ECEg535/EE507:-Instrumentation Eng’g 1
By Sintayehu Challa
CHAPTER 5: OUTPUT PRESENTATION UNIT
5.1 INTRODUCTION
All analogue electrical and electronic instruments can be broadly classified into two categories,
namely:
(1) Instruments with pointer movements, and
(2) Instruments with graphical displays.
All analogue electrical and electronic meters belong to the first category. Included in this very
wide group category are the basic electromechanical instruments used as panel board or
industrial instruments, and the portable electronic instruments. There are four types of these
meters built around the following principles of operations:
1. The permanent magnet moving coil (PMMC) movement mechanism, which responds
to average or direct current only;
2. The moving iron vane (MIV), fixed coil movement mechanism, which responds to d-c
or a-c currents.
3. The moving magnet fixed coil (MMFC) movement, which again responds to direct
current only; and
4. The electrodynamometer (EDM) movement with one fixed coil and one moving coil,
which again responds to direct or alternating currents.
We shall next examine briefly the principle of operation of each of the above type of
instruments by emphasising the input-output response relationships as discussed in the
generalised static and dynamic performance characteristics of instruments. Particular attention
will be devoted to the PMMC as this meter movement is still the heart of many simple and
complex electronic instrumentation systems.
ECEg535/EE507:-Instrumentation Eng’g 2
By Sintayehu Challa
5.2 DYNAMICS OF POINTER METER MOVEMENTS
Electromechanical instruments have a number of common features in that they basically
involve pointer movements rotated against fixed scales. As illustrated in Fig. 5-1, the moving
elements of these instruments are mounted on rigid supports (called spindles) which rest
pivoted on jewel cups or bronze bearings. To minimise friction losses, the pivots and bearings
are polished as thoroughly as possible. The force giving rise to deflecting torques is typically
exerted on current carrying conductors, and in principle these can be determined using the laws
of electromagnetic dynamics. Thus starting from the instantaneous force acting on a charge of
q columb moving with a velocity V in an electric field of strength E and magnetic field of flux
density B, the basic force law yields
F = q (E + v x B) newtons
If E is neglected in comparison to the cross product vxB, we can set
F = q ( v x B )
Fig.5.1
ECEg535/EE507:-Instrumentation Eng’g 3
By Sintayehu Challa
We consider first the movements of the PMMC. If two parallel conductors (carrying equal but
oppositely direct currents) are placed in a magnetic field, then two force couples giving rise to
a torque will result as shown in Fig. 5-2, which represents the top view of a PMMC.
The deflecting torque will then be given by
Fdi =τ
where d is the distance between the lines of action of the two couple forces.
This deflecting torque will be opposed by three other torque acting on the moving assembly,
namely:
1. A restoring or control torque which will be directly proportional to the angular
displacement θ, and is also directly proportional to the spring constant K of the spiral
spring, and thus can be set equal to K;
2. An accelerating torque which is dependent on the moment of inertial J of the moving
assembly, and with the angular acceleration 2
2
dt
d θ
3. A damping torque, which is dependent on the angular, speeddt
Cdθ C being an effective
damping coefficient.
Fig. 5.2
ECEg535/EE507:-Instrumentation Eng’g 4
By Sintayehu Challa
Hence, the torque equation
∑ = 0τ
Can be written as
θθθ
τ
θθθ
τ
kdt
dC
dt
dJ
elyequivalentor
kdt
d
dt
dJ
i
i
++=
=−−−
2
2
2
2
0
The above equation represents the dynamic performance of a second order instruments.
5.3 THE PERMANENT MAGNET MOVING COIL (PMMC)
The permanent magnet moving coil (PMMC) movement mechanism, which responds to
average or direct current only; the dynamic performance of second order instruments of
PMMC is given by
θτ ki =
5.4 THE MOVING IRON VANE (MIV)
While the above procedure for obtaining the dynamic equation of the PMMC is in principle
valid for all electromechanical instruments, it could pose difficulties in determining the
magnetic forces, which give rise to the deflecting torque. An alternative procedure for
determining τi from energy relations is, however, found helpful for practically all
electromechanical and electrostatic pointer movements. The principle behind this method is
that torque can be related to rotational systems. Consider the top-view representation of a
repulsion type of moving iron vane movement shown in Fig. 5-2.
ECEg535/EE507:-Instrumentation Eng’g 5
By Sintayehu Challa
Let θ, i, di, L, and dL denote, respectively, the angular displacement of the moving vane, the
current applied to the fixed coil, a small increase in current, the self-inductance of the coil, and
a small variation in L. During the time that the moving vane carrying the pointer is being
deflected, an emf will be induced in the coil, which can be expressed by
dt
dLi
dt
diL
Lidt
de
+=
= )(
Which are due to changes in i and L, respectively. The electrical work done against this
opposing emf will be given by
dLidiLi
dteiWdone
2+=
= ∫
This work must be equal to the change of stored energy in the whole system. This comprises
of the change in potential energy of the spring which is Kθ, and the change in energy stored in
the inductance of the coil due to current and inductance changes such that
LidLLdiiWstored
22
2
1)()(
2
1=+++= τθ
Fig.5.3
ECEg535/EE507:-Instrumentation Eng’g 6
By Sintayehu Challa
Equating the above Equations, it can thus be shown that the deflecting torque will be
θθ
θθτ
d
dLiK
orWWd
d
d
dWdonestoredi
2
2
1
,)(
=
−==
Where K is the spring constant. which is generally valid for any rotating system. Simply
states that the deflection torque is proportional to the product of the square of the current and
the change in inductance due to the rotation of the moving vane from its equilibrium position.
5.5 THE ELECTRODYNAMOMETER (EDM)
The same procedure can be followed to describe the principle of operation of an EDM. The
basic construction of an electrodynamic instrument consists of a moving coil mounted the
magnetic field of two series-connected stator coils. Let il and i2 denote the current in the stator
and moving coil, respectively. The conditions relating the deflecting torque to the magnetising
currents il and i2 is very similar to the torque relationship applying to the moving-iron vane
instrument, but the important variable here is the mutual inductance M between the fixed and
moving coils, and the stored magnetic energy i1 i2 M. Starting from the expression for the
induced emf
dt
dMi
dt
diM
dt
dMi
dt
diM
eee i
22
11
2
+++=
+=
where the first and second terms resepresent the emfs induced in the moving coil, and the third
and fourth terms give the emfs induced in the fixed coil. By considering the change in stored
energy, we obtain
dMiiMdiiW istored 2112 ++= θτ
It can thus be shown that the deflecting torque for the EDM movement is given by
θτ
d
dM
K
iii
2
21=
ECEg535/EE507:-Instrumentation Eng’g 7
By Sintayehu Challa
Ignoring the transient parts of the solutions of the basic equations for the three-meter
movements, the steady state solutions (θss) for the three basic electromechanical meter
movements can then be summarised by the following relationships:
(i) For the PMMC,
KIxinputtConsC
iNBldss === tanθ
(ii) For the MIV,
22
2
)(tan2
KIinputxtConsC
ddli
SS ===θ
θ
(iii) for the EDM,
2121
21
)()(tan IKIinputxinputxtConsC
ddMii
SS ===θ
θ
where C is stiffness of the suspension spring.
5.6 ANALOG ELECTRONIC INSTRUMENTS
All analogue electronic instruments are built around the PMMC as the basic display
mechanism. Such instruments are capable of measuring a-c signals (i.e voltages, currents and
non-electrical quantities) which cannot be measured by other standard electrical instruments.
As shown in Fig. 5-5, the input signal qi (t) is fed to a signal conditioning circuit before it is
applied to a PMMC instrument which contains different scales for different quantities on a
single plate.
ECEg535/EE507:-Instrumentation Eng’g 8
By Sintayehu Challa
The signal conditioning circuit involves the use of a suitable range selector or attenuator as
well as an interconnection of active (i.e amplifying) and non-linear passive (i.e. rectifying)
devices. Depending on the electronic circuitry, an analogue electronic meter belongs to one of
the following instruments:
1. Average responding meter
2. RMS responding meter;
3. peak responding rectifier
4. peak-to-peak responding rectifier
5.6.1 AVERAGE RESPONDING METER
The deflection of the PMMC meter is proportional to the average value for this purpose the
instrument are used for dc responding either voltages or currents to safe values. This can be
realised in practice by using very large series of resistors (for voltage ), and very small shunt
paths (for current ).
Fig.5.4
ECEg535/EE507:-Instrumentation Eng’g 9
By Sintayehu Challa
5.6.2 RECTIFIER METERS
As shown in Fig. 5-5, rectifier meters can be built with the use of half-wave or full-wave
rectifying devices. In both circuits, currents can only flow in one direction. The deflection of
the PMMC meter is proportional to the average value of the rectified current (or voltage).
Assuming that the input current is sinusoidal, then the average current Iav will be
WAVEFULLFORII
WAVEHALFFORII
RMSAVE
RMSAVE
9.0
45.0
=
=
Fig.5.5
ECEg535/EE507:-Instrumentation Eng’g 10
By Sintayehu Challa
Typically, the meter scale can also be calibrated in terms of rms values of sinusoidal voltages
(or currents). However, the meter indications will be incorrect when non-sinusoidal voltages
are measured.
5.6.3 PEAK AND PEAK-TO-PEAK RESPONDING METERS
As shown in Figs 5-6, the peak-responding meter is built with the use of a capacitor and a
diode.
Fig.5.6
ECEg535/EE507:-Instrumentation Eng’g 11
By Sintayehu Challa
It can be seen that
Vo(t) = Vin + Vc
where Vin can be taken as a sinusoidal voltage with
Vin(t) = Vm sin ωt
and Vc = Vm is the voltage to which the capacitor C will be charged when the diode D is
conducting during the negative half of the input voltage. The primary difference between the
peak-responding meter and the average responding meter is the use of a storage capacitor with
the rectifying meter. The average voltage read by the meter is given by
m
T
oavO VdttVT
tV == ∫ )(1
)(0
Which is equal to the peak amplitude of Vin.
The peak-responding voltmeter is able to measure signals of frequencies up to hundreds of
megahertz. However for unsymmetrical waveforms, the reading will be in error. To overcome
this problem, a capacitor and a diode can be added as shown in Fig. 5-7 to obtain a peak-to-
peak response.
Fig.5.7
ECEg535/EE507:-Instrumentation Eng’g 12
By Sintayehu Challa
Fig.5.8
ECEg535/EE507:-Instrumentation Eng’g 13
By Sintayehu Challa
INSTRUMENTS WITH GRAPHICAL DISPLAYS.
5.7 GENERAL PURPOSE CATHODE RAY OSCILLOSCOPES
The oscilloscope, or more generally the cathode ray oscilloscope (CRO), is the most versatile
electronic measuring instrument. The CRO can mainly be regarded as a true analog instrument
in that it serves both as a voltmeter, and as an electrical-to-optical transducer showing a
waveform display of the signal under measurement. Still more importantly, the CRO is an
indispensable instrument for the measurements of frequency, time duration, and phase
differences at audio and higher frequencies The basic structures of a general-purpose cathode
ray tube (CRT) are illustrated in Fig. 5-9 and 5-10. It consists of five major parts, namely: (i)
power supply, (ii) vertical amplifier, (iii) triggering circuit, (iv) time base circuit, (v) gate
amplifier, (vi) horizontal amplifier, and (vii) the cathode ray tube(CRT). The power supply
unit indicated in Fig. 5-9 provides d-c voltages to the various electrodes inside the CRT, and
also to the amplifier and time-base circuits.
Laboratory oscilloscopes are classified in many ways. Usually, the distinctions are based
either on frequency-response capability or on CRT characteristics. Thus there are low-
frequency oscilloscopes (DC to 10 MHz for vertical amplifier response), high frequency
Oscilloscopes (sometimes capable of capturing and displaying single-shot phenomena of less
Fig.5.9
ECEg535/EE507:-Instrumentation Eng’g 14
By Sintayehu Challa
than 1-n sec rise time), and sampling oscilloscopes which reconstruct very high frequency
signals up to 18 GHz - repetitive waveforms on a dot-sample basis. In terms of the
characteristics of the display screen, there are standard refreshed phosphor oscilloscopes, and
storage oscilloscopes,
Depending upon the type of CRT used. With the foregoing general introduction, we shall next
try to examine in some detail important features and principles of operation of the CRO under
the following major topics:
- Beam Generation, Focusing and Deflection inside a CRO,
- Saw-tooth Sweep Waveform Generation and Waveform Display
- Large Input Signal Attenuation and Attenuators
- Bandwidth and Rise-Time features of CRO's
- Applications.
Without going too much into details, we shall thus attempt to understand firmly the underlying
principles and techniques which explain the basic operations and working conditions of
general-purpose CRO's in particular, and other similar devices.
Fig.5.10
ECEg535/EE507:-Instrumentation Eng’g 15
By Sintayehu Challa
5.8 BEAM GENERATION, FOCUSING AND DEFLECTION
Generation and focusing of an electron beam inside a CRT is illustrated in Fig. 5-11. An
electron gun assembly mounted at the base of the CRT produces the electron beam. In this
section of the tube, continuous beams of electrons emitted by a heated cathode subsequently
pass through a series of electrodes. Starting from the control grid, followed by the accelerating
grid, the intensity of the beam of the beam is adjusted to produce a very thin line that will
strike the CRT screen with sufficient energy. In a laboratory type CRO, the diffused beam is
first narrowed into a pencil-sharp ray using an electrostatic method of focusing as illustrated in
Fig. 5-11.
Fig.5.11
ECEg535/EE507:-Instrumentation Eng’g 16
By Sintayehu Challa
Fig 5.12 Electrostatic focusing in a CRT
Further, the focused beam spot striking the CRT screen can be deflected vertically and
horizontally by means of voltages applied to pairs of vertical and horizontal deflection plates
that are placed at right angles to each other, and the focused beam.
There are two methods employed in the deflections of electron beams in CRTs:
A) In laboratory oscilloscopes, the beam passes through two pairs of orthogonal mounted
deflection plates, which deflect the beam vertically up and downwards, and horizontally
sideways. These are called, respectively, vertical and horizontal deflection plates. Time
varying and direct current potentials are applied across each plate pair, and the deflection
method is called electrostatic deflection.
B) Another method of beam deflection, which is called magnetic deflection, uses cross-aligned
coils. This method of deflection is practically used in the CRTs found in television and word-
processing terminals.
The CROs used in measurement systems are built with electrostatic deflection mechanisms. As
illustrated in Fig. 5-13, the important parameters in the deflection mechanisms are
D = Separation or spacing of deflection plates at right angles to the beam;
Lp = Effective length of deflection plates int he direction of the focused electron beam;
S = Length along beam from center of deflecting plates to the CRT screen;
ECEg535/EE507:-Instrumentation Eng’g 17
By Sintayehu Challa
Va = Beam voltage or the potential to which the electron beam has been accelerated
when it enters the deflection region,
Vd = Deflecting voltage between plates,
D = final deflection distance on the CRT screen, either horizontally or vertically;
Z = Coordinate along central axis of the CRT
ECEg535/EE507:-Instrumentation Eng’g 18
By Sintayehu Challa
Fig. 5.13 Principle of electrostatic deflection in a CRT
Detailed analyses on the electrodynamics of the fields and accelerating voltages finally lead
))((a
dP
V
V
d
LSD =
From the above Equation, the deflection sensitivity is defined as the ratio of D to Vd in
millimeters per volt (or centimeters per volt)
De = D/Vd (millimeters/volt)
Which is independent of both the deflecting voltage Vd and the ratio e/m, but varies inversely
with the accelerate voltage Va. Inversely the deflection factor Va/D of a CRT
Is defined by
G = 1/De
Can thus be defined in volts/centimeter for standard measurements, or in volts/millimeter for
very sensitive deflections. Typical values of deflection sensitivities range from 1.0mm per volt
to 0.1 mm per volt, and with corresponding deflection factors ranging from 10 volts per
centimeter to 100 volts per centimeter.
5.9 SWEEP WAVEFORM GENERATION AND DISPLAY
To use an oscilloscope properly, it is first very essential to know when or why the various
selector switches are operated. The functions of the four important circuits (i.e-vertical input
circuits, horizontal amplifier circuits, triggering circuits, and sweep generator circuits) are
therefore discussed below.
ECEg535/EE507:-Instrumentation Eng’g 19
By Sintayehu Challa
5.9.1 VERTICAL INPUT CIRCUITS
A typical functional diagram of vertical input circuits is illustrated in Fig. 5-14
The AC/DC switch short-circuits the blocking capacitor when the oscilloscope is to be used to
display signals from d-c to an upper frequency limit determined by the bandwidth of the
vertical amplifier. A calibrated attenuator placed before the vertical amplifier controls the
vertical sensitivity or deflection factor in VOLTS/DIVISION.
Fig.5.14Block – diagram representation of vertical, input circuits
5.9.2 TIME BASE CIRCUIT AND WAVEFORM DISPLAY
The most frequent application of an oscilloscope is to display a changing signal y(t), as a
function of time. Thus oscilloscopes are provided with time base or sweep-generator circuit
which produce saw-tooth voltages as illustrated in Fig. 5-15.
The waveform illustrated rises linearly from "A" to "B" during the active sweep period T1,
during which time the electron beam moves across the CRT screen with constant horizontal
velocity. the slope of this waveform and hence the horizontal beam velocity is adjusted by
setting the time base to the desired TIME DIVISION. During the retrace period, T2, the
electron beam is returned to its starting point and the CRT is "blanked out" during this flyback
time. To obtain a stable display on the CRT it is necessary that the start of each sweep by
synchronized to the signal being displayed. In other words, the frequency of the signal (y-
input) must be equal to or be integer multiples of the sweep frequency (i.e. l/Tl). Otherwise a
stable display will not be obtained.
ECEg535/EE507:-Instrumentation Eng’g 20
By Sintayehu Challa
In conjunction with the sweep circuits, there is also a STABILITY control mechanism, which
controls the sensitivity of the sweep circuit to an input pulse. In TRIG (i.e. triggered)
operation, this control is set so that the sweep will not operate unless a triggering pulse is
received. The synchronized (SYNC) operation for stability control is adjusted such that
repetitive sweeps are generated. The function of the sweep circuit is shown in Fig.5-16. In
many respects, the horizontal amplifier circuits are similar to the vertical amplifier circuits.
However, the gain and bandwidth of the horizontal amplifier are usually lower than those of
the vertical amplifier. Just as the vertical amplifier is provided with different gain settings for
different input signal levels, the horizontal amplifier gain can also be adjusted to magnify or
expand the horizontal sweep display. This magnification is obtained by increasing the gain of
the amplifier by a variable setting or by a certain factor, which in effect increases the display
area width by the same factor.
Fig.5.16 Functional diagram of a sweep circuit
Fig. 5.15 Typical output of a time-base circuit
ECEg535/EE507:-Instrumentation Eng’g 21
By Sintayehu Challa
5.9.3 TRIGGERING CIRCUITS
These circuits enable the sweep generator circuits to start the sweep generation in coincidence
with selected time spots. Triggering signals can be selected from one of the following three
sources:
a) Internal source from the vertical amplifier;
b) External source from an externally applied signal;
c) Line supply, which is a line frequency of 50 or 60 HZ, and usually obtained
from a transformer internal to the oscilloscope unit.
These are shown abbreviated as INT, EXT, and LINE on the front panel of a CRO. The
internal source, whereby the triggering signal is derived from one of the vertical inputs
is most frequently used, while line triggering is used to display signals which contain
components at line frequency or multiples of line frequency.
When the two triggering sources are not satisfactory, external-triggering sources will be
required. In standard service or laboratory oscilloscopes, one frequently finds triggering
controls or MODE selector switches for +VE or -Ve slope TRIGGERING TRIG LEVEL,
AUTO, TV LINE, and TV FRAME. The selection of the triggering point is illustrated in Fig.
5-17.
ECEg535/EE507:-Instrumentation Eng’g 22
By Sintayehu Challa
Fig. 5.17 Selection of triggering level.
Consider the Y-input signal (i.e. vertical amplifier input) to be a triangular waveform. Then,
the triggering point can be chosen either from the + ve going or -ve going parts of the signal.
Additionally, the TRIG LEVEL required to obtain the pulse, which triggers the sawtooth
waveform generator can also, be adjusted. In the AUTO mode, the TRIG LEVEL is not used
and the circuit slope selector still operates as indicated above. The AUTO mode is usually
preferred as this minimizes the number of adjustments, which must be made to obtain a stable
display on the CRT screen.
The triggering modes TV LINE and TV FRAME are only found in oscilloscopes used
additionally for servicing television equipment, and they are used for synchronizing a
composite video waveform display with the horizontal or vertical synchronizing pulses,
respectively. It should be mentioned here that more complex triggering switches are also used
in advanced oscilloscopes.
ECEg535/EE507:-Instrumentation Eng’g 23
By Sintayehu Challa
5.9.4 SAW TOOTH VOLTAGE GENERATION
The sweep waveform illustrated in Fig. 5-17 is generated with the use of circuits, which can be
represented as simple RC circuits as shown in Fig. 5-18.
The switch SW represents an electronic switching circuit which can alternatively connect the
capacitor to position 1 for charging, and to position 2 for discharging through resistors R1 and
R2, respectively. If the electronic switch is running free, then the sweep also becomes free-
running. Typical output waveforms are presented in Figs. 5-19(a) and (b) for different duration
of sweep or scan times.
Actual saw-tooth waveforms can be described by
Vc = V(1-e-t/R1C
)
in which the voltage levels Vm, Vm1 and Vm2 are determined by the values of the capacitor C
for fixed values of R1 and R2. To obtain T2<<T1, then R2C<<R1C. For different sweep rates,
different capacitors are then connected to the switch SW, which is physically built from a fast
active electronic switch such as unijunction transistor or a combination of bistable
multivibrators.
Fig. 5.18 Equivalent circuit for a
saw-tooth waveform generator
ECEg535/EE507:-Instrumentation Eng’g 24
By Sintayehu Challa
5.9.5 CALIBRATED SWEEPS
To obtain reasonably accurate displays of applied waveforms, it is necessary to maintain a
precise relation between the electron beam and the sweep generation. This means that the
sweep voltage must be highly linear and adjustable between exactly known limits as illustrated
in Fig. 5-20. The one important use of such a calibrated sweep is for time calibrations during
measurements.
Very dependable oscilloscopes actually have two calibrated sweeps. One is produced by the
sawtooth voltage generator, and is called the main sweep. The second sweep is produced by a
similar but delayed circuit, thus giving a waveform, which is behind that of the main sweep.
The speed of the delayed waveform is usually arranged to be 2,5, and 10 times that of the main
Fig.5.20 Linearized sweep waveform needed for calibration
Fig.5.19
ECEg535/EE507:-Instrumentation Eng’g 25
By Sintayehu Challa
sweep. Consequently, two time axes are produced in advanced oscilloscopes, each using a
time scale of its own.
5.9.6 MULTIPLE TRACE DISPLAYS
In many CRO measurements, it is necessary to compare one waveform (signal) with another
waveform simultaneously. To facilitate such applications, dual (and multiple) trace displays
are obtained using one of the following three methods: a) dual beam oscilloscope, (b) alternate
mode, and (c) chopped mode. These are illustrated in Fig.5-21.
Fig 5.21 Examples of multiple traces : (a) dual beam oscilloscope; (b) dual traces with alternate mode;
(c) dual traces with chopped mode
The use of two electron guns complete with two independent sweep generators can produce an
instrument known as a dual beam oscilloscope. The same effect may also be produced by a
single electron gun, but with the output being split into two independent controllable electron
beams. Employing a single electron gun, a double trace can also be produced by switching the
Y deflection plates from one input signal to another in an alternate mode of operation
(Fig. 5-21(b). By building dot traces of two (or more) channel inputs as shown in Fig. 5-21(c),
the instrument can also be operated in a chopped mode. In such a mode, the Y deflection
plates are switched from one input signal amplifier to another at a rate, which is faster than the
standard sweep rate.
ECEg535/EE507:-Instrumentation Eng’g 26
By Sintayehu Challa
APPENDIX FOR CHAPTER 5
5.9.7 APPENDIX A 5.1 CHARACTERISTICS OF CRT SCREEN PHOSPHORS
Type
Trace Type
or
Fluorescence
PER-
SISTENCE
Relative
Writing
Speed
Application
P1
P2
P4
P7
P11
P31
P33
P39
Yellow-green
Blue-green
White
Blue
Blue-Violet
Yellow-green
Orange
Green
Medium
Medium
Medium to
Medium short
Medium
Medium
Medium short
Very long
Medium to
medium long
35%
70%
75%
95%
100%
75%
7%
4 0%
CROs, Radar
CROs
Black & white television
Radar, Medical
Photographic recording
General purpose
replacement for P1
Radar
Medical, graphics
ECEg535/EE507:-Instrumentation Eng’g 27
By Sintayehu Challa
Note:- 1. By Fluorescence of a phosphor is meant the light emission observed when an
electron beam hits the phosphor.
2. By Phosphorescence is meant the colour of the light left after the phosphor has
been stimulated into light emission and the excitation has been removed.
While the phosphorescences of most phosphor screens are the same as their
fluorescences types P2 and p7 have yellow-green phosphorescences.
3. By Persistence is meant an approximate measure of the time it takes for the
phosphorescence to decay to l.e of the excitation level during fluorescence.
For short-persistence, decay time is less than 1 msec. If the decay-time is
less than 2 sec., the phosphor is said to have medium-persistence.
Phosphors with decay times of the order of a minute or more are said to have
long-persistence.
4. Writing-speed is a measure of the fastest deflection rate of a single beam trace
(or single-short wave-form display) that is visible on a film when a
photograph is taken of a waveform display with P11 taken as a reference
phosphor.
ECEg535/EE507:-Instrumentation Eng’g 28
By Sintayehu Challa
DATA ACQUISITION SYSTEMS
In transducer instrumentation, data acquisition systems are used to measure and record
information for later study and analysis. The recorded data takes the form of signals with either
continuous or pulse waveforms. Accordingly, data acquisition systems are divided into analog
and digital recorders. The important analog data storing devices are: (1)X-Y recorders, (2)
graphical recorders with moving coil movements, and (3) analog magnetic tape recorders.
Digital recording, reproduction and processing of massive data, which is the standard feature
of computing systems, is commonly made with instrumentation magnetic tapes and disks,
5.9.8 X-Y RECORDERS
X-Y recordings are similar to CRT waveform displays, with the x-coordinate provided by a
ramp generator. The Y- input receives low-frequency analog signals. Accuracy and resolution,
functions of an X-Y recorder’s electronic and mechanical characteristics, determine the static
performance of such an instrument. The slew speed and acceleration in response to the capture
of rapid and transient signal input determine the dynamic response. Other important features,
which are also common to other graphical recorders, include chart size, number of pens, time
base capability, preamplifiers and filters.
5.9.9 GRAPHICAL RECORDERS
Graphical recorders measure variations of electrical or non-electrical quantities with respect to
time taking place over many seconds, minutes, hours, or days. In many instances, the recording
is needed to check the performance of industrial processes or electrical power generation and
distribution systems. Mostly, these recorders are built around a moving coil instrument where
Fig.5.22
ECEg535/EE507:-Instrumentation Eng’g 29
By Sintayehu Challa
the indicating pointer is replaced by a writing pen movement on a graduated paper moving at a
constant speed. As shown in Fig. 5-22, the tip of the pointer leaves marks on the paper thereby
forming a permanent record of the amplitude of the input signal with respect to time.
Maximum sensitivity is of the order of 4mV/cm within a narrow bandwidth of d-c to about
10Hz. Important features of the basic graphical recorders are: (1) input impedance, (2) time
scale, (3) event markers, and (4) writing mechanism. The use of amplifier ensures the input
impedance to the recorder is maintained relatively high. The writing mechanism can be an ink
pen with a capillary feed system, or heated stylus recording the variations of the input (Vin) on
a heat sensitive paper. Other recorders, commonly known as photographic or ultraviolet (UV)
recorders, use a light beam as a pointer leaving traces on photographic papers. Recorders
known as multi-channel recorders contain a number of writing pens in all making marks
simultaneously on a wide roll of paper thereby permitting easy comparison of several
simultaneous functions.
5.9.10 MAGNETIC TAPE RECORDERS
The principle of recording with instrumentation magnetic tapes is illustrated in Fig. 5-23.
There are three main components: (1) a core with a small nonmagnetic gap, (2) a coil wound
on the core, and (3) a thin magnetic coating sitting on a base. The latter can be a wire, but in
most application it consists of a flexible plastic tape.
Fig.5.23
ECEg535/EE507:-Instrumentation Eng’g 30
By Sintayehu Challa
The magnetic coating is a thin layer of iron oxide(Fe2°3) particles, and the core is made of
laminated steel alloys. The core assembly carrying the winding is called a tape recording head.
With current (1) flowing in the coil, a magnetic flux will bridge cross the non-magnetic gap,
thus magnetizing the iron oxide particles as they pass the gap. Because the magnetic coating is
purposely selected for its high remanence, the iron particles will remain magnetized in the
direction of tape travel with the magnitude of flux impressed upon the tape as it moves in front
of the non-magnetic gap. Hence, a recording of the applied signal will have been realized.
When the same tape is passed through the front gap of a similar playback head, it will cause
variations in the reluctance of the winding, and thereby inducing a voltage which ideally is
required to be a faithful reproduction of the recorded signal. A functional diagram of a
complete magnetic tape-recorder is shown .in Fig. 5-24. While two different heads are needed
for accurate instrumentation recording and reproduction, only one combined head is used for
coarse instrumentation works as well as for audio recordings. With one head, one needs only to
switch from a record to reproduce (playback) mode. At the same time the direction of tape
travel has accordingly to is set by the sense of rotation of the motor moving the tape transport.
Fig.5.24
ECEg535/EE507:-Instrumentation Eng’g 31
By Sintayehu Challa
Four factors contribute to the accuracy of magnetic tape recording, particularly for
instrumentation and data processing purposes. These are: (1) the magnetization characteristics
of the magnetic recording medium, (2) the tape speed, (3) the bandwidth of the recorded signal,
and (4) the gap width. All these factors are considered together to minimize errors and
distortions.
Fig.5.25
ECEg535/EE507:-Instrumentation Eng’g 32
By Sintayehu Challa
5.9.11 DIGITAL RECORDING
Magnetic tapes are also commonly used for digital recording of binary data encoded. in terms
of "1s" or "0s", both in instrumentation and data processing systems. Three broad methods of
digital recording can be briefly described as follows:
a) Return to zero (RZ) systems which employ one direction of saturation for" 1" ,
while magnetization to saturation in the opposite direction then represents "0" .
b) Non-return to zero (NRZ) systems which regard changes in the direction of
magnetization as "1", while no changes in magnetization are taken as "0".
c) Phase encoding (FE) systems in which one sense of magnetization is regarded
as" 1" , while the opposite sense is regarded as "0" .
For each method, the density .of recording is measured in terms of so many bits per tape
length. The unit often used in bits per inch (BPI), typical values being 8000 BPI, and 1600
BPI for NRZ and phase encoded systems, respectively. The recorded binaries represent
decimals or alphanumeric characters using one of the standard binary codes. Also, the record
and reproduce amplifiers are very simple, and the only requirements are maintenance of
accurate tape and head alignments. For computer application, recording is mostly made in a
block form, meaning that a block of information is recorded at one time, in place of
conventional or incremental recording where data is continuously stored on the tape. As the
recorded binaries are typically obtained by digitizing input analog signals, the latter can be
recovered by use of digital to analog converters. Alternative recording systems which are
gaining wide applications use the same principles as already described for magnetic recording,
but with the flexible tapes replaced by rigid disk packs, diskettes or floppy disks, all operated
with use of movable heads to scan magnetically coated surfaces. Storage capacities measured
in kilo (i.e. thousand), millions, and possibly
billions of bytes (i.e 8-bit words), and access times ranging from 35 to 100 ms are the two
important features of these recording devices. Data recorded on these devices is therefore
much more voluminous than those recorded in tapes, and can also be accessed much
ECEg535/EE507:-Instrumentation Eng’g 33
By Sintayehu Challa
faster. Further, magnetic tapes are also commonly used for very large digital data
recording and processing.