+ All Categories
Home > Documents > Final report on AFRIMETS comparison of Liquid in Glass [email protected] 4 Apr 2016 NMISA...

Final report on AFRIMETS comparison of Liquid in Glass [email protected] 4 Apr 2016 NMISA...

Date post: 20-Apr-2020
Category:
Upload: others
View: 1 times
Download: 0 times
Share this document with a friend
18
AFRIMETS.T-S5 Supplementary comparison Liquid-in-glass thermometers -35 °C to 250 °C Final report (16 August 2017) Page 1 of 18 Final report on AFRIMETS comparison of Liquid in Glass thermometer calibrations from -35 °C to 250 °C (Final report) 16 August 2017 1. Introduction ..................................................................................................................................................... 1 2. Participants ..................................................................................................................................................... 1 3. Detailed description of the devices ................................................................................................................. 2 4. Procedure for measurement ........................................................................................................................... 2 5. Comparison reference value (RV) .................................................................................................................. 2 6. Results ............................................................................................................................................................ 2 7. Evaluation of results ....................................................................................................................................... 8 8. Conclusions and recommendations ............................................................................................................... 8 9. Appendix: Uncertainty budgets .....................................................................................................................10 1 Introduction The purpose of this interlaboratory comparison (ILC) is to test the proficiency / evaluate the measurement capabilities of the participant National Metrology Institutes (NMIs), in the field of calibration of liquid-in-glass thermometers (LIGTs), in order to support their submissions of Calibration and Measurement Capabilities (CMCs) in the field of Thermometry to the BIPM Key Comparison Database (KCDB). The comparison forms part of the project “Strengthening Quality Infrastructure for Trade Enhancement and Consumer Protection in the SADC Region in accordance with the TBT Annex to the SADC Trade protocol” (Project No.: PN 2012.2297.5 – 95088). In accordance with the terms of the project, participation in the ILC was restricted to those members of the SADCMET sub-RMO (Regional Metrology Organisation) who are signatories of the CIPM Mutual Recognition Arrangement (MRA), namely, Botswana, Mauritius, Namibia, Seychelles, South Africa, Zambia and Zimbabwe. 2 Participants Institute Contact person Email Measuremen t period National Metrology Institute of South Africa (NMISA) Dr Efrem Ejigu [email protected] 13 Jan 2016
Transcript
Page 1: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

AFRIMETS.T-S5 Supplementary comparison Liquid-in-glass thermometers -35 °C to 250 °C Final report (16 August 2017)

Page 1 of 18

Final report on AFRIMETS comparison of Liquid in Gl ass

thermometer calibrations from -35 °C to 250 °C (Final report)

16 August 2017

1. Introduction ..................................................................................................................................................... 1

2. Participants ..................................................................................................................................................... 1

3. Detailed description of the devices ................................................................................................................. 2

4. Procedure for measurement ........................................................................................................................... 2

5. Comparison reference value (RV) .................................................................................................................. 2

6. Results ............................................................................................................................................................ 2

7. Evaluation of results ....................................................................................................................................... 8

8. Conclusions and recommendations ............................................................................................................... 8

9. Appendix: Uncertainty budgets ..................................................................................................................... 10

1 Introduction

The purpose of this interlaboratory comparison (ILC) is to test the proficiency / evaluate the measurement

capabilities of the participant National Metrology Institutes (NMIs), in the field of calibration of liquid-in-glass

thermometers (LIGTs), in order to support their submissions of Calibration and Measurement Capabilities

(CMCs) in the field of Thermometry to the BIPM Key Comparison Database (KCDB).

The comparison forms part of the project “Strengthening Quality Infrastructure for Trade Enhancement and

Consumer Protection in the SADC Region in accordance with the TBT Annex to the SADC Trade protocol”

(Project No.: PN 2012.2297.5 – 95088). In accordance with the terms of the project, participation in the ILC

was restricted to those members of the SADCMET sub-RMO (Regional Metrology Organisation) who are

signatories of the CIPM Mutual Recognition Arrangement (MRA), namely, Botswana, Mauritius, Namibia,

Seychelles, South Africa, Zambia and Zimbabwe.

2 Participants

Institute Contact person Email Measuremen

t period

National Metrology Institute of South

Africa (NMISA)

Dr Efrem Ejigu [email protected] 13 Jan 2016

Page 2: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

AFRIMETS.T-S5 Supplementary comparison Liquid-in-glass thermometers -35 °C to 250 °C Final report (16 August 2017)

Page 2 of 18

Scientific & Industrial Research &

Development Centre - National

Metrology Institute (SIRDC-NMI)

Mr Blessing

Chibaya

[email protected] 19 Jan 2016

Zambia Bureau of Standards (ZABS) Mr Victor Mwazi [email protected] 22 Jan 2016

Botswana Bureau of Standards

(BOBS)

Mr Tebogo

Kajane

[email protected] 25 Jan 2016

Namibian Standards Institute (NSI) Mr Victor

Mundembe

[email protected] 12 Feb 2016

NMISA 26 Feb 2016

Mauritius Standards Bureau (MSB) Mr Christian Ng

Ha Kwong

[email protected] 4 Apr 2016

NMISA 20 Jun 2016

Seychelles Bureau of Standards (SBS) Mr Gilbert

Madeleine

[email protected] 12 Sep 2016

3 Detailed description of the devices

The Units Under Test (UUTs) were mercury-in-glass thermometers designed for total immersion of the liquid

column.

Manufactur er Model Range Serial no

Amarell L 33 102 (solid-stem) -38 to +50: 0.2 °C 1469700

N 60 076 (solid-stem) +98 + 152: 0.1 °C 35765

N 60 008 (enclosed-scale) +198 + 252: 0.1 °C 5

4 Procedure for measurement

Participants were requested to perform all measurements as close to total immersion as the local heat

source allowed. Any deviation from total immersion was to be taken into account in the uncertainty budget, or

corrected for (Emergent Liquid Column correction). The temperature indicated by the transfer artefact was

requested to be within 2 °C of the nominal temperature.

5 Comparison reference value (RV)

As NMISA is accredited by SANAS to perform LIGT calibrations, with uncertainties from 0.015 °C to 0.05 °C

over the range of the ILC, NMISA's results were chosen to be used as Comparison Reference Values (RV).

In the analysis below, NMISA's initial corrections are used as the Reference Values (RV) and each

participant's corrections as the Lab Values (LV).

Page 3: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

AFRIMETS.T-S5 Supplementary comparison Liquid-in-glass thermometers -35 °C to 250 °C Final report (16 August 2017)

Page 3 of 18

Though this approach deviates from the usual philosophy of NMI-level comparisons, namely that all

participants' results contribute to the Reference Value, the present “proficiency testing” approach is

considered more reasonable, as the Measurement Capabilities of the participants other than NMISA are

largely untested. Though NMISA does not yet have published CMCs for LIGT calibration in the KCDB, and

no Key Comparison has or will take place in this industrial thermometry field, the institute has performed

similar measurements in APMP.T-S6 of August 2009: this ILC involved calibration of an industrial PRT by

comparison with SPRTs in liquid baths, and NMISA achieved |LV-RV| ≤ 0.004 °C and U (LV) ≤ 0.008 °C over

the range -50 °C to 400 °C. (Its SPRT CMCs in the KCDB are supported by its July 2000 APMP.T-K3 results

from -39 °C to 420 °C, namely, |LV-RV| < 0.001 °C and U(LV) ≤ 0.001 °C.)

6 Results

The Reference Values differ somewhat from those used to prepare the Peer Review reports issued to

participants from 19 January to 7 April, as Emergent Liquid Column (ELC) corrections of -0.02 °C to +0.05 °C

were only applied to NMISA's results for the present report. The only significant changes (|En| <1 to >1, or

vice versa) are as follows:

BOBS, 250 °C: LV-RV = 0.08 °C to 0.13 °C (En = 0.8 to 1.5)

MSB, 200 °C: LV-RV = -0.10 °C to -0.05 °C (En = -1.2 to -0.7)

Participant: NMISA Meas date: 14/Jan/2016 Cal'ed by: T Madiba

Actual Indicated Correction U(LV) RV U(RV) LV-RV

temp (°C) temp (°C) (°C) (°C) (°C) (°C) (°C) En

-35.040 -35.020 -0.020 0.030 -0.020 0.030 0.000 0.0

-30.050 -30.020 -0.030 0.030 -0.030 0.030 0.000 0.0

-0.020 -0.010 -0.010 0.030 -0.010 0.030 0.000 0.0

49.980 50.010 -0.030 0.030 -0.030 0.030 0.000 0.0

99.990 100.010 -0.020 0.030 -0.020 0.030 0.000 0.0

149.990 150.030 -0.040 0.030 -0.040 0.030 0.000 0.0

200.030 200.040 -0.010 0.030 -0.010 0.030 0.000 0.0

249.990 250.050 -0.060 0.030 -0.060 0.030 0.000 0.0

Participant:

SIRDC-

NMI Meas date: 19/Jan/2016 Cal'ed by:

B Chibaya & J

Madziwa

Actual Indicated Correction U(LV) RV U(RV) LV-RV

temp (°C) temp (°C) (°C) (°C) (°C) (°C) (°C) En

-35.003 -35.000 -0.003 0.250 -0.020 0.030 0.017 0.1

Page 4: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

AFRIMETS.T-S5 Supplementary comparison Liquid-in-glass thermometers -35 °C to 250 °C Final report (16 August 2017)

Page 4 of 18

Participant:

SIRDC-

NMI Meas date: 19/Jan/2016 Cal'ed by:

B Chibaya & J

Madziwa

Actual Indicated Correction U(LV) RV U(RV) LV-RV

temp (°C) temp (°C) (°C) (°C) (°C) (°C) (°C) En

-29.814 -29.800 -0.014 0.250 -0.030 0.030 0.016 0.1

-0.005 0.000 -0.005 0.250 -0.010 0.030 0.005 0.0

50.003 50.000 0.003 0.250 -0.030 0.030 0.033 0.1

99.764 99.900 -0.136 0.600 -0.020 0.030 -0.116 -0.2

150.061 150.000 0.061 0.600 -0.040 0.030 0.101 0.2

-0.010 0.030

250.569 250.600 -0.031 0.820 -0.060 0.030 0.029 0.0

Participant: ZABS Meas date: 22/Jan/2016 Cal'ed by:

B Fisambo, M

Nyambe & V Mwazi

Actu al Indicated Correction U(LV) RV U(RV) LV-RV

temp (°C) temp (°C) (°C) (°C) (°C) (°C) (°C) En

-34.980 -35.000 0.020 0.130 -0.020 0.030 0.040 0.3

-50 0 50 100 150 200 250-0.8

-0.4

0.0

0.4

0.8

AFRIMETS.T-S5: LIGT (Hg, total immersion), 2016

SIRDC-NMI U(RSS) U(RSS)

Temperature (°C)

LV -

RV

(°C

)

Page 5: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

AFRIMETS.T-S5 Supplementary comparison Liquid-in-glass thermometers -35 °C to 250 °C Final report (16 August 2017)

Page 5 of 18

Participant: ZABS Meas date: 22/Jan/2016 Cal'ed by:

B Fisambo, M

Nyambe & V Mwazi

Actu al Indicated Correction U(LV) RV U(RV) LV-RV

temp (°C) temp (°C) (°C) (°C) (°C) (°C) (°C) En

-29.980 -30.000 0.020 0.130 -0.030 0.030 0.050 0.4

0.000 0.000 0.000 0.120 -0.010 0.030 0.010 0.1

50.040 49.977 0.063 0.130 -0.030 0.030 0.093 0.7

99.610 99.600 0.010 0.100 -0.020 0.030 0.030 0.3

149.990 150.000 -0.010 0.080 -0.040 0.030 0.030 0.4

201.070 201.100 -0.030 0.060 -0.010 0.030 -0.020 -0.3

Participant: BOBS Meas date: 25/Jan/2016 Cal'ed by: N Nkgare

Actual Indicated Correction U(LV) RV U(RV) LV-RV

temp (°C) temp (°C) (°C) (°C) (°C) (°C) (°C) En

-34.874 -35.000 0.126 0.130 -0.020 0.030 0.146 1.1

-29.889 -30.000 0.111 0.130 -0.030 0.030 0.141 1.1

0.016 0.000 0.016 0.130 -0.010 0.030 0.026 0.2

-50 0 50 100 150 200 250-0.8

-0.4

0.0

0.4

0.8

AFRIMETS.T-S5: LIGT (Hg, total immersion), 2016

ZABS U(RSS) U(RSS)

Temperature (°C)

LV -

RV

(°C

)

Page 6: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

AFRIMETS.T-S5 Supplementary comparison Liquid-in-glass thermometers -35 °C to 250 °C Final report (16 August 2017)

Page 6 of 18

Participant: BOBS Meas date: 25/Jan/2016 Cal'ed by: N Nkgare

Actual Indicated Correction U(LV) RV U(RV) LV-RV

temp (°C) temp (°C) (°C) (°C) (°C) (°C) (°C) En

50.140 50.200 -0.060 0.130 -0.030 0.030 -0.030 -0.2

100.250 100.200 0.050 0.080 -0.020 0.030 0.070 0.8

150.313 150.300 0.013 0.090 -0.040 0.030 0.053 0.6

200.325 200.300 0.025 0.090 -0.010 0.030 0.035 0.4

250.368 250.300 0.068 0.080 -0.060 0.030 0.128 1.5

Participant: NSI Meas date: 12/Feb/2016 Cal'ed by:

G Mabakeng & H

Shikomba

Actual Indicated Correction U(LV) RV U(RV) LV-RV

temp (°C) temp (°C) (°C) (°C) (°C) (°C) (°C) En

-0.020 0.030

-29.750 -29.950 0.200 0.200 -0.030 0.030 0.230 1.1

0.140 0.040 0.100 0.200 -0.010 0.030 0.110 0.5

50.160 50.190 -0.030 0.200 -0.030 0.030 0.000 0.0

-50 0 50 100 150 200 250-0.8

-0.4

0.0

0.4

0.8

AFRIMETS.T-S5: LIGT (Hg, total immersion), 2016

BOBS U(RSS) U(RSS)

Temperature (°C)

LV -

RV

(°C

)

Page 7: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

AFRIMETS.T-S5 Supplementary comparison Liquid-in-glass thermometers -35 °C to 250 °C Final report (16 August 2017)

Page 7 of 18

Participant: NSI Meas date: 12/Feb/2016 Cal'ed by:

G Mabakeng & H

Shikomba

Actual Indicated Correction U(LV) RV U(RV) LV-RV

temp (°C) temp (°C) (°C) (°C) (°C) (°C) (°C) En

100.140 100.080 0.060 0.200 -0.020 0.030 0.080 0.4

149.870 149.937 -0.067 0.200 -0.040 0.030 -0.027 -0.1

Participant: NMISA Meas date: 26/Feb/2016 Cal'ed by: T Madiba

Actual Indicated Correction U(LV) RV U(RV) LV-RV

temp (°C) temp (°C) (°C) (°C) (°C) (°C) (°C) En

-0.040 -0.010 -0.030 0.030 -0.010 0.030 -0.020 -0.5

99.980 100.010 -0.030 0.030 -0.020 0.030 -0.010 -0.2

200.010 200.040 -0.030 0.030 -0.010 0.030 -0.020 -0.5

Participant: MSB Meas date: 4/Apr/2016 Cal'ed by: C Ng Ha Kwong

Actual Indicated Correction U(LV) RV U(RV) LV-RV

temp (°C) temp (°C) (°C) (°C) (°C) (°C) (°C) En

-50 0 50 100 150 200 250-0.8

-0.4

0.0

0.4

0.8

AFRIMETS.T-S5: LIGT (Hg, total immersion), 2016

NSI U(RSS) U(RSS)

Temperature (°C)

LV -

RV

(°C

)

Page 8: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

AFRIMETS.T-S5 Supplementary comparison Liquid-in-glass thermometers -35 °C to 250 °C Final report (16 August 2017)

Page 8 of 18

Participant: MSB Meas date: 4/Apr/2016 Cal'ed by: C Ng Ha Kwong

Actual Indicated Correction U(LV) RV U(RV) LV-RV

temp (°C) temp (°C) (°C) (°C) (°C) (°C) (°C) En

-0.020 0.030

-0.030 0.030

0.010 0.000 0.010 0.040 -0.010 0.030 0.020 0.4

49.990 50.000 -0.010 0.040 -0.030 0.030 0.020 0.4

100.000 100.000 0.000 0.070 -0.020 0.030 0.020 0.3

149.960 150.010 -0.050 0.070 -0.040 0.030 -0.010 -0.1

199.960 200.020 -0.060 0.070 -0.010 0.030 -0.050 -0.7

Note: The liquid column of the 198 °C to 252 °C thermometer separated during return shipment from

Mauritius to South Africa, and was not able to be repaired. For this reason, no 200 °C verification appears

below. Visual inspection by MSB (and close agreement of their 200 °C measurement with the RV) suggests

that their 200 °C result may be included in the ILC (i.e., that it is unaffected by the fault).

Participant: NMISA Meas date: 20/Jun/2016 Cal'ed by: S du Clou

Actual Indicated Correction U(LV) RV U(RV) LV-RV

temp (°C) temp (°C) (°C) (°C) (°C) (°C) (°C) En

-50 0 50 100 150 200 250-0.8

-0.4

0.0

0.4

0.8

AFRIMETS.T-S5: LIGT (Hg, total immersion), 2016

MSB U(RSS) U(RSS)

Temperature (°C)

LV -

RV

(°C

)

Page 9: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

AFRIMETS.T-S5 Supplementary comparison Liquid-in-glass thermometers -35 °C to 250 °C Final report (16 August 2017)

Page 9 of 18

Participant: NMISA Meas date: 20/Jun/2016 Cal'ed by: S du Clou

Actual Indicated Correction U(LV) RV U(RV) LV-RV

temp (°C) temp (°C) (°C) (°C) (°C) (°C) (°C) En

-0.020 -0.010 -0.010 0.030 -0.010 0.030 0.000 0.0

100.000 100.010 -0.010 0.030 -0.020 0.030 0.010 0.2

Participant: SBS Meas date: 12/Sep/2016 Cal'ed by: H Richmond

Actual Indicated Correction U(LV) RV U(RV) LV-RV

temp (°C) temp (°C) (°C) (°C) (°C) (°C) (°C) En

100.000 100.020 -0.020 0.034 -0.020 0.030 0.000 0.0

150.002 150.020 -0.018 0.036 -0.040 0.030 0.022 0.5

Note: Owing to logistical reasons, no closing measurement could be made to confirm artefact stability after

measurement at SBS - however, the results obtained by SBS suggest that the artefact was in good condition

at the time of measurement, and that SBS results may reasonably be compared to the Reference Values.

7 Evaluation of results

NMISA's 26 Feb and 20 June verification checks agree with their initial measurements within 0.02 °C,

-50 0 50 100 150 200 250-0.8

-0.4

0.0

0.4

0.8

AFRIMETS.T-S5: LIGT (Hg, total immersion), 2016

SBS U(RSS) U(RSS)

Temperature (°C)

LV -

RV

(°C

)

Page 10: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

AFRIMETS.T-S5 Supplementary comparison Liquid-in-glass thermometers -35 °C to 250 °C Final report (16 August 2017)

Page 10 of 18

indicating that the artefacts were stable during the ILC. U(RV)/U(LV) ≤ 1 at all points (indeed,

U(RV)/U(LV) ≤ 0.5 at all points except MSB's 0 °C and 50 °C), indicating that RV is sufficiently accurate to

test most participants' measurement capabilities rigorously.

8 Conclusions and recommendations

The artefacts were hand-carried from South Africa to the first four participants. (Hand-carriage to Mauritius

and Seychelles was not possible, as mercury-in-glass thermometers may not be carried in passengers'

luggage on aero planes.) The comparison coordinator visited each of these four institutes for two days,

performing an on-site peer review of the evidence supporting the laboratory's Thermometry measurement

capabilities. During that time, laboratory staff completed their ILC measurements, calculated uncertainties

and handed their results to the coordinator.

This method of artefact circulation had significant advantages, namely:

i. The risk of damage to the artefacts was greatly reduced. (This risk is otherwise significant, as shown

by damage to one artefact on the return flight from Mauritius, and destruction of a second artefact

during shipment by courier from South Africa to Seychelles. The latter incident made it impractical for

Seychelles Bureau of Standards to participate in the present ILC.)

ii. The time required for shipment (particularly, Customs clearance) was shortened from approximately

one month (as experienced during past regional ILCs) to two days per journey.

iii. Delays in the transmission and analysis of results were avoided.

The present ILC has largely avoided the delays and damage that affected the 2001 LIGT ILC SADC.T-P1,

while covering a much larger temperature range than the 0 °C to 50 °C of that ILC. It should be of significant

value to participants, both in support of Thermometry CMC submissions to the AFRIMETS TC/WG-T, and to

satisfy the proficiency testing requirements of accreditation bodies.

As a guide to participants, the smallest CMC which can be supported by this interlaboratory comparison, for

a particular participant, at a particular point, is the larger of U (LV) and |LV-RV| at that point.

Page 11: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

9 Appendix: Uncertainty budgets

Note: As the values of dominant uncertainty components estimated by a participant are often similar at the various comparison temperatures, a typical uncertainty

budget for each participant (the one for 100 °C) is reproduced below, rather than all their budgets.

Page 12: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

NMISA:

Rtp (Ω)25.579,155 Rs (Ω) dRs/dT/Rs (Ω/Ω/K)ESDM (°C)

99.99°C Std: PRT 25.579,188 380Ω -5.E-07 0.0062

REF STD: Value Unit Divisor Sensitivity coeff ui(k=1) (°C) D.o.f.PRT:

Cal of std thermom 0.002,00 °C 2.000 1 0.001,00 500.0 0.0

Drift of std thermom 0.001,40 °C 1.732 1 0.000,81 500.0 0.0

Self-heating

Stability of Rtp 0.000,02 Ω 1.732 13.900 0.000,13 500.0 0.0

Uncert of ref temp (WTP or ice pt) 0.000,30 °C 2.000 1.390 0.000,21 500.0 0.0

Elec. measuring instrument 0.0E+00 Ω/Ω 2.000 3800.0 0.000,00 41.0 0.0

Elec. measuring instrument 3.E-05 3.E-05 Ω 2.000 10.0 0.000,15 31.0 0.0

Ref resistor: cal

Ref resistor: temp coeff 0.2 °C 1.732 -3.E-07 0.000,00 500.0 0.0

HEAT SOURCE:

Temp diff betw std & UUT: melt/freeze coincidence, horiz temp gr 0.0030 °C 1.732 1 0.001,73 500.0 0.0

Control cycle of heat source: Zero if averaged over several cyclesUUT

LIG:Resolution 8 8 0.013 °C 1.732 1.0 0.007,22 500.0 0.0Pressure dependence 102.8 102 0.8 kPa 1.732 0.001 0.000,46 500.0 0.0ELC(cal) - ELC(use) 0.5 0.5 0.250 °C 1.732 0.00842 0.001,22 500.0 0.0Uncert of ELC temperature 0.1285714286 10.03 °C 1.732 0.00016 0.000,92 500.0 0.0

Repeatability 0.006 °C 1.000 1.00 0.006,22 2.0 0.0Reproducibility 0.003 °C 1.732 1.00 0.001,59 500.0 0.0

uc(k=1) (°C): 0.010,03Eff deg of freedom 13.4 8.E-10

t(eff d.o.f.): 2.21U(k=2) (°C): 0.022,16

BMC (°C): 0.030,00

Page 13: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

SIRDC-NMI:

Source of uncertainty Value Unit Probability Divisor Sensitivity Uncertainty (k=1)

distribution (°C / unit) (°C)

Uncert of reference std 0.0015 ° C NORMAL K=2 2 1 0.0008 500

Uncert of Resistance Measuring Device 0.0005 Ω NORMAL K=2 2 2.5 0.0008 500

Resolution of STD 0.0001 ° C RECTANGULAR 1.7320508076 1 0.0001 5 00

Drift of reference std 0.003 Ω RECTANGULAR 1.7320508076 2.5 0.00435 00

Stability of R(0.01°C) or R(0°C): ref PRT 0.0024 Ω RECTANGULAR 1.7320508076 2.5 0.0035 2 0

Uncert of ref temperature (WTP or ice point): ref thermometer 0.05 ° C RECTANGULAR 1.7320508076 1 0.0289 2 0

Temperature gradient in bath/furnace/FP cell 0.03 ° C RECTANGULAR 1.7320508076 1 0.0173 2 0

Resolution of UUT 0.1 ° C RECTANGULAR 1.7320508076 1 0.0577 5 00

Uncertainty of ref temperature (WTP or ice point): UUT 0.05 ° C RECTANGULAR 1.7320508076 1 0.0289 5 0 0

Repeatability 0 ° C Normal K=1 1 1 0.0000 5 0 . 0

UUT: Emergent Liquid Columnn 0.02048 ° C RECTANGULAR 1.7320508076 1 0.0118 20

Vertical Temperature Uniformity 0.515 ° C RECTANGULAR 1.7320508076 1 0.2973 20

E f fe c t iv e d e g r e e s o f f r e e d o m 2 2 .5 4 8 5 7 1 6 9

S t u d e n t t 's e f fe c t iv e d e g r e e s o f f r e e d o m 2 .1

Combined uncertainty (k=1): 0.3064

Expanded Uncertainty (k=2) 0.6128

D e g r e e s o f

f r e e d o m

Page 14: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

ZABS:

100.00°C UncertaintySource of uncertainty Value Unit Probability Divisor Sensitivity contribution

distribution (k=1) REFERENCE STANDARDCalibration of standard thermometer 0.002000 °C Normal 2.00 1.E+00 0.0010000Repeatability of standard thermometer 0.003569 °C Normal 1.00 1.E+00 0.0035690Drift of standard thermometer 0.006000 °C Rectangular 1.73 1.E+00 0.0034641Resolution of standard thermometer 0.000050 °C Rectangular 1.73 1.E+00 0.0000289Uncertainty of Electrical Measuring Equipment 0.010000 °C Rectangular 1.73 1.E+00 0.0057735HEAT SOURCEBath Stability 0.004000 °C Rectangular 1.73 1.E+00 0.0023094Temperature gradient of bath 0.005000 °C Rectangular 1.73 1.E+00 0.0028868UNIT UNDER TEST (UUT)Resolution of test thermometer 0.050000 °C Rectangular 1.73 1.E+00 0.0288675Repeatability of test thermometer 0.000000 °C Normal 1.00 1.E+00 0.0000000Emergent Stem Temperature Correction 0.000000 °C Rectangular 1.73 1.E+00 0.0000000

Combined uncertainty (k=1): 0.0301010

2.0Expanded uncertainty U(k=2): 0.0602021

Page 15: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

BOBS:

Page 16: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

NSI:

Page 17: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

MSB:

U n it Va lu e D is trib u tio D iv is or S e n s itiv ity C on trib u ti

c a lib ra tion of s td °C 0 .0 1 5 n orm a l 2 1 0.0075 9 5 2 0 0 1 .5 8 E -1 1

drift of standard °C 0 .0 1 5 rec ta n g u la 1 .7 3 2 0 5 1 1 0.008660254 9 5 2 0 0 2 .8 1 E -1 1

Resolution UUT °C 0.01 rec ta n g u la 1 .7 3 2 0 5 1 1 0.005773503 9 5 2 0 0 5 .5 6 E -1 2

Repeatability UUT °C 0 rec ta n g u la 1 .7 3 2 0 5 1 1 0 9 5 2 0 0 0

Temp Gradient bath °C 0 .0 5 rec ta n g u la 1 .7 3 2 0 5 1 1 0.028867513 9 5 2 0 0 3 .4 7 E -0 9

Stability of bath °C 0 .0 0 1 2 2 4 rec ta n g u la 1 .7 3 2 0 5 1 1 0.000706677 9 5 2 0 0 1 .2 5 E -1 5

Job c od e MT/2016/14/0374Ite m MIG Thermometer 100 °C c om b in e d u n c e rta in t 0.031597722 °C ex p a n d e d u0.063195445

C lie n t LMC Solutions 2 8 3 .0 5 3

Relia b ility

(% )

D e g re e s

of

fre e d om u i4(y )/v i

e f fe c tiv e

d e g re e of

fre e d om

Page 18: Final report on AFRIMETS comparison of Liquid in Glass ...CNgHaKwong@msb.intnet.mu 4 Apr 2016 NMISA 20 Jun 2016 Seychelles Bureau of Standards (SBS) Mr Gilbert Madeleine mems-sbs@email.sc

SBS:

Uncertainty Budget at 100 ºC

DescriptionDivisor Unit Unit Comments

B Normal 0.018 2 0.0090000 ºC 1 ºC 0.009 1E^10

B Rectangular 0.0005 1.732 0.0002887 ºC 1 ºC 0.0002886836 1E^10

A Normal 0.00012 1 0.0001225 ºC 1 ºC 0.0001224745 4A Normal 0 1 0.0000000 ºC 1 ºC 0 4

B Rectangular 0.02 1.732 0.0115470 ºC 1 ºC 0.0115470054 1E^10

B Rectangular 0.01 1.732 0.0057735 ºC 1 ºC 0.0057735027 1E^10 obtained from manufacturer

B Rectangular 0.009 1.732 0.0051962 ºC 1 ºC 0.0051961524 50

B Rectangular 0 1.732 0 ºC 1 ºC 0 1E^10

B Rectangular 0.01 1.732 0.0057735 ºC 1 ºC 0.0057735027 1E^11 obtained from manufacturerStdard Therm Therm UT Combined Standard Uncertainty 0.0175527302

1 100.001 100.020 Effective Degrees of freedom veff 6510.58

2 100.001 100.020 Level of confidence 95%

3 100.001 100.020 Coverage factor k 1.96

4 100.002 100.020 Expanded Uncertainty 0.0344033512

5 100.002 100.020 Expression of the measurement value

Experimental Standard Deviation 0.0002738613 0

ESDM 0.0001224745 0

Uncertainty Analysis Type

Expected Value

Probability Distribution

Uncertainty Estimate

Standard Uncertainty

u(xi)

Sensitivity Coefficient

ci

Uncertainty Contribution

u(yi)

Degrees of Freedom

νi

us1 : Calibration uncertainty of standard PRT probe & Indicator

us-res : Resolution uncertainty of standard thermometer PRT

us-rep : Repeatability uncertainty of standard thermometer PRT

u5: Repeatability uncertainty of Thermometer Under Test

u6: Resolution uncertainty of Thermometer Under Test

u7: Stability uncertainty of the Bath temperature

u8: Drift of Standard ThermometerNew standard thermometer(1 month) used calibration uncertainty divided by 2

u9: Emergent Liquid columnuut was totally immersed to calibartion point and raised 6mm above surface when taking the reading

u10: Bath inhomogeneity


Recommended