Date post: | 18-Nov-2014 |
Category: |
Documents |
Upload: | api-3727411 |
View: | 210 times |
Download: | 0 times |
BA 237C/07/en/01.0151504342
Software version:1.0 or higher
Liquisys MCLM 223 FTransmitter for Conductivity
Hauser+EndressThe Power of Know How
Operating Instructions
CAL REL
REL1
REL1
ALARMREL2
REL2
ENDRESS+HAUSERLIQUISYS M
Need information on the instrument?Please read the following chapters:
Safety
General information
You wish to install the instrument.The required steps are described in these chapters:
Installation
You wish to operate or reconfigure the instrument.The operating concept is explained in these chapters:
Operation
Instrument configuration
Interfaces
Need help with problems or maintenance?
Maintenance and troubleshooting
Diagnosis and corrective maintenance
Accessories
Appendix
Technical data
Index
%Ω°C
LM2XFE00.CHP
Table of Contents1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.1 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.2 Storage and transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.3 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.4 Packaging and disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.5 Product structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42.1 Intended application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42.2 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42.3 Installation, start-up, operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42.4 Monitoring and safety features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52.5 Immunity to interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52.6 Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3 Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63.1 Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73.3 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73.4 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83.5 Sensor installation and cable connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124.1 Operator interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124.3 Key functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134.4 Auto / manual mode of operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144.5 Operating concept. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154.6 Access codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174.7 Display during measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174.8 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5 Instrument configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185.1 Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205.2 System configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205.3 Current outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225.4 Monitoring functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 245.5 Relay configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 275.6 Temperature compensation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 295.7 Concentration measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 325.8 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 345.9 E+H Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 375.10 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 375.11 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 385.12 Determination of temperature coefficient. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 415.13 Remote measuring range switching (MRS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
7 Maintenance and troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 467.1 Definition of terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 467.2 Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 467.3 Troubleshooting common problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 467.4 Problem elimination based on error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
8 Diagnosis and corrective maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 518.1 Definition of terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 518.2 Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 518.3 Diagnosis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 518.4 Corrective maintenance of Liquisys M CLM 223 F . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 538.5 Spare parts orders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 568.6 “Optoscope” service equipment with “Scopeware” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 568.7 Corrective maintenance of measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
9 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
10 Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
11 Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
12 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Liquisys M CLM 223 F Table of contents
Endress+Hauser 1
1 General information
1.1 Symbols used
Warning:
This symbol alerts to hazards whichmay cause serious injuries as well asdamage to equipment if ignored.
Note:
This symbol indicates importantitems of information.Ignoring this information may resultin malfunction.
1.2 Storage and transport
The packaging material used to store ortransport the transmitter must provide shockprotection. Optimal protection is provided bythe original packaging materials.
Conformance with the ambient conditions(see technical data) must be assured.
1.3 Unpacking
Verify that the packaging and contents areundamaged! Inform the post office or freightcarrier of any damage.Damaged merchandise must be retained untilthe matter has been settled.
Check that the delivery is complete andagrees with the shipping documents and yourorder (refer to nameplate for type andversion).
The delivery includes:
• Transmitter Liquisys M CLM 223 F• Operating instructions BA 237C/07/en• 1 set of plug-in screw terminals• 2 fastening clips
Keep the original packaging materials forfuture storage or shipping of the instrument.
If you have any questions, consult yoursupplier or the Endress+Hauser sales agencyin your area (see back cover of theseoperating instructions for addresses).
1.4 Packaging and disposal
Package the assembly properly for reuse at alater point in time. Optimal protection isprovided by the original packaging materials.Observe local regulations for disposal.
General information Liquisys M CLM 223 F
2 Endress+Hauser
1.5 Product structure
You can identify the instrument version by theorder code on the nameplate.
1310
85-4
B
ENDRESS+HAUSERLIQUISYS M
order code / Best.Nr.serial no. / Ser.-Nr. Codes:
::
::
::
:::
prot. class / Schutzartambient temp. / Umgebungstemperatur
output 1 / Ausgang 1output 2 / Ausgang 2mains / Netz
measuring range / Messbereichtemperature / Temperatur
CLM 223-CF0110123456
0...2000 mS/cm
conductivity cond./ Leitfähigkeit kond.
-35...250 °C
0/4...20 mA0/4...20 mA230 VAC 50/60 Hz 7,5 VA
-10...+55 °CIP54 / IP30
TYPL223F.CDR Fig. 1.1 Nameplate of CLM 223 F
Liquisys M S CLM 223 F
Version
CF Conductivity measurement (two-electrode sensor)IF Conductivity measurement (inductive sensor)
Power supply
0 230 V AC1 115 V AC5 100 V AC8 24 V AC/DC
Measurement output
0 Conductivity1 Conductivity and temperature3 PROFIBUS-PA4 PROFIBUS-DP5 Conductivity with HART6 Conductivity and temperature with HART
Contacts
05 No additional contacts10 2 contacts (limit)
complete order codeCLM223-
LM2XFE01.CHP
Liquisys M CLM 223 F General information
Endress+Hauser 3
2 Safety
2.1 Intended application
Liquisys M CLM 223 F is a field-tested andreliable transmitter used to determine theconductivity of liquid media.
Liquisys M CLM 223 F is particularly suitablefor use in the following areas of application:
• Concentration control• CIP plants control• Phase separation• Product quality assurance• Wash-mobiles and cleaning plants
2.2 General safety instructions
This device has been manufactured for safeoperation according to the state of the art inengineering and conforms to the applicableregulations and European standards (seeTechnical data). It has been designedaccording to EN 61010-1 and has left themanufacturer’s works in perfect condition.
However, if used improperly or for purposesother than the intended purpose, it may bedangerous, e.g. due to incorrect connection.
Warning:
• Operating this instrument in anyway other than as described inthese instructions may compromisethe safety and function of themeasuring system and is thereforeimpermissible.
• The notes and warnings in theseinstallation and operatinginstructions must be strictlyadhered to!
2.3 Installation, start-up, operation
Warning:
• This device may only be installed,connected electrically,commissioned, operated andserviced by properly trainedpersonnel authorized by thesystem operator.
• The personnel must be familiarwith these operating instructionsand must adhere to theinstructions described therein.
• Make sure that the power supplyratings match the data specifiedon the nameplate before youconnect the instrument to a powersource.
• A clearly identified mainsdisconnecting device must beinstalled close to the instrument.
• Live components can be touchedthrough the vent slots in thehousing and the openings on therear of the housing. Do not inserttools, wires, etc., in these slots.
• Check that all connections havebeen properly made beforepowering up the system!
• Damaged equipment that may bedangerous must not be operatedand should be clearly identified asbeing defective.
• Any troubleshooting of themeasuring system is to beperformed exclusively byauthorized, trained personnel.
• If faults cannot be remedied, theinstrument must be removed fromservice and secured to preventaccidental start-up.
• Repairs not described in theseoperating instructions may only beperformed at the manufacturer’sworks or by the Endress+HauserService Organization.
Safety Liquisys M CLM 223 F
4 Endress+Hauser
2.4 Monitoring and safety features
Safety features
The transmitter is protected against externalinfluences and damage by the followingdesign measures:
• Rugged housing• UV resistance
Monitoring features
In the event of a system error or power failure,an alarm condition is signalled via a fault-signalling contact.
2.5 Immunity to interference
This instrument has been tested according tothe applicable European standards forindustrial applications with regard toelectromagnetic compatibility. It is protectedagainst electromagnetic interference bydesign measures.
Warning:
The specified immunity tointerference only applies for devicesconnected as outlined in theseoperating instructions.
2.6 Declaration of conformity
Liquisys M CLM 223 F has been developedand manufactured in accordance withcurrently valid European standards anddirectives.
LM253E02.CHP
Liquisys M CLM 223 F Safety
Endress+Hauser 5
3 Installation
The following procedure should be followedfor a complete measuring system installation:
• Installation or attachment of transmitter(see chapter 3.3)
• Selection and connection of cable andsensor (see chapters 3.4, 3.5 and 9)
• Installation is followed by start-up (seechapter 5).
3.1 Measuring system
The complete measuring system comprises:
• The Liquisys M CLM 223 F transmitter• A sensor with or without an integrated
temperature sensor• A measuring cable CYK 71 (conductive
measurement) or measuring cable CLK 5(cable for inductive measurementpermanently attached to sensor)
Optional:
• Extension cable CYK 71 (conductive)or CLK 5 (inductive)
• Junction box VBM
ENDRESS+HAUSERLIQUISYS M
CAL REL
REL1
REL1
ALARMREL2
REL2E
ENDRESS+HAUSERLIQUISYS M
CAL REL
REL1
REL1
ALARMREL2
REL2E
L2X3FSYS.CDR
CLS 21(conductive sensor)with integr. Pt 100
CLS 50(inductive sensor)with integr. Pt 100
Liquisys M S CLM 223
Liquisys M S CLM 223
Fig. 3.1
Complete measuringsystems consisting ofLiquisys M S CLM 223 Fwith measuring cable andconductivity sensor
Installation Liquisys M CLM 223 F
6 Endress+Hauser
3.2 Dimensions
3.3 Mounting
The instrument is attached using the suppliedtensioning screws (see Fig. 3.3).The required overall installation depth isapprox. 165 mm.
ENDRESS+HAUSERLIQUISYS F
EINBAU2.CDR
96
613
990
Fig. 3.2 Liquisys M S CLM 223 F
MONT9.CDR
92+0.5
92+
0.5
90
57
33
approx.25
max. 45
96
92
1
2
3
139 6
3
Fig. 3.3
Attachment ofLiquisys M S CLM 223 F1 Wall of control
cabinet2 Gasket3 Tensioning screws
ENDRESS+HAUSERLIQUISYS F
EINBAU1.CDR
LM2XFE03.CHP
Liquisys M CLM 223 F Installation
Endress+Hauser 7
3.4 Electrical connection
Connection diagram
The connection diagram depicted in Fig. 3.4shows the connections for an inductive orconductive sensor (dashed lines).
The connection of the various sensors isshown in more detail in Figs. 3.7 to 3.9.
Note:
The instrument has protection class IIand is generally operated withoutprotective earth connection.
11
12
13
47
48
49
57
58
59
15 V85
86
10-50 V
81
82
10-50 V
93
94
41
42
43
Cd
mA31
32
mA33
34
S
84
83
15
16
84
PE
83
L+
L–N
L∼ =
PE
ANLF01.CDR
Aux. voltageoutput
Digitalinput 2(MRS)
Digitalinput 1
(HOLD / MRS)
Signal output 2temperature
Signal output 1conductivity
Temperature
Sensor(conductive)
Sensor(inductive)
Alarm(de-energized
contactposition)
Relay 1(de-energized
contactposition)
Relay 2(de-energized
contactposition)
Mains
* optional
*
*
Ground terminal(only at version -CF)
Fig. 3.4
Electrical connectionof Liquisys M S CLM 223F(all inputs and outputsconnected)
Installation Liquisys M CLM 223 F
8 Endress+Hauser
Instrument connections
Power supply of digital inputs
REL1
REL2
REL3
REL4
NC
NC
49
COM
COM
COM
COM
59
54
55
56
52
51
53
58
57
48
47
131084-4B
LM-AN02.CDR
Groundterminal
Fig. 3.5Connection compartmentstickers for CLM 223 F
15 V85
86
81
82
93
94
HOLD / MRS
*
HOLD / MRS
*
*: Function »MRS 4 measuring ranges«: switches are digital coded (s. table chp. 5.13). DIGI_EIN.CDR Fig. 3.6Power supply of digitalinputs
LM2XFE03.CHP
Liquisys M CLM 223 F Installation
Endress+Hauser 9
3.5 Sensor installation and cable connection
Measuring cable connection
The conductivity sensors are connected usinga special, shielded multi-core cable.Termination instructions are supplied with themeasuring cables.Use junction box VBM (see chapter 9) toextend the measuring cable.
Structure and termination of measuring cables
Special measuring cable required for connection of conductivity sensors
Sensor type Cable Extension
Two-electrode sensor with or withouttemperature sensor Pt 100 CYK 71 Junction box VBM
+ CYK 71
Inductive sensor CLS 50, CLS 52 Cable permanentlyattached to sensor
Junction box VBM+ CLK 5
Maximum cable length
Conductive conductivitymeasurement
max. 100 m with CYK 71 (corresponds to 10 nF). Atk=1 you have to be aware of lower precision for lineresistances bigger than 5 Ohm in the range of 200mS/cm. In that case please use a cable with agreater cross section.
Inductive conductivity measurement max. 55 m (with CLK 5 and sensor cable)
KABEL.CDR
Inner conductor
Green (11)
White (12)
Yellow (13)
CYK 71
Semiconductorscreen
Screen
Green (11)
White (1)
Yellow (13)
Brown(unused)
Semiconductorscreen
White
Red
CLK 5
Brown(unused)
Black
(83)
(84)
(84) (83)
(15)(16)
Fig. 3.7
Structure of specialmeasuring cables CYK 71(left) and CLK 5 (right)
Installation Liquisys M CLM 223 F
10 Endress+Hauser
Examples for connection
Note:
The sensor screen must be groundedto ensure faultless operation of theconductive instrument. For thatpurpose, a PE distribution board isavailable. This is located on thecover frame of the CLM 223 F.The distribution board has noconnection to the transmitter. It hasto be connected with PE when thetransmitter is installed.
gn wt ye
gn wt ye bk
bk
ϑ
83 15 16 84
23 1
11 12 13
CYK 71
PE
LM2X3CDE.CDR Fig. 3.8
Connection of conductivesensors (CLS 15, CLS 19,CLS 20, CLS 21)
wt rd
CLK 5 (fixed cable)
ϑ
84 15 83 1611 12 13S
gn wt yeorbn
bl rd rdbl
wt rd
CLK 5 (fixed cable)
ϑ
84 15 83 1611 12 13S
gn wt yeorbn
bl rd rdbl
LM2X3IDE.CDR Fig. 3.9
Connection of inductivesensors (left: CLS 50,right: CLS 52)
LM2XFE03.CHP
Liquisys M CLM 223 F Installation
Endress+Hauser 11
4 Operation
4.1 Operator interface
4.2 Display
LED indicators
Liquid crystal display
9999mS/cm
R2
On value
ENDRESS+HAUSERLIQUISYS F
DISPL1-E.CDR
Liquid crystal displayfor display of measuredvalues andconfiguration data
4 main control keys forcalibration andinstrument configuration
Key for switchingbetween automatic /manual operation
LED indicators forlimit contactor relays(switching state)
LED indicator for alarmfunction
Display of activecontact and key forrelay switching inmanual mode
Field for user labelling
Fig. 4.1Operating elements ofLiquisys M S CLM 223 F
Indication of current operating mode: “auto” (green LED) or “manual” (yellow LED)
Indication for relay controlled in “manual” mode (red LED)
Indicates the state of relays 1 and 2.LED green: measured value is within permissible limits, relay is inactive.LED red: measured value is outside permissible limits, relay is active.
Alarm indication for continuous limit violation, temperature sensor failure or system errors(see error list in chapter 7)
9999mS/cm
R2
On value
DISPL3-E.CDR
Error indicator
Sensor symbolflashes duringactive calibration
Measuring modeindicator (normaloperation)
Calibration modeindicator
Setup modeindicator(configuration)
Hold mode indicator(outputs reflect lastcurrent status)
On instruments withcommunication:indication for receptionof a message.Permanently visible inburst mode (with HART)
Function coding display
In measuring mode:main meas. value.In setup / calibr. mode:e.g. parameter adjusted
In measuring mode:secondary meas. value.In setup / calibr. mode:e.g. parameter setting
Fig. 4.2 Liquid crystal display
Operation Liquisys M CLM 223 F
12 Endress+Hauser
4.3 Key functions
CAL key
When the CAL key is pressed, the instrumentprompts for the calibration access code (fixedsetting: 22 for calibration or any other numberto view the calibration data). Acknowledgewith the CAL key to proceed. Use the CALkey to continue through the calibrationprocess.
Note:
The calibration data set in functiongroup C is used for calibration.
ENTER key
The ENTER key has several functions:• Calling up the setup menu from the
measuring mode• Storage (acknowledgement) of data
entered in the setup mode• Starting calibration (same function as
CAL key).
PLUS key
MINUS key
The PLUS and MINUS keys have the followingfunctions:• Selection of function groups• Setting parameters and numeric values
(the setting speed increases when the keyis held down)
• Relay operation in manual mode(see chapter 4.4).
In the measuring mode, the functions of thePLUS and MINUS keys are (see chapter 4.7):• Temperature display in °F (PLUS)• Suppression of temperature display (PLUS)• Display of the uncompensated conductivity
value (PLUS)• Selection of error display (MINUS).
REL key
The REL key is used in manual mode toswitch between the relays and the manualstart cleaning function.
AUTO key
This key is used to toggle between theautomatic and manual modes of operation.
Escape function
Press the PLUS and MINUS keyssimultaneously to return to the measuringmode (in calibration mode: at the end ofcalibration). Press the PLUS and MINUSkeys again to return to the main menu.
Locking the hardware
Access to field operation can be locked forcommunication via HART or Profibus.
Press the PLUS key and the ENTER key at thesame time to lock the instrument. The codeprompt displays the code 9999.
Unlocking the hardware
To unlock, press the CAL and MINUS keys atthe same time. The code prompt displays thecode 0.
LM2XFE04.CHP
Liquisys M CLM 223 F Operation
Endress+Hauser 13
4.4 Auto / manual mode of operation
Auto mode
In this mode of operation, the relaysare controlled by the transmitter.
Manual mode
In this mode all relays can becontrolled manually.
Switching to manual mode
The instrument is switched to themanual mode by pressing thefollowing keys:
1. Press AUTO key.
2. Enter code 22.
3. Select relay or function. Press theREL key to toggle between therelays. The display shows theselected relay in the second line.
4. Set the relays. Switch on withPLUS, switch off with MINUS. Therelay state remains in effect until itis actively reset.
5. Press AUTO key for returning tothe measuring mode.
Note:
• Enable the manual mode byentering access code “22”.
• The operating mode remains ineffect even after a power failure.
• The manual mode takesprecedence over any otherautomatic function (hold).
• Hardware locking in the manualmode is not possible.
• The manual settings remain ineffect until they are actively reset.
• Error code E102 is signalled in themanual mode.
Operation Liquisys M CLM 223 F
14 Endress+Hauser
4.5 Operating concept
Operating modes
Note:
• The user can put the functions andcontacts on hold during calibrationand configuration (see chapter 5.8,function S2); the duration of thehold period may also be changed.
• If you do not press a key in setupmode for approx. 15 min, then theLCD returns automatically tomeasuring mode. An active holdfunction (hold at setup) is thenreset.
Code
Selectfunction groupswith PLUS orMINUS key.
Calibration mode:Run throughcalibration routine
Measuring mode:Standard operating state,indicates currentmeasured values
Setup mode:Access to allconfigurationsettings
Code input:Use Code 22 to changesettings in functiongroups.For all other codes,you can only readsettings but notchange them.Code 22 is also requiredfor calibration.
SETMOD-E.CDR Fig. 4.3Description of operatingmodes
LM2XFE04.CHP
Liquisys M CLM 223 F Operation
Endress+Hauser 15
Menu structure
The configuration and calibration functionsare arranged in a menu structure by functiongroups.
The function groups are selected in the setupmode with the PLUS and MINUS keys. TheENTER key is used to move from one functionto the next within a function group. The PLUSand MINUS keys are used for option selectionand editing. Selections must be confirmed bypressing the ENTER key. This also moves thecursor to the next function.Pressing the PLUS and MINUS keys at thesame time terminates programming (return tomain menu).
Note:
• If a change is made but notconfirmed by pressing the ENTERkey, the previous setting isretained.
• See the appendix of theseoperating instructions for anoverview of the Liquisys MCLM 223 F menu structure.
HOLD function: »freezing« the outputs
The current output can be »frozen« in thesetup mode and during calibration, i.e. thelast current value (resp. the in field S3 definedvalue) is constantly output. The display showsthe »HOLD« message (see chapter 5.8, p. 34for hold settings).
• During HOLD in the automatic mode thecontacts will go to the idle position(if factory settings are active).
• At CC, timer and external HOLD HOLDalways is active. Adjustments in the fieldsS2 and S3 will be ignored.
• An alarm delay will be reset to »0«.• This function can also be activated
externally via the HOLD input(see connection diagram in Fig. 3.4,digital input 1).
• The manual HOLD (field S3) also remainsactive after a power failure.
EE E EEE E E
-
BEDSCHE.CDR
Functions(parameter selection, value entry)
Function groupsMove forward and backward
with the PLUS and MINUS keys
Move from one function to thenext with the ENTER key
Fig. 4.4
Schematic representationof Liquisys M S CLM 223F menu structure
Operation Liquisys M CLM 223 F
16 Endress+Hauser
Hold function: “freezes” the outputs
The current output is “frozen” in the setupmode and during calibration, i.e. the lastcurrent value is constantly output. The displayshows the “HOLD” message (see chapter 5.8for hold settings).
Note:
• During hold, all contacts will go totheir normal positions.
• An active hold has priority over all otherfunctions.
• Any alarm delay accumulated will be resetto “0”.
• The hold function can also be activatedexternally via the hold input (seeconnection diagram in Fig. 3.4; digitalinput 1).
• The manual hold (field S3) remains activeeven after a power failure.
4.6 Access codes
All instrument access codes are fixed, i.e.they cannot be modified. Three access codesare distinguished (cf. Fig. 4.3):
• Any code: Read mode access, i.e. allsettings can be viewed but not modified(access with ENTER/CAL key, see Fig. 4.3).
• Code 22: Access to calibration and offsetmenus (access with CAL key, see Fig. 4.3).
• Code 22: Access to configuration menusfor instrument configuration and usersettings (access with ENTER key, seeFig. 4.3).
• See chapter 4.3 for hardware locking andunlocking.
4.7 Display during measurement
The measurement display can be individuallyadapted to user requirements.
Settings accessed with PLUS key:
• Press the PLUS key to display thetemperature in °F instead of °C.
• Press the PLUS key a second time tosuppress the temperature display.
• Press the PLUS key a third time to displaythe uncompensated conductivity value.
• Press the PLUS key a fourth time to returnto the standard display.
Settings accessed with MINUS key:
• Press the MINUS key to display the firsterror message.
• Press the MINUS key repeatedly to displaythe following error messages (up to ten).If no further errors exist, the display returnsto the measuring mode.
Note:
Function group F (alarm, seechapter 5.4.1) can be used todefine the alarm functions for eachindividual error code.
4.8 Calibration
Please refer to chapter 5.11 for the calibrationprocedure.
LM2XFE04.CHP
Liquisys M CLM 223 F Operation
Endress+Hauser 17
5 Instrument configuration
After power-up (connection to power), theinstrument performs a self-test and thenenters the measuring mode.
Now it can be configured and calibrated forthe first time.
The following function groups are available onthe Liquisys M CLM 223 F:
Setup mode
SETUP 1 (A) see chapter 5.2.1 SETUP 2 (B) see chapter 5.2.2 OUTPUT (O) see chapter 5.3 ALARM (F) see chapter 5.4.1 CHECK (P) see chapter 5.4.2 RELAY (R) see chapter 5.5 ALPHA TABLE (T) see chapter 5.6 CONCENTRATION (K) see chapter 5.7 SERVICE (S) see chapter 5.8 E+H SERVICE (E) see chapter 5.9 INTERFACE (I) see chapter 5.10 TEMPERATURE
COEFFICIENT (D) see chapter 5.12 MRS (M) see chapter 5.13
Calibration mode
CALIBRATION (C) see chapter 5.11
Selecting and locating functions is facilitatedby a code displayed for each function in aspecial display field (Fig. 5.1).The structure of this coding is shown inFig. 5.2. The first column indicates thefunction group as a letter (see function groupdesignations). The functions in the individualgroups are counted from the top to thebottom and from the left to the right.
Factory settings
When the instrument is switched on for thefirst time, the factory settings are in effect. Thefollowing table provides an overview of allmajor settings.
Please refer to the description of theindividual functions in chapter 5 for all otherfactory settings (the factory settings areprinted in bold face there).
2s
F3
Err.DelayDISPLAY4.CDR
Function display:The displayed code indicatesthe function position in thefunction group.
Additional information
Fig. 5.1Example of display insetup mode
A A1 A111
A121
A131 A132 A133
MATRIX.CDRFig. 5.2 Function coding
Type of measurement Inductive or conductive conductivity measurement(acc. to order), temperature measurement in °C
Temperature compensation type Linear with reference temperature 25 °C
Temperature compensation Automatic (ATC on)
Limit for controller 1 + 2 2000 mS/cm
Hold Active during configuration and calibration
Measuring range 10 µS/cm ... 2 S/cm (no adjustment neccessary,floating adjustment.
Instrument configuration Liquisys M CLM 223 F
18 Endress+Hauser
The customer-specific settings can beentered in the free fields provided for thispurpose in the operating menus (chapter 11,Appendix).
Alarm contact (Fail-safe contact)
Normal operating state:
• Instrument in operation• No error message present
(alarm LED green)
→ Relay energised→ Contact 42/43 closed
Alarm state:
• Error message present (alarm LED red) or• Instrument defective or dead
(alarm LED off)
→ Relay de-energised→ Contact 41/42 closed
Current outputs 1 and 2* 4 ... 20 mA
Current output 1: measured valuefor 4 mA signal current 0 µS/cm
Current output 1: measured valuefor 20 mA signal current 2000 mS/cm
Current output 2*: temperaturevalue for 4 mA signal current* 0.0 °C
Current output 2*: temperaturevalue for 20 mA signal current* 150.0 °C
*If equipped accordingly
RELAIS2.CDR
Normal operating state Alarm state
41
43
42
41
42
43
Fig. 5.3
Recommended fail-safecircuit for the alarmcontact
LM2XF5A.CHP
Liquisys M CLM 223 F Instrument configuration
Endress+Hauser 19
5.1 Start-up
After power-up (connection to power supplyestablished), the user must make thefollowing selections in the specified functiongroups:
• Function group SERVICE (S)S1: Select language and exit functiongroup (not required for English).
• Function group SETUP 1 (A)Adjust all the parameters in this group; seechapter 5.2.1.
• Function group SETUP 2 (B)Adjust all the parameters in this group; seechapter 5.2.2.
Other configuration options are explained inthe chapters to follow for each menu.
5.2 System configuration
The system is configured using the functiongroups SETUP 1 and SETUP 2. Themeasurement type and sensor are selectedhere, and the settings for temperaturemeasurement are made.
All the parameters in these two functiongroups are to be configured to avoidmeasuring errors or failure to measure at all.
Note:
For the measuring range 1(MR1)only the following settings canbe adjusted in the fields M1-M15 inthe same way as shown in Setup 1and 2.
5.2.1 Setup 1
Coding Field Selection or rangeFactory setting
Display Info
AFunction groupSETUP 1
Initial display in the functiongroup SETUP 1.
A1Operating modeselection
cond = conductiveconc = concentration
Display varies according toinstrument version:– cond/conc– ind/conc
Any change in operatingmode causes anautomatic reset ofuser settings.
A2Selection ofconcentration unitto be displayed
%ppmmg/lTDSnone
A3Selection ofdisplay format forconcentration unit
XX.xxX.xxxXXX.xXXXX
A
SETUP 1
cond A1
Oper.Mode
% A2
Conc.Unit
XX.xx A3
Format
Instrument configuration Liquisys M CLM 223 F
20 Endress+Hauser
5.2.2 Setup 2
Coding Field Selection or rangeFactory setting
Display Info
A4Selection of unit tobe displayed
auto, µS/cm, mS/cm,S/cm, µS/m, mS/m,S/m
When “auto” is selected, themaximum resolution possible isautomatically selected.
A5Enter cell constantfor sensorconnected
0.0025... 5.9 ... 99.99Refer to the sensor qualitycertificate for the exact cellconstant.
A6
Edit Installationfactor
0.10 ... 1 ... 5.00
This is where the installationfactor is edited. The correctfactor is determined in C1(3), seechapter 5.11 or referring to thediagram in Fig. 3.6.
Edit Cableresistance
0Ω0 ... 99.99Ω
A7Enter measuredvalue damping
11 ... 60
Measured value damping causesaveraging over the specifiednumber of individual measuredvalues. It is used, for example, tostabilise the display withapplications that fluctuate a greatdeal. There is no damping if “1”is entered.
Coding Field Selection or rangeFactory setting
Display Info
BFunction groupSETUP 2
Initial display in the functiongroup SETUP 2.
B1Selection oftemperaturesensor
Pt100Pt1k = Pt 1000NTC30fixed
If set to “fixed”:no temperature measurement, afixed temperature value isentered instead.
B2
Selection oftemperaturecompensationtype
lin = linearTab = table 1 ... 4NaCl = common salt
(IEC 60746)none
This option is not displayed forconcentration measurement.
auto A4
Unit
5.900 A5
Cellconst
1/cm
A6
InstFac
1.00
A6
Ohm cable
0@
1 A7
Damping
B
SETUP 2
B1
ProcTemp.
Pt100
B2
TempComp.
lin
LM2XF5A.CHP
Liquisys M CLM 223 F Instrument configuration
Endress+Hauser 21
5.3 Current outputs
The function group OUTPUT is used toconfigure the individual outputs.
Furthermore, a current output value can besimulated to check the current outputs (O2 (2)).
Coding Field Selection or rangeFactory setting
Display Info
B3Enter temperaturecoefficient α
2.1 %/K0.0 ... 20.0 %/K
Only if B2 = lin.Any table defined is not active inthis case.
B4Enter processtemperature
25.0 °C–10.0 ... 150.0 °C
Only if B1 = fixed.
This value can only be specifiedin °C.
B5Temperaturesensor calibration
Entry and display ofreal temperature–10.0 ... 150.0 °C
This entry is used to calibrate thetemperature sensor to anexternal measurement.Omitted if B1 = fixed.
B6Enter temperaturedifference (offset)
0.0 °C–5.0 ... 5.0 °C
The offset is the differencebetween the temperaturesmeasured and entered.
Omitted if B1 = fixed.
B7Enter referencetemperature
25.0 °C–35 ... +250 °C
Coding Field Selection or rangeFactory setting
Display Info
OFunction groupOUTPUT
Initial display in the functiongroup OUTPUT.
O1Selection ofcurrent output
Out1Out2
A different characteristic can beselected for each output.
O2 (1)Choose linearcharacteristic orsimulation
lin = linear (1)sim = simulation (2)
The slope of the characteristicmay be positive or negative.
B3
Alpha val
2.10%/K
B4
ProcTemp.
25.0°C
B5
RealTemp.
0.0°C
B6
TempOffs.
0.0°C
B7
RefTemp.
25.0°C
O
OUTPUT
O1
Sel Out
Out1
O2
Sel. Type
lin
Instrument configuration Liquisys M CLM 223 F
22 Endress+Hauser
Coding Field Selection or rangeFactory setting
Display Info
O211Selection ofcurrent range
4–20 mA0–20 mA
O212
0/4 mA value;entercorrespondingmeasured value
Cond.: 0.00 µS/cmConc.: 0.00 %Temp.: 0.0 °Centire meas. range
Enter the measured valuecorresponding to the minimumcurrent value (0/4 mA) at thetransmitter output.Display format from A3.(Spreading: see Technical data.)
O213
20 mA value;entercorrespondingmeasured value
Cond.: 2000 mS/cmConc.: 99.99 %Temp.: 150.0 °Centire meas. range
Enter the measured valuecorresponding to the maximumcurrent value (20 mA) at thetransmitter output.Display format from A3.(Spreading: see Technical data.)
O2 (2)Current outputsimulation
lin = linear (1)sim = simulation (2)
The simulation is terminated byselecting (1).
O221Enter simulationvalue
current value0.00 ... 22.00 mA
The current value entered here isoutput through the current output.
O211
Sel.Range
4-20
O212
0/4 mA
0.00µS/cm
O213
20 mA
2000mS/cm
O2
Sel. Type
sim
O221
Simulat.
4.00mA
LM2XF5A.CHP
Liquisys M CLM 223 F Instrument configuration
Endress+Hauser 23
5.4 Monitoring functions
The monitoring functions are used to definevarious alarms and to set output contacts.Each individual error can defined to beeffective or not (at the contact or as an errorcurrent). Moreover, sensor polarisation (P1)
can be detected, or the measuring signal canbe checked to ascertain that the sensor isworking properly (it must supply plausiblevalues). An alarm condition can be defined toactivate a cleaning function (F8).
5.4.1 Alarm
Coding Field Selection or rangeFactory setting
Display Info
FFunction groupALARM
Alarm function settings.
F1Selection ofcontact type
Stead = steadycontact
Fleet = fleetingcontact
The contact type selected hereonly applies to the alarm contact.
F2Selection of timeunit
smin
F3 Enter alarm delay0 s (min)0 ... 2000 s (min)
Depending on the unit selectedin F2, the alarm delay is enteredin s or min.The alarm delay does not affectthe LED; it indicates the alarmimmediately.
F4 Selection of errorcurrent
22 mA2.4 mA
This selection must be madeeven if all error messages aresuppressed in F5.
The error current must notlie within the current range(field O211).
F5 Selection of error11 ... 255
This is where the errors areselected that are to trigger analarm signal. The errors areselected via the error number.Please refer to the table inchpt. 7 for the error numbers.The factory settings remain ineffect for all errors not edited(see factory settings in chp. 7.4).
F
ALARM
F1
Cont.Type
Stead
F2
Time Unit
s
F3
Err.Delay
0s
22mA F4
Err.Curr
F5
Sel.Error
1
Instrument configuration Liquisys M CLM 223 F
24 Endress+Hauser
5.4.2 Check
Polarisation detection
Polarisation effects in the interface betweenelectrode and measuring solution limit themeasuring range of conductive conductivitysensors. The Liquisys M CLM 223 F
transmitter has the ability to detectpolarisation effects using an intelligent signalevaluation process.
PCS alarm (Process Check System)
This function is used to examine themeasuring signal for changes. If themeasuring signal is constant for a specificperiod of time (several measured values), analarm is issued. This type of sensor behaviourmay be caused by soiling, etc.
Note:
An active PCS alarm is automaticallycleared when the measuring signalchanges.
Coding Field Selection or rangeFactory setting
Display Info
F6Set alarm contactto be effective forselected error
yesno
If set to “no”, all the other alarmsettings (e.g. alarm delay) arealso deactivated. The settingsthemselves are retained.This setting only applies to theerror selected in F5.Factory setting is no starting withE080!
F7Set error currentto be effective forselected error
noyes
The error current selected in F4becomes effective or issuppressed when it occurs.This setting only applies to theerror selected in F5.
F8Return to menu orselect next error
next = next error<—-R
If <—-R is selected, the programreturns to F. If next is selected,the program returns to F5.
LIVE.CDR
Constant measuring signal= alarm is triggered
t
Fig. 5.4 PCS alarm (live-check)
yes F6
Rel.Assg
no F7
Curr.Assg
F8
Select
next
LM2XF5A.CHP
Liquisys M CLM 223 F Instrument configuration
Endress+Hauser 25
Coding Field Selection or rangeFactory setting
Display Info
P
Function groupCHECK(only with functionextension)
Settings for sensor and processmonitoring.
P1
Switchpolarisationdetection on or off(conductive only)
offon
Polarisation only occurs withconductive sensors. Polarisationis detected but not compensated.(Error no.: E071.)
P2Set PCS alarm(live-check)
off1h2h4h
This function is used to examinethe measuring signal. An alarm issignalled if it does not change forthe period selected here.Monitoring limit: 0.3 % of meanvalue over period of timeselected.(Error no.: E152.)
P
CHECK
P1
Pol.Detec
off
P2
PCS alarm
off
Instrument configuration Liquisys M CLM 223 F
26 Endress+Hauser
5.5 Relay configuration
The relay contacts in the Liquisys MCLM 223 F can be configured as required.Switch-on and switch-off points and pickupand dropout delays can be defined for thelimit contactor. Moreover, an alarm thresholdcan be set to issue an error message.
Please refer to Fig. 5.5 for a graphicrepresentation of the contact states of anyrelay and of the alarm contact.
When the measured value increases (maxfunction), the relay contact is closed at time t2when the switch-on point has been exceeded(t1) and the pickup delay (t2 – t1) has expired.When the alarm threshold (t3) is reached andthe alarm delay (t4 – t3) also has expired, thealarm contact is switched.
When the measured value decreases, thealarm contact is reopened when themeasured value drops below the alarmthreshold (t5), and the relay contact alsoopens (t7, after the dropout delay t7 – t6).When the pickup and dropout delays are setto 0 s, the switch-on and switch-off points areidentical to the contact switching points.The picture below shows the switching statusfor a minimum contact.
t1
t1
t2
t2
t3
t3
t4
t4
t5
t5
t6
t6
t7
t7
t
t
Alarm threshold
Switch-on pointSwitch-off point
Meas. valueSwitch-on point > switch-off point: Maximum function
Switch-on point < switch-off point : Minimum function
Contact ON
Alarm ON
Alarm OFF
Contact OFF
Alarm threshold
Switch-on pointSwitch-off point
Contact ON
Alarm ONAlarm OFF
Contact OFF
GRENZ-E.CDR Fig. 5.5
Relationships betweenswitch-on and switch-offpoints and pickup anddropout delays
LM2XFE5B.CHP
Liquisys M CLM 223 F Instrument configuration
Endress+Hauser 27
Coding Field Selection or rangeFactory setting
Display Info
RFunction groupRELAY
Relay contacts can be selectedand adjusted.
R1Selection ofcontact to beconfigured
Rel1Rel2
R2
Switch limitfunction forselected contactoff or on
OffOn
R3Enter contactswitch-on point
cond: 2000 mS/cmind: 2000 mS/cmconc: 99.99 %entire meas. range
Never set switch-on point andswitch-off point to the same value!
Only the operating modeselected in A1 appears.
R4Enter contactswitch-off point
cond: 2000 mS/cmind: 2000 mS/cmconc: 99.99 %entire meas. range
The switch-off point entry selectsa max contact (switch-off point< switch-on point) or a mincontact (switch-off point> switch-on point), therebyimplementing an always requiredhysteresis function (see Fig. 5.5).
R5 Enter pickup delay0 s0 ... 2000 s
R6Enter dropoutdelay
0 s0 ... 2000 s
R7Enter alarmthreshold (as anabsolute value)
cond: 2000 mS/cmind: 2000 mS/cmconc: 99.99 %entire meas. range
When the alarm threshold isviolated, the transmitter issues analarm with an error message anderror current (note alarm delayand alarm allocation).If you define a min contact, thenyou must set the alarm thresholdto a less value than the switch-offpoint!
R
RELAY
Rel1 R1
Sel.Relay
Off R2
Function
2000mS/cm
R3
On value
2000mS/cm
R4
Off value
0s
R5
On delay
0s
R6
Off delay
2000mS/cm
R7
A.Thresh
Instrument configuration Liquisys M CLM 223 F
28 Endress+Hauser
5.6 Temperature compensation
The temperature compensation needs onlyto be performed in the conductivity mode(selection of mode in field A1, selection oftemperature compensation in field B2).
The temperature coefficient α specifies therelative change in conductivity per degree oftemperature change. It depends on thechemical composition of the medium and thetemperature itself.
In order to compensate for this dependence,three different compensation types can beselected in the Liquisys M CLM 223 F.
Linear compensation
The change between two temperature pointsis considered to be constant, i.e. α = const.The α value can be edited for the linearcompensation type. The referencetemperature is 25 °C.
NaCl compensation
The NaCl compensation (according toIEC 60746) is based on a fixed nonlinearcurve that defines the relationship betweenthe temperature coefficient and thetemperature. This curve is used for lowerconcentrations of approx. 0.1 ... 5 % NaCl.
Conductivity(uncompensated)
1
α
Temperature TKOMPLIN.CDR Fig. 5.7
Linear temperaturecompensation
Temperature [°C]
2.7
2.5
2.3
2.1
Tem
pera
ture
coef
ficie
ntα
[%/K
]
NACLKOMP.CDR
0 96 14448
Fig. 5.6 NaCl compensation
LM2XFE5C.CHP
Liquisys M CLM 223 F Instrument configuration
Endress+Hauser 29
Temperature compensation with table
In order to use the alpha-table function fortemperature compensation, the followingconductivity data of the process medium tobe measured are required:
Value pairs of temperature T and conductivity κfor T0 = 25 °C and for temperatures occurringin the process.
For the temperatures relevant in your process,you calculate the α values using the followingformula (not to calculate an alpha value for25 °C would neither make sense nor couldyou edit a table with this value!):
In the measuring instrument, the T/α valuepairs received in this way are edited into thetable, and measurement can be started.
Note:
You can determine the temperaturecoefficient α with the help of chapter
5.12, too.
T1 T0 T2 T
κ2
κ1
κ
κ( )T0
T1 T2 T
α1
α2
Required data Calculated valuesα
TAB01E.CDRFig. 5.8
Required data andcalculated α valuesfor temperaturecompensation with table
ακ
κ κ= ⋅
−−
≠100
0
0
00T
T T
T TT T
b g
b g b g; .
Instrument configuration Liquisys M CLM 223 F
30 Endress+Hauser
Coding Field Selection or rangeFactory setting
Display Info
TFunction groupALPHA TABLE
T1 Selection of table11 ... 4
Selection of table to be edited.
T2Selection of tableoption
readedit
T3Enter number oftable value pairs
11 ... 10
Up to 10 value pairs can beentered in the α table. These arenumbered from 1 ... 10 and canbe edited individually or insequence.
T4Selection of tablevalue pair
11 ... number of table
value pairsasign
The functional chain T4 ... T6 isautomatically passed as often asset in T3. “asign” is displayed asthe last step. After confirmation,the display jumps to T7.
T5Enter temperaturevalue (x value)
0.0 °C–35.0 ... 250.0 °C
The temperature values musthave a minimum distance of 1 K.Factory setting for x value ofvalue pairs in table:0.0 °C; 10.0 °C; 20.0 °C; 30.0 °C ...
T6Enter temperaturecoefficient α(y value)
2.10 %/K0.00 ... 20.00 %/K
T7Message whethertable status is okay
yesno
Back to T.
If no, then set table correctly (allprevious settings are retained) orback to measuring mode (thenthe table is invalid).
T
ALPHA TAB
T1
editCurve
1
T2
Sel.Table
read
T3
No.Elem.
1
T4
Sel.Elem.
1
T5
Temp.val.
0.0°C
T6
alpha val
2.10%/K
T7
Status ok
yes
LM2XFE5C.CHP
Liquisys M CLM 223 F Instrument configuration
Endress+Hauser 31
5.7 Concentration measurement
The Liquisys M CLM 223 F can convert fromconductivity values to concentration values.For that purpose, the operating mode is set toconcentration measurement (s. field A1).
Subsequently, the basic data for thecalculation of concentration must be enteredinto the measuring instrument. The datarequired for the most common substances arealready stored in your instrument. In field K1,you can select one of these substances.
It is also possible to determine theconcentration of a sample that is not stored inthe instrument. In this case you need theconductivity characteristics of the medium. Ifthese data are not available in data sheets,then you can easily determine thecharacteristics by yourself: Prepare samples
of the medium in concentrations occurring inthe process. Measure the uncompensatedconductivity of these samples at temperaturesalso occurring in the process.
Variable process temperature:If these variations shall be taken into accountfor concentration measurement, then theconductivity of each prepared sample mustbe measured for at least two differenttemperatures (minimum and maximumtemperature of the process).
Constant process temperature:Measure the differently concentrated samplesat this process temperature.
Finally you should have received measuringdata looking as follows:
The characteristics received from themeasuring points must rise or fall strictlymonotonically in the range of processconditions. That means they must not haveeither maxima or minima, or ranges ofconstant behaviour. Curve types like inFig. 5.11 are therefore impermissible.
Concentration C Concentration CCmin Cmin
T1
T1
T2
T2
T3
T3
Cmax Cmax
Con
duct
ivity
Con
duct
ivity
Meas. rangeMeas. range
Meas. point Meas. point
Variable process temperature:
KONZ1E.CDRFig. 5.9
Measuring data in case ofvariable temperature(qualitative representation)
Concentration C
T constant
Cmin Cmax Concentration C
T constant
Cmin Cmax
Con
duct
ivity
Meas. range
Con
duct
ivity
Meas. range
Constant process temperature:
KONZ2E.CDRFig. 5.10
Measured data in case ofconstant temperature(qualitative representation)
Con
duct
ivity
Concentration
Con
duct
ivity
Concentration
KONZ3E.CDRFig. 5.11 Impermissible curve types
Instrument configuration Liquisys M CLM 223 F
32 Endress+Hauser
Value entry
In the fields K6 to K8, the three parameters(value triplet with conductivity, temperatureand concentration) are now entered for eachmeasured sample.
Variable process temperature:At least two samples, i.e. two differentconcentrations, are required. For each ofthese samples, values for at least twotemperatures (minimum distance 0.5 °C) mustbe entered. The temperature values of sampleswith different concentrations must be equal ineach case. This results in a minimum of fourvalue triplets.
Constant process temperature:At least two value triplets must be entered inthe instrument. The temperature values of theentered triplets must be equal.
Note:
If the conductivity or temperaturevalues measured during operationare out of the measuring rangedefined in the concentration table(s. Fig. 5.9 and 5.10), then theaccuracy falls off essentially, and theinstrument generates an alarmmessage. Therefore the processlimits should be taken into accountfor the calculation of characteristics.
If in case of rising characteristic anadditional value triplet with 0 µS/cmand 0% is entered, then operationwith sufficient accuracy and withouterror message is possible from thestart of measuring range.
Coding Field Selection or rangeFactory setting
Display Info
KFunction groupCONCENTRATION
Four fixed and four editableconcentration fields are stored inthis function group.
K1
Selection ofconcentrationcurve to be usedto calculate thedisplay value
NaOH 0 ... 15%,H2SO4 0 ... 30%,H3PO4 0 ... 15%,HNO3 0 ... 25%User 1 ... 4
K2Selection ofcorrection factor
10.5 ... 1.5
If required, select a correctionfactor (only available for the usertable).
K3Selection of tableto be edited
11 ... 4
When editing a curve, anothercurve should be used tocalculate the current displayvalues (see K1).
K4Selection of tableoption
readedit
This selection applies to allconcentration curves.
K5Enter number oftriplets
14 ... 16
Each triplet consists of twonumeric values.
K
CONCENTRA
K1NaOH
akt.Curve
K21
Conc.Fact
K31
editCurve
K4read
Table
K5
No.Elem.
1
LM2XFE5D.CHP
Liquisys M CLM 223 F Instrument configuration
Endress+Hauser 33
5.8 Service
Coding Field Selection or rangeFactory setting
Display Info
K6 Selection of triplet11 ... number of
triplets in K4Any triplet can be edited.
K7
Enteruncompensatedconductivity valuefor K6
0.0 ... 9999 mS/cm
The functional chain K6 ... K9 isautomatically passed as often asset in K5. Then the display jumpsto K10.
K8Enterconcentrationvalue for K6
0.00 ... 99.99 % Measuring unit selected in A2.
K9Enter temperaturevalue for K6
–35.0 ... 250.0 °C
K10Display tablestatus after check
yesno
Back to K.
If no, then set table correctly (allprevious settings are retained) orback to measuring mode (thenpressing »+« and »–« at a time.The new data are dropped then).
Coding Feld Selection or rangeFactory setting
Display Info
SFunction groupSERVICE
S1 Selection oflanguage
ENG = EnglishGER = GermanFRA = FrenchITA = ItalianNEL = DutchESP = Spanish
This field must be configuredonce during start-up. Then youcan exit S1 and continue.
S2 HOLD effect froz. = last valuefix = fixed value
“froz.”:Display of last value beforeactivation of hold.“fix”:When hold is active, the fixedvalue entered in S3 is displayed.
K6
Sel.Elem.
1
K7
conduct.
0.0µS/cm
K8
concentr.
0.0%
K8
Temp.
0.0°C
K10
Status ok
yes
S
SERVICE
ENG S1
Language
froz. S2
Holdeffec
Instrument configuration Liquisys M CLM 223 F
34 Endress+Hauser
Coding Feld Selection or rangeFactory setting
Display Info
S3Enterfixed value forcurrent output
00 ... 100 %(of currentoutput value)
Only if S2 = fixed value
S4 Hold configuration
S+C =setup andcalibration
CAL =calibrationSetup =setupnone =no hold
S = setup = parametrizationC = calibration
S5 Manual holdOffOn
S6Enter hold delaytime
10 s0 ... 999 s
S7Enter SW upgraderelease code forFood package
00000000 ... 9999
The release code for Foodpackage is already entered byEndress+Hauser. You only haveto edit it when the central modulehas been changed.
S8Order number isdisplayed
When you upgrade theinstrument, the order code is notchanged. The delivery state isdisplayed.
S9Serial number isdisplayed
S10
Reset to factorysettings
noSens = sensor dataFacty= factory
settings
Facty= All data except for thelanguage (field S1) arecleared and reset to thefactory settings!
Sens = Sensor data arecleared (temp. offset,Airset value, cell constant,installation factor)
S3
Fixed Val
0
S+C S4
Auto HOLD
Off S5
Man. HOLD
10 S6
Cont.Time
s
0000 S7
FoodCode
order S8
CD0005
SerNo S9
xxxxxxxx
no S10
S.Default
LM2XFE5D.CHP
Liquisys M CLM 223 F Instrument configuration
Endress+Hauser 35
Calibration of inner resistance
If you use a conductive sensor in higherconductivities (> 45 mS/cm) the sensor’sinner resistance should be determined.
Otherwise a measuring error of more than 1 %could occur in some cases.
When performing the following menu fieldsyou determine the inner resistance and storeit in the measuring transmitter. Theconductivity values will be compensated withthis value.
By use of inductive sensors the calibration isnot necessary.
Procedure:
• Switch off power supply• Uninstall panel-mounted instrument resp.
open field instrument.• Connect the test resistance (included in
delivery) with terminals 83/84.• Perform the menu fields S12-S14.
Note:
• The cable resistance is notconsidered.
• Do not use another than the testresistance you obtained with yourdelivery.
Coding Feld Selection or rangeFactory setting
Display Info
S11Performinstrument test
noDispl = display test
S12
Calibration ofinner resistance ofconductivesensors
offon
Starting the calibration of theinner resistance of conductivesensors.
S13Status ofresistancedetermination
waito.k.E xxx
Countdown: 30s.If the status is not o.k., an errrorE xxx will be displayed.
no S11
Test
off S12Error
Ri-Calib.
wait S13
Status30s
yes S14
Store
Instrument configuration Liquisys M CLM 223 F
36 Endress+Hauser
5.9 E+H Service
5.10 Interfaces
Coding Field Selection or rangeFactory setting
Display Info
EFunction groupE+H SERVICE
E1Selection ofmodule
Contr = controller (1)Trans = transmitter (2)MainB = mainboard (3)Sens = sensor (4)
E111E121E131E141
Softwareversionis displayed
Cannot be edited.
E112E122E132E142
Hardwareversionis displayed
Cannot be edited.
E113E123E133E143
Serial number isdisplayed
Cannot be edited.
E114E124E134E144
Moduleidentificationis displayed
Cannot be edited.
Coding Field Selection or rangeFactory setting
Display Info
IFunction groupINTERFACE
I1 Enter address
AddressHART: 0 ... 15orPROFIBUS: 1 ... 126
For communication only.
E
E+H SERV
ContrE1
Select
xx.xx E111
SW-Vers.
xx.xx E112
HW-Vers.
SerNo E113
xxxxxxxx
LSG E114
Module ID
I
INTERFACE
126 I1
Address
LM2XFE5D.CHP
Liquisys M CLM 223 F Instrument configuration
Endress+Hauser 37
5.11 Calibration
This function group is used to calibrate thetransmitter. Two different types of calibrationare possible:
• Calibration by means of measurement of acalibration solution of a known conductivity.
• Calibration by entry of the exact cellconstant of the conductivity sensor.
Note:
• For conductive sensors, onlythe menu positions C121 toC126 are relevant.
• The instrument is automaticallyswitched to hold during calibration(factory setting).
Note:
• If the calibration procedure isaborted by pressing the PLUS andMINUS keys simultaneously(return to C114, C126 or C136) orif the calibration is faulty, then theprevious calibration data arereinstated. A calibration error isindicated by the “ERR” messageand flashing of the sensor symbolon the display. In this case, repeatcalibration!
• The end of calibration is followedby a return to the measuringmode. During the hold delay time(field S4), the hold symbol isdisplayed.
Coding Field Selection or rangeFactory setting
Display Info
CFunction groupCALIBRATION
C1 (1)
Calibration ofinductive sensorswith a ring-shaped opening
Airs = Airset (1)Cellc = cell constant (2)InstF = installationfactor (3)
The calibration of the inductivesensor is to be performed in air.The sensor must be dry.
Remove inductive sensor from the liquid and dry completely.
C111Residual couplingStart calibration(Airset)
current meas. value Start calibration with CAL.
C112Residual couplingis displayed(Airset)
–80.0 ... 80.0 µSResidual coupling of measuringsystem (sensor and transmitter).
C
CALIBRAT
AirS C1
Calibrat
0.0 C111
AirSet
µS/cm
5.3 C112
AirSetVal
µS/cm
Instrument configuration Liquisys M CLM 223 F
38 Endress+Hauser
Coding Field Selection or rangeFactory setting
Display Info
C113Calibration statusis displayed
o.k.E xxx
If the calibration status is notokay, then the second displayline shows an explanation of theerror.
C114Store calibrationresults?
yesnonew
If C113 = E xxx, then only no ornew.If new, then return to C.If yes/no, then return to“Measurement”.
C1 (2)Calibration of cellconstant
Airs = Airset (1)Cellc = cell constant (2)InstF = installationfactor (3)
Immerse sensor in calibration solution.
The sensor should be immersedat a sufficient distance a from thevessel wall (installation factor hasno influence if a > 15 mm).
C121Enter calibrationtemperature (MTC)
25.0 °C–35.0 ... 250.0 °C
Only exists if B1 = fixed.
C122 Enter α value ofcalibration solution
2.10 %/K0.00 ... 20.00 %/K
This value is specified with allE+H calibration solutions.
C123
Enter correctconductivity valueof calibrationsolution
current meas. value0.0 ... 9999 mS/cm
The practical range is dependson the sensor, i.e. the calibrationsolution should be at approx.40% of the measuring range ofthe sensor used. The display isalways in mS/cm.
C124Calculated cellconstant isdisplayed
0.1 ... 5.9 ... 9.99 cm–1 The calculated cell constant isdisplayed and entered in A5.
C125Calibration statusis displayed
o.k.E xxx
If the calibration status is notokay, then the second displayline shows an explanation of theerror.
o.k. C113
Status
yes C114
Store
Cellc C1
Calibrat
25.0 C121
ProcTemp.
°C
2.10%/K
C122
alpha val
10.30 C123
Real val.
mS/cm
5.900 C124
Cellconst
1/cm
o.k. C125
Status
LM2XFE5D.CHP
Liquisys M CLM 223 F Instrument configuration
Endress+Hauser 39
Coding Field Selection or rangeFactory setting
Display Info
C126Store calibrationresults?
yesnonew
If C125 = E xxx, then only no ornew.If new, then return to C.If yes/no, then return to“Measurement”.
C1 (3)
Calibration withsensor adaptationfor inductivesensors
Airs = Airset (1)Cellc = cell constant (2)InstF = installationfactor (3)
Calibration of inductive sensorwith compensation of wallinfluence.
The inductive sensor remains at the place of installation.
C131Enter calibrationtemperature (MTC)
25.0 °C–35.0 ... 250.0 °C
Only exists if B1 = fixed.
C132Enter α value ofthe mediummeasured
2.10 %/K0.00 ... 20.00 %/K
C133Enter correctconductivity valueof the medium
current meas. value0.0 ...9999 mS/cm
Determine the correctconductivity value via areference measurement.
C134Calculatedinstallation factoris displayed
10.10 ... 5.00
C135Calibration statusis displayed
o.k.E xxx
If the calibration status is notokay, then the second displayline shows an explanation of theerror.
C136Store calibrationresults?
yesnonew
If C135 = E xxx, then only no ornew.If new, then return to C.If yes/no, then return to“Measurement”.
yes C126
Store
InstF C1
Calibrat
25.0 C131
MTC temp.
°C
2.10 C132
alpha val
%/K
10.30 C133
Real val.
mS/cm
1 C134
InstFact
o.k. C135
Status
yes C136
Store
Instrument configuration Liquisys M CLM 223 F
40 Endress+Hauser
5.12 Determination of temperature coefficient
If you have enough conductivity data thetemperature coefficient can also bedetermined by using the formula on page 29.
Otherwise please follow the instructions below:
• Immerse the sensor in the medium(minimum distance from wall: 15 mm;concentration has to be in the process-typical range).
• Meter the uncompensated conductivity at25 °C and note down this value and theexact temperature.
• Then perform the menu »D« (see below).Important: use the medium andtemperature prevalent in your process.
• With the temperature and α values you canedit the alpha-table for temperaturecompensation (see menu fields T1-T7 onpage 30).
Note:
If you only want to use onetemperature coefficient, please referto field B3 on page 22.
Coding Field Selection or rangeFactory setting
Display Info
DFunction groupTEMPERATURECOEFFICIENT
D1Entry ofcompensatedconductivity
current value0 ... 9999
Edit the measured(uncompensated) conductivityvalue here you have determinedat 25 °C in the medium used inthe process.
D2Display ofuncompensatedconductivity
current value0 ... 9999
The uncompensated conductivityis displayed here (no edit).
D3Display of currenttemperature
current value–35 ... +250
The current tempeature isdisplayed here (no edit).Please note down this value.
D4 Display of α valuedetermined
Please note down this value.
D
Det.alpha
2000 D1
µS/cm
Cond.comp
2077 D2
µS/cm
Cond.unc.
60.0 D3
°C
Meas.temp
2.20 D4
%/K
alpha val
LM2XFE5D.CHP
Liquisys M CLM 223 F Instrument configuration
Endress+Hauser 41
5.13 Remote measuring range switching (MRS)
For the remote measuring range switchingfunction complete parameter sets can bedefined for up to 4 substances.
Individual settings for each parameter set:
• Operating mode (conductivity ortemperature)
• Temperature compensation• Current output (main parameter and
temperature)• Concentration table• Limit functions
Assignment of digital inputs
The Liquisys M CLM 223 F 223 transmitterhas 2 digital inputs. They can be defined infield M1 as follows:
Note:
For the measuring range 1(MR1)only the following settings canbe adjusted in the fields A1-A7 andB1-B6 (Setup 1 and 2) in the sameway.
Settings of 4 parameter sets
Example: CIP cleaning
Adjustment infield M1
Function of digital inputs
M1 = 0 MRS not active. The digital input 1 can be used for external hold.
M1 = 1 The digital input 2 can be used to switch 2 measuring ranges(parameter sets). The digital input 1 can be used for external hold.
M1 = 2 The digital inputs 1 and 2 can be used to switch 4 measuring ranges(parameter sets). This is the setting used in the following example.
Digital input 1 0 0 1 1
Digital input 2 0 1 0 1
Coding /operation
field
Parameter setMedium
1Beer
2Water
3Lye
4Acid
M4 Mode Conductivity Conductivity Concentration Concentration
M8, M9 Currentoutput
1 ... 3mS/cm
0.1 ... 0.8mS/cm
0 ... 10% 0 ... 5%
M6 Temp. comp. User tab.1 Linear — —
M5 Conc. tab. — — NaOH User tab.
M10, M11 Limits on: 2.3 mS/cmoff: 2.5 mS/cm
on: 0.7 µS/cmoff: 0.8 µS/cm
on: 2%off: 2.1%
on: 1.5%off: 1.6%
Instrument configuration Liquisys M CLM 223 F
42 Endress+Hauser
Coding Field Selection or rangeFactory setting
Display Info
M
Function groupREMOTE MEAS.RANGESWITCHING
M1Select function ofbinary inputs
10, 1, 2
0 = no MRS1 = 2 measuring range selectable
via digital input 2.Digital input 1 for hold.
2 = 4 meas. ranges selectablevia digital inputs 1+2
M2Selection / displayof activeparameter set
21 ... 4 if M1=0
Without MRS (M1 = 0):Selection of active parameter setWith MRS ( M1 = 1 or 2):Display of active parameter set
M3Selection ofparameter set tobe edited
11 ... 4 if M1 = 01 ... 2 if M1 = 11 ... 4 if M1 = 2
Selection of parameter set to beedited (independent of activeparameter acc. to M2)
M4Selection ofoperating mode
cond = conductivityconc = concentration
The operating mode can beindividually selected for eachparameter set.
M5Selection ofmedium
NaOHH2SO4H3PO4HNO3User1 ... 4
Only available if M4 = conc
M6Selection oftemperaturecompensation
linNaClTab 1 ... 4none
Only available if M4 = cond
M7 Entry of α value2.1 %/K0 ... 20 %/K Can only be entered if M6 = lin
M8
0/4 mA value:Entry ofcorrespondingmeasured value
cond: 0 ... 2000 mS/cmconc: 0 ... 99.99 %
For concentration:edit Unit in field A2edit Format in field A3
M
MRS
1 M1
Bin.Input
2 M2
Act. MR
1 M3
Edit MR
cond. M4
Oper.Mode
NaOH M5
Conc.Tab.
lin M6
TempComp
2.10 M7
%/K
Alpha val
0 M8
µS/cm
0/4 mA
LM2XFE5D.CHP
Liquisys M CLM 223 F Instrument configuration
Endress+Hauser 43
Coding Field Selection or rangeFactory setting
Display Info
M9Entry of measuredvalue for 20 mAvalue
cond: 0 ... 2000 mS/cmconc: 0 ... 99.99 %
For concentration:edit Unit in field A2edit Format in field A3
M10Selection ofcontact to beconfigured
Rel1Rel2
M11Switch function ofrelais of or on
OffOn
All settings are retained.
M12Entry of switch-onpoint for limit
cond: 0 ... 2000 mS/cmconc: 0 ... 99.99 %
For concentration:edit Unit in field A2edit Format in field A3
M13 Entry of switch-offpoint for limit
cond: 0 ... 2000 mS/cmconc: 0 ... 99.99 %
For concentration:edit Unit in field A2edit Format in field A3.With editing the switch-off pointeither a max contact (switch-offpoint < switch-on point) or a mincontact (switch-off point >switch-on point) is selected anda always necessary hysteresisrealized (see fig. 5.5)
M14 Enter pickup delay0 s0 ... 2000 s
M15Enter dropoutdelay
0 s0 ... 2000 s
2000 M9
mS/cm
20 mA
Rel1 M10
Sel.Relay
Off M11
Function
2000 M12
mS/cm
PV on
2000 M13
mS/cm
PV off
0s
M14
On Delay
0s
M15
Off Delay
Instrument configuration Liquisys M CLM 223 F
44 Endress+Hauser
6 Interfaces
For instruments equipped with acommunication interface, please refer toseparate operating instructionsBA 212C/07/en (HART®) orBA 213C/07/en (PROFIBUS®).
LM2XFE06.CHP
Liquisys M CLM 223 F Interfaces
Endress+Hauser 45
7 Maintenance and troubleshooting
7.1 Definition of terms
Maintenance means that all measures whichwill guarantee the safety of operation andreliability of the entire measuring system aretaken in due time.
Maintenance on the CLM 223 F includes:
• Calibration (see chpt. 5.11)• Cleaning of assembly and sensor• Checking of cables and connections.
Troubleshooting means that the cause of theproblem is determined and eliminated.Troubleshooting refers to measures that canbe performed without intervention in theinstrument (refer to chpt. 8, Correctivemaintenance, for defects of the instrument).
Troubleshooting of the CLM 223 F and themeasuring system is performed with the aid ofthe troubleshooting tables in chpt. 7.3 and 7.4.
7.2 Safety instructions
Warning:
Please be aware of effects work per-formed on the instrument might haveon the process control system or theprocess itself.
Warning:
When removing the sensor duringmaintenance or calibration, pleaseconsider potential hazards due topressure, high temperatures andcontamination.
7.3 Troubleshooting common problems
Problem Possible cause Remedy Equipment needed, spare parts
Instrument is notworking / display: 9999
– Instrument is locked Unlock instrument see chapter 4.3
Display deviates fromreference measurement
– Calibration faulty Calibrate instr. acc. to chpt. 5.11 Cal. solution or sensor certificate
– Sensor soiled Clean sensor see chapter 8.7.1
– Incorrect temperaturemeasurement
Check temperature value oninstrument and reference unit
Temp. meas. instr., precisionthermometer
– Incorrect temp. compensation Check compensation mode (none /ATC / MTC) and compensationtype (linear/subst./user table)
Please note: transmitter hasseparate calibration and operatingtemperature coefficients.
– Reference instrument calibrationfaulty
Calibrate reference instrument oruse calibrated instrument.
Calibration solution, operatinginstructions of reference instrument
– Incorrect ATC setting onreference instrument
Compensation mode andcompensation type must beidentical on both instruments.
Oper. instr. of reference instrument
Implausible measuredvalues in general:– continual meas. value
overflow– meas. value always
000– meas. value too low– meas. value too high– meas. value frozen– incorrect current
output value
– Short / moisture in sensor Check sensor see chapter 8.7.3
– Short in cable or junction box Check cable and junction box see chapter 8.7.4
– Interruption in sensor Check sensor see chapter 8.7.3
– Interrupt. in cable or junction box Check cable and junction box see chapter 8.7.4
– Incorrect cell constant setting Check cell constant Sensor nameplate or certificate
– Incorrect output assignments Check assignments of measuredvalue to current signal
– Incorrect output function Check 0-20 / 4 -20 mA selectionand curve shape (linear / table)
– Air cushion in assembly Check assembly and installation
– Incorrect temp. measurement/temperature sensor defective
Check instrument with equivalentresistance/ check Pt 100 in sensor
Pt 100 simulation s. chapter 8.7.2 /Pt 100 test s. chapter 8.7.3
– Transmitter module defective Test with new module Diagn. and spare parts s. chpt. 8
– Impermissible instrumentoperating state (no response tokey actuation)
Switch instrument off and back on EMC problem: check groundingand line routing if problem persistsor call E+H Service to test.
Maintenance and troubleshooting Liquisys M CLM 223 F
46 Endress+Hauser
Problem Possible cause Remedy Equipment needed, spare parts
Incorrect temperaturevalue
– Incorrect sensor connection Verify connections usingconnection diagram; three-wireconnection mandatory
Connection diagram in chpt. 3.4
– Measuring cable defective Check cable for interruption/short/shunt
Ohmmeter; also see chpt.7.2 / 7.3
– Incorrect sensor type Select sensor type on instrument(field B1)
Incorrect conductivitymeas. value in process
– No / incorrect temperaturecompensation
ATC: select compensation type;linear: set correct
coefficient.MTC: set process temperature
– Incorrect temperaturemeasurement
Check temperature value Reference instrument, thermometer
– Bubbles in medium Suppress bubble formation:– gas bubble trap– counterpressure (cover)– bypass measurement
– Incorrect sensor orientation Centre hole in sensor must point inmedium flow direction
Turn sensor in flange.
– Flow rate too high (may causebubbles)
Reduce flow or chooselow-turbulence mounting position
– Interference current in medium Ground medium close to sensor;remove/repair source
Most frequent cause of currents inmedium: defective submergedmotors
– Sensor soiled or coated Clean sensor (s. chpt. 8.7.1 ) Heavily soiled media: use spraycleaning
Measured valuefluctuates
– Measuring cable interference Connect cable screen acc. toconnection diagram
s. chpt. 3.5
– Signal output line interference Check line routing, try separate linerouting
Separate routing of signal outputand meas. input lines
– Interference currents in medium Eliminate source of interference orground medium close to sensor
– Contact switched-off Activate limit contactor s. field R2
Limit contact does notwork
– Pickup delay setting too long Shorten pickup delay s. field R5
– “Hold” function active “Autom. hold” during calibration,“hold” input activated;“hold” via keyboard active
s. fields S2 to S5
Limit contact workscontinuously
– Dropout delay setting too long Shorten dropout delay s. field R6
– Control loop interruption Check measured value, currentoutput, actuators, chemical supply
No conductivity currentoutput signal
– Line open or short-circuited Disconnect line and measuredirectly on instrument
mA meter 0–20 mA
– Output defective s. Diagnosis in chpt. 8.3
Fixed current outputsignal
– Current simulation active Switch off simulation s. field O22
– HOLD active Control HOLD configuration
– Impermissible processor systemoperating state
Switch instrument off and back on EMC problem: check installation,screen, grounding if problempersists/ call E+H Service to test.
Incorrect current outputsignal
– Incorrect current assignment Check current assignment:0–20 mA or 4–20 mA?
Field O211 / 212
– Total load in current loopexcessive (> 500 Ω)
Disconnect output and measuredirectly on instrument
Ohmmeter
– EMC (interference coupling) Disconnect both output lines andmeasure directly on instrument
Use shielded lines, ground screenson both sides, route line in otherduct if necessary.
LM2XFE07.CHP
Liquisys M CLM 223 F Maintenance and troubleshooting
Endress+Hauser 47
Problem Possible cause Remedy Equipment needed, spare parts
No temperature outputsignal
– Instrument does not have 2ndcurrent output
Refer to nameplate for variant;change LSCH-x1 module if nec.
Module LSCH-x2,s. chpt. 8.4.4
– HOLD active Control HOLD configuration
– Instrument withPROFIBUS-PA/-DP
PA/DP instruments have no currentoutput!
No HARTcommunication
– No central HART module Verify by looking at nameplate:HART = -xxx5xx and -xxx6xx
Upgrade to LSCH-H1 / -H2
– Current output < 4 mA
– No or wrong DD(device description)
– HART interface missing
– Instrument not registered withHART server
– Load too low (load > 230 Ωrequired)
For further informationsee BA 212C/07/en,
“HART – Field communication”– HART receiver (e.g. FXA 191) not
connected via load
– Incorrect device address(addr. = 0 for single operation,addr. > 0 for multi-drop operation)
– Line capacitance too high
– Line interference
– Several devices set to sameaddress
Set addresses correctly Communication not possible withseveral devices set to same address.
– No PA/DP module Verify by looking at nameplate:PA = -xxx3xx, DP = -xxx4xx
Upgrade to LSCP module
No PROFIBUScommunication
– Wrong SW version(without PROFIBUS)
– Commuwin (CW) II:CW II version and instrumentSW version incompatible
– No or wrong DD/DLL
– Incorrect baud rate setting forsegment coupler in DPV-1 server
For further informationsee BA 213C/07/en
“PROFIBUS-PA/-DP –Field communication”
– Incorrect station (master)address or duplicate address
– Incorrect station (slave) address
– Bus line not terminated
– Line problems(too long, cross section toosmall; not shielded, screen notgrounded, wires not twisted)
Use cables recommended by E+H
– Bus voltage too low (bus supplyvoltage typ. 24 V DC for non-Ex,13.5 V DC for Ex)
Voltage at instrument PA/DPconnector must be at least 9 V.
– Several devices set to sameaddress
Set addresses correctly Communication not possible withseveral devices set to same address.
Maintenance and troubleshooting Liquisys M CLM 223 F
48 Endress+Hauser
7.4 Problem elimination based on error messages
Error messages
The Liquisys M CLM 223 F transmittercontinually monitors its own functions. If theinstrument detects a defect, the error numberappears on the display.
This error number is displayed underneaththe main value unit display. If several errorsare detected, these can be called up with theMINUS key.
Errorno. Cause Measures
Contact Error current
Fact. User Fact. User
E001 EEPROM memory error 1. Switch instrument off and back on2. Set to factory settings (S11)3. Load hardware-compatible software
(with Optoscope, s. chpt. 8.6)4. If problem persists, return instrument
to your local Endress+Hauser salesagency for repair or replaceinstrument.
On E003: invalid configuration, download attempt wasperformed with invalid parameter set.
On E004: wrong software version has been loaded orunsuitable module has been exchanged
On E007: at inductive version sensor is connected in thewrong way
yes no
E002
Instrument not calibrated,calibration data invalid,no user data or user datainvalid(EEPROM error)Software does not matchhardware (central module)
yes no
E003 Download error yes no
E004
Instrument softwareincompatible withhardware version ofmodule
yes no
E007Transmitter malfunctionSoftware does not matchtransmitter version
yes no
E008 Sensor or sensorconnection faulty
Check sensor and sensor connection (s. chpt. 8.7.2 or callE+H Service).
yes no
E010
No temperature sensorconnected or temperaturesensor short-circuited(temperature sensorfaulty)
Check temperature sensor and connections; if necessary,check instrument with temperature simulator.
yes no
E025 Limit for Airset offsetexceeded
Repeat Airset (in air) or replace sensor. Clean and dry cellbefore Airset (only with inductive)
yes no
E036 Sensor calibration rangeexceeded
Clean and recalibrate sensor; if necessary, check measuringcell and connections.
yes no
E037 Below calibration range ofsensor
yes no
E040 Test resistance rangeexceeded.
Check, if test resistance is installed and all right. yes no
E045 Calibration aborted Recalibrate. yes no
E049Calibration range ofinstallation factorexceeded
Check pipe diameter, clean sensor and repeat calibration(only with inductive)
yes no
E050 Below calibration range ofinstallation factor
Check pipe diameter, clean sensor and repeat calibration(only with inductive)
yes no
E055 Below measuring rangeof main parameter
Immerse sensor in conductive medium (inductive: performAirset)
yes no
E057 Measuring range of mainparameter exceeded
Check measurement, control and connections (simulations. chpt. 8.7.2).
yes no
E059 Below temperaturemeasuring range
yes no
E061 Temperature measuringrange exceeded
yes no
LM2XFE07.CHP
Liquisys M CLM 223 F Maintenance and troubleshooting
Endress+Hauser 49
Errorno. Cause Measures
Contact Error current
Fact. User Fact. User
E063 Below current outputrange 1
Check measured value and current output assignment(function group O).
yes no
E064 Current output range 1exceeded
Check measured value and current output assignment(function group O).
yes no
E065 Below current outputrange 2
Check measured value and current output assignment. yes no
E066 Current output range 2exceeded
yes no
E067 Alarm threshold for limitcontactor exceeded
Check measured value, limit setting, alarm threshold andmetering devices.
yes no
E071 Polarisation Clean Sensor; use higher cell constant yes no
E077 Temperature outsideα value table range
Check measurement and tables. yes no
E078 Temperature outsideconcentration table
yes no
E079 Conductivity outsideconcentration table
yes no
E080 Current output 1 param.range too small
Spread current output. no no
E081 Current output 2 param.range too small
Spread current output. no no
E100 Current simulation active no no
E101 Service function yes Switch service function off orswitch instrument off and back on.
no no
E102 Manual mode active no no
E106 Download yes Wait for download to end. no no
E116 Download error Repeat download. no no
E150Distance between temp.values in α value table toosmall
Enter correct values in α value table (minimum distancebetween temperature values of 1 K required).
no no
E152Live Check alarm Check sensor and connection.
Check processor (flow).Check sensor installation on air bubbles
no no
Maintenance and troubleshooting Liquisys M CLM 223 F
50 Endress+Hauser
8 Diagnosis and corrective maintenance
8.1 Definition of terms
Diagnosis refers to the identification ofinstrument malfunctions and defects.
Corrective maintenance means• replacement of parts diagnosed to be
defective• testing of instrument and measuring
system function• restoration of complete functionality.
Diagnosis based on the error table below anddepending on difficulty and measuringequipment at hand is to be performed by:• trained operator personnel• operator’s electricians• company responsible for system
installation/ operation• E+H ServicePlease refer to the tables in chapters 8.4 and8.5 for identification of spare parts required.
8.2 Safety instructions
Warning:
• Disconnect the instrument fromthe power source before openingit up.Work under tension may only beperformed by trained electricians.
• Switched contacts may besupplied from separate circuits.These circuits must also bede-energized before work on theterminals is performed.
Caution: ESD!
• Electronic components aresensitive to electrostaticdischarges. Personal protectivemeasures, such as discharge viaPE or permanent grounding usinga wrist strap, are to be taken.
• For your own safety, use onlyoriginal spare parts. Original partswill guarantee functionality,accuracy and reliability afterrepairs.
8.3 Diagnosis
The table below will help you diagnoseproblems and specifies the spare partsrequired.
Please refer to chapters 8.4.3 and 8.5.3 forinformation on the exact designations of thespare parts and their installation.
Problem Possible cause Tests and/or remedial measures Equipment, spare parts, personnel
Display dark, no LEDsactive
– No mains voltage Check if mains voltage is available Electrician / e.g. multimeter
– Wrong supply voltage / voltagetoo low
Compare mains voltage and ratingon nameplate
Operator (utility companyspecification or multimeter)
– Connection fault Terminal not tightened;insulation clamped in terminal;wrong terminal used
Electrician
– Fuse blown Replace fuse,first compare mains voltage andrating on nameplate
Electrician /see drawings in chapters 8.4.1 and8.5.1 for correct fuses
– Power supply unit defective Replace power supply unitusing correct variant
On-site diagnosis by E+H Service(substitute power unit required)
– Central module defective Replace central moduleusing correct variant
On-site diagnosis by E+H Service(substitute central module required)
Display dark, LED(s)active
– Central module defective(module: LSCH/LSCP)
Replace central module On-site diagnosis by E+H Service(substitute central module required)
LM2XFE08.CHP
Liquisys M CLM 223 F Diagnosis and corrective maintenance
Endress+Hauser 51
Problem Possible cause Tests and/or remedial measures Equipment, spare parts, personnel
Display showsmeasured value but– value does not change
and / or– instrument cannot be
operated
– Instrument, or module notproperly installed in instrument
Reinstall module Refer to assembly drawings inchapters 8.4.1 and 8.5.1.
– Impermissible operating systemstate
Switch instrument off and back on Possible EMC problem: if problempersists, call E+H Service to haveinstallation checked.
Instrument gets hot – Incorrect voltage / too high Compare mains voltage and ratingon nameplate
– Power supply unit defective Replace power supply unit Can only be diagnosed by E+HService
Incorrect meas. Cdand/or temperaturevalue
– Transmitter module defective(module: MKIC),please perform tests and takemeasures acc. to chapter 7.3
Test measuring inputs:– connect resistor in place of
conductivity sensor,see table in chpt. 8.7.2
– Connect 100 Ω resistorto terminals 11 / 12 + 13= display 0 °C
If test fails: replace module (usingcorrect variant), refer to explodedviews in chapters 8.4.1 and 8.5.1.
Current output,incorrect current
– Not calibrated correctly Test with built-in current simulation,connecting mA meter directly tocurrent output
If simulation value is incorrect:recalibration at factory or newLSCxx module required.
If simulation value is correct:check current loop for load andshunts.
– Load excessive
– Shunt / short-circuit to frame incurrent loop
Disconnect current output, connectmA-meter directly to the currentoutput
– Incorrect mode of operation Check whether 0–20 mA or4–20 mA has been selected
No current output signal – Current output stage defective(module: LSCH)
Test with built-in current simulation,connecting mA meter directly tocurrent output
If test fails:Replace central module(using correct variant)
– Wrong module The Profibus-PA or -DP modules donot have current outputs.
No HART orPROFIBUS-PA/-DPinterface function
– Wrong central module HART: LSCH-H1 or -H2PROFIBUS-PA/-DP: LSCP module,s. field E112
Replace central module;operator or E+H Service
– Wrong software Instrument SW version: s. field E111
– Bus problem Remove some instruments and testagain
Consult E+H Service
Diagnosis and corrective maintenance Liquisys M CLM 223 F
52 Endress+Hauser
8.4 Corrective maintenance of Liquisys M CLM 223 F
8.4.1 Exploded view
230V
100V
115V
Item 420aTerminal for signalinputs/outputs (external)
Item 420bTerminal for Pt 100,alarm relay, supply (internal)
Item 240Conductivity transmitter MKIC
Item 400Specialscrew
Item 340End frame
223_EXLF.CDR
Item 15 / 20Power supply unit LSGA / LSGD
End framelatches
Item 320Back plate
Item 40 ... 91Central moduleLSCH / LSCP
Item 330Housingassembly
Item 430Push-on terminal stripfor LSR1-2 module
Item 290 / 300Optional relaymodule LSR1-2
Module rotated180° forrepresentation!
Item 345Groundterminal
Item 15Power supply unit LSGA
100 - 230 V AC
Mains voltageselection
FuseJumper
LM2XFE08.CHP
Liquisys M CLM 223 F Diagnosis and corrective maintenance
Endress+Hauser 53
8.4.2 Dismantling of CLM 223 F
• Consider potential effects on processwhen removing the instrument from service!
• First pull off the terminal block (item 420b)on the rear of the instrument to de-energizethe instrument.
• Then pull off the terminal blocks (items420a and 430 if applicable) on the rear.Now you can dismantle the instrument.
• Push the end frame latches (item 340)inward and pull off the frame towards therear.
• Loosen the special screw (item 400) byturning it counterclockwise.
• Remove the complete electronics blockfrom the housing. The modules areplugged together mechanically and can beeasily separated:– Simply pull the processor/display module
off towards the front– Pull the back plate tabs outward slightly
to remove the lateral modules.
8.4.3 Assembly of CLM 223 F
• Reverse the dismantling sequence forassembly.
• Hand-tighten the special screw withoutusing a tool.
• Incorrect assembly is not possible!A module block which has been pluggedtogether incorrectly cannot be inserted inthe housing.
Diagnosis and corrective maintenance Liquisys M CLM 223 F
54 Endress+Hauser
8.4.4 Spare parts for CLM 223 F
Item Designation Name Function Order number
15 Power supply unit LSGA 100/115/230 V AC 51500317
20 Power supply unit LSGD 24 V AC + DC 51500318
290 Relay module LSR1-2 2 relays 51500320
40 Conductive central module LSCH-S1 1 current output 51506379
50 Conductive central module LSCH-S2 2 current outputs 51506380
60 Conductive central module LSCH-H1 1 current output + HART 51506381
70 Conductive central module LSCH-H2 2 current outputs + HART 51506382
80 Conductive central module LSCP-PA PROFIBUS-PA /no current output! 51506383
90 Conductive central module LSCP-DP PROFIBUS-DP /no current output! 51506384
41 Inductive central module LSCH-S1 1 current output 51506385
51 Inductive central module LSCH-S2 2 current outputs 51506386
61 Inductive central module LSCH-H1 1 current output + HART 51506387
71 Inductive central module LSCH-H2 2 current outputs + HART 51506389
81 Inductive central module LSCP-PA PROFIBUS-PA /no current output! 51506391
91 Inductive central module LSCP-DP PROFIBUS-DP /no current output! 51506392
240 Conductivity transmitter MKIC Cond. + temp. input 51501206
340 End frame PROFIBUS-DP End frame with D-Subminconnector 51502513
330, 340 Housing assembly
With front membrane, keytappets, gasket, special
screw, tensioning brackets,all nameplates
51501075
310, 320,340, 400 Mechanical housing parts Back plate, side wall, end
frame, special screw 51501076
420a, 420b Terminal strip set, complete,standard + HART
Terminal strip setstandard + Hart,
inputs/outputs, supply,alarm relay
51501203
420a, 420b Terminal strip set, complete,PROFIBUS-PA
Terminal strip setPROFIBUS-PA,
inputs/outputs, supply,alarm relay
51502126
420a, 420b Terminal strip set, complete,PROFIBUS-DP
Terminal strip setPROFIBUS-DP,
inputs/outputs, supply,alarm relay
51502493
430 Terminal strip Terminal strip for relay module 51501078
345 Ground terminal strip PE and screen connections 51501086
LM2XFE08.CHP
Liquisys M CLM 223 F Diagnosis and corrective maintenance
Endress+Hauser 55
8.4.5 Special case: replacement of central module
Note:
All the editable data are reset to thefactory settings following centralmodule replacement.
Proceed as described below after centralmodule replacement:• If possible, record the user settings of the
instrument, e.g.:– Calibration data– Conductivity and temperature current
assignments– Relay function selections– Limit settings– Monitoring functions– Interface parameters
• Dismantle the instrument as described inchpt. 8.4.2.
• Refer to the part no. of the central moduleto determine whether the new module hasthe same part no. as the old one.
• Assemble the instrument with the newmodule as described in chapter 8.4.3.
• Start up the instrument and test its basicfunctions (e.g. meas. value andtemperature display, operation viakeyboard).
• Enter the instrument serial number:– Read the instrument serial number
(“ser-no.”) on the nameplate.– Enter this number in the fields E116 (year,
one digit), E117 (month, one digit), E118(sequence number, four digits).
– Field E119 displays the complete numberfor verification; acknowledge with ENTERor abort and re-enter.
Please note: The serial number can onlybe entered – and only once – in the caseof a brand-new module from the factorywith serial number! Make sure that yourentry is correct before confirming withENTER!
• Restore the user settings of the instrument.
8.5 Spare parts orders
Spare parts are to be ordered from your localE+H Sales Agency. See the back cover ofthese operating instructions for addresses.Specify the order numbers listed in chapter8.4.4.
To be on the safe side, you should alwaysspecify the following data with your spareparts orders:
• Instrument order code (order code)• Serial number (ser-no.)• Software version where available
Refer to the nameplate for the order code andserial number.The software version is displayed in fieldE111 when the instrument processor systemis functional.
8.6 “Optoscope” service equipment with “Scopeware”
The Optoscope in conjunction with the“Scopeware” software opens the followingpossibilities without removing or opening theLiquisys M CLM 223 F and without makingan electrical connection to the instrument:• Documentation of the instrument settings in
conjunction with Commuwin II• Software update by the service technician• Upload/download a hex dump to duplicate
configurations.
The Optoscope acts as an interface betweenthe Liquisys M CLM 223 F and PC / Laptop.Information is transferred via the opticalinterface on the Liquisys M CLM 223 F andvia the RS 232 interface on the PC / Laptop.
Handling and operation are described in theOptoscope operating instructions. Therequired Windows “Scopeware” softwarecomes in the scope of supply.
The Optoscope is also designed for theMycom CxM 152 and MyPro CxM 431. It issupplied with all the required accessories ina sturdy case.
Order number of the Optoscope: 51500650
Diagnosis and corrective maintenance Liquisys M CLM 223 F
56 Endress+Hauser
8.7 Corrective maintenance of measuring system
8.7.1 Conductivity sensors
Soiling detected on conductivity sensorscan be removed as follows:
• Oily and greasy coatingsClean with detergent (fat solvent, e.g.alcohol, acetone, poss. detergent).
Warning:
Hands, eyes and clothes are to beprotected when using the cleaningagents described below.
• Limestone deposits or metal hydroxidecoatingsLoosen coatings with diluted hydrochloricacid (3%), then rinse thoroughly.
• Coatings containing sulphide(from flue gas desulphurizing plantsor sewage treatment plants)Use mixture of hydrochloric acid (0.5%)and thiourea (8%), then rinse thoroughly.
• Coatings containing protein (food industry)Use mixture of hydrochloric acid (0.5%)and pepsin (1%), then rinse thoroughly.
8.7.2 Simulation of conductive sensors for instrument test
A measuring instrument for conductiveconductivity is checked by replacing theprocess and temperature sensor withresistors. The accuracy of the simulationdepends on the accuracy of the resistors.
Refer to the table on the right for conductivity(cell constant k set to nominal valueaccording to column 2). Otherwise: display ofCD[mS/cm] = k ⋅ 1/R[kΩ].
Refer to the table below for temperature (notemperature offset set on Liquisys MCLM 223 F).Connect temperature equivalent resistanceusing three-wire arrangement!
• Use the “CD test adapter” service kit toconnect decade resistors to the connectorof a conductivity measuring cable in placeof a conductivity sensor.Order number: 51500629.
Pt 100 equivalent resistances:
Temperature Resistance
–20 °C 92.13 Ω
–10 °C 96.07 Ω
0 °C 100.00 Ω
10 °C 103.90 Ω
20 °C 107.79 Ω
25 °C 109.73 Ω
50 °C 119.40 Ω
80 °C 130.89 Ω
100 °C 138.50 Ω
200 °C 175.84 Ω
All resistance values are to be multiplied by10 for a Pt 1000 temperature sensor.
Conductivity equivalent resistances:
Resist. R Cell const. k Display forCD
10 Ω1 cm–1 100 mS/cm
10 cm–1 1000 mS/cm
100 Ω0.1 cm–1 1 mS/cm
1 cm–1 10 mS/cm
10 cm–1 100 mS/cm
1000 Ω0.1 cm–1 0.1 mS/cm
1 cm–1 1 mS/cm
10 cm–1 10 mS/cm
10 kΩ
0.01 cm–1 1 µS/cm
0.1 cm–1 10 µS/cm
1 cm–1 100 µS/cm
10 cm–1 1 mS/cm
100 kΩ0.01 cm–1 0.1 mS/cm
0.1 cm–1 1 µS/cm
1 cm-1 10 µS/cm
1 MΩ0.01 cm–1 0.01 µS/cm
0.1 cm–1 0.1 µS/cm
1 cm–1 1 µS/cm
10 MΩ0.01 cm–1 0.001µS/cm
0.1 cm–1 0.01 µS/cm
LM2XFE08.CHP
Liquisys M CLM 223 F Diagnosis and corrective maintenance
Endress+Hauser 57
8.7.3 Simulation of inductive sensors for instrument test
An inductive sensor cannot be simulated byresistors.
However, the overall system comprising theCLM 223-IF and inductive sensor can bechecked using equivalent resistances. Notecell constant k (e.g. knominal = 2 forCLS 50, knominal = 5.9 for CLS 52).
For an accurate simulation, the actual cellconstant (can be read in field C124) is to beused to calculate the display value:display of CD[mS/cm] = k ⋅ 1/R[kΩ].
Reference values for simulation of CLS 50 at25 °C:
Simulationresistance R
Cellconstant k
Conductivitydisplay
6.8 Ω 5.90 cm–1 868 mS/cm
33 Ω 5.90 cm–1 178.8 mS/cm
330 Ω 5.90 cm–1 17.88 mS/cm
3.3 kΩ 5.90 cm–1 1.788 mS/cm
• Performance of simulation:Pull a cable through the sensor openingand then connect, e.g. to a decade resistor.
8.7.4 Checking of conductive conductivity sensors
• Measuring surface connection:The measuring surfaces are connecteddirectly to the connections of the sensorconnector.Check with ohmmeter for < 1 Ω
• Measuring surface shunt:A shunt between the measuring surfaces isnot allowed.Check with ohmmeter for > 20 MΩ
• Temperature sensor shunt:A shunt between the measuring surfacesand temperature sensor is not allowed.Check with ohmmeter for > 20 MΩ.
• Temperature sensor:Refer to the sensor nameplate for thetemperature sensor type.The sensor can be checked at the sensorconnector with an ohmmeter:– Pt 100 at 25 °C = 109.79 Ω– Pt 1000 at 25 °C = 1097.9 Ω– NTC 10 k at 25 °C = 10 kΩ
• Connection:Check terminal wiring (interchanged?) andtightness of clamping screws in the case ofsensors with screw terminals (CLS 12/13).
8.7.5 Checking of inductive conductivity sensors
The following specifications apply to sensorsCLS 50 and CLS 52:
• Testing of transmitting and receiving coils(white and red coaxial cables, measure-ment between int. conductor and screen):– ohmic resistance approx. 0.5 ... 2 Ω– induct. approx. 260 ... 450 mH (at 2 kHz)
• Testing of coil shunt:A shunt between the coils (from coax redto coax white) is not allowed.Check with ohmmeter for > 20 MΩ
• Testing of temperature sensor:Use the table in chapter 8.7.2 to check thePt 100.The resistance values between the greenand white wires and between the greenand yellow wires should be identical.
• Testing of temperature sensor shunt:Shunts between the temperature sensor(green, white or yellow) and the coils (redor white coaxial cable) are not allowed.Check with ohmmeter for > 20 MΩ.
8.7.6 Checking of line extension and junction box
• Use the methods described in chapter8.8.2 or 8.8.3 to perform a quick functionalcheck from the sensor connector(conductive sensors) or from the sensor(inductive sensor) to the measuringinstrument.Decade resistors are best connected withthe “CD test adapter” service kit (ordernumber: 51500629).
• Check junction boxes for:– Moisture (influence in low conductivity
range; dry junction box, replace gaskets,use desiccant bag)
– Correct line connections– Outer screen connections– Tightness of clamping screws
Diagnosis and corrective maintenance Liquisys M CLM 223 F
58 Endress+Hauser
9 Accessories
Connection accessories
• Junction box VBMJunction box for extension of measuringcable connection between sensor andinstrument.Material: cast aluminiumIngress protection: IP 65Order no.: 50003987
• Extension cable CYK 71Non-terminated measuring cable fortwo-electrode conductivity sensors withtemperature sensor (by the metre)Order no.: 50085333
• Extension cable CLK 5Non-terminated measuring cable forinductive sensors (by the metre)Order no.: 50085473
Software upgrade(serial number of instrument to be specifiedwith order)
Two-relay cardOrder no.: 51500320
Note:
The desiccant bag in the junctionbox must be checked and replacedat regular intervals which depend onambient conditions in order toprevent inaccurate measurementdue to moisture bridges in themeasuring line.
•
VBM.CDR
125
113
62.5
5280
10approx.
136
26.8
55
Fig. 9.1Dimensions ofjunction box VBM
LM2XFE09.CHP
Liquisys M CLM 223 F Accessories
Endress+Hauser 59
Conductivity sensors
Please refer to the corresponding TIs fordetailed technical data.
Conductive
k = 0,1
0,01 0,1
µS/cm
1,0 10 100 1 10
mS/cm
100 10002000
10000
k = 0,01
k = 1
CLS 12 / 15 / 19
CLS 12 / 15 / 19
CLS 50
CLS 52
k = 1,98
k = 5,9
Inductive
Ultrapure-water
Pure water Service water Wastewater
Concentrate
k = 0,1
CLS 16
CLS 21
CD-CELL.CDRFig. 9.2
Overview of apllicationranges of E+Hconductivity sensors
Accessories Liquisys M CLM 223 F
60 Endress+Hauser
10 Technical data
General specifications Manufacturer Endress+Hauser
Product designation Liquisys M CLM 223 F
Input Measured variables conductivity, concentration, temperature
Minimum distance for 0 / 4 ... 20 mA signal
Conductivity measurement meas. value 0 ... 19.99 µS/cm: 2 µS/cmmeas. value 20 ... 199.9 µS/cm: 20 µS/cmmeas. value 200 ... 1999 µS/cm: 200 µS/cmmeas. value 2 ... 19.99 mS/cm: 2 mS/cmmeas. value 20 ... 2000 mS/cm: 20 mS/cm
Concentration measurement no minimum distance
Conductive conductivity measurement
Measuring range conductivity: 0 ... 400 mS/cm (uncompensated)concentration: 0 ... 9999 %
Usable cell constant k = 0.0025 ... 99.99 cm–1
Maximum cable length to sensor 100 m
Measuring frequency 170 ... 2000 Hz
Inductive conductivity measurement
Measuring range 0 ... 2000 mS/cm (uncompensated)
Usable cell constant k = 0.0025 ... 99.99 cm–1
Maximum cable length to sensor 55 m (CLK 5)
Measuring frequency 2 kHz
Temperature measurement
Temperature sensor Pt 100, Pt 1000, NTC
Measuring range –35 ... +250 °C
Temperature offset range ±5.0 °C
Temperature compensation
Compensation types linear, NaCl, table
Range –35 ... +250 °C
Reference temperature 25 °C
Digital inputs 1 and 2
Voltage 10 ... 50 V
Current consumption max. 10 mA
%Ω°C
LM2XFE11.CHP
Liquisys M CLM 223 F Technical data
Endress+Hauser 61
Output Conductivity signal output
Current range 0 / 4 ... 20 mA, galvanically isolated; error current 2.4 / 22 mA
Load max. 500 Ω
Maximum resolution 700 digits/mA
Output range adjustable
Separation voltage max. 350 Vrms / 500 V DC
Overvoltage protection acc. to EN 61000-4-5:1995
Temperature signal output (optional)
Current range 0 / 4 ... 20 mA, galvanically isolated
Load max. 500 Ω
Maximum resolution 700 digits/mA
Output range adjustable, ∆ 10 ... ∆ 100 % of upper range value
Separation voltage max. 350 Vrms / 500 V DC
Overvoltage protection acc. to EN 61000-4-5:1995
Auxiliary voltage output
Output voltage 15 V ± 0.6 V
Output current max. 10 mA
Contact outputs (potential-free changeover contacts)
Switching current with ohmic load (cos ϕ = 1) max. 2 A
Switching current with inductive load (cos ϕ = 0.4) max. 2 A
Switching voltage max. 250 V AC, 30 V DC
Switching power with ohmic load (cos ϕ = 1) max. 500 VA AC, 60 W DC
Switching power with inductive load (cos ϕ = 0.4) max. 500 VA AC, 60 W DC
Limit contactor
Pickup / dropout delay 0 ... 2000 s
Alarm
Function (switchable) steady / fleeting contact
Alarm threshold adjustment range conductivity / temperature / concentration:entire measuring range
Alarm delay 0 ... 2000 s (min)
Accuracy Conductivity measurement
Display deviation1 max. 0.5 % of meas. value ± 4 digits
Repeatability max. 0.2 % of meas. value ± 2 digits
Measurement deviation1, conductivity signal output 0.75 % of current output range
Temperature measurement
Measured value resolution 0.1 °C
Display deviation1 max. 1.0 % of measuring range
Measurement deviation1, temperature signal output max. 1.25 % of current output range
%Ω°C Technical data Liquisys M CLM 223 F
62 Endress+Hauser
Ambient conditions Ambient temperature (nom. operating conditions) –10 ... +55 °C
Ambient temperature (limit operating conditions) –20 ... +60 °C
Storage and transport temperature –25 ... +65 °C
Relative humidity (nom. operating conditions) 10 ... 95 %,non-condensing
Ingress protection IP 54 (front), IP 30 (housing)
Electromagnetic compatibility interference emission and interference immunityacc. to EN 61326:1997 / A1:1998
Physical data Dimensions (H × W × D) 96 × 96 × 145 mm
Installation depth approx. 165 mm
Weight
Measured value display LC display, two lines, five and nine digits, with statusindicators
Materials
Housing of panel-mounted unit polycarbonate
Front membrane polyester, UV-resistant
Power supply Supply voltage 100 / 115 / 230 V AC +10 / –15 %, 48 ... 62 Hz24 V AC/DC +20 / –15 %
Power consumption max. 7.5 VA
Mains fuse fine-wire fuse, medium time-lag, 250 V / 3.15 A
1According to IEC 746-1, for nom. operating conditions
Subject to modifications.
%Ω°C
LM2XFE11.CHP
Liquisys M CLM 223 F Technical data
Endress+Hauser 63
11 Appendix
Sto
reca
libra
tion
resu
lts
; no;
new
yes
Ent
ry o
fva
lue
ofca
libra
tion
solu
tion
0.00
...20
.00
%/K
α
2.10
%/K
Ent
ry o
fca
libra
tion
tem
pera
ture
(if B
1=
fixed
)
–35.
0...
+25
0.0
C25
.0C°
°
Ent
ry o
f cor
rect
cond
uctiv
ity v
alue
ofca
libra
tion
solu
tion
0.0
µS/c
m...
9999
mS
/cm
curr
ent
mea
s.va
lue
.
C13
3
Dis
play
of c
alcu
late
din
stal
latio
nfa
ctor
0.10
...5.
01.
0
Dis
play
of
calib
ratio
nst
atus
o.k.
;E
---
Fun
ctio
ngr
oup
CA
LIB
RAT
ION
C
Cal
ibra
tion
Inst
F=
inst
alla
tion
fact
or
Cel
lc=
cell
cons
tant
Airs
=A
irset
C1
(3)
C1
(2)
C1
(1)
C13
6
Res
idua
l cou
plin
gS
tart
cal
ibra
tion
curr
ent
mea
s.va
lue
Dis
play
of
resi
dual
coup
ling
valu
e
0.0
...80
.0µ
S/c
mC
112
Dis
play
of
calib
ratio
nst
atus
o.k.
;E
---
.
C11
3
Sto
reca
libra
tion
resu
lts
; no;
new
yes
C11
4C
111
C13
2C
131
C13
4C
135
F unc
tion
grou
p
SE
TUP
1
ME
AS
.VA
LUE
DIS
PLA
Y
Con
duct
ivity
and
tem
pera
ture
(°C
)
A
Sel
ectio
nof
oper
atin
gm
ode
conc
=co
ncen
tratio
nco
nd=
cond
uctiv
ity
A1
Sel
ectio
nof
dis
play
unit
µS/c
m;m
S/c
m;S
/cm
;µS
/m;m
S/m
;S/m
auto
;
Dis
play
of
cond
uctiv
ity (u
ncom
pens
ated
)co
ncen
tratio
n
A4
Sel
ectio
nof
dis
play
form
at(if
A1
= co
nc)
X.x
xx;
; XX
X.x
;XX
XX
XX
.xx
Dis
play
of
cond
uctiv
ity
A3
Sel
ectio
nof
dis
play
unit
; mg/
l;%
;TD
S;
none
ppm
Dis
play
of
cond
uctiv
ity a
ndte
mpe
ratu
re(°
F )S
elec
tion
with
PLU
Ske
y
Sel
ectio
nw
ith M
INU
Ske
y
Edit mode:Code 22
E key
Edit mode:Code 22
CAL key
A2
Err
ordi
spla
y(u
p to
10
erro
rs)
Err
---
Dis
play
of c
urre
ntpa
ram
eter
set
(onl
yw
ith M
RS
)
Ent
ry o
f cel
l con
stan
t
0.00
25...
...99
.99
1/cm
5.9
A5
ind.
:Ent
ry o
f ins
talla
tion
fact
or01
......
5.00
cond
.:en
try
ofca
ble
resi
stan
ce0.
00...
99.9
9
1.00
ΩA
6
En t
ry o
fm
e asu
red
valu
ed a
mpi
ng
(no
dam
ping
)1
...60
1
A7
Fun
ctio
ngr
oup
SE
TU
P2
B
Sel
ectio
nof
tem
pera
ture
mea
sure
men
t
Pt1
k(=
Pt 1
000)
NT
C30
(=N
TC
30k
Pt1
00
Ω)
fixed
B1
Sele
ctio
no
fte
mpera
ture
com
pen
satio
nty
pe
no
ne
NaC
I=
com
mon
salt
Tab
=ta
ble
1..
. 4
lin=
linea
r
.
B2
Ent
ry o
fva
lue
(if B
2=
linea
r)
0.00
...20
.00
%/K
α
2.10
%/K
B3
Ent
ry o
f cor
rect
proc
ess
tem
pera
ture
(if B
1=
fixed
)
25.0
°C–3
5.0
°C...
+25
0.0
°C
B4
Tem
pera
ture
sens
orof
fset
(not
ifB
1=
fixed
)
–35.
0...
+25
0.0
En
try
of
actu
alte
mp
.°C
B5
Dis
play
of
tem
pera
ture
diffe
renc
e(n
otif
B1
=fix
ed)
0.0
°C–5
.0...
5.0
°C
B6
Sto
reca
libra
tion
resu
lts
; no;
new
yes
C12
6
Ent
ry o
fca
libra
tion
tem
pera
ture
(if B
1=
fixed
)
–10.
0...
+15
0.0
C25
.0C
.
°°
C12
1
Ent
ry o
fva
lue
ofca
libra
tion
solu
tion
0.00
...20
.00
%/K
α
2.10
%/K
.
C12
2
Ent
ry o
f cor
rect
cond
uctiv
ity v
alue
ofca
libra
tion
solu
tion
0.0
mS
/cm
...9
999
mS
/cm
curr
ent
mea
s.va
lue
.
C12
3
Dis
play
of c
alcu
late
dce
llco
nsta
nt
0.1
...9.
991/
cm
C12
4
Dis
play
of
calib
ratio
nst
atus
o.k.
;E
---
C12
5
Fie
ldfo
ren
try
ofus
er s
ettin
g
Appendix Liquisys M CLM 223 F
64 Endress+Hauser
Fun
ctio
ngr
oup
ALA
RM
F
Sel
ectio
nof
cont
act t
ype
=st
eady
cont
act
Fle
et=
fleet
ing
cont
act
Ste
adF
1
Fun
ctio
ngr
oup
CH
EC
K
P
Sel
ectio
nof
unit
for
alar
mde
lay
min
s;F
2
Ent
ry o
fal
arm
dela
y
...20
00s
(min
)(d
epen
dent
onF
2)0
F3
Det
erm
inat
ion
ofer
ror
curr
ent
2.4
mA
22m
AF
4
Sel
ectio
nof
erro
rnu
mbe
r
1...
255
1F
5
Set
ala
rmco
ntac
t to
beef
fect
ive
; no
yes
F6
Set
err
orcu
rren
tto
beef
fect
ive
yes
no
;F
7F
8
Sel
ect "
next
err
or"
orre
turn
tom
enu
Rn
ext
=n
ext
erro
r←
Fun
ctio
ngr
oup
OU
TP
UT
O
Sel
ectio
nof
Cur
rent
outp
ut
; Out
2O
ut
1
O1
Sel
ectio
nof
char
acte
ristic
sim
=si
mul
atio
n
lin=
linea
r
Ent
ry o
fsi
mul
atio
nva
lue
0...
22.0
0m
Acu
rren
tva
lue
O22
1O
2(2
)
O2
(1)
Sel
ectio
nof
curr
ent r
ange 0-20
mA
4-20
mA
;
O21
1O
212
Ent
ry o
f20
mA
valu
e
entir
em
eas.
rang
e
2000
mS
/cm
; 99.
99%
;15
0.0
°C
O21
3
Ent
ry o
f0/
4m
Ava
lue
entir
em
eas.
rang
e0
µS
/cm
;0%
;0
°C
P2
PC
SA
larm
setti
ng(li
vech
eck)
/ 1h
/ 2h
/ 4h
off
Mon
itorin
glim
it0.
3%
ofm
ean
valu
eov
ertim
een
tere
d
Sw
itch
pola
risat
ion
dete
ctio
non
orof
f
ono
ff;
P1
R
Fun
ctio
ngr
oup
RE
LAY
(onl
y if
MR
S)
Sel
ectio
nof
cont
act
tobe
conf
igur
ed
Rel
2R
el1
R1
Sw
itch
limit
func
tion
for
sele
cted
cont
act o
ffor
on
onoff
R2
Sel
ectio
nof
cont
act
switc
h-on
poin
t
entir
em
eas.
rang
e
9999
mS
/cm
; 200
0m
S/c
m;
9999
%
Sel
ectio
nof
cont
act
switc
h-of
f poi
nt
entir
em
eas.
rang
e
9999
mS
/cm
; 200
0m
S/c
m;
9999
%
R3
R4
Pic
kup
dela
yse
tting
0...
2000
s0
s
Dro
pout
dela
yse
tting
0...
2000
s0
s
R5
Set
ting
of a
larm
thre
shol
d(a
s an
abso
lute
valu
e)
entir
e m
eas.
rang
e
9999
mS
/cm
;200
0m
S/c
m;
9999
%
R6
R7
Fun
ctio
ngr
oup
ALP
HA
TAB
LEE
ntry
of n
umbe
rof
valu
epa
irs in
tabl
e
1...
101
Sel
ectio
nof
tabl
es
1...
4(>
1on
lyw
ithM
RS
)
1
Sel
ectio
nof
tabl
eop
tion
edit
read
T2
Mul
tiplic
atio
nfa
ctor
for
conc
entr
atio
nva
lue
ofa
user
tabl
e
0.5
...1.
51
K2
T1
Sel
ectio
nof
tabl
eva
lue
pair
1...
num
ber
from
R3
asig
n
1
T4
Ent
ry o
fte
mpe
ratu
reco
effic
ient
(yva
lue)
0.00
...20
.00
%/K
α
2.10
%/K
Ent
ry o
fte
mpe
ratu
reva
lue
(xva
lue)
–35.
0...
250.
0°C
0.0
°C
Out
put
tabl
est
atus
o.k
.
noye
s;
T5
T6
T7
TT
3
Fun
ctio
ngr
oup
CO
NC
EN
TR
ATIO
N
K
Sel
ectio
nof
activ
eco
ncen
trat
ion
tabl
e
; HS
O ;
HP
O ;
HN
OU
ser
...4
NaO
H 1
24
34
3
K1
Sel
ectio
nof
tabl
es
1...
41
K3
Sel
ectio
nof
tabl
eop
tion
edit
read
K4
Ent
ryof
num
ber
of v
alue
pairs
inta
ble
1...
161
Sel
ectio
nof
tabl
eva
lue
pair
1...
num
ber
from
K5
1
K5
Ent
ry o
fun
com
pens
ated
cond
uctiv
ity v
alue
0.0
...99
99m
S/c
m0.
0µ
S/c
mK
6
Ent
ry o
f ass
ocia
ted
conc
entr
atio
nva
lue
0...
99.9
9%
0.00
%
K7
Ent
ry o
f ass
ocia
ted
tem
pera
ture
valu
e
0.0
°C–3
5.0
...+
250.
0°C
K8
Out
put
tabl
est
atus
o.k
.
noye
s;
K9
K10
LM2XFE11.CHP
Liquisys M CLM 223 F Appendix
Endress+Hauser 65
Fun
ctio
ngr
oup
SE
RV
ICE
Fun
ctio
ngr
oup
DE
TE
RM
IN. O
FT
EM
PE
RAT
UR
EC
OE
FF
ICIE
NT
Fun
ctio
ngr
oup
ME
AS
. RA
NG
ES
WIT
CH
ING
(MR
S)
S D M
Sel
ectio
nof
lang
uage GE
RIT
A;F
RA
ES
P;N
EL
EN
G;
Ent
ryof
com
pens
ated
cond
uctiv
ity
0...
9999
curr
ent
valu
e
Sel
ectio
nof
bin
ary
inpu
tsfo
rM
RS
0...
22
Sel
ectio
nof
tem
pera
ture
com
pens
atio
n
none
;N
aCl;
Tab
1...
4if
M4=
cond
lin;
S1
D1
M1
M6
Sof
twar
eve
rsio
n
SW
vers
ion
E11
1
Har
dwar
eve
rsio
n
HW
vers
ion
E11
2
Dis
play
of
seria
l num
ber
E11
3
Sof
twar
eve
rsio
n
SW
vers
ion
E12
1
Har
dwar
eve
rsio
n
HW
vers
ion
E12
2
Dis
play
of
seria
l num
ber
E12
3
Sof
twar
eve
rsio
n
SW
vers
ion
E13
1
Har
dwar
eve
rsio
n
HW
vers
ion
E13
2
Dis
play
of
seria
l num
ber
E13
3
HO
LDco
nfig
urat
ion
none
=no
HO
LD
Set
up=
durin
gse
tup
CA
L=
dur.
calib
ratio
n
S+C
=d
uri
ng
setu
pan
dca
libra
tio
n
Ent
ryof
curr
ent t
empe
ratu
re
–35
...+
250
°Ccu
rren
tva
lue
Sel
ectio
nof
para
met
erse
t
1...
4if
M1=
01
...2
if M
1=1
1
Ent
ryof
mea
sure
dva
lue
for
0/4
mA
valu
eco
nd.:
0...
2000
mS
/cm
conc
.:0
...99
99.9
99%
Uni
t:A
2F
orm
at: A
3
Ent
ryof
mea
sure
dva
lue
for
20m
Ava
lue
cond
.:0
...20
00m
S/c
mco
nc.:
0...
9999
.999
%U
nit:
A2
For
mat
: A3
Ent
ryof
lim
itsw
itch-
onpo
int
cond
.:0
...20
00m
S/c
mco
nc.:
0...
9999
.999
%U
nit:
A2
For
mat
: A3
Ent
ryof
lim
itsw
itch-
off p
oint
cond
.:0
...20
00m
S/c
mco
nc.:
0...
9999
.999
%U
nit:
A2
For
mat
: A3
S4
D3
M3
M8
Ent
ryof
fixed
valu
e(o
nly
if fix
ed)
0...
100
%of
20
or16
mA
0
Sel
ectio
nof
HO
LDef
fect
fix=
fixed
valu
efr
oz
=la
stva
lue
Dis
play
of
unco
mpe
nsat
edco
nduc
tivity
0...
9999
curr
ent
valu
e
Dis
play
of c
urre
ntpa
ram
eter
set
1...
4if
M1=
04
Ent
ryof
alph
ava
lue
0...
20%
/K2.
1
S3
S2
D2
M2
M7
Man
ual H
OLD
onoff
Ent
ryof
HO
LDdw
ell p
erio
d
0...
999
s10
Dis
play
of
alph
ava
lue
dete
rmin
ed
2.10
%/K
Sel
ectio
nof
ope
r.m
ode
conc
=co
ncen
trat
ion
1...
2if
M1=
11
...4
if M
1=2
con
d=
con
du
ctiv
ity
Sel
ectio
nof
med
ium
HS
O;
HP
O; H
NO
;U
ser
1...
2if
M1=
1U
ser
1...
4if
M1=
2
NaO
H;
24
34
3
S5
D4
M4
M9
M5
M10
M12
M11
M13
S6
Dis
play
of
orde
rnu
mbe
r
S9
Dis
play
of
seria
l num
ber
S8
S10
Fun
ctio
ngr
oup
INT
ER
FAC
EI
Fun
ctio
ngr
oup
E+
HS
ER
VIC
E
E
Ent
ryof
add
ress
HA
RT
:...
15P
RO
FIB
US
:1...
012
6
Mod
ule
sele
ctio
n
Mai
nB=
mai
nboa
rd
Tran
s=
tran
smitt
er
Con
tr=
cont
rolle
r
I1
E1(
3)
E1(
2)
E1(
1)
Dis
play
of
mod
ule
iden
tific
atio
n
E11
4
Dis
play
of
mod
ule
iden
tific
atio
n
E12
4
Dis
play
of
mod
ule
iden
tific
atio
n
E13
4
Inst
rum
ent r
eset
Sen
s=
sens
orda
ta;
Fact
y=
fact
ory
setti
ngs
no
;S
tart
inst
rum
ent t
est
;D
ispl
ayn
o
S11
M14
M15
Ent
erpi
ckup
dela
y
0...
2000
s0
s
Ent
erdr
opou
t del
ay
0...
2000
s0
s
Sel
ectio
nof
con
tact
tobe
conf
igur
edR
el1
Rel
2
Sw
itch
func
tion
ofco
ntac
t off
oron
On
Off
Ent
erS
Wup
grad
ere
leas
eco
defo
rfo
odpa
ckag
e
0000
...99
9900
00
S7
Appendix Liquisys M CLM 223 F
66 Endress+Hauser
12 Index
!9999 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
AA1 ... A7 . . . . . . . . . . . . . . . . . . . . . . . . . 20-21Access codes . . . . . . . . . . . . . . . . . . . . . . . 17Accessories . . . . . . . . . . . . . . . . . . . . . . 59-60ALARM (F) . . . . . . . . . . . . . . . . . . . . . . . . . . 24Alarm contact. . . . . . . . . . . . . . . . . . . . . 19, 25Alarm delay . . . . . . . . . . . . . . . . . . . . . . . . . 24Alarm threshold . . . . . . . . . . . . . . . . . . . . . . 28ALPHA TABLE (T) . . . . . . . . . . . . . . . . . . . . 31Alpha value . . . . . . . . . . . . . . . . . . . . . . 29, 41Alpha value entry . . . . . . . . . . . . . . . . . . . . . 22Assembly of CLM 223 . . . . . . . . . . . . . . . . . 54AUTO key. . . . . . . . . . . . . . . . . . . . . . . . . . . 13Auto mode . . . . . . . . . . . . . . . . . . . . . . . . . . 14
BB1 ... B6 . . . . . . . . . . . . . . . . . . . . . . . . . 21-22
CC1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38-40Cable length. . . . . . . . . . . . . . . . . . . . . . . . . 10CAL key . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13Calibration . . . . . . . . . . . . . . . . . . . . . . . 17, 38Calibration mode . . . . . . . . . . . . . . . . . . . . . 18Calibration of inductive sensors . . . . . . . . . 38Calibration of temperature sensor . . . . . . . . 22Calibration with sensor adaptation . . . . . . . 40CHECK (P) . . . . . . . . . . . . . . . . . . . . . . . . . . 26CLK 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . 10, 59Code 9999 . . . . . . . . . . . . . . . . . . . . . . . . . . 13Coding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Complete installation . . . . . . . . . . . . . . . . . . . 6Complete measuring system. . . . . . . . . . . . . 6CONCENTRATION (K). . . . . . . . . . . . . . . . . 33Concentration measurement . . . . . . . . . . . . 32Conductivity measuring cells. . . . . . . . . . . . 60Connection diagram . . . . . . . . . . . . . . . . . . . 8Connection, examples . . . . . . . . . . . . . . . . . 11Contact selection . . . . . . . . . . . . . . . . . . . . . 24Corrective maintenance of CLM 223. . . . . . 53Corrective maintenance of measuring system .............................57Current outputs . . . . . . . . . . . . . . . . . . . . . . 22CYK 71 . . . . . . . . . . . . . . . . . . . . . . . . . . 10, 59
DD1 ... D4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Declaration of conformity. . . . . . . . . . . . . . . . 5Definition of terms . . . . . . . . . . . . . . . . . . . . 51Delivery, scope of . . . . . . . . . . . . . . . . . . . . . 2Determination of temperature coefficient / al-pha value . . . . . . . . . . . . . . . . . . . . . . . . . . . 41Diagnosis . . . . . . . . . . . . . . . . . . . . . . . . 51-58Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . 7Dismantling of CLM 223 . . . . . . . . . . . . . . . 54Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Display during measurement. . . . . . . . . . . . 17Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Dropout delay . . . . . . . . . . . . . . . . . . . . . . . 28
EE+H SERVICE (E) . . . . . . . . . . . . . . . . . . . . 37E1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Electrical connection. . . . . . . . . . . . . . . . . . . 8ENTER key. . . . . . . . . . . . . . . . . . . . . . . . . . 13Error codes . . . . . . . . . . . . . . . . . . . . . . . . . 49Error current . . . . . . . . . . . . . . . . . . . . . . 24-25Error log. . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Error messages . . . . . . . . . . . . . . . . . . . . . . 46Error selection . . . . . . . . . . . . . . . . . . . . . . . 24Escape function. . . . . . . . . . . . . . . . . . . . . . 13Extension cable CLK 5 . . . . . . . . . . . . . . . . 59Extension cable CYK 71 . . . . . . . . . . . . . . . 59Extension of measuring cable. . . . . . . . . . . 10
FF1 ... F8 . . . . . . . . . . . . . . . . . . . . . . . . . . 24-25Factory settings. . . . . . . . . . . . . . . . . . . . . . 18Freezing of outputs . . . . . . . . . . . . . . . . . . . 17Function coding. . . . . . . . . . . . . . . . . . . . . . 18Function group . . . . . . . . . . . . . . . . . . . . . . 16Fuse. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
GGeneral . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3General safety instructions . . . . . . . . . . . . . . 4
HHardware version . . . . . . . . . . . . . . . . . . . . 37HART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Hold delay time . . . . . . . . . . . . . . . . . . . . . . 35HOLD function. . . . . . . . . . . . . . . . . . . . . 16-17
II1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Immunity to interference . . . . . . . . . . . . . . . . 5Inquiries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Installation . . . . . . . . . . . . . . . . . . . . . . . 4, 6-11Instrument configuration . . . . . . . . . . . . . 18-26Instrument connections. . . . . . . . . . . . . . . . . 9Intended application . . . . . . . . . . . . . . . . . . . 4INTERFACE (I). . . . . . . . . . . . . . . . . . . . . . . 37Interfaces. . . . . . . . . . . . . . . . . . . . . . . . . . . 45
JJunction box VBM . . . . . . . . . . . . . . . . . 10, 59
KK1 ... K10. . . . . . . . . . . . . . . . . . . . . . . . . 33-34Key functions. . . . . . . . . . . . . . . . . . . . . . . . 13Keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
LLanguage selection. . . . . . . . . . . . . . . . . . . 34LED indicators . . . . . . . . . . . . . . . . . . . . . . . 12Linear characteristic . . . . . . . . . . . . . . . . . . 22Liquid crystal display . . . . . . . . . . . . . . . . . 12Locking the hardware . . . . . . . . . . . . . . . . . 13
LM2XFE12.CHP
Liquisys M CLM 223 F Index
Endress+Hauser 67
MM1 ... M11 . . . . . . . . . . . . . . . . . . . . . . . . 43-44Maintenance and troubleshooting . . . . . 46-50Manual hold . . . . . . . . . . . . . . . . . . . . . . . . . 35Manual mode . . . . . . . . . . . . . . . . . . . . . . . . 14Measurement display . . . . . . . . . . . . . . . . . . 17Measuring cable . . . . . . . . . . . . . . . . . . . . . . 10Measuring cable connection . . . . . . . . . . . . 10Measuring mode. . . . . . . . . . . . . . . . . . . . . . 18Measuring system. . . . . . . . . . . . . . . . . . . . . . 6Menu structure . . . . . . . . . . . . . . . . . . . . . . . 16MINUS key . . . . . . . . . . . . . . . . . . . . . . . 13, 17Monitoring features . . . . . . . . . . . . . . . . . . . . . 5Monitoring functions . . . . . . . . . . . . . . . . . . . 24Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
OO1 ... O2 . . . . . . . . . . . . . . . . . . . . . . . . . 22-23Operating concept . . . . . . . . . . . . . . . . . . . . 15Operating mode . . . . . . . . . . . . . . . . . . . . . . 15Operation . . . . . . . . . . . . . . . . . . . . . . 4, 12-17Operator interface. . . . . . . . . . . . . . . . . . . . . 12Optoscope . . . . . . . . . . . . . . . . . . . . . . . . . . 56Order number . . . . . . . . . . . . . . . . . . . . . . . . 35OUTPUT (O) . . . . . . . . . . . . . . . . . . . . . . . . . 22
PP1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26P2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Packaging. . . . . . . . . . . . . . . . . . . . . . . . . . . . 2PCS alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Pickup delay . . . . . . . . . . . . . . . . . . . . . . . . . 28PLUS key. . . . . . . . . . . . . . . . . . . . . . . . . 13, 17Polarisation detection . . . . . . . . . . . . . . . . . . 25Process temperature . . . . . . . . . . . . . . . . . . 22Product structure . . . . . . . . . . . . . . . . . . . . . . 3PROFIBUS . . . . . . . . . . . . . . . . . . . . . . . . . . 37Programming . . . . . . . . . . . . . . . . . . . . . . . . 16
RR1 ... R7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28REL key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13RELAY (R) . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Relay configuration . . . . . . . . . . . . . . . . . . . . 27Replacement of central module . . . . . . . . . . 56
SS1 ... S11 . . . . . . . . . . . . . . . . . . . . . . . . 34, 36Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5Safety features . . . . . . . . . . . . . . . . . . . . . . . . 5Safety instructions . . . . . . . . . . . . . . . 4, 46, 51Scopeware . . . . . . . . . . . . . . . . . . . . . . . . . . 56Selection of time unit . . . . . . . . . . . . . . . . . . 24Self-test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Sensor adaptation . . . . . . . . . . . . . . . . . . . . 40Sensor installation . . . . . . . . . . . . . . . . . . . . 10Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51SERVICE (S) . . . . . . . . . . . . . . . . . . . . . . 20, 34SETUP 1 (A) . . . . . . . . . . . . . . . . . . . . . . . . . 20SETUP 2 (B) . . . . . . . . . . . . . . . . . . . . . . 20-21Setup mode . . . . . . . . . . . . . . . . . . . . . . . . . 18Shipping documents . . . . . . . . . . . . . . . . . . . 2Simulation. . . . . . . . . . . . . . . . . . . . . . . . . . . 23Software version. . . . . . . . . . . . . . . . . . . . . . 37Spare parts for CLM 223 . . . . . . . . . . . . . . . 55Spare parts orders . . . . . . . . . . . . . . . . . . . . 56Start-up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Start-up menu. . . . . . . . . . . . . . . . . . . . . . . . 20Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Structure of coding. . . . . . . . . . . . . . . . . . . . 18Structure of measuring cables. . . . . . . . . . . 10Switch-off point of contact . . . . . . . . . . . . . . 28Switch-on point of contact . . . . . . . . . . . . . . 28Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Symbols used. . . . . . . . . . . . . . . . . . . . . . . . . 2System configuration . . . . . . . . . . . . . . . . . . 20
TT1 ... T7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31Technical data . . . . . . . . . . . . . . . . . . . . 61-63Temperature coefficient . . . . . . . . . . . . . . . . 29Temperature compensation. . . . . . . . . . . . . 21Temperature sensor. . . . . . . . . . . . . . . . . . . 21Tensioning screws . . . . . . . . . . . . . . . . . . . . . 7Termination of measuring cables . . . . . . . . 10Time unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Troubleshooting . . . . . . . . . . . . . . . . . . . . . . 46Troubleshooting common problems . . . . . . 46
UUnlocking the hardware. . . . . . . . . . . . . . . . 13Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Index Liquisys M CLM 223 F
68 Endress+Hauser
Europe
Austria Endress+Hauser Ges.m.b.H.WienTel. (01) 880 56-0, Fax (01) 88056-35
BelarusBelorgsintezMinskTel. (01 72) 50 84 73, Fax (0172) 508583
Belgium / Luxembourg Endress+Hauser N.V.BrusselsTel. (02) 248 0600, Fax (02) 24805 53
BulgariaINTERTECH-AUTOMATIONSofiaTel. (02) 6648 69, Fax (02) 9 63 1389
Croatia Endress+Hauser GmbH+Co.ZagrebTel. (01) 663 7785, Fax (01) 66378 23
CyprusI+G Electrical Services Co. Ltd.NicosiaTel. (02) 4847 88, Fax (02) 4846 90
Czech Republic Endress+Hauser GmbH+Co.PrahaTel. (0 26) 6 7842 00, Fax (0 26) 67841 79
Denmark Endress+Hauser A/SSøborgTel. (70) 1311 32, Fax (70) 1321 33
EstoniaELVI-AquaTartuTel. (7) 4416 38, Fax (7) 44 1582
Finland Endress+Hauser OyEspooTel. (09) 867 6740, Fax (09) 86 7674 40
France Endress+Hauser S.A.HuningueTel. (3 89) 69 6768, Fax (389) 6948 02
Germany Endress+Hauser Messtechnik GmbH+Co.Weil am RheinTel. (0 7621) 975-01, Fax (0 7621) 975-5 55
Great Britain Endress+Hauser Ltd.ManchesterTel. (01 61) 2 865000, Fax (01 61) 9 981841
GreeceI & G Building Services Automation S.A.AthensTel. (01) 924 1500, Fax (01) 92217 14
HungaryMile Ipari-ElektroBudapestTel. (01) 261 5535, Fax (01) 26155 35
IcelandBIL ehfReykjavikTel. (05) 6196 16, Fax (05) 6196 17
IrelandFlomeaco Company Ltd.KildareTel. (0 45) 86 8615, Fax (045) 8681 82
Italy Endress+Hauser Italia S.p.A.Cernusco s/N MilanoTel. (02) 9219 21, Fax (02) 9210 7153
LatviaRino TKRigaTel. (07) 3150 87, Fax (07) 3150 84
LithuaniaUAB “Agava”KaunasTel. (07) 2024 10, Fax (07) 2074 14
Netherlands Endress+Hauser B.V.NaardenTel. (0 35) 6 958611, Fax (0 35) 6 9588 25
Norway Endress+Hauser A/STranbyTel. (0 32) 85 98 50, Fax (032) 8598 51
Poland Endress+Hauser Polska Sp. z o.o.K. WarszawyTel. (0 22) 7 201090, Fax (0 22) 7 2010 85
PortugalTecnisis, LdaCacémTel. (0 21) 4 267290, Fax (0 21) 4 2672 99
RomaniaRomconseng S.R.L.BucharestTel. (01) 410 16 34, Fax (01) 4 1125 01
Russia Endress+Hauser Moscow OfficeMoscowTel. (0 95) 1 587564, Fax (0 95) 1 5898 71
SlovakiaTranscom Technik s.r.o.BratislavaTel. (7) 4488 86 84, Fax (7) 44 8871 12
Slovenia Endress+Hauser D.O.O.LjubljanaTel. (0 61) 15 19 2217, Fax (0 61) 15 19 2298
Spain Endress+Hauser S.A.Sant Just DesvernTel. (0 93) 4 803366, Fax (0 93) 4 7338 39
Sweden Endress+Hauser ABSollentunaTel. (08) 555116 00, Fax (08) 55 51 1655
Switzerland Endress+Hauser AGReinach/BL 1Tel. (0 61) 7 157575, Fax (0 61) 7 1116 50
TurkeyIntek Endüstriyel Ölcü ve Kontrol SistemleriIstanbulTel. (0212) 2 75 1355, Fax (0212) 2 66 2775
UkrainePhotonika GmbHKievTel. (44) 268 81, Fax (44) 2 69 08
YugoslaviaMeris d.o.o.BeogradTel. (11) 444 19 66, Fax (11) 4 4419 66
Africa
EgyptAnasiaHeliopolis/CairoTel. (02) 417 90 07, Fax (02) 4 1790 08
MoroccoOussama S.A.CasablancaTel. (02) 241338, Fax (02) 4026 57
South Africa Endress+Hauser Pty. Ltd.SandtonTel. (0 11) 2 628000, Fax (0 11) 2 6280 62
TunisiaControle, Maintenance et RegulationTunisTel. (01) 793077, Fax (01) 7885 95
America
Argentina Endress+Hauser Argentina S.A.Buenos AiresTel. (01) 145 22 7970, Fax (01) 145 22 7909
BoliviaTritec S.R.L.CochabambaTel. (042) 5 69 93, Fax (042) 509 81
Brazil Samson Endress+Hauser Ltda.Sao PauloTel. (011) 5031 3455, Fax (011) 5031 3067
Canada Endress+Hauser Ltd.Burlington, OntarioTel. (905) 6 81 9292, Fax (9 05) 6 819444
Chile Endress+Hauser Chile Ltd.SantiagoTel. (02) 32130 09, Fax (02) 3 213025
ColombiaColsein Ltda.Bogota D.C.Tel. (01) 23676 59, Fax (01) 6 104186
Costa RicaEURO-TEC S.A.San JoseTel. (02) 96 1542, Fax (02) 961542
EcuadorInsetec Cia. Ltda.QuitoTel. (02) 26 9148, Fax (02) 461833
GuatemalaACISA Automatizacion Y Control Industrial S.A.Ciudad de Guatemala, C.A.Tel. (03) 34 5985, Fax (03) 327431
Mexico Endress+Hauser S.A. de C.V.Mexico CityTel. (5) 568 24 05, Fax (5) 56874 59
ParaguayIncoel S.R.L.AsuncionTel. (021) 2139 89, Fax (021) 226583
UruguayCircular S.A.MontevideoTel. (02) 92 5785, Fax (02) 929151
USA Endress+Hauser Inc.Greenwood, IndianaTel. (317) 5 35-7138, Fax (317) 5 35-8498
VenezuelaControval C.A.CaracasTel. (02) 94409 66, Fax (02) 9 444554
Asia
China Endress+Hauser Shanghai
Instrumentation Co. Ltd.ShanghaiTel. (021) 5490 2300, Fax (021) 5490 2303
Endress+Hauser Beijing OfficeBeijingTel. (010) 6834 4058, Fax (010) 6834 4068
Hong Kong Endress+Hauser HK Ltd.Hong KongTel. 25 283120, Fax 286541 71
India Endress+Hauser (India) Pvt Ltd.MumbaiTel. (022) 8 52 1458, Fax (0 22) 8 521927
IndonesiaPT Grama BazitaJakartaTel. (21) 79750 83, Fax (21) 7 975089
Japan Sakura Endress Co., Ltd.TokyoTel. (0422) 5406 13, Fax (04 22) 55 0275
Malaysia Endress+Hauser (M) Sdn. Bhd.Petaling Jaya, Selangor Darul EhsanTel. (03) 73348 48, Fax (03) 7 338800
PakistanSpeedy AutomationKarachiTel. (021) 772 2953, Fax (021) 7 73 6884
Philippines Endress+Hauser Philippines Inc.Metro ManilaTel. (2) 37236 01-05, Fax (2) 41219 44
Singapore Endress+Hauser (S.E.A.) Pte., Ltd.SingaporeTel. 5 668222, Fax 566 68 48
South Korea Endress+Hauser (Korea) Co., Ltd.SeoulTel. (02) 6 5872 00, Fax (02) 6 59 2838
TaiwanKingjarl CorporationTaipei R.O.C.Tel. (02) 27 183938, Fax (02) 2713 41 90
Thailand Endress+Hauser Ltd.BangkokTel. (2) 99678 11-20, Fax (2) 99678 10
VietnamTan Viet Bao Co. Ltd.Ho Chi Minh CityTel. (08) 8 3352 25, Fax (08) 8 33 5227
IranPATSA Co.TehranTel. (021) 875 4748, Fax (021) 8 74 7761
IsraelInstrumetrics Industrial Control Ltd.Tel-AvivTel. (03) 6 4802 05, Fax (03) 6 47 1992
JordanA.P. Parpas Engineering S.A.AmmanTel. (06) 4 6432 46, Fax (06) 4 64 5707
Kingdom of Saudi ArabiaAnasia Ind. AgenciesJeddahTel. (02) 6 7100 14, Fax (02) 6 72 5929
LebanonNetwork EngineeringJbeilTel. (9) 94 4080, Fax (9) 5480 38
Sultanate of OmanMustafa & Jawad Science & Industry Co. L.L.C.RuwiTel. 60 20 09, Fax 60 70 66
United Arab EmiratesDescon Trading EST.DubaiTel. (04) 2 6536 51, Fax (04) 2 65 3264
YemenYemen Company for Ghee and Soap IndustryTaizTel. (04) 23 0664, Fax (04) 21 2338
Australia + New Zealand
AustraliaALSTOM Australia Ltd.MilperraTel. (02) 97 747444, Fax (02) 9774 46 67
New ZealandEMC Industrial Group Ltd.AucklandTel. (09) 4 1551 10, Fax (09) 4 15 5115
All other countries
Endress+Hauser GmbH+Co.Instruments International
D-Weil am RheinGermanyTel. (076 21) 9 75-02, Fax (076 21) 9 75-345
BA 237C/07/en/01.01Printed in Germany / CV 5 / DT
Members of the Endress+Hauser Group
Hauser+EndressThe Power of Know How51504342