Whereas the Parliament of India has set out to provide a practical regime of right to information for citizens to secure access to information under the control of public authorities, in order to promote transparency and accountability in the working of every public authority, and whereas the attached publication of the Bureau of Indian Standards is of particular interest to the public, particularly disadvantaged communities and those engaged in the pursuit of education and knowledge, the attached public safety standard is made available to promote the timely dissemination of this information in an accurate manner to the public.
इंटरनेट मानक
“!ान $ एक न' भारत का +नम-ण”Satyanarayan Gangaram Pitroda
“Invent a New India Using Knowledge”
“प0रा1 को छोड न' 5 तरफ”Jawaharlal Nehru
“Step Out From the Old to the New”
“जान1 का अ+धकार, जी1 का अ+धकार”Mazdoor Kisan Shakti Sangathan
“The Right to Information, The Right to Live”
“!ान एक ऐसा खजाना > जो कभी च0राया नहB जा सकता है”Bhartṛhari—Nītiśatakam
“Knowledge is such a treasure which cannot be stolen”
“Invent a New India Using Knowledge”
है”ह”ह
IS 5182-12 (2004): Methods for measurement of airpollution, Part 12: Polynuclear aromatic hydrocarbons(PAHs)in air particulate matter [CHD 32: Environmental Protectionand Waste Management]
IS 5182 (Part 12) :2004
f777@T m-%
wwmaq~
m12a@awimdm&m-pJJ@m- (-)
(w%TT@%JT)Indian Standard
METHOD FOR MEASUREMENT OFAIR POLLUTION
PART 12 POLYNUCLEAR AROMATIC HYDROCARBONS (PAHs)IN AIR PARTICULATE MATTER
( First Revision)
ICS 13.040.20
0 BIS 2004
BUREAU OF IN DIAN STANDARDSMANAK BHAVAN, 9 BAHADUR SHAH ZAFAR MARC
NEW DELHI 110002
November 2004 Price Group 4
Environment Protection and Waste Management Sectional Committee, CHD 32
FOREWORD
This Indian Standard (Part 12) (First Revision) was adopted by the Bureau of Indian Standards, after the draftfinalized by the Environment Protection and Waste Management Sectional Committee had been approved bythe Chemical Division Council.
Pol ycyclic aromatic hydrocarbons (PAHs) an ubiquitous environmental pollutant that tepresent the largest classof suspected chemical carcinogens, PAHs are compound of two or more benzene rings with adjacent ringssharing two carbon atoms, non-aromatic rings may also be present. Physico-chemical characteristics of PAHsare highly governed by their molecular weight. Resistance to oxidation and reduction, varpour pressure andaqueous volubility, for instance, decrease with increasing molecular weight.
The PAHs can be formed from both natural and anthropognic sources. With the exception of anthracene, whichwas used in the dye industry, PAHs have no commercial use except for research purposes. Apart from smallamounts of geothermal and biosynthetic origin (synthesized by some algae, bacteria, fungi and plants), PAHsare mainly anthropogenic. Heating, power plants using fossil fuels, industrial process, incineration of industrialand domestic wastes, forest fires, volcanic activity, vehicle exhausts, asphalt pavements, and in general, allincomplete combination at high temperature and pyrolytic processes involving fossils fuels or more generally,materials containing carbon and hydrogen, are major sources of PAHs.
Human exposure to PAHs occurs principally by inhalation of tobacco smoke and polluted air, ingestion ofcontaminated and processed food and water, or by dermal contact with soots, tars and oils. The greatest exposureis likely to take place in the workplace of certain facilities like tar production plants, coking plants, asphaltproduction plants, coal-gasification sites, smoke-houses, municipal trash incinerators via air. In air, PAHs mayexist adbsorbed on carrier particles or in free state.
This standard was first published in 1974 based on ASTM D 2682: 1972 ‘Method of test for polynucleararomatic hydrocarbons in air particulate matter’. This revision has been necessitated in view of the latestdevelopments which have taken place in the sampling techniques and the need to have results for ambient airquality expressed as nanogram per cubic metre and for fugitive and source emissions results expressed asmicro-gram per cubic metre. Gas chromatographic (GC) method using flame ionization detector (FID) isprovided in this standard. Other analytical methods like HPLC and GC-MS maybe considered by the committeefor formulation later on.
Considerable assistance has been derived from ISO/DIS 12884:1997 ‘Air quality Determination of gas andparticle-phase polycyclic aromatic hydrocarbons in ambient air collection on sorbent-backed filters with gaschromatographic/mass spectrometric analysis’.
The composition of the Committee responsible for formulation of this standard is given at Annex B.
1n reporting the results of a test or analysis made in accordance with this standard, if the final value, observedor calculated, is to be rounded off, it shall be done in accordance with IS 2 : 1960 ‘Rules for rounding offnumerical values (revised)’.
IS 5182 (Part 12): 2004
Indian Standard
METHOD FOR MEASUREMENT OFAIR POLLUTION
PART 12 POLYNUCLEAR AROMATIC HYDROCARBONS (PAHs)IN AIR PARTICULATE MATTER
~ First Revision)
1 SCOPE
1.1 This standard (Pall 12) describes a method forsampling and analysis procedure for PAHs in ambientair and fugitive emission.
1.2 It does not measure vapour-phase PAH, for thatpolyurethane foam plugs may be used after the filterfor sampling.
2 PRINCIPLE
This method is designed to collect particulate phasePAHs in ambient air and fugitive emissions and todetermine individual PAH compounds. It is based onhigh volume (- 1.2 m3/min) sampling method capableof detecting sub ng/m3 concentration of PAH with atotal sample volume -480 m3/ of air over a period of8 h with same filter. It involves collection from airparticulate on a fine particle (glass-fibre) filter usinghigh volume sampler for total suspended particulatematter (TSPM) or respirable dust sampler forrespirable suspended particulate matter (RSPM orPM1O) and subsequent analysis by GasChromatograph (GC) using Flame Ionization Detector(FID). If sampling period is extended to 24 h withoutchanging the filter, it may enhance sample loss due tovolatility or reactions of PAHs on collection media.
3 INTERFERENCES
3.1 The panicle phase PAH maybe lost from particlefi Iter during sampling due to resorption andvolatilization especially during summer months atambient temperature of 30°C and above.
3.2 The method interference may be caused bycontaminations in low grade filter, solvent, andreagent, if used.
3.3 Glassware shall be properly cleaned(acid-washed) followed by solvent rinsing prior to use.
3.4 Matrix interferences may be caused bycontaminants, that is, hydrocarbons and other
organics that are co-extracted from sample. In thiscase clean-up by column chromatography shall berequired besides identification and confirmation ofindividual analyte followed by mass-spectrometer.
4 SAMPLE PRESERVATION
Sample should be wrapped in an aluminium-foil andshould be stored in a refrigerator at 4°C in dark placeto avoid photo-oxidation of PAHs for a period up totwo months. However, sample extracts may be stroredin dried form for a longer period.
5 APPARATUS
5.1 Ultrasonicator, with compact tank/bath of 4.5litre capacity and producing -40 kHz frequency forextraction.
5.2 Rotary Evaporator, buchi-type.
5.3 Silica-Gel Column, 200 mm length, 5 mminternal diameter with teflon stopcock.
5.4 GC-FID with Capillary Column
5.5 Syringes, 1 @to 10 ~1.
5.6 Flask and Beakers, 5-ml, lo-ml, 25-ml, 50-mland 250-ml capacity.
5.7 Variable Volume Micro-Pipettes, 0.5 ml and1.0 ml capacity.
6 REAGENTS
6.1 All solvents to be used should be of reagent grade.
6.1.1 Toluene, ultra-residue grade.
6.1.2 Cyclohexane, ultra-residue grade.
6.1.3 Tri-phenyl Benzene, ultra-residue grade.
6.1.4 Solid PAHs Compounds, high purity (seeAnnex A) to prepare the standard PAH solution.
6.1.5 Activated Silica Gel (60-100 meshes),chromatography grade.
1
IS 5182 (Part 12): 2004
7 PROCEDURE
7.1 Sampling
Collect sample through a high volume-sampler (HVS)using glass fibre (EPM — 2000) filter paper perferablyWhatmart or equivalent) at a flow rate of -1.2 m3/minover an extended period of time usually 8 h forambient air.
NOTE — A multi-point calibration of flow control system shatlbe conducted every 3 months (maximum 6 months) usingstandard calibration orifice/device.
7.2 Sample Processing and Extraction
Cut/punched at least 30 percent of total sample of theexposed filter paper or measured fraction of it intosmall strips/circtrlar pieces in a beaker/flask of 250-mlcapacity. Add tri-phenyl benzene, an internal standardat this stage for recovery test. Add about 100 ml oftoluene for extraction and keep beakers in ultrasonicbath for 30 min (or for 6 using Soxhlet extractionapparatus). Filter the extracts into evaporative flaskof 250 ml with the help of Whatrnan filter paperNo. 20 or filter-disc. Repeat the extraction twiceand combine extractants.
7.3 Sample Concentration
Evaporate the toluene extracts using rotary evaporatorwith water bath as cool as possible (temperature notexceeding 40”C). Do not evaporate up to total dryness.It should be stopped at near dryness (less than 1 ml,visible). Add 2.0 ml of toluene to rinse the wall ofevaporation flask and transfer extract into a beaker of5 ml capacity.
NOTE — Samples extraction should preferably be carried outwithin a month of sampling.
7.4 Clean-Up and Enrichment
It is performed using silica gel column having length200” mm, and inner diameter (ID) 0.5 cm. Pour aslurry of 3 g deactivated silica gel (60-100 mesh size)in cyclohexane into the column. Eltrte toluenefollowed by cyclohexane through the column forconditioning. Now introduce sample extract(concentrated, 2.0 or 3.0 ml) at the top of silicacolumn. Collect the PAH fraction with about 5 mlof cyclohexane. Collect all the eluants into a rotaryevaporator flask. Add another 30 ml of cyclohexane[o the column to elute all organics of interest. Collectall fractions into the flask and reduce to about 1 ml.Finally transfer into 5 ml capacity beaker/vials, dryand store in a dark and cool place.
7.5 Gas Chromatographic Analysis
Dilute the extracted residue and make up to 0.5 ml or1 ml. Inject 1 pl or 2 IA into GC-FID for analysis.
7.6 Gas Chromatography Conditions
7.6.1 Gas chromatography equipped with percentionization detector (FID), a split injector and capillarycolumn (Phase cross linked 5 percent phenyl,methyl-silicone) :25 m length, 0.2 mm inner diameter(ID), 0.33 ~m film thickness with following GCconditions:
Injection — Port — Temperature : 320°CFID — Temperature : 320”C
Oven — Temperature — Programme : Initialtemperature 140°C, hold for 3 min with :
Rate Final Temperature Hold forDeg/min “c min
Ramp A 6 250 6
Ramp B 10 300 5
Total run time :36 min
NOTE — Temperature programming may vary to suit betterresolution of peaks and subsequently also the total run timebased upon column features.
7.6.2 Carrier Gas
Nitrogen linear flow 30 ml/min (column flow 1 :2ml/min, makeup flow 28.8 ml/min), split ratio 1:20approximate y.
7.6.3 Gases for FID
Hydrogen flow 30 ml/min and airflow approximately300 ml/min
7.6.4 Type of Liner
Glass liner with some glass wool.
7.6.5 Integrator Setting
Chart speed :0.5 crnfmin; Attenuation : Zero; PeakWindow : Zero; Range : 8; Thresh : Zero; AreaRejection : 50; Baseline : 10 Ref RT window :-0.08; Non-Ref RT window :-0.08
7.6.6 Calibration of GC
7.6.6.1 Stock standard solution
Use solid PAH standards (10 mg/10 ml of toluene) toobtain the stock solution of PAH.
7.6.6.2 Working standard solution
From stock solution, make a working standardsolution of PAH by taking 625 1A of each individualPAH compound into a 25-ml volumetric flask andmake up with toluene (concentration should be25 ~g/ml). This PAH working solution is used forcalibration of GC.
7.6.6.3 Internal calibration
Each of the calibration standards shall contain theappropriate internal standards as specifiedconcentration. Anal yze (1 or 2 PI) of each standard
2
solution and plot the area ratio of analyte and the
corresponding internal standard against the
concentration for each compound and internalstandard.
7.6.6.4 The response factor (RF) foreachanalyteiscalculated using the following equation:
RF=(Asx C,s)/(Aisxcs)
where
A. = area of theanalyte (sample/standard) to bemeasured,
A,, = <area of lhe internal standard,
~, = concentration of the internal standard(ng/pi), and
C, = concentration of the analyte (sample/standard) to be measured (ng/f.d).
7.6.6.5 Add internal standards to all calibrationstandards and sample extracts analysed by GC. If theRF is constant over the working range (< 20 percentRSD), assume the RF to be invariant and use theaverage RF for calculations. Alternatively, plot theresults a calibration curve of response ratios, that isAJAi~ VS. RF.
7.6.6.6 Verify the working calibration curve or RFon each working day by the measurement of one ormore calibration standards. If the response for anyparameter varies from the predicted response by morethan i 20 percent, repeat the test using a freshcalibration standard. Alternatively, prepare a newcalibration curve. The relative retention times for eachcompound in each calibration run shall agree within0.03 relative retention time units.
7.6.6.7 When the ratio ( R ) of the retention time (R7)
of the unknown analyte, Tu to that of the correspondinginternal standard (where R = Rtu / Rtj ) is used toidentify the anal yte, the ratio of retention times (Rs)
from the sample chromatogram shall not be greaterthan 0.4 percent of the retention time ratio (Rc) fromthe chromatogram of the calibration standard. Thevalue of R shall not be more than 2 or less than 0.5.The retention index of the sample analyte andcorresponding standard shall agree within *2percent.
7.6.6.8 External calibration
When an analyte has been identified, thequantification of that analyte will be based on peakarea response of respective compounds with respect toworking calibration standard, that is, calibrationfactor (the ratio of response to the amount of massinjected). The retention times of various PAHscompounds obtained under the above GC conditionare given in Fig. 1.
IS 5182 (Part 12): 2004
7.6.6.9 Sample analysis
Dilute the extracted residue and make up to 0.5 ml or1 ml. Inject 1 pl or 2 @ into GC- FID for analysisand record the resulting peak size in area units / (orpeak height when overlapping peaks cause errors inarea integration). Identify the compounds usingexternal calibration procedures from the samplechromatograms.
NOTES1 Confirmation may be required on second GC column or byother technique, that is GC-MS if overlaps of peaks occur.2 Validation of GC system is performed through the use of midrange standards. If it falls outside its daily retention timewindow, the system is out of control. The cause of problem isdetected and error is rectified. A new calibration is to beperformed whenever a system is adjusted.
7.6.6.10 Field blank
Use atleast 10 percent of the samples or minimum ofone per batch of samples as field blank.
8 CALCULATION
8.1 Calculate the concentration in (rig/@) of eachidentified analyte in the sample extract (CJ asfollows:
cs = (AS Xcis)/(Al$X RF)
where
A, = area count of characteristic analytesample/peak being measured,
Ais = area count of characteristic internalstandardlpeak, and
C,, = concentration of internal standard.
8.2 Calculate the air volume from the periodic flowreading taken during sampling using the followingequation:
V = Average flow rate of sampling, m3/min x T
where
v = total sample volume at ambientconditions, in m3; and
T = elapsed sampling time, in min.
8.3 The volume of air sampled (VS) may optionallybe converted to standard conditions of temperatureand pressure (25°C and 101 kpa) using the followingequation:
V,= VX (P. / 101)X [298/(273+ T.)]
where
v=
Pa =
T, =
total sample volume under ambientconditions, in m3;
ambient pressure, in kPa; and
ambient temperature, in “C.
3
IS 5182 (Part 12) :2004
8.4 The concentration, in ng/m3, of each analyte in the anticipated concentrations of the targeted compoundsair sampled is given by: in the air to be pulled through the sampling medium.
C=csxvelvsAllow the spiked filter or sorbent to dry for about onehour in a clean, light-protected area prior to pulling air
where through the system.
V. = final volume of extract, in p]; and
V, = volume of air sampled (see 8.3).
9 DETECTION LIMIT
The minimum detectable concentration in term of BaPfor a sampling period of 8 h (with about 480 m3 of airpassed) will be 2 ng per cubic meter assuming 0.5 mlas the final volume of sample extract after clean-upand detectable concentration of 2 ng/pl of that sampleextract. High resolution mass-spectrometry or highpressure liquid chromatography can improvesensitivity down to lng/m3.
10 DETERMINATION OF SAMPLINGEFFICIENCY (SE)
10.1 Confirm the efficiency of the sampler for thetargeted PAH under the conditions anticipated in thefield prior to the initiation of any samplingprogram me. Determination of the efficiency isparticularly important if sampling periods exceeding24 are planned. Acceptable performance may beestablished by determining sampling efficiency.
10.2 Determine sampling efficiency by spiking asolution of the compounds of interest (or arepresentative selection that includes the most volatilePAH) onto a clean filter, then pulling through theassembled sampling module a volume of airequivalent to the maximum volume that will besampled.
10.3 Add the spiking solution for SE determinationdropwise to the filter, so as to uniformly load it andavoid oversaturation. Spiking levels shall correspondto atleast 3 times but not more than 10 times at
10.4 The sampling rate and sampling period shall bethe same as that planned for the programme. Ambienttemperatures during the test shall also approximatethose expected in the field, especially whenwarm-weather conditions are anticipated.
11 QUALITY ASSURANCE TEST
11.1 Recovery efficiencies (RE) of theisotonically-labelled PAH surrogates added to thesamples prior to extraction and analysis shall beclosely monitored to assure the effectiveness ofsample work-up and analytical procedures. Thesurrogate recoveries should fall between 75-125percent. Discard the samples for which surrogaterecoveries are less than 50 percent or more than 150pefcent.
11.2 Approximately 10 percent of the sample extractsshall be subjected to duplicate GC analysis to assureacceptable analytical precision.
12 EXPRESSION OF RESULTS
Express results for ambient air quality as nanogram percubic metre (ng/m3) of air whereas for fugitive andsource emissions express results in micro-gram percubic metre of air.
13 PRECISION
Precision and uncertainty under normal condition canbe expected to be t30 to *45 percent. Precision willvary with sample volume and analyte concentration.An overall mean standard deviation of about k15 tof20 percent is acceptable air samples over a period for
different PAH concentration range of up to 50 ng/m3.
4
IS 5182 (Part 12): 2004
1 TO LUENE SOLVENT PEAK 1.616
2dOlNAPHTHALENE
9.224
12.980 FLUORANTHENE
18.711 BENZO(a)ANTHRACENE
18.820 CHRYSENE
22.789 8ENZO(b) FLUORANTHENE
22.859‘BENZO(k) FLUORANTHENE
SENZO(Q]PYRENE 24049(a) PYRENE 2S.131
TRIPHENYL 6EN2ENE (INTL. STD.} 26-016
lNDENO(cd) PYRENE 31.581DIBENZO (ah) ANTHRACENES 31.820
[
OENZO(gh’1) PERYLENE 32.809
STOP
FIG. 1 CHROMATOGRAM OF STANDARD PAH MIXTURE
IS 5182 (Part 12): 2004
ANNEX A
slNo.
(1)i)
ii)iii)iv)v)
vi)vii)
viii)ix)x)
xi)xii)
xiii)xiv)xv)
xvi)
(Clause 6.1.4)PURITY OF POLYCYCLIC AROMATIC HYDROCARBONS
(SOLID COMPOUNDS FOR MAKING CALIBRATION STANDARDS)
Polycyclic AromaticHydrocarbons (PAHs)
(2)NaphthalenePhenanthreneAnthraceneFluoranthenePyreneChryseneBenzo (a) anthraceneBenzo (b) fluorantheneBenzo (k) fluorantheneBenzo (e) pyreneBenzo (a) pyrenePeryleneBenzo (ghi) peryleneDibenzo (ah) anthracenesIndeno (cd) pyreneCoronene
Molecular Formula Purity/Grade
(3)cloHt3
C14H1O
C14HI0
C16H1O
C16HI0
C1gH12
c18H12c20H,2
C20H12
C20H12
C20H12
C20H12
C22H12
c22H14
C22H12
c24H12
(4)99+ %99+ %99+ %98+ %98+ %95+ %99+ %99+ %98+ %99+ %98+ %98+ %98+ %98+ %99+ %98+ %
Molecular Weight
(5)128178178202202228228252252252252252276278276300
NOTE — Alternatively readymade Reference PAH Standard Mixture in liquid form available in ampmdes with traceable purity maybe obtained.
6
IS 5182 (Part 12): 2004
ANNEX B
(Foreword)
COMMITTEE COMPOSITION
Environment Protection and Waste Management Sectional Committee, CHD 32
Or.gunizarion
In personal capacity (40/14, C. R. Park, New Delhi 110019)
Bhabh Atomic Research Centre, Mumbai
Bhamt Heavy Electrical Limited, HardwarCement Manufacturers’ Association, New DelhiCcntml Fuel Research Institute, Dhanbad
C’enma] Lcathei Research Institute, ChennstiCentral Mining Rescwch Institute, DhanbadCentral Pollution Control Board, New Delhi
Confederation of Indian Industries (CI1), New Delhi
Crop Care Federation of India, Ncw DelhiDepartment of Civil Engineering, Indian Institute of Technology,
New Delhi
Dcpamnent of Science & Technology (TIFAC), New Delhi
Delhi College of Engineering, Delhi
l)mectorate General Factory Advice Service and Labour Institute,Mumbai
Directorate General of Health Services, New DelhiEngineers India Limited, New Delhi
Envirotech Instruments Private Limited, New Delhi
Food Research & Analysis Centre, New Delhi
Gujtrat Pollution Control Board, AhmedabadHindustan Lever Limited, Mumbai
Indian Chemical Manufacturers’ Association, Mumbai
Itndian (hsncil of Agricultural Research, New Delhilndian Council of Medical Research, New DelhiIndian Institute of Packaging, MumbaiIndian Oil Corpomtion Limited, Faridabad1PCL, Vadodra
industrial Toxicology Research Centre, LucknowMinistry of Defence (R & D), DRDO, New DelhiMinistry of Environment& Forests, New DelhiMinistry of Non-conventional Energy Sources, New DelhiTown & Country Planning Organization, Ministry of Urban
Development, New DelhiMunicipal Corporation of Greater Mumbai, Mumbai
National Institute of Occupational Health (ICMR), Ahmedabad
Notional Environmental Engineering Research Institute, Nagpur
National Productivity Council, New Delhi
N~!tional Thermal Power Corporation Limited, New Delhi
NCCBM, New Delhi
Representative(s)
PROF DILIPBISWAS (C/NU”nnatt)DR V. D. PURANtK
DR T. N. MAHADEVEN(Alfernafe)DR N, G. SHRIVASTAVADR K, C. NARANGDR GULAB SINGH
DR L. C. RAM (Al?emafe)DR S. RAJAMANIDR B, K. TEWARYDR S. D. MAKHtJ.4Nl
DR C. S. SHARMA (Alternate 1)DR S. K. TYAGI (Alternafe H)
StiRtA. K. GHOSE
SHRIR. P. SHARMA (Afremate)SHRIP. N. PARMESHAWARANDR MUKESHKHARE
DR ATUL Mt’tTAL (Alfemate I)DR ARVINDNEMA (Ahemate II)
DR P. SANJEEVARAODR G. SRINIVASAN(Alternate)
SHRIRAKESHMEHROTRASHRIV. K. MtNGCHA (Afternate 1)DR (SHRIMATI)A. MANDAL (Alternate 11)
SHRIS. S. GAUTAMSHRIM. R. RAJPUT (Alternate)
DR (SHRIMATI)MADHURISHARMASHRIB. B. LAL
SHRISUSHEELSADH (Alternate)SHRIRAKESHAGARWAL
DR RAIENDRAPRASAD (Alternate)DR S. K. SAXENA
DR R. PRABHAKARAN(Alternate)KUMARIP, S. SHAHSHRIB. B, DAVE
SHRIADITYAJHAVAR (Ahernde)SHRIV. N. DAS
SHRIA. A. PANJWANI(Ahernafe)DR R. C, MAHL!HWARI
SHRIH. N. SAIYAOSHRIS. C. ADAKREPRESENTATIVESHRIP. VIJAYRAGHAVAN
DR J. D. DESAI (Alternate)DR S. K. BHARGAVASHRIJ, C, KAFOORREPRESENTATIVESHRIVINODKUMARJAINSHRIK. K. JOADDER
DR SANTOSHVIDHYADHARAN (A[temate)DEPUTYCITY ENGINEERCIVIL (ENVT)
DEPUTYEXECUTIVEENGINEER(EhIISStONlNVENTORYGROLJP)(Alternate)
DR V. KRISHNAMURTHYDR A. K. MUKHERJEE(Alternate)
DR V. L PANDtT
DR TAPANNANDY (Allemate)SHRIR. C. MONGA
DR A. K. SAXENA (Alternate)SHRIR. GOPAL
SHRIM. S. BHAGWATDR S. N. PATI (Aftemate)
7
IS 5182 (Part 12): 2004
Organization Representative(s)
Reliance Indusrriea Limited,Mumbai REPiw3ENTAmvE
Shnram Institute of Industrial Research, New Delhi SHRIV. G. K. NAtRDR lAODI.SHKUMAR (Alternate)
SGS India Limited, Chennai SHRtS. RAW
Steel Authority of India Limited, New Delhi SHRIJ. KUMARDR MEENAKSHIKAKKAR (Alternate)
Thapar Centre for Industrial Research& Development, Patiala DR MAHESHWARROYDR S. K. CHAKRABORTY(Aher-nde)
The Fertilizer Association of India, New Delhi DR (SHRIMATI)B. SWAMINATHANDR S. NAND (Afternate)
BIS Directorate General DR U. C. SRtVASTAVA,Director & Head (CHD)[Representing Director General (Ex-oficio)]
Member-SecretarySHRIN. K. PAL
Director (Chemical), BIS
8
r --
Bureau of Indian Standards
B IS is a statutory institution established under the Bureau of Zndian Standards Act, 1986 to promoteharmonious development of the activities of standardization, marking and quality certification
and attending to connected matters in the country.
Copyright
B[S has the copyright of all its publications. No part of these publications may be reproduced in
of goods
any formwithout the prior permission in writing of BIS. This does not preclude the free use, in the course ofimplementing the standard, of necessary details, such as symbols and sizes, type or grade designations.
Enquiries relating to copyright be addressed to the Director (Publications), BIS.
Review of Indian Standards
Amendments are issued to standards as the need arises on the basis of comments. Standards are also reviewedperiodically; a standard along with amendments is reaffirmed when such review indicates that no changes areneeded; if the review indicates that changes are needed, it is taken up for revision. Users of Indian Standardsshould ascertain that they are in possession of the latest amendments or edition by referring to the latest issue of‘131S Catalogue’ and ‘Standards: Monthly Additions’.
This Indian Standard has been developed from Doc : No. CHD 32 (1 108).
Amendments Issued Since Publication
Amend No. Date of Issue Text Affected
BUREAU OF INDIAN STANDARDS
}Ieadquarters :
Manak Bhavan, 9 Bahadur Shah Zafar Marg, New Delhi 110002 Telegrams : ManaksansthaTelephones :23230131,23233375,2323 9402 (Common to all offices)
Regional Offices : Telephone
Central :
Eastern :
!Jorthem :
Southern :
Western :
Manak Bhavan, 9 Bahadur Shah Zafar Marg
{
23237617NEW DELHI 110002 232338.41
1/14 C. LT. Scheme VII M, V. I. P. Road, Kankurgachi
{
23378499,23378561KOLKATA 700054 23378626,23379120
SCO 335-336, Sector 34-A, CHANDIGARH 160022
{
603843609285
C.I.T. Campus, IV Cross Road, CHENNAI 600113
{
22541216,2254144222542519,22542315