Date post: | 23-Dec-2015 |
Category: |
Documents |
Upload: | anthony-byrd |
View: | 223 times |
Download: | 0 times |
CH 908: Mass SpectrometryLecture 5
Ionization sources, EI, MALDI, and ESI
Prof. Peter B. O’Connor
Objectives for this lecture
• Ionizing molecules
• The many attempts to ionize biomolecules– Direct Chemical Ionization– Plasma desorption– FAB– LDI
• ESI and MALDI
MS Block Diagram
Ion Source
Mass Separator
Detector
Chromatography
Inlet
Sample Cleanup
Computer
DataReduction
Bioinformatics
FragmentationMethod
Mass Spectrometers do not measure mass, they measure mass/charge ratio.
A simple, laser desorption ion source.
Sample
Laser
Ionization• Photons
• Electrons
• Fast Neutral Molecules
• Other Ions.
• Heat
• Combination methods
Because most molecules prefer to be electrically neutral, ionization requires addition or removal of one or more charges particles (usually electrons or protons). This can be done by hitting the sample with one of the following.
Exceptions: samples that naturally prefer to be ions or are somehow stabilized as ions.
EI Mass Spectrum of an acetylated and reduced peptide
Field ionization/ field desorption MS Field ionization/ field desorption MS (FI/FDMS) (FI/FDMS) H. D. Beckey, H. D. Beckey, IJMSIPIJMSIP, 1969, 1969
H. R. Schulten, Int. J. Mass Spectrom. Ion Phys., 32, 97-283, 1979
Emitters for field ionization/ field Emitters for field ionization/ field desorption MS (FI/FDMS)desorption MS (FI/FDMS)
C. E. Costello, electron microscopy courtesy of JEOL , 1979
Field desorption MS of bradykininField desorption MS of bradykinin
S. Asante-Poku, W. G. Wood, and D. E. Schmidt, Jr. Biomed. Mass Spectrom, 2, 121, 1975
Field desorption MS of Vitamin BField desorption MS of Vitamin B1212
H.-R. Schulten and H. M. Schiebel Naturwissenschaften, 65, 223, 1978
Methods for “soft” ionizationMethods for “soft” ionization
• Direct chemical ionization MS (DCIMS)• Field ionization/ field desorption MS (FI/FDMS)• Secondary ionization MS (SIMS)• 252Cf Plasma desorption (PDMS)• Fast atom bombardment (FABMS)• Liquid secondary ionization MS(LSIMS)• Laser desorption MS (LDMS)• Electrospray ionization MS (ESIMS)Electrospray ionization MS (ESIMS)• Matrix-assisted laser desorption/ionization MS Matrix-assisted laser desorption/ionization MS
(MALDI MS)(MALDI MS)
[Direct] chemical ionization MS (DCIMS)[Direct] chemical ionization MS (DCIMS)
Frank H. Field and Burnaby MunsonASMS Distinguished Contribution, 1996
252Cf Plasma desorption (PDMS)
ASMS Distinguished Contribution, 1990
Ronald D. Macfarlane
R. D. Macfarlane, Anal. Chem., 55, 1247A-1264A, 1983
252Cf Plasma desorption (PDMS) of porcine trypsin (23,463 Da)
Peter Roepstorff
B. Sundquist, P. Roepstorff, et al., Science, 226, 696 – 698, 1984
Per Håkansson
Michael BarberASMS Distinguished Contribution, 1991
Fast atom bombardment (FABMS)Fast atom bombardment (FABMS)
M. Barber, R. S. Bordoli, G. J. Elliott, R. D. Sedgwick and A. N. Tyler, Anal. Chem., 54, 645A-657A, 1982
Fast atom bombardment mass spectra of (a) Fast atom bombardment mass spectra of (a) vitamin Bvitamin B1212 and (b) the coenzyme of vitamin B and (b) the coenzyme of vitamin B1212
M. Barber, R. S. Bordoli, R. D. Sedgwick and A. N. Tyler, Nature, 293, 270-275, 1981
(a) (b)
Fast atom bombardment mass spectra: Fast atom bombardment mass spectra: (a) human insulin and (b) proinsulin(a) human insulin and (b) proinsulin
M. Barber, R. S. Bordoli, G. J. Elliott, N. J. Horoch and B. N. Green, Biochem. Biophys. Res. Commun, 110, 753-757, 1983
A. Dell and H. R. Morris, Biochem. Biophys. Res. Commun, 106, 1456-1462, 1982
(a) (b)
Laser
Sample
Target
Figure 2. Laser Desorption plus Electron Impact Ionization
+
+
+
+ +
+
cathode
+ -Anode
e- e-e-
Two Special Cases• Matrix Assisted Laser Desorption/Ionization (MALDI)1,2
• Analyte is mixed with a "Matrix" and dried down on a surface• Laser hits the spot and desorbs matrix and analyte together• some of the analyte ends up charged
• Electrospray Ionization (ESI)3,4
•Solution containing the sample is sprayed out a needle into a large electric field. Droplets are formed and are sucked down into the mass spectrometer. They evaporate and leave charges on the analytes.
1. Hillenkamp, F.; Karas, M.; Beavis, R. C.; Chait, B. T., Matrix-Assisted Laser Desorption/Ionization Mass Spectrometry of Biopolymers Anal. Chem. 1991, 63, 1193A-1203A.
2. Karas, M. I.; Bachmann, D.; Bahr, U.; Hillenkamp, F., Matrix-Assisted Ultraviolet Laser Desorption of Non-Volatile Compounds Int. J. Mass Spectrom. Ion Processes 1987, 78, 53-68.
3. Fenn, J. B.; Mann, M.; Meng, C. K.; Wong, S. F.; Whitehouse, C. M., Electrospray ionization for mass spectrometry of large biomolecules Science 1989, 246, 64-71.
4. Whitehouse, C. M.; Dreyer, R. N.; Yamashita, M.; Fenn, J. B., Electrospray interface for LCMS Anal. Chem. 1985, 57, 675-679.
ESI and MALDI
Prof. John Fenn
Developed Electrospray
Prof. Dr. Ing. Franz HillenkampProf. Dr. Michael Karas
Developed MALDI
Nobel prize, 2002
+++
++
+ ++ + + + + ++ + +
+ + + ++ +++++
+ + +++
Robert A. Millikan (Nobel prize in physics, 1923)
Oil Drop Experiment
A simple ESI ion source.
Electrospray Ionization
• Electrospray Ionization (ESI)•Solution containing the sample is sprayed out a needle into a large electric field. Droplets are formed and are sucked down into the mass spectrometer. They evaporate and leave charges on the analytes. •Sample is a flowing liquid - compatible with LC•flow rates typically in the <1 μl/min range•sample cleanup is required!•sensitivity (Best case <1 amol, normal case 10-1000 fmol)•Multiple charging•upper mass range not really limited, but big masses get confusing if you don't have sufficient mass resolution•extremely soft ionization method - allows ionization of non-covalently bound species (e.g. protein complexes)
Generic Atmospheric Pressure Ionization Source/Mass Spectrometer System
Vacuum system with multiple stages:1st stage - free-jet expansion, molecular-beam stage2nd stage - ion optics3rd stage - mass analyzer
ElectrosprayMajor Components and Processes in ES-MS
ElectrosprayEmitter
AtmosphericPressure Region
(The Gap)
Atmosphereto VacuumInterface
MassAnalyzer
Žshrinkage/subdivisonof droplets
Ž gas-phase ionproduction
Ž gas-phasereactions
- ionization- charge permutation
Ž finaldesolvation
Ž fragmentation
Ž chargepermutation
Ž production ofcharged dropletsat emitter tip
Taylor cone
Spray ~1 microliter/min
In “nanospray”, flow rates of ~1 nl/min are used. The taylor cone and plume become invisible because the droplets are in the 100 nm diameter range, and sensitivity goes up due to greatly reduced space charge and improved capture efficiency.
A few examples of electrospray mass spectra…
~80 Da difference~25 Da wide
Bad mass spectrometer?
Peroxynitrite
MI=21284.4032 (21284.5511)
1000 1200 1400 1600 1800Mass/Charge (m/z)
16+
17+
18+
15+
14+
MI=21284.4080 (21284.5511)
1185 1230Mass/Charge (m/z)
1000 1200 1400 1600 1800Mass/Charge (m/z)
21291 21303 21291 21307
21+20+
800 900 1000 1100 1200 1300 1400 1500Mass/Charge (m/z)
22+23+24+
25+
19+ 18+
17+
16+
MI=21282.3880 (21282.53543)
21291 21303
21290 21304
MI=21282.3422 (21282.53543)
p21RasNormal distribution
DTT treated distribution
Zhao, C.; Sethuraman, M.; Clavreul, N.; Kaur, P.; Cohen, R. A.; O'Connor, P. B. A Detailed Map of Oxidative Post-translational Modifications of Human p21ras using Fourier Transform Mass Spectrometry Anal. Chem. 2006, 78, 5134-5142.
Calmodulin (Entire protein)
Micro-TOF
884.7075
933.7842
988.6485
1050.3689
1120.3351
1200.2786
1292.5220
1400.1262
1527.3098
1680.0067
+MS, 0.3-1.5min #(16-91)
0.0
0.5
1.0
1.5
4x10Intens.
800 900 1000 1100 1200 1300 1400 1500 1600 1700 m/z
Carbonic anhydrase (Entire protein Micro-TOF
429.2474
543.4134593.1336
659.8101
880.5196
1075.9647
1117.3119
1210.3295
1262.9047
1320.2672
1383.1014
1452.1785
1528.59651613.4771
1747.1915
908.0058
937.2675
968.4701
1001.8412
1037.5745
1161.9558
830.2614
807.2280
785.4459
764.7936
714.6845
854.6555
+MS, 0.4-1.2min #(23-70)
0
1000
2000
3000
4000
5000
6000
Intens.
400 600 800 1000 1200 1400 1600 1800 m/z
819.8634
956.4497
1147.5278
1433.7548
+MS, 0.0-0.3min #(3-16)
0
2
4
6
4x10Intens.
600 800 1000 1200 1400 1600 1800 m/z
Insulin (Entire protein)
Micro-TOF
659.8
714.7
779.5
857.4
952.6
1071.5
+MS, 0.1-1.3min #(7-79)
0.0
0.2
0.4
0.6
0.8
1.0
5x10Intens.
500 600 700 800 900 1000 1100 1200 1300 1400 m/z
Ubiquitin (Entire protein)
Micro-TOF
MALDI
• Matrix Assisted Laser Desorption/Ionization (MALDI)• Analyte is mixed with a "Matrix" and dried down on a surface• Laser hits the spot and desorbs matrix and analyte together• some of the analyte ends up charged "Lucky Survivor Model"• pulsed beam - nicely couples with pulsed mass analyzers• sensitivity (best case ~1 amol, normal case 10-1000 fmol)• more tolerant of salts than ESI, less cleanup required• upper mass limit defined by
1) metastable decay (PSD)2) detector
• relatively soft, but generally not soft enough for non-covalent complexes
MALDI mass spectrometry
Laser
++
+
Most commonly, Laser is a <10 nsec pulse at 337 nm (N2 laser) or 355 (frequency tripled Nd:YAG). 50 nsec pulse at 2.94 (Er:YAG) is also used.
ESI and MALDI
Prof. John Fenn
Developed Electrospray
Prof. Dr. Ing. Franz HillenkampProf. Dr. Michael Karas
Developed MALDI
Nobel prize, 2002Nobel prize, 2002
193 nm 200 nm 300 nm 400 nm 800 nm 1 µm 3µm 10µm wavelength
6.3 eV 6,1 eV 4 eV 3 eV 1.5 eV 0.4 eV 0.01 eV photon energy h
ultraviolet visible infrared
Excimer lasers ND:YAG laser OPO-lasers
ArF KrF XeCl ER:YAG laser CO2 laser 193 nm 248 nm 308 nm 2.94 µm
4xf 3xf Nitrogen laser 1xf 266 nm 353 nm 337 nm 1.06 µm 9.1-10.6 µm
Figure 1. Wavelength and photon energy of frequently used desorption lasers
2D Sample stage.
Figure 3. Laser Microprobe for Mass Analysis (LaMMA) configuration
Illumination
Laser
Camera
Ocular system
Ion Optics
TOF Analyzer
1
2
f1 f2 f3
Laser
Targetd1
d2
Figure 2. Laser focusing optics for adjustment of the spot size on the target
Infinite Conjugation
A.
B.
Figure 4. Two of the first MALDI spectra. A) the matrix effect, desorbing Alanine mixed with Tryptophan allowed observation of the Alanine peaks at 10x lower laser fluence than possible with pure Alanine. B) the first MALDI of a protein, showing that desorption/ionization could occur with no observable fragmentation.
Karas, M.; Bachmann, D.; Hillenkamp, F. Influence of the wavelength in high-irradiance ultraviolet laser desorption mass spectrometry of organic molecules Anal Chem 1985, 57, 2935-2939.
Karas, M. I.; Bachmann, D.; Bahr, U.; Hillenkamp, F. Matrix-assisted ultraviolet laser desorption of non-volatile compounds Int J Mass Spectrom Ion Processes 1987, 78, 53-68.
Peptide digests
Laser
Temperature
Sample
Target
Figure 1. Laser Desorption/Ionization
+
+
+
+ +
+
PEG (entire compound)
MALDI
MALDI of a 0.6 kDa Polymer
MALDI of a 1.6 kDa Polymer1581.914
1405.8121758.019
1361.785 1802.049
1846.0741317.760
1890.0981273.735
1934.1241229.709
1978.1431185.686
2022.1691141.659
2066.1871097.626 2110.212
1053.5862198.245
0
2000
4000
6000
8000Inte
ns.
[a.u
.]
1000 1200 1400 1600 1800 2000 2200m/z
MALDI of a 13 kDa polymer
0
250
500
750
1000
1250
Inte
ns.
[a.u
.]
10000 11000 12000 13000 14000 15000 16000m/z
Calmodulin (Entire protein)
MALDI
16794.350
0
500
1000
1500
2000
Inte
ns.
[a.u
.]
10000 12000 14000 16000 18000 20000 22000 24000 26000
m/z
Other laser wavelengths
In-gel digest experiment
In-gel digest experiment
Self Assessment
1. list 3 methods that people used to try to ionize proteins before 1985.
2. What two methods were discovered in 1985-1987 that revolutionized the study of biomolecules via mass spectrometry? How do they work?
3. What’s a good matrix for proteins/peptides in MALDI? What’s the structure? What laser wavelengths are used for MALDI?
4. How do you determine the mass of a protein from an ESI mass spectrum? 2 methods (at least).
Fini…
CH908: Mass spectrometryLecture 5 – Ionization sources.
Typical MALDI-TOF mass spectra of small proteins
First MALDI-TOF mass spectrum of a glycoprotein.
ElectronsThe oldest, most controllable method of ionizing most species is to smash electrons into them.
A few important ionization techniques using this method• Electron Impact
• >10 eV electrons • Gas phase samples only• Typically used for GC/MS• Deposits lots of energy into the molecule - causing extensive fragmentation
• Electron capture ionization •< 5 eV electrons get captured by the molecule• relatively gentle, but only works for some species with high electron affinity (like PCB's)
PhotonsUsing light is the next most reliable method. Some commonly used techniques are:
• Photoionization•Uses high energy photons (UV typically) to knock an electron off the molecule.• Very wavelength dependent - every molecule has particular wavelengths for ionization• Resonance Enhanced Multiphoton Ionization (REMPI)• Only used for gas phase samples• Very controllable energy deposition
• Laser Desorption/Ionization (LD/I)• Absorbs laser energy into electronic and vibrational modes of the molecule blasting it off of a surface (typically steel)• Not very controllable in energy deposition• Large suppression effects• Very small fraction of the desorbed molecules are ionized.
C60 IR-LDI
26-FEB-200112:50:46.00# Scans : 1# Zero Fills : 2
ScaleFactor : 25.64
0
10
20
30
40
50
60
70
80
90
100
110
500 750 1000 1250 1500 1750 2000
Buda v1.2 C:\IONSPEC\FTDATA\PO010226.002MALDI : C60 spectrum, IR laser
Inte
nsity
(ar
bitr
ary
units
)
Mass/Charge (m/z)
26-FEB-200113:43:30.03# Scans : 1# Zero Fills : 2
ScaleFactor : 37.55
0
10
20
30
40
50
60
70
80
90
100
110
500 750 1000 1250 1500 1750 2000
Buda v1.2 C:\IONSPEC\FTDATA\PO010226.004MALDI : C60 spectrum, IR laser, adjusting timing
Inte
nsity
(ar
bitr
ary
units
)
Mass/Charge (m/z)
26-FEB-200115:15:42.13# Scans : 1# Zero Fills : 2
ScaleFactor : 1.39
0
10
20
30
40
50
60
70
80
90
100
110
500 750 1000 1250 1500 1750 2000
Buda v1.2 C:\IONSPEC\FTDATA\PO010226.009MALDI : ir laser, c60
Inte
nsity
(ar
bitr
ary
units
)
Mass/Charge (m/z)
26-FEB-200115:15:42.13# Scans : 1# Zero Fills : 2
ScaleFactor : 1.39
0
10
20
30
40
50
60
70
80
90
100
110
718 719 720 721 722 723 724 725
Buda v1.2 C:\IONSPEC\FTDATA\PO010226.009MALDI : ir laser, c60
Inte
nsity
(ar
bitr
ary
units
)
Mass/Charge (m/z)
Neutral Molecules
High velocity neutral molecules can be used to generate ions.
• Fast Atom Bombarment (FAB) • Cesium ion gun (with deflector)• Used heavily in the early 80's• high energy deposition, very good for lipids
Other Ions
• Secondary Ion Mass Spectrometry (SIMS)• Cesium ion gun (with no deflector)• Not generally used for liquid/gaseous samples• Commonly used for imaging of solids• Deposits lots of energy into the molecule - causing extensive fragmentation
• Massive Cluster Bombardment• clusters (droplets) of glycerol ions• lower energy deposition• not used much, mostly in fundamentals work
Heat
• Thermal Desorption•Solid samples placed directly on a surface that is rapidly ramped up in temperature until the samples leap off the surface.•Mostly an experimental technique that got dropped with the advent of ESI/MALDI methods.
• Pyrolysis•determining species desorbed as a function of temperature.
Combined Methods
• Chemical Ionization (CI)•EI is used to ionize a background gas, and the background gas then undergoes a charge-transfer reaction to the analyte. •e.g. CH4 CH4
+· CH5+ + CH3
· A+H+ + CH4
• "somekind of" Desorption with "somekind of" postionization• for example, laser desorption with EI post ionization. • these are mostly fundamental methods used in physical chemistry experiments.
• ICP-MS, plasma desorption,
Electrospray is a soft ionization method.
Protein folding and protein complexes are accessible
Myoglobin natural theoretical isotope distribution
17,07017,06017,05017,040
100
90
80
70
60
50
40
30
20
10
0
Monoisotopic peak
C769 H1215 N209 O221 S4
13C = 1.1%2H=0.015%15N = 0.36%17O=0.04%18O=0.2%33S=0.76%34S=4.2%
ESI-FTMS of a tryptic digest
08-AUG-200014:22:33.98# Scans : 20# Zero Fills : 1
ScaleFactor : 9.66
635
8291137
15561332
2274
0
10
20
30
40
50
60
70
80
90
100
110
500 750 1000 1250 1500 1750 2000 2250 2500
Buda v1.2 C:\IONSPEC\FTDATA\PO000808.002Electrospray : aldolase 107 eap digested with trypsin
Inte
nsi
ty (
arb
itra
ry u
nits
)
Mass/Charge (m/z)
670 680 690 700
ESI-FTMS of a tryptic digest can give extremely high mass accuracy