9390 Digital Weight Indicator Version 1.1 Installation Manual
Transcript
Page 1: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

9390Digital Weight Indicator

Version 1.1

Installation Manual

Page 2: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page
Page 3: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

Copyright © 2001 Interface, Inc.. All rights reserved. Printed in the United States of America. Specifications subject to change without notice.

Contents

About This Manual ................................................................................................................................... 11.0 Introduction.................................................................................................................................. 1

1.1 Operating Modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.2 Front Panel Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.3 Display Annunciators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.4 Indicator Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

1.4.1 Toggle Gross/Net/Piece Count Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31.4.2 Toggle Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31.4.3 Zero Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31.4.4 Acquire Tare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31.4.5 Remove Stored Tare Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31.4.6 Print Ticket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31.4.7 Acquire Parts Sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

2.0 Installation ................................................................................................................................... 42.1 Unpacking and Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42.2 Enclosure Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42.3 Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

2.3.1 Cable Grounding. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52.3.2 Load Cells. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52.3.3 Serial Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

2.4 Enclosure Reassembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62.5 Battery Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62.6 AC Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72.7 Board Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72.8 Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

3.0 Configuration.............................................................................................................................. 103.1 Configuration Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

3.1.1 Revolution Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103.1.2 EDP Command Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103.1.3 Front Panel Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

3.2 Menu Structures and Parameter Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123.2.1 Configuration Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133.2.2 Format Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153.2.3 Calibration Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173.2.4 Serial Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183.2.5 Program Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193.2.6 Print Format Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 213.2.7 Version Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

4.0 Calibration.................................................................................................................................. 224.1 Front Panel Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 224.2 EDP Command Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 234.3 Revolution™ Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

5.0 EDP Commands.......................................................................................................................... 245.1 The EDP Command Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

5.1.1 Key Press Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 245.1.2 Reporting Commands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 255.1.3 The RESETCONFIGURATION Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 255.1.4 Parameter Setting Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 255.1.5 Normal Mode Commands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

Version 1.1, October 2001

Page 4: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

5.2 Saving and Transferring Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 275.2.1 Saving Indicator Data to a Personal Computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 275.2.2 Downloading Configuration Data from PC to Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

6.0 Counting Operations................................................................................................................... 287.0 Print Formatting ......................................................................................................................... 29

7.1 Print Formatting Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 297.2 Customizing Print Formats. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

7.2.1 Using the EDP Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 307.2.2 Using the Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 307.2.3 Using Revolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

8.0 Appendix .................................................................................................................................... 328.1 Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 328.2 Continuous Output (Stream) Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 328.3 ASCII Character Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 338.4 Front Panel Display Characters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 358.5 Conversion Factors for Secondary Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 368.6 Digital Filtering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

8.6.1 DIGFLx Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 378.6.2 DFSENS and DFTHRH Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 378.6.3 Setting the Digital Filter Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

8.7 Test Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 388.8 Software Revision History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 398.9 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

9390 Limited Warranty........................................................................................................................... 41

ii 9390 Installation Manual

Page 5: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

he

e

sthewo

rtse

te

ledy.d

ale,

errt aletor

tice,een

deiss

yor

About This ManualThis manual is intended for use by service techniciansresponsible for installing and servicing 9390 digitalweight indicators.

This manual applies to indicators using Version 1.1 ofthe 9390 software. See Section 8.8 on page 39 for asummary of software changes included in this release.

Configuration and calibration of the indicator can beaccomplished using the indicator front panel keys, theEDP command set, or the Revolution™ configurationutility. See Section 3.1 on page 10 for informationabout configuration methods.

Some procedures described in thismanual require work inside the indicatorenclosure. These procedures are to beperformed by qualified service personnelonly.

The Operator Card included with this manualprovides basic operating instructions for users of t9390. Please leave the Operator Card with theindicator when installation and configuration arcomplete.

1.0 IntroductionThe 9390 is a single-channel digital weight indicatorhoused in a NEMA 4X/IP66-rated stainless steelenclosure. The indicator front panel consists of a 1 in(25 mm), six-digit, liquid crystal (LCD) display andsix-button keypad. Features include:

• Drives up to four 350Ω or eight 700Ω loadcells

• Supports 4- and 6-wire load cell connections• Electronic data processing (EDP) port for full

duplex, RS-232 communications at up to19200 bps

• Powered by 6 C batteries for completeportability

• AC adapter for 115 or 230 VAC power• Configurable standby mode limits indicator

power consumption when scale is inactive• Piece count mode for basic parts counting

functions.See Section 8.9 on page 40 for detailed specifications.

1.1 Operating ModesThe 9390 has several modes of operation:

Normal (weighing) modeNormal mode is the “production” mode of theindicator. The indicator displays gross or netweights as required, using the LCD annunciatorsdescribed in Section 1.3 on page 2 to indicatescale status and the type of weight valuedisplayed. Once configuration is complete and alegal seal is affixed to the back of the indicator,the 9390 can operate only in normal mode.

Piece count modeIn piece count mode, the indicator display showthe number of parts on the scale rather than weight of those parts. Piece count mode has tsubmodes: • Count display mode displays the current pa

count and allows ticket printing using thCFMT print format.

• Sample acquisition mode is used to calibrathe indicator for parts counting.

Operator access to piece count mode is disabwhen the indicator is shipped from the factorSee Section 6.0 on page 28 for detaileinformation about counting operations.

Setup modeMost of the procedures described in this manurequire the indicator to be in setup modincluding configuration and calibration.To enter setup mode, remove the large fillisthead screw from the enclosure backplate. Insescrewdriver or a similar tool into the access hoand press the setup switch once. The indicadisplay changes to show the word CONFIG.

Test modeTest mode provides a number of diagnosfunctions for the 9390 indicator. Like setup modtest mode is entered using the setup switch. SSection 8.7 on page 38 for more informatioabout entering and using test mode.

Standby modeStandby mode is a configurable low-power moused to extend battery life when the indicator inactive. In standby mode, the LCD display showthe word STNDBY; press any key to exit standbmode. See Section 3.2.5 on page 19 finformation about configuring standby mode.

Warning

Introduction 1

Page 6: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

the keyect digitsn about

n 6.0 on

De

C)ew

w

e

t of

dsry asofrts

ut

1.2 Front Panel KeypadFigure 1-1 shows the 9390 front panel keys and the key functions assigned in normal mode.

The symbols shown above the keys in Figure 1-1 (representing up, down, enter, left, right) describe functions assigned in setup mode. In setup mode, the keys are used to navigate through menus, selwithin numeric values, and increment/decrement values. See Section 3.1.3 on page 11 for informatiousing the front panel keys in setup mode.

Labels under the keys in Figure 1-1 describe the key functions used in piece count mode. See Sectiopage 28 for more information about counting operations.

Figure 1-1. Front Panel Key Functions in Normal Mode

1.3 Display AnnunciatorsThe 9390 display uses a set of LCD annunciators toprovide additional information about the value beingdisplayed. Figure 1-2 on page 3 shows theseannunciators and their meanings.

GThe displayed weight is a gross weight.

NThe displayed weight is a net weight.

Scale is at standstill or within the specified motionband. Some operations, including zero, tare, andprinting functions, can only be done when thestandstill symbol is shown.

Center of zero. Gross weight is wi th in 0.25graduations of zero. This annunciator lights when thescale is zeroed.

Indicates that a tare has been entered. The tare valuecan be entered by pressing the TARE key on thekeypad or by sending the KTARE EDP command. SeeSection 5.0 on page 24 for information about the EDPcommand set.

Indicates that the battery charge is low. lb, kg, PCThree arrow annunciators on the right side of the LCdisplay indicate the units associated with thdisplayed value. The bottom arrow annunciator (Pis lit when the indicator is in piece count mode. Thtop and middle arrows, labeled lb and kg, shoprimary and secondary units, as described below:• If the displayed weight is in pounds, the top arro

(lb) is lit; if kilograms, the middle arrow (kg) islit.

• If the primary unit of weight is pounds, the middl(kg) arrow is lit for secondary units—or, if theprimary unit is kilograms, the top (lb) arrow is lifor secondary units, unless the secondary unitweight is the same as the primary unit.

• If neither primary nor secondary units are pounor kilograms, the top arrow is used as a primaunits annunciator and the middle arrow is usedthe secondary units annunciator. A set replacement overlay decals is included in the pakit for units other than pounds and kilograms.

See Section 3.2.2 on page 15 for information aboconfiguring primary and secondary display units.

T

ZERO GROSSNET

TARE UNITS PRINT

TB/G Units

POWER

ENTER SAMPLEMODE PRINT

I/O

Switch betweenprimary and

secondary units

Send data toserial port

Set gross weightto zero

Switch betweengross, net, and

piece count mode

Acquire tare

NORMAL MODE KEY FUNCTIONS

2 9390 Installation Manual

Page 7: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

or

n

l

fors

ge

ofize

50,

le,

y

ay

Figure 1-2. 9390 Display Annunciators

1.4 Indicator OperationsBasic 9390 operations are summarized below:

1.4.1 Toggle Gross/Net/Piece Count ModePress the GROSS/NET key to switch the display modebetween gross, net, and piece count modes. If a tarevalue has been entered or acquired, the net value is thegross weight minus the tare. If access to piece countmode is disabled, pressing the GROSS/NET keyswitches between gross and net modes only.

Gross mode is shown by the G annunciator; net modeis shown by the N annunciator. Piece count mode isshown by the LCD annunciator adjacent to the PC atthe right edge of the display. 1.4.2 Toggle UnitsPress the UNITS key to switch between primary andsecondary units. The units annunciator arrow on theright side of the LCD display points to the unitsidentifier.1.4.3 Zero Scale

1. In gross mode, remove all weight from thescale and wait for the standstill annunciator( ).

2. Press the ZERO key. The center of zero( ) annunciator lights to indicate thescale is zeroed.

1.4.4 Acquire Tare1. Place container on scale and wait for the

standstill annunciator ( ).2. Press the TARE key to acquire the tare weight

of the container.

3. Net weight is displayed and the annunciator lit to show the tare value wasacquired.

1.4.5 Remove Stored Tare Value1. Remove all weight from the scale and wait f

the standstill annunciator ( ).

2. Press the ZERO key. The annunciatorgoes off, indicating the tare value has beeremoved.

1.4.6 Print Ticket1. Wait for the standstill annunciator ( ).2. Press the PRINT key to send data to the seria

port.1.4.7 Acquire Parts Sample

1. Place empty parts container on scale. Wait the standstill annunciator ( ), then presTARE to acquire the tare weight of thecontainer.

2. Press MODE to enter piece count mode.3. Press the SAMPLE (UNITS) key to enter

sample acquisition mode. 4. The indicator display shows the messa

Addnnn, where nnn is the sample quantity tobe placed on the scale. Add the number parts shown or choose a different sample sby pressing the SAMPLE key to scroll throughthe selectable sample quantities (5, 10, 20, 100).

5. Once the sample quantity is on the scapress ENTER to calibrate the indicator forcounting the new parts. The indicator displashows the message –CNT– as it acquires thesample weight, then switches to count displmode and shows the part quantity.

Gross mode

Net mode

Standstill

Center of zero

Tare in system

Low battery

lb

kg

PC

Displayed unitsSecondary

Primary

Piece count mode

Introduction 3

Page 8: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

Nin

r

,

),d

d

ect

C onoldhect

late

r serial

2.0 InstallationThis section describes procedures for connecting loadcell and serial communications cables to the 9390indicator. Instructions for battery and CPU boardreplacement are included, along with assemblydrawings and parts lists for the service technician.

Use a wrist strap to ground yourself andprotect components from electrostaticdischarge (ESD) when working insidethe indicator enclosure.

2.1 Unpacking and AssemblyImmediately after unpacking, visually inspect the9390 to ensure all components are included andundamaged. The shipping carton should contain theindicator with attached tilt stand, this manual, and aparts kit. If any parts were damaged in shipment,notify Interface, Inc. and the shipper immediately.

The parts kit contains the items listed below:

• 3-position (PN 15888) and 7-position (PN23165) pluggable terminal blocks for load celland serial communications connectors.

• One 8-32NC x 7/16 fillister head screw (PN30623). This screw occupies the hole belowthe setup switch access screw on the indicator

backplate (see Figure 2-4 on page 6).• Four 8-32NC x 3/8 machine screws (P

14862) for the indicator backplate (see #10 Figure 2-7 on page 9).

• Five neoprene washers (PN 45042) fobackplate screws included in the parts kit.

• Four rubber bumpers (“feet”) for the tilt standPN 42149.

• Two each of grounding clamps (PN 53075external tooth lock washers, (PN 15133), ankep nuts (PN 14626) for cable shielgrounding against the enclosure.

2.2 Enclosure DisassemblyThe indicator enclosure must be opened to connload cell and communications cables.

Power-off the indicator and disconnect the Aadapter, if necessary. Place the indicator face-downan antistatic work mat. Remove the screws that hthe backplate to the enclosure body, then lift tbackplate away from the enclosure. Disconnebattery cable to connector J3, then set the backpassembly aside.

2.3 Cable ConnectionsThe 9390 provides two cord grips for cabling into the indicator: one for the load cell cable, the other focommunications.

Figure 2-1. 9390 Backplate, Showing Key Component Locations

Caution

Battery coverSerial communicationscable cord grip

Load cell cablecord grip

AC adapterport cover

4 9390 Installation Manual

Page 9: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

x,toromll-1.),ingn,

tor

veecttor

ader

2nd for.4

2.3.1 Cable GroundingExcept for the power cord, all cables routed throughthe cord grips should be grounded against theindicator backplate. Do the following to groundshielded cables:

• Use the lockwashers, clamps, and kep nutsprovided in the parts kit to install groundingclamps on the backplate studs adjacent to cordgrips. Install grounding clamps only for cordgrips that will be used; do not tighten nuts.

• Route cables through cord grips and groundingclamps to determine cable lengths required toreach cable connectors. Mark cables to removeinsulation and shield as described below:• For cables with foil shielding, strip insulation

and foil from the cable half an inch (15 mm)past the grounding clamp (see Figure 2-2).Fold the foil shield back on the cable wherethe cable passes through the clamp. Ensuresilver (conductive) side of foil is turnedoutward for contact with the groundingclamp.

• For cables with braided shielding, strip cableinsulation and braided shield from a point justpast the grounding clamp. Strip another halfinch (15 mm) of insulation only to expose thebraid where the cable passes through theclamp (see Figure 2-2).

• For load cell cables, cut the shield wire just pastthe grounding clamp. Shield wire function isprovided by contact between the cable shieldand the grounding clamp.

• Route stripped cables through cord grips andclamps. Ensure shields contact groundingclamps as shown in Figure 2-2. Tightengrounding clamp nuts.

• Finish installation using cable mounts and ties tosecure cables inside of indicator enclosure.

Figure 2-2. Grounding Clamp Attachment for Foil-Shielded and Braided Cabling

2.3.2 Load CellsTo attach cable from a load cell or junction boremove connector J1 from the board. The connecplugs into a header on the board. Connect cable frthe load cell or junction box through the load cecable cord grip to connector J1 as shown in Table 2If using 6-wire load cell cable (with sense wiresremove jumpers JP1 and JP2 before reinstallconnector J1 (see Figure 2-3). For 4-wire installatioleave jumpers JP1 and JP2 on.

When connections are complete, reinstall connecJ1 on the board.

2.3.3 Serial CommunicationsTo attach serial communications cables, remoconnector J2 from the board (see Figure 2-3). Conncommunications cable through cord grip to connecJ2 as shown in Table 2-2.

Once cables are attached, reconnect J2 to the heon the board.

The 9390 serial port supports full duplex RS-23communications for connections to printers, PCs, aother attached devices. See Section 3.0 on page 10general configuration information; see Section 3.2on page 18 for serial port configuration.aa Cord grip

Insulated cable

Foil (silver side out) Grounding clamp

Shield wire (cut)aa

Length of foil before foldingback on cable insulation

Cut insulation herefor foil-shielded cables

Braid

Cut insulation herefor braided cables

NOTE: Install lockwashersfirst, against backplate,under grounding clampaa Cord grip

Insulated cable

Foil (silver side out) Grounding clamp

Shield wire (cut)aa

Length of foil before foldingback on cable insulation

Cut insulation herefor foil-shielded cables

Braid

Cut insulation herefor braided cables

NOTE: Install lockwashersfirst, against backplate,under grounding clamp

J1 Pin Function

1 +SIG

2 –SIG

3 +SENSE

4 –SENSE

5 SHIELD (see NOTE below)

6 +EXC

7 –EXC

NOTES: • SHIELD wire connection not used. Use grounding

procedure described in Section 2.3.1 on page 5.• For 6-wire connections, remove jumpers JP1 and

JP2.

Table 2-1. J1 Pin Assignments

J2 Pin Function

1 RS-232 TxD

2 RS-232 Ground

3 RS-232 RxD

Table 2-2. J2 Pin Assignments

Installation 5

Page 10: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

e

n

lllb

Figure 2-3. 9390 CPU Board

2.4 Enclosure ReassemblyOnce cabling is complete, position the backplate overthe enclosure, reconnect battery cable, and reinstallthe backplate screws. Use the torque pattern shown inFigure 2-4 to prevent distorting the backplate gasket.Torque screws to 15 in-lb (1.7 N-m).

Figure 2-4. 9390 Enclosure Backplate

2.5 Battery InstallationTo install or replace batteries, do the following:

1. Remove the four thumbscrews that hold thbattery cover to the enclosure backplate.

2. Install six alkaline C-cell batteries as showin Figure 2-5.

3. Replace the battery cover and reinstathumbscrews. Torque thumbscrews to 15 in-(1.7 N-m).

Figure 2-5. Battery Installation

R12

C34

C33

D3

D4

RN

1

C40

C38C

36

C39

C35

R23

C72

C73

C71

C74C78

C76

R27

R28

C81

C83

R24

R25

R31

D5F1

C80

EM

I1

R46

R45

R44

R40

R39

R38

R37

C92

C91

C90

R34

R33

R43

C84

C85

U15

C77

VR

2

D1

R2

C31

C32

R1

C1

L2

C28

U1

VR1L1

C13

C6

R7 C10

C9

R26 C79

C75

U6

C14

R32

R29

C82

C2

XT1

C4

C8C7

C3

C5

R6

R5

R3

R4

C41

R15

R14

C37

C42

Q1

C11

C12

R30

C86

C87

C15

U12

U10

U7

U11

R8

C17

C18

C19

R9

C16

C20

C27

U9

C88

C89

VR4

U8

R49

R50

C95

R48

R42

R41

C94

C93

R35

R36

R47

RICE LAKE WEIGHING SYSTEMS

1

RESET

AGND

DGND

+SIG

–SIG

+SENSE

–SENSE

SHIELD

+EXC

–EXC

1

2

3

4

5

6

7

LOAD CELLCONNECTOR

J1

JP2

JP1

D2

J5Keypad Connector

1

J4

J2

TxD

GND

RxD

1

2

3

SERIAL PORT

1

J3

BatteryInput

ACAdapter

Input

1

J6 To SetupSwitchA/D Conver ter

GND

GND

1 J7

VR3

U4

Microcontroller

U5FLASH RAM

U2 U3

Display Drivers

R11

C29

R13

C30

RN

2

+

+

+

C26

C23

C25

R10

C21

C22

C24

+

Setup switchaccess screw

1

3

4

5 6

2

7

Torque backplate screwsto 15 in-lb (1.7 N-m)

Torque pattern

8

+

+ +

+––

––

––

+ +

6 9390 Installation Manual

Page 11: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

he

ned

ee.

he

.ve allce4.

2.6 AC AdapterThe AC adapter can be used to provide power to theindicator if battery power is low and AC power isavailable. Use only the adapter supplied with theindicator.

To use the AC adapter, do the following:

1. Remove the top screw from the AC adapterport cover.

2. Loosen the bottom screw, then rotate the portcover to expose the AC adapter port (seeFigure 2-6).

3. Reinstall the top screw in the backplate.4. Plug AC adapter from power outlet to the

indicator adapter port.Close the adapter port cover and reinstall screws asdescribed in Section 2.4 when not using the ACadapter.

The 9390 enclosure cannot be rated forNEMA 4X/IP66 applications with the ACadapter port cover open.

Do not use the AC adapter in washdown environments.

Figure 2-6. AC Adapter Port Cover

2.7 Board RemovalIf you must remove the 9390 CPU board, use tfollowing procedure:

1. Disconnect power to the indicator. Loosecord grips and remove backplate as describin Section 2.2 on page 4.

2. Unplug all connections to the CPU board. SFigure 2-3 on page 6 for connector locations

3. Remove the four nuts from the corners of tCPU board.

4. Remove the CPU board from the enclosureTo replace the CPU board, reverse the aboprocedure. Be sure to reinstall cable ties to securecables inside the indicator enclosure. Replabackplate and torque screws as shown in Figure 2-

Caution

Installation 7

Page 12: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

2.8 Replacement PartsTable 2-3 lists replacement parts for the 9390, including all parts referenced in Figure 2-7 on page 9.

Ref Number PN Description (Quantity)

1 49520 Battery cover thumb screws (4)

2 47939 Battery cover (1)

3 49557 Battery cover gasket (1)

4 49909 Battery holder assembly (1)

5 19538 Cable grip plug (1)

6 42640 Setup switch access screw, 1/4 x 28NF x 1/4 (1)

7 15626 Cable grips, PG9 (2)

8 44676 Sealing washer for setup switch access screw, #14 (1)

9 30375 Nylon seal rings for cable grips (2)

10 14862* Screws, 8-32NC x 3/8 (6)

11 45042* Sealing washers, #8 (8)

12 47940 Enclosure backplate (1)

13 15627 Locknuts, PCN9 (2)

14 49558 Backplate gasket (1)

15 14621 Kep nuts, 6-32NC hex (4)

16 49954 Display and CPU board assembly (1)

18 48790 Overlay membrane panel (1)

19 49953 Screw, 8-32NC x 1/2 (1)

20 49906 AC adapter port cover (1)

21 49904 AC adapter port gasket (1)

23 53073 Tilt stand (1)

24 30342 Wing knobs for tilt stand (2)

25 15144 Nylon washers for tilt stand, 1/4 x 1 x 1/16 (2)

26 49908 AC adapter connector assembly (1)

27 15635 Brass spacers for board mounting (4)

28 49905 Battery cover pads (2)

29 14825 Phillips pan head screws, 18-8 SST (2)

30 49910 Setup switch assembly (1)

32 15631* Cable tie (1)

33 15650* Cable tie mount (1)

34 — Serial tag (1)

36 15130 Internal lock washers (8)

38 14626 Kep nuts (2)

39 45043 Ground wire (1)

40 15134 Lock washers (2)

— 49978 AC adapter, 115 VAC (1)

49977 AC adapter, 230 VAC (1)

* Additional parts included in parts kit.

Table 2-3. Replacement Parts

8 9390 Installation Manual

Page 13: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

Figure 2-7. 9390 Assembly

20

6

8

RIC

E LA

KE

WEI

GH

ING

SYS

TEM

SM

OD

EL:

IQ+3

90-D

CS/

N:

nMAX

:

10 0

00III

/IIIL

2

0 00

0 III

HD

34

7/2X

21

1/4X

2

3

4

5

9/2X

10/6X

11/8X

12

13/2X

14

15/4X

16

17

18

19

23

24/2X26

27/4X

25/2X

28

29/2X

30

33

32

39

36/8X

38/2X

40/2X

Installation 9

Page 14: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

ngsor,

sst

f,ns

hep

ion

snsninge

the ordPntII

leeanto

terre

3.0 ConfigurationTo configure the 9390 indicator, the indicator must beplaced in setup mode. The setup switch is accessed byremoving the large fillister head screw on theenclosure backplate. Switch position is changed byinserting a screwdriver into the access hole andpressing the switch.

When the indicator is placed in setup mode, the wordCONFIG is shown on the display. The CONFIG menuis the first of seven main menus used to configure theindicator. Detailed descriptions of these menus aregiven in Section 3.2. When configuration is complete,return to the CONFIG menu and press the (ZERO)key to exit setup mode, then replace the setup switchaccess screw.

3.1 Configuration MethodsThe 9390 indicator can be configured by using thefront panel keys to navigate through a series ofconfiguration menus or by sending commands orconfiguration data to the EDP port. Configurationusing the menus is described in Section 3.1.3.

Configuration using the EDP port can beaccomplished using either the EDP command setdescribed in Section 5.0 or the Revolution™

configuration utility.

3.1.1 Revolution ConfigurationThe Revolution configuration utility provides thepreferred method for configuring the 9390 indicator.Revolution runs on a personal computer to setconfiguration parameters for the indicator. WhenRevolution configuration is complete, configurationdata is downloaded to the indicator.

Figure 3-1. Sample Revolution Format Display

Revolution supports both uploading and downloadiof indicator configuration data. This capability allowconfiguration data to be retrieved from one indicatedited, then downloaded to another.

To use Revolution, do the following:

1. Install Revolution on an IBM-compatiblepersonal computer running Windows® 3.11 orWindows 95. Minimum system requirementare 8MB of extended memory and at lea5MB of available hard disk space.

2. With both indicator and PC powered ofconnect the PC serial port to the RS-232 pion the indicator EDP port.

3. Power up the PC and the indicator. Use tsetup switch to place the indicator in setumode.

4. Start the Revolution program.Figure 3-1 shows an example of one of the Revolutconfiguration displays.

Revolution provides online help for each of itconfiguration displays. Parameter descriptioprovided in this manual for front panel configuratiocan also be used when configuring the indicator usRevolution: the interface is different, but thparameters set are the same.

3.1.2 EDP Command ConfigurationThe EDP command set can be used to configure 9390 indicator using a personal computer, terminal,remote keyboard. Like Revolution, EDP commanconfiguration sends commands to the indicator EDport; unlike Revolution, EDP commands can be seusing any external device capable of sending ASCcharacters over a serial connection.

EDP commands duplicate the functions availabusing the indicator front panel and provide somfunctions not otherwise available. EDP commands cbe used to simulate pressing front panel keys, configure the indicator, or to dump lists of paramesettings. See Section 5.0 on page 24 for moinformation about using the EDP command set.

10 9390 Installation Manual

Page 15: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

when the

gure 3-2).

rirection

3.1.3 Front Panel ConfigurationThe 9390 indicator can be configured using a series of menus accessed through the indicator front panelindicator is in setup mode. Table 3-1 summarizes the functions of each of the main menus.

Figure 3-2. Front Panel Key Functions in Setup Mode

Four front panel keys are used as directional keys to navigate through the menus in setup mode (see FiThe UNITS ( ) and PRINT ( ) keys scroll left and right (horizontally) on the same menu level; ZERO ( ) andGROSS/NET ( ) move up and down (vertically) to different menu levels. The TARE key ( ) serves as an Entekey for selecting parameter values within the menus. A label over each of these keys identifies the dprovided by the key when navigating through the setup menus.

Figure 3-3. Setup Mode Menu Navigation

Menu Menu Function

CONFIG Configuration Configure grads, zero tracking, zero range, motion band, overload, tare function, and digitalfiltering parameters.

FORMAT Format Set format of primary and secondary units, display rate.

CALIBR Calibration Calibrate indicator. See Section 4.0 on page 22 for calibration procedures.

SERIAL Serial Configure serial port.

PROGRM Program Set power-up and standby modes, counting scale functions, regulatory mode, unit ID andconsecutive number values.

P FORMT Print Format Set print format used for gross, net, and counting scale tickets. See Section 7.0 on page 29for more information.

VERSION Version Display installed software version number.

Table 3-1. 9390 Menu Summary

T

ZERO GROSSNET

TARE UNITS PRINT

TB/G Units

POWER

ENTER SAMPLEMODE PRINT

I/O

Move LEFT /Previous

Move RIGHT /Next

Move UP /Increment Value

Move DOWN /Decrement Value

ENTER Value

SETUP MODE KEY FUNCTIONS

1st LevelParameter

Default value

2nd LevelParameter

Value Value Value

1st LevelParameter

2nd LevelParameter

When moving through values below the first menu level, press to return to the levelabove. Press or to move to the next parameter on the level above.

Configuration 11

Page 16: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

tructure,hoices are

associated

ction 8.4

To select a parameter, press or to scroll left orright until the desired menu group appears on thedisplay, then press to move down to the submenu orparameter you want. When moving through the menuparameters, the default or previously selected valueappears first on the display.

To change a parameter value, scroll left or right toview the values for that parameter. When the desiredvalue appears on the display, press to select thevalue and move back up one level.

To edit numeric values, use the navigation keys toselect the digit and to increment or decrement thevalue. Press to enter the numeric value, then use the

or key to adjust the decimal point position, ifnecessary. Press again to save the value and exit theparameter (see Figure 3-4).

Figure 3-4. Editing Procedure for Numeric Values

3.2 Menu Structures and Parameter DescriptionsThe following sections provide graphic representations of the 9390 menu structures. In the actual menu sthe settings you choose under each parameter are arranged horizontally. To save page space, menu cshown in vertical columns. The factory default setting appears at the top of each column.

Most menu diagrams are accompanied by a table that describes all parameters and parameter values with that menu. Default parameter values are shown in bold type.

NOTE: Some characters cannot be accurately represented on the 7-segment front panel LCD display. Seon page 35 lists the 7-segment character set and the alphanumerics they represent.

When editing numeric values, press or to change thedigit selected. Press or to increment or decrement thevalue of the selected digit. Press to save the value.

2 0 0 0 0 0

After entering the value, press or to change the decimalpoint location. Press again to save the value and exit theparameter.

2 0 0 0 0 0

12 9390 Installation Manual

Page 17: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

3.2.1 Configuration Menu

Figure 3-5. Configuration Menu

CONFIG Menu

Parameter Choices Description

Level 2 submenus

GRADS 10000number

Graduations. Specifies the number of full scale graduations. The value entered must be inthe range 1–999 999 and should be consistent with legal requirements and environmentallimits on system resolution.

To calculate GRADS, use the formula, GRADS = Capacity / Display Divisions.

Display divisions for primary and secondary units are specified on the FORMAT menu.

ZTRKBN OFF0.5D1D3D

Zero track band. Automatically zeroes the scale when within the range specified, as longas the input is within the ZRANGE and scale is at standstill. Selections are ± displaydivisions. Maximum legal value varies depending on local regulations.

ZRANGE 1.9%100%

Zero range. Selects the range within which the scale can be zeroed. The 1.9% selection is± 1.9% around the calibrated zero point, for a total range of 3.8%. Indicator must be atstandstill to zero the scale. Use 1.9% for legal-for-trade applications.

Table 3-2. Configuration Menu Parameters

XXXXXXXXXXXXXX

GRADS

10000 OFF

1D

0.5D

3D

ZTRKBN ZRANGE

1.9%

100%

OFF

2D

1D

3D

MOTBAN

10D

5D

20D

FS+2%

FS+9D

FS+1D

FS

OVRLOA

1

2

4

8

DIGFL1

32

16

64

number

1

2

4

8

DIGFL2

32

16

64

1

2

4

8

DIGFL3

32

16

64

BOTH

PBTARE

NOTARE

KEYED

TAREFN

VERSPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

8OUT

32OUT

16OUT

64OUT

DFSENS DFTHRH

2OUT

128OUT

4OUT

2DD

NONE

5DD

20DD

10DD

50DD

200DD

100DD

250DD

15HZ

3.75HZ

30HZ

7.5HZ

SMPRAT

Configuration 13

Page 18: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

MOTBAN 1D2D3D5D10D20DOFF

Motion band. Sets the level, in display divisions, at which scale motion is detected. Ifmotion is not detected for 1 second or more, the standstill symbol lights. Someoperations, including print, tare, and zero, require the scale to be at standstill. Maximumlegal value varies depending on local regulations.

If OFF is selected, ZTRKBN is disabled.

OVRLOA FS+2%FS+1DFS+9DFS

Overload. Determines the point at which the display blanks and an out-of-range errormessage is displayed. Maximum legal value varies depending on local regulations.

SMPRAT 15HZ30HZ3.75HZ7.5HZ

Sample rate. Selects measurement rate, in samples per second, of the analog-to-digitalconverter. Lower sample rate values provide greater signal noise immunity.

DIGFL1DIGFL2DIGFL3

4816326412

Digital filtering. Selects the digital filtering rate used to reduce the effects of mechanicalvibration from the immediate area of the scale.

Choices indicate the number of A/D conversions that are averaged to obtain thedisplayed reading. A higher number gives a more accurate display by minimizing theeffect of a few noisy readings, but slows down the settling rate of the indicator. SeeSection 8.6 on page 37 for more information on digital filtering.

DFSENS 8OUT16OUT32OUT64OUT128OUT2OUT4OUT

Digital filter cutout sensitivity. Specifies the number of consecutive readings that must falloutside the filter threshold (DFTHRH parameter) before digital filtering is suspended.

DFTHRH NONE2DD5DD10DD20DD50DD100DD200DD250DD

Digital filter cutout threshold. Specifies the filter threshold, in display divisions. When aspecified number of consecutive scale readings (DFSENS parameter) fall outside of thisthreshold, digital filtering is suspended. If NONE is selected, the filter is always enabled.

TAREFN BOTHNOTAREPBTAREKEYED

Tare function. Enables or disables push-button and keyed tares. Possible values are:

BOTH: Both push-button and keyed tares are enabledNOTARE: No tare allowed (gross mode only)PBTARE: Push-button tares enabledKEYED: Keyed tare enabled

CONFIG Menu

Parameter Choices Description

Table 3-2. Configuration Menu Parameters (Continued)

14 9390 Installation Manual

Page 19: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

3.2.2 Format Menu

Figure 3-6. Format Menu

FORMAT Menu

Parameter Choices Description

Level 2 submenus

PRIMAR DECPNTDSPDIVUNITS

Specifies the decimal position, display divisions, and units used for the primary units. SeeLevel 3 submenu parameter descriptions.

SECNDR DECPNTDSPDIVUNITSMULT

Specifies the decimal position, display divisions, units, and conversion multiplier used forthe secondary units. See Level 3 submenu parameter descriptions.

DSPRAT 250MS500MS750MS1500MS2500MS1SEC2SEC3SEC4SEC6SEC8SEC

Display rate. Sets the update rate for displayed values. Values are in milliseconds (MS) orseconds (SEC).

Table 3-3. Format Menu Parameters

XXXXXXXXXXXXXX

DSPDIV MULTUNITSDECPNT

0.453595D

2D

1D number

PRIMAR SECNDR

6SEC

4SEC

3SEC

2500MS

2SEC

1500MS

750MS

500MS

250MS

8SEC

DSPRAT

DSPDIV UNITSDECPNT

OZ

KG888880

888888 1D

5D

2D

LB

88888.8

8888.88

888.888

88.8888

8.88888 888880

888888

88888.8

8888.88

888.888

88.8888

8.88888

T

TN

NONE

G

TN

OZ

KG

G

T

LB

NONE

VERSPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

1SEC

Configuration 15

Page 20: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

Level 3 submenus

Primary Units (PRIMAR Parameter)

DECPNT 8888888888808.8888888.8888888.8888888.8888888.8

Decimal point location. Specifies the location of the decimal point or dummy zeroes in theprimary unit display. Value should be consistent with local legal requirements.

DSPDIV 1D2D5D

Display divisions. Selects the minimum division size for the primary units displayed weight.

UNITS LBKGOZTNTGNONE

Specifies primary units for displayed and printed weight. Values are: LB=pound;KG=kilogram; OZ=ounce; TN=short ton; T=metric ton; G=gram.

NOTE: Indicators sold outside North America are configured with KG for both primary andsecondary units.

Secondary Units (SECNDR Parameter)

DECPNT 88888.88888888888808.8888888.8888888.8888888.88

Decimal point location. Determines the location of the decimal point or dummy zeros inthe secondary unit display.

DSPDIV 5D1D2D

Display divisions. Selects the value of minimum division size of the displayed weight.

UNITS KGOZTNTGLBNONE

Specifies secondary units for displayed and printed weight. Values are: KG=kilogram;OZ=ounce; TN=short ton; T=metric ton; G=gram; LB=pound.

MULT 0.45359Enter other choices via keyboard

Multiplier. Specifies the conversion factor by which the primary units are multiplied toobtain the secondary units. The default is 0.45359, which is the conversion factor forchanging pounds to kilograms. See Section 8.5 on page 36 for a list of multipliers.

To toggle between primary and secondary units, press the UNITS key.

FORMAT Menu

Parameter Choices Description

Table 3-3. Format Menu Parameters (Continued)

16 9390 Installation Manual

Page 21: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

3.2.3 Calibration MenuSee Section 4.0 on page 22 for calibration procedures.

Figure 3-7. Calibration Menu

CALIBR Menu

Parameter Choices Description

Level 2 submenus

WZERO — Calibrate zero, display and edit the zero calibration A/D count value.

WVAL — Display and edit the test weight value.

WSPAN — Calibrate span, display and edit the span calibration A/D count value.

REZERO — Press Enter to remove an offset value from the zero and span calibrations.

Use this parameter only after WZERO and WSPAN have been set. See Section 4.1 onpage 22 for more information about using this parameter.

Table 3-4. Calibration Menu Parameters

XXXXXXXXXXXXXX

WZERO

*CAL*Display and editzero calibrationA/D count value

Display and edittest weight value

WVAL

*CAL*Display and editspan calibrationA/D count value

WSPAN

VERSPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

*CAL*Press Enter to

remove offset fromzero and span

calibrations

REZERO

Configuration 17

Page 22: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

3.2.4 Serial MenuSee Section 8.2 on page 32 for information about the 9390 serial data format.

Figure 3-8. Serial Menu

XXXXXXXXXXXXXX

EOLDLY

000000

number

BITS TERMINBAUD

CR

9600

7ODD

7EVEN

8NONE CR/LF

4800

2400

1200

600

300

LFT

OFF

STREAM

INDUST

VERSPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

19200

SERIAL Menu

Parameter Choices Description

Level 2 Submenus

BAUD 960019200300600120024004800

Baud rate. Selects the transmission speed for the EDP port.

BITS 8NONE7EVEN7ODD

Selects number of data bits and parity of data transmitted from the EDP port.

TERMIN CR/LFCR

Termination character. Selects termination character for data sent from the EDP port.

EOLDLY 0number

End-of-line delay. Sets the delay period, in 0.1-second intervals, from when a formatted line isterminated to the beginning of the next formatted serial output. Value specified must be in therange 0-255, in tenths of a second (10 = 1 second).

STREAM OFFLFTINDUST

Selects whether continuous data output is streamed from the serial port. Select LFT for legal-for-trade applications in which the streamed data must match the current displayed weight. SelectINDUST to stream the latest weight data, regardless of the value displayed.

See Section 8.2 on page 32 for information about the 9390 continuous data format.

Table 3-5. Serial Menu Parameters

18 9390 Installation Manual

Page 23: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

3.2.5 Program Menu

Figure 3-9. ProgramMenu

XXXXXXXXXXXXXX

PWRUPM

GO NONE

OIML

REGULA

000000

number

CONSNU

DELAY

VERSPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

000000

number

CONSTU

NTEP

CANADA

1

number

UIDSTNDBY COUNT

SBYDLY LSTMOD ACCESS SPLSIZ

10

20

50

100

5

DSABLE

ENABLE

0

number

YES

NO

PROGRM Menu

Parameter Choices Description

Level 2 submenus

PWRUPM GODELAY

Power up mode. In GO mode, the indicator goes into operation immediately after a brief powerup display test.

In DELAY mode, the indicator performs a power up display test, then enters a 30-secondwarm up period. If no motion is detected during the warm up period, the indicator becomesoperational when the warm up period ends; if motion is detected, the delay timer is reset andthe warm up period repeated.

STNDBY SBYDLYLSTMOD

Specifies standby mode delay and whether indicator returns to the last operating mode orperforms a power-up sequence when standby mode ends. See Level 3 submenu parameterdescriptions.

COUNT ACCESSSPLSIZ

Specifies whether operator has access to piece count mode and the default sample size usedfor parts counting. See Level 3 submenu parameter descriptions.

REGULA NONEOIMLNTEPCANADA

Regulatory mode. Specifies the regulatory agency having jurisdiction over the scale site.

OIML and CANADA modes require the scale load to be at zero before clearing a tare.

Table 3-6. Program Menu Parameters

Configuration 19

Page 24: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

CONSNU 000000number

Consecutive numbering. Allows sequential numbering for print operations. The consecutivenumber value is incremented following each print operation.

The initial value of this parameter is set to the start up value specified on the CONSTUparameter. Changing either CONSTU or CONSNU immediately resets the consecutive numberused for printing.

CONSTU 000000number

Consecutive number start up value. Specifies the initial consecutive number (CONSNU) valueused when the indicator is powered on.

UID 1number

Unit ID. Specify a unit identifier for the indicator in the range 1–999999. The unit ID can beadded to print ticket formats to identify the indicator used to generate the ticket.

Level 3 submenus

SBYDLY 0number

Standby mode delay. Specifies the number of minutes the indicator must be inactive beforeentering standby mode. Valid values are 0 (off) or 1–120 minutes.

The indicator enters standby mode if no key presses, serial communications, or scale motionoccur for the length of time specified on this parameter. Set this parameter to 0 to disablestandby mode.

LSTMOD YESNO

Last mode. Specifies whether the indicator resumes operation in the last mode it was in beforeentering standby mode (YES) or performs the standard power-up sequence (NO).

ACCESS DSABLEENABLE

Operator access to piece count mode. Specify DSABLE if piece count mode will not be used.With access disabled, pressing the GROSS/NET (MODE) key toggles between gross and netmodes only.

SPLSIZ 1020501005

Sample size. Specify the default sample size used for counting scale operations. Sample sizecan be changed in counting mode during sample acquisition.

PROGRM Menu

Parameter Choices Description

Table 3-6. Program Menu Parameters (Continued)

20 9390 Installation Manual

Page 25: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

ametersmber.

3.2.6 Print Format MenuSee Section 7.0 on page 29 for information about custom print formatting.

Figure 3-10. Print Format Menu

3.2.7 Version MenuThe VERS menu is used to check the software version installed in the indicator. There are no parassociated with the Version menu: when selected, the indicator displays the installed software version nu

Figure 3-11. Version Menu

XXXXXXXXXXXXXXXXXXXXX

GFMT

Display first 6characters of format

Same as GFMT

NFMT

VERSPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

Display and editactive character and

ASCII value

Delete activecharacter

Scroll right in format stringScroll left in format string

Increment ASCII value of active characterDecrement ASCII value of active character

Press to insert a spacebefore the active character

CFMT

XXXXXXXXXXXXXXXXXXXXX

Softwareversion

VERSPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

Configuration 21

Page 26: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

nht

-2 toon

e

anpt

are

heeto

gsttisns

e,

re

4.0 CalibrationThe 9390 can be calibrated using the front panel, EDP commands, or the Revolution™ configuration utility. Eachmethod consists of the following steps:

• Zero calibration• Entering the test weight value• Span calibration• Optional rezero calibration for test weights using hooks or chains.

The following sections describe the calibration procedure for each of the calibration methods.

Figure 4-1. Calibration (CALIBR) Menu

4.1 Front Panel CalibrationTo calibrate the indicator using the front panel, do thefollowing:

1. Place the indicator in setup mode (displayreads CONFIG) and remove all weight fromthe scale platform. If your test weights requirehooks or chains, place the hooks or chains onthe scale for zero calibration.

2. Press until the display reads CALIBR (seeFigure 4-1). Press to go to zero calibration(WZERO).

3. With WZERO displayed, press to calibratezero. The indicator displays *CAL* whilecalibration is in progress. When complete, theA/D count for the zero calibration isdisplayed. Press again to save the zerocalibration value and go to the next prompt(WVAL).

Figure 4-2. Editing Procedure for Numeric Values

4. With WVAL displayed, place test weights othe scale and press to show the test weigvalue. Use the procedure shown in Figure 4to enter the actual test weight, then press save the value and go to span calibrati(WSPAN).

5. With WSPAN displayed, press to calibratespan. The indicator displays *CAL* whilecalibration is in progress. When complete, thA/D count for the span calibration isdisplayed. Press again to save the spcalibration value and go to the next prom(REZERO).

6. The rezero function is used to remove calibration offset when hooks or chains aused to hang the test weights. • If no other apparatus was used to hang t

test weights during calibration, removthe test weights and press to return the CALIBR menu.

• If hooks or chains were used durincalibration, remove these and the teweights from the scale. With all weighremoved, press to rezero the scale. Thfunct ion adjusts the zero and spacalibration values. The indicator display*CAL * wh i le the ze ro and spancalibrations are adjusted. When completthe adjusted A/D count for the zerocalibration is displayed. Press to entethe value, then press to return to thCALIBR menu.

7. Press until the display reads CONFIG, thenpress to exit setup mode.

XXXXXXXXXXXXXX

WZERO

*CAL*Display and editzero calibrationA/D count value

Display and edittest weight value

WVAL

*CAL*Display and editspan calibrationA/D count value

WSPAN

VERSPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

*CAL*Press Enter to

remove offset fromzero and span

calibrations

REZERO

When editing numeric values, press or to change thedigit selected. Press or to increment or decrement thevalue of the selected digit. Press to save the value.

2 0 0 0 0 0

After entering the value, press or to change the decimalpoint location. Press again to save the value and exit theparameter.

2 0 0 0 0 0

22 9390 Installation Manual

Page 27: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

thet

te

htedO

re

it

ue

rale

anen.

eets

inn

4.2 EDP Command CalibrationTo calibrate the indicator using EDP commands, theindicator EDP port must be connected to a terminal orpersonal computer. See Section 2.3.3 on page 5 forEDP port pin assignments; see Section 5.0 on page 24for more information about using EDP commands.

Once the indicator is connected to the sending device,do the following:

1. Place the indicator in setup mode (displayreads CONFIG) and remove all weight fromthe scale platform. If your test weights requirehooks or chains, place the hooks or chains onthe scale for zero calibration.

2. Send the WZERO EDP command to calibratezero. The indicator displays *CAL* whilecalibration is in progress.

3. Place test weights on the scale and use WVAL command to enter the test weighvalue in the following format:

WVAL=nnnnnn<CR>

4. Send the WSPAN EDP command to calibraspan. The indicator displays *CAL* whilecalibration is in progress.

5. To remove an offset value, clear all weigfrom the scale, including hooks or chains usto hang test weights, then send the REZEREDP command. The indicator displays *CAL*while the zero and span calibrations aadjusted.

6. Send the KUPARROW EDP command to exsetup mode.

4.3 Revolution™ CalibrationTo calibrate the indicator using Revolution, theindicator EDP port must be connected to a PC runningthe Revolution configuration utility.

Use the following procedure to calibrate the indicator:

1. Select Calibrate Indicator from the Revolutionmain menu.

2. On the Indicator Calibration display, selectthe indicator model and communications port,then click OK.

3. Revolution uploads calibration data from theindicator then presents the information in adisplay like that shown in Figure 4-3.

Figure 4-3. Revolution Calibration Display

4. Enter the Value of Test Weight to be used forspan calibration then click OK.

5. The Zero Calibration dialog box prompts yoto remove all weight from the scale. Clear thscale and click OK to begin zero calibration.NOTE: If your test weights require hooks ochains, place the hooks or chains on the scfor zero calibration.

6. When zero calibration is complete, the SpCalibration dialog box prompts you to plactest weights on the scale for span calibratioPlace tests weights on the scale then click OK.

7. When span calibration is complete, thRezero dialog box prompts you to removweights from the scale. Remove the weighthen click OK.

8. When calibration is complete, the NewSettings fields of the Indicator Calibrationdisplay are filled in. Click Exit to save the newvalues and return to the Revolution mamenu; to restore the previous calibratiovalues, click Restore Settings.

Calibration 23

Page 28: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

5.0 EDP CommandsThe 9390 indicator can be controlled by a personalcomputer or remote keyboard connected to theindicator EDP port. Control is provided by a set ofEDP commands that can simulate front panel keypress functions, display and change setup parameters,and perform reporting functions. The EDP portprovides the capability to print configuration data or tosave that data to an attached personal computer. Thissection describes the EDP command set andprocedures for saving and transferring data using theEDP port.

5.1 The EDP Command SetThe EDP command set can be divided into fivegroups: key press commands, reporting commands,the RESETCONFIGURATION special functioncommand, parameter setting commands, and transmitweight data commands.

When the indicator processes an EDP command, itresponds with the message OK. The OK responseverifies that the command was received and has beenexecuted. If the command is unrecognized or cannotbe executed, the indicator responds with ??.

The following sections list the commands andcommand syntax used for each of these groups.

5.1.1 Key Press CommandsKey press EDP commands (see Table 5-1) simulatepressing the keys on the front panel of the indicator.These commands can be used in both setup andweighing mode. Several of the commands serve as“pseudo” keys, providing functions that are notrepresented by a key on the front panel.

For example, to enter a 15-pound tare weight usingEDP commands:

1. Type K1 and press ENTER (or RETURN).2. Type K5 and press ENTER.3. Type KTARE and press ENTER.

The indicator displays the symbol when a tare hasbeen entered and shifts the display to net mode.

Command Function

KZERO Press the ZERO key

KGROSSNET Press the GROSS/NET key

KGROSS Go to gross mode (pseudo key)

KNET Go to net mode (pseudo key)

KCOUNT Go to piece count mode (pseudo key)

KTARE Press the TARE key

KUNITS Press the UNITS key

KPRIM Go to primary units (pseudo key)

KSEC Go to secondary units (pseudo key)

KPRINT Press the PRINT key

KLEFTARROW In setup mode, move left in the menu

KRIGHTARROW In setup mode, move right in the menu

KUPARROW In setup mode, move up in the menu

KDOWNARROW In setup mode, move down in the menu

K0 Press number 0 (zero)

K1 Press number 1

K2 Press number 2

K3 Press number 3

K4 Press number 4

K5 Press number 5

K6 Press number 6

K7 Press number 7

K8 Press number 8

K9 Press number 9

KDOT Press the decimal point (.)

KENTER Press the ENTER key

Table 5-1. Key Press EDP Commands

24 9390 Installation Manual

Page 29: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

T)Tonnd

N

oron

being

oden

e.ou

you

5,

5.1.2 Reporting CommandsReporting commands (see Table 5-2) send specificinformation to the EDP port. These commands can beused in both setup mode and normal mode.

5.1.3 The RESETCONFIGURATION CommandThe RESETCONFIGURATION command can beused to restore all configuration parameters to theirdefault values. Before issuing this command, theindicator must be placed in setup mode or test mode(press and hold setup switch for approximately fourseconds to enter test mode).

Sending this command, followed by the RS (RESEEDP command, is equivalent to using the DEFLfunction on the TEST menu. See Section 8.7 page 38 for more information about test mode ausing the TEST menu. NOTE: All load cell calibrationsettings are lost when the RESETCONFIGURATIOcommand is run.

5.1.4 Parameter Setting CommandsParameter setting commands allow you to displaychange the current value for a particular configuratiparameter (Tables 5-3 through Table 5-8).

Current configuration parameter settings can displayed in either setup mode or normal mode usthe following syntax:

command<ENTER>

Most parameter values can be changed in setup monly. Use the following command syntax whechanging parameter values:

command=value<ENTER>

where value is either a number or a parameter valuUse no spaces before or after the equal (=) sign. If ytype an incorrect command, the display reads ??.Changes to the parameters do not take effect until exit setup mode.

For example, to set the motion band parameter totype the following:

MOTBAND=5D<ENTER>

Command Function

DUMPALL List all parameter values

VERSION Write 9390 software version

RS Reset the indicator

P Write current displayed weight with units identifier. The P command returns the current displayed weight value to the EDP port, along with the units identifier. If the indicator is in an underrange or overload condition, the weight value is replaced with ###### (overload) or %%%%%% (underrange).

S Write one frame of stream format

Table 5-2. EDP Reporting Commands

Command Description Values

GRADS Graduations 1–999 999

ZTRKBND Zero track band OFF, 0.5D, 1D, 3D

ZRANGE Zero range 1.9%, 100%

MOTBAND Motion band 1D, 2D, 3D, 5D, 10D, 20D, OFF

OVRLOAD Overload FS+2%, FS+1D, FS+9D, FS

SMPRAT Sample rate 15HZ, 30HZ, 3.75HZ, 7.5HZ

DIGFLTR1DIGFLTR2DIGFLTR3

Digital filtering 1, 2, 4, 8, 16, 32, 64

DFSENS Digital filter cutout sensitivity 2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT

DFTHRH Digital filter cutout threshold NONE, 2DD, 5DD, 10DD, 20DD, 50DD, 100DD, 200DD, 250DD

TAREFN Tare function BOTH, NOTARE, PBTARE, KEYED

Table 5-3. CONFIG EDP Commands

EDP Commands 25

Page 30: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

Command Description Values

PRI.DECPNT Primary units decimal position 8.88888, 88.8888, 888.888, 8888.88, 88888.8, 888888, 888880

PRI.DSPDIV Primary units display divisions 1D, 2D, 5D

PRI.UNITS Primary units LB, KG, OZ, TN, T, G, NONE

SEC.DECPNT Secondary units decimal position 8.88888, 88.8888, 888.888, 8888.88, 88888.8, 888888, 888880

SEC.DSPDIV Secondary units display divisions 1D, 2D, 5D

SEC.UNITS Secondary units LB, KG, OZ, TN, T, G, NONE

SEC.MULT Secondary units multiplier 0.00000–999999

DSPRATE Display rate 250MS, 500MS, 750MS, 1500MS, 2500MS, 1SEC, 2SEC, 3SEC, 4SEC, 6SEC, 8SEC

Table 5-4. FORMAT EDP Commands

Command Description Values

WZERO Zero calibration —

WVAL Test weight value test_weight_value

WSPAN Span calibration —

REZERO Rezero —

LC.CD Set deadload coefficient value

LC.CW Set span coefficient value

Table 5-5. CALIBR EDP Commands

Command Description Values

EDP.BAUD Baud rate 300, 600, 1200, 2400, 4800, 9600, 19200

EDP.BITS Data bits/parity 8NONE, 7EVEN, 7ODD

EDP.TERMIN Termination character CR/LF, CR

EDP.EOLDLY End-of-line delay 0–255 (0.1-second intervals)

EDP.STREAM Stream format OFF, LFT, INDUST

Table 5-6. SERIAL EDP Commands

Command Description Values

PWRUPMD Power up mode GO, DELAY

SBYDLY Standby delay 0–120

LSTMOD Last mode YES, NO

CNTMOD Counting mode ENABLE, DSABLE

SAMPSIZ Default sample size 5, 10, 20, 50, 100

REGULAT Regulatory compliance NONE, OIML, NTEP, CANADA

CONSNUM Consecutive number 0–999 999

CONSTUP Consecutive number start-up value 0–999 999

UID Unit identifier 0–999 999

Table 5-7. PROGRM EDP Commands

26 9390 Installation Manual

Page 31: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

transmit

beisithr is

thetheonsthehe

er.

ntnot

ehes

oneteSed

5.1.5 Normal Mode CommandsThe serial transmit weight data commands (see Table 5-9) transmit data to the EDP port on demand. Theweight data commands are valid only in normal operating mode.

5.2 Saving and Transferring DataConnecting a personal computer to the 9390 EDP portallows you to save indicator configuration data to thePC or to download configuration data from the PC toan indicator. The following sections describe theprocedures for these save and transfer operations.

5.2.1 Saving Indicator Data to a Personal ComputerConfiguration data can be saved to a personal computerconnected to the EDP port. The PC must be running acommunications program such as PROCOMMPLUS®.See Section 2.3.3 on page 5 for information about serialcommunications wiring and EDP port pin assignments.

When configuring the indicator, ensure that the valuesset for the BAUD and BITS parameters on the SERIALmenu match the baud rate, bits, and parity settingsconfigured for the serial port on the PC.

To save all configuration data, place the indicator insetup mode and send the DUMPALL EDP command tothe indicator. The 9390 responds by sending allconfiguration parameters to the PC as ASCII-formattedtext.

5.2.2 Downloading Configuration Data from PC toIndicator

Configuration data saved on a PC or floppy disk candownloaded from the PC to an indicator. Thprocedure is useful when a number of indicators wsimilar configurations are set up or when an indicatoreplaced.

To download configuration data, connect the PC to EDP port as described in Section 5.2.1. Place indicator in setup mode and use the PC communicatisoftware to send the saved configuration data to indicator. When transfer is complete, calibrate tindicator as described in Section 4.0 on page 22.

NOTES: • Calibration settings are included in th

configuration data downloaded to the indicatoIf the receiving indicator is a direct replacemefor another 9390 and the attached scale is changed, recalibration is not required.

• When downloading configurations that includchanged serial communications settings, edit tdata file to place the serial communicationchanges at the end of the file. Communicatibetween the PC and indicator will be lost oncthe indicator receives settings for baud ra(BAUD parameter) or data bits and parity (BITparameter) that do not match those configurfor the PC.

Command Description Values

GFMT Gross demand print format string See Section 7.0 on page 29 for detailed information

NFMT Net demand print format string

CFMT Counting mode print format string

Table 5-8. PFORMT EDP Commands

Command Description Response Format

SX Start EDP streaming OK or ??

EX Stop EDP streaming OK or ??

XG Transmit gross weight in displayed units nnnnnn UU

where nnnnnn is the weight value, UU is the units.

NOTE: In count mode, displayed units = primary units; non-displayed units = secondary units

XN Transmit net weight in displayed units

XT Transmit tare weight in displayed units

XG2 Transmit gross weight in non-displayed units

XN2 Transmit net weight in non-displayed units

XT2 Transmit tare weight in non-displayed units

XC Transmit current part count nnnnnn

Table 5-9. Normal Mode EDP Commands

EDP Commands 27

Page 32: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

28 9390 Installation Manual

6.0 Counting OperationsThe 9390 provides a piece count mode that allows theindicator to serve as a portable counting scale. Inpiece count mode, the indicator display shows thenumber of parts on the scale rather than the weight ofthose parts.

NOTE: Legal-for-trade applications require product tobe sold by weight rather than an equivalent parts count.Operator access to piece count mode is disabled whenthe indicator is shipped from the factory.

To enable operator access to piece count mode, set theACCESS parameter to ENABLE (COUNT submenuon the PROGRM menu) or send the CNTMOD=ENABLE EDP command.

Piece count mode has two submodes:

• Count display mode displays the current partscount and allows ticket printing using theCFMT print format.

• Sample acquisition mode is used to calibratethe indicator for parts counting.

Count Display ModePiece count mode is entered by pressing the MODE(GROSS/NET) key from normal weighing mode. Incount display mode, the display shows the number ofparts rather than weight, the arrow annunciator at theright side of the LCD display points to piece count(PC), and pressing the PRINT key sends the CFMTprint format data string to the serial port.

If a sample weight has not yet been acquired, theindicator switches automatically to sample acquisitionmode, described below.

Sample Acquisition ModeSample acquisition mode is used to select the samplequantity and calibrate the indicator for the parts beingcounted.

To set a new sample weight, do the following:

1. Place empty parts container on scale. Wait forstandstill, then press TARE to acquire the tareweight of the container.

2. Press MODE to enter piece count mode.3. From count display mode, press the SAMPLE

(UNITS) key to enter sample acquisitionmode.

4. The indicator display shows the messageAddnnn, where nnn is the sample quantity tobe placed on the scale. Add the number ofparts shown or choose a different sample sizeby pressing the SAMPLE key to scroll throughthe selectable sample quantities (5, 10, 20, 50,100).

5. Once the sample quantity is on the scale,press ENTER to calibrate the indicator forcounting the new parts. The indicator displayshows the message –CNT– as it acquires thesample weight, then switches to count displaymode and shows the part quantity.If the weight of the sample quantity isinsufficient for piece counting, the indicatorautomatically requests the next higher samplequantity by showing the Addnnn messageagain. If the sample quantity is at maximum(100) and the sample weight is still notsufficient, the indicator shows the messageCNTERR.

To view gross and net weight of parts, press MODE toswitch from count display mode to normal weighingmode.

Figure 6-1. Front Panel Key Functions in Piece Count Mode

T

ZERO GROSSNET

TARE UNITS PRINT

TB/G Units

POWER

ENTER SAMPLEMODE PRINT

I/O

New sample /change sample quantity

Send data toserial port

(CFMT format)

Not used

Switch between gross, net,and piece count mode

Acquire tare /acquire sample

COUNTING MODE KEY FUNCTIONS

Page 33: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

lyfor>,

7-

dree

ory>,

gD

ted

>3ea

ors),

7.0 Print FormattingThe 9390 provides three print formats, GFMT, NFMT,and CFMT that determine the format of the printedoutput when the PRINT key is pressed or when aKPRINT EDP command is received. If no tare hasbeen entered, the GFMT print format is used; if a tarehas been entered, NFMT is used; in counting mode,CFMT is used.

Each print format can be customized to include up to300 characters of information, such as company nameand address, on printed tickets. You can use theindicator front panel (PFORMT menu), EDPcommands, or the Revolution™ configuration utility tocustomize the print formats.

7.1 Print Formatting CommandsTable 7-1 lists commands you can use to format thegross and net print formats. Commands included inthe format strings must be enclosed between < and >delimiters. Any characters outside of the delimitersare printed as text on the ticket. Text characters caninclude any ASCII character that can be printed by theoutput device.

The default GFMT and NFMT print formats use onthe new line (<NL>) command and the commands gross, net, and tare weights in displayed units (<G<N>, and <T>).

The default 9390 print formats are shown in Table2:

NOTES:

• The <G2>, <N2>, and <T2> commands listein Table 7-1 print the gross, net, and taweights in non-displayed units—that is, in thunits not currently displayed on the indicator.In count count mode, formatting commands fdisplayed units (<G>, <N>, <T>) use primarunits, non-displayed units commands (<G2<N2>, <T2>) use secondary units.

• ID numbers included in the print format strin(<ID> command) must be set using the UIEDP command.

• The 300-character limit of each print formastring includes the output field length of thprint formatting commands, not the commanlength. For example, if the indicator isconfigured to show a decimal point, the <Gcommand generates an output field of 1characters: the 10-character weight valu(including decimal point), one space, and two-digit units identifier.

7.2 Customizing Print FormatsThe following sections describe procedures fcustomizing the GFMT, NFMT, and CFMT formatusing the EDP port, the front panel (PFORMT menuand the Revolution configuration utility.

Command Description

<G> Gross weight in displayed units

<G2> Gross weight in non-displayed units

<N> Net weight in displayed units

<N2> Net weight in non-displayed units

<T> Tare weight in displayed units

<T2> Tare weight in non-displayed units

<C> Current piece count

<ID> Unit identifier

<CN> Consecutive number

<NLnn> New line (nn = number of termination (<CR/LF> or <CR>) characters)*

<SPnn> Space (nn = number of spaces)*

<SU> Toggle weight data format (formatted/unformatted)**

Gross, net, and tare weights are 10 digits in length, including sign, followed by a space and a two-digit units identifier. Total field length with units identifier is 13 characters.

ID and consecutive number (CN) fields are 1–6 characters in length, as required.

* If nn is not specified, 1 is assumed. Value must be in the range 1–99.

Table 7-1. Print Format Commands

** After receiving an SU command, the indicator sends unformatted data until the next SU command is received. Unformatted data omits decimal points, leading and trailing characters.

Format Default Format String Sample Output

GFMT <G> GROSS<NL> 2046.81 LB GROSS

NFMT <G> GROSS<NL><T> TARE<NL><N> NET<NL>

4053.1 LB GROSS15.6 LB TARE4037.5 LB NET

CFMT <C> PIECES<NL> 512 PC PIECES

NOTE: In OIML and CANADA modes, the letters PT (preset tare) are automatically inserted after the printed tare weight.

Table 7-2. Default Print Formats

Command Description

Table 7-1. Print Format Commands

Print Formatting 29

Page 34: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

e

rgyouto

nthe

ersthe asII

thee

7.2.1 Using the EDP Port With a personal computer, terminal, or remotekeyboard attached to the 9390 EDP port, you can usethe EDP command set to customize the print formatstrings.

To view the current setting of a format string, type thename of the string (GFMT, NFMT, or CFMT) andpress ENTER. For example, to check the currentconfiguration of the GFMT format, type GFMT andpress ENTER. The indicator responds by sending thecurrent configuration for the gross format:

GFMT=<G> GROSS<NL>

To change the format, use the GFMT, NFMT, orCFMT EDP command followed by an equals sign (=)and the modified print format string. For example, toadd the name and address of a company to the grossformat, you could send the following EDP command:

GFMT=MLC SHIPPING<NL>1357 WEST HOWARD ROAD<NL>BIGTOWN<NL2><G> GROSS<NL>

A ticket printed using this format might look like thfollowing:

MLC SHIPPING1357 WEST HOWARD ROADBIGTOWN

27862 LB GROSS

7.2.2 Using the Front PanelIf you have no access to equipment focommunication through the EDP port or are workinat a site where such equipment cannot be used, can use the PFORMT menu (see Figure 7-1) customize the print formats.

Using the PFORMT menu, you can edit the priformat strings by changing the decimal values of tASCII characters in the format string.

NOTE: Lower-case letters and some special charactcannot be displayed on the 9390 front panel (see ASCII character chart on page 33) and are shownblanks. The 9390 can send or receive any ASCcharacter; the character printed depends on particular ASCII character set implemented for threceiving device.

Figure 7-1. PFORMT Menu, Showing Alphanumeric Character Entry Procedure

XXXXXXXXXXXXXXXXXXXXX

GFMT

Display first 6characters of format

Same as GFMT

NFMT

VERSPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

Display and editactive character and

ASCII value

Delete activecharacter

Scroll right in format stringScroll left in format string

Increment ASCII value of active characterDecrement ASCII value of active character

Press to insert a spacebefore the active character

CFMT

30 9390 Installation Manual

Page 35: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

7.2.3 Using RevolutionThe Revolution configuration utility provides a printformatting grid with a tool bar. The grid allows you toconstruct the print format without the formattingcommands (<NL> and <SP>) required by the frontpanel or EDP command methods. Using Revolution,you can type text directly into the grid, then selectweight value fields from the tool bar and place themwhere you want them to appear on the printed ticket.

Figure 7-2 shows an example of the Revolution printformatting grid.

Figure 7-2. Revolution Print Format Grid

Print Formatting 31

Page 36: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

sts these

ameter

8.0 Appendix

8.1 Error MessagesThe 9390 provides a number of front panel error messages to assist in problem diagnosis. Table 8-1 limessages and their meanings.

8.2 Continuous Output (Stream) FormatFigure 8-1 shows the continuous output format sent to the 9390 serial port when the STREAM par(SERIAL menu) is set to either LFT or INDUST.

Figure 8-1. Continuous Output Data Format

Error Message Description Solution

ADCERR A/D not responding Call Interface, Inc. Service.

ADCKSM A/D coefficient checksum error

CFCKSM Configuration checksum error Use TEST menu to perform DEFLT (restore defaults) procedure, then recalibrate load cells.

CNTERR Count error; insufficient sample size Increase sample size.

DSPERR Display error Call Interface Service.

EEPERR EEPROM write error

LCCKSM Load cell calibration checksum error Recalibrate load cells.

LCPERR Load cell power failure Replace batteries. If error still shown, call Interface Service.

PFCKSM Print format checksum error Check print formats for errors.

RAMERR RAM test failed Call Interface Service.

TARERR EEPROM read error Use TEST menu to perform DEFLT (restore defaults) procedure, then recalibrate load cells.

VIREE EEPROM error Use TEST menu to perform DEFLT (restore defaults) procedure, then recalibrate load cells and reconfigure indicator.

------ Overflow error Weight value too large (> 999999) to be displayed or gross value exceeds overload limit. Check configuration.

______ A/D underrange Weight value too small (< –99999) to be displayed or A/D reading < –2.5 mV. Check scale for binding or damage.

STNDBY Standby mode Press any key to return to operating mode.

Table 8-1. 9390 Error Messages

<STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM>

ASCII 02decimal

Polarity:<space> = Positive<–> = Negative

Weight: 7 digits, right-justified, dummyzeroes, decimal point with no leadingzeroes except for leading zero immediatelypreceding the decimal point. Leadingzeroes transmitted as spaces.

L = poundsK = kilogramsT = metric tons (T)G = grams<space> =

OZ, TN, NONE

G = GrossN = Net

Status:<space> = validI = InvalidM = MotionO = Over/under range

<CR> <LF>or <CR>

32 9390 Installation Manual

Page 37: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

ings one output

splay is

8.3 ASCII Character ChartUse the decimal values for ASCII characters listed in Tables 8-2 and 8-3 when specifying print format strthe 9390 PFORMT menu. The actual character printed depends on the character mapping used by thdevice. The 9390 can send or receive any ASCII character value (decimal 0–255), but the indicator dilimited to the set described in Section 8.4 on page 35.

Table 8-2. ASCII Character Chart (Part 1)

Control ASCII Dec Hex ASCII Dec Hex ASCII Dec Hex ASCII Dec Hex

Ctrl-@ NUL 00 00 space 32 20 @ 64 40 ` 96 60

Ctrl-A SOH 01 01 ! 33 21 A 65 41 a 97 61

Ctrl-B STX 02 02 “ 34 22 B 66 42 b 98 62

Ctrl-C ETX 03 03 # 35 23 C 67 43 c 99 63

Ctrl-D EOT 04 04 $ 36 24 D 68 44 d 100 64

Ctrl-E ENQ 05 05 % 37 25 E 69 45 e 101 65

Ctrl-F ACK 06 06 & 38 26 F 70 46 f 102 66

Ctrl-G BEL 07 07 ’ 39 27 G 71 47 g 103 67

Ctrl-H BS 08 08 ( 40 28 H 72 48 h 104 68

Ctrl-I HT 09 09 ) 41 29 I 73 49 i 105 69

Ctrl-J LF 10 0A * 42 2A J 74 4A j 106 6A

Ctrl-K VT 11 0B + 43 2B K 75 4B k 107 6B

Ctrl-L FF 12 0C , 44 2C L 76 4C l 108 6C

Ctrl-M CR 13 0D - 45 2D M 77 4D m 109 6D

Ctrl-N SO 14 0E . 46 2E N 78 4E n 110 6E

Ctrl-O SI 15 0F / 47 2F O 79 4F o 111 6F

Ctrl-P DLE 16 10 0 48 30 P 80 50 p 112 70

Ctrl-Q DC1 17 11 1 49 31 Q 81 51 q 113 71

Ctrl-R DC2 18 12 2 50 32 R 82 52 r 114 72

Ctrl-S DC3 19 13 3 51 33 S 83 53 s 115 73

Ctrl-T DC4 20 14 4 52 34 T 84 54 t 116 74

Ctrl-U NAK 21 15 5 53 35 U 85 55 u 117 75

Ctrl-V SYN 22 16 6 54 36 V 86 56 v 118 76

Ctrl-W ETB 23 17 7 55 37 W 87 57 w 119 77

Ctrl-X CAN 24 18 8 56 38 X 88 58 x 120 78

Ctrl-Y EM 25 19 9 57 39 Y 89 59 y 121 79

Ctrl-Z SUB 26 1A : 58 3A Z 90 5A z 122 7A

Ctrl-[ ESC 27 1B ; 59 3B [ 91 5B 123 7B

Ctrl-\ FS 28 1C < 60 3C \ 92 5C | 124 7C

Ctrl-] GS 29 1D = 61 3D ] 93 5D 125 7D

Ctrl-^ RS 30 1E > 62 3E ^ 94 5E ~ 126 7E

Ctrl-_ US 31 1F ? 63 3F _ 95 5F DEL 127 7F

Appendix 33

Page 38: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

Table 8-3. ASCII Character Chart (Part 2)

ASCII Dec Hex ASCII Dec Hex ASCII Dec Hex ASCII Dec Hex

Ç 128 80 á 160 A0 192 C0 α 224 E0

ü 129 81 í 161 A1 193 C1 β 225 E1

é 130 82 ó 162 A2 194 C2 Γ 226 E2

â 131 83 ú 163 A3 195 C3 π 227 E3

ä 132 84 ñ 164 A4 196 C4 Σ 228 E4

à 133 85 Ñ 165 A5 197 C5 σ 229 E5

å 134 86 ª 166 A6 198 C6 µ 230 E6

ç 135 87 º 167 A7 199 C7 τ 231 E7

ê 136 88 ¿ 168 A8 200 C8 Φ 232 E8

ë 137 89 169 A9 201 C9 Θ 233 E9

è 138 8A ¬ 170 AA 202 CA Ω 234 EA

ï 139 8B 1/2 171 AB 203 CB δ 235 EB

î 140 8C 1/4 172 AC 204 CC ∞ 236 EC

ì 141 8D ¡ 173 AD 205 CD φ 237 ED

Ä 142 8E « 174 AE 206 CE ∈ 238 EE

Å 143 8F » 175 AF 207 CF ∩ 239 EF

É 144 90 176 B0 208 D0 ≡ 240 F0

æ 145 91 177 B1 209 D1 ± 241 F1

Æ 146 92 178 B2 210 D2 ≥ 242 F2

ô 147 93 179 B3 211 D3 ≤ 243 F3

ö 148 94 180 B4 212 D4 ⌠ 244 F4

ò 149 95 181 B5 213 D5 ⌡ 245 F5

û 150 96 182 B6 214 D6 ÷ 246 F6

ù 151 97 183 B7 215 D7 ≈ 247 F7

ÿ 152 98 184 B8 216 D8 ° 248 F8

Ö 153 99 185 B9 217 D9 • 249 F9

Ü 154 9A 186 BA 218 DA 250 FA

¢ 155 9B 187 BB 219 DB 251 FB

£ 156 9C 188 BC 220 DC 252 FC

¥ 157 9D 189 BD 221 DD 2 253 FD

Pts 158 9E 190 BE 222 DE 254 FE

ƒ 159 9F 191 BF 223 DF 255 FF

34 9390 Installation Manual

Page 39: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

390 front

8.4 Front Panel Display CharactersFigure 8-2 shows the 7-segment LCD character set used to display alphanumeric characters on the 9panel.

Figure 8-2. 9390 Display Characters

#

$

%

&

'

(

)

*

+

,

-

. ./

0

1

2

3

4

5

6

7

8

9

:

;

<

=

>

?

@

Aa

Bb

Cc

Dd

Ee

Ff

Gg

Hh

Ii

Jj

Kk

Ll

Mm

Nn

Oo

Pp

Qq

Rr

Ss

Tt

Uu

Vv

Ww

Xx

Yy

Zz

[

\

]

^

_

`

|

~

!

"

Appendix 35

Page 40: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

8.5 Conversion Factors for Secondary UnitsThe 9390 has the capability to mathematically converta weight into many different types of units andinstantly display those results with a press of theUNITS key. Secondary units can be specified on theFORMAT menu using the SECNDR parameter, or byusing EDP commands.

• To configure secondary units using the frontpanel menus, use Table 8-4 to find theconversion multiplier for the MULTparameter. For example, if the primary unit ispounds and the secondary unit is short tons,set the MULT parameter to 0.00050, thenpress to save the value.For secondary units with a conversion factorof 10 or more, enter the value as describedabove without regard to decimal position,then use the or key to move the decimalpoint to the correct position. Press a secondtime to save the value. For example, if the primary unit is ounces andthe secondary unit is grams, the conversionfactor shown in Table 8-4 is 28.3495. To enterthis value using the menus, first enter 2.83495for the MULT parameter, then use the left andright front panel keys to adjust the decimalpoint to 28.3495. Press again to save thevalue and exit the MULT parameter.

• To configure secondary units using EDPcommands, use the Table 8-4 to find theconversion multiplier for the SEC.MULTcommand. For example, if the primary unit ispounds and the secondary unit is short tons,send the EDP command SEC.MULT=0.0005<CR> to set the multiplier for thesecondary units.

• Long tons and grains units listed in Table 8-4cannot be directly specified as primary orsecondary units on the 9390 indicator. Forthese or other unlisted units of weight, specifyNONE on the UNITS parameter.

NOTE: Ensure that the secondary decimal pointposition is set appropriately for the scale capacity inthe secondary units. If the converted value requiresmore digits than are available, the indicator willdisplay an overflow message (OVERFL).

For example, if the primary units are short tons,secondary units are pounds, and the secondarydecimal point is set to 8888.88, the indicator willoverflow if 5 tons or more are applied to the scale.With 5 tons applied, and a conversion factor of 2000,the secondary units display needs five digits to the leftof the decimal point to display the 10000 lb secondaryunits value.

Primary Unit x Multiplier Secondary Unit

grains 0.06480 grams

0.00229 ounces

0.00014 pounds

0.00006 kilograms

ounces 437.500 grains

28.3495 grams

0.06250 pounds

0.02835 kilograms

pounds 7000.00 grains

453.592 grams

16.0000 ounces

0.45359 kilograms

0.00050 short tons

0.00045 long tons

0.00045 metric tons

short tons 2000.00 pounds

907.185 kilograms

0.89286 long tons

0.90718 metric tons

grams 15.4324 grains

0.03527 ounces

0.00220 pounds

0.00100 kilograms

kilograms 15432.4 grains

35.2740 ounces

1000.00 grams

2.20462 pounds

0.00110 short tons

0.00098 long tons

0.00100 metric tons

metric tons 2204.62 pounds

1000.00 kilograms

1.10231 short tons

0.98421 long tons

long tons 2240.00 pounds

1016.05 kilograms

1.12000 short tons

1.01605 metric tons

Table 8-4. Conversion Factors

36 9390 Installation Manual

Page 41: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

reshold jump

toor

lter

veer

yfll

ff

8.6 Digital FilteringThe 9390 uses averaged digital filtering to reduce the effect of vibration on weight readings. Adjustable thand sensitivity functions allow quick settling by suspending filter averaging, allowing the weight reading toto the new value. Figure 8-3 shows the digital filter parameters on the CONFIG menu.

Figure 8-3. Digital Filtering Parameters on the Configuration (CONFIG) Menu

Figure 8-4. Flow Diagram for 9390 Digital Filters

8.6.1 DIGFLx ParametersThe first three digital filtering parameters, DIGFL1,DIGFL2, and DIGFL3, are configurable filter stagesthat control the effect of a single A/D reading on thedisplayed weight. The value assigned to eachparameter sets the number of readings received fromthe preceding filter stage before averaging (seeFigure 8-4).

The overall filtering effect can be expressed bymultiplying the values assigned to the three filterstages:

DIGFL1 x DIGFL2 x DIGFL3

For example, if the filters are configured asDIGFL1=4, DIGFL2=8, DIGFL3=8, the overallfiltering effect is 256 (4 x 8 x 8). With thisconfiguration, each A/D reading has a 1-in-256 effecton the displayed weight value. Setting the filters to 1effectively disables digital filtering (1 x 1 x 1 = 1).

8.6.2 DFSENS and DFTHRH ParametersThe three digital filters can be used by themselveseliminate vibration effects, but heavy filtering alsincreases settling time. The DFSENS (digital filtesensitivity) and DFTHRH (digital filter threshold)parameters can be used to temporarily override fiaveraging and improve settling time:

• DFSENS specifies the number of consecutiscale readings that must fall outside the filtthreshold (DFTHRH) before digital filteringis suspended.

• DFTHRH sets a threshold value, in displadivisions. When a specified number oconsecutive scale readings (DFSENS) faoutside of this threshold, digital filtering issuspended. Set DFTHRH to NONE to turn othe filter override.

1

2

4

8

DIGFL1

32

16

64

1

2

4

8

DIGFL2

32

16

64

1

2

4

8

DIGFL3

32

16

64

8OUT

32OUT

16OUT

64OUT

DFSENS DFTHRH

2OUT

128OUT

4OUT

2DD

NONE

5DD

20DD

10DD

50DD

200DD

100DD

250DD

DIGFL1=n1 DIGFL2=n2 DIGFL3=n3

3rd StageFilter Average

2nd StageFilter Average

1st StageFilter Average

DisplayedValue

n32nd Stage

Filter Averages

n21st Stage

Filter Averages

n1A/D Readings

Appendix 37

Page 42: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

eses

terd

ldn

hernt-et

st

e 9390,

ote that,

8.6.3 Setting the Digital Filter ParametersFine-tuning the digital filter parameters greatlyimproves indicator performance in heavy-vibrationenvironments. Use the following procedure todetermine vibration effects on the scale and optimizethe digital filtering configuration.

1. In setup mode, set all three digital filters(DIGFL1, DIGFL2, DIGFL3) to 1. SetDFTHRH to NONE. Return indicator tonormal mode.

2. Remove all weight from the scale, then watchthe indicator display to determine themagnitude of vibration effects on the scale.Record the weight below which all but a fewreadings fall. This value is used to calculatethe DFTHRH parameter value in Step 4.For example, if a heavy-capacity scaleproduces vibration-related readings of up to50 lb, with occasional spikes to 75 lb, record50 lb as the threshold weight value.

3. Place the indicator in setup mode and set the

digital filters (DIGFLx) to eliminate thevibration effects on the scale. (LeavDFTHRH set to NONE.) Reconfigure anecessary to find the lowest effective valufor the DIGFLx parameters.

4. With optimum values assigned to the DIGFLxparameters, calculate the DFTHRH paramevalue by converting the weight value recordein Step 2 to display divisions:

threshold_weight_value / DSPDIV

In the example in Step 2, with a threshoweight value of 50 lb and a display divisiovalue of 5D: 50 / 5D = 10. DFTHRH shouldbe set to 10DD for this example.

5. Finally, set the DFSENS parameter higenough to ignore transient peaks. Longtransients (typically caused by lower vibratiofrequencies) will cause more consecutive ouof-band readings, so DFSENS should be shigher to counter low frequency transients.Reconfigure as necessary to find the lowe

8.7 Test ModeIn addition to normal and setup modes, test mode provides a number of diagnostic functions for thincluding:

• Display raw A/D count• Reset configuration parameters to default values• Print indicator configuration

To enter test mode, press and hold the setup switch until the front panel display shows the word TEST. After aboutthree seconds, the test mode display automatically shifts to the first test menu function, A/DTST.

Figure 8-5 shows the Test Menu structure; Figure 8-6 shows the front panel key functions in test mode. Nbecause the Test Menu functions are all on a single menu level, the GROSS/NET ( ) key has no function. Pressthe ZERO ( ) key to exit test mode.

Table 8-5 on page 39 summarizes the test menu functions.

Figure 8-5. Test Menu

A/DTSTTEST PRTCFGDEFLT

38 9390 Installation Manual

Page 43: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

g from

Figure 8-6. Front Panel Key Functions in Test Mode

8.8 Software Revision HistoryThe following list summarizes the principal software changes made for Version 1.1:

• New SMPRAT parameter on the CONFIG menu provides selectable A/D conversion rate rangin3.75–30 Hz.

• A/D gain (ADGAIN) and A/D offset (ADOFFS) functions removed from TEST menu.

TEST Menu

Function Description

A/DTST Display A/D test

Press and hold Enter key to display raw count from A/D converter.

DEFLT Default parameters

Press and hold the setup switch, then press the Enter key to reset configuration and calibration parameters tofactory default values. This function is equivalent to using the RESETCONFIGURATION EDP command followedby the RS (reset) EDP command.

PRTCFG Print configuration

Press the Enter key to print the indicator configuration to the serial port. This function is equivalent to using theDUMPALL EDP command.

Table 8-5. Test Menu Functions

T

ZERO GROSSNET

TARE UNITS PRINT

TB/G Units

POWER

ENTER SAMPLEMODE PRINT

I/O

Move LEFT

Move RIGHTEXIT

Not used

ENTER

TEST MODE KEY FUNCTIONS

Appendix 39

Page 44: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

8.9 SpecificationsPowerPower 9 VDC, provided by 6 x “C” cell alkaline batteries

or AC adapter (9VDC, 300 mA, center-positive output connector)

Power Consumption 100 mA (using 4 x 350Ω load cells)

Analog SpecificationsFull Scale Input Signal Up to 17 mVExcitation Voltage 5 ± 0.25 VDC,

4 x 350Ω or 8 x 700Ω load cellsSense Amplifier Differential amplifier with

4- and 6-wire sensingAnalog Signal

Input Range –2.5 mV/V – 22.5 mV/VAnalog Signal

Sensitivity 0.3 µV/graduation minimum, 1.5 µV/grad recommended

Input Impedance 200 MΩ, typicalNoise (ref to input) 0.3 µV p-p with digital filters at 4-4-4

Internal Resolution 220 (approx. 1 000 000) countsDisplay Resolution 999 999 ddMeasurement Rate 15 measurements/sec, nominalInput Sensitivity 155 nV per internal countSystem Linearity Within 0.01% of full scaleZero Stability 155 nV/°C, maximumSpan Stability 4.0 ppm/°C, maximumCalibration Method Software, constants stored in

EEPROMCommon Mode

Voltage 2.5 ±.25 VDCCommon Mode

Rejection 140 dB minimum @ 50 or 60 HzNormal Mode

Rejection 90 dB minimum @ 50 or 60 HzInput Overload ± 12 V continuous, static discharge

protectedRFI Protection Signal, excitation, and sense lines

protected by capacitor bypass

Digital SpecificationsMicrocomputer Phillips P51XAG30 microcontroller @

12.288 MHzDigital Filters 3 filters, software selectable

Serial CommunicationsEDP/Printer Port Full duplex RS-232

19 200, 9600, 4800, 2400, 1200, 600, 300 bps; 8 data bits, no parity, or 7 data bits, even or odd parity

Operator InterfaceDisplay 6-digit LCD display: 7-segment, 1.0 in

(25 mm) digitsAdditional Symbols Annunciators for gross, net, standstill,

center of zero, tare, low battery. annunciators for units (lb/kg), piece

count mode (PC). Adhesive labels provided for units other than lb and kg.

Keyboard 6-key flat membrane panel

EnvironmentalOperating Temperature–10 to +40°C (legal);

–10 to +50°C (industrial)Storage Temperature –25 to +70°CHumidity 0–95% relative humidityAltitude 2000 m (6500 ft) maximum

EnclosureEnclosure Dimensions (including battery compartment cover)

9.0 in x 5.5 in x 4.1 in 23 cm x 14 cm x 10.4 cm

Weight 1.42 Kg (3.13 lb) Rating/Material NEMA 4X/IP66, stainless steel

Certifications and Approvals

9.00

5.526.42

11.04

3.24

4.09

1.62

•N

ATIO

NAL CONFERENC

E•

ON

WE

IGH T S A N D M E A

SU

RE

S

40 9390 Installation Manual

Page 45: Installation Manual · 1.3 Display Annunciators The 9390 display uses a set of LCD annunciators to provide additional information about the value being displayed. Figure 1-2 on page

riginal/OEMrials and

ationsfectiveair or

toin a

d or

r its

is

de

9390 Limited WarrantyInterface, Inc. warrants that all Interface equipment and systems properly installed by a Distributor or OEquipment Manufacturer (OEM) will operate per written specifications as confirmed by the Distributorand accepted by Interface, Inc. All systems and components are warranted against defects in mateworkmanship for two years.

Interface, Inc. warrants that the equipment sold hereunder will conform to the current written specificauthorized by Interface, Inc.. Interface, Inc. warrants the equipment against faulty workmanship and dematerials. If any equipment fails to conform to these warranties, Interface, Inc. will, at its option, repreplace such goods returned within the warranty period subject to the following conditions:

• Upon discovery by Buyer of such nonconformity, Interface will be given prompt written notice witha detailed explanation of the alleged deficiencies.

• Individual electronic components returned to Interface for warranty purposes must be packagedprevent electrostatic discharge (ESD) damage in shipment. Packaging requirements are listed publication, Protecting Your Components From Static Damage in Shipment, available from theInterface, Inc. Equipment Return Department.

• Examination of such equipment by Interface confirms that the nonconformity actually exists, anwas not caused by accident, misuse, neglect, alteration, improper installation, improper repairimproper testing; Interface, Inc. shall be the sole judge of all alleged non-conformities.

• Such equipment has not been modified, altered, or changed by any person other than Interface oduly authorized repair agents.

• Interface will have a reasonable time to repair or replace the defective equipment. Buyer responsible for shipping charges both ways.

• In no event will Interface, Inc. be responsible for travel time or on-location repairs, includingassembly or disassembly of equipment, nor will Interface be liable for the cost of any repairs maby others.

THESE WARRANTIES EXCLUDE ALL OTHER WARRANTIES , EXPRESSED OR IMPLIED , INCLUDINGWITHOUT LIMITATION WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULARPURPOSE . NEITHER INTERFACE , INC . NOR DISTRIBUTOR WILL , IN ANY EVENT , BE LIABLE FORINCIDENTAL OR CONSEQUENTIAL DAMAGES .

INTERFACE , INC . AND BUYER AGREE THAT INTERFACE , INC .’S SOLE AND EXCLUSIVE LIABILITYHEREUNDER IS LIMITED TO REPAIR OR REPLACEMENT OF SUCH GOODS . IN ACCEPTING THISWARRANTY , THE BUYER WAIVES ANY AND ALL OTHER CLAIMS TO WARRANTY .

SHOULD THE SELLER BE OTHER THAN INTERFACE , INC ., THE BUYER AGREES TO LOOK ONLY TO THESELLER FOR WARRANTY CLAIMS .