Confocal Microscopy and Atomic Force Microscopy (AFM) of biofilms of biofilms A very brief primer...
Transcript
Page 1: Confocal Microscopy and Atomic Force Microscopy (AFM) of … · 2014. 1. 15. · Disadvantages of Confocal Microscopy Requires fluorescent samples Uses laser illumination (expensive,

Confocal Microscopy andAtomic Force Microscopy (AFM)of biofilmsof biofilmsA very brief primer...

Page 2: Confocal Microscopy and Atomic Force Microscopy (AFM) of … · 2014. 1. 15. · Disadvantages of Confocal Microscopy Requires fluorescent samples Uses laser illumination (expensive,

Fundamentals of Confocal MicroscopyBased on a conventional fluorescence microscope

Fluorescent MicroscopeFluorescent Microscope

Arc Lamp Laser

ConfocalConfocal MicroscopeMicroscope

Excitation DiaphragmExcitation Filter

Excitation Pinhole

Excitation Filter

OcularPMT

Objective

Emission Filter

Objective

Emission Pinhole

EmissionFilter

15 March 2010Confocal and Atomic Force Microscopy2 DTU Sytems Biology, Technical University of Denmark

Page 3: Confocal Microscopy and Atomic Force Microscopy (AFM) of … · 2014. 1. 15. · Disadvantages of Confocal Microscopy Requires fluorescent samples Uses laser illumination (expensive,

3D reconstruction

# z sections =#images

y

z

x

15 March 2010Confocal and Atomic Force Microscopy3 DTU Sytems Biology, Technical University of Denmark

Page 4: Confocal Microscopy and Atomic Force Microscopy (AFM) of … · 2014. 1. 15. · Disadvantages of Confocal Microscopy Requires fluorescent samples Uses laser illumination (expensive,

Pseudomonas putidall i d ithcells mixed with

Acinetobacter cells in a microbial biofilm

Ch i t t l 1998 A l E i

15 March 2010Confocal and Atomic Force Microscopy4 DTU Sytems Biology, Technical University of Denmark

Christensen et al. 1998. Appl. Environ. Microbiol. 64: 2247-55

Page 5: Confocal Microscopy and Atomic Force Microscopy (AFM) of … · 2014. 1. 15. · Disadvantages of Confocal Microscopy Requires fluorescent samples Uses laser illumination (expensive,

Benefits of Confocal Microscopy

Reduced blurring of the image from light scatteringg g g g Increased effective resolution Improved signal to noise ratio Clear examination of thick specimens Z-axis scanning (3D-reconstruction possible) Magnification can be adjusted electronically X-Y resolution: ~200-250 nm (Ernst Abbe)

Disadvantages of Confocal Microscopy

Requires fluorescent samples Requires fluorescent samples Uses laser illumination (expensive, few wavelengths) Instrument expensive to acquire and run Z-resolution typically >500 nm

15 March 2010Confocal and Atomic Force Microscopy5 DTU Sytems Biology, Technical University of Denmark

Page 6: Confocal Microscopy and Atomic Force Microscopy (AFM) of … · 2014. 1. 15. · Disadvantages of Confocal Microscopy Requires fluorescent samples Uses laser illumination (expensive,

15 March 2010Confocal and Atomic Force Microscopy6 DTU Sytems Biology, Technical University of Denmark

Haagensen et al. 2007. J. Bacteriol. 189:28-37

Page 7: Confocal Microscopy and Atomic Force Microscopy (AFM) of … · 2014. 1. 15. · Disadvantages of Confocal Microscopy Requires fluorescent samples Uses laser illumination (expensive,

61.6

8 M

icro

bio

l. 5

0:

al.

2003.

Mol.

Kla

use

n e

t a

15 March 2010Confocal and Atomic Force Microscopy7 DTU Sytems Biology, Technical University of Denmark

Page 8: Confocal Microscopy and Atomic Force Microscopy (AFM) of … · 2014. 1. 15. · Disadvantages of Confocal Microscopy Requires fluorescent samples Uses laser illumination (expensive,

15 March 2010Confocal and Atomic Force Microscopy8 DTU Sytems Biology, Technical University of Denmark

Page 9: Confocal Microscopy and Atomic Force Microscopy (AFM) of … · 2014. 1. 15. · Disadvantages of Confocal Microscopy Requires fluorescent samples Uses laser illumination (expensive,

New developments in confocal microscopy

• MP (multi-photon) or two-photon confocal microscopy: Two or more photons from a long ( p ) p py p gwavelength illumination at a time excites a fluorophore. High resolution in Z-axis (practally down to ~200 nm)

• White Laser (Leica): A continous wave ”white” laser (tunable from 470-670 nm, with up to 8 lines simultaneously)y)

• STED (Stimulated Emission Depletion)(Leica): A new method where fluourescence is depleted around the area of interest (see next slide)

• Superresolution structured microscopy (Zeiss): A methods where images are rotated and • Superresolution structured microscopy (Zeiss): A methods where images are rotated and combined to create a moire pattern which is deconvolved to create a high res image.

• Photo-Activated localization microscopy (Zeiss): Sequential illumination and localization of fluorophores combined with computational reconstruction of high res images.

• Raman-confocal microscopy (Leica, under development): Confocal microscope combined with raman spectroscope. Enables localized determination of [changes in] concentrations of metabolites etc.

15 March 2010Confocal and Atomic Force Microscopy9 DTU Sytems Biology, Technical University of Denmark

Page 10: Confocal Microscopy and Atomic Force Microscopy (AFM) of … · 2014. 1. 15. · Disadvantages of Confocal Microscopy Requires fluorescent samples Uses laser illumination (expensive,

STED (Stimulated Emission Depletion)• In a Leica TCS STED microscope the sample is illuminated by two pulsed

laser beams tightly synchronizedlaser beams, tightly synchronized.

• The 635 nm wavelength excites the fluorophores of the sample the same way a conventional confocal system does. The excitation laser pulses are directly followed by a ring shaped illumination of a Ti:Sapphire Infrared laser 730-780 nm).

• This pulse inhibits/depletes the fluorescence in the outer regions of the • This pulse inhibits/depletes the fluorescence in the outer regions of the illuminated spot.

• The result: A smaller fluorescence spot that allows much more accurate i th ith th th d i f d li ht X Y 90 scanning than with other methods using focused light. X-Y res: <90 nm

15 March 2010Confocal and Atomic Force Microscopy10 DTU Sytems Biology, Technical University of Denmark

Page 11: Confocal Microscopy and Atomic Force Microscopy (AFM) of … · 2014. 1. 15. · Disadvantages of Confocal Microscopy Requires fluorescent samples Uses laser illumination (expensive,

Sample STED image

Confocal image STED image

15 March 2010Confocal and Atomic Force Microscopy11 DTU Sytems Biology, Technical University of Denmark

Images: Prof. Dr. T. Lang, Univ. Of Bonn, germany and Leica Microsystems

Page 12: Confocal Microscopy and Atomic Force Microscopy (AFM) of … · 2014. 1. 15. · Disadvantages of Confocal Microscopy Requires fluorescent samples Uses laser illumination (expensive,

Atomic Force Microscopy (AFM)• A method to record the topographic property of a surface

Ph i l i t ti ith th f i • Physical interaction with the surface is necessary

15 March 2010Confocal and Atomic Force Microscopy12 DTU Sytems Biology, Technical University of Denmark

Page 13: Confocal Microscopy and Atomic Force Microscopy (AFM) of … · 2014. 1. 15. · Disadvantages of Confocal Microscopy Requires fluorescent samples Uses laser illumination (expensive,

The principle

15 March 2010Confocal and Atomic Force Microscopy13 DTU Sytems Biology, Technical University of Denmark

Page 14: Confocal Microscopy and Atomic Force Microscopy (AFM) of … · 2014. 1. 15. · Disadvantages of Confocal Microscopy Requires fluorescent samples Uses laser illumination (expensive,

15 March 2010Confocal and Atomic Force Microscopy14 DTU Sytems Biology, Technical University of Denmark

Page 15: Confocal Microscopy and Atomic Force Microscopy (AFM) of … · 2014. 1. 15. · Disadvantages of Confocal Microscopy Requires fluorescent samples Uses laser illumination (expensive,

Anne Louise Frost m.fl. 2007 (unpublished)

15 March 2010Confocal and Atomic Force Microscopy15 DTU Sytems Biology, Technical University of Denmark

Page 16: Confocal Microscopy and Atomic Force Microscopy (AFM) of … · 2014. 1. 15. · Disadvantages of Confocal Microscopy Requires fluorescent samples Uses laser illumination (expensive,

Benefits of AFM• Extemely high resolution (down to atomic level few Å), but typically 10-

20 nm)20 nm)• No staining required• Possible to measure attractive/replusive forces

Disadvantages of AFM• Extremely high resolution (very low ”focal depth”)

S ll• Small image area• Sample must be ”flat” and tightly fixed• Very difficult to work in humid or wet environments• Need direct contact to sample• Need direct contact to sample• Imaging depends on tip shape• Slow

15 March 2010Confocal and Atomic Force Microscopy16 DTU Sytems Biology, Technical University of Denmark