Dr. Jack Driscoll, PID Analyzers, LLC Paper #270-17P Sunday, Feb. 28, 2010 Pittcburg Conf. on Anal. Chem. Orlando, FL Feb. 28, 2010 PID Analyzers, LLC copyright 2010
Transcript
Page 1: Dual Detector Tcd Fuv For Coal Mine Analsis

Dr. Jack Driscoll, PID Analyzers, LLCPaper #270-17P

Sunday, Feb. 28, 2010Pittcburg Conf. on Anal. Chem.

Orlando, FL Feb. 28, 2010

PID Analyzers, LLC copyright 2010

Page 2: Dual Detector Tcd Fuv For Coal Mine Analsis

Presentation Background There are more than 20,000 coal mines in China China relies heavily on the use of coal for power

generation and will be opening a new coal fired power plant every --- months to keep up with their power demands

China is in the process of closing down many of the smaller mines and focusing on the larger ones

Now the need exists to improve the monitoring of the air in coal mines

The method selected was GC with a TCD & FUV

PID Analyzers, LLC copyright 2010

Page 3: Dual Detector Tcd Fuv For Coal Mine Analsis

Target Compounds for Coal MinesCompound

O2 % TCD FUV

N2 % TCD

CH4 h- ppm TCD FUV

NO ppm FUV

CO ppm FUV

NO2 ppm FUV

SO2 ppm FUV

VOC’s ppm FUV

PID Analyzers, LLC copyright 2010

Page 4: Dual Detector Tcd Fuv For Coal Mine Analsis

PID Analyzers, LLC copyright 2010

UV photodiode

Page 5: Dual Detector Tcd Fuv For Coal Mine Analsis

FUV Detector Schematic

PID Analyzers, LLC copyright 2010

Page 6: Dual Detector Tcd Fuv For Coal Mine Analsis

PID Analyzers, LLC copyright 2010

Page 7: Dual Detector Tcd Fuv For Coal Mine Analsis

FUV Sensitivity vs Absorption Coefficient

PID Analyzers, LLC copyright 2010

Page 8: Dual Detector Tcd Fuv For Coal Mine Analsis

Absorption coefficients & estimated detection limits in the Far UV

Compound Abs coeff (cm-1) Det. Limit (ppm)

CO 3 300

CO2 6 140

O2 6.3 130

NO 82 10

NO2 400 2

CH4 485 2

NO2 500 2

H2S 900 1

SO2 1000 1

PID Analyzers, LLC copyright 2010

Page 9: Dual Detector Tcd Fuv For Coal Mine Analsis

FUV Range

PID Analyzers, LLC copyright 2010

Page 10: Dual Detector Tcd Fuv For Coal Mine Analsis

Calibration Curve for Methaneppm CH4 Peak Ht

25000 539173

12500 302620

6250 144479

3100 81241

1550 38061

775 19000

500 9800

388 7500

40 810

PID Analyzers, LLC copyright 2010

y = 21.831x + 6121.1R² = 0.9964

0

100

200

300

400

500

600

0 5000 10000 15000 20000 25000 30000

FUV

Pea

k H

eigh

t

Thou

sand

s

ppm Methane

Page 11: Dual Detector Tcd Fuv For Coal Mine Analsis

PID Analyzers, LLC copyright 2010

Page 12: Dual Detector Tcd Fuv For Coal Mine Analsis

PID Analyzers, LLC copyright 2010

Page 13: Dual Detector Tcd Fuv For Coal Mine Analsis

PID Analyzers, LLC copyright 2010

Page 14: Dual Detector Tcd Fuv For Coal Mine Analsis

PID Analyzers, LLC copyright 2010

Page 15: Dual Detector Tcd Fuv For Coal Mine Analsis

PID Analyzers, LLC copyright 2010

Page 16: Dual Detector Tcd Fuv For Coal Mine Analsis

PID Analyzers, LLC copyright 2010

Page 17: Dual Detector Tcd Fuv For Coal Mine Analsis

TCD PrecisionO2 % N2 %20.29 79.55

20.99 80.7

20.83 80.39

20.69 80.24

20.55 80.15

20.86 80.48

20.58 80.1

21.06 80.41

20.88 80.54

20.61 80.23

20.54 79.97

20.64 80.24

20.75 80.24

21.26 80.4

20.4 79.54

20.6 79.92

20.83 79.42

20.77 79.5

21.08 79.72

20.35 79.25

20.84 79.51

Avg. 20.73333 80.02381

std. dev. 0.2482 0.425869

CV % 1.197107 0.532178

PID Analyzers, LLC copyright 2010

Page 18: Dual Detector Tcd Fuv For Coal Mine Analsis

FUV Precision for methane

FUV % O2

21.01

21.35

21.66

21.16

20.24

21.28

20.45

Ave 21.02143

Std dev 0.506472

CV % 2.409315

PID Analyzers, LLC copyright 2010

Page 19: Dual Detector Tcd Fuv For Coal Mine Analsis

PID Analyzers, LLC copyright 2010

FUV/PIDIn-Series

Page 20: Dual Detector Tcd Fuv For Coal Mine Analsis

PID Analyzers, LLC copyright 2010

FUV/FIDIn-Series

Page 21: Dual Detector Tcd Fuv For Coal Mine Analsis

Summary The FUV/TCD in series has been evaluated for the analysis of

gases in coal mines. The TCD is used to detect high levels of gases > 500 ppm and the FUV to detect low ppm levels of contaminants such as CH4, NO, NO2, SO2…

A combination molecular sieve 5A and HSQ column in-series is used for the separation. The O2, N2, CH4 and CO are only slightly retained on the HSQ but are baseline separated on the mole sieve (MS). Once the separation is complete on the MS, the columns are backflushed and the components on the HSQ are eluted to the detector.

The use of detectors in-series for analysis of samples provides a dynamic range of approximately 10 exp 6 for the two detectors

PID Analyzers, LLC copyright 2010