DRÄGER SAFETY HYGIENE CONFERENCE CHICAGO USA 18.05.2006 FRANK ARNOLD • SLIDE NO.1 GAS DETECTION TUBES FOR FUMIGANTS 18.05.2006 CHICAGO USA
Transcript
Page 1: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.1

GAS DETECTION TUBES FOR FUMIGANTS

18.05.2006 CHICAGO USA

Page 2: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.2

SHORT-TERM TUBES WORLDWIDE

• Worldwide experience since more than 65 years

• More than 200 different types of gas detection Tubes

• Easy to use• Reasonable price• Direct reading• single hand operation

Page 3: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.3

GAS DETECTION TUBES

Draeger

Water Vapor 1/b

n=

2

15

10

5

3

1

Airflow directional arrow

Tube nomenclature indicates gas to be measured

Determine reading using the length of color stain

Number of pump strokesBatch: ARHM - 0111

(Aim towards pump inlet)

Quality control batch number

8101781

Part number

Hermetically-sealed glass tube housing

Page 4: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.4

DRAEGER ACCURO PUMP

-- --

Pump stroke counter

Airflow directional arrow

End of stroke indicator

(Facing pump inlet)

Scissors & spring mechanisms provides uniform pump stroke

Durable EPDM rubber housing

Exhaust port

Allen screw for pump maintenance

Wrist strap

Ceramic tube opener

Bellows-style pump

AREL - F012

Quality control batch number

Page 5: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.5

Draeger Water Vapor 1/b

n=

2

1510531

Batch: ARHM - 0111

8101781

A SPECIFIC CHEMICAL REACTION SHOWS A VISIBLE STAIN

Page 6: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.6

GAS DETECTION PUMP

Fumigation is an important market for gas detectiontubes since more than 20 years

New fumigants coming up in the last 5 years because:

•Phase Out of Methyl Bromide by the UN (Montreal protocol)

•Increasing Resistances against Phosphine

Page 7: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.7

STATUS OF FUMIGATION

Phase Out of Methyl Bromide by the UN is the start forinventing a big variety of new fumigants

Page 8: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.8

STATUS OF FUMIGATIONSEGMENTS OF FUMIGATION

Agricultural: Post Harvest Fumigation (PH3)

Page 9: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.9

STATUS OF FUMIGATIONSEGMENTS OF FUMIGATION

Agricultural: Preplant Soil Fumigation (Methyl Bromide)

Page 10: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.10

STATUS OF FUMIGATIONSEGMENTS OF FUMIGATION

Architecture: Structural fumigation of houses and churches for pest control (Methyl Bromide, will be replaced by i.e. Sulfuryl Fluoride)

Page 11: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.11

STATUS OF FUMIGATIONSEGMENTS OF FUMIGATIONTransport/Quarantine: Fumigations of Containers (different fumigants)

Page 12: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.12

STATUS OF FUMIGATIONSEGMENTS OF FUMIGATION

Transport/Quarantine/quality: Fumigations of Containers (different fumigants)Fumigation of bulk ships (PH3)Fumigation of wood packaging material (Methyl Bromide)Fumigation of warehouses

Page 13: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.13

NEW REGULATIONS

New important issues were set up for achievingglobal transport and global trade:

• Eurep Gapstandards and procedures for the global certification of Good Agricultural Practices

• IPPC– ISPM 15

Page 14: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.14

WOOD PACKAGING MATERIAL

IPPC International plant protection convention

is attached directly to the UNO, 132 states have ratified this contract since 2004

www.ippc.int

ISPM 15 International Standard for Phytosanitary Measures (ISPM 15)

Page 15: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.15

ISPM !%

The compliance with ISPM 15 standards is tied into 2 types of treatment:

§ Heating Treatment (HT): dunnageshould be treated at a minimum temperature of 56° C for at least 30 minutes.

§ Methyl Bromide (MB) Fumigation: dunnage must be fumigated with Methyl Bromide

NOTE: ISPM 15 refer to all types of wood from coniferous wood to hardwood

Page 16: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.16

ISPM 15

Page 17: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.17

STATUS OF FUMIGATION

• This leads to an increase in fumigated containers worldwide

• It has to be fumigated with Methyl Bromide, which should be prohibited (at the moment no other gas is recommended)

Page 18: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.18

Exposure Limits OSHA PEL 20 ppm, 80 mg/m³ACGIH TLV: 1 ppm TWANIOSH IDLH: 250 ppm

METHYL BROMIDE

Page 19: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.19

METHYLBROMIDE

• Methyl Bromide 0.2/a 0.2 - 8 ppm • Methyl Bromide 0.5/a 0,5 - 30 ppm • Methyl Bromide 3/a 3 - 100 ppm • Methyl Bromide 5/b 5 - 50 ppm• Petroleum Hydrocarbons 100/a 500 - 2,500 ppm

Page 20: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.20

SO2F2 AS A SUCCESSOR OF METHYL BROMIDE

• Exposure Limits • OSHA 5 ppm, 20 mg/m3 TWA • (ACGIH) TLV: 5 ppm, 21 mg/m3 TWA

10 ppm, 42 mg/m3 STEL • NIOSH 5 ppm, 20 mg/m3 TWA

10 ppm, 40 mg/m3 STEL IDLH 200 ppm

• LEL: Noncombustible Liquid

Page 21: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.21

SO2F2 AS A SUCCESSOR OF METHYL BROMIDE

• Odorless Gas• Very toxic• In use since 1950• 3 production places worldwide• Sold by

and Solvay

Page 22: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.22

New Tube Sulphuryl Fluoride 1/a

•Detection limit: 0.8 ppm (!!!)•Scale: 1,3,5 ppm•Part number: 8103471First tube in this detection range, replaces the old tubeCheapest detection system for Sulfuryl Fluoride

NEW TUBESSULPHURYL FLUORIDE 1/A

Page 23: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.23

• NIOSH 0.1 ppm (0.7 mg/m3) TWA • OSHA PEL: 0.1 ppm (0.7 mg/m3) TWA • Original IDLH: 4 ppm • LEL: Noncombustible Liquid • Description of Substance:

– Colorless to faint-yellow– oil liquid– intensely irritating odor.

CHLOROPICRIN AS A SUCCESSOR OF METHYL BROMIDE

Page 24: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.24

New Tube Chloropicrin 0,1/a

•Detection limit: 0,1 ppm•Range: 0,1-2 ppm•Part number: 8103421

•Chloropicrin is one of the recommended successors for Methyl Bromide•Already in use as an odorant for Methyl Bromide and Sulfuryl Fluoride

NEW TUBESCHLOROPICRIN 1/a

Page 25: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.25

PHOSPHINE

A large range of tubes are offered for different applications:

Phosphine 0.01/a 0.1 - 1 ppmPhosphine 0.1/a 0.1 - 4 ppm Phosphine 0.1/b 1 - 15 ppmPhosphine 1/a 1 - 100 ppm Phosphine 25/a 25 -10,000 ppmPhosphine 50/a 50 - 1,000 ppm

PhosphineCurrent TLV = 0.3 ppm, 1ppm STELLab Range = 0.3 ppm – 3.5 ppmIDLH = 200 ppm

Problem:

Resistances are coming up

More and more insect are notaffectable with phosphine

Page 26: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.26

GASES WORLDWIDE ON TEST

• Ethyl Formate• Ethylen Dibromide• Dinitrile• Carbonylsulfide• MITC• More will follow…………………..

Page 27: GAS DETECTION TUBES FOR FUMIGANTS - AIHA Home

DRÄGER SAFETY • HYGIENE CONFERENCE CHICAGO USA

18.05.2006 • FRANK ARNOLD • SLIDE NO.27

THANK YOU FOR YOUR ATTENTION!