Ian Marshall, Maurits A Jansen, Yuehui Tao,
Gavin D Merrifield, Gillian A Gray
University of Edinburgh
Accelerated mouse cardiac imaging using threefold undersampling and kt-BLAST reconstruction
3847
Computer 35
Introduction
Mouse kt-BLAST cardiac imaging 3847
Why acceleration?
- shorter scan times, reduced costs,
- improved patient compliance
- reduced animal burden
Parallel imaging now widely used clinically
(some) kt-undersampling for human cardiac scanning
Limited translation of techniques to rodent imaging
- would improve animal welfare (3Rs)
Study design
6 healthy + 2 MI adult mice
Cardiac MRI scans in short-axis plane
Standard CINE sequence
3-fold accelerated ktBLAST sequence + training data
Comparison of image quality and cardiac indices
EDV, ESV, EF, EDM, ESM
Mouse kt-BLAST cardiac imaging 3847
Scanning details
7T scanner (Agilent Technologies)
39 mm quad coil
Respiratory gating and cardiac triggering
CINE Gradient Echo sequence
- (12 frames, 192×192, FOV 30 mm, 3 Averages)
kt-BLAST sequence (Tsao MRM 2003)
- 3× undersampling (linear grid)
- 16 central training rows
Mouse kt-BLAST cardiac imaging 3847
Image analysis
One analyst, blinded to type of images
Subjective quality (scale 1-5)
Endo- and epicardial borders drawn manually
Cardiological indices derived
Mouse kt-BLAST cardiac imaging 3847
Results summary
Cardiological indices (σ) wrt fully sampled images
“Ratios” Bland-Altman bias
EDV 0.94 (0.07) -5 (8) μL
ESV 1.06 (0.14) 2 (6) μL
EF 0.93 (0.08) -5 (5) %
EDM 1.11 (0.10) 13 (12) mg
ESM 1.01 (0.06) 1 (8) mg
Image quality: 4.3±1.2 (full) versus 3.8±1.4 (ktBLAST)
Mouse kt-BLAST cardiac imaging 3847
Discussion
kt-BLAST images have similar “quality”
“Ratios” biased up to 11% (for EDM)
- corresponding Bland-Altman bias 13 mg
Observer variability 3-10% (Heijman JMRI 2008)
MI mice (large hearts and small EF)
Overall acceleration 2.4×
Compare with Compressed Sensing (Wech JMRI 2011)
Mouse kt-BLAST cardiac imaging 3847
Conclusion
Mouse kt-BLAST cardiac imaging 3847
Accelerated cardiac imaging of mice is feasible
Threefold acceleration gives acceptable results
This enables- increased throughput- improved animal welfare