PH, 50 Hz, 4 FLAShipping Weight Ocean: 805 lbs. (365 kg) Air/Container: 780 lbs. (354 kg) Motor: 700 lbs. (318 kg)

Continuing research and improvements may result in specification changes at any time.

Stainless steel drawers hold inventory boxes for easy access. Stainless steel drawer dividers; the Model 8097 (8103) Blower keeps cold air circulating. Bottle includes temperature freezer holds seven dividers. (Stock #189570)probe to simulate product temperature.

PLASMA STORAGE FREEZER

UL Listed and CSA Certified, the Model 8536 (8540) Plasma Freezer features cold wall refrigerationand five inches of foamed-in-place urethane insulation for the ultimate safe storage of plasma, FactorVIII and rare types.

PRODUCT FEATURES

• Storage capacity is 300 plasma units. The storage capacity may vary depending on the configuration of the plasma storage containers or size of box used.

• The unit maintains a -40°C operating temperature. Freezer design protects inventory against warm-up during power failures, voltage fluctuations and repeated door openings.

• Optional stainless steel plasma storage racks slide out for easy reference and quick access. No inner doors are needed. Handling is reduced and cleaning and rearranging are fast and convenient. Durable, stainless steel interior is easy

to clean.

• Key lock, stabilizing feet and factory installed dual wheel casters are standard.

Model 8536 (8540)

9

Page 10: Bloodbank Equipment

• Independent electronic controller is used to set actual cabinet temperature.

• Six inch, seven day, inkless chart recorder is standard.

• Temperature controller, recorder and Enviro-Scan monitor probes are located in cabinet to provide actual chamber temperature where product is stored. Probes are positioned opposite the door hinges in the lower left

front corner of the chamber and will sound the alarm before product sees a rise in temperature.

• To detect and respond to high and low voltage situations, a built-in automatic voltage compensator is standard.

• Custom-designed inventory systems can be manufactured by Forma to accommodate your individual plasma storage needs.

Optional stainless steel storage racks hold individual plasma or cryoprecipitate units for fast identification and easy inventory arrangement.

ENVIRO-SCAN™ MICROPROCESSOR MONITORING SYSTEM

• Enviro-Scan incorporates a 40 character LCD digital readout, audible and visual indicators and a sensi-touch key pad to keep you informed of all freezer functions.

• The module digitally displays actual chamber tempera-ture and control setpoint, readable to the nearest

degree C. High and low alarm points can be quickly and easily programmed. Access code can be entered toprevent unauthorized adjustments.

• Audible and visual over and undertemperature alarms warn the technician if cabinet temperature extends beyond alarm limits. Display indicates high limit or low limit.

• Alarm can be temporarily silenced or defeated with the touch pad. The system rings back as a reminder to reset when the cabinet temp-

erature returns to normal.

• Exclusive alarm test system saves valuable time when checking alarm system readiness. Cabinet warm-up and cool-down can be safely simulated using the touch key pad located on the

monitor. The entire alarm circuit is tested. Both tests take only seconds and can be performed without opening the freezer.You can quickly and conveniently verify the high limit/low limit setpoint, assuring that the alarm system is ready if need-ed.

• Audible and visual alarms alert user to power failure. Battery back-up supplies power to the monitor during power failure (minimum 72 hour capacity if battery is fully charged). Monitor reports remaining battery capacity and includes a

low battery capacity alarm.

• Audible and visual door ajar alarm. Light flashes when door is open. Alarm will sound after 30 seconds.

• Monitors the refrigeration system for clogged condenser. Alerts operator of condition. An audible alarm is continuously sounded for 12 seconds every 15 minutes until the condition is corrected.

• An RS-232 port permits computer interface and monitoring of temperature and alarms. Remote alarm contacts andmillivolt output signal are standard features.

• Forma Scientific recommends the connection of external alarms for added product protection.

• Please refer to Forma’s Non-CFC 8500 Series ULT Freezers brochure for further information about the Enviro-Scan Microprocessor Monitoring System.

10

Page 11: Bloodbank Equipment

SPECIFICATIONS

Model 8536 (8540)Operating Temperature -40°C (-40°F)Interior Dimensions 23.0"W x 51.1"H x 19.3"F-B (58.4cm x 130.8cm x 49cm)Exterior Dimensions 33.3"W x 78.8"H x 31.0"F-B (84.6cm x 200.1cm x 78.7cm)Capacity 13 cu. ft. (368 liters). Holds 300 single donor plasma units or 600 cryoprecipitate units in 20 racks.

Total capacity -- 5 Stock #189010 and 15 Stock #183011 plasma inventory racks (optional). Inventory system to accommodate wedge shaped plasma storage boxes - Stock #189197. Twenty racks total per unit.

Shelves 4 stainless steel shelvesAlarm Enviro-Scan Microprocessor Monitoring SystemRefrigeration 1/2 HP air-cooled compressor, Non-CFC refrigerant.Recorder 6", 7-day, single pen, inkless recorder.Insulation Minimum 5" Non-CFC, foamed-in-place urethane in cabinet sides, 5" in floor, 4.5" in door.Shipping Weight Ocean: 900 lbs. (408 kg) Air/Container: 802 lbs. (364 kg) Motor: 662 lbs. (300 kg)Electrical 8536: 120 VAC, 1 PH, 60 Hz, 12.0 FLA 8540: 200-240 VAC, 1 PH, 50 Hz, 6.0 FLAAutomatic LOW: Cut In: 110V, Cut Out: 115V, Volts Boost: 10Voltage Cut In: 210V, Cut Out: 220V, Volts Boost: 18Compensator HIGH: Cut In: 125V, Cut Out: 120V, Volts Buck: 10

Cut In: 235V, Cut Out: 230V, Volts Buck: 18

Continuing research and improvements may result in specification changes at any time.NOTE: Model 8536 UL Listed, CSA Certified and bears the CE Mark.

RED CELL STORAGE FREEZERS

Forma's Red Cell Storage Freezers ensure product protection to -86°C. ULListed and CSA Certified, these chest and upright freezers are designed withproven cascade refrigeration systems and feature a minimum of 5" Non-CFC, foamed-in-place urethane insulation.

PRODUCT FEATURES

• Three capacities handle the needs of any blood bank operation. The 20 cu. ft. chest freezer holds up to 309 standard red cell canisters.

Model 8590 Model 8560 Model 8570 (8574) The 6.7 cu. ft. chest freezer holds up to 124 units. And the 13 cu. ft. upright freezer holds up to 198 units.

• Flat freezer section in Model 8560 eliminates canister bulges and holds six additional canisters.

• Chest models feature a flexible gasket that tightly seals the counterbalanced lid against the cabinet, while upright doors seal against a triple-point silicone gasket.

• For maximum energy savings, the cascade refrigeration system runs only on demand, then shuts off. Step-starting compressors reduce initial power drain.

• Independent electronic controller allows temperature to be set from -50ºC to -86ºC.

• Temperature controller, recorder and Enviro-Scan probes, located inside the freezer chamber, show actual temperature where product is stored. Probes are positioned opposite the door hinge in the lower left front corner of the chamber on the upright freezer. Chest freezer probes are located opposite the lid hinge in the upper left front corner. Probes are posi-tioned to sound the alarm before the stored product sees the rise in temperature.

• Stainless steel interior holds removable stainless steel inventory system. Inventory system (excludes canisters) is included standardwith the freezer.

• Factory installed dual wheel casters, stabilizing feet, inkless chart recorder and key lock are standard.

• Stock #828260 Storage Racks allow the Model 8560 Red Cell Storage Freezer to be used for plasma storage. Forma also manufactures customized inventory systems to permit plasma storage in chest units.

11

Page 12: Bloodbank Equipment

• To detect and respond to voltage fluctuations, a built-in voltage compen-sator is standard on every Forma Red Cell Storage Freezer. The auto-matic voltage compensator continuously monitors and adjustsincoming power to ensure compressor operation is within designparameters. For peace-of-mind performance, a panel mountedlight alerts operator to incoming voltage adjustments.

• Remote alarm contacts and millivolt output signal are standard features.

• Six inch, seven day inkless recorder charts product temperature for permanent records.

ENVIRO-SCAN™ MICROPROCESSOR MONITORING SYSTEM

• Enviro-Scan incorporates a 40 character LCD digital readout, audible and visual indicators and a sensi-touch key pad to keep you informed of all freezer functions.

• The module digitally displays actual chamber temperature and control setpoint,readable to the nearest degree C. High and low alarm points can be quickly and easily programmed. Access code can be entered toprevent unauthorized adjustments.

• Audible and visual over and undertemperature alarms warn the technician if cabinet temperature extends beyond alarm limits. Display indicates high limit or low limit.

• Alarm can be temporarily silenced or defeated with the touch pad. The system rings back as a reminder to reset when the cabinet temperature returns to normal.

• Exclusive alarm test system saves valuable time when checking alarm system readiness. Cabinet warm-up and cool-down can be safely simulated using the touch key pad located on the monitor.

The entire alarm circuit is tested. Both tests take only seconds and can be performed without opening the freezer. You can quickly and conveniently verify the high limit/low limit setpoint, assuring that the alarm system is ready if needed.

• Audible and visual alarms alert user to power failure. Battery back-up supplies power to the monitor during power failure (minimum 72 hour capacity if battery is fully charged). Monitor reports remaining battery capacity and includes a low battery capacity alarm.

• Audible and visual door/lid ajar alarm. Light flashes when door or lid is open. Alarm will sound after 30 seconds.

• Monitors the refrigeration system for clogged condenser. Alerts operator of condition. An audible alarm is continuously sounded for 12 seconds every 15 minutes until the condition is corrected.

• An RS-232 port permits computer interface and monitoring of temperature and alarms. Remote alarm contacts andmillivolt output signal are standard features.

• Forma Scientific recommends the connection of external alarms for added product protection.

• Please refer to Forma’s Non-CFC 8500 Series ULT Freezers brochure for further information on the Enviro-Scan Microprocessor Monitoring System.

12

Page 13: Bloodbank Equipment

6.7 cu. ft. chest freezer holds red cell canisters in a The upright red cell freezers hold canisters in stainless steel removable inventory rack. separate compartments for easy access.

Compartments also hold optional plasma racks. (Shown are Stock #4000354)

SPECIFICATIONS

Model 8560 Model 8570 (8574) Model 8590Operating Temperature -50°C (-58°C) to -86°C (-122.8°F)

Interior 23.0"W x 51.5”H x 19.3"F-B 38.0"W x 16.5"H x 18.5"F-B 66.5"W x 28.0"H x 18.5"F-B

Dimensions (58.4 cm x 130.8 cm x 49.0 cm) (96.5 cm x 41.9 cm x 47.0 cm) (168.9 cm X 71.1 cm X 47.0 cm)

Exterior 33.3"W x 78.8"H x 31.0"F-B 48.0"W x 41.3"H x 28.8”F-B 93.0"W x 40.5"H x 28.8"F-B

Dimensions (84.6 cm x 200.1 cm x 78.7 cm) (121.9 cm x 104.9 cm x 73.2 cm) (236.2 cm x 102.9 cm x 73.2 cm)

Add 3.0" (7.6cm) to width for handle/hinge Add 4.6" (11.7cm) to height for control Add 4.6" (11.7cm) to height for control

Add 7.0" (17.8cm) to F-B for control Add 6.6" (16.8cm) to F-B for control Add 6.6" (16.8cm) to F-B for control

panel/wall spacer panel/wall spacer panel/wall spacer

Capacity 13 cu.ft. (368 liters). Holds 6.7 cu.ft. (190 liters). Holds 124 20 cu.ft. (566 liters). Holds 309

198 red cell canisters, plus flat standard red cell canisters. standard red cell canisters.

freeze capacity for 6 canisters.

Note: Capacities based on minimum fill volume. Variances in volume may alter capacity.

Alarm Enviro-Scan Microprocessor Monitoring System

Refrigeration Two 1/2 HP air-cooled compressors, Two 1/3 HP air-cooled compressors, Two 1/2 HP air-cooled compressors,

cascade system, Non-CFC refrigerants cascade system, Non-CFC refrigerants cascade system, Non-CFC refrigerants

Recorder 6", 7-day, single pen, recorder.

Insulation Min. 5" Non-CFC, foamed-in-place Min. 5" Non-CFC, foamed-in-place Min. 5" Non-CFC, foamed-in-place

urethane; 4.5" in door urethane; 2" in lid with 1" sub-lids urethane; 2" in lid with 1" sub-lids

Electrical 208-230 VAC, 1 PH, 60 Hz, 12 FLA or 8570: 110-120 VAC, 1 PH, 60 Hz, 14 FLA 208-230 VAC, 1 PH, 60 Hz,12 FLA or

200-240 VAC, 1 PH, 50 HZ, 12 FLA 8574: 200-240 VAC, 1 PH, 50 Hz, 8 FLA 200-240 VAC, 1 PH, 50 Hz, 12 FLA

Shipping Weight Ocean: 1000 lbs. (454 kg) Ocean: 600 lbs. (272 kg) Ocean: 1200 lbs. (544 kg)

Air/Container: 964 lbs. (437 kg) Air/Container: 565 lbs. (256 kg) Air/Container: 1087 lbs. (493 kg)

Motor: 801 lbs. (363 kg) Motor: 492 lbs. (223 kg) Motor: 970 lbs. (440 kg)

Automatic LOW: Cut In: 110V, Cut Out: 115V, Volts Boost: 10

Voltage Cut In: 210V, Cut Out: 220V, Volts Boost: 18

Compensator HlGH: Cut In: 125V, Cut Out: 120V, Volts Buck: 10

Cut In: 235V, Cut Out: 225V, Volts Buck: 18

Continuing research and improvements may result in specification changes at any time. NOTE: Models 8560, 8590 UL Listed (Model 8570 UL pending). Models 8560, 8570, 8574, 8590 CSA Certified. Model 8560 bears the

CE Mark.NOTE: If recorder and inventory system are not required, request information on Forma's ultra-low temperature freezers (available

in various capacities).

13

Page 14: Bloodbank Equipment

14

CHEST PLASMA/BLAST FREEZERForma's plasma/blast freezers allow rapid freezing of single donor, apheresedand fresh frozen plasma. Rapid freezing also preserves Factor VIII.

Forma offers eight models in 4 cu. ft. capacities, including 220V and 440Vmodels, with air or water-cooled compressors. The 4 cu. ft. chest blast freez-er freezes up to (30) 650ml of pooled units, or 96 single donor units in 75 minutes.

PRODUCT FEATURES

• The 5 HP air or water-cooled compressor can be located upto 50 feet* from the main freezer cabinet, eliminating noise

and heat from the lab. Compressor is mounted on a stand complete with tubular frame and leveling feet.

• When the compressor is located outside the facility, subjected to the elements and cold climate, an optional winterizing package (crank case heater and prop-up valve) and weatherproof enclosure are recommended.

• Copper-tubed, copper-finned evaporator maximizes cooling efficiency.

• Stainless steel chamber holds up to 16 optional compression racks (Stock #600002), six plasma units per rack. Racks leave “proof of freeze” imprint and can be transferred to Forma's Model 2386 (2388) Cryo Bath for immediate

production of cryoprecipitate.

• Two styles of optional pans hold pooling bottles or flexible bags for freezing pooled or apheresed plasma.

• Optional stainless steel flat freeze racks are available in two styles to hold single units of plasma for quick freezing.

Stock #600029 Flat Freeze Racks hold 6 units of plasma. The 4 cu. ft. chest blast freezer holds 16 racks. This rack may also be transferred to Forma's Model 2386 (2388) Cryo

Bath for immediate production of cryoprecipitate.

Stock #600031 Flat Freeze Racks accommodate 12 units of plasma. The 4 cu. ft. chest blast freezer holds 8 racks.

*Distance from blast freezer cab-inet to compressor can be extend-ed

based on refrigeration line siz-ing and compressor location.

Model 8098 (8099)

Control panel features direct-setting indicating temperature con-troller and LED digital display.

Forma’s optional flat freeze racks can be utilized for freezing plasma. Air circulates freelythroughout the chamber surrounding all units.Plasma compression racks are also available.

Page 15: Bloodbank Equipment

PRODUCT FEATURES

• Two 1/4 HP blowers produce maximum air circulation throughout the cabinet. Safety switch cuts off blowers when the lid is opened.

• 5" Non-CFC, foamed-in-place, urethane insulation maintains the ultra-cold temperature. Sub-lids minimize temperature loss when lid is opened.

• Factory installed casters and leveling feet are standard on the blast freezer cabinet.

• Blast freezer cabinet includes drain line.

SPECIFICATIONS

Model 8087 (8088) Model 8089 (8091) Model 8094 (8095) Model 8098 (8099)Operating Temperature -50°C (-58°F)

Interior Dimensions 42.5"W x 11.25"H x 14.4"F-B (108.0cm x 28.6cm x 36.6cm)

Exterior Dimensions 61.0"W x 40.3"H x 28.5"F-B (154.9cm x 102.4cm x 72.4cm), add 7.0" (17.8cm) to back for motor

Compressor Stand 60.0"x 30.0"x 36.0"H (152.4cm x 76.2cm x 91.4cm)

Capacity 4 cu. ft. (113.4 liters)

Temperature Monitor Electronic digital temperature display/controller

Refrigeration Remote Water Cooled 5 HP Remote Air Cooled 5 HP Remote Water Cooled 5 HP Remote Air Cooled 5 HP

(Non-CFC Refrigerant) compound compressor compound compressor compound compressor compound compressor

Insulation Min. 5” Non-CFC, foamed-in-place urethane in cabinet; 2" in lid; 1” polystyrene sub-lids

Cabinet Electrical Models 8087, 8089, 8094, 8098: 120 VAC, 1 PH, 2 Wire, 50/60 Hz, 5.8 FLA

Models 8088, 8091, 8095, 8099: 208-240 VAC, 1 PH, 2 Wire, 50/60 Hz, 3 FLA

Compressor Electrical Model 8087: 187-253 VAC, Model 8089: 414-506 VAC, Model 8094: 414-506 VAC, Model 8098: 187-253 VAC,

3 PH, 3 Wire, 60 Hz, 19.6 FLA 3 PH, 3 Wire, 60 Hz, 12 FLA 3 PH, 3 Wire, 60 Hz, 8.3 FLA 3 PH, 3 Wire, 60 Hz, 23 FLA

Model 8088: 180-242 VAC, Model 8091: 342-418 VAC, Model 8095: 342-418 VAC, Model 8099: 180-242 VAC,

3 PH, 3 Wire, 50 Hz, 19.6 FLA 3 PH, 3 Wire, 50 Hz, 12 FLA 3 PH, 3 Wire, 60 Hz, 8.3 FLA 3 PH, 3 Wire, 50 Hz, 23 FLA

Cabinet Shipping Weight Ocean: 840 lbs. (381 kg)

Air/Container: 800 lbs. (363 kg)

Motor- 800 lbs. (363 kg)

Compressor Shipping Weight Ocean: 740 lbs. (336 kg)

Air/Container: 700 lbs. (317 kg)

Motor: 700 lbs. (317 kg)

Continuing research and improvements may result in specification changes at any time.NOTE: Periodic defrosting of the cabinet is necessary for proper operation of the unit. When freezing is completed and the product removed,a defrost cycle can be manually initiated to shut down the compressor and keep the cabinet frost-free. When defrosting is completed, theuser turns off the cabinet.NOTE: Blast freezer cabinet, compressor and remote refrigeration lines require installation. For assistance, contact Forma's Customer Service Department for details.

15

Compressor, with mounting stand, can be located up to 50 feet away from the freezer cabinet. Compressor and remote refrigeration lines require installation. Compressor enclosure is optional.

Page 16: Bloodbank Equipment

UPRIGHT PLASMA/BLAST FREEZER

Forma Scientific's Model 8001 (8003) Upright Plasma/Blast Freezer offers space savings,performance and versatility. This 4.2 cu.ft. freezing unit requires only 8.6 sq. ft. of floor space. The blast freezer capabilities allow freezing of 99 single plasma units in 75 minutes.

PRODUCT FEATURES

• Unit is complete with direct-setting indicating temperature controller and digital display for maintaining desired operation.

• Interior is constructed of heavy gauge stainless steel. Powder coated exterior provides a uniform surface and is resistant to chipping and rust. Cabinets meet ASTM Standard B117-85 for salt spray testing that exceeds 1000 hours.

• Standard with 11 stainless steel wire baskets, complete with removable roll-out drawers. Baskets can be removed from drawers for loading. (For additional stainless steel wire baskets, order Stock #600032.)

• 5" Non-CFC, foamed-in-place, urethane insulation surrounds cabinet; door has 4.5" of insulation.

• For loading and unloading convenience, bottom shelf is 24" from building floor, while top shelf is 58" from building floor.

• Fastory installed casters and leveling feet are standard.

• 5 HP compressor with Non-CFC refrigerant, air or water-cooled, includes stand with tubular frame and leveling feet. It is designed to be remote mounted up to 50 feet* from the cabinet.

• Copper-finned, copper-tubed evaporator maximizes cooling efficiency.

• Triple point silicone gasket ensures tight door seal.

• Single draw door latch with handle allows easy access to the freezer chamber. Key lock is standard.

• Stainless steel sloped drain pan located in bottom of blast freezer chamber. Drain connection is provided at rear of cabinet.

• Externally mounted blower motor (not pictured) provides efficient airflow.

• At time of order select Compressor/Condensing Unit. (See Specifications)

• When the compressor is located outside the facility, subjected to the elements and cold climate, an optional winterizingpackage (crank case heater and prop-up valve) and weather-proof enclosure are recommended.

*Distance from blast freezer cabinet to compressor cabinet can be extended based on refrigeration line sizing and location of compressor.

16

Model 8001 (8003)

Page 17: Bloodbank Equipment

SPECIFICATIONS

Model 8001 (8003)Operating Temperature -50°C (-58°F)

Interior Dimensions 23.0"W x 38.5"H x 16.9"F-B (less interior rack assembly and shelves) (58.4cm X 97.8cm X 42.9cm)Plasma freezing area: 19.5"W x 24.2"H x 15.9"F-B (4.2 cu.ft.)Stainless steel wire shelf loading area: 19.5"W x 15.9"F-B (2.14 sq. ft.)Dimension from shelf to shelf: 3.5" (8.9cm)

Exterior Dimensions 33.3"W x 79.0"H x 37.0"F-B (84.6cm x 200.7cm x 94.0cm). Top mounted blower adds 11.0” (27.9cm) to external height.*Compressor Stand 60.0" x 30.0" x 36.0"H (152.4cm x 76.2cm x 91.4cm)

Capacity 4.2 cu. ft. (118.9 liters)

Temperature Control Electronic digital temperature display/controller

Refrigeration Remote air-cooled or water-cooled 5 HP compressor, Non-CFC Refrigerant

Insulation 5" Non-CFC, foamed-in-place urethane in cabinet; 4.5" in door.Cabinet Electrical Model 8001: 120 VAC, 1 PH, 50/60 Hz, 5 FLA Model 8003: 208-230 VAC, 1 PH, 50/60 Hz, 3 FLA

Compressor Electrical Stock # Stock #189259 187-253 VAC, 3 PH, 3 Wire, 60 Hz, 23 FLA, air-cooled 189261 414-506 VAC, 3 PH, 3 Wire, 60 Hz, 12 FLA, air-cooled189259 180-242 VAC, 3 PH, 3 Wire, 50 Hz, 23 FLA, air-cooled 189262 414-506 VAC, 3 PH, 3 Wire, 60 Hz, 8.8 FLA, water-cooled189260 187-253 VAC, 3 PH, 3 Wire, 60 Hz, 19.6 FLA, water-cooled 189276 342-418 VAC, 3 PH, 3 Wire, 50 Hz, 12 FLA, air-cooled189260 150-242 VAC, 3 PH, 3 Wire, 50 Hz, 19.6 FLA, water-cooled 189277 342-418 VAC, 3 PH, 3 Wire, 50 Hz, 8.8 FLA, water-cooled

Shipping Weight Cabinet:Ocean: 950 lbs. (431 kg) Air/Container: 890 lbs. (404 kg) Motor: 770 lbs. (349 kg)

Compressor:Ocean: 740 lbs. (336 kg) Air/Container: 700 lbs. (317 kg) Motor: 700 lbs. (317 kg)

Continuing research and improvements may result in specification changes at any time.*Unit is shipped without installation of the blower motor to allow for easier maneuvering through doorways.NOTE: Periodic defrosting of the cabinet is necessary for proper operation of the unit. When freezing is completed and the product removed, a defrost cycle can be manually initiated to shut down the compressor and keep the cabinet frost-free. When defrosting is com-pleted, the user turns off the cabinet.NOTE: Blast freezer cabinet, compressor and remote refrigeration lines require installation. For assistance, contact Forma's

Inside Sales Support Department for details.

Control panelfeatures direct-setting indicat-

ing temperature Cabinet features stainless steel drawers with removablecontroller and LED digital display. wire shelves. (Wire Shelf Stock #600032) 17

Page 18: Bloodbank Equipment

CRYOPRECIPITATE BATH

Forma Scientific's Cryoprecipitate Bath controls thawing at +4°C and improves yields by 50%over conventional air thaw methods. The CH/P® temperature control system maintains precisetemperature for consistent day-to-day operations.

PRODUCT FEATURES

• Direct digital temperature control displays setpoint and operating temperature in tenths of a degree.

• Fine-tune adjustment selects temperature setpoint with exceptional accuracy.

• Seventeen gallon tank holds five optional stainless steel compression racks (Stock #600002) or four optional flat freeze racks (Stock #600029). Optional plasma rack holder available (Stock #185012).

• Fifteen units thaw in approximately 90 minutes. Thawing time is extended if an overwrap is used.

• Stainless steel tank resists corrosion and staining. A 3/4", durable, stain-resistant DuPont Corian top surrounds the work area.

• Stainless steel lid standard.

• Drain, with shut-off valve, can be connected to existing plumbing for easy draining and cleaning of tank.

• Unit includes dual wheel casters for complete mobility.

• Forma can custom design baths, such as a 50-gallon thawing bath, for increased thawing capabilities.

Cryo bath controls include LCD digital readout, coarse and fine setpoint

adjustment, power and refrigeration switches, and indicator lights for

17 gallon tank holds 5 plasma compression racks. power, refrigeration, heat and cooling.

SPECIFICATIONS

Model 2386 (2388)Operating Temperature +4°C (+39.2°F)

Temperature Uniformity ±0.02°CUsable Interior Dim. 21.0"W x 12.0"H x 13.5"F-B (53.3cm x 30.5cm x 34.3cm)Exterior Dimensions 34.3"W x 34.5"H x 22.3"F-B (87.1cm x 87.6cm x 56.6cm) Capacity 17 gallons (64.3 liters) Heat Source 12" long flanged heaterRefrigeration 1/4 HP air-cooled compressorCirculation Stainless steel pumpElectrical 2386: 120 VAC, 1 PH, 60 Hz, 12 FLA 2388: 220 VAC, 1 PH, 50 Hz, 6 FLATemperature Monitor Digital temperature readoutShipping Weight Ocean: 396 lbs. (180 kg) Air/Container: 370 lbs. (168 kg) Motor: 304 lbs. (138 kg)

Continuing research and improvements may result in specification changes at any time.

18

Model 2386 (2388)

Page 19: Bloodbank Equipment

THAWING BATH

The Forma Thawing Bath is designed for rapid and uniform thawing of frozen plas-ma and red cell units.

PRODUCT FEATURES

• Unit thaws one to eight plasma units in 25 minutes. Thawing time is extended if an overwrap is used.

• Up to four red cell units thaw in only 12 minutes.

• Overtemperature safety thermostat protects product during thawing.

• Adjustable 30-minute timer rings when thawing is completed.

• Stainless steel filter screen protects pump intake from debris such as labels.

• Unit comes complete with choice of two stainless steel standard size or one wedge shaped thawing rack. Unit will also accommodate one optional red cell canister rack.

• Stainless steel tank features coved corners for easy cleaning.

• Stainless steel lid and stem thermometer standard.

• Drain line, with shut-off valve, can be connected to existing plumbing.

Thawing bath controls include timer, fixed and variable temp-erature control, overtemperature safety limit, power and pump switches.

SPECIFICATIONS

Model 2032 (2036)Operating Temperature +5°C above ambient to +60°C (+140°F) Control: ±0.4°C

Usable Interior Dim. 16.7"W x 8.0"H x 12.8"F-B (42.4cm x 20.3cm x 32.5cm)

Exterior Dimensions 30.5"W X 16.3"H X 16.4"F-B (77.5 cm X 41.4 cm X 41.7 cm)

Capacity 7.6 gallons (28.8 liters)

Temperature Monitor Overtemperature safety thermostat

Heat Source Low watt density silicone heater

Circulation Stainless steel pump

Electrical 2032: 120 VAC, 1 PH, 60 Hz, 11 FLA

2036: 220 VAC, 1 PH, 50 Hz, 7 FLA

Shipping Weight Ocean: 145 lbs. (66 kg)

Air/Container: 134 lbs. (61 kg)

Motor: 79 lbs. (36 kg)

Continuing research and improvements may result in specification changes at any time.NOTE: Model 2032 CSA Certified.

19

Model 2032 (2036)

Standard size plasma racks hold up to eight units; perforated design allows for complete circulation. Wedgeshaped racks accommodate 12 storage boxes.

Page 20: Bloodbank Equipment

PLATELET INCUBATORS

Don't rely on a fluctuating building environment for storing your platelets. Select the +22°C protection of Forma's exclusive platelet incubators. CSA Certified, Models 3603 (3609) and 3606 (3612) improve the viability of platelet concentrates.

PRODUCT FEATURES

• Single and dual shelf capacity incubators hold platelet rotators and agitators for continual mixing at +22°C.

• Heavy-duty shelves roll out for easy access to platelet packs. Model 3603 (3609) includes two shelves. Model 3606 (3612) includes one.

• Forced-air circulation provides temperature uniformity within ±0.5°C from top to bottom. High density insulation surrounds the cabinet for temperature stability.

• Independent temperature controller allows temperature to be set at a fixed +22°C or controlled across a range from +15°C to +40°C.

• Standard 6", 7-day inkless chart recorder keeps track of all temperature levels for permanent blood bank records.

• Durable stainless steel interior wipes down for easy cleaning. Port with plug measures 3/8" diameter.

• Recessed, chamber mounted electrical outlets provided for platelet motion devices.

MICROPROCESSOR PERFORMANCE MONITOR/ALARM

• Microprocessor Performance Monitor/Alarm indicates temperature and alarm conditions on a two line, 40 character,alphanumeric readout in one low-voltage module.

• Monitor displays chamber temperature to a tenth of a degree C in an LCD digital readout, and also indicates high and lowlimit temperature conditions.

• Two temperature/alarm probes monitor chamber temperature and are conveniently located and easily removed for alarmsystems checks outside the cabinet. Probes are located in thermally-weighted product simulation bottles.

• Ringback alarm system, with audible and visual alarms, alerts technician to power loss

and deviations beyond limits. Alarm can be temporarily silenced with the touch of a button. When silenced, the system will ring

back every 30 minutes as a reminder to reset the alarm. Independent alarm contacts are provided for high and low alarms.

• Factory-set high/low alarms. Alarm setpoints are easily reprogrammed.

• Alarm silence button also silences door ajar alarm. System rings back every 30 seconds until door is closed. Warning light remains on until door is closed.

• Performance monitor tracks number of door openings in a 24 hr. period. Nine temperature alarm occurrences are stored in themonitor’s memory and may be reviewed at any time by the operator.

• Auto-charging nickel cadmium battery (minimum 72 hours) keeps alarm system operational during power failure.

• High and low alarm test system checks the entire alarm circuit at the touch of a button without opening the incubator door. Both temperature probes can be tested.

• RS-232 port permits computer interface and monitoring of temperature and alarms.

20

Model 3603 (3609) Model 3606 (3612)

The control panel on Model 3603 (3609) features a power switch, circuit breaker, fixed orvariable temperature control, Microprocessor Performance Monitor/Alarm and an inkless chart recorder.

Page 21: Bloodbank Equipment

21

SPECIFICATIONS

Model 3603 (3609) Model 3606 (3612)Operating Temperature +15°C to +40°C (+59°F to +104°F) with a fixed setpoint at +22°C (+71°F)

Control: ±0.1°CUniformity: ±0.5°C at +22°C (+71°F)

Interior Dimensions 37.0"W x 30.3"H x 16.5"F-B 37.0"W x 15.0"H x 16.5"F-B(94.0cm x 77.0cm x 41.9cm) (94.0cm x 38.1cm x 41.9cm)

Exterior Dimensions 39.0"W x 47.0"H x 22.0"F-B 39.0"W x 29.0"H x 22.0"F-B(99.1cm x 119.4cm x 55.9cm) (93.0cm x 74.0cm x 42.0cm)

Capacity 2 solid, roll-out shelves 1 solid, roll-out shelf Alarm Microprocessor Performance Monitor/Alarm SystemRefrigeration 1/6 HP air-cooled compressor, Non-CFC refrigerantCirculation Forced draft Electrical 3603: 120 VAC, 1 PH, 60 Hz, 6.5 FLA 3606: 120 VAC, 1 PH, 60 Hz, 6.5 FLA

3609: 220 VAC, 1 PH, 50 Hz, 4 FLA 3612: 220 VAC, 1 PH, 50 HZ, 4 FLA Models include four each 1.5 Amp outlets Models include two each 1.5 Amp outlets

Recorder 6", 7-day, single pen inkless recorderShipping Weight Ocean: 505 lbs. (229 kg) Ocean: 400 lbs. (181 kg)

Air/Container: 470 lbs. (213 kg) Air/Container: 365 lbs. (166 kg)Motor: 470 lbs. (213 kg) Motor: 295 lbs. (134 kg)

Continuing research and improvements may result in specification changes at any time.

ACCESSORIES - BLOOD BANK REFRIGERATORS

Description Order No. Model Model Model Model3881 (3889) 3883 (3887) 3626 (3629) 3866/3868 (3867)

Caster Dolly, Customer Installed. 189649 •Casters, Factory Installed. 189653 • •Casters, Factory Installed. 189905 •Casters, Customer Installed. 189654 • • Black Laminate Countertop.

Converts model to freestanding 189003 •unit, Customer Installed.

Lexan Inner Door Kit, Factory

Installed. Non-hinged, mounts 189187 •to drawer front.

Lexan Inner Door Kit, Factory

Installed. Non-hinged, mounts 189571 •to drawer front.

6", 7-day Inkless Chart Recorder, 201005 (120V)

Customer Installed. 201010 (220V)•

Lexan Drawer Dividers.

Holds 50 bags per drawer. 505054 •Refrigerator holds 12 dividers.

Lexan Drawer Dividers.

Holds 35 bags per drawer. 505057 •Refrigerator holds 6 dividers.

Lexan Drawer Dividers.

Holds 56 bags per drawer. 505058 •Refrigerator holds 7 dividers.

Bottle With Stopper* 890138 • • • •Thermometer, Right Angle 940561 • • •-5°C to +30°C

Thermometer, Right Angle 940246 •-40°C to +30°C

* Two right angle thermometers in bottles with stoppers included with Models 3881, 3889, 3883, 3887, 3866, 3868 and 3867. One included with 3626 and 3629.

Page 22: Bloodbank Equipment

Stainless steel plasma storage racks for use in Forma freezers.

ACCESSORIES - BLOOD COMPONENT STORAGE FREEZERS

Model Model Model Model Model ModelDescription Order No. 8097 (8103) 8536 (8540) 8070 8590 8570 (8574) 8560

Small Plasma Storage Rack.

Freezer holds 5 small plasma racks.

Each rack holds 12 plasma 189010 •boxes. One small rack and 3

large racks (189011) per shelf.

Large Plasma Storage Rack.

Freezer holds 15 large plasma racks.

Each rack holds 16 plasma boxes. 189011 •Three large racks and 1 small rack

(189010) per shelf.

Plasma Inventory System.

Accommodates wedge-shaped 189197 •plasma boxes. Freezer holds

20 racks.

Plasma Storage Rack.

Holds 12 plasma boxes per rack. 828260 •Freezer holds up to 10 racks,

2 per shelf.

Stainless Steel Drawer Divider. 189570 •Freezer holds 7 dividers.

Lexan Inner Door 189194 •Kit (attached to drawers)

Casters, Customer Installed. 189651 •6", 7-day Inkless Chart Recorder,

Customer Installed. 201005 •Black Laminate Countertop.

Converts model to freestanding unit, 830015 •Customer Installed.

Thermometer, Right Angle. 940246 • • •-40°C to +30°C

Plain Blood Bag Swing Arm Canisters. 4000354 • • •Dimpled Blood Bag Swing Arm Canisters. 4000355 • • •Plasma Storage Rack. 828367 17.3 cu. ft. upright freezer holds 19 racks. Holds 16 boxes per rack. 23.0 cu. ft. upright freezer holds 24 racks.

Continuing research and improvements may result in specification changes at any time. NOTE: Forma Scientific provides standard a 6", 7-day inkless chart recorder on red cell freezers. An optional strip chart recorder

is available upon request. Please contact our Inside Sales Support Department for ordering information.

22

Page 23: Bloodbank Equipment

Plasma flat freeze racks for use in chest style freezers. Plasma compression racks for use in chest style plasma/blast freezers also available. (Shown above: L-R, Stock #600031 & #600029).

ACCESSORIES - BLAST FREEZERS

Description Order No. Model Model Model Model Model 8087 (8088) 8089 (8091) 8094 (8095) 8098 (8099) 8001 (8003)

Pooling Bottle Pan for 4 cu.ft. units. 189005 • • • • •Holds up to 30 Fenwal pooling bottles.

Pooling Bag Pan for 4 cu. ft. units 189006 • • • • •Compressor Stand Enclosure for 4 cu. ft. units 189007 • • • • •Flat Freeze Rack. Holds 6 units of frozen plasma. 600029 • • • • •Designed to be stacked. 4 cu. ft. unit will hold 16.

Flat Freeze Rack. Holds 12 single plasma units. 600031 • • • • •Designed to be stacked. 4 cu. ft. unit will hold 8.

Stainless Steel Wire Basket.600032 •Plasma Compression Rack. Holds 6 single donor

plasma bags. 4 cu. ft. unit holds 16 racks. 600002 • • • • •Also for use in Model 2386 (2388) Cryoprecipitate

Bath, holds 5 racks.

Winterizing Package. For 208V, 3 PH, 4 and 4.2 cu. ft. 189028 •air-cooled blast freezer compressor.

Winterizing Package. For 480V, 3 PH, 4 and 4.2 cu. ft. 189029 •air-cooled blast freezer compressor.

ACCESSORIES - THAWING BATHS

Description Order No. Model 2032 (2036) Model 2386 (2388)Plasma Thawing Rack 185048 •(holds 8 standard size plasma boxes)

Plasma Thawing Rack 185055 •(holds 12 wedge-shaped plasma boxes)

Red Cell Canister Rack 185004 •Compression Rack Holder 185012 •Plasma Compression Rack 600002 •Stainless Steel Flat Thawing Rack 600029 •Thermometer, +20°C to +60°C 942030 •

23

Page 24: Bloodbank Equipment

ACCESSORIES - SAFE TEMP GLOVES

Item Order No.

Wrist Length Gloves Small, one pr. 189573Medium, one pr. 189574Large, one pr. 189575

Mid-Arm Length Gloves Medium, one pr. 189561Large, one pr. 189562Extra large, one pr. 189563

Elbow Length Gloves Medium, one pr. 189564Large, one pr. 189565Extra large, one pr. 189566

Shoulder Length Gloves Medium, one pr. 189567Large, one pr. 189568Extra large, one pr. 189569

Red cell canister rack holds 4 standard red cell canisters. Model 2032 (2036) Plasma Thaw Bath holds one rack. Stock #185004

ACCESSORIES - SAFETY BACK-UP SYSTEMSCSA Certified, the CO2 and LN2 Back-up Systems utilize an electronic controller with a platinum temperature probe. The temperature

probe monitors the cabinet temperature within the freezer chamber. If the power fails, or a component malfunctions, the CO2 or LN2 Back-up System takes over, automatically injecting CO2 or LN2.

The back-up systems are factory set, but have a temperature range of -30°C to -140°C which can be field adjusted through the direct setting temperature controller. Batteries continually recharge during normal field operation to provide power to the back-up system in the event of a power failure. A minimum 72 hour battery life (based on a full charge) is provided. The back-up system is complete with test switch and power monitor light.

Item Order No. DescriptionCO2 Back-up System 115V, 60 Hz 189050 Designed to inject CO2 into freezer compartment if temperature rises above preset

safety level. Automatic switching prohibits CO2 injection when freezer door is opened. System can be connected to multiple liquid CO2 supply bottles. CO2

back-up system is factory set at -60°C.

CO2 Back-up System 220V, 50 Hz 189055

Deluxe CO2 Back-up System 189051 Same as CO2 Back-up System but includes CO2 wall clamp, hose and necessary 115V, 60 Hz fittings.

Deluxe CO2 Back-up System 189056220V, 50 Hz

LN2 Back-up System 115V, 60 Hz 189052 Will maintain a temperature below that of a CO2 back-up system. Can be connected to multiple LN2 supply bottles. LN2 loss is limited to normal tank bleed-off. Factory set at -60°C.

LN2 Back-up System 220V, 50 Hz 189057

Back-up System AccessoriesWall Clamp 950316 CO2 bottle holder for securing tank to wall in accordance with OSHA regulations. High Pressure CO2 Hose 991505 Kink resistant type 316 stainless steel with stainless steel liner covered with double-

braided stainless steel sheath. 1/4" NPT type 316 stainless steel female fittings. Maximum working pressure 2500 PSIG in 6 ft. lengths.

24

Forma Scientific Safe Temp Gloves are available in various sizes. Select wrist, mid-arm, elbow and shoulder lengths.

Page 25: Bloodbank Equipment

25

ACCESSORIES - GENERAL

Monitor/Alarm System • Model 1535 interfaces with as many as 24 products (channels) to

monitor equipment conditions up to 2000 ft. away.

• A two line, 40 character LCD readout continusouly updates and displays temperature, CO2, relative humidity, power and remote alarm contact closure.

• Programmable audible/visual alarms (with delay and ringback) alert the operator to deviations in monitored equipment status.

• A dot matrix printer records data, time, unit number, channel status and designator in user specified intervals from 1 minute to 24 hours and provides an automatic printout when a monitored unit goes in or out of alarm.

• An easy-to-use, six button keypad and menu prompting provide simple programming steps for the operator.

• Non-volatile memory retains programming parameters in the event of a power outage.

• Daisy chain wiring allows easy setup and equipment can be added or removed from the system as needs change without interrupting communications.

• Modules accept analog signals, dry contacts and RS-232 outputs or have their own RTD probes. The modules convert input signals to RS-485 for transmission to Monitor/Alarm System.

• An optional floppy disk drive can be factory installed for time stamped data and alarm storage.

• In the event of an alarm, a signal can be transmitted to the optional Sensaphone Telephone Dialing System which will continuously dial up to four different pre-programmed numbers in sequence until one user is reached.

Model 1535

Dimensions 14.0"W x 6.0"H x 13.8"F-B (35.6cm x 15.2cm x 34.9cm)

Electrical 90-125V, 50/60 Hz, 1 PH, 0.7 FLA or 180-250V, 50/60 Hz, 1 PH, 0.35 FLA

Weight 20 lbs. (9.1 kg)

Monitor/Alarm System Accessories

Stock # Description

190456 Cryogenic Converter Module, RTD Probe and Dry Contacts, -200°C to -90°C190457 Converter Module, RTD Probe and Dry Contacts, -90°C to +70°C190458 Converter Module, RS-232 Port and Dry Contacts190459 Converter Module, (4) Analog Signals and (2) Dry Contacts430222 4 Wire Cable, 1 ft.190470 Wall Mounting Bracket190477 Factory Installed Floppy Disk Drive (3.5"/1.44MB)197019 (10) Dot Matrix Paper Rolls197066 Printer Ribbon

Model 1535

Page 26: Bloodbank Equipment

Stock #400047 Sensaphone Telephone Dialing System

• Monitors alarms, electrical power and ambient temperature where the

system is located.

• Can be connected to the remote alarm contacts on Forma products or

the 1535 Monitor/Alarm System to alert user of alarm conditions.

• Will automatically call up to four parties in sequence until the warning

message (stated in English) is received and acknowledged. All messages

are pre-programmed.

• Allows the technician to call the lab or office and receive the status of

all monitored conditions. The system automatically hangs up after

completion of call.

Stock # 400047

Dimensions 7.8"W x 2.0"H x 8.6"F-B (19.8cm x 5.1cm x 21.9cm)

Electrical 115V, 60 HzWeight 2 lbs. (0.9kg)

Stock #290003 Temperature/Power Monitor

• Attaches to wall or cabinet for single unit monitoring of power supply and temperature from -90°C to +40°C.• Audible and visual alarms alert user to over and undertemperature conditions and power loss.• Alarms can be silenced by switching to standby.• Battery back-up power keeps monitor operational even during power failures.• Self-contained probe avoids reliance on remote contacts.• Monitor includes power contacts. Can be connected to Stock #400011 and 400040 Remote Alarm Audible Signals

or other monitoring devices.• 105-125V, 60 Hz.

Stock #400048 (400049) Digital Temperature/Power Monitors

• Provides single unit monitoring of power supply and temperature from -130°C to +99°C.

• Digital LCD readout displays temperature.• Audible and visual alarms for power failure and high and low temperature conditions.• High and low alarm parameters are user adjustable.• Alarms can be silenced and ring back as a reminder to reset when conditions return

to normal.• High temperature alarms features test system.• Recharging nickel cadmium battery provides back-up protection.• Millivolt output signal and dry remote alarm contacts are included. Monitor can be

connected to Stock #400011 and 400040 (400041) Remote Alarm Audible Signals or other monitoring devices.

• Stock #400048: 115V, 60 Hz Stock #400049: 220V, 50 Hz

Stock #400011 Remote Alarm Audible Signal

• For areas where an alarm might not be readily heard, Forma’s Remote Alarm Audible Signal extends an existing alarm system to another location.

• Lightweight unit attaches to any wall or cabinet and operates off a plug-in charger module; switches to battery back-up if the power fails.

• Easily connects to Stock #290003 and 400048 Temperature/Power Monitors, as well as other standard Forma products. • 115V, 60 Hz

26

Page 27: Bloodbank Equipment

Stock #400040 (400041) Remote Alarm Audible/Visual Signals

• For areas where an alarm might not be readily seen or heard, Forma’s Remote Alarm Audible/Visual Signal extends an existing alarm system to another location.• Audible and visual alarms indicate deviations on monitored equipment.• Lightweight unit attaches to any wall or cabinet and operates off a plug-in charger module; switches to battery back-up

if the power fails.• Alarms can be silenced and ring back as a reminder to reset when conditions return to normal.• Easily connects to Stock #290003 and 400048 (400049) Temperature/Power Monitors, as well as other standard Forma

products.• Stock #400040: 115V, 60 Hz Stock #400041: 220V, 50 Hz

Stock #853227 Pocket Digital Thermometer

• Provides instantaneous LCD readouts of temperatures from -90°C to +60°C.• Platinum probe indicates temperature changes in seconds.• Fingertouch control switches unit on and off.• Additional Stock #292001 Probes are available to be installed in all monitored equip-ment; quick connect wires link probe to thermometer for easy movement from one

probe to another.• Operates on a 9-Volt battery.

Stand Alone Voltage Compensators

Item Order No. Description120V, 1 PH, 50/60 Hz, 15 Amp Receptacle 8960Maximum Load: 12 AmpsComplete with NEMA 5-15 Plug and Receptacle LOW: Cut In: 110V, Cut Out: 115V, Volts Boost: 10 120V, 1 PH, 50/60Hz, 20 Amp Receptacle 8961 HIGH: Cut In: 125V, Cut Out: 120V, Volts Buck: 10Maximum Load: 15 AmpsComplete with NEMA 5-20 Plug and Receptacle220V, 1 PH, 50/60 Hz, 15 Amp Receptacle 8962 LOW: Cut In: 210V, Cut Out: 220V, Volts Boost: 18 Maximum Load: 12 Amps HIGH: Cut In: 235V, Cut Out: 230V, Volts Buck: 18Complete with NEMA 6-15 Plug and Receptacle

CERTIFICATION

UL Listing and CSA Certification attests to the electrical and mechanical safety of all blood bank equipment. Unit construction and elec-trical characteristics are evaluated by engineers from the Underwriters Laboratories and the Canadian Standards Association. Selectedblood bank models also bear the CE Mark for compliance with the Electromagnetic Compatibility Directive.

Many tests are performed on Forma's blood bank units before they can carry the UL and CSA markings. Demand UL Listed and CSACertified blood bank products for your laboratory. Copies of UL, CSA and EMC documentation are available from Forma Scientific forcompliance verification.

ISO 9001

As a global leader in innovation, Forma Scientific continues its pursuit of excellence by maintaining ISO 9001-1994 registration. Thisadvanced quality system reinforces Forma's commitment to providing the highest quality products available.

The ISO (International Standards Organization) 9000 series is a set of international requirements for establishing and maintaining quality management systems. It is based on the widely recognized quality assurance standards of the British StandardsInstitute (BSI). Continued compliance is assured by ongoing internal audits and periodic audits by the assessing agency.

27

Page 28: Bloodbank Equipment

• Cell Culture Incubators• Ultra-Low Temperature Freezers• Laminar Airflow and

Biological Safety Cabinets• Cryopreservation Equipment• Blood Bank Equipment

• Orbital Shakers• Programmable Centrifuges• Lab Washers and Dryers• Lab Refrigerators and Freezers• Environmental Rooms• Custom-Designed Equipment

Box 649, Marietta, OH 457501-614-373-4763 • Fax: 1-614-373-6770

USA and Canada 1-800-848-3080International Distributors Worldwide

An ISO 9001 CompanyRev.010/95


Recommended