BMC Medical Co., Ltd. Nov. 2011 Training Course
Transcript

BMC Medical Co., Ltd.

Nov. 2011

Training Course

2

Table of Contents

1.CATALOGUE

2.INSTALL& DISASSEMBLE

3.SET UP

3

1. CATAOLGUE

4

RESmart BPAP (bilevel)

5

2. INSTALL & DISASSEMBLE

Install

Heater Plate

Indicator Light

Outlet Port

Water Chamber

Fill Line

Install Water tank

Component : Water tank, Humidifier, Device.

1

2

3

Disconnect the tubing from the water chamber. Press down

on the water chamber and slide it out of the humidifier platform.

Open the silica gel cover of water chamber and discard any remaining water.

Fill a solution of warm water and a mild dishwashing detergent into the chamber, cover the silica gel cover, rock the chamber a few minutes, then outflow the solution. Rinse the chamber several times with clean water and allow to air dry.

Fill the water chamber and close the silica gel cover. Inspect the water chamber for any leaks or damage.

Replace the water chamber if any damage is present.

•Clearing Water tank

DISASSEMBLE

Open the cover of the filter carefully. And then take it out.

DISASSEMBLE

Press the button at the bottom of the main machine, then pull the heated humidifier.

SET UP

3. SET UP

A. USER SET UP

B. MAINTENANCE SET UP

SET UP

Control Button

BPAP SETUP

Verify/Change the RESmart USER settings. To start the setup menu, press and hold the Ramp button until the setup menu appear (about 5 seconds) when system standby.

IMPORTANT! Press the Ramp Button to enter change status and press +/- User Buttons to change the setting. Press Ramp Button again to confirm the setting.

Press +/- User Buttons to access the next setting.

BPAP SETUP

Setting the Humidifier When the optional InH2™ Heated Humidifier is used, this setting can change the temperature and humidity of airflow between level 0 to 5. Default setting is 3.

BPAP SETUP

Setting the Altitude The altitude setting can be changed between level 0 to 2. Default setting is 0.

0 = less than 2,460 ft. (<750 m)

1 = 2,460 to 4,921 ft. (750 m to 1500 m)

2 = 4,924 to 8,202 ft. (1501 m to 2500 m)

*over 8,202 ft. = The airflow pressure may not be accurate. Contact your home care provider to have your pressure setting adjusted.

BPAP SETUP

Setting Auto ON When the RESmart BPAP is standby and wearing mask, your deep breath will start airflow automatically. Default setting is Enable.

BPAP SETUP

Setting Auto Off When the mask is removed, the RESmart BPAP will stop airflow automatically. Default setting is Enable.

Note: We advice you can set Auto Off is Disable. Most machine may appear this kind of problem, because apnea and mask off  is not to be judged at pressure lower than 7 hpa

BPAP SETUP

Setting the LCD Light This setting allows you to adjust the LCD light among Auto, On and Off. Default setting is Auto.

Auto: LCD light will be turned on when any button pressing and turned off after few seconds.

On: LCD light will be turned ON always.

Off: LCD light will be turned OFF always.

BPAP SETUP

Setting the Alert

This setting allows you to turn on/off the patient disconnect alert. When a large, continuous air leak (such as mask removal) has been detected in the circuit, this setting enables/disables the audible alert (a beeping sound). This setting is only effective when the Auto Off setting is disabled. (This function is NOT available on certain model.) Default setting is Disable.

BPAP SETUP

Setting the Delay Off When the optional InH2™ Heated Humidifier is used, this setting allows the air flow continue for about 15 minutes at very low pressure (about 2hPa) after RESmart Device is turned off. This will blow off the left vapor and protect the RESmart Device. Default setting is Enable.

BPAP SETUP

Setting the Sensitive(Only 25A) The Sensitive Setting is only available on Auto S mode. This setting can be set from 1(At level 1, the treatment pressure will increase slowly) to 5(More faster than level 1). Default setting is level 3.

BPAP SETUP

Setting the RiseTime

Rise time settings are adjustable by the patient and the provider from 0 (off) to 3. Rise time varies from 150 msec to 400msec dependent on patient setting and the pressure differential from IPAP to EPAP. The rise time setting should be adjusted according to the patient comfort level. Default setting is 2 (disable).

BPAP SETUP

Setting the Date The current date is displayed and can be modified if needed.

Note: Check and confirm this Date Setting frequently to ensure the usage blog is correct.

BPAP SETUP

Use Days The number of nights the RESmart was used for therapy for more than 4 hours will appear. This screen is only for reference. Your home care provider may periodically ask you for this information.

BPAP SETUP

I sense This function will adjust trigger’s sensitivity of

inspiration. The setting rang is from 1 to 8. 8 is highest sensitivity.(1.83 vision later)

BPAP SETUP

E sense This function will adjust trigger’s sensitivity of

expiration. The setting rang is from 1 to 8. 8 is highest sensitivity.(1.83 vision later)

BPAP SETUP

BPAP SETUP

Exit the Settings The settings are complete. Press the +/- user button to access the SAVE setting and Ramp Button to save all changes and exit the settings menu.

BPAP SETUP

BPAP MAINTENANCE SET UP

.ENTER MAINTENANCE MENU

BPAP SETUP

Entering the Menu:

Power off the machine

Press and hold the Pressure Start/Stop Button and Ramp Button while connecting the power cord into the AC inlet. After several seconds the Maintenance Menu will appear on display screen as below.

Mode CPAP

BPAP SETUP

IMPORTANT! Press the Ramp Button to enter change status and press +/- User Buttons to change the setting. Press Ramp Button again to confirm the setting.

BPAP SETUP

1.The mode-selection menu.

BPAP 25: “CPAP”, “S Mode”

BPAP 25A: “CPAP”, “S Mode”, “Auto S”

2. Navigating the Menu Press +/- User Buttons to access the next setting

as below

BPAP SETUP

CPAP mode:

Treat PManual PRampInit PUnitUse TimeSave

S mode:

EPAPIPAPManual PRampInitEPAP UnitUse TimeSave

Auto S mode:

EPAPIPAP Max IPAPRampInit EPAP

UnitUse TimeSave

BPAP SETUP

Treat P(Only CPAP Mode): The prescription pressure after ramp time. Setting range is 4-20hpa. Adjustment step is 0.5hpa. Default setting is 8hpa.

BPAP SETUP

EPAP: (for S mode and Auto S mode)

Expiratory Positive Airway Pressure . This pressure is set as the prescription pressure or less. Setting range is 4-25hPa. Adjustment step is 0.5hPa. Default setting is 6hPa. EPAP data modification is related with IPAP data as well. When EPAP is revised, IPAP will be changed too, the pressure difference value will keep the same.

BPAP SETUP

IPAP: (for S mode and Auto S mode)

Inspiratory Positive Airway Pressure . This pressure is set as the prescription pressure or less. Setting range is 4-25hPa. Adjustment step is 0.5hPa. Default setting is 8hPa. IPAP can be revised separately, when you revise IPAP, EPAP will keep the same as before.

BPAP SETUP

Max IPAP:

This setting is the maximal pressure which IPAP can reach. Default setting is 20hPa.

BPAP SETUP

Manual Pressure:

This setting only for CPAP&S mode. This setting allows user to adjust Treatment Pressure by oneself. Setting range is 0-2hpa/cm H2O. Adjustment step is 0.5hpa/cm H2O. Default setting is 0.

For example: If this setting is 1hpa/cm H2O and Treatment Pressure setting is 8hpa/cm H2O, user may change the treatment pressure between 7-9hpa/cm H2O by pressing +/- User Button.

BPAP SETUP

Ramp :

This setting is the ramp timer from starting pressure to prescription pressure. The device gives airflow at the relative lower pressure of Init P from start point of Ramp time, and increases the output pressure to the prescribed pressure in the last 5 minutes during Ramp. Setting range is 0-60mins with the adjustment step of 5mins. Default setting is 10mins. .

BPAP SETUP

Ramp Starting Pressure:

The starting pressure after blower is turned on. Setting range is 4hPa to BI-LEVEL Treatment Pressure. Adjustment step is 0.5hPa. Default setting is 4hPa. This feature will reduce the pressure and then gradually increase (ramp) the pressure to the prescription pressure setting so you can fall asleep more comfortably.

BPAP SETUP

Unit:

The unit of the output pressure, selectable between hpa and cmH2O.

BAPA SETUP

Use Time: Total hours this RESmart has been used will appear. This number is only for reference.

BPAP SETUP

Save and Exit: The settings are complete. Press the +/- user button to access the SAVE setting and Ramp Button to save all changes and exit the Maintenance Menu.

IMPORTANT! To make all changes active, press the +/- User Button to access the SAVE setting and Ramp Button to save all changes and exit the Maintenance Menu.

Hide function

The device can be read using information(AHI, SNI, Compliance, P95…) of last night, when you press Ramp button then press + or – button at the device standby mode.(1.79 version later and Only for BPAP 25A)