14
Keller America Inc User’s Guide Rev. 07/19
Transcript
Page 1: Keller America Inc User s Guide Rev. 07/19 · eller America Inc ser’s uide Rev. eller America Inc ser’s uide Rev. IV. Product Overview - Level Transmitters Acculevel - The Acculevel

Keller America Inc User’s Guide Rev. 07/19

Page 2: Keller America Inc User s Guide Rev. 07/19 · eller America Inc ser’s uide Rev. eller America Inc ser’s uide Rev. IV. Product Overview - Level Transmitters Acculevel - The Acculevel

Keller America Inc User’s Guide Rev. 07/19 Table of ContentsI. CompanyInformation..................................................................................2II. ContactInformation……......……………………………………………………………………..2III. ProductApplicability.………....………………………………………………………………….2IV. ProductOverview…………………………………………………………………………………….3V. Lightning/SurgeProtection.………………………………………………………………………4VI. Options/Accessories.………………………………………………………………………………5VII. Installation……....………………………………………………………………………………………6VIII. Useandcare...…………….…………………………………………………………………………..8IX. TermsandConditions…....…………………………………………………………………………9X. Appendix a. Generic2-wireCurrentLoopConfiguration………………………………A-1 b. Generic3-wireVoltageOutputConfiguration…………………………..A-1 i. Econoline 1. 4-20mA,Cable..................………………………………….......A-2 2. 4-20mA,mPmConnector..................………………………A-2 3. 0.5-4.5VDC,Cable................…………………………………..A-2 4. 0.5-4.5VDC,mPmConnector................……………………A-2 ii. Valueline 1. 4-20mA,Cable................…………………………………………A-3 2. 4-20mA,DIN43650Connector................…………………A-3 3. 4-20mA,MIL-C-26482Connector.................……………A-4 4. 0-5/10VDC,Cable.................……………………………………A-3 5. 0-5/10VDC,DIN43650Connector..................……………A-3 6. 0-5/10VDC,MIL-C-26482Connector.................………..A-4 iii. Preciseline 1. 4-20mA+RS485,Cable................…………………………….A-5 2. 0-5/10VDC+RS485,Cable.................………………………..A-53.SDI-12+RS485,Cable.........................................…………A-5 4. 4-20mA+RS485,MIL-C-26482Connector...................A-6 5. 0-5/10VDC+RS485,MIL-C-26482Connector.....………..A-6 6. SDI-12+RS485,DIN43650Connector........................…A-5 iv. Levelgage 1. 4-20mA,Cable................…………………………………………A-7 2. 0-5/10VDC,Cable.................………………………………….…A-7 v. LevelRat 1. 4-20mA+RS485,Cable..............………………………………A-7 2. 0-5/10VDC+RS485,Cable..............…………………………A-7 vi. Acculevel 1. 4-20mA+RS485,Cable.………………..............………………A-7 2. 0-5/10VDC+RS485,Cable..............…………………………A-7 3. SDI-12+RS485,Cable..................................................A-8 vii. Microlevel/Nanolevel 1. 4-20mA+RS485,Cable...………………………………...........A-8 c. SubmersibleLevelTransmitters i. InstallationTips.………………………………………..……………A9-11

Page 3: Keller America Inc User s Guide Rev. 07/19 · eller America Inc ser’s uide Rev. eller America Inc ser’s uide Rev. IV. Product Overview - Level Transmitters Acculevel - The Acculevel

Keller America Inc User’s Guide Rev. 07/19 Keller America Inc User’s Guide Rev. 07/19

IV. Product Overview - Level Transmitters

Acculevel-TheAcculevelfromKellerisahighaccuracysubmersibleleveltransmitter.IncorporatingallofthefeaturesoftheLevelgageincludingtheNSF61andNSF372certifications,theAcculevelhasadditionalfeaturesthatprovideenhancedfunctionality,includingdualoutputs;i.e.,oneanalog(4…20mA,0…5VDC,0…10VDC)andonedigital(RS485).TheRS485interfaceprovideslevelandtemperatureinformationandallowstheusertorescaletheanalogoutputfrom10…110%ofthebasicrange.Additionally,theAcculeveliscapableofgreateraccuracythantheLevelgage,withastandardaccuracyof0.25%BRTEBandanoptional0.1%BRTEB.TheAcculevelcanbemanufacturedin316Lstainlesssteelortitaniumforincreasedresistancetocorrosion,notablyfromseawater.

Acculevel SDI -TheAcculevelSDIisaversionoftheAcculevelhighaccuracysubmersibleleveltransmitterthatisavailablewithSDI-12orRS485digitalcommunicationprotocols.Itisavailablewithstandardaccuracyof±0.1%FSTEBinrangesupto900ftWCandoptionalUSGSOSWaccuracyforrangesupto70ftWC.TheAcculevelSDIcanbemanu-facturedinstandard316Lstainlesssteeloroptionaltitaniumforincreasedresistancetocorrosion,notablyfromseawater,carriestheNSF61andNSF372approvals.

Levelgage-TheKellerLevelgageisageneralpurposesubmersibleleveltransmitteravailablein1%and0.5%BRTEBaccuracy.Constructedfrom316Lstainlesssteel,itisideallysuitedformostfreshwatermonitoringapplications.Withstandardconstruction(i.e.EPDMandXLPE)materials,theLevelgagehasbeencertifiedtoNSF61andNSF372standardsforuseindrinkingwater.Itisavailableincustomrangesfrom0…3ftWCto0…900ftWC.Cablelengthsarecuttouserspecifications.TheLevelgageisalsoavailableinvariousoutputconfigurations,including4…20mA,0…5VDC,and0…10VDC.

LevelRat -TheKellerLevelRatisspecificallydesignedtoprovide0.5%BRTEBaccuracyforlevelindicationingrease-ladenenvironmentscommonlyfoundinliftstationsservingcommercialconcerns.TheLevelRatdesigntakesastepawayfromthelargeandbulkyprotectioncagedesignsbyutilizingauniquediaphragmmaterialthatoffersthenon-stickpropertiesofTeflonandprovidesaddedresistancetodamagefromdebris.TheLevelRatprovidesdualoutputs;yourchoiceofanalogoutputs(4…20mA,0…5VDC,or0…10VDC)plusonedigital(RS485)forlevelrangesfrom0…5ftWCto0…100ftWC.

Microlevel-TheMicrolevelcombinesthecapabilityoftheAcculevelwithaclassleading0.63”(16mm)outsidediameter.TheMicrolevelisavailableinfullscalerangesfrom0…3ftWCthrough0…900ftWCandprovides0.25%BRTEBaccuracy.LiketheAcculevel,italsoprovidesboth4…20mAanalogandRS485Digitaloutput.TheRS485interfaceprovideslevelandtemperatureinformation.Italsoallowstheusertorescaletheanalogoutputfrom10…110%ofthebasicrange.

Nanolevel-TheNanolevelisuniquelydesignedforhighaccuracymeasurementofverylowpressureranges.TheNanolevelisavailableinfullscalerangesfrom0…4inWCthrough0…10ftWCandisavailablewithstandardaccuracy0.25%BRTEBand0.1%BRTEBoptional.LiketheAcculevel,italsoprovidesboth4…20mAanalogandRS485Digitaloutput.TheRS485interfaceprovideslevelandtemperatureinformation.Italsoallowstheusertorescaletheanalogoutputfrom10…110%ofthebasicrange.

Page3

I. Company Information

KellerAmericaIncistheNorthAmericansubsidiaryofKellerAGfürDruckmesstechnik,whichcontinuesastheleaderinlevelandpressuresensing.Fromhumblebeginnings,Kellernowprovidesitsproductsandexpertisethroughaworldwidenetworkoffacto-ry-trainedprofessionals.

Withtheadventofinexpensive,miniaturizedmicroprocessors,Kellercontinuestoad-vancethestate-of-the-artinfullypackagedpressuretransmitters,withTotalErrorBandperformancenotpossiblejustafewshortyearsago.Annualsensorproductionnowexceeds1millionpieces.

DoingbusinesswithKelleriseasyandmostinquiriesareansweredwithin24hours.StandarddeliveryonU.S.-assembledproductsis3daysARO.Weofferconvenienttermswithapprovedcreditandacceptmostmajorcreditcards.

AtKeller,providingcustomerswithsuperiorvalueisawayoflife.

II. Contact Information

KellerAmerica,Inc.351BellKingRdNewportNews,Virginia,23606P:877-253-5537F:757-596-6659Web: www.kelleramerica.comEmail: [email protected]

III. Product Applicability:Unlessotherwisenoted,thismanualcontainsuserinformationforthefollowinglevelandpressureinstrumentationprovidedbyKellerAmerica:Acculevel™,Levelgage™,LevelRat™,Microlevel™,Econolevel™,andNanolevel™submersibleleveltransmitters;Econoline™,Valueline™,andPreciseline™abovegroundpressuretransmitters.

Electronic copies of this user’s guide can be found at: http://www.kelleramerica.com/manuals-and-software/manuals/Keller_America_Users_Guide.pdf

Page2

Page 4: Keller America Inc User s Guide Rev. 07/19 · eller America Inc ser’s uide Rev. eller America Inc ser’s uide Rev. IV. Product Overview - Level Transmitters Acculevel - The Acculevel

Keller America Inc User’s Guide Rev. 07/19 Keller America Inc User’s Guide Rev. 07/19

VI. Options /Accessories:

Titanium Construction -OptionalfortheAcculevelandPreciseline.Highlyrecommend-edforbrackishandseawaterapplications.

External Surge Protector-Recommendedforlightning/surgeprotectionofuser’spowersupply/readout/PLC,etc.TobeusedinconjunctionwithLevelgage,Acculevel,LevelRat,Econolevel,Microlevel,Nanolevel,Valueline,orPreciselinetransmitters.

½” NPT Pipe conduit fitting -½”NPTmalefitting,allowsrigidmountingto½”conduitforLevelgage,Acculevel,orLevelRatsubmersibletransmitters.ItcanalsobeaddedtoourPreciselineandValuelinepressuretransmitters.

Drying Tube Assembly-Cleartubefilledwithindicatingdesiccant,attachesdirectlytocableventtube,interceptswatervapor.Highlyrecommendedwhenoperatingisinhighhumidityconditions.Mustbeperiodicallyrenewedasdesiccantbecomessaturated,turningcolorfromblue(dry)topink(saturated).

Bellows Assembly-Alternativetothedryingtube,thisaneroidbellowsattachestocableventtubeandrequiresnoperiodicmaintenance.Notethattheuseofthebellowsresultsinaclosedreferencepressuresystemsubjecttozeroshifterrorsinducedbychangingtemperaturesofupto0.0004psi/ºC.Notrecommendedforusewithaccuracyrequirementsof±0.25%FSorbetter,orranges<2psi.

Electrical Connectors -mPm393isoptionalfortheEconoline,DIN43650andMIL-C26482areoptionalforValuelineandPreciseline.ThemPm393andMIL-C26482matingconnectorsfeaturesoldercupswhiletheDIN43650incorporatesscrewterminals.

RS485 Converter/Cable -FortheuserwhowishestocommunicatewiththeirAcculevel,LevelRat,Nanolevel,Microlevel,orPreciselinetransmitterviatheRS485digitalinterface.AvailableinRS232andUSBversions,alongwithcableadapters,asneeded,andCCS30Software(alsoavailablefordownload,freeofchargefromwww.kelleramerica.com).Consultfactoryfordetails.

Termination Enclosure-Convenientoptioncomplementinggauge-typepressure/leveltransmitters,whereitisdesiredtoterminatethetransmittercableclosetothemea-surementpoint.ItincludesaNEMA4Xclearfrontenclosure(7.9X4.7X3.5inches)withtwo,liquid-tightcablefittings(onein,oneout),aterminalstrip,andampleroomformountingadryingtubeorbellowsassembly.

Stabilizing Weight-ZincpropshaftanodeadaptedtofitØ21mmO.D.ofLevelgage,Acculevel,andLevelRatsubmersibleleveltransmitters.Aidsincorrosionresistanceaswellashelpsensurethatthecableremainstautinturbulentconditions.

Cable Hanger-Singleeyefor0.230”diametercable.Designedtogripthecablejacketandprovideasuspensionpointforinstallationofcabledtransmitters.

Pressure Test Adapters -Designedtofacilitatecalibrationandmaintenancetesting,thesespecialtyadaptersallowforconnectionofthevariousKellerleveltransmitterstoapressurecalibrator.

Page5

Econolevel-TheEconolevelbyKellerisalowcostgeneralpurposesubmersibleleveltransmitterdesignedforhighvolumeapplications.Itisavailableinnominalpressurerangesfortankandwelllevelmonitoringapplications.TheEconolevelprovidesa4…20mAanalogoutputthatiscompatiblewithmostSCADAsystemsandPLCs.Mini-mumorderquantitiesapply.

IV. Product Overview cont’d - Pressure Transmitters

Econoline-TheEconolinebyKellerisalowcostgeneralpurposepressuretransmitter.Itisavailableinnominalpressurerangesbetween5psiand10,000psiandwithathreaded¼”NPTpressureconnection.Idealfortankandpipepressuremonitoringapplication,itprovideseither4…20mAor0.5…4.5VDC(ratiometricornon-ratiometric)analogoutputsthatarecompatiblewithmostSCADAsystemsandPLCs.

Valueline-TheValuelinebyKellerisahighlyversatileandaccuratepressuretransmitter.Itisavailableincustompressurerangesfrom0…2through0…15,000PSIandyourchoiceofelectricalconnectionsincludingattachedcable,DIN43650,andMIL-C-26482connec-tors.Theattachedcableversionisfullysubmersible,makingittheidealchoicewheninstalledinavalvepitorotherenvironmentpronetoperiodicflooding.TheValuelinealsocarriestheNSF61andNSF372approvalsfordrinkingwaterapplications.

Preciseline -ThePreciselineisourmostaccurateabove-groundpressuretransmitter.Withstandardaccuracyof0.25%BRTEBandoptional0.1%BRTEB,a¼”NPTpressureconnection,anddualoutputs(analoganddigital)thePreciselineoffersversatilityindesignandapplication.LiketheAcculevel,theanalogoutputcanberescaledviatheRS485interface.TheRS485interfacealsoallowstemperaturedatatoberetrievediftheapplicationdemands.LiketheValueline,thePreciselineisincludedintheNSF61and372approvals.

Preciseline SDI-ThePreciselineSDIishighlyprecisepressuretransmitterthatisavail-ablewithSDI-12orRS485digitalcommunicationprotocols.Itisavailablewithstandardaccuracyof±0.1%FSTEBinrangesupto900ftWCandoptionalUSGSOSWaccuracyforrangesupto70ftWC.ItcomeswithyourchoiceofattachedcableorMIL-C-26482connectorelectricalconnections.Thecableversionisfullysubmersible,makingittheidealchoiceinenvironmentspronetoperiodicflooding,whiletheNSF61andNSF372approvalsmakethePreciselineSDIsuitableforuseindrinkingwaterapplications.

V. Lightning / Surge Protection:

Asofthiswriting,Kelleristheonlypressure/leveltransmittermanufacturerintheworldtoprovide,atnoadditionalcharge,alifetimewarrantyagainstdamagecausedbylight-ningand/orelectricalsurge.Thisguaranteeappliestothe4…20mAoutputversionofthefollowingproducts:Levelgage,Acculevel,LevelRat,Microlevel,Valueline,andPreciseline.Thiswarrantyapplieseveninthecaseofadirectlightningstrike.Theuser’sliabilityislimitedtoshippingcoststoKellerAmerica.

Page4

Page 5: Keller America Inc User s Guide Rev. 07/19 · eller America Inc ser’s uide Rev. eller America Inc ser’s uide Rev. IV. Product Overview - Level Transmitters Acculevel - The Acculevel

Keller America Inc User’s Guide Rev. 07/19 Keller America Inc User’s Guide Rev. 07/19

Thecalculationsbelowareusefulinidentifyingtheminimumsupplyvoltageneededforacurrentloopwithadditionallineimpedancesorconversely,themaximumcurrentloopimpedanceallowedforagivensupplyvoltage.

Minimumsupplyvoltagewithlightningprotectionoptioninstalled: 11VDC+(Totalloopimpedancex0.022)

Minimumsupplyvoltagewithoutlightningprotectionoptioninstalled: 8VDC+(TotalloopImpedancex0.022)

Maximumallowableloopimpedancewithlightningprotectionoption installed: (SupplyVDC–11VDC)/0.022

Maximumallowableloopimpedancewithoutlightningprotectionoption installed: (SupplyVDC–8VDC)/0.022

Transmitter Anchoring-ItisrecommendedthatKellersubmersibletransmittersbeinstalledinastillingwellorattachedtorigidconduitviaaconduitfittingintegraltothetransmitter,inordertopreventdamagetothetransmitterfromimpactwithimmovableobjects.Itisnotadvisabletotiethetransmittertoapumportopiping,asanyproblemwiththetransmittercouldrequirethatthepumpbepulledfromtheinstallation.

Someapplicationsrequirethetransmittertobesuspendedwithoutaprotectivestillingwellorconduitattachment.Inallinstallations,careshouldbetakentopreventdamagetothesubmersiblecable.

Transmitter Submersion -Damagetosubmersiblecablecanleadtofailureofthetrans-mitter.Kelleremploysaruggedcablejacketmaterialstominimizetheriskofcutsandabrasion.Still,takecarewhenloweringyourtransmitterintothewell,makingsurethecabledoesnotdragoversharpedges.Avoiddroppingthetransmitterfromthesurface.Donotsubmergeelectricalconnectionendofcable.

Condensation protection -Kellerhasoptimizedthesizeofthecableventtominimizetheoccurrenceofwatervaporincursion.Inareasofhighhumidity,itmaybedesirabletouseaDryingTubeAssembly(desiccant)orBellowsAssemblytopreventwatervaporfromenteringtheventtube.ContactKellerAmericafororderinginformation.

Bending of Cable-Ourjacketedcableisquiteflexible.However,caremustbetakentoensuretheventtubeintegraltothecableisnotcrimpedwhenbendingthecabletosuityourinstallation.Itisrecommendedthatthecablenotbebenttoaradiussmallerthan1inch.

Cable Compression -Manyusersemployacompressionfittingtosecureourcableasitentersajunctionbox.Caremustbetakenthatthefittingisnotovertightened,causingdamage to the cable and/or crimping the vent tube.

Page7

Digital Display / Meter -TheTRIDENTdisplaybyPrecisionDigitalisaversatileandcosteffectiveprocessmeterprovidinganeasytouseandreaddisplay/controllerforyourlevelandpressuremeasurementapplication.Formoreinformationandtechnicalsup-port,pleasecontactPrecisionDigitalat(800)343-1001oronlineatwww.predig.com.

VII. Installation:

Thefollowingisimportantinstallationandgeneralmaintenanceinformationforsub-mersibletransmitters.PleasecontactKellerAmericaforadditionalinstruction.

Supply Voltage-Adequatesupplyvoltageiscriticaltoensureproperoperationof4-20mApressureandleveltransmitters.Withouttheminimumrequiredvoltageavailableatthetransmitter,thetransmitterwillnotoutputthecorrectanalogvalue.

Manyanalogtransmitterswillappeartooperateproperlyevenwhenthesupplyvoltageisnotadequatetopowertheloopwhenthetransmittershouldbeoutputting20mA.Forexample,a10voltsupplymayappeartobeenoughtopowerananalogtransmitterwhenitisoutputting4mAwithzeropressureapplied,butasthetransmitter’soutputincreaseswithincreasingpressure,voltagedropsacrossotherdevicesintheloop(ana-loginputdevices,cableand/orexternalbarrierdevices)mayreducethesupplyvoltagetothetransmitterandpreventitfromprovidingthecorrectoutputaboveacertaininputpressure/levelthreshold.

KellerAmerica’s4-20mApressureandleveltransmittersfeaturemicroprocessorbasedsignalconditioning.Duringpowerup,thecircuitperformsachecksequencewhichde-termineswhetherthereissufficientsupplyvoltagetopoweralldevicesontheloop,bysettingtheoutputto~110%ofthemaximumvalue,i.e.,~22.5mA.Ifthesupplyvoltageisnotsufficienttosupply22.5mAtothecircuit,thenthemaximumpossiblecurrentwillbeseenontheanalogoutput,e.g.,17mA,andthetransmitterwillnotinitiatenormaloperation.

Thebenefitofthistechnologyisthattotalloopimpedanceisaccountedforpriortoplacingtheequipmentintoservice,preventingfalseindicationswhenthevoltagesupplyisinsufficienttosupporttheloopwithmaximumpressureapplied.

Mostcurrentloopscontainanaloginputdevices,indicatorsorothercomponentshavinginputimpedanceswhichmustbeconsideredwhencalculatingthesupplyvoltagesneed-ed.Forexampleananaloginputdevicewitha250OhmInputImpedancewillrequireanadditionalvoltageof5.5VDC(250x0.022=5.5)inadditiontotheminimumsupplyvoltagenecessaryforthetransmittertooperateproperlyovertheentirerange.

Page6

Total LoopImpedance

IndicatorAnalog Input

Device

Cable

ResistanceIS Barrier

4 - 20mA outputLevel / Pressure

Transmitter

Page 6: Keller America Inc User s Guide Rev. 07/19 · eller America Inc ser’s uide Rev. eller America Inc ser’s uide Rev. IV. Product Overview - Level Transmitters Acculevel - The Acculevel

Keller America Inc User’s Guide Rev. 07/19 Keller America Inc User’s Guide Rev. 07/19

theinput/nosecaptothedeviceisclear.Finishbystirringinfreshwater.Wipedrywithasoftragortowel.

Asoapscumandhard-waterstainremovermayalsobeused,ifnecessary,butonlyaftercompatibilitywithanyo-ringsealsintheKELLERproductisdetermined.Followthesolventmanufacturer’srecommendationsforsafehandling.

IX. Terms and Conditions:

Offers -Unlessotherwiseindicated,thepricessetforthinthisquotationarevalidfor60daysfromthedateofthequotationandapplytoproductsthatarescheduledforship-mentwithintwelvemonthsfromthedateoftheBuyer’spurchaseoratKellerAmerica,Inc.shippingcapabilitiesatthetimetheorderisentered,whicheverislater.

Acceptance of order-Anypurchaseordertobeacceptedhastorefertoaquotation,whetherdeliveredbysurfacemail,e-mailorfacsimile.Alloralordersmustbeconfirmedinwriting.Byreferringtothisquotation,BuyeracceptsandadoptstheGeneralTermsandConditionsofSale(“TermsandConditions“)totheexclusionofanyadditionalordifferenttermsappearinginBuyer’spurchaseorderandwaivesanyrightBuyermayhavetoenforceanysuchadditionalordifferentterms.OurConfirmationofOrderorInvoicevalidatesorders.Ordersforcustom-configuredproductsarebuilttoyourspecifi-cationandthereforemaynotbecanceledoncetheorderhasbeenconfirmed.

Date of shipment-TheestimatedshippingcapabilitystatedonthequotationisgivensolelyfortheBuyer’sinformationanddoesnotconstituteacommitmenttodeliverproductsinaccordanceherewith.Buyermayrequestaspecificshippingdateorshippingschedule.KellerAmerica,Inc.willscheduleshipmentsbaseduponBuyer’srequestandKellerAmerica,Inc.’sshippingcapabilitiesatthetimeBuyer’spurchaseorderispro-cessed,atwhichtimeKellerAmerica,Inc.willissuetoBuyeraformalAcknowledgmentofOrderthatwillindicatetheestimatedshippingdate(s).Aftertheshipmentshavebeenscheduled,BuyermaynotcancelorpostponeascheduledshipmentunlessBuyersub-mitsitsrequestinwritingandKellerAmerica,Inc.consentsbyissuinganewacknowl-edgment.

Anyrequesttocancelortorescheduletheshipmentthatisreceivedlessthan21daysfromthedatescheduledfortheshipmentoftheproductscoveredbytherequestmayberejectedasuntimely,ormaybeaccepteduponpaymentofthecancellationchargeof20%orareschedulingchargeof10%ofthesalespriceshownonthisquotationatKellerAmerica,Inc.’soption.KellerAmerica,Inc.willuseitsbesteffortstoshiponorevenbe-foretheestimatedshippingdatesindicated,butwillnotbeliableforanydelayorfailuretodeliver.KellerAmerica,Inc.shallnotbeliableforanyspecial,incidentalorconsequen-tialdamagesresultingfordeliverydelaysorinabilitytodeliver. AllgoodsormaterialssuppliedbyKellerAmerica,Inc.remainitspropertyuntiltotalpay-mentisreceived.AllsuchgoodsandmaterialsareatthesoleriskoftheBuyerandintheeventofbeingdamaged,destroyedorlostafterdelivery.KellerAmerica,Inc.isentitledtoreceivethetotalpaymentofthosegoods.

Page9

Position Sensitivity-Thetransmittershouldbeinstalledinaverticalposition,other-wiseitmayexhibitanoffset.Ifthetransmittermustbeinstalledinanypositionotherthanvertical,measuretheoutputwithnopressureappliedpriortoconnectiontoyourdisplay,PLC,orcontroller.Usethemeasuredvalueforyourzeropoint.

VIII. Use and Care:

Safe Handling-SafehandlingofKELLERpressuremeasurementdevicesisaccomplishedifanominalamountofcareistaken.Thingstoavoidare:•Sharpimpactagainsthardsurfaces•Contactwithchemicalsknowntobecorrosivetothematerialsofconstruction•ProbingofpressuresensingmembranewithANYTHING

Limits of Pressure -KELLERpressuresensors,transducersandtransmittersaredesignedtowithstandacertainamountofoverpressurewithoutdamageorcalibrationshift.Itcanrangefrom15Xforthelowerpressurerangesto1.1Xforthehighestranges.Thisvalueisdifferentforeachproductandisreferredtointheproductliteratureas“Overpressure”.Itistheuser’sresponsibilitytoensurethattheproperKELLERproductischo-senfortheparticularpressureconditionsexpected.

Environmental conditions-EachKELLERproductisdesignedtobecompatiblewithaparticularenvironment.Itistheuser’sresponsibilitytoensurethattheKELLERproductisnotexposedtoanenvironmentalconditionforwhichitisnotdesigned.Thesecondi-tionscanincludeoperatingtemperaturerangeandexposuretohigh-pressurewaterjets,medianotcompatiblewiththematerialsofconstruction,submergenceoftransmittersnotdesignedforthatpurpose,orpotentiallyexplosiveatmospheres.AKELLERengi-neercanhelptheuserdeterminethecorrectchoiceofenclosuretosuittheparticularapplication.

Electrical conditions -EachKELLERproductisdesignedtooperateproperlywithinaspecificrangeofelectricalconditions.Thespecificproductlabeldefinestherating(s),ifany,thatappliestotheproducttowhichitisaffixed.Alltransmittersaredesignedtowithstandreversepolarityaswellasovervoltagetoacertainextent.Itistheuser’sresponsibilitytoensurethatallelectricalconnectionsaremadetotheKELLERproductsinaccordancewithKELLERrecommendationsaswellaslocalelectricalcode.Wirecolorsorconnectorpin-outsareeitherprintedonthelabelaffixedtotheproductorprovidedseparately.

Cleaning-WARNING!Undernocircumstancesshouldthemembrane,pressureinputport,ornosecaptotheKELLERdevicebeprobedwithanyobject.Damagetothesens-ingmembraneispermanentand,inmostcases,requiresrepairorreplacement.

Regardingmedia-isolatedproducts,shouldthepressureinputornosecaptothesensor,transducer,ortransmitterbecomefouled,itmaybecleanedinthefollowingmanner:

Thedeviceshouldbeslowlyloweredmembrane-endintoasolutionofwarm,soapywater.Careshouldbetakennottosubmergetheentiredevice,unlessitisspecificallydesignedforcontinuoussubmergence.Agitateinthesolutionfor20-30secondsoruntil

Page8

Page 7: Keller America Inc User s Guide Rev. 07/19 · eller America Inc ser’s uide Rev. eller America Inc ser’s uide Rev. IV. Product Overview - Level Transmitters Acculevel - The Acculevel

Keller America Inc User’s Guide Rev. 07/19 Keller America Inc User’s Guide Rev. 07/19

uponpaymentofalltravelexpensesbyBuyer.

WarrantyperiodfortheLevelgage,LevelRat,Acculevel,Microlevel,Preciseline,andVal-uelineproductsis24(twenty-four)monthsfromdateofshipment.Warrantyperiodforallotherproductshereinis12(twelve)monthsfromdateofshipment.Warrantyrepaircoversallapplicablepartsandlabor.Thiswarrantyisgiveninlieuofanyotherwarranty,expressorimplied.KellerAmerica,Inc.explicitlydisclaimsanyimpliedwarrantiesofmerchantabilityandfitnessforaparticularpurpose.Therearenowarranties,expressorimplied,thatextendbeyondthedescriptionherein.ThesoleandexclusiveremedyforanyclaimsagainstKellerAmerica,Inc.shallbethewarrantydescribedinthisdocument.Alldamages,directorconsequential,limitedtothedescribedwarrantyareexcluded.

Repairs-UnlessotherwiserequestedinwritingbytheBuyer,allproductsreturnedtoKellerAmerica,Inc.underthetermsofthewarrantywillbecheckedandanalyzedinor-dertodeterminethecauseofthedefault(s)claimedbytheBuyer.AreportwillthenbesubmittedtotheBuyerpointingoutthenatureofthedefault(s),thepartyresponsibleforthedefault(s)andthequotationoftherepair,ifneeded.Forfurtherrepairinstruc-tionsgotowww.kelleramerica.com.

Complaints-AllclaimsordisputesmustbemadeinwritingtoKellerAmerica,Inc.amaximumof60daysfromreceiptofthegoods,includingdiscoveryoffaultsnotpre-viouslyapparent.Ifthewarrantyclaimsarejustified,KellerAmerica,Inc.isfreeatitsdiscretiontorepair,replaceorissuecredit.Nofurthercompensationfordamageswillbemade.

AnydisputesorclaimsofBuyermustbeinitiatedinapropercourtorotheradjudicativebody,asapplicable,withinone(1)yearfromthedateofshipmentbyKellerAmerica,Inc.,oritsrepresentative(s),orsuchclaimshallbedeemedinvalidorexpiredandcannotberenewed.Totheextentallowed,thislimitationperiodshalltrumpanyapplicablestatutorylimitationsperiodthatmaystatealongerperiod.

Limitation of responsibility -Ourresponsibilityconcernsthecorrectfunctionofourproductsonly.Itcannotbeextendedtothewholesysteminwhichtheyareused.Ourresponsibilityislimitedtothereplacement,repairorreimbursementofthegoodsweagreearedefectiveornon-conforming.Theclaimmustbeinwritingwithin60daysfromreceiptofthegoods.

Place of jurisdiction / applicable law -ThecontractmadebyacceptanceofthisoffershallbedeemedmadeintheStateofVirginiaandshallbegovernedbyandconstruedinaccordancewiththelawsofthatstatewithoutreferencetoorapplicationofanyconflictsoflawsprinciplesandwithoutconsiderationoftheplaceofexecution.BuyerexpresslyagreestothesubjectmatterandpersonaljurisdictionoftheCircuitCourtsfortheCityofNewportNews,Virginia,orthefederalDistrictCourtfortheEasternDistrictofVirginia,NewportNewsDivision.

Assignment-NeitherthisoffernoranycontractresultingtherefrommaybeassignedortransferredinwholeorpartwithoutthepriorwrittenconsentofKellerAmerica,Inc.

Page11

Price -Unlessotherwisespecifiedinthequotation,ourpricesareinUSdollars(USD).Theyareexclusiveoffreightcostsandofallstateandlocalsales,use,excise,privilegeandsimilartaxes.SuchtaxesimposedonKellerAmerica,Inc.orwhichKellerAmerica,Inc.hasadutytocollectinconnectionwiththesaleordeliveryoftheproductsde-scribedonthequotationshallbepaidbyBuyerandwillappearasseparateitemsontheinvoice.

Shipment / freight -DeliveriesareF.O.B.NewportNews,VA.TitleandriskoflossshallpasstoBuyerupontenderoftheproductsbyKellerAmerica,Inc.toacommoncarrier.InabsenceofspecificwritteninstructionfromBuyer,KellerAmerica,Inc.willselectthecommoncarrier,butKellerAmerica,Inc.shallnottherebyincuranyliabilityinconnec-tionwithshipment.Buyershallberesponsibleforanyfreightcharge.Declaredvalueforeachshipmentwillbeamaximumof$100regardlessoftheactualvalueofthegoods,notwithstandingwritteninstructionsfromtheBuyerspecifyingahigherdeclaredvalue.Iftheproductsareshippedfreightprepaid,BuyershallpayKellerAmerica,Inc.theap-propriatefreightcharges,whichwillbeshownasseparateitemsontheinvoice.

Payment -TermsareindicatedonourConfirmationofOrderandonourInvoices.KellerAmerica,Inc.willsubmitaninvoicetoBuyerforeachshipmentatthetimeofshipment.Exceptasotherwiseprovidedonthequotation,BuyershallpaytheamountinvoicedbyKellerAmerica,Inc.within30daysfromthedateoftheinvoice.IfinthejudgmentofKellerAmerica,Inc.thefinancialconditionorpaymentrecordofBuyeratanytimedoesnotjustifyshipmentunderthepaymenttermsspecifiedabove,KellerAmerica,Inc.mayrefusetoshipunlessitreceivespaymentinadvance,oratitsoption,paymentupondelivery.2%interestpermonthischargedonoverdueaccounts.

Documentation -Theinformationgiveninourdocumentation,printedmatter,datasheetsandpricelistsiswithoutcommitment.Thisinformationspecifiestheproductbutisnowarranty,unlessagreedinwriting.

Warranties -KellerAmerica,Inc.warrantsthattheproductsthatitsellsaredeliveredfreefromdefectsinmaterialandworkmanship.KellerAmerica,Inc.’sliabilityunderthiswarrantyislimitedtoreplacingorrepairingorissuingacreditnote,atitsoption,foranyproductwhichisreturnedtothefactory,transportationchargesprepaid,andwhichisdeterminedbyKellerAmerica,Inc.tobedefective.Thiswarrantydoesnotapplytobatteriesandaccumulatorsandtoanyproductwhichhasbeensubjectedtoordam-agedduetomisuse,misapplication,negligenceoraccident,orwhichhasbeenrepairedoralteredwithoutexpresspriorandwrittenconsentfromKellerAmerica,Inc.Inordertoobtainserviceunderthetermsofthewarranty,BuyermustnotifyKellerAmerica,Inc.ofanydefectsbeforetheexpirationofthewarrantyperiodandmakesuitablearrangementsfortheperformanceofservices.Inallcases,BuyershallberesponsibleforpackagingandshippingtheproductstotheKellerAmerica,Inc.plantwithshippingchargesprepaid.KellerAmerica,Inc.shallpayforthereturnofanyproductstoBuyeriftheshipmentistoalocationwithinthecontinentalUSA.Buyershallberesponsibleforpayingallshippingduties,taxes,andotherchargesforproductsreturnedtoanyotherlocation.KellerAmerica,Inc.willprovideon-siteserviceonlyuponprioragreementand

Page10

Page 8: Keller America Inc User s Guide Rev. 07/19 · eller America Inc ser’s uide Rev. eller America Inc ser’s uide Rev. IV. Product Overview - Level Transmitters Acculevel - The Acculevel

Keller America Inc User’s Guide Rev. 07/19 Keller America Inc User’s Guide Rev. 07/19

X: Appendix

+ SUPPLY

OUTPUT/GROUND

2-WIRE

TRANSMITTER

POWER

SUPPLY

DISPLAY/

CONTROL

+

-

-

+

2-WIRE CURRENT LOOP

3-WIRE VOLTAGE OUTPUT

+ SUPPLY

+ OUTPUT

3-WIRE

TRANSMITTER

POWER

SUPPLY

DISPLAY/

CONTROL

+

-

-

+

GROUND

A-1

PleaserefertoSectionVII:Installationforminimumsupplyvoltageinformation.

NoassignmentortransferinviolationofthisprovisionshallbevalidorbindingonKellerAmerica,Inc.

Attorney’s fees-UponanybreachofthisAgreement,thereasonableattorney’sfeesandcostsofthesubstantiallyprevailingparty,whetherbylitigationorsettlement,shallbepaid by the breaching party.

Page12

Page 9: Keller America Inc User s Guide Rev. 07/19 · eller America Inc ser’s uide Rev. eller America Inc ser’s uide Rev. IV. Product Overview - Level Transmitters Acculevel - The Acculevel

Keller America Inc User’s Guide Rev. 07/19 Keller America Inc User’s Guide Rev. 07/19

NOTES: 1: FOR TRANSMITTERS WITH LIGHTNING/SURGE PROTECTION, CONNECT DRAIN WIRE TO A LOW-IMPEDANCE EARTH GROUND. FOR TRANSMITTERS WITHOUT LIGHTNING/ SURGE PROTECTION, THIS PIN IS CONNECTED TO THE TRANSMITTER CASE. 2: DO NOT SUBMERSE ELECTRICAL TERMINATION IN LIQUID. IT WILL DAMAGE THE INSTRUMENT.

VALUELINE w/ATTACHED CABLE...4-20mA

VALUELINE w/ATTACHED CABLE...0-5...10VDC

VALUELINE w/DIN43650...4-20mA

VIEW OF REAR OF TRANSMITTER

VALUELINE w/DIN43650...0-5...10VDCVALUELINE

+ OUTPUT

VIEW OF REAR OF TRANSMITTER

3

12

3

12

DRAIN

BLACK: + SUPPLY

WHITE: OUTPUT/GROUND

DRAIN

BLACK: + SUPPLY

WHITE: GROUND

RED: + OUTPUT

+ SUPPLY

OUTPUT/GROUND

SEE NOTE 1.

+ SUPPLY

GROUND

A-3

12

3

4

12

3

4

DRAIN/CASE GROUND

BLACK: + SUPPLY

WHITE: OUTPUT/GROUND

DRAIN

BLACK: + SUPPLY

WHITE: GROUND

RED: + OUTPUT

ECONOLINE w/CABLE...4-20mA

ECONOLINE w/mPm393...4-20mA

ECONOLINE w/CABLE...0,5-4,5VDC

ECONOLINE w/mPm393...0,5-4,5VDC

CASE

+ SUPPLY

OUTPUT/GND

VIEW OF REAR OF TRANSMITTER

CASE

+ SUPPLY

GROUND

VIEW OF REAR OF TRANSMITTER

+ OUTPUT

A-2

NOTES: 1: DO NOT SUBMERSE ELECTRICAL TERMINATION IN LIQUID. IT WILL DAMAGE THE INSTRUMENT.

Page 10: Keller America Inc User s Guide Rev. 07/19 · eller America Inc ser’s uide Rev. eller America Inc ser’s uide Rev. IV. Product Overview - Level Transmitters Acculevel - The Acculevel

Keller America Inc User’s Guide Rev. 07/19 Keller America Inc User’s Guide Rev. 07/19

DRAIN (NOTE 1)

BLACK: + SUPPLY

WHITE: OUTPUT/GROUND

BLUE: RS485A

YELLOW: RS485B

DRAIN (NOTE 1)

BLACK: + SUPPLY

WHITE: GROUND

RED: + OUTPUT

BLUE: RS485A

YELLOW: RS485B

PRECISELINE w/ATTACHED CABLE...4-20mA + RS485

PRECISELINE w/ATTACHED CABLE...0-5...10VDC + RS485

DRAIN (NOTE 1)

BLACK: GROUND

WHITE: SDI-12

RED: + VCC

BLUE: RS485A

YELLOW: RS485B

PRECISELINE SDI w/ATTACHED CABLE...SDI-12+ RS485

A-5

NOTES: 1: FOR TRANSMITTERS WITH LIGHTNING/SURGE PROTECTION, CONNECT DRAIN WIRE TO A LOW-IMPEDANCE EARTH GROUND. FOR TRANSMITTERS WITHOUT LIGHTNING/ SURGE PROTECTION, THIS PIN IS CONNECTED TO THE TRANSMITTER CASE. 2: DO NOT SUBMERSE ELECTRICAL TERMINATION IN LIQUID. IT WILL DAMAGE THE INSTRUMENT.

AB

CD

E

F

AB

CD

E

F

+ SUPPLY

OUTPUT/GROUND

VALUELINE W/MIL-C-26482...4-20mA

REAR VIEW OF TRANSMITTER

VALUELINE W/MIL-C-26482...0-5...10VDC

+ SUPPLY

GROUND

+ OUTPUT

REAR VIEW OF TRANSMITTER

E: DRAIN

A-4

NOTES: 1: FOR TRANSMITTERS WITH LIGHTNING/SURGE PROTECTION, CONNECT DRAIN WIRE TO A LOW-IMPEDANCE EARTH GROUND. FOR TRANSMITTERS WITHOUT LIGHTNING/ SURGE PROTECTION, THIS PIN IS CONNECTED TO THE TRANSMITTER CASE. 2: DO NOT SUBMERSE ELECTRICAL TERMINATION IN LIQUID. IT WILL DAMAGE THE INSTRUMENT.

Page 11: Keller America Inc User s Guide Rev. 07/19 · eller America Inc ser’s uide Rev. eller America Inc ser’s uide Rev. IV. Product Overview - Level Transmitters Acculevel - The Acculevel

Keller America Inc User’s Guide Rev. 07/19 Keller America Inc User’s Guide Rev. 07/19

NOTES: 1: FOR LIGHTNING/SURGE PROTECTION TO BE EFFECTIVE, CONNECT DRAIN WIRE TO A LOW-IMPEDANCE EARTH GROUND. 2: LIGHTNING PROTECTION NOT AVAILABLE ON ECONOLEVEL. DRAIN WIRE CONECTION NOT REQUIRED. 3: DO NOT SUBMERSE ELECTRICAL TERMINATION IN LIQUID. IT WILL DAMAGE THE INSTRUMENT.

LEVELGAGE & ECONOLEVEL w/ATTACHED CABLE...4-20mA

LEVELGAGE w/ATTACHED CABLE...0-5...10VDC

ACCULEVEL & LEVELRAT w/ATTACHED CABLE...4-20mA + RS485

ACCULEVEL & LEVELRAT w/ATTACHED CABLE...0-5...10VDC + RS485

DRAIN (NOTE 1,2)

BLACK: + SUPPLY

WHITE: OUTPUT/GROUND

DRAIN (NOTE 1)

BLACK: + SUPPLY

WHITE: GROUND

RED: + OUTPUT

DRAIN (NOTE 1)

BLACK: + SUPPLY

WHITE: OUTPUT/GROUND

BLUE: RS485A

YELLOW: RS485B

DRAIN (NOTE 1)

BLACK: + SUPPLY

WHITE: GROUND

RED: + OUTPUT

BLUE: RS485A

YELLOW: RS485B

A-7

PRECISELINE w/MIL-C-26482...4-20mA + RS485

REAR VIEW OF TRANSMITTER

PRECISELINE w/MIL-C-26482...0-5...10VDC + RS485

REAR VIEW OF TRANSMITTER

A: + SUPPLY

C: OUTPUT/GROUND

A: + SUPPLY

C: GROUND

B: + OUTPUT

D: RS485A

F: RS485B

D: RS485A

F: RS485B

E: DRAIN

AB

CD

E

F

AB

CD

E

F

PRECISELINE SDI w/MIL-C-26482...SDI-12 + RS485

REAR VIEW OF TRANSMITTER

A: + SUPPLY

C: GROUND

B: SDI-12F: RS485B

D: RS485A

AB

CD

E

F

A-6

NOTES: 1: FOR TRANSMITTERS WITH LIGHTNING/SURGE PROTECTION, CONNECT DRAIN WIRE TO A LOW-IMPEDANCE EARTH GROUND. FOR TRANSMITTERS WITHOUT LIGHTNING/ SURGE PROTECTION, THIS PIN IS CONNECTED TO THE TRANSMITTER CASE. 2: DO NOT SUBMERSE ELECTRICAL TERMINATION IN LIQUID. IT WILL DAMAGE THE INSTRUMENT.

Page 12: Keller America Inc User s Guide Rev. 07/19 · eller America Inc ser’s uide Rev. eller America Inc ser’s uide Rev. IV. Product Overview - Level Transmitters Acculevel - The Acculevel

Keller America Inc User’s Guide Rev. 07/19 Keller America Inc User’s Guide Rev. 07/19

A-9

MINIMUM PIPE BEND RADIUS

Acculevel/ LevelgageØ0.83”

MicrolevelØ0.63”

Schedule40

1”IPS(Ø1.315x0.133”w) 22” 18”

1.5”IPS(Ø1.900”x0.145”w) 4” 5”

2”IPS(Ø2.375x0.154”w) 3” 3”

Schedule80

1”IPS (Ø1.315x0.179”w) 22” 22”

1.5”IPS(Ø1.315x0.200”w) 4” 6”

2”IPS(Ø2.375x0.218”w) 4” 4”

NOTES: 1: FOR LIGHTNING/SURGE PROTECTION TO BE EFFECTIVE, CONNECT DRAIN WIRE TO A LOW-IMPEDANCE EARTH GROUND. 2: DO NOT SUBMERSE ELECTRICAL TERMINATION IN LIQUID. IT WILL DAMAGE THE INSTRUMENT.

MICROLEVEL & NANOLEVEL w/ATTACHED CABLE...4-20mA + RS485

DRAIN (NOTE 1)

BLACK: + SUPPLY

WHITE: OUTPUT/GROUND

BLUE: RS485A

YELLOW: RS485B

DRAIN (NOTE 1)

BLACK: GROUND

WHITE: SDI-12

RED: + VCC

BLUE: RS485A

YELLOW: RS485B

ACCULEVEL SDI w/ATTACHED CABLE...SDI-12+ RS485

A-8

Page 13: Keller America Inc User s Guide Rev. 07/19 · eller America Inc ser’s uide Rev. eller America Inc ser’s uide Rev. IV. Product Overview - Level Transmitters Acculevel - The Acculevel

Keller America Inc User’s Guide Rev. 07/19 Keller America Inc User’s Guide Rev. 07/19

A-11

Signal cable

Transmitter

Vent tube

Bellows Assembly

Drying Tube Assembly

VENT TUBE TERMINATION - DRYING TUBE OR BELLOWS

Remove yellowcap after installation

NOTES: 1: FOR TRANSMITTERS WITH LIGHTNING/SURGE PROTECTION, CONNECT DRAIN WIRE TO A LOW-IMPEDANCE EARTH GROUND. FOR TRANSMITTERS WITHOUT LIGHTNING/SURGE PROTECTION, THIS PIN IS CONNECTED TO THE TRANSMITTER CASE. 2: DRYING TUBE ASSEMBLY - CLEAR TUBE FILLED WITH INDICATING DESICCANT, ATTACHES DIRECTLY TO CABLE VENT TUBE, INTERCEPTS WATER VAPOR. HIGHLY RECOMMENDED WHEN OPERATING IS IN HIGH HUMIDITY CONDITIONS. MUST BE PERIODICALLY RENEWED AS DESICCANT BECOMES SATURATED, TURNING COLOR FROM BLUE (DRY) TO PINK (SATURATED). REMOVE YELLOW CAP FROM FREE END AFTER INSTALLATION. 3: BELLOWS ASSEMBLY - ALTERNATIVE TO THE DRYING TUBE, THIS ANEROID BELLOWS ATTACHES TO CABLE VENT TUBE AND REQUIRES NO PERIODIC MAINTENANCE. RECOMMENDED WHERE A SLIGHT SACRIFICE IN ACCURACY CAN BE TOLERATED.

A-10

Page 14: Keller America Inc User s Guide Rev. 07/19 · eller America Inc ser’s uide Rev. eller America Inc ser’s uide Rev. IV. Product Overview - Level Transmitters Acculevel - The Acculevel

Electronic copies of this user’s guide can be found at: http://www.kelleramerica.com/manuals-and-software/manuals/Keller_America_Users_Guide.pdf