+ All Categories
Home > Documents > HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and...

HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and...

Date post: 20-Mar-2021
Category:
Upload: others
View: 5 times
Download: 0 times
Share this document with a friend
38
HotBlock200 Digestion Systems Operation & Instruction Manual 800-343-5319 or 843-881-6560 www.environmentalexpress.com ©2018 Environmental Express, Inc. All rights reserved. August 2018 1065EE5_MAN.1
Transcript
Page 1: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

HotBlock™ 200 Digestion Systems Operation & Instruction Manual

800-343-5319 or 843-881-6560 www.environmentalexpress.com©2018 Environmental Express, Inc. All rights reserved.

August 2018 1065EE5_MAN.1

Page 2: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

HotBlock™ 200: Operation and Instruction Manual Contents

Contents General Information

Warranty and Repair Policy 1 Declaration of Conformity 2

About Your HotBlock™ 200 Part Numbers / Model Information / Specifications 3

Getting Started Installation / Safety / Maintenance 4-6Operation Using the HotBlock™ 200 Controller 7

Parts and Supplies Accessories 8

Adaptation of EPA Methods for Use with Your HotBlockTM

Method 200.2, Revision 2.8 9-10Method 200.7, Revision 4.4 11-12Method 200.8 13-14Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16Method 365.1 for Phosphorus 17Method 3050B 18-19NIOSH Method 7303 19-20Sample Preparation for Lead Analysis using the GhostWipe® Wipes 20-21Method 3060A for use with the HotBlock™ and StirBase™ Systems 22-23

Appendix A:

A1-A13Programming your HotBlock Controller

Page 3: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

HotBlock™ 200: Operation and Instruction Manual Warranty and Repair Policy

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 1

Limited Warranty

The Environmental Express HotBlock™ 200 is warranted against defects in materials and workmanship when used in accordance with applicable instructions, for a period of one year from the date of shipment. This warranty extends to parts, labor, and any approved transportation charges. This warranty applies only to damage or failure caused by normal laboratory use. The warranty is limited to product repair. If Environmental Express is unable to repair the HotBlock™ 200, the customer may, at his or her option, receive a replacement unit or a full refund. Operating the HotBlock™ 200 at temperatures higher than 200°C will void the warranty. In no event shall Environmental Express have any obligation to make repairs, replacements or corrections required, in whole or in part, as the result of (i) normal wear and tear, (ii) accident, disaster or event of force majeure, (iii) abuse, neglect, misuse, fault or negligence of or by customer, (iv) use of the product in a manner for which it was not designed, (v) causes external to the product such as, but not limited to, power failure or electrical power surges, (vi) improper storage and handling of the product, (vii) use of the product in combination with equipment or software not supplied by Environmental Express, (viii) ordinary maintenance, (ix) alterations, repairs or installations that have not been performed by Environmental Express or its authorized representative or (x) failure to maintain product in accordance with Environmental Express’ written instructions. Environmental Express makes no other warranty, expressed or implied for this product with respect to merchantability, fitness for a particular use or any other matter and expressly disclaims all other warranties. Environmental Express is not liable for any consequential, special, indirect or compensatory damages arising from use of, or in conjunction with this product. the maximum liability of Environmental Express (whether by reason of breach of contract, tort, indemnification, or otherwise, but excluding liability of seller for breach of warranty (the sole remedy for which shall be as otherwise provided herein)) shall be the invoice price of this product.

Repair Policies

Under Warranty Repair: If the HotBlock™ 200 should fail to operate as warranted within the warranty period (one year from date of shipment), Environmental Express will repair it and ship it back to the customer at Environmental Express’ expense. The remainder of the warranty period will be honored from the original ship date. Environmental Express will bear the cost of ground transportation both to and from the customer’s location, and bear the cost of any parts, labor and cleanup required. However, if it is determined that the damage to the HotBlock™ 200 was caused by negligence or improper use or by another excluded cause as set forth above, this warranty will not apply. The warranty is also void if the system is used beyond its intended purpose or in the event of any unauthorized repair. In such cases, reasonable and customary repair charges will apply. Repair charges will be quoted prior to work being done.

Out of Warranty Repair: If the HotBlock™ 200 fails after the warranty period has lapsed, the repair procedure is as follows: First, notify an Environmental Express Technical Service Representative of product’s failure and place an order for repair. Whenever possible, our customer service technician will walk you through possible troubleshooting scenarios which may enable you to repair your block on-site. If on-site repair is not possible, the customer may return the non-working unit to Environmental Express using appropriate shipping containers and insurance. Repair charges will be assessed and estimated prior to work being done. Repair charges will include all freight costs as well as reasonable and customary charges for parts and labor.

Note: This warranty does not apply to any consumable items associated with the HotBlock™ 200 system.

Loaner HotBlock™ 200 system MAY be available during the repair period. There are only a limited number of these units. A reasonable charge for “cleanup” will be charged if a loaner is issued. The customer will be responsible for all shipping charges associated with a loaner unit.

Page 4: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

HotBlock™ 200: Operation and Instruction Manual Declaration of Conformity

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 2

Call 800-343-5319 or 843-881-6560 www.environmentalexpress.com 2345 A Charleston Regional Pkwy • Charleston, SC 29492

The manufacturer, Environmental Express, 2345A Charleston Regional Parkway, Charleston, SC 29492 declares that the following products, HotBlock™ Catalog Numbers, SSC2015-96, SC2015-96V240, SC2050-36, SC2050-36V240, SC2050-54, SC2050-54V240, SC2100-25, SC2100-25V240, SC2100-35, and SC2100-35V240 are in conformity with:

Standard for Safety Electrical Equipment for Measurement, Control, and Laboratory Use; Part 1 General Requirements, UL 61010-1, CAN/CSA-C22.2 No. 61010-1, 2nd Edition, Issued 12 July, 2004 with revisions through and including 28 October, 2008; Equipment for Measurement, Control, and Laboratory Use Part 2-010: Particular Requirements for Laboratory Equipment for the Heating of Materials, IEY 61010-2-010, 2nd Edition, Issued 1 June, 2003, Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use - Part 2-010:Particular Requirements for Laboratory Equipment for the Heating of Materials, CSA C22.2.61010.2.01

Environmental Express, Inc. declares that all HotBlocks conform with the essential requirements of the applicable EC directives.

Signed: ________________________________________

David Shepard, General Manager

4000040

Page 5: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 3

HotBlock™ 200: Operation and Instruction Manual Part Numbers / Model Information / Specifications

The HotBlock 200 Series, offers programmable metals digestions with touch-screen remote control.

Above, the SC2050-54 HotBlock 200 is shown. The controllers (shown right) can be installed either via magnets on the outside of the hood (above) or installed in an available electrical connection in the hood (below).

The separate controller can be used to operate one (SC1831) or two (SC1832) blocks. See individual block descriptions below. HotBlock 200 components are manufactured with the same rigorous specifications as our original HotBlocks. Polycarbonate transfer racks are included.

HotBlock 200 Heating Block Specifications:

Temperature Range: to 200°C

Temperature Resolution: 0.1°C

Temperature Uniformity: ±0.2°C

Electrical: 110V*, 9A per block

Internal Thermocouple: RTD

All HotBlocks are also available in 240V.

Number of Digestion Description Volume Samples Cup Model Racks Included Footprint Catalog #

HotBlock 200 Controller

(networks any two HotBlock 200 blocks) — — — — — SC1831/

SC1832

36-well HotBlock 200 Heating Block 50mL 36 SC475 2, 18-place 15"W x 15"L SC2050-36

54-well HotBlock 200 Heating Block 50mL 54 SC475 3, 18-place 15"W x 21.5"L

SC2050-54

96-well HotBlock 200 Heating Block 15mL 96 SC415 3, 32-place 15"W x 21.5"L

SC2015-96

25-well HotBlock 200 Heating Block 100mL 25 SC490 1, 10-place; 1, 15-place 15"W x 15"L SC2100-25

35-well HotBlock 200 Heating Block 100mL 35 SC490 2, 10-place; 1, 15-place 15"W x 21.5"L

SC2100-35

Page 6: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 4

HotBlock™ 200: Operation and Instruction Manual Installation, Safety & Maintenance

About Your Environmental Express HotBlock™ 200 Series

Environmental Express HotBlock 200 Series provides an efficient method of digesting and storing water, wastewater, soil and sludge samples for metals analysis. These innovative digestion systems allow samples to be digested in a corrosion-free environment using an external touch screen controller. Samples are handled in a small area with minimal radiant heat loss. Users should be aware of potential dangers from heating certain types of compounds. Such hazards may include explosion or the release of toxic or flammable gases.

Definitions/Markings

Each HotBlock™ 200 displays certain markings and symbols. All personnel working with the HotBlock™ 200 should have an understanding of the following symbols and definitions:

• V = voltage• ~ = alternating current• Hz = frequency• A = amperes

This symbol means Caution Hot Surface. The surface of the HotBlock™ may be too hot to safely touch with bare hands.

This symbol means Read and become familiar with instructions before operation of instrument.

Installation of your HotBlock™ 200:

1. Plug the controller into an appropriate outlet. The HotBlockTM 200 must be plugged into a 15 or 20 AMP receptacle. Thecontrollers can be mounted to the side of a metal hood using the included magnets. The black controller mountingplate can also be removed from the housing and installed in an empty electric receptacle on the fume hood.Ensure installation in a receptacle is done by a qualified electrician.

2. The cord running from the HotBlock™ plugs into the corresponding receptacle on the front of the controller. If usingtwo blocks, you must use SC1832 to have two receptacles. Attach the cord from the second block to theappropriate receptacle.

3. The HotBlock or blocks should be placed in a vented enclosure while the controller must remain outside theenclosure for ease of operation and to keep the controller removed from the corrosive digestion environment. (Seeinstallation requirements in the next section).

Failure to locate your HotBlock™ 200 controller outside the fume hood will void your product warranty.

Installation Requirements

Locate the HotBlock™ 200 under a fume hood with a minimum face velocity of 100fpm, and allow a minimum of 2" of space on all sides. The following environmental conditions should be observed:

• Ambient temperature range: 5-45°C• Ambient relative humidity: 0-100%RH• Altitude: sea level to 2500 meters

HotBlock 200 is rated as Pollution Degree 2 and Installation Category 2.

The HotBlock 200 Controller MUST be located outside the fume hood. Failure to do so will void your product warranty.

Page 7: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 5

HotBlock™ 200: Operation and Instruction Manual Installation, Safety & Maintenance

Electrical Requirements

• Required Voltage 120 volts, ~60Hz, Current - 9A per block (all HotBlock™ 200 systems are alsoavailable in 240V)

• Power should not vary greater than ±10%. Connecting to the power supply must be done with the supplied heat-resistantpower cord or equivalent.

• For safety reasons, a separate power receptacle should be provided for HotBlock™ 200 system. Do not use extension cordsor outlet adaptors. Make certain that power outlets are earth-grounded at the grounding pin.

Potential Hazards

The HotBlock™ 200 should only be operated by properly trained personnel using standard laboratory safety practices.

• Use extreme caution when operating the HotBlock™ 200 digestion blocks. Plastic and graphite surfaces of the blocks maybe too hot to safely touch with bare hands.

• The HotBlock™ 200 components contain electrical circuits and devices and compounds operating at dangerous voltages.Contact with these circuits, devices and components can cause serious injury or painful electric shock.

• Proper grounding is essential to avoid a potentially serious electric shock hazard. Ensure that there is an internal groundconnection between the metal base of the system and the 3-pin earth-grounded receptacle.

• For safety reasons a separate power outlet receptacle should be provided for each HotBlock ™ 200 system. Do not useextension cords or outlet adaptors. Make certain each power outlet is earth-grounded at the grounding pin.

• Power requirements for individual blocks may be found on our website.

• Application of the wrong supply voltage can create a fire hazard and a potentially serious shock hazard, and could seriouslydamage the HotBlock™ 200 system. See specifications for individual components.

• Users should be aware of potential dangers from heating certain types of compounds. Such dangers may include therelease of toxic or flammable gases or explosion.

Maintenance

• Any service inquiries should be directed to Environmental Express Inc. Technical Service Department.

• After each use, clean exterior surfaces with a damp sponge to remove acid residue.

• For acid spills, sponge with a diluted solution of sodium bicarbonate followed by distilled water. Acid that is spilleddirectly into the digestion wells should be neutralized and removed.

• Before using any cleaning or decontamination methods except those recommended, check with Environmental Expressto confirm the proposed method will not damage your HotBlock™ 200.

• Avoid excessive spills, as liquid allowed to overflow into the HotBlock™ 200 casing can seriously damage electroniccomponents.

Page 8: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 6

HotBlock™ 200: Operation and Instruction Manual Installation, Safety & Maintenance

HotBlockTM Temperature Settings:

The pre-set factory “set point” temperature of your HotBlockTM is 106°C. Factory tests have shown that this temperature is “sea level safe.” Liquids in uncovered polypropylene tubes should not boil at this setting. Please note that the set point of the block is not the same as the temperature of the liquids being digested. The block temperature should be optimized for the specific digestion. The temperature of liquid contents of the digestion cup will vary according to:

• The material being digested

• The number of samples being digested

• The air movement of the digestion area

• The addition of a watch glass or reflux cap

If watch glasses are being used, the control point temperature should be lowered approximately 10°C to avoid boiling.

Note The maximum sample temperature tolerance for our polypropylene digestion vessels is 130°C.

Remember that the temperature display (current block temperature) is not the temperature of the sample. Sample temperature will usually be 10-15° less than the display temperature.

Safe-Sample™ Temperature Protection:

Your HotBlockTM is protected from runaway temperatures by a fail-safe alarm system. In the unlikely event that the heating system fails to respond to the controller, the Safe-Sample™ system will automatically shut the system off and display a visible alarm.

This alarm sequence occurs if the actual temperature of the block reaches a temperature that is fifteen degrees higher than the set point temperature. If this should occur, the HotBlockTM will stop heating, preventing the loss of samples. The HotBlockTM must be turned off, then turned back on to reset the alarm.

Page 9: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 7

HotBlock™ 200: Operation and Instruction Manual Operation

Easy, Intuitive Operation Using the HotBlock™ 200 Controller The HotBlock™ 200 has two options for the controller, catalog number SC1831 provides operational

control of one block or catalog number SC1832 provides operation control of two blocks simultaneously. The two blocks may be any of the HotBlock™ 200 series.

To change the set point of the HotBlock press the up or down keys. The temperature may be changed at any time that a profile is not running. Be cautious when lowering the set point of the HotBlock. If the set point is lowered to a temperature more than 15°C below the current HotBlock temperature an alarm will display and power to the block will be cut off.

To use the ramp (have the HotBlock™ increase the temperature to a new set point) and soak (hold the temperature at a set point for a specified length of time) features you must create a profile for the controller to follow. Follow the instructions below to create a profile for operating the HotBlock™ in an automated manner.

1. From the home screen on the controller press and hold the Advance Key for approximately fiveseconds. The profile prompt will appear in the lower display and the profile number (e.g. ) will appear in the upper display.

2. Press the Up or Down key to change to another profile if needed (1 to 4).

3. Press the Advance key to move to the selected profile’s first step.4. Press the Up or Down key to move through and select the step type.

5. Press the Advance key to move through the selected step settings.6. Press the Up or Down key to change the step settings.7. Press the Infinity Key at any time to return to the step number prompt.8. Press the Infinity Key again to return to the profile number prompt.9. From any point press and hold the Infinity Key for two seconds to return to the home screen.

To start a profile, follow the instructions below.

1. From the home screen on the controller press and hold the Advance Key to locate the Profile Start

and select the file or step number to start. The upper display will show and the lower display willshow .

2. Press the Up or Down key to choose the file or step number.

3. Press the Advance key to select the Profile Action Request. The upper display will showand the lower display will show .

4. Press the Up or Down key to select Profile Start. The upper display will show and the lower display will show .

5. Press the Infinity Key to return Home. The Profile will start.

To end a profile before the scheduled completion, follow the instructions below.

1. Press the Advance key to select the Profile Action Request. The upper display will showand the lower display will show .

2. Press the Up or Down key to select End. The upper display will show and the lower display will show .

3. Press the Infinity Key to return Home. The Profile will end.

Page 10: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 8

HotBlock™ 200: Operation and Instruction Manual Parts & Supplies

Polypropylene Disposable Screw-cap Digestion Cups Eliminate Transfer Steps.

• Available in 15mL, 50mL and 100mL volumes

• Certified for low metals background

• Volume certified

• Lined caps ensure a leak-proof seal

• Accurate graduations eliminate the need for graduated cylinders and transfer steps

These carefully engineered digestion vessels provide premium performance in your HotBlock®, HotBlock™ 200. Cups are molded of clarified homopolymer polypropylene, providing a higher working temperature and greater chemical resistance than commonly used co-polymer polypropylene.

Accurate, molded-in graduations allow analysts to easily reproduce volumes to within 0.5%. Vessels serve as graduated cylinders, as well as storage containers. After digestion is complete, the sample can be brought to volume and capped for storage, eliminating time-consuming transfer steps. The result is a “one-cup” system that greatly reduces labor and costs associated with metals preparation.

Screw caps are available lined with polyethylene-faced foam liners for a leak-proof seal (green and white caps only) or as a plug style closure (natural caps only).

Each box of cups comes with a certification of volume accuracy, as well as a Metals Background Analysis Report.

Digestion Cup Specifications:

Description Total Capacity Graduations Quantity Catalog #

15mL Cup (clear cap, no liner) 18mL to 15mL 1000 SC415

50mL Cup (green cap, liner) 68mL to 50mL 500 UC475-GN

50mL Cup (white cap, liner) 68mL to 50mL 500 UC475-W

50mL Cup (natural cap, no liner)

68mL to 50mL 500 UC475-NL

50 mL Ultimate Clean cup (natural cap, no liner)

68mL To 50mL 150 UCC000-50

100mL Cup (white cap, liner) 125mL to 100mL 225 SC490

Ribbed Watch Glasses allow partial reflux during digestion.

Ribbed watch glasses are available to fit all Environmental Express digestion vessels. Polypropylene construction is naturally free of contaminants. Watch glasses have a center stem to aid in handling and decrease the potential for contamination.

For a complete list of consumables and accessories, visit the Metals Digestions section of our website.

Description Digestion Cup Quantity Catalog #

19.5mm Watch Glass SC415 1000 SC417

44mm Watch Glass SC475/SC480-W 1000 SC505 52mm Watch Glass SC490 500 SC610

Page 11: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 9

HotBlock™ 200: Operation and Instruction Manual EPA Method 200.2, Revision 2.8

Adaptation for EPA Method 200.2, Revision 2.8 for use with the Environmental Express HotBlock® Digestion System Revised 12.13

1.0 Scope and Application 1.1 The following procedures have been written as an aid to EPA digestion

procedure 200.2 for use with the Environmental Express HotBlock. EPA Method 200.2 is for the determination of total recoverable analytes in groundwater, surface water, drinking water, wastewater and (with the exception of silica) in solid samples such as sediment, sludge and soil. Use EPA Method 200.2 for reference while following the sample preparation steps outlined below. A complete list of elements appropriate for analysis is included in EPA Method 200.2. Analysis can be performed by flame atomic absorption, GF/AA, ICP and ICP-MS.

2.0 Apparatus and Materials 2.1 HotBlock for Metals Digestions — Model Number SC100, SC150*, SC154,

SC181, SC182*, SC189, SC191, or SC192* 2.2 Polypropylene Digestion Vessels — Catalog # SC475 (50mL) or # SC490

(100mL) or # SC415 (15mL) 2.3 Ribbed Watch Glasses, Catalog # SC505 or Reflux Caps, Catalog # SC506

(for use with SC490 vials for HotBlock SC150, use Ribbed Watch Glasses, Catalog # SC610)

2.4 FilterMate — Catalog # SC0401 (or appropriate FilterMate® or FlipMate®) for sample filtration if necessary *Note: For all procedures, when using the block with the SC490 digestion vials with

a 100mL sample, double the volume of all reagents and acids added.

3.0 Procedure, Aqueous Sample Preparation 3.1 Mix sample thoroughly to achieve homogeneity. For each digestion

procedure, transfer a 50mL sample (or appropriate volume for your lab) into the SC475 polypropylene vessel. For best results, weigh the sample directly in the vessel on a tared balance.

3.2 Add 1mL (1:1) HNO3 and 0.5mL (1:1) HCl, then swirl. Heat the sample in the HotBlock at 95°C for 2.5 hours without boiling. Place a ribbed watch glass (SC505) on top of the digestion vessel. Note: When using a watch glass, adjust the control point temperature of the HotBlock so that a 50mL, 5% acid solution is heated to 85°C BEFORE placing the watch glass over the sample. Laboratory tests have proven that the addition of the ribbed watch glass will add approximately 10°C to the sample temperature, bringing the sample temperature up to 95°C.

3.3 Using the polycarbonate transfer racks, remove samples from the HotBlock and allow them to cool.

3.4 After cooling, dilute samples to 50mL with DI Water. 3.5 If necessary, filter with SC0401 (or appropriate FilterMate® or FlipMate®)

to remove insoluble material. Note: The filtration step should be performed slowly with little pressure placed on the plunger. If excessive back pressure occurs, stop filtration and allow sediments to “settle out”. Applying pressure to the plunger may cause sample “blow through” allowing sediment to pass through the filter into the digestate.

Page 12: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 10

HotBlock™ 200: Operation and Instruction Manual EPA Method 200.2, Revision 2.8

4.0 Procedure, Solids Sample Preparation 4.1 Sieve a dried sample using a 5-mesh polypropylene sieve and grind in a

mortar and pestle. Weigh a representative sample of 0.5 ±0.01g into a digestion vessel.

4.2 Add 2mL of (1+1) HNO3 and 5mL of (1+4) HCl. Cover with a ribbed watch glass (SC505) or reflux cap (SC506) and place in the HotBlock at 95°C.

4.3 Heat sample for 30 minutes. 4.4 (Optional) Although step 3 is the final heating step for EPA Method 200.2,

if the sample is suspected of having a high concentration of organic compounds, it is recommended to complete this step: Add 2mL of 30% H2O2 to the well-cooled sample. Allow the exothermic reaction to occur (approximately 10 minutes) and place the sample back in the HotBlock at a temperature of 10° less than the original set point for an additional 30 minutes. The reaction with the H2O2 raises the sample temperature. Boiling should not occur if the temperature of the HotBlock is lowered. Note: H2O2 helps aid in the breakdown of high organic compounds in the sample thus creating a more complete digestion.

4.5 Using the polycarbonate transfer racks, remove samples from the HotBlock and allow them to cool completely.

4.6 Bring sample volume to 50mL with DI water. 4.7 If necessary, filter with SC0401 (or appropriate FilterMate® or FlipMate®)

to remove insoluble material. Note: The filtration step should be performed slowly with little pressure placed on the plunger. If excessive back pressure occurs, stop filtration and allow sediments to “settle out”. Applying pressure to the plunger may cause sample “blow through” allowing sediment to pass through the filter into the digestate.

All QC samples, concentration limitations, elemental lists and reagent specifications are addressed in depth in EPA Method 200.2. Safety concerns are also part of the full method. Follow the instructions listed in EPA Method 200.2. These steps should only be used as a guide to help improve the performance of your HotBlock.

Page 13: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 11

HotBlock™ 200: Operation and Instruction Manual EPA Method 200.7, Revision 4.4

Adaptation for EPA Method 200.7, Revision 4.4 for use with the Environmental Express HotBlock Digestion System Revised 12.13

1.0 Scope and Application 1.1 The following procedures have been written as an aid to EPA digestion

procedure 200.7 for use with the Environmental Express HotBlock. EPA Method 200.7 is for the determination of total recoverable analytes in groundwater, surface water, drinking water, wastewater, and (with the exception of silica) in solid samples such as sediment, sludge and soil. Use EPA Method 200.7 for reference while following the sample preparation steps outlined below. A complete list of elements appropriate for analysis is included in EPA Method 200.7. Analysis can be performed by ICP.

2.0 Apparatus and Materials 2.1 HotBlock for Metals Digestions — Model Number SC100, SC150*, SC154,

SC181, SC182*, SC189, SC191, or SC192* 2.2 Polypropylene Digestion Vessels — Catalog # SC475 (50mL) or # SC490

(100mL) or # SC415 (15mL) 2.3 Ribbed Watch Glasses, Catalog # SC505 or Reflux Caps, Catalog # SC506

(for use with SC490 vials for HotBlock SC150, use Ribbed Watch Glasses, Catalog # SC610)

2.4 FilterMate — Catalog # SC0401 (or appropriate FilterMate® or FlipMate®) for sample filtration if necessary *Note: For all procedures, when using the block with the SC490 digestion vials with

a 100mL sample, double the volume of all reagents and acids added.

3.0 Procedure, Total Recoverable Elements, Aqueous Sample (EPA Method 200.7, Paragraph 11.2) 3.1 Mix sample thoroughly to achieve homogeneity. For each digestion

procedure, transfer a 50mL sample (or appropriate volume for your lab) into the SC475 polypropylene vessel. For best results, weigh the sample directly in the vessel on a tared balance.

3.2 Add 1mL 1:1 HNO3 and 0.5mL 1:1 HCl and swirl. Heat in the HotBlock at a sample temperature of 85°C and evaporate to about 10mL taking care that the cup does not go dry. This will take approximately 2 hours.

3.3 Cover the vessel with a disposable watch glass (catalog # SC505) and reflux an additional 30 minutes. Note: When using a watch glass, adjust the set point temperature of the HotBlock so that a 50mL, 5% acid solution is heated to 85°C BEFORE placing the watch glass over the sample. Laboratory tests have proven that the addition of the ribbed watch glass will add approximately 10°C to the sample temperature, bringing the sample temperature up to 95°C.

3.4 Using the polycarbonate transfer racks, remove samples from the HotBlock and allow them to cool. Bring samples up to volume with DI water.

Page 14: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 12

HotBlock™ 200: Operation and Instruction Manual EPA Method 200.7, Revision 4.4

3.5 If necessary, filter with SC0401 (or appropriate FilterMate® or FlipMate®) to remove insoluble material. Note: The filtration step should be performed slowly with little pressure placed on the plunger. If excessive back pressure occurs stop filtration and allow sediments to “settle out”. Applying pressure to the plunger may cause sample “blow through” allowing sediment to pass through the filter into the digestate.

4.0 Procedure, Total Recoverable Elements, Solids Sample (EPA Method 200.7, Paragraph 11.3) 4.1 For each digestion procedure, transfer 0.5g ± 0.01g of a dried sieved

sample into the SC475 polypropylene vessel. 4.2 Add 2.0mL (1:1) HNO3 and 5mL (1:4) HCl and swirl. Cover the vessel with

a disposable watch glass (catalog # SC505) and reflux an additional 30 minutes at a sample temperature of 85°C. Note: When using a watch glass, adjust the set point temperature of the HotBlock so that a 50mL, 5% acid solution is heated to 85°C BEFORE placing the watch glass over the sample. Laboratory tests have proven that the addition of the ribbed watch glass will add approximately 10°C to the sample temperature, bringing the sample temperature up to 95°C.

4.3 Allow to cool and bring to 50mL volume with DI water. 4.4 Filter with SC0401 (or appropriate FilterMate® or FlipMate®) to remove

insoluble material.

All QC samples, concentration limitations, elemental lists and reagent specifications are addressed in depth in EPA Method 200.7. Safety concerns are also part of the full method. Follow the instructions listed in EPA Method 200.7. These steps should only be used as a guide to help improve the performance of your HotBlock.

Page 15: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 13

HotBlock™ 200: Operation and Instruction Manual EPA Method 200.8

Adaptation for EPA Method 200.8, for use with the Environmental Express HotBlock Digestion System Revised 12.13

1.0 Scope and Application 1.1 The following procedures have been written as an aid to EPA digestion

procedure 200.8 for use with the Environmental Express HotBlock. EPA Method 200.8 is for the determination of total recoverable analytes in groundwater, surface water, drinking water, wastewater, and (with the exception of silica) in solid samples such as sediment, sludge and soil. Use EPA Method 200.8 for reference while following the sample preparation steps outlined below. A complete list of elements appropriate for analysis is included in EPA Method 200.8. Analysis can be performed by ICP-MS.

2.0 Apparatus and Materials 2.1 HotBlock for Metals Digestions — Model Number SC100, SC150*, SC154,

SC181, SC182*, SC189, SC191, or SC192* 2.2 Polypropylene Digestion Vessels — Catalog # SC475 (50mL) or # SC490

(100mL) or # SC415 (15mL) 2.3 Ribbed Watch Glasses, Catalog # SC505 or Reflux Caps, Catalog # SC506 (for

use with SC490 vials for HotBlock SC150, use Ribbed Watch Glasses, Catalog # SC610)

2.4 FilterMate — Catalog # SC0401 (or appropriate FilterMate® or FlipMate®) for sample filtration if necessary *Note: For all procedures, when using the block with the SC490 digestion vials with

a 100mL sample, double the volume of all reagents and acids added.

3.0 Procedure, Aqueous Sample Preparation -- Dissolved Analytes (EPA Method 200.8, Paragraph 11.1) 3.1 For the determination of dissolved analytes in ground and surface waters,

pipet a 20mL or greater aliquot of filtered, acid-preserved sample into the SC475 digestion vessel.

3.2 Add an appropriate volume of (1+1) HNO3 to adjust the acid concentration of the aliquot to approximate a 1% (v/v) nitric acid solution. If the direct addition procedure is being used, add internal standards, cap and mix. Note: If a precipitate is formed during acidification, transport or storage the sample aliquot must be treated using procedures for Total Recoverable Analytes.

4.0 Procedure, Aqueous Sample Preparation — Total Recoverable Analytes (EPA Method 200.8, Paragraph 11.2) Note: This section applies to water samples containing turbidity of greater than 1 NTU.

4.1 For each digestion procedure, transfer 50mL of well-mixed, unfiltered, acid-preserved sample into the SC475 polypropylene vessel.

4.2 Add 1.0mL (1+1) HNO3 and 0.5mL of (1+1) HCl and swirl. Heat in the HotBlock at a sample temperature of 85°C. The HotBlock set temperature should be approximately 105°C. The temperature of a reference blank should be tested to ensure correct temperature.

Page 16: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 14

HotBlock™ 200: Operation and Instruction Manual EPA Method 200.8

4.3 Reduce the volume of the sample aliquot to about 10mL at this temperature. This should take about 2.5 hours.

4.4 Place a ribbed watch glass (SC505) over the digestion vessel to reduce additional evaporation. Note: When using a watch glass, adjust the set point temperature of the HotBlock so that a 50mL, 5% acid solution is heated to 85°C BEFORE placing the watch glass over the sample. Laboratory tests have proven that the addition of the ribbed watch glass will add approximately 10°C to the sample temperature, bringing the sample temperature up to 95°C.

4.5 Gently reflux the sample for 30 minutes. 4.6 Using the polycarbonate transfer racks, remove samples from the HotBlock

and allow them to cool. 4.7 Add reagent water to bring the sample to the 50mL mark on the digestion

vessel. Cap and mix. 4.8 If necessary, filter with SC0401 (or appropriate FilterMate® or FlipMate®)

to remove insoluble material. Note: The filtration step should be performed slowly with little pressure placed on the plunger. If excessive back pressure occurs, stop filtration and allow sediments to “settle out.” Applying pressure to the plunger may cause sample “blow through” allowing sediment to pass through the filter into the digestate.

4.9 Prior to analysis, adjust the chloride concentration by pipetting 20mL of the prepared solution into another SC475 digestion vessel and bring up to 50mL volume with reagent water. If the dissolved solids in this solution are >0.2%, additional dilution may be necessary to prevent clogging of the extraction and/or skimmer cones. Note the dilution factor for concentration calculations. If the direct addition procedure is being used, add internal standards, cap and mix.

5.0 Procedure, Solid Sample Preparation 5.1 For each digestion procedure, transfer 0.5g ± 0.01g of a dried sieved

sample into the SC475 polypropylene vessel. 5.2 Add 2.0mL (1:1) HNO3 and 5mL (1:4) HCl and swirl. Cover the vessel with

a disposable watch glass (catalog # SC505) and reflux an additional 30 minutes at a sample temperature of 85°C. Note: When using a watch glass, adjust the set point temperature of the HotBlock so that a 50mL, 5% acid solution is heated to 85°C BEFORE placing the watch glass over the sample. Laboratory tests have proven that the addition of the ribbed watch glass will add approximately 10°C to the sample temperature, bringing the sample temperature up to 95°C.

5.3 Allow to cool and bring to 50mL volume with DI water. 5.4 Filter with SC0401 (or appropriate FilterMate) to remove insoluble material.

All QC samples, concentration limitations, elemental lists and reagent specifications are addressed in depth in EPA Method 200.8. Safety concerns are also part of the full method. Follow the instructions listed in EPA Method 200.8. These steps should only be used as a guide to help improve the performance of your HotBlock.

Page 17: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 15

HotBlock™ 200: Operation and Instruction Manual Adaptation of EPA Methods 245.1, 7470, 7471

Adaptation of EPA Methods for Mercury Digestions 245.1, 7470, 7471 for use with the Environmental Express HotBlock Digestion System Revised 12.13

1.0 Scope and Application 1.1 The following procedures have been written as an aid to EPA Digestion

Methods 245.1, 7470 and 7471 for use with the Environmental Express HotBlock. EPA Methods 245.1 and 7470 are for the determination of mercury in aqueous samples and TCLP extracts. EPA Method 7471 is for the determination of mercury in soil, oil and sediment. Use the full EPA Methods for reference while following the sample preparation steps outlined below.

2.0 Apparatus and Materials 2.1 HotBlock for Metals Digestions — Model Number SC100, SC150*, SC154,

SC181, SC182*, SC189, SC191, or SC192* 2.2 Polypropylene Digestion Vessels — Catalog # SC475 (50mL) or # SC490

(100mL) or # SC415 (15mL) 2.3 Ribbed Watch Glasses, Catalog # SC505 or Reflux Caps, Catalog # SC506 (for

use with SC490 vials for HotBlock SC150, use Ribbed Watch Glasses, Catalog # SC610) *Note: For all procedures, when using the block with the SC490 digestion vials with

a 100mL sample, double the volume of all reagents and acids added.

3.0 Procedure, Aqueous Sample Preparation 3.1 Add 20mL of well-mixed sample to a SC475 digestion vessel or 2mL of

TCLP extract and 18mL water. Note: This is a 10x dilution for instruments that can achieve the detection limit including the dilution, otherwise digest 20mL of TCLP extract.

3.2 Add 20mL of standard solutions or appropriate amount of standard spiking solution to give desired concentration when diluted to 20mL.

3.3 Add 0.5mL of concentrated HNO3. 3.4 Add 1.0mL of concentrated H2SO4. 3.5 Mix thoroughly. Add 3mL of 5% KMnO4 and let stand for 15 minutes. If

sample does not maintain purple or brown color, add additional portions to all samples, blanks and standards until the color persists for at least 15 minutes.

3.6 Add 1.6mL of 5% persulfate solution. Place watch glass or reflux cap on top of digestion vessel to allow pressure to vent while minimizing evaporative loss.

3.7 Digest the sample for 2 hours at 95°C. 3.8 Remove samples and let cool to room temperature. 3.9 Add 1.2mL of sodium chloride/hydroxyl amine hydrochloride solution.

Page 18: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 16

HotBlock™ 200: Operation and Instruction Manual Adaptation of EPA Methods 245.1, 7470, 7471

4.0 Procedure, Soil, Oil and Sediments, 7471A: 4.1 Weigh 0.25-0.30g of homogenized sample into a tube. For best results,

weigh the sample directly in the vessel on a tared balance. 4.2 Add 5mL DI water and 5mL aqua regia. 4.3 Heat for 2 minutes at 95 ±3°C. 4.4 Allow samples to cool and add 25mL DI water. 4.5 Add 7.5mL of KMnO4 and let stand for 15 minutes. If sample does not

maintain purple or brown color, add additional portions to all samples, blanks and standards until the color persists for at least 15 minutes. Heat samples at 95 ±3°C for 30 minutes. Note: When using a watch glass, adjust the set point temperature of the HotBlock so that a 50mL sample is heated to 85°C BEFORE placing the watch glass over the sample. Laboratory tests have proven that the addition of the ribbed watch glass will add 10°C to the sample temperature, bringing the sample temperature up to 95°C.

4.6 Let samples cool to room temperature and add 3mL of Sodium Chloride-Hydroxylamine Sulfate solution.

All QC samples, concentration limitations, elemental lists and reagent specifications are addressed in depth in EPA Methods 245.1, 7470 and 7471. Safety concerns are also part of the EPA Methods. These steps should only be used as a guide to help improve the performance of your HotBlock.

Page 19: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 17

HotBlock™ 200: Operation and Instruction Manual Adaptation of EPA Method 365.1

Adaptation of EPA Method 365.1, Determination of Phosphorous by Semi-Automated Colorimetry, for use with the Environmental Express HotBlock Digestion System Revised 12.13

1.0 Scope and Application 1.1 The following procedures have been written as an aid to EPA Method 3651

for use with the Environmental Express HotBlock. EPA Method 365.1 is for the determination of phosphorous in drinking water, surface water and saline water and in domestic and industrial wastes. Use EPA Method 365.1 for reference while following the sample preparation steps outlined below. Range is 0.01 to 0.5 mg/L.

2.0 Apparatus and Materials 2.1 HotBlock for metals digestions — Model Numbers SC100, SC150*, or SC154 2.2 Polypropylene Digestion Vessels — Catalog # SC475 (or SC490 for use with

HotBlock SC150) 2.3 FilterMate — Catalog # SC0401 (or appropriate FilterMate® or FlipMate®)

for sample filtration if necessary *Note: For all procedures, when using the SC150 block with the SC490 digestion vials with

a 100mL sample, double the volume of all reagents and acids added.

3.0 Procedure, Aqueous Sample Preparation 3.1 For each digestion procedure, transfer 50mL of sample (or appropriate

volume for your lab) into the SC475 polypropylene vessel. 3.2 Add 1mL of H2SO4 solution and swirl. (11 N H2SO4 — Slowly add 310mL

concentrated H2SO4 to 600mL distilled water. When cooled dilute to 1 liter.) 3.3 Add 0.4g of ammonium persulfate. 3.4 Heat the sample in the HotBlock at approximately 100°C for 40 minutes.

The sample should slightly boil. Note: Adjust the HotBlock temperature so that the sample obtains a temperature of 100°C.

3.5 Using the polycarbonate transfer racks, remove samples from the HotBlock and allow them to cool.

3.6 After cooling, dilute to 50mL with DI Water. 3.7 If necessary, filter with SC0401 (or appropriate FilterMate® or FlipMate®)

to remove insoluble material. Note: The filtration step should be performed slowly with little pressure placed on the plunger. If excessive back pressure occurs, stop filtration and allow sediments to “settle out.” Applying pressure to the plunger may cause sample “blow through” allowing sediment to pass through the filter into the digestate.

All QC samples, concentration limitations, elemental lists and reagent specifications are addressed in depth in EPA Method 365.1. Safety concerns are also part of the full method. Follow the instructions listed in EPA Method 365.1. These steps should only be used as a guide to help improve the performance of your HotBlock.

Page 20: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 18

HotBlock™ 200: Operation and Instruction Manual Adaptation of EPA Method 3050B

Adaptation of EPA Method 3050B for use with the Environmental Express HotBlock® Digestion System Revised 12.13

1.0 Scope and Application 1.1 The following procedures have been written as an aid to EPA Method

3050B for use with the Environmental Express HotBlock. EPA Method 3050B is for the preparation of sediment, sludge, and soil samples for analysis. Use EPA Method 3050B for reference while following the sample preparation steps outlined below. A complete list of elements appropriate for analysis is included in EPA Method 3050B. Analysis can be performed by flame atomic absorption, GF/AA, ICP and ICP-MS.

2.0 Apparatus and Materials 2.1 HotBlock for Metals Digestions — Model Numbers SC100, SC150*, or

SC154 2.2 Polypropylene Digestion Vessels — Catalog # SC475 (or SC490 for use

with HotBlock SC150) 2.3 Ribbed Watch Glasses, Catalog # SC505 or Reflux Caps, Catalog # SC506

(for use with SC490 vials for HotBlock SC150, use Ribbed Watch Glasses, Catalog # SC610)

2.4 FilterMate® — Catalog # SC0401 (or appropriate FilterMate or FlipMate®) for sample filtration if necessary *Note: When using the SC150 block with the SC490 digestion vials with a 100mL sample,

double the volume of all reagents and acids added.

3.0 Procedure, Sample Preparation 3.1 Mix sample thoroughly to achieve homogeneity and sieve if appropriate.

For each digestion procedure, weigh to the nearest 0.01 gram and transfer a 0.5-1-gram aliquot (or appropriate amount for your lab) to the SC475 polypropylene digestion vessel. For best results, weigh the sample directly in the vessel on a tared balance.

3.2 Add 5mL (1:1) HNO3 + DI Water and swirl. Cover with a ribbed watch glass or reflux cap and heat the sample in the HotBlock at 95±5°C for 15 minutes without boiling. Note: If using the watch glass, adjust the HotBlock temperature so that a 50mL, 5% acid solution is heated to 85°C. Laboratory tests have proven that the addition of the ribbed watch glass will add approximately 10°C to the sample temperature.

3.3 Allow the sample to cool, then add 2.5mL concentrated HNO3 and reflux at 95°C for 30 minutes. Repeat this step until NO brown fumes are given off by the sample.

3.4 Heat sample with the ribbed watch glass to a volume of about 5mL or for 2 hours at 95±5°C. Do not allow the sample to boil or go dry. Neither of these aspects should occur if the temperature of the HotBlock™ is adjusted correctly.

3.5 Using the polycarbonate transfer racks, remove samples from the HotBlock and cool completely (it is very important that samples are cooled completely).

Page 21: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 19

HotBlock™ 200: Operation and Instruction Manual Adaptation of EPA Method 3050B

3.6 Add 1mL DI Water and 1.5mL of 30% H2O2 slowly. Allow an exothermic reaction to occur. Wait 5-10 minutes and place samples back in the HotBlock with the ribbed watch glasses in place. If effervescence starts to occur lift the samples out of the HotBlock and allow the reaction to continue. Do not let the samples foam out of the vessel. (Reducing the HotBlock display temperature by 10°C should reduce the effervescence while maintaining the sample temperature).

3.7 Continue to add 0.5mL of H2O2 (no more than 5mL total) until the sample remains unchanged in color (no longer than 30 minutes). Continue heating for a total of 2 hours or until volume has been reduced to approximately 5mL.

3.8 For the analysis of samples for FLAA or ICP-AES, add 5mL concentrated HCl to each sample and cover with a ribbed watch glass and reflux at 95°C for 15 minutes. For GFAA and ICP-MS digestions, skip this step.

3.9 After cooling, dilute to 50mL with DI Water. 3.10 If necessary, filter with SC0401 (or appropriate FilterMate® or FlipMate®)

to remove insoluble material.

All QC samples, concentration limitations, elemental lists and reagent specifications are addressed in depth in EPA Method 3050B. Safety concerns are also part of the full method. Follow the instructions listed in EPA Method 3050B. These steps should only be used as a guide to help improve the performance of your HotBlock.

Adaptation of NIOSH Method 7303 for use with the Environmental Express HotBlock Digestion System Revised 12.13

1.0 Scope and Application 1.1 The following procedures have been written as an aid to NIOSH Method

7303 for use with the Environmental Express HotBlock. Method 7303 is for the preparation of cellulose ester membrane filters commonly used in air sampling devices. Use EPA Method 7303 for reference while following the sample preparation steps outlined below. A complete list of elements appropriate for analysis is included in Method 7303. Analysis can be performed by ICP or AA.

2.0 Apparatus and Materials 2.1 HotBlock for Metals Digestions — Model Numbers SC100, SC150*, or SC154 2.2 Polypropylene Digestion Vessels — Catalog # SC475 (or SC490 for use with

HotBlock SC150) 2.3 Ribbed Watch Glasses, Catalog # SC505 or Reflux Caps, Catalog # SC506 (for

use with SC490 vials for HotBlock SC150, use Ribbed Watch Glasses, Catalog # SC610)

2.4 FilterMate® — Catalog # SC0401 (or appropriate FilterMate® or FlipMate®) for sample filtration if necessary *Note: When using the SC150 block with the SC490 digestion vials with a 100mL sample,

double the volume of all reagents and acids added.

Page 22: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 20

HotBlock™ 200: Operation and Instruction Manual NIOSH Method 7303

3.0 Procedure, Filter Preparation 3.1 Remove filter from cassette holder and fold into quarters taking care not to

lose any sample. 3.2 Place in SC475 digestion vessel and add 1.25mL HCl. 3.3 Cover with plastic watch glass or reflux cap. Place in HotBlock and heat at

a sample temperature of 95°C for 15 minutes. Note: If using the watch glass, adjust the HotBlock temperature so that a 50mL, 5% acid solution is heated to 85°C. Laboratory tests have proven that the addition of the ribbed watch glass will add approximately 10°C to the sample temperature.

3.4 Remove samples from the HotBlock and cool for 5 minutes. 3.5 Remove watch glass and add 1.25mL HNO3. Replace watch glass and return

to HotBlock at sample temperature of 95°C for 15 minutes. 3.6 Remove the sample from the HotBlock and cool for at least 5 minutes.

Rinse watch glass into vessel. Discard watch glass. 3.7 Dilute to 25mL final volume with distilled, deionized type II water.

All QC samples, concentration limitations, elemental lists and reagent specifications are addressed in depth in Method 7303. Safety concerns are also part of the full method. Follow the instructions listed in Method 7303. These steps should only be used as a guide to help improve the performance of your HotBlock.

Sample Preparation for Lead Analysis with the GhostWipe® Wipes Revised 12.13

Methodology Note: The following procedure has been written as an aid for use with the Environmental Express HotBlock and strictly follows HUD (Housing and Urban Development) guidelines for Lead in Dust Wipes Appendix A-5.0 which references NIOSH 7082.

1.0 Scope and Application 1.1 This application deals with the preparation of samples using lead wipes,

specifically the GhostWipe from Environmental Express. The sample is heated in the presence of Nitric and Hydrochloric acids to dissolve the wipe and all lead compounds into solution. Analysis can be performed by flame atomic absorption (FLAA) or inductively coupled plasma technique (ICP).

2.0 Apparatus and Materials 2.1 HotBlock for Metals Digestions — Model Numbers SC100, SC150*, or SC154 2.2 GhostWipe wipes — Catalog # SC4210 or SC4250 2.3 Polypropylene Digestion Vessels — Catalog # SC475 (SC490 for HotBlock

SC150) 2.4 Ribbed Watch Glasses, Catalog # SC505 or Reflux Caps, Catalog # SC506 (for

use with SC490 vials for HotBlock SC150, use Ribbed Watch Glasses, Catalog # SC610)

Page 23: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 21

HotBlock™ 200: Operation and Instruction Manual Using the GhostWipe

2.5 FilterMate® or FlipMate® — optional if sample does not completely dissolve *Note: For all procedures, when using the SC150 block with the SC490 digestion vials with

a 100mL sample, double the volume of all reagents and acids added.

3.0 Reagents 3.1 Concentrated Reagent-grade HNO3 (5mL repipet dispenser recommended) 3.2 Concentrated Reagent-grade HCl (5mL repipet dispenser recommended)

4.0 Procedure 4.1 Transfer the wipe to the SC475 digestion cup. It is recommended that the

wipe is sent to the sampling area with a cup and wipe so no transfer steps are required. The wipe must be transported in a rigid-walled container according to the sampling procedure ASTM E1728.

4.2 Add 10mL DI Water, 2mL HNO3 and 2mL HCl. Cover with a ribbed watch glass.

4.3 Heat the sample for 45 minutes at 95°C. Note: When using a watch glass, adjust the set point temperature of the HotBlock so that a 50mL solution is heated to 85°C BEFORE placing the watch glass over the sample. Laboratory tests have proven that the addition of the ribbed watch glass will add approximately 10°C to the sample temperature, bringing the sample temperature up to 95°C.

4.4 Using the polycarbonate transfer racks, remove samples from the HotBlock and allow them to cool for approximately 10 minutes. Bring the sample to 50mL volume. If excess amounts of undigested material remain, filter the sample using a 2.0µm PTFE FilterMate®.

4.5 Cap and mix well.

According to AIHA Policies 2001- Revision Nov. 2000, one LCS, one matrix spike and one duplicate must be run every 20 samples. “The LCS shall be a solid matrix material with an established concentration obtained from a source independent of the instrument calibration and traceable to NIST or other similar reference material. Liquid spikes may not be used for preparing LCSs”.

Environmental Express does offer NIST-traceable spiked GhostWipe wipes (Catalog # SC4252). The range is 175 to 200µg of Lead per wipe.

Page 24: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 22

HotBlock™ 200: Operation and Instruction Manual Adaptation of EPA Method 3060A

Adaptation of EPA Method 3060A for use with the HotBlock™ and StirBase™ Systems Revised 12.13

1.0 Scope and Application 1.1 The following procedures have been written as an aid to EPA Method

3060A for use with the Environmental Express HotBlock and StirBase. EPA Method 3060A is an alkaline digestion procedure for extracting hexavalent chromium [Cr(6+)] from soluble, adsorbed, and precipitated forms of chromium compounds in soils, sludges, sediments, and similar waste materials. Use EPA Method 3060A for reference while following the sample preparation steps outlined below. The quantification of Cr6+ in Method 3060A digests should be performed using a suitable technique with appropriate accuracy and precision.

2.0 Apparatus and Materials 2.1 HotBlock for sample digestion – Model Numbers SC100, SC154, SC150,

or SC151 2.2 StirBase stirring device – Catalog # SC160 2.3 Polypropylene Digestion Vessels – Catalog # SC475 (or SC490 for use with

the SC150 or SC151 HotBlocks) 2.4 Ribbed Watch Glasses – Catalog # SC505 (or SC610 for use with the SC150

or SC151 HotBlocks) 2.5 Reflux Caps – Catalog # SC506 (used as an alternative to the SC505 only) 2.6 Stir Bars – Catalog # SC168 2.7 FilterMate – Catalog # SC0407 (or appropriate FilterMate or FlipMate®)

for sample filtration if necessary 2.8 FlipMate – Catalog # SC0301 (or appropriate FilterMate or FlipMate®)

for sample filtration if necessary

3.0 Procedure, Solid Sample Preparation 3.1 Place 2.5 ±0.10 g of the field-moist sample into a clean and labeled

digestion vessel. The sample should have been mixed thoroughly before the aliquot is removed.

3.2 For the specific sample aliquot being spiked, the spike material should be added directly to the sample aliquot at this point.

3.3 Add 50mL±1mL of digestion solution to each sample using a graduated cylinder, and also add approximately 400 mg of magnesium chloride and 0.5mL of 1.0M phosphate buffer. For analytical techniques that can correct for oxidation/reduction of Cr, the addition of Mg(2+) is optional. Cover all samples with watch glasses or reflux caps.

3.4 Stir the samples continuously (unheated) for at least five minutes using the StirBase.

3.5 Heat the samples to 90 - 95ºC, then maintain the samples at 90 - 95ºC for at least 60 minutes with continuous stirring.

3.6 Gradually cool, with continued agitation, each vessel to room temperature.

Page 25: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • 23

Adaptation of EPA Method 3060

3.7 Filter each sample using a 0.45µm FlipMate (SC0601). 3.8 Adjust the pH of the digestate accordingly to the method being used

for analysis. If a precipitate forms, the sample should be filtered as in step 3.7.

3.9 Adjust the sample volume to 100mL with reagent water. Mix well. The sample digestates are new ready to be analyzed.

All QC samples, concentration limitations, interferences, and reagent specifications are addressed in depth in EPA method 3060A. Safety concerns are also part of the full method. Follow the instructions listed in EPA method 3060A. These steps should only be used as a guide to help improve the performance of your HotBlock and StirBaseTM.

HotBlock™ 200: Operation and Instruction Manual

Page 26: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

Appendix A:

Programming Your HotBlock Controller

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • A1

Page 27: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

Navigating the Profiling Page

Note:

Some of these menus and parameters may not appear, depending on the controller's op­tions. See model number information in the Appendix for more information. If there is only one instance of a menu, no submenus will appear.

Profile Setup First, consider some foundational profile setup features that once configured, will apply to all configured profiles. The screen shot below (EZ-ZONE Configurator software) graphically shows the settings (shaded green)that will apply to all profiles; e.g., if Guaranteed Soak is not en­abled here this feature will not be available in any individual profile configuration.

Some of those features that apply to all profiles are listed below with a brief description of their function.

- Ramping Type (Time or Edlt�evlceSettlngaOn-llne-ModelNumber: PM882FC-CCFCCAA Rate) which changes the Click. Menu in the tree ta view ondedit its settings. Click Finish ta sove •nd exit.

profile set point based -· - ----Psramete, Menus rParame=t=en...::S=etu�. G=lo=bal�1�===::;----i-"Pa=ram=et •::;.;r H=•'�-----1

on a set interval of 8 �·z:!PM DioplayUntts IF 3 SelecttheRamplngType

time 0r Set rate. ' r� hlalog� AClinefrequency j60Hz 3 : �i� ' �--�=. ; 1+1 - Value c, Ramping Type !Rate - - · ·•· .:J

- Profile Type (Set Point i : =� ". -,. .. Profile Type .... IPr_oce_••_-_. ·�· ;:J'""'y

or Process) determines l l� CUp.J f. Gllllrenl-SonllEnable 1011 .:J

, �: _, Guorenteed Sook Devort,on 1 Is "F

whether a step (any i : ��NI® GuerenteedSookDeviotion2 Is PIOOOSS

Step Changing the set ! �l Ft.ndiool<.ey Soun:elnstnnceA ,...15 __ _ ' El �1 t Sourcelnst11nceB 16

point) Of a profile Will ' r•: � ,,. •:i' PowerOffTime >-�----; lfl Rea Trr.e Clock

b · b · th ' Conununicll1ions Led Action jBath .:Jegm y usmg e pro- � � Zone Ion 3cess value (Process) or �-Prtlle-

• =�, Channel Ion ;:'.I th l t l d l t

'DisplayPeira j2

e as C ose - oop se Displlly Time 1-0 --

point (Set Point). Us11<SellingsSave !None 3 - Guaranteed Soak En­

able, when set to on makes this feature avail-

User Settings ReslOftl !None 3

Rongo: Not ,l,ppUcable

Use Rom!J1no Tn,e to haYO 1"I ramping sel · po1ntchanaeotasetRate0<...,aset1nte1va1

ornme.

(022036)

able in all profiles. If < fU::.</If?i:. ',c!• .-::::::"�=P"="'::-:-""r-=·''•=· =----------- ------ ----·---Guaranteed Soak Enable � � -is on, use Guaranteed Soak Deviation 1 to 2 to set the value for the corresponding loop. Set the deviation or band above or below the working set point where this condition must be met before the profile can proceed.

Note:

Changes made to profile parameters in the Profiling Pages will be saved and take effect on the next pass through the step. Changes made in the Profile Status page effect the cur­rent step being executed and do not update the step setting in the profiling page. Chang­ing profiles should only be changed by knowledgeable personnel and with caution.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • A2

Page 28: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

Once these global profile features are configured, the next step will require navigation to the Profiling Page. Here, each desired ramp and soak profile will be configured.

To navjgate to the ProfUe Page from the front panel, follow the steps below:

1. From the Home Page, press and hold the Advance Key @ for approximately five seconds.The profile prompt Pro F will appear in the lower display and the profile number (e.g. P /)appears in the upper display.

2. Press the Up o or Down o key to change to another profile (1 to 4).

3. Press the Advance Key @ to move to the selected profiles first step.

4. Press the Up o or Down o keys to move through and select the step type.

5. Press the Advance Key @ to move through the selected step settings.

6. Press the Up o or Down o keys to change the steps settings.

7. Press the Infinity Key � at any time to return to the step number prompt.

8. Press the Infinity Key � again to return to the profile number prompt.

9. From any point press and hold the Infinity Key� for two seconds to return to the HomePage.

If using EZ-ZONE Configurator software, simply click on the plus OickoMenuinthetreetoviewondedititssetlings. Clickfinishtosoveondexit.

sign next to Profiles in - - ----------Parameter Menus Parameters: Profile: Profile 1 Ste 1 ParametM Hel

the left hand column, a ez.zoNEPM r=.;;;..;..:.�-'--':""'-'-';:1=======�::;-- SelectWhlltTypethloC.-wmee e� ���-�� � -

as shown in the screen ti-J Malogqu TorgetSetPointloopl 1200 'F

shot below. ;! :O":.. Torge!SetPointLoop2 i-lo --- Process t•. r,g,o1 l/0 Hours lo

Use St,p Type to selee! wllat this slop will do. The parameter Mst updal&S In a""" secoo<1,; atter 11le SHIP Type setllng 1s cl!anged.

Notice in the screen shot to the right some fields or parameters are not selectable (grayed out) based on the Step Type that is selected.

Starting a Profile There are several ways to start a profile. Some of the examples that follow requires thatcertain optional hard­ware be available on the control. If you are uncertain as to how your control is

t!l Cortrol Ltx,p ,,.i Minutes ,-lo ---[fl 01.tpu If: ,_ Seconds lo I� Mah Rote ;...IO --- 'F 1;: Speciajo.,p., -f� fu1aion � Weit for Process lnstonce 11

b � Glabol 1 Woit For Process 1 ;...!200___ 'F w. Conriuication, Woil Event 1 jOff iJ

l'1 =rneClock WoitEvent2 i-!Off----=3=;

�: Fad. ·""

...,; Dey of Week ! Sunday iJ

8 r Proffel9:� ..

JumpStep 1 1 MOiie 1 Step 1 Jump Count j1 : P!oflo 1 Step 2 ,-----, , Protlo t Step 3 End Type !Off � : P!oflolSlep� Event! jOff iJ '. ::::: Event2 ,-!Off--�i.J..,

: P!oflo t S,,p 7 . Prdle t Step 8 i Pn:He 1 �ep 9 ; Prdlo1Sl,p10

PR&251,p 1¥ Profle.351,p r+J Pn:fle-49:�

•M-SHll>rs,lo..-.ct.anemptystap at can be used to erase a step In lhe protte.

'" • A lime step ramps to a Ta,get set P<lllit o,,e a spedlled lime span and malolalns up to two E....i sllllas fo<tl>e desigoalled Ume.

• A Rate step ramps� process value to tht Tlf!let set PointwllnO<Jt e,ce,,dolg tne Rate, while maiotlining up to two E""'1t outputs_

• A SOIi< slop maintains 1he last Target set Point an<! up to two E""'1t sto1eo for 11\e designated Ume.

• A -lor Event step WIii watt for up to two Walt E"""° 10 be sat!sfled while maintaining up to two E"°"' outputs

----· • A-lor Procese step wtll waltforlhe process w+ue to matdl ttie Watt for Process

kle, while matntalolng up to two E­ouq,uts.

• A Waltfof Procns « E- step WIii W1ttf01 ·­lhe l)rOCl!SS value to match 11\e Walt for Process value, ana wall IOI an1 speetned E­lnpu! conditions to be satlmd.

• A _tor lloUI mp wm waltfor11le process tue to match the Wait for PrOCMs vafue aoc

up to two WIit E- to be satisfied .mHe maintaining up to two E...-.t ou1puts.

....,I R-•noe-, N-01-,..,

-pll-cabl-0------- ���":!"' ... 1:'80:J:n-:,"wattfor a spedfled

CopJL Settin�s ... · I • A Slate step maifltalns conlrol loops and :� ---- ----------- --

equipped, compare the part number of your control to the "Ordering Information" page found in the Appendix of this Users Guide.

Three ways to start a profile:

- Function Key, Digital Input or Profile Request

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • A3

Page 29: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

Configuring the Function Key to Start and Stop a Profile

1. Navigate to the Setup Page and then the Function menu. From the Home Page, press and

hold the o or Down o key for approximately six seconds where the upper display will show

R , and the lower display will show 5 E c .

2. Press the Up o or Down o key to navigate to the Function FU n menu.

3. Press the Advance Key ® to enter this menu. The upper display will show I and the lower

display will show FU n •

4. Press the Advance Key ® to select the level. The upper display will show h , 9 h and the

lower display will show L Eu .

5. Press the Up o or Down o keys to select the level that will start the profile (high or low).

6. Press the Advance Key ® to select the function. In this example, select Profile Start / Stop

P.5 c 5.

7. Press the Advance Key ® to select the function instance (Profile to start).

8. Return to the Home Page by pressing and holding the Infinity Key � for approximately

three seconds.

Note:

The state of the El-Function Key (high or low) is maintained with each successive push of

the key.

Configuring a Digital Input to Start and Stop a Profile

1. Navigate to the Setup Page and then the Digital 1/0 menu. From the Home Page, press andhold the o or Down o key for approximately six seconds where the upper display will show

R , and the lower display will show 5 E c .

2. Press the Up o or Down o key to navigate to the Digital 1/0 menu. Upper display will

show d , a and the lower display will show 5 E c.

3. Press the Advance Key ® where the first available digital instance will be displayed in the

upper display.

4. Press the Up o or Down o key to select the input of choice.

5. Press the Advance Key ® to select the direction (input or output). In this example, select

Dry Contact , [ o n •

6. Select the level (high or low) that will activate the function by pressing the Advance Key ®where the upper display will show h , 9 h and the lower display will show LE u.

7. Press the Up o or Down o keys to select the level that will start the profile (high = closed

or low = open).

8. Press the Advance Key ® to select the function F n. In this example, select Profile Start /

Stop P.5 c 5.

9. Press the Advance Key ® to select the function instance (Profile to start).

10. Return to the Home Page by pressing and holding the Infinity Key � for approximatelythree seconds.

Starting a Profile from the Operations Page

1. Navigate to the Operations Page and then the Profile Status menu. From the Home Page,

press and hold the o or Down o key for approximately three seconds where the upper dis­play will show R , and the lower display will show o PE,-.

2. Press the Up o or Down o key to navigate to the Profile Status P. 5 c R menu.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • A4

Page 30: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

3. Press the Advance Key® to enter this menu. The upper display will show I and the lowerdisplay will show P. St:,.

4. Press the Up o or Down o keys to select the Profile or Step to start. In this example select1.

5. Press the Advance Key ® to select the Profile Action Request. The upper display will shown o n f and the lower display will show P. R [,.

6. Press the Up o or Down o keys to select the Profile start. The upper display will showP, o F and the lower display will show P. R [ ,.

Note:

As soon as the Green Advance Key ® is pressed (step 7 below) the designated Profile or Step (as determined in step 4 above) will start.

7. Press the Advance Key ® to select whether Event 1 will be on or off. The upper displaywill show o F F and the lower display will show En t: I.

Note:

This setting will temporally override the profile configuration.

8. Press the Up o or Down o keys to select whether Event 1 will be on or off. This will imme­diately drive the Event to the specified state regardless of the Profile configuration.

9. Press the Advance Key ® to select whether Event 2 will be on or off. The upper displaywill show o FF and the lower display will show f n t: 2.

10. Press the Up o or Down o keys to select whether Event 2 will be on or off. This will im­mediately drive the Event to the specified state regardless of the Profile configuration.

Note:

The event state will be as left when the profile ended and may be toggled at the profile

status menu.

11. Press the Advance Key ® to see the current Jump Count. The upper display will show Oand the lower display will show J [.

12. Return to the Home Page by pressing and holding the Infinity Key e for approximatelythree seconds.

Ending a Profile from the Operations Page

1. Navigate to the Operations Page and then the Profile Status menu. From the Home Page,press and hold the o or Down o key for approximately three seconds where the upper dis­play will show R , and the lower display will show o PE,·.

2. Press the Up o or Down o key to navigate to the Profile Status P. St: R menu.

3. Press the Advance Key ® to enter this menu. The upper display will show I and the lower

display will show P. St:,.

4. Press the Advance Key ® to select the Profile Action Request. The upper display will shownon E and the lower display will show P. R [ r.

6. Press the Up o or Down o keys to select the End. The upper display will show En d and

the lower display will show P R [,.

7. Press the Advance Key ® to end the Profile.

8. Return to the Home Page by pressing and holding the Infinity Key � for approximatelythree seconds.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • A5

Page 31: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

Starting a Profile from the Home Page

1. When at the Home Page, press the Advance Key ® to locate Profile Start and select thefile or step number to start. The upper display will show I and the lower display will showP.51::: I.

2. Press the Up O or Down O key to choose the file or step number.

3. Press the Advance Key ® to select the Profile Action Request. The upper display will shown o n f and the lower display will show P. R [ i.

4. Press the Up O or Down O keys to select the Profile Start. The upper display will showPro F and the lower display will show P. R [ §.

5. Press the Infinity Key to return Home. The Profile will Start

Ending a Profile from the Home Page

1. Press the Advance Key ® to select the Profile Action Request. The upper display will shown o n f and the lower display will show P. R [ i •

2. Press the Up O or Down O keys to select the End. The upper display will show En d andthe lower display will show P. R [ 1.

3. Press the Infinity Key to return Home. The Profile will End.

Profiling Parameters p I

Pr oF Profile (1 to 4) I

P I Profile [1 to 4] Step (1 to 40)

5.1: '::JP Step Type

1:.5 p I Target Set Point Loop 1

1:.5P2 Target Set Point Loop 2

hoUr Hours

r'7 ,n Minutes

SE[ Seconds

rRl:E Rate

LJP Wait For Process Instance

LJ.P I Wait For Process 1

LJE I Wait For Event 1

LJE.2 Wait for Event 2

doLJ Day of Week

JS Jump Step

J[ Jump Count

End End Type

E nl: I Event 1

Enl:2 Event 2

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • A6

Page 32: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

Profiling Page

Parameter Name Display Range Default

Description

p I

ProF

Profiling Menu

Pl Profile [1 to 4] Step 1 to 1 0 [profile 1]

P1 to Select a step to edit or view. 11 to 20 [profile 2]

P4 21 to 30 [profile 3]

P4 31 to 40 [profile 4]

S.L: '::IP Step Type US/:P Unused Step Unused

S.typ Select a step type. (50)

Note: SoRH Soak (87)

Prior to selecting the Step L Jf Wait For Event

(144) Type consider whether or

L JP r Wait For Pro-not profiles will be based

cess (209) on time or rate of change.

L J.b o Wait For Both By default, profiles are

(210) configured for Time L: , •

Jl Jump (116) Therefore, Rate will not be available here. If it is

End End (27)

desired to base profiles on [ L a [ Wait For Time

rate of change, navigate (1543)

to the Setup Page and then I: , Time (143)

the Global Menu where rRl:f Rate (81)

Ramping Type can be changed from Time to Rate.

L:.5 p I Step Type Parameters -1, 999.000 to 0.0 ° F or

t.SP1 Target Set Point Loop 1 9, 999.000 ° F or units When Step Type is Time or units -18 ° CRate, enter the closed loop -1, 128 toset point for loop 1 to ramp 5,537.000 ° Cto for this step.

Note:

Modbus Relative Address

- - .. -

Instance 1

Map 1 Map 2 2570 4500

Offset to next in-stance (Map 1 equals +50, Map 2equals +100)

Instance 1

Map 1 Map 2 2572 4502

Offset to next in-stance (Map 1 equals +50, Map 2equals +100)

CIP

Class Instance Attribute hex (dee)

- .. - -

0x79 (121) 1 to 40

1

0x79 (121) 1 to 40

2

Pa-ram-eter

ID

... ... - -

21001

21002

Data Type and

Access **

- - - -

uint RWE

float RWE

Some values will be rounded off to fit in the four-character display. Full values can be read with other inter-faces.

** R: Read, W: Write, E: EEPROM, S: User Set

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • A7

Page 33: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

Profiling Page

Parameter Name Display Range Default

Description

1:..5P2 Step Type Parameters -1,999.000 to 0.0°

F or

t.SP2 Target Set Point Loop 2 9, 999.000°

F or units When Step Type is Time en- units -18

°

(ter the closed loop set point -1, 128 tofor loop 2 to ramp to for this 5,537.000

°

C

step.

hoUr Step Type Parameters Oto 99 0

hoUr Hours Select the hours (plus Min-utes and Seconds) for a timed step.

r'7 ,nStep Type Parameters Oto 59 0

Min Minutes When Step Type is Time, Soak, or Wait For Time enter Minutes (plus Hours and Sec-onds) for this step.

Sf[ Step Type Parameters Oto 59 0

SEC Seconds When Step Type is Time, Soak, or Wait For Time enter Seconds (plus Hours and Min-utes) for this step.

Note:

Modbus Relative Address

Instance 1

Map 1 Map 2 - - - - 4554

Offset to next in-stance Map 2 equals

+100

Instance 1 Map 1 Map 2 2574 4504

Offset to next in-stance (Map 1 equals +50, Map 2equals +100)

Instance 1

Map 1 Map 2 2576 4506

Offset to next in-stance (Map 1 equals +50, Map 2equals +100)

Instance 1

Map 1 Map 2 2578 4508

Offset to next in-stance (Map 1 equals +50, Map 2equals +100)

CIP Class

Instance Attribute hex (dee)

0x79 (121) 1 to 40

0x1C (28)

0x79 (121) 1 to 40

3

0x79 (121) 1 to 40

4

0x79 (121) 1 to 40

5

Pa-Data T ype

ram-and

eter Access

ID **

21028 float RWE

21003 uint RWE

21004 uint RWE

21005 uint RWE

Some values will be rounded off to fit in the four-character display. Full values can be read with other inter-faces.

** R: Read, W: Write, E: EEPROM, S: User Set

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • A8

Page 34: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

Profiling Page

Parameter Name Display Range Default

Description

, Rl:E Step Type Parameters 0 to 9, 999.000°

F or 0.0

rAtE Rate units per minute

When Step Type is Rate, en- 0 to 5,555.000°

C per ter the rate for ramping in minute

degrees or units per minute.

LJ.P Step Type Parameters 1 or 2 1

W.Pi Wait For Process Instance When Step Type is Wait for Process or Wait For Both, enter which analog input specified by Wait For Process 1 must be met before pro-ceeding in profile.

LJ.P I Step Type Parameters -1, 999.000 to 0.0°

F or

W.P1 Wait For Process 1 9,999.000°

F or units When Step Type is Wait for units -18.0

°

CProcess or Wait For Both, -1, 128.000 to

enter wait for process value 5,537.000°

C

on analog input specified by Wait For Process Instance before pro-ceeding in profile.

Note:

Modbus Relative Address

Instance 1

Map 1 Map 2 2580 4510

Offset to next in-stance (Map 1 equals +50, Map 2equals +100)

Instance 1

Map 1 Map 2 2598 4528

Offset to next in-stance (Map 1 equals +50, Map 2equals +100)

Instance 1

Map 1 Map 2 2590 4520

Offset to next in-stance (Map 1 equals +50, Map 2equals +100)

CIP Class

Instance Attribute hex (dee)

0x79 (121) 1 to 40

6

0x79 (121) 1 to 40

0x0F (15)

0x79 (121) 1 to 40

0x0B (11)

Pa-ram-eter ID

21006

21015

21011

Data Type and

Access **

float RWE

uint RWE

float RWE

Some values will be rounded off to fit in the four-character display. Full values can be read with other inter-faces.

** R: Read, W: Write, E: EEPROM, S: User Set

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • A9

Page 35: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

Profiling Page

Parameter Name Display Range Default

Description

LJE. I Step Type Parameters oFF Off (62) Off

WE.1 Wait Event 1 on On (63) When Step Type is Wait for nonf None (61) Event or Wait For Both, select the event state that must be satisfied during this step.

Note:

Wait Event 1 can be mapped to any available digital input (5 - 12). Navi-gate to the Setup Page un-der the Global Menu to find and modify Source Instance A 5. , R (Event 1) and Source Instance B 5. ,b (Event 2).

LJE.2 Step Type Parameters oFFOff(62) Off

WE.2 Wait Event 2 on On (63) When Step Type is Wait for nonf None (61) Event or Wait For Both, select the event state that must be satisfied during this step.

Note:

Wait Event 2 can be mapped to any available digital input (5 - 12). Navi-gate to the Setup Page un-der the Global Menu to find and modify Source Instance A 5. , R (Event 1) and Source Instance B 5. ,b (Event 2).

doLJ Step Type Parameters Ed Every Day (1567) Sunday

dow Day of Week L J d Week days When Step Type is Wait for (1566) Time, the profile waits until 5 u n Sunday (1565) this Day of Week along with r'1on Monday (1559) Hours, Minutes and Seconds l:uf Tuesday (1560) time of day is met. L J f d Wednesday

(1561) I: h LJ r T hursday (1562) Fr , Friday (1563)

5 RI: Saturday ( 1564)

Note:

Modbus Relative Address

Instance 1

Map 1 Map 2 2586 4516

Offset to next in-stance (Map 1 equals +50, Map 2equals +100)

Instance 1

Map 1 Map 2 2588 4518

Offset to next in-stance (Map 1 equals +50, Map 2equals +100)

Instance 1

Map 1 Map 2 - ... - "' 4580

Offset to next in-stance Map 2 equals +100)

CIP Pa-

Data Class Type

Instance ram-

and Attribute

eter Access

hex (dee) ID

**

0x79 (121) 21009 uint 1 to 10 RWE

9

0x79 (121) 21010 uint 1 to 40 RWE 0xA(10)

0x79 (121) 21041 uint 1 to 40 RWE

0x29 (41)

Some values will be rounded off to fit in the four-character display. Full values can be read with other inter-faces.

** R: Read, W: Write, E: EEPROM, S: User Set

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • A10

Page 36: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

Profiling Page

Parameter Name Display Range Default

Description

JS Step Type Parameters 1 to 40 0

JS Jump Step

When Step Type is Jump, this specifies which step to jump back to. Jump Step must be a lower step num-ber than the current step number.

J[ Step Type Parameters Oto 9,999 0

JC Jump Count

When Step Type is Jump, this specifies the number of jumps to repeat. A value of 0 creates an infinite loop. Loops can be nested four deep.

End Step Type Parameters a FF Control Mode Off

End End Type set to Off (62) When Step Type is End, this Hal d Hold last specifies what the control- closed-loop set ler will do when this profile point in the profile ends. (47)

US Er User, reverts to previous set point (100)

E nl:: I Step Type Parameters oFF Off (62) Off

Ent1 Event 1 on On (63) When Step Type is not Un-used Step, select whether Event Output 1 or 2 is on or off during this step.

Note:

Modbus Relative Address

Instance 1

Map 1 Map 2 2592 4522

Offset to next in-stance (Map 1 equals +50, Map 2equals +100)

Instance 1

Map 1 Map 2 2594 4524

Offset to next in-stance (Map 1 equals +50, Map 2equals +100)

Instance 1

Map 1 Map 2

2596 4526

Offset to next in-stance (Map 1 equals +50, Map 2equals +100)

Instance 1 Map 1 Map

2 2582 4512

Offset to next in-stance (Map 1 equals +50, Map 2equals +100)

CIP Class

Instance Attribute hex (dee)

0x79 (121) 1 to 40

0xC (12)

0x79 (121) 1 to 40

0xD (13)

0x79 (121) 1 to 40

0xE (14)

0x79 (121) 1 to 40

7

Pa-Data Type

ram-and

eter Access

ID **

21012 uint RWE

21013 uint RWE

21014 uint RWE

21007 uint RWE

Some values will be rounded off to fit in the four-character display. Full values can be read with other inter-faces.

** R: Read, W: Write, E: EEPROM, S: User Set

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • A11

Page 37: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

Profiling Page

CIP Pa-

Data

Parameter Name Modbus Class Type

Display Range Default Relative Instance ram-

and Description

Address Attribute eter

Access hex (dee)

ID **

Enl:.2 Step Type Parameters oFF Off (62) Off Instance 1 0x79 (121) 21008 uint

Ent2 Event 2 on On (63) Map 1 Map 2 1 to 40 RWE When Step Type is not Un- 2584 4514 8

used Step, select whether Event Output 1 or 2 is on or Offset to

off during this step. next in-stance (Map 1 equals +50, Map 2equals +100)

Note:

Some values will be rounded off to fit in the four-character display. Full values can be read with other inter-faces.

** R: Read, W: Write, E: EEPROM, S: User Set

Display Step Type Description

USl:.P Step Types

UStP Unused Step This is an empty step that can be used to plan for future steps to be inserted or temporarily deactivate a step in a profile.

Change step type back when the step should be active again.

C I Step Types

ti Time If Ramping Type found in the Global Menu of the Setup Page is set for Time, the control loop will follow set point over the specified time. If two loops of control are present then they will both follow independent set points over the specified time. The state of up to 2 event outputs may be set or maintained.

rRl:.E Step Types

rAtE Rate If Ramping Type found in the Global Menu of the Setup Page is set for Rate, specify the rate of change in degrees or units per minute. The state of up to 2 event outputs may be set or main-tained.

SoRH Step Types

SoAh Soak A Soak Step maintains the last Target Set Points for the desig-nated time. The state of up to 2 event outputs may be set or maintained.

[Lo[ Step Types

CLoC Wait For Time A Wait for Time Step is available with the real-time calendar clock feature. This allows the program to wait for a specified day and time before proceeding to the next step. Used to have the profile execute steps everyday or only weekdays. The state of up to 2 event outputs may be set or maintained.

Parameters in Step Type

I: 9.5 /

I: 9.52

hoUr

r'1 ,n

SE[

Enl:: I

Enl::2

I: 9.5 /

1::9.52

rRl::E

E nl:: I

Enl::2

hoUr

r'1 ,n

SE[

Enl:: I

Enl::2

hoUr

r'1 ,n

SE[

doLJ

Enl:: I

Enl::2

.. - - -

Target Set Point Loop 1

Target Set Point Loop 2

Hours

Minutes

Seconds

Event 1

Event 2

Target Set Point Loop 1

Target Set Point Loop 2

Rate

Event 1

Event 2

Hours

Minutes

Seconds

Event 1

Event 2

Hours

Minutes

Seconds

Day of Week

Event 1

Event 2

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • A12

Page 38: HotBlock 200 Digestion Systems · 2018. 8. 17. · Method 200.8 13-14. Methods 245.1, 7470, and 7471 for Mercury Digestions 15-16. Method 365.1 for Phosphorus 17. Method 3050B 18-19.

Display Step Type Descr iption Parameter s in Step Type

LJf Step Types L J f. / Wait Event 1

W.E Wait For Event L J f.2 Wait Event 2 A Wait for Event Step will wait for the two Wait for Event states

E nl: I Event 1 (1 to 2) to match the specified state. The state of up to 2 event

En I: 2 Event 2 outputs may be set or maintained.

LJPr Step Types LJ.P I Wait for Process 1

W.Pr Wait For Process LJ.P2 Wait for Process 2 A Wait for Process Step will wait for Process Value 1 or 2 to

E nl: I Event 1 match the Wait for Process Value. The state of up to 2 event

Enl:2 Event 2 outputs may be set or maintained.

LJbo Step Types LJ.P I Wait for Process 1

W.bo Wait For Both LJ.P2 Wait for Process 2 A Wait For Process and Event Step will wait for Process Value 1

LJE. I Wait Event 1 or 2 to match the Wait for Process 1 value, and/or the two Wait

LJE.2 Wait Event 2 Event states to match the specified state. The state of up to 2 event outputs may be set or maintained. E nl: I Event 1

fnl:2 Event 2

JL Step Types JS Jump Step

JL Jump J[ Jump Count A Jump step will repeat previous steps a number of times

E nl: I Event 1 designated in Jump Count. Jumps can be nested up to four

fnf:2 Event 2 deep. The state of up to 2 event outputs may be set or main-tained.

End Step Types End End Type

End End E nl: I Event 1 An End Step will end the profile and set the control modes and

fnl:2 Event 2 set points to match the End Type. The state of up to 2 event outputs may be set or maintained. The event outputs will not be set off unless specifically stated in this step. If a profile does not have an End Step, the profile continues until step 40, then stops and maintains the last set points and control modes.

800-343-5319 or 843-881-6560 www.environmentalexpress.com Environmental Express • A13


Recommended