Date post: | 07-Jul-2018 |
Category: |
Documents |
Upload: | azima-abdelrhaman |
View: | 213 times |
Download: | 0 times |
of 25
8/18/2019 Curves.pdf
1/25
Curves
1. Patient ventilator-dyssynchrony
2. ECGs
3. Capnographs
4. Arterial waveforms
5. Flow-volume loops
6. ICP waveforms
8/18/2019 Curves.pdf
2/25
8/18/2019 Curves.pdf
3/25
Patient-ventilator dyssynchrony
1. Trigger
2. Flow
3. Exhalation
Case
An old age lady with the history of COPD is ventilated in the ICU. When he sedation is being
weaned off she becomes agitated, restless and sweaty.
What are the differential diagnoses?
Pain
Anxiety
Hunger
Mechanical complications: Pneumothorax, sputum plug. Haemorrhage, cardiac
ischemia/APO, tube dislodgement
Pulmonary embolism
Dys-synchrony with the ventilator
What is the next step in her managements?
Resuscitation
Finding out the cause of the problem
ASSESSMENT
Examination
work of breathing
respiratory pattern
audible sounds (e.g. cuff leak, stridor, wheeze)
chest findings (e.g. hyperexpansion, dullness, crackles)
8/18/2019 Curves.pdf
4/25
Monitor
vital signs
ETCO2
SpO2
Ventilator
waveforms
alarms
Chest x-ray
MANAGEMENT
Resuscitation
address life threats
disconnect patient from ventilator and replace with BVM if required
Address patient factors
treat patients respiratory pathology affecting resistance and/ or compliance (e.g. sputum,
bronchospasm, chest wall eschar, pneumothorax)
treat other patient factors (e.g. hunger, pain, weakness, sleep ,sedation, nutrition, physiotherapy)
Correct problems with the endotracheal tube
kinking
obstruction (e.g. secretions blocking)
impingement on carina or between cords
Correct problems with the ventilator
choose appropriate ventilator
choose appropriate mode
ensure sensitivity is not too low or high
choose appropriate ventilator rate set appropriate flow rate
check that patient isn’t auto-triggering (cardiogenic oscillations, high sensitivity, circuit
leaks, water condensation in the circuit)
sedate patient to reduce agitation
taking over ventilation if fatigue is apparent
8/18/2019 Curves.pdf
5/25
Address ineffective triggering
address PEEPi: — apply increased externally applied PEEP — decrease tidal volume andrespiratory rate — increase expiratory time — bronchodilators
address weakness: — nutrition — reduce sedation — physiotherapy
adjust trigger sensitivity threshold (may lead to inappropriate triggering)
Exhalation dysynchrony
treat underlying patient factors (e.g. COPD, asthma)
adjust exhalation sensitivity or change to time-control cycling between inspiration andexpiration or change to a volume-cycled mode
8/18/2019 Curves.pdf
6/25
Figure 1 Genuine patient triggering (middle) and auto-triggering (right) and with the ventilator
set to pressure trigger and PEEP>0
Autotriggering
triggering of the ventilator in the absence of inspiratory muscle contractionmay result from:
circuit leakcoughing/hiccupswater in circuitseizurescardiogenic oscillations
trigger sensitivity set too low
most common cause for this is a circuit leak. This results in a fall in circuitpressure (figure 1) or a discrepancy between flow leaving the ventilator andreturning to the ventilator.
8/18/2019 Curves.pdf
7/25
Figure 2. Trigger delay indicated by a negative deflection in pressure waveform prior to delivery
of breath and a change in slope of flow waveform. Shaded area on pressure waveform
represents work of triggering
Figure 3. Trigger failure
8/18/2019 Curves.pdf
8/25
Oesophageal pressure (Peso) and flow tracings in the context of patient inspiratory effort. In the
Peso tracing, note that in addition to the effort which starts an inspiratory cycle, there is a second
effort that proves ineffective (hollow arrow).
Oesophageal pressure monitoring is more sensitive than ventilator flow measurement in the
assessment of ineffective triggering.
Causes
1. Patient factors
high PEEPi (must generate enough effort to overcome PEEPi)
weakness
incorrect ventilator settings
ventilator dysfunction
2. ventilator factors
o high level of pressure support or high tidal volume causing gas trapping and
intrinsic PEEP (see above)
8/18/2019 Curves.pdf
9/25
o expiratory asynchrony with delayed opening of exhalation valve
PEEPi is one of the cause of ineffective triggering
Double triggering
Causes
inspiratory time too short> controlled modes
8/18/2019 Curves.pdf
10/25
inspiratory flow rate too low
set tidal volume low> ARDS
coughing and hiccups
Figure 4. Inadequate inspiratory flow
Another example of flow dyssynchrony> inadequate inspiratory flow
Clinically, this presents as upper airway obstruction with signs such as accessory muscle
use,paradoxical breathing, tracheal tug.
Flow related. Low flow or too high flow, also called flow dyssynchrony
8/18/2019 Curves.pdf
11/25
(in constant flow modes (eg volume preset assist control) the inspiratory flow rate may be
inadequate to match the patient’ s attempted inspiratory flow rate).
inspiratory asynchrony may be reduced by use of proportional assist ventilation (PAV) or NAVA
Figure 5. Change in flow waveform characteristic of premature inspiratory to
expiratory cycling in pressure support mode > Premature inspiratory-expiratory
cycling
Termination dyssynchrony. Display of flow (top) and pressure (bottom) versus
time reveals an example of “delayed termination dyssynchrony” (solid arrow) as the
patient attempts exhalation before completion of the inspiratory breath, and “prematuretermination dyssynchrony” (dotted arrows) as the patient makes an inspiratory effort
early in the expiratory phase.
8/18/2019 Curves.pdf
12/25
Arterial waveforms
SPV
Arterial pressure rises during inspiration and falls during expiration due to changes in intra-
thoracic pressure secondary to positive pressure ventilation. > 10 is considered significant.
CAUSES OF INCREASED SYSTOLIC PRESSURE VARIATION
•Hypovolaemia
•Tamponade •Constrictive pericarditis
•LV dysfunction
•Massive PE
•Bronchospasm
•Dynamic hyperinflation
•Pneumothorax
•Raised intrathoracic pressure
•Raised intraabdominal pressure
The utility of SVV is limited in the following settings:
•small tidal volumes (tidal volume must be at least 8 mL/kg)
•spontaneous breathing (patient must have 100% controlled mechanical ventilations at a fixed
rate)
•ARDS and low lung compliance (false negatives more likely)
8/18/2019 Curves.pdf
13/25
•PEEP (may increase SVV)
•arrhythmia (R-R interval must be regular on ECG)
•low heart rate/respiratory rate ratio
•open chest
•right ventricular systolic dysfunction •norepinephrine (may decrease SVV)
•vasodilators (may increase SVV)
•b-blocker medication
Over-damped waveform is slurred -looking and under-damped is overshooting.
8/18/2019 Curves.pdf
14/25
Under-damped waveform overestimates the systolic pressure and under-estimate the diastolic
pressure but the MAP is unaffected. Long stiff tubing and increased vascular resistance are the
causes of under-damping.
Over-damped waveform under-estimates the systolic and over-estimate the diastolic pressures
but again MAP is unaffected. Causes include air bubbles, compliant tubing, catheter kink,blood clots/fibrin, stopcocks, no fluid or low flush bag pressure.
Calibrating (‘zeroing’)
ensure the transducer pressure tubing and flush solution are correctly assembled and freeof air bubbles
place transducer at level of the right atrium
‘off to patient, open to air (atmosphere)’
press ‘zero’ -> sets atmospheric pressure as zero reference point
whenever patient position is altered the transducer height should be altered
Square wave test
aka fast flush test
snap flush to generate square wave
check for oscillations as an indicator of the harmonic characteristics of the system
usually only 1 oscillation before returning to baseline
2 or more oscillations before returning to baseline (underdamped)
if no oscillations (overdamped – response speed is too slow)
A good art line trace has a distinct dicrotic notch, and after the fast flush test there are two
oscillations only.
ACCURACY AND MEASUREMENT ERRORS
Conditions that must be met to ensure accuracy
cannula properly placed within the lumen of an unobstructed artery (ie. no spasm,
thrombus, atheroma proximal to cannula)
cannula not kinked or obstructed
cannula connected by short, rigid, wide-bore tubing to the transducer
no air bubbles in tubing or transducer
interface from fluid to transducer accurately transmits deflections
transducer has adequate frequency response (natural frequency > 100Hz)
transducer is leveled and zeroed to desired point (ie. left atrium)
no zero drift
monitor calibrated accurately
Common sources of error
8/18/2019 Curves.pdf
15/25
bubbles in catheter-transducer system -> decreased resonant frequency
clotting in arterial catheter
elastic walls causes increased damping
cannula won’t flush – kinked, clotted, tissued
OTHER INFORMATION
Information other than blood pressure can be obtained:
pulse rate and rhythm
effects of dysrhythmia on perfusion
ECG lead disconnection
continuous cardiac output using pulse contour analysis
specific wave form morphologies might be diagnostic
— e.g. slow rising = AS, pulsus alternans = tamponade
pulse pressure variation (suggests fluid responsiveness)
steeper upstroke of pulse pressure = increased contractility area under upstroke = SV
steep downstroke = low SVR
Advantages of using MAP rather than SBP/DBP
least dependent on measurement site or technique (whether invasive or not)
least altered by damping
determines tissue blood flow via autoregulation
COMPLICATIONS
Pain
thrombosis and distal ischaemia
infection
increased diagnostic blood loss and anemia
retrograde air embolism
inadvertent drug/air injection
haematoma (+/- nerve compression)
retroperitoneal haematoma (femoral)
bowel perforation (femoral)
vessel damage may lead to stricture and prevent future AV fistula formation for
haemodialysis
pseudo-aneurysm
arterial dissection
arteriovenous fistula
EVIDENCE
8/18/2019 Curves.pdf
16/25
A 2014 observational study using propensity matching based on the Project IMPACT
database found no mortality benefit for use of arterial catheters in medical ICU patients
requiring mechanical ventilation.
8/18/2019 Curves.pdf
17/25
Capnographs
A, Prolonged phase II, increased α angle, and steeper phase III suggest bronchospasm or airway
obstruction.
B, Expiratory valve malfunction resulting in elevation of the baseline, and the angle between the
alveolar plateau and the downstroke of inspiration is increased from 90°. This is due torebreathing of expiratory gases from the expiratory limb during inspiration.
C, Inspiratory valve malfunction resulting in rebreathing of expired gases from inspiratory limb
during inspiration .
D, Capnogram with normal phase II but with increased slope of phase III. This capnogram is
observed in pregnant subjects under general anesthesia .
E, Curare cleft: Patient is attempting to breathe during partial muscle paralysis. Surgical
movements on the chest and abdomen can also result in the curare cleft. (You have maybe 3minutes to sedate the patient before they begin to waken or start to fight the tube.)
F, Baseline is elevated as a result of carbon dioxide rebreathing.
G, Esophageal intubation resulting in the gastric washout of residual carbon dioxide andsubsequent carbon dioxide will be zero.
H, Spontaneously breathing carbon dioxide waveforms where phase III is not well delineated.
I, Dual capnogram in one lung transplantation patient. The first peak in phase III is from the
transplanted normal lung, whereas the second peak is from the native disease lung. A variation ofdual capnogram (steeple sign capnogram – dotted line) is seen if there is a leak around the
8/18/2019 Curves.pdf
18/25
sidestream sensor port at the monitor. This is because of the dilution of expired PCO2 with
atmospheric air.
J, Malignant hyperpyrexia where carbon dioxide is raising gradually with zero baseline
suggesting increased carbon dioxide production with carbon dioxide absorption by the soda lime.
K, Classic ripple effect during the expiratory pause showing cardiogenic oscillations. Theseoccur as a result of to-and-for movement of expired gases at the sensor due to motion of the
heartbeat during expiratory pause when respiratory frequency of mechanical ventilation is low.
Ripple effect like wave forms also occur when forward flow of fresh gases from a source duringexpiratory pause intermingles with expiratory gases at the sensor.
L, Sudden raise of baseline and the end-tidal PCO2 (PETCO2) due to contamination of the
sensor with secretions or water vapor. Gradual rise of baseline and PETCO2 occurs when sodalime is exhausted.
M, Intermittent mechanical ventilation (IMV) breaths in the midst of spontaneously breathing patient. A comparison of the height of spontaneous breaths compared to the mechanical breaths
is useful to assess spontaneous ventilation during weaning process.
N, Cardiopulmonary resuscitation: capnogram showing positive waveforms during eachcompression suggesting effective cardiac compression generating pulmonary blood.
O, Capnogram showing rebreathing during inspiration. This is normal in rebreathing circuitssuch as Mapleson D or Bain circuit.
Ref. Kodali, Bhavani Shankar Anesthesiology. 118(1):192-201, January 2013.
8/18/2019 Curves.pdf
19/25
The examiner may ask you : 1). Indications of monitoring the ICP 2) Different methods
available. 3). Management of raised ICP.
8/18/2019 Curves.pdf
20/25
Inflation of the balloon is triggered by the the beginning of diastole, which correlates with the
middle of the T-wave.
The balloon is timed to deflate at the very end of diastole. This correlates with the R-wave on the
ECG, and this is the most commonly used trigger for balloon deflation.
8/18/2019 Curves.pdf
21/25
Unassisted Peak Systolic Pressure? A Diastolic Augmentation / Augmented Peak Diastolic Pressure? D
Balloon Aortic/ Assisted End Diastole Pressure? F
Assisted Peak Systolic Pressure? E
Unassisted End Diastolic Pressure? B
Balloon inflation? C
What features seen on the aortic pressure waveform indicate optimal timing and function of
the IABP?
Optimal timing and function ofIABP has the following features:
inflation of the balloon occurs at the dicrotic notch (forming the sharp ‘V’)
the slope of rise of augmented diastolic waveform is straight and parallel to the
systolic upstroke
the augmented peak DBP at balloon deflation exceeds or is equal to end-systolic BP
the balloon aortic end-diastolic BP (DBP at balloon deflation) is lower than the
preceding unassisted end-DBP by 15-20 mmHg
the assisted SBP (following a cycle of balloon inflation) is lower than the previousunassisted SBP by 5 mmHg
The timing of IABP inflation and deflation should be manually adjusted to achieve these optimalwaveform characteristics if they are absent.
8/18/2019 Curves.pdf
22/25
Deflation of the IAB is too early – it occurred well before the onset left ventricular contraction and the
start of systole.
Waveform features:
There is a sharp drop in pressure immediately following the peak of diastolicaugmentation (peak B).
Diastolic augmentation may be suboptimal but it is difficult to confirm in the absence of
a pressure scale or comparison to an unassisted waveform.
Assisted aortic end-diastolic BP may be sub-optimally increased (trough C), but it is
difficult to say based on the information given. With early deflation a widened U-shaped
trough is typically seen. Early deflation can lead to an assisted aortic end-diastolic BP
that equals or exceeds the unassisted aortic end-diastolic BP (trough F), although this isnot the case in this scenario.
Assisted SBP (peak D) is the same or higher than the unassisted SBP (peak A) – it should be slightly less.
8/18/2019 Curves.pdf
23/25
Inflation of the IAB is occurring too early – it occurred prior to aortic valve closure (indicated by the
absence of a dicrotic notch).
Waveform features:
Diastolic augmentation (peak B) encroaches on the peak corresponding to unassisted
systole (peak A) – the two peaks have merged and are barely distinguishable.
There is no ‘sharp V’ or dicrotic notch between peaks A and B.
8/18/2019 Curves.pdf
24/25
Deflation of the IAB is occurring too late – deflation is occurring as the aortic valve is opening
Waveform features:
The peak corresponding to diastolic augmentation (peak C) is widened.
Assisted aortic end-diastolic BP (trough E) is the same as, not lower than, the unassisted
aortic end-diastolic BP (trough G).
The upstroke of assisted systolic BP (peak F) has a gentle gradient resulting in a
prolonged rise
Triggering of the IABP is usually set according to the patient’s ECG tracing. When an R wave is detected
the balloon is triggered to automatically start inflating in the middle of the T wave.Triggering can be impaired if the patient develops an arrhythmia, is paced or has a poor ECG trace.
If the IABP cannot be successfully triggered using the R waves from the ECG trace, pacingspikes or the arterial blood pressure waveform may used for triggering. Alternatively, the pump
can be set at an intrinsic rate without any physiological triggering. The IABP may automatically
use these other modes as default backups if ECG-triggered timing is not possible.If the patient
remains hemodynamically stable despite the IABP malfunctioning it might be a good time to
8/18/2019 Curves.pdf
25/25
think about removing the IABP… Treating any underlying causes of arrhythmia is also a good
idea!
Inflation of the IAB is occurring too late – it occurred well after the dicrotic notch (B).
Waveform features:
Inflation of the IAB occurs after the dicrotic notch ( B)
Absence of a sharp V at the point of IAB inflation (should occur at B – the dicrotic notch)
The augmented DBP (peak C) is less than the unassisted SBP (peak A) – it should be
higher