    INDICEINTRODUCCIN......................................................................................................................................................03INSPECCIN INICIAL AL RECIBIMIENTO...................................................................................................04

    EMBALAJE........................... .................... ................. ............................................................................04INSTALACIN.......................................................................................................................................................04PUNTOS.......................... ....................... ................. ...............................................................................05PUNTOS (AHMC 13)........................ ................... ................... ..................... ........................................06SITIO................... ..................... ................... .................. ........................................................................07NIVELACIN DEL EQUIPO................... .................. ...................... .................. ...............................07

    CONOZCA TU EQUIPO.......................... .................... .................... .................. .....................................................08ABASTECIMIENTO DE AGUA........................... ................... ..................... .....................................09CIERRE DE LA PUERTA......... .................. ...................... .................. ................... ...............................09CIERRE DE LA PUERTA (AHMC13)..................... .................. ...................... ....................................11SENSOR DE LA PUERTA.................................................................................................................12

    PANEL DE MANDO VERSIN DISPLAY + LED`S........... .................. ...................... ................. 13TECLA DE SELECCIN...................... .................... ................... ..................... ...... .........................14TECLA DE PARTIDA........................... ................... ..................... ...................................15LED`S INDICATIVOS............................... ................... ..................... ...........................16FASE DE CALENTAMIENTO.................................. ................... ..................................17FASE DE ESTERILIZACIN......................... .................. ............................................18FASE DE DESCARGA............. ...................... .................. ..................... ...........................19FASE DE SECADO........................................................................................................20FIN DE CICLO..... .................. ...................... .................. .................... ..............................21PANEL DE MANDO VERSIN DISPLAY ALFANUMRICO.............................. .22

    TECLA DE SELECCIN.............................. .................... ................................................23

    TECLA DE PARTIDA...................... ....................... ................. .........................................24ENTRADA DE AGUA......................... ...................... .................. ......................................25FASE DE CALENTAMIENTO.................................. ................... ..................................25FASE DE ESTERILIZACIN......................... .................. ............................................26FASE DE DESCARGA............. ...................... .................. ..................... ...........................28FASE DE SECADO.......... ..................... ................... ................... .....................................28FIN DE CICLO..... .................. ...................... .................. .................... ..............................29DESPRESURIZACIN MANUAL................... .................. .................... .......................30DESCARGA DE AGUA MANUAL............................... ................... ................................30

    PREVACUO - BOMBA DE VACUO................. .................... ................... ..................... .................. .........................31OPERACIN............................ .................... .................... .................. ........................................................................31TABLA DE REFERENCIAL DE CICLOS.................... .................. ..................... ................... ................................34TABLA DE REFERENCIAL DE CICLOS BV.................. .................... .................... .................. ............................35FUNCIONES ESPECIALES................................ .................. .................... .................... .........................................36MANTENIMIENTO PREVENTIVO.................................. ...................... .................. ...........................................37ANTES DE LLAMAR LA ASISTENCIA TCNICA..................... .................. ...................... .................. ...........43ALARMAS.......................... .................... .................... .................. ..............................................................................45ESPECIFICACIONES TCNICAS.......................... .................. ...................... .....................................................46TRMINOS DE GARANTA........................ ...................... .................. ..................................................................47ASISTENCIA TCNICA.....................................................................................................................................48

  • 7/25/2019 Manual de Operaciones Autoclaves

    3/47

    3

    AO CLIENTE

    EL CLIENTE DE LA ATENCIN, PGINAS 40 LOS 50 ESRESPONSABILIDADES DEL USUARIO.

    NO UTILIZA ESTE EQUIPO ANTES DE LA LECTURA Y DELACUERDO DE ESTE MANUAL

    CUALQUIER DUDA ENTRA EN CONTACTO CON EL SACSERCON, PARA EL TELFONO:

    (55-11) 2149 - 1733

  • 7/25/2019 Manual de Operaciones Autoclaves

    4/47

    4

    INTRODUCCIN

    Usted acaba de dar un paso muy importante para su seguridad y la de sus pacientes. Laautoclave es el punto de partida para cualquier proceso serio de esterilizacin deinstrumentos y materiales mdicos u odontolgicos.

    Se hace necesario siempre repetir que los consultorios son ambientes de riesgo, en mayoro menor grado, dependiendo de su especialidad.

    Otro paso acertado fue la eleccin de la marca sercon que detiene la mejor tecnologa,calidad y seguridad bajo este aspecto, usted acaba de adquirir un equipo de primera lnea,teniendo en vista, por lo tanto, obtener resultados optimizados y libres de problemas.

    Las principales ventajas de las autoclaves sercon son la: rapidez de sus ciclos, su

    seguridad, su precisin y la facilidad de operacin.Las autoclaves compactas de mesa son equipos proyectados para esterilizar instrumentos

    y materiales medicoquirrgicos, en clnicas mdicas, consultorios odontolgicos, urgencias,laboratorios, etc.

    Las autoclaves son esterilizadores por medio de vapor saturado hmedo como agenteesterilizante.

    Este manual se propone a suministrar al usuario una comprensin general sobre elfuncionamiento de la autoclave; la indicacin de las mejores maneras de operarla, y an,los cuidados que deben ser tomados para su observacin, su seguridad y garanta del

    producto.Tras leer este manual, operar la autoclave ser una tarea fcil, sin embargo, por tratarsede un equipo construido con componentes de alta tecnologa, para el perfectofuncionamiento de este producto, busque siempre la asistencia tcnica autorizada sercon,para el esclarecimiento de cualquier duda o mantenimiento correctivo necesario en suequipo.

    Siempre desconecte el cable de la enchufe antes de un trabajo de mantenimiento oconservacin.

    Cuando el equipo est mucho tiempo fuera de uso, retire toda agua del aljibe y tambintodo material en el contenido, seque el aparato por dentro y por fuera y si posible ,cbralo.

  • 7/25/2019 Manual de Operaciones Autoclaves

    5/47

    5

    ESTERILIZACIN

    La confeccin segura y estril de productos hospitalarios requiere atencin en lascaractersticas del producto y en el mtodo de esterilizacin y controles.

    Un producto estril es aquel que est libre de microorganismos viables. Artculosproducidos bajo control de las condiciones de fabricacin, pueden previamente a laesterilizacin poseer microorganismo.

    Tales productos son, por definicin, no estriles. El objetivo del proceso de esterilizacines destruir los contaminantes microbiolgicos de estes productos no estriles.

    La destruccin de microorganismos por agentes fsicos y qumicos sigue una leyexponencial.

    Por consiguiente se puede calcular una probabilidad finita de microorganismossobrevivientes independiente de la dosis o tratamiento de esterilizacin.

    La probabilidad de sobrevivi enca es una funcin del nmero y tipos (especie) demicroorganismos presentes en el producto, la letalidad del proceso de esterilizacin, y enalgunas instancias, del ambiente en el cual los microorganismos existen durante eltratamiento.

    El resultado de esterilidad en un articulo individual, en una poblacin de productosesterilizados, no puede ser garantizada en absoluto.

    La probabilidad de no-esterilidad en cada unidad individual del producto es deducidamatemticamente. Por ejemplo, con una probabilidad de 10-6, la probabilidad de no-esterilizacin de una unidad del producto es menor o igual a uno en un milln.

    Esterilizacin es un ejemplo de proceso especial porque su eficacia no puede serverificada por inspeccin o teste en el producto. Por este motivo, deben ser validadosantes del uso los procesos de esterilizacin, la rutina de monitorizacin del proceso ymantenimiento realizados en el equipo.

    Esterilizadores de vapor con vacuo, cuando de su construccin, montaje y uso, debenatender las determinaciones legales y directrices reconocidas de la tcnica, refirindosea:

    Normas sobre vasos de presin Reglamentos de prevencin de accidentes Norma sobre calderas y las respectivas tcnicas en vigor. Manual del fabricante

    Los usuarios de esterilizadores de vapor con vacuo, antes de poner el equipo en marcha,deben informarse respecto a las determinaciones concernientes a la operacin y laobligacin de una eventual autorizacin.

  • 7/25/2019 Manual de Operaciones Autoclaves

    6/47

    6

    Definiciones

    Mezcla aire-vapor: mezcla homognea de aire y vapor saturado usado paraesterilizacin.

    El aire es utilizado para compensar la presin generada dentro de recipientes

    tapados, que exceden la presin del vapor saturado.Carga microbiana: Poblacin de microorganismos viables en una materia prima,componente, producto final y/o un embalaje.

    Certificacin:Proceso documentado de revisin y aprobacin efectuado como paso finaldel programa, permitiendo la liberacin del producto.

    Valor D: tiempo de exposicin necesario, tras el proceso alcanzar las condicionespredeterminadas para causar de 1-log o 90% en la poblacin de un determinadomicroorganismo.

    Controles ambientales: controles establecidos en las reas de fabricacin de losproductos para control de la carga microbiana.

    Estos pueden incluir filtros de aire y de fluidos, desinfeccin de superficies,informes personales y procedimientos administrativos.

    Valor F0: valor F calculado a 121,1 oc (250 of) con el valor z de 10 K y valor D de 1 min.

    Materias de construccin: materiales utilizados en la composicin del equipo deesterilizacin.

    Desafi microbiolgico:indicadores biolgicos, paquete de test de indicador biolgico, oproducto inoculado que contiene una poblacin conocida de microorganismos y que puedenser utilizados para el test del ciclo de esterilizacin.

    Calor hmedo:calor originado a partir del agua en la forma lquida, o del vapor saturadobajo presin.

    Esterilizacin por calor hmedo: proceso que utiliza calor hmedo para la produccin deproductos estriles.

    Embalaje primario: elemento del sistema que mantiene la esterilidad del producto.

    Letalidad del proceso: capacidad del proceso de esterilizacin en destruirmicroorganismos.

    Esto puede ser determinado por la medicin de la muerte microbiana o por elestablecimiento y medida de los parmetros fsicos requeridos.

    Sistema de cargamento de producto: mecanismo utilizado para acondicionamiento de losproductos embalados, para la esterilizacin.

    El sistema de cargamento de productos debe protegerlos de daos y permitiracceso uniforme por el agente esterilizante.

  • 7/25/2019 Manual de Operaciones Autoclaves

    7/47

    7

    Comisionamiento: obtencin y documentacin de evidencias de que el equipo fuesuministrado e instalado acorde con sus especificaciones y que las funciones estn dentrode los lmites predeterminados cuando operado de acuerdo con las instrucciones deoperacin.

    Recomisionamento: Repeticin de parte o de todos los testes necesarios para elComisionamiento con la intencin de reconfirmar la fiabilidad del proceso.

    Revalidacin: Repeticin de parte o de todos los testes de validacin con la intencin dereconfirmar la fiabilidad del proceso.

    Vapor saturado:Vapor de agua a una temperatura correspondiente al punto de ebullicindel lquido original a una determinada presin.

    Carga simulada del producto: Carga que es utilizada como alternativa con relacin a lacarga del producto real, y que representa un desafo igual o mayor para el proceso.

    Estril: Ausencia de microorganismos viables. En la prctica, ninguna afirmacin semejante respecto a la ausencia de

    microorganismos puede ser probada (vea esterilizacin)

    Esterilizacin: Proceso validado usado para proveer un producto libre de todas lasformas de microorganismos viables.

    En el proceso de esterilizacin, la naturaleza de la muerte de losmicroorganismos es descripta por una funcin exponencial. Por lo tanto, lapresencia de microorganismos viables en cualquier tem individual puede ser

    expresa en trminos de probabilidad. Aunque esta probabilidad pueda serreducida a un nmero muy pequeo, nunca puede ser reducida a cero.

    Ciclo de esterilizacin: secuencia automtica de fases operacionales realizada en elesterilizador.

    Desarrollo del proceso de esterilizacin:estudios conducidos para el desarrollo de unproceso reproducible por el cual el producto pueda ser esterilizado, sin daos, dentro deuna probabilidad deseada de no-esterilidad.

    Validacin: procedimiento documentado para obtencin, registro e interpretacin de

    resultados deseados para el establecimiento de un proceso que deber consistentementesuministrar productos, cumpliendo especificaciones predeterminadas.

    La validacin es cubierta por tres actividades, Comisionamiento, verificacin delas especificaciones del proceso y calificacin de performance.

    Valor z: nmero de grados de temperatura requerida para el cambio de 1-Log en le valorD.

  • 7/25/2019 Manual de Operaciones Autoclaves

    8/47

    8

    INSPECCIN INICIAL AL RECEBIMIENTO

    Embalaje

    Usted recibi el equipo a contento acorde con nuestra instruccin anexa a la factura,observe el modo como el equipo fue embalado y guarde el embalaje hasta que la unidadsea inspeccionada. La inspeccin fsica consiste en la busca de seales de daos, talescomo, superficie del panel araada, botones rotos, etc. Si el dao es aparente, registreuna reclamacin con el transportador y el fabricante por medio de su representante oresponsable por el producto hasta el momento de la entrega. Garantizamos que elproducto fue cuidadosamente inspeccionado antes de su expedicin.

    INSTALACIN

    Antes de instalar su autoclave, verifique si las condiciones de la red elctricaexistentes en el sitio, tales como voltaje, cableado, estn conforme la tabla(vea pgina 41)

    Utiliza cableado y disyuntor exclusivo para el equipo.Aterramiento: utilice asta de aterramiento exclusiva con disruptor de tierra inferior a10 (nunca utilice el neutro de su red elctreica como aterramiento).

    La autoclave debe ser instalada en sitio bien ventilado, encima de un mueble de fuerteestructura, obedeciendo una distancia mnima de 15 cm de paredes y objetos que puedanobstruir su ventilacin, adems de sitios protegidos de rayos solares, luvvia o equipos queirradien magnetismo (mocroondas, antenas de radio o tlefonos mviles, televisores, etc).

    La empresa no se responsabiliza por cualquier incidente en el caso aterramientoincorrecto del equipo..

  • 7/25/2019 Manual de Operaciones Autoclaves

    9/47

    9

    INSTALACIN - PUNTOS

    L IG A R A V O LTA G EM D E M O N S TR A D A

    N A ETIQ UETA A D ESIVA

    L IG A R A F IO TE R RA

    1 5 c m d ed is t n c iad a p a re d e

    1 5 c m d e d is t n c iad a p a re d e

    C HAVE G ERAL

    FIO TERRA

    C A BO D E A LIM E N TA O

    VISTA TRASERA

    OBS.:LA ENCHUFE UTILIZADO PARA LA ALIMENTACIN DEL EQUIPO DEBER SEREXCLUSIVA, SIN EXTENSIONES O LADRN (TEES)

    LA BASE DONDE SERAPOYADO EL EQUIPONO PUEDE SER DEMATERIALINFLAMABLE

    RESERVATRIO

    MANILLA

    PANEL DE CONTROL

    PUERTA

    ENCHUFE

    TAPA DELGABINETE

    TORNILLO DE FIJACIN DE LATAPA DEL GABINETE

    PANEL DE CONTROL

    DISPLAY

    TECLA DE PARTIDA

    TECLA DE SELEO( SELECIN)

    PLUG

    RESERVATRIO DEGUA

    PLACA DEIDENTIFICACIN

  • 7/25/2019 Manual de Operaciones Autoclaves

    10/47

    10

    PANORAMA

    OBS.:LA ENCHUFE UTILIZADOPARA LA ALIMENTACIN DELEQUIPO DEBER SEREXCLUSIVA, SIN

    EXTENSIONES O LADRN

    DETALLE DE INSTALACIN PARA EQUIPOS CON BOMBA DE VACUO (AHMC5,AHMC6, AHMC10 Y AHMC13)

    INSTALACIN PUNTOS (AHM 54 litros)

  • 7/25/2019 Manual de Operaciones Autoclaves

    11/47

    11

    LINHA DE NVEL DEESCOAMENTO

    3GRAUS

    INSTALACIN SITIO

    INSTALACIN NIVELACIN DEL EQUIPO

    Nuestro equipo ya sale de la fbrica con la inclinacin ajustada y debe ser instalado enbase nivelada.

    Caso haya necesidade de nivelar el equipo para que el agua desage totalmente para eldren, utilice el sistema de patas niveladoras. Conforme la figura abajo

    NO DEJEESPUESTO AL SOL

    NO DEJE EXPUESTOA LA HUMEDAD

    NO PUEDESUFRIR CADAS

    NO DEIXE

    EXPOSTO NODEJE EXPUESTO A

    ENERGAELETROMAGNTIC

    PS NIVELANTES

    NO DEJE EXPUESTO

    A ANTENASEMISORAS DE

    ONDAS DE RADIO,TELFONOS

    MVILES O TV

    EJECUTE EL

    ATERRAMIENTO!

    GUA

  • 7/25/2019 Manual de Operaciones Autoclaves

    12/47

    12

    CONOZCA SU EQUIPO

    Este es un equipo que trabaja con presin, energa elctrica, vapor y otros, por lo tanto no

    permita el manuseo por personas sin el conocimiento integral de las instruccionescontenidas en este manual.

    A

    SERCON

    S PPARTIDASELEO

    AMPLIAO DO FILTRO

    CMARA INTERNA

    RESISTNCIA

    TRABA

    GUARNIO

    MANILLA

    PUERTA

    ALOJAMIENTO DE LA

    TRABA

    PAINEL DECOMANDO

    DREN DEL ALJIBE

    SENSOR DE LAPUERTA

    PT 100

    PARAFUSO DAGUARNIO

    PANEL DEMANDOVERSINDISPLAYALFANUMRICODE CRISTALLQUIDO

    BASE DE LA CMARACMARA INTERNA

    RESISTNCIA

  • 7/25/2019 Manual de Operaciones Autoclaves

    13/47

    13

    Autoclave de 54 litros

    MANIPULO

    HASTES DE FECHAMIENTO

    EMPAQUETADURA

    PUERTA

    ALOJAMENTO DE LAS

    HASTES DE FECHAMIENTO

    CARRO INTERNO

    SALIDA DE GUA DABOMBA DE VACIO

    RESERVORIO

    ENTRADA DE GUA DABOMBA DE VACIO

  • 7/25/2019 Manual de Operaciones Autoclaves

    14/47

    14

    ABASTECIMIENTO DE AGUA (RESERVORIO PLSTICO)

    Solamente abastezca su equipo con la puerta abierta.

    Ojo: durante el ciclo es normal que el agua baje su nivel de formaconsiderable; siendo que , despus del ciclo, ella volver al aljibeVerifique y reabastezca el aljibe antes de comenzar otro ciclo

    CIERRE DE LA PUERTA

    Gire la traba a conforme el dibujo abajo hasta la posicin correcta de trabamiento (1er.paso).

    UTILICE SIEMPRE AGUA DESTILADA DE FARMACIA

    VERIFIQUE EL NVEL DE AGUA IDEAL

    TUBO DE LA PURGA

    6 cm

  • 7/25/2019 Manual de Operaciones Autoclaves

    15/47

    15

    Tras la conclusin del tem anterior; gire en el sentido horario la manilla b (2 paso)hasta estancar (no hace falta que se utilice de fuerza, solamente hasta sentir una levepresin).

    Caso la posicin de la traba a est en la posicin incorrecta, ser accionada una alarma,indicando que la puerta no est cerrada correctamente (alarma 0).

  • 7/25/2019 Manual de Operaciones Autoclaves

    16/47

    16

    CIERRE DE LA PUERTA (AHM 54 LITROS)

    PARA CERRARSE LA PUERTA DEL EQUIPO SE DEBE ARRIMAR LA TAPA EN EL FLANJE,

    COMO ILUSTRADO EN LA FIGURA AL LADO Y LUEGO PORCEDER EL GIRO DEL

    VOLANTE CENTRAL EN SENTIDO HORARIO, HASTA QUE SE OBTENGA FIRMEZA EN EL

    GIRO Y EL PEREFECTO CIERRE DE LAS ASTAS DE APOYO.

    Si en el cierre perfecto de lapuerta es obedecido a ella, tendrvaciar de la agua caliente o delvapor, causando a la gente queinscribir directo en contacto fsicocon estas sustancias. Nunca abrirla uerta durante el ciclo

    ATENCINATENCINATENCINATENCIN

  • 7/25/2019 Manual de Operaciones Autoclaves

    17/47

    17

    SENSOR DE LA PUERTA

    INSTALADO DETRS EL PANEL FRONTAL AUXILIA EN LA SEGURIDAD DEL EQUIPO,PREVINIENDO POSIBLES FALLAS EN EL MANUSEO Y CIERRE DE LA PUERTA

    UTILIZACIN: TRANSMITE AL SISTEMA DE CONTROL CORRECTO TRABAMIENTO DE LAPUERTA.

    POSI OFECHADA

    POSIOABERTA

    POSIO

    DE FECHAR INCORRETA

    LINGETA

    MICRO SWIT

    IMPORTANTEADEMS DE HACER ELTRABAMIENTO DE LA

    PUERTA SE DEBE GIRAR LAMANILLA PARA TENER UNBUEN ESTANCO.

  • 7/25/2019 Manual de Operaciones Autoclaves

    18/47

    18

    PANEL DE MANDO

    DISPLAYCRISTAL LQUIDOSUMINISTRA MENSAJES DE PRESIN,TEMPERATURA, CICLOS EN OPERACIN,ALARMAS, ETC

    TECLA SELEO (S)(SELECCIN):SELECCIONA EL CICLO DESEADO DEACUERDO CON EL MATERIAL A SERESTERILIZADO

    TECLA PARTIDA (P):COMIENZA EL CICLO PREPROGRAMADOPOR LA TECLA SELEO (SELECCIN)

  • 7/25/2019 Manual de Operaciones Autoclaves

    19/47

    19

    DESCRIPCIN DE FUNCIONAMIENTO

    TECLA DE SELEO (SELECCIN)

    sSELEO

    PERMITE ELEGIR UNO DE LOS OCHO CICLOS DE ESTERILIZACIN Y LAS FUNCIONESESPECIALES, PARA ESTO ES SLO HACER CLIC REPETIDAMENTE, SIGUIENDO EN EL DISPLAY

    LA EXHIBICIN DE LOS MENSAJES QUE IDENTIFICAN LOS CICLOS DESEADOS, EJEMPLOS:

    FUNES ESPECIAIS

    SELECCIONE:< S >DESPRESURIZACIN

    SELECCIONE:< S >DESCARGA DE AGUA

    SELECCIONE:< S >CICLO 6

    SELECCIONE:< S >CICLO 7 SELECCIONE:< S >CICLO 8

    SELECCIONE:< S >CICLO 1

    SELECCIONE:< S >CICLO 5

    SELECCIONE:< S >CICLO 2

    SELECCIONE:< S >CICLO 3

    SELECCIONE:< S >CICLO 4

    SELECCIONE:< S >SECADO

  • 7/25/2019 Manual de Operaciones Autoclaves

    20/47

    20

    TECLA DE PARTIDA

    ESTA TECLA TIENE COMO FUNCIN, COMENZAR EL CICLO ELEGIDO POR LATECLA (SELECCIN.

    OJO: LA PUERTA DEBE STAR TRABADA Y ESTANCADA Y LA TEMPERATURADE LA CMARA INTERNA DEBE ESTAR ABAJO DE 70 PARA PRESIONARSELA TECLA PARTIDA.

    AL ENCENDER SU EQUIPO EN LA LLAVE GENERAL, SURGIR LA PANTALLA ABAJO,INDICANDO EL MODELO DE LA AUTOCLAVE Y LA VERSIN DEL PROGRAMA.

    modelo da la Autoclave Versin del programa

    DESPUS DE 2 SEGUNDOS SER EXHIBIDA LA PANTALLA DE SELECCIN DE CICLO,CORRESPONDIENTE AL LTIMO CICLO EJECUTADO.

    1. TECLEE < S > HASTA APARECER EL CICLO DESEADO.

    2. TECLEE PARA COMENZAR EL CICLO.

    3. SI ACASO LA PUERTA NO EST CORRECTAMENTE CERRADA EL SISTEMA EMITIR UN BIPINTERMITENTE Y EL MENSAJE ALARMA 00 SER EXHIBIDO EN EL VISOR, CENTELLENADO LAPRIMERA LNEA.

    4. SI ACASOLA TEMPERATURA EST ARRIBA DE 70 C, EL SISTEMA EMITIR UN BIPINTERMITENTE Y EL MENSAJE ALARAM 09 SER EXHIBIDO EN EL VISOR CENTELLEANDO LAPRIEMRA LNEA, INDICANDO QUE EL APARATO AN NO ALCANZ LA TEMPERATURA DEENFRIAMIENTO NECESARIA PARA COMENZAR EL CICLO. ESTA SITUACIN OCURRE DEBIDOAL CALENTAMIENTO DEL CICLO ANTERIOR.

    PPARTIDA

    AUTOCLAVEMOD.AHMC**/V**

    SELECCIONE:

    CICLO 1

    PARTIDA:

    025

    0,00

    >

    PORTA ABERTA / FMP

    >

    TEMP. > 70 C

  • 7/25/2019 Manual de Operaciones Autoclaves

    21/47

    21

    ENTRADA DE AGUASI TODO EST CORRECTO PARA COMENZAR EL CICLO, APARECER EN EL DISPLY EL MENSAJEABAJO;

    LA ENTRADA DE AGUA ES LIBERADA PARA LA CMARA INTERNA DURANTE 5 MINUTOS Y LAVLVULA SALIDA DE VAPOR ES ABIERTA PARA SACAR LA PRESIN EXISTENTE EN LACMARA INTERNA.

    EL DISRUPTOR DE CALENTAMIENTO QUEDA APAGADA DURANTE ESTA FASE.

    ESTAS PANTALLAS NOS MUESTRAN LA FASE DEL CICLO ENTRADA DE AGUA, EL NMERO

    DEL CICLO DE ESTERILIZACIN SELECCIONADO CICLO 6, EL TIEMPO QUE FALTA PARATERMINAR LA FASE (EN MINUTOS Y SEGUNDOS / MM SS) Y LA TEMPERATURA Y LA PRESINEXISTENTES EN LA CMARA.

    CALENTAMIENTO

    DESPUS DE LA FASE DE ENTRADA DE AGUA, EL SISTEMA COMENZAR EL CALENTAMIENTODE LA CMARA INTERNA, HACIENDO CON QUE EL DISRUPTOR CALIENTE EL AGUA, AS COMOABRE LA VLVULA DE SALIDA DE VAPOR PARA PERMITIR LA EXPULSIN DEL AIRE FRO QUEPUEDA ESTAR CONTENIDO EN EL INTERIOR DE LA CMARA.

    LA VLVULA DE ENTRADA DE AGUA SIGUE CERRADA DURANTE ESTA FASE. EN EL VISOR SONEXHIBIDAS DOS PANTALLAS ALTERNADAS TEMPORIZADAS EN 1 SEGUNDO.

    ESTAS PANTALLAS NOS MUESTRAN LA FASE DEL CICLO CALENTAMIENTO EL NMERO DELCICLO SELECCIONADO CICLO 6 Y LA TEMPERATURA Y LA PRESIN EXISTENTE EN LACMARA.

    EN LA FASE DE CALENTAMIENTO PUEDEN OCURRIR HASTA 3 ALARMAS:ALARMA 01

    SI NO HAY AGUA EN LA CMARA O HAY FALLA EN EL SENSOR DE PRESIN O DETEMPERATURA, EL SISTEMA EMITIR UN BIP INTERMITENTE Y EL MENSAJE ALARMA01 SER EXHIBIDO EN EL DISPLAY LCD, CENTELLEANDO LA PRIMERA LNEA.(DETECTADO SI SUBE 8 GRADOS O MS EN 4 SEGUNDOS O SI A 108 C EST CONPRESIN MENOR QUE 0,15 KGF/CM.

    CICLO 6 025AGUARDE . . . 0,00

    AQUECIMENTO 025

    >FALTA GUA / FST

    CICLO 6 0 25C04:54 0,00

    ENTRADA GUA 025C

  • 7/25/2019 Manual de Operaciones Autoclaves

    22/47

    22

    ALARMA 02

    SI LA CAMARA INTERNA NO EST CALENTANDO EL SISTEMA EMITIR UN BIPINTERMITENTE Y EL MENSAJE ALARMA 2 SER EXHIBIDO EN EL DISPLAY LCD,

    CENTELELANDO LA PRIMERA LNERA (DETECTADO S LA TEMPERATURA S LA TEMPERATURAMEDIDA A CADA 3 MINUTOS SEA IGUAL O MENOR QUE LA MEDIDA EN LOS 3 MINUTOSANTERIORES.

    ALARMA 07

    SI NO ALCANZA LA TEMPERATURA Y LA PRESIN RELATIVA AL CICLO SELECCIONADO EN 25MINUTOS, EL SISTEMA EMITIR UN BIP INTERMITENTE Y EL MENSAJE ALARMA 07 SEREXHIBIDO EN EL DISPLAY LCD, CENTELLENADO LA PRIMERA LNEA (DETECTADO S DESPUSDE 25 MINUTOS NO ALCANCE LOS CICLOS 1 A 3 (T=121 C E P>=1,08 KGF/CM O EN LOS CICLOS 4A 7 (T=127 C E P>=1,50 KGF/CM.

    ESTERILIZACINDESPUS DE LA FASE DE CALENTAMIENTO, CUANDO SE ALCANCE LA TEMPERATURA YPRESIN RELATIVA AL CICLO SELECCIONADO, EL SISTEMA VERIFICA SI LA RELACINTEMPERATURA X PRESIN EST COERENTES.

    EN CASO AFIRMATIVO EL SISTEMA COMENZAR LA TEMPORIZACIN DE LA FASE DEESTERILIZACIN, CONFORME EL TIEMPO DE ESTERILIZACIN RELATIVO AL CICLOSELECCIONADO.

    EN EL VISOR SON EXHIBIDAS DOS PANTALLAS ALTERNADAS TEMPORIZADAS EN 1SEGUNDO:

    C C

    ESTAS PANTALLAS NOS MUESTRAN LA FASE DEL CICLO ESTERILIZACIN, EL NMERO DELCICLO SELECCIONADO CICLO 6, EL TIEMPO QUE FALTA PARA TERMINAR LA FASE Y LATEMPERATURA Y LA PRESIN EXISTENTE EN LA CMARA.

    >FALHA AQUECIMEN..

    >TEMPO AQUEC.ULT.

    CICLO 6 12704:54 1,56ESTERILIZ. 127

  • 7/25/2019 Manual de Operaciones Autoclaves

    23/47

    23

    ALARMA 03

    SI LA PRESIN EST ARRIBA DE 20% CON RELACIN A LA TEMPERATURA DEL CICLOSELECCIONADO, DURANTE 6 SEGUNDOS EL SISTEMA EMITIR UN BIP INTERMITENTE Y ELMENSAJE ALARMA 03 SER EXHIBIDO EN EL DISPALY LCD, CENTELLEANDO LA PRIMERA

    LNEA.

    ALARMA 04

    SI LA PRESIN EST ABAJO DE 20% CON RELACIN A LA TEMPERATURA DEL CICLOSELECCIONADO, DURANTE 6 SEGUNDOS, EL SISTEMA EMITIR UN BIP INTERMITENTE Y ELMENSAJE ALARMA 4 SER EXHIBIDO EN LE VISOR, CENTELLEANDO LA PRIMERA LNEA.

    ALARMA 05

    SI LA TEMPERATURA EST ARRIAB DE 5% DE LA TEMPERATURA RELATIVA AL CICLOSELECCIONADO, DURANTE 6 SEGUNDOS, EL SISTEMA EMITIR UN BIP INTERMITENTE Y ELMENSAJE ALARMA 05 SER EXHIBIDO EN EL VISOR, CENTELLLEANDO LA PRIMERA LNEA

    ALARMA 06

    SI LA TEMPERATURA EST ABAJO DE 5% DE LA TEMPERATURA RELATIVA AL CICLOSELECCIONADO, DURANTE 6 SEGUNDOS, EL SISTEMA EMITIR UN BIP INTERMITENTE Y ELMENSAJE ALARMA 06 SER EXHIBIDO EN EL VISOR CENTELLEANDO LA PRIEMRA LNEA.

    >SOBRE-PRESSAO

    >SUB-PRESSAO

    >SOBRE-TEMPERATURA

    >SOB-TEMPERATURA

  • 7/25/2019 Manual de Operaciones Autoclaves

    24/47

    24

    DESCARGA DE AGUADESPUS DE LA FASE DE ESTERILIZACIN, TIENE INICIO LA FASE DE DESCARGA DE AGUA,EN ESTA FASE EL SISTEMA UTILIZA LA PRESIN INTERNA DE LA CMARA PARA IMPULSAREL AGUA DE VUELTA AL ALJIBE (VLVULA DE ENTRADA DE AGUA ABIERTA) ESTA FASE

    TERMINA CUANDO LA PRESIN ALCANZA 0,10 KGF/CM, CERRANDO LA VLVULA DE ENTRADADE AGUA EN EL VISOR SON EXHIBIDAS DOS PANTALLAS ALTERNADAS TEMPORIZADAS EN 1

    SEGUNDO.

    C C

    ESTAS PANTALLAS NOS MUESTRAN LA FASE DEL CICLO DESCARGA, EL NMERO DEL CICLOSELECCIONADO CICLO 6 Y LA TEMPERATURA Y PRESIN EXISTENTES EN LA CMARA

    ALARMA 08

    SI LA FASE DE DESCARGA DE AGUA ULTRAPASE EL TIEMPO DE 4 MINUTOS, EL SISTEMAEMITIR UN BIP INTERMITENTE Y EL MENSAJE ALARMA08 SER EXHIBIDO EN EL VISOR,CENTELLEANDO LA PRIMERA LNEA.

    SECADODESPUS DE LA FASE DE DESCARGA DE AGUA, SI EL CICLO SELECCIONADO ES EL CICLO 1,2,5 6 TIENE INICIO LA FASE DE SECADO. EL SISTEMA COMIENZA LA TEMPORIZACIN DE LAFASE DE SECADO CONFORME EL TIEMPO DE SECADO RELATIVO AL CICLO SELECIONADO,ABRIENDO LA VLVULA DE SALIDA DE VAPOR. SI ES EL CICLO 3,4 7, PASA PARA LA FASEFIN DE CICLO.

    ESTAS PANTALLAS NOS MUESTRAN LA FASE DEL CICLO SECADO, EL NMERO DEL CICLOSELECCIONADO CICLO 6. EL TIEMPO QUE FALTA PARA TERMINAR LA FASE Y LATEMPERATURA Y PRESIN EXISTENTE EN LA CMARA.

    CICLO 6 125AGUARDE . . . 0,97

    DESCARGA 125AGUARDE . . . 0 97

    >TEMPO DESCARG.ULT.

    CICLO 6 10104:54 0,00

    SECAGEM. 10104:55 0,00

  • 7/25/2019 Manual de Operaciones Autoclaves

    25/47

    25

    FIN DE CICLODESPUS DE LA FASE DE SECADO, TERMINA EL CICLO DE ESTERILIZACIN, SEALADO PORFIN DE CICLO. EL SISTEMA EMITE UN BIP CONTINUO POR 1 MINUTO, ABRIENDO LA VLVULADE SALIDA PARA IGUALAR LA PRESIN INTERNA A LA EXTERNA, Y EN VISOR SON

    EXHIBIDAS DOS PANTALLAS ALTERNADAS TEMPORIZADAS EN 1 SEGUNDO.

    ESTAS PANTALLAS NOS MUESTRAN LA FASE FIN DE CICLO , EL NMERO DEL CICLOSELECCIONADO CICLO 6, EL TIEMPO QUE FALTA PARA TERMINAR LA FASE Y LA

    TEMPERATURA Y PRESIN EXISTENTE EN LA CMARA.AL FINAL DEL CICLO, ABRA IMEDIATAMENTE SU PUERTA PUES SI L SE ENFRA CON LACMARA CERRADA CRIAR VACUO EN LA CMARA INTERNA, AN CON LA TRABA EN LAPOSICIN DESTRABADA NO SER POSIBLE ABRIR LA PUERTA, UTILICE ENTONCES LAFUNCIN DESPRESURIZACIN MANUAL.

    NOTAS IMPORTANTES

    EN EL CASO DE CADA DE ENERGA ELCTRICA DURANTE LA EJECUCIN DEL CICLO, ELSISTEMA REASUMIR LA EJECUCIN EN EL PUNTO INTERUMPIDO SI LOS PARMETROSFSICOS DE LOS PROCESOS FUERON ALCANZADOS, TAN PRONTO HAYA RESTABLECIMIENTODE LA MISMA.

    EL SISTEMA EN REPOSO, O SEA, FUERA DE EJECUCIN DE CICLO, A CADA MINUTO, ABRE LAVLVULA DE SALIDA DE VAPOR PARA IGUALAR LA PRESIN INTERNA A LA EXTERNA,DURANTE 30 SEGUNDOS, VOLVIENDO A CERRAR DESPUS DE ESTE TIEMPO.

    PARA CANCELARSE UN CICLO EN EJECUCIN ES

    NECESARIO PRESIONAR SIMULTNEAMENTE LA TECLAPARTIDA [P] E A TECLA SELECIN [S].

    CICLO 6 09500:34 0,00

    FIM DE CICLO 09500:35 0,00

  • 7/25/2019 Manual de Operaciones Autoclaves

    26/47

    26

    DESPRESURIZACIN MANUAL

    SI ACASO INTENTE ABRIR LA PUERTA DE LA CMARA Y ELLA MISMO CON LA TRABA EN LAPOSICIN TRABADA NO ABRE (VACUO), PRESIONE LA TECLA SELECCIN [S] HASTAAPARECER LA PANTALLA

    LUEGO, PRESIONE LA TECLA PARTIDA [P], PARA INICIAR LA DESPRESURIZACIN MANUAL, ELSISTEMA HAR LA DESPRESURIZACIN AUTOMTICAMENTE, ABRIENDO LA VLVULA DESALIDA DURANTE 4 MINUTOS. DESPUS DE 1 MINUTO YA SE PUEDE ABRIR LA PUERTA.

    EN EL DISPLAY LCD SE EXHIBE LA PANTALLA ENSEGUIDA. ELLA NOS MUESTRA EL CICLOMANUAL DESPRESURIZACIN, EL TIEMPO QUE FALTA PARA TERMINAR Y LA PRESINEXISTENTE EN LA CMARA.

    OJO: SI LA AUTOCLAVE TIENE AGUA EN LA CMARA INTERNA Y LA PRESIN INTERNAMAYOR QUE 0,50 KGF/CM, HAGA ANTES UNA DESCARGA DE AGUA MANUAL Y DESPUS LADESPRESURIZACIN MANUAL.

    DESCARGA DE AGUA MANUALSI DESPUS DE CANCELAR UN CICLO PROVOCADO POR UNA ALARMA O NO, SI DESEAELIMINAR AGUA DE LA CMARA, PRESIONE LA TECLA SELECCIN [S] HASTA APARECER LAPANTALLA:

    ENSEGUIDA PRESIONE LA TECLA PARTIDA [P], PARA COMENZAR LA DESCARGA DE AGUAMANUAL. NOTE QUE PARA QUE OCURRA LA DESCARGA ES NECESARIO QUE LA PRESININTERNA DE LA CMARA EST EN POR LO MENOS 0,50 kgf/cmY QUE LA DESCARGA TERMINACUANDO LA PRESIN LLEGA A 0,10 KG EN EL VISOR ES EXHIBIDA LA PANTALLA ENSEGUIDA.ELLA NOS MUESTRA EL CICLO MANUAL DESCARGA DE AGUA Y LA PRESIN EXISTENTE ENLA CMARA.

    RECOMENDACIN: SI LA AUTOCLAVE TIENE AGUA EN LA CMARA INTERNA Y LAPRESIN MENOR QUE0,50 KGF/CM, EJECUTE CUALQUIER CICLO DE 1 A 7 HASTA QUE LA

    FASE DE CALENTAMIENTO ALCANCE PRESIN ARRIBA DE 0,50 KGF/CM. LUEGO CANCELE EL

    SELECIONE:DESPRESSURIZA O

    PARTIDA:

    0250,00

    SELECIONE:DESCARGA DE GUA

    PARTIDA:

    0250 00

    DESPRESSURIZA O03:34 0,00

    DESCARGA DE GUA0 95

  • 7/25/2019 Manual de Operaciones Autoclaves

    27/47

    27

    CICLO PRESIONANDO LAS TECLAS PARTIDA [P] Y SELECCIN[S] AL MISMO TIEMPO, LUEGOEJECUTE DESCARGA MANUAL.

    PREVACUO EQUIPOS CON BOMBA DE VACUO (BV)CON LA ENTRADA Y SALIDA DE AGUA CERRADAS (MANDOS EFECTUADOS POR EL SISTEMADEL APARATO), SE ENCIENDE LA BOMBA DE VCUO DURANTE 2 MINUTOS GENERANDO

    VACUO EN LA CMARA INTERNA, CON CONTROL MEDIO DE 0,43 Kgf/cm E 0,68 kgf/(cm), OSEA, ENTRE 31,57 cm de Hg Y 50 cm de Hg.

    NOTE QUE SON EXHIBIDAS DOS PANTALLAS ALTERNADAS TEMPORIZADAS EN 1 SEGUNDO.

    OPERACIN

    ANTES DE CUALQUIER OPERACIN, RECORDAMOS QUE ELFUNCIONAMIENTO DEL APARATO SE DA NICAMENTE A TRAVS DELA TECLAS

    sPPARTIDA SELEO

    HAY MATERIALES QUE DEBIDO A SU NATURALEZA (CRISTALES/ LQUIDOS/ SINTTICOS),POSEEN SENSIBILIDAD A ALTA TEMPERATURA, NO PUDIENDO SUFRIR EL PROCESO DESECADO. SE LES RECOMIENDA LOS CICLOS 3,4 Y 7, QUE NO TIENEN EL PROCESO ARRIBAMENTADO. ATENCIN: SIEMPRE SE DEBE CONSULTAR LAS INFORMACIONES SOBRE LANATURALEZA Y LA RESISTENCIA DE LOS MATERIALES A SER ESTERILIZADOS CONSTANDOESTAS INFORMACIONES EN SUS EMBALAJES O PROSPECTOS, EVITANDO DE ESTA MANERAPERJUICIOS Y CONSERVANDO SU EQUIPO.

    PR V CUO 02501:06 -0.50

    CICLO 6 02501:05 -0.50

  • 7/25/2019 Manual de Operaciones Autoclaves

    28/47

    28

    VERIFIQUE SI EL APARATO EST CONECTADO AL ENCHUFE COMPATIBLE

    VERIFIQUE EL NIVEL DE AGUA DEL ALJIBE (SIEMPRE DESTILADA) VERIFIQUE COMO EST EMBALADO EL MATERIAL

    EL CICLO DE OPERACIN

    CORRESPONDE AL TIPO DE MATERIAL YEL MODO COMO ES EMBALADO

  • 7/25/2019 Manual de Operaciones Autoclaves

    29/47

    29

    COLOQUE EL MATERIAL A SER ESTERILIZADO EN LA CMARA INTERNA, NO REMUEVALOS ANAQUELES Y LA CHAPA DE PROTECCIN DEL DISRUPTOR (EL MATERIAL DEBETENER DIMENSIONES COMPATIBLES CON LA CAPACIDAS DE ALMACENAMIENTO DE LOSANAQUELES)

    EFECTE LA LIMPIEZA DE LA GUARNICIN DE LA PUERTA Y DE LA BASE DE LA CMARACON LCOHOL 70%

    CIERRE LA PUERTA CORRECTAMENTE (CONFORME PGINA 8) ENCIENDA LA LLAVE GENERAL QUE SE POSICIONA EN LA PARTE TRASERA DE LA

    AUTOCLAVE (VEA PGINA 5) SELECCIONA EL CICLO DESEADO (VEA TABLA PGINAS 31 Y 32) D PARTIDA, EL CICLO TRANCURRIR AUTOMTICAMENTE AL FINAL DEL CICLO EL MENSAJE END O FINAL DE CICLO APARECER EN EL VISOR,

    PERMITIENDO LA APERTURA DE LA PUERTA. MANTENGA LA PUERTA ENTREABIERTA POR CINCO MINUTOS. RETIRE EL MATERIAL

  • 7/25/2019 Manual de Operaciones Autoclaves

    30/47

    30

    TABLA REFERENCIAL DE CICLOSTemperatura mxima: 127 C

    SUPERFICIE ESPESSURA LIQIDOS ESTERILIZACIN SECADO

    1EN CAJAS DE AO INOXIDABLE PERFORADA; LIMAS/ BROCAS MATERIALSDE CORTE EM GENERAL 121121121121 20'20'20'20' 5'5'5'5'

    2

    ENVOLVIDOS EN PROTECIN O SOBRE PARILLAS DE AO INOXIDABLEPERFORADA (EMPAQUES QUIRURGICOS/ CAMPOS/ SOLAMENTE PARATERMOSSENSIBLES) BIOGRAFO ROTATIVA, INSTRUMENTAL CON CABLE DEPLSTICO/ ESPEJOS N 5/ LIMAS/ BROCAS/ MATERIALES DE CORTE EN

    GENERAL.

    121121121121 30'30'30'30' 10'10'10'10'

    3 GOMAS/ LIQIDOS(JAMS INFLAMABLES)/GUANTES DE LATEX 121121121121 30'30'30'30' 0000

    4

    CICLO DE EMERGNCIA, RECOMENDADO PARA MATERIALES QUE NOTIVERAM CONTACTO CON EL PACIENTE, MAS QUE SOLAMENTE TUVEVENCIMIENTO EN LE VALIDADE DE ESTERILIZACIN Y NECESITA SERESTERILIZADOS RAPIDAMENTE

    127127127127 5'5'5'5' 0000

    5INSTRUMENTALES ACOMODAOS EN CAJAS DE AO INOXIDABLEPERFURADO. 127127127127 10'10'10'10' 5'5'5'5'

    6ENVOLVIDOS EN EMPAQUES GRADO QUIRRGICO/ CAMPOS/INSTRUMENTABLES DE METAL EN GENERAL/ GASES/ ALGODN/ TECIDOS/VIDRIOS/ GOMAS.

    127127127127 12'12'12'12' 10'10'10'10'

    7 LIQIDOS/ VIDRIOS/ GOMAS (RESISTENTES A TEMPERATURA) 127127127127 15'15'15'15' 0000

    8ENVOLVIDOS EN PROTECIONES (EMPAQUES/ GRADO QUIRRGICO/ CAMPO)INSTRUMENTALES DE METAL EN GENERAL 127127127127 12'12'12'12' 15'15'15'15'

    SECSECADO COMPLEMENTAR - CASO EL MATERIAL AINDA PERMANEZAHMEDO (NO DEBE SER EFECTUADA EN VIDRIOS/ GOMASTERMOSENSIBLES Y LQUIDOS)

    ---- ---- 15'15'15'15'

    TERMOSENSIBLES - MATERIALES QUE NO SOPORTAN ALTAS TEMPERATURA

    NO TERMOSENSIBLES - MATERIALES QUE SUPORTAM ALTAS TEMPERATURA

    SUPERFCIE - MATERIAILES NO EMPAQUETADOS, EXPUESTO DIRECTAMENTE NO NECESITAN PENETRACINLIQUIDOS - MATERIALES LIQUIDOS CON RESERVATRIOS ABERTOS (JAMS INFLAMABLES)ESPESURA - MATERIALES ENVOLVIDOS EM PROTECIONES COMO EMPAQUES O CAMPOS, NECESITAN DE PROTCIN

    CONCULTE EL EMPAQUE EL MANULA DEL MATERIAL A SER ESTERILIZADO, VERIFICANDO SUS CARACTERISTICAS FSICAS

    Y LA SUYA RESISTNCIA AL PROCESO DE ESTERILIZAIN A VAPOR Y LA TEMPERATURA DEL CICLO ELIGIDO, SI NO TIVIER INFORMACIONES NA

    EMPAQUE, CONSULTE EL FABRICANTE DEL PRODUCTO

    JAMSUTILIZE PAPEL CRAFT OTRO EMPAQUE IMPERMEABLE AL VAPOR O QUE SE DEGENERE PELA UMIDADE Y LA TEMPERATURA

    DEL PROCESO PARA EMPAQUETAR O ENVOLVER LOS MATERIALES A SERN ESTERILIZADOS

    TABELA DE REFERNCIA PARA ESTERILIZACIN SIN BOMBA DE VACO

    CICLO MATERIALES A SEREM ESTERILIZADOSTRATAMIENTO

    CTIEMPOS

    TERMOSENS

    VEIS

    ESPECIFICACIONES

    OBS:

    NOTERMOSNSVEIS

  • 7/25/2019 Manual de Operaciones Autoclaves

    31/47

    31

    TABLA REFERENCIAL DE CICLOSMODELO: CON BOMBA DE VACO (BV)Temperatura mxima: 127 C

    SUPERFICIE ESPESSURA LIQIDOS ESTERILIZACIN SECADO

    1EN CAJAS DE AO INOXIDABLE PERFORADA; LIMAS/ BROCAS MATERIALSDE CORTE EM GENERAL 121121121121 20'20'20'20' 5'5'5'5'

    2

    ENVOLVIDOS EN PROTECIN O SOBRE PARILLAS DE AO INOXIDABLEPERFORADA (EMPAQUES QUIRURGICOS/ CAMPOS/ SOLAMENTE PARA

    TERMOSSENSIBLES) BIOGRAFO ROTATIVA, INSTRUMENTAL CON CABLE DEPLSTICO/ ESPEJOS N 5/ LIMAS/ BROCAS/ MATERIALES DE CORTE ENGENERAL.

    121121121121 30'30'30'30' 10'10'10'10'

    3 GOMAS/ LIQIDOS(JAMS INFLAMABLES)/GUANTES DE LATEX 121121121121 30'30'30'30' 0000

    4

    CICLO DE EMERGNCIA, RECOMENDADO PARA MATERIALES QUE NOTIVERAM CONTACTO CON EL PACIENTE, MAS QUE SOLAMENTE TUVEVENCIMIENTO EN LE VALIDADE DE ESTERILIZACIN Y NECESITA SERESTERILIZADOS RAPIDAMENTE

    127127127127 5'5'5'5' 0000

    5INSTRUMENTALES ACOMODAOS EN CAJAS DE AO INOXIDABLEPERFURADO. 127127127127 10'10'10'10' 5'5'5'5'

    6ENVOLVIDOS EN EMPAQUES GRADO QUIRRGICO/ CAMPOS/INSTRUMENTABLES DE METAL EN GENERAL/ GASES/ ALGODN/ TECIDOS/VIDRIOS/ GOMAS.

    127127127127 12'12'12'12' 10'10'10'10'

    7 LIQIDOS/ VIDRIOS/ GOMAS (RESISTENTES A TEMPERATURA) 127127127127 15'15'15'15' 0000

    8ENVOLVIDOS EN PROTECIONES (EMPAQUES/ GRADO QUIRRGICO/ CAMPO)INSTRUMENTALES DE METAL EN GENERAL 127127127127 12'12'12'12' 15'15'15'15'

    SECSECADO COMPLEMENTAR - CASO EL MATERIAL AINDA PERMANEZAHMEDO (NO DEBE SER EFECTUADA EN VIDRIOS/ GOMASTERMOSENSIBLES Y LQUIDOS)

    ---- ---- 15'15'15'15'

    TERMOSENSIBLES - MATERIALES QUE NO SOPORTAN ALTAS TEMPERATURA

    NO TERMOSENSIBLES - MATERIALES QUE SUPORTAM ALTAS TEMPERATURA

    SUPERFCIE - MATERIAILES NO EMPAQUETADOS, EXPUESTO DIRECTAMENTE NO NECESITAN PENETRACINLIQUIDOS - MATERIALES LIQUIDOS CON RESERVATRIOS ABERTOS (JAMS INFLAMABLES)

    ESPESURA - MATERIALES ENVOLVIDOS EM PROTECIONES COMO EMPAQUES O CAMPOS, NECESITAN DE PROTCIN

    CONCULTE EL EMPAQUE EL MANULA DEL MATERIAL A SER ESTERILIZADO, VERIFICANDO SUS CARACTERISTICAS FSICAS

    Y LA SUYA RESISTNCIA AL PROCESO DE ESTERILIZAIN A VAPOR Y LA TEMPERATURA DEL CICLO ELIGIDO, SI NO TIVIER INFORMACIONES NA

    EMPAQUE, CONSULTE EL FABRICANTE DEL PRODUCTO

    JAMSUTILIZE PAPEL CRAFT OTRO EMPAQUE IMPERMEABLE AL VAPOR O QUE SE DEGENERE PELA UMIDADE Y LA TEMPERATURA

    DEL PROCESO PARA EMPAQUETAR O ENVOLVER LOS MATERIALES A SERN ESTERILIZADOS

    TERMOS

    ENSVEIS

    ESPECIFICACIONES

    OBS:

    NOTERMOSN

    SVEIS

    TABELA DE REFERNCIA PARA ESTERILIZACIN SIN BOMBA DE VACO

    CICLO MATERIALES A SEREM ESTERILIZADOSTRATAMENTO

    CTIEMPOS

  • 7/25/2019 Manual de Operaciones Autoclaves

    32/47

    32

    TABLA REFERENCIAL DE CICLOSTemperatura mxima: 134 C (ENCOMENDA)

    SUPERFICIE ESPESSURA LIQIDOS ESTERILIZAO SECAGEM

    1EN CAJAS DE AO INOXIDABLE PERFORADA; LIMAS/ BROCAS MATERIALSDE CORTE EM GENERAL 121121121121 20'20'20'20' 5'5'5'5'

    2

    ENVOLVIDOS EN PROTECIN O SOBRE PARILLAS DE AO INOXIDABLEPERFORADA (EMPAQUES QUIRURGICOS/ CAMPOS/ SOLAMENTE PARATERMOSSENSIBLES) BIOGRAFO ROTATIVA, INSTRUMENTAL CON CABLE DEPLSTICO/ ESPEJOS N 5/ LIMAS/ BROCAS/ MATERIALES DE CORTE EN

    GENERAL.

    121121121121 30'30'30'30' 10'10'10'10'

    3 GOMAS/ LIQIDOS(JAMS INFLAMABLES)/GUANTES DE LATEX 121121121121 30'30'30'30' 0000

    4

    NOTERMOSN

    CICLO DE EMERGNCIA, RECOMENDADO PARA MATERIALES QUE NOTIVERAM CONTACTO CON EL PACIENTE, MAS QUE SOLAMENTE TUVEVENCIMIENTO EN LE VALIDADE DE ESTERILIZACIN Y NECESITA SERESTERILIZADOS RAPIDAMENTE

    127127127127 5'5'5'5' 0000

    5INSTRUMENTALES ACOMODAOS EN CAJAS DE AO INOXIDABLEPERFURADO. 127127127127 10'10'10'10' 5'5'5'5'

    6ENVOLVIDOS EN EMPAQUES GRADO QUIRRGICO/ CAMPOS/INSTRUMENTABLES DE METAL EN GENERAL/ GASES/ ALGODN/ TECIDOS/VIDRIOS/ GOMAS.

    127127127127 12'12'12'12' 10'10'10'10'

    7 LIQIDOS/ VIDRIOS/ GOMAS (RESISTENTES A TEMPERATURA) 127127127127 15'15'15'15' 0000

    8INSTRUMENTALES ACOMODAOS EN CAJAS DE AO INOXIDABLEPERFURADO. 134134134134 4'4'4'4' 7'7'7'7'

    9ENVOLVIDOS EN PROTECIONES (EMPAQUES/ GRADO QUIRRGICO/ CAMPO)INSTRUMENTALES DE METAL EN GENERAL 134134134134 7,5'7,5'7,5'7,5' 7'7'7'7'

    SECSECADO COMPLEMENTAR - CASO EL MATERIAL AINDA PERMANEZAHMEDO (NO DEBE SER EFECTUADA EN VIDRIOS/ GOMASTERMOSENSIBLES Y LQUIDOS)

    ---- ---- 15'15'15'15'

    TERMOSENSIBLES - MATERIALES QUE NO SOPORTAN ALTAS TEMPERATURA

    NO TERMOSENSIBLES - MATERIALES QUE SUPORTAM ALTAS TEMPERATURASUPERFCIE - MATERIAILES NO EMPAQUETADOS, EXPUESTO DIRECTAMENTE NO NECESITAN PENETRACINLIQUIDOS - MATERIALES LIQUIDOS CON RESERVATRIOS ABERTOS (JAMS INFLAMABLES)

    ESPESURA - MATERIALES ENVOLVIDOS EM PROTECIONES COMO EMPAQUES O CAMPOS, NECESITAN DE PROTCIN

    CONCULTE EL EMPAQUE EL MANULA DEL MATERIAL A SER ESTERILIZADO, VERIFICANDO SUS CARACTERISTICAS FSICAS

    Y LA SUYA RESISTNCIA AL PROCESO DE ESTERILIZAIN A VAPOR Y LA TEMPERATURA DEL CICLO ELIGIDO, SI NO TIVIER INFORMACIONES NA EN EL

    EMPAQUE, CONSULTE EL FABRICANTE DEL PRODUCTO

    JAMSUTILIZE PAPEL CRAFT OTRO EMPAQUE IMPERMEABLE AL VAPOR O QUE SE DEGENERE PELA UMIDADE Y LA TEMPERATURA

    DEL PROCESO PARA EMPAQUETAR O ENVOLVER LOS MATERIALES A SERN ESTERILIZADOS

    TABELA DE REFERNCIA PARA ESTERILIZACIN SIN BOMBA DE VACO

    CICLO MATERIALES A SEREM ESTERILIZADOSTRATAMENTO

    CTEMPOS

    TERMOSENS

    VEIS

    ESPECIFICACIONES

    OBS:

  • 7/25/2019 Manual de Operaciones Autoclaves

    33/47

    33

    TABLA REFERENCIAL DE CICLOSMODELO: CON BOMBA DE VACO (BV)Temperatura mxima: 134 C (ENCOMENDA)

    SUPERFICIE ESPESSURA LIQIDOS ESTERILIZAO SECAGEM

    1EN CAJAS DE AO INOXIDABLE PERFORADA; LIMAS/ BROCAS MATERIALSDE CORTE EM GENERAL 121121121121 20'20'20'20' 5'5'5'5'

    2

    ENVOLVIDOS EN PROTECIN O SOBRE PARILLAS DE AO INOXIDABLEPERFORADA (EMPAQUES QUIRURGICOS/ CAMPOS/ SOLAMENTE PARATERMOSSENSIBLES) BIOGRAFO ROTATIVA, INSTRUMENTAL CON CABLE DEPLSTICO/ ESPEJOS N 5/ LIMAS/ BROCAS/ MATERIALES DE CORTE ENGENERAL.

    121121121121 30'30'30'30' 5'5'5'5'

    3 GOMAS/ LIQIDOS(JAMS INFLAMABLES)/GUANTES DE LATEX 121121121121 30'30'30'30' 0000

    4

    NOTERMOSN

    CICLO DE EMERGNCIA, RECOMENDADO PARA MATERIALES QUE NOTIVERAM CONTACTO CON EL PACIENTE, MAS QUE SOLAMENTE TUVEVENCIMIENTO EN LE VALIDADE DE ESTERILIZACIN Y NECESITA SERESTERILIZADOS RAPIDAMENTE

    127127127127 5'5'5'5' 0000

    5INSTRUMENTALES ACOMODAOS EN CAJAS DE AO INOXIDABLEPERFURADO. 127127127127 10'10'10'10' 5'5'5'5'

    6ENVOLVIDOS EN EMPAQUES GRADO QUIRRGICO/ CAMPOS/INSTRUMENTABLES DE METAL EN GENERAL/ GASES/ ALGODN/ TECIDOS/VIDRIOS/ GOMAS.

    127127127127 12'12'12'12' 5'5'5'5'

    7 LIQIDOS/ VIDRIOS/ GOMAS (RESISTENTES A TEMPERATURA) 127127127127 15'15'15'15' 0000

    8 INSTRUMENTALES ACOMODAOS EN CAJAS DE AO INOXIDABLEPERFURADO. 134134134134 4'4'4'4' 5'5'5'5'

    9ENVOLVIDOS EN PROTECIONES (EMPAQUES/ GRADO QUIRRGICO/ CAMPO)INSTRUMENTALES DE METAL EN GENERAL 134134134134 7,5'7,5'7,5'7,5' 5'5'5'5'

    SECSECADO COMPLEMENTAR - CASO EL MATERIAL AINDA PERMANEZAHMEDO (NO DEBE SER EFECTUADA EN VIDRIOS/ GOMASTERMOSENSIBLES Y LQUIDOS)

    ---- ---- 15'15'15'15'

    TERMOSENSIBLES - MATERIALES QUE NO SOPORTAN ALTAS TEMPERATURA

    NO TERMOSENSIBLES - MATERIALES QUE SUPORTAM ALTAS TEMPERATURA

    SUPERFCIE - MATERIAILES NO EMPAQUETADOS, EXPUESTO DIRECTAMENTE NO NECESITAN PENETRACINLIQUIDOS - MATERIALES LIQUIDOS CON RESERVATRIOS ABERTOS (JAMS INFLAMABLES)

    ESPESURA - MATERIALES ENVOLVIDOS EM PROTECIONES COMO EMPAQUES O CAMPOS, NECESITAN DE PROTCIN

    CONCULTE EL EMPAQUE EL MANULA DEL MATERIAL A SER ESTERILIZADO, VERIFICANDO SUS CARACTERISTICAS FSICAS

    Y LA SUYA RESISTNCIA AL PROCESO DE ESTERILIZAIN A VAPOR Y LA TEMPERATURA DEL CICLO ELIGIDO, SI NO TIVIER INFORMACIONES NA EN EL

    EMPAQUE, CONSULTE EL FABRICANTE DEL PRODUCTO

    JAMSUTILIZE PAPEL CRAFT OTRO EMPAQUE IMPERMEABLE AL VAPOR O QUE SE DEGENERE PELA UMIDADE Y LA TEMPERATURA

    DEL PROCESO PARA EMPAQUETAR O ENVOLVER LOS MATERIALES A SERN ESTERILIZADOS

    TERMOSENSVEIS

    ESPECIFICACIONES

    OBS:

    TABELA DE REFERNCIA PARA ESTERILIZACIN SIN BOMBA DE VACO

    CICLO MATERIALES A SEREM ESTERILIZADOSTRATAMENTO

    CTEMPOS

  • 7/25/2019 Manual de Operaciones Autoclaves

    34/47

    34

    FUNCIONES ESPECIALES

    DESCARGA DE AGUA MANUAL (H2O):

    DESPUS DEL CANCELAMIENTO DE UN CICLO OCASIONADO POR UNA ALARMA O PORHABER PRESIONADO LAS TECLAS Y

    SIMULTNEAMENTE, USTED PODRR EXTRAEREL AGUA DE LA CMARA INTERNA, PARA ESO PRESIONE LA TECLA , HASTA QUEAPAREZCA EN EL DISPLAY EL MENSAJE H20 O Y PRESIONELA TECLA

    .AGUARDE HASTA QUE EL MENSAJE O SE APAGUE YVUELVA A MOSTRAR EL LTIMO CICLO EJECUTADO.(PARA VERIFICAR SI TODA AGUA FUE DRENADA, OBSERVE SI ELLA VOLVI A SU NIVELNORMAL DENTRO DEL ALJIBE).

    OJO: PARA QUE ESTE PROCEDIMIENTO TENGA RESULTADO SATISFACTORIO, LAPRESIN EN LA CMARA INTERNA DEBER SER DE POR LO MENOS 0,5KGF/cm.

    DESPRESURIZACIN MANUAL (OPEN):CASO HAYA NECESIDAD DE LA APERTURA DE LA PUERTA DESPUS DE ALGN CICLO, Y ESTAEST EMPACADA, SE DEBE DESPRESURIZAR LA CMARAINTERNA, PULSANDO LA TECLA HASTA APARECER EN EL DISPLAY EL MENSAJE OPENO DESPRESSURIZAO MANUAL,(DESPRESURIZACIN MANUAL). EL SISTEMAAUTOMTICAMENTE LIBERAR LA PUERTA.

    RESET:EN CASO DE CADAS DE ENERGA OCASIONALES O VARIACIONES DE LA MISMA, ELDISPLAY PUEDE DEMOSTRAR INFORMACIONES RARAS O CONGELAR, PUES EL SISTEMAINTENTA RECUPERAR LAS INFORMACIONES SOBRE EL CICLO REALIZADO ANTERIORMENTEAL PROBLEMA DE OSCILACIN DE ENERGA, DE ESTE MODO SE PUEDE PROCEDER ELRESTABLECIMIENTO DE LAS FUNCIONES.

    OPCIN 1:APAGUE LA LLAVE GENERAL DEL EQUIPO, PULSE LA TECLA Y VUELVA A

    ENCENDER EL EQUIPO (PULSANDO LA TECLA ), AGUARDE EL BIP Y QUE EL DISPLAY SEAPAGUE, LUEGO SUELTE LA TECLA, ESTE PROCEDIMIENTO EFECTA UNA LIMPIEZA ENSU MEMORIA, HACINDOLO FUNCIONAR DE MANERA CORRECTA.

    OPCIN 2: SI ACASO EL PROCEDIMIENTO ANTERIOR NO ALCANCE EL RESULTADOESPERADO, PRESIONE SIMULTNEAMENTE LAS TECLAS Y , ENCIENDA Y APAGUEEL APARATO EN LA LLAVE GENERAL EN MEDIA DE SIETE VECES, SOLTANDO LAS TECLASDESPUS DE ENCENDER Y APAGAR.

  • 7/25/2019 Manual de Operaciones Autoclaves

    35/47

    35

    MANTENIMIENTO PREVENTIVO

    PARA EFECTURSE CUALQUIER TIPO DE MANTENIMIENTO INTERNO EN NUESTRO EQUIPO,SE DEBE SEGUIR COMO EN EL DIBUJO ABAJO PARA LA RETIRADA DEL GABINETE,

    DEBIENDO SEGUIR EL GUIN ABAJO PARA UNA MEJOR PERFORMANCE DEL EQUIPO Y PARAQUE LAS CONDICIONES DE ORIGINALIDAD SEAN GARANTIZADAS.

  • 7/25/2019 Manual de Operaciones Autoclaves

    36/47

    36

    MANTENIMIENTO PREVENTIVO

    A DIARIO:

    EFECTE LA LIMPIEZA DE LA CONEXIN DE ESTANCO DE LA PUERTA (GUARNICIN) Y DELBORDE DE LA CMARA INTERNA UTILIZANDO LCOHOL 70% Y UNA FRANELA(RECOMENDAMOS QUE ESTE PROCEDIMIENTO SEA HECHO ANTE DE CADA CICLO A SEREJECUTADO).

    SEMANALMENTE:

    CAMBIE EL AGUA DESTILADA CONTENIDA EN EL ALJIBE Y HAGA UNA LIMPIEZA INTERNAADEMS DEL TEM ANTERIOR.

    MENSUALMENTE:

    LIMPIEZA Y DESOBSTRUCCIN DE LOS FILTROS Y VLVULAS, HAGA UN TEST DEFUNCIONAMIENTO DE LA VLVULA DE SEGURIDAD, PUJANDO SU ARGOLLA CINCO VECES,ADEMS DE LOS TEM ANTERIORES.

    SEMESTRALMENTE:

    LIMPIEZA Y DESOBSTRUCCIN DE LAS TUBERAS Y COMPONIENTES HIDRULICOS,VERIFICACIN DEL CIERRE DE LA PUERTA, ADEMS DE LOS TEM ANTERIORES.

    ANUALMENTE:DESIMPREGNACIN DE LOS ELEMENTOS HIDRULICOS, TEST Y EVALUACINHIDROSTTICA, COTEJO DE LOS INSTRUMENTOS DE PROTECCIN Y CONTROL,VALIDACIN DE LOS ELEMENTOS DE SEGURIDAD Y CONTROL, AJUSTE DEL SISTEMA DECIERRE, REAPRETADURA DE LOS ELEMENTOS DE FIJACIN

    OJO.: TODO Y CUALQUIER TIPO DE LIMPIEZA O MANTENIMIENTO,DEVE SER EFECTUADO CON EL EQUIPO DESCONECTADO DE LA RED

    ELCTRICA.

    EL NO-CUMPLIMIENTO INTEGRAL DE ESTES TEM PUEDE CAUSARFALLA EN EL PROCESO, DISMINUICIN CONSIDERABLE DE LA VIDA

    TIL DEL EQUIPO E INCLUSO ACCIDENTES.

  • 7/25/2019 Manual de Operaciones Autoclaves

    37/47

    37

    MANTENIMIENTO PREVENTIVO

    LA LIMPIEZA ES DE ESENCIAL IMPORTANCIA PARA SU EQUIPO.

    LOS RESIDUOS DE GRASA Y SUCIEDAD, PUEDEN SER FCILMENTE REMOVIDOS CON JABNNEUTRO Y AGUA, (PERO SIEMPRE DANDO NFASIS AL USO DEL ALCOHOL 70%) CONAUXILIO DE UN ESTROPAJO O CEPILLO BLANDO SIEMPRE QUE POSIBLE, RECORDANDO QUENUNCA SE DEBE USAR ESTROPAJO METLICO, LO QUE PUEDE GENERAR DETRITOS YCAUSAR INCRUSTACIONES, SIENDO ACELERADOS POR LAS CONDICIONES TRMICAS EN LACMARA, ADEMS DE GENERAR RAYAS EN LA CMARA QUE ALMACENARAN SUCIEDAD. LACMARA DEBE SER TOTALMENTE ENJUGADA Y SECADA DESPUS DEL LAVADO.

    LIMPIEZAS A DIARIO MANTENDRN LA SUPERFICIE BRILLANTE Y AYUDARN A PREVENIREL COMIENZO DE UN PROCESO DE CORROSIN GALVNICA

    CONTACTOS CON METALES DIFERENTES DEBEN SER EVITADOS SIEMPRE QUE POSIBLE,ESTO AYUDAR A PREVENIR LA CORROSIN GALVNICA, CUANDO ESTN PRESENTESSOLUCIONES CIDAS O SALINAS.

    EL AGUA UTILIZADA PARA LA ALIMENTACIN DEL EQUIPO DEBE SER DESTILADA OEXENTA DE CLORETOS, FLUORETOS, OTROS SALES MINERALES Y METALES, PUES

    ARRASTRAN IMPUEREZAS PROVENIENTES DE LAS TUBERAS Y ALJIBES QUE HARNDEPSITOS PRINCIPALMENTE EN LA CMARA INTERNA, PUDIENDO ATACARLA YDISMINUIR SU VIDA TIL.

    CON LA LIMPIEZA CORECTA, ESTES DEPSITOS SERN REMOVIDOS YEL PROCESO DE CORROSIN SER INHIBIDO, GARANTIZANDO LA

    LONGEVIDAD DE SU CMARA.

  • 7/25/2019 Manual de Operaciones Autoclaves

    38/47

    38

    MANTENIMIENTO PREVENTIVO

    EL FILTRO TIENE LA FUNCIN DE IMPEDIR QUE DETRITOS Y RESIDUO SEAN IMPELIDOS ALSISTEMA HIDRULICO DEL EQUIPO, LO QUE PUEDE OCASIONAR TUPICIONES Y OTROSDAOS LOCALIZADO DENTRO DE LA CMARA INTERNA (FIG 8) L PUEDE SER RETIRADO PARALIMPIEZA CON GRAN FACILIDAD Y PRACTICIDAD, ES SLO REMOVERLO, TOMANDO LASDEBIDAS PRECAUCIONES.

    CMARA INTERNA

    DISRUPTOR

    FILTRO

    PT 100

    OJO: SE PUEDE USAR PARAAUXILIO DE LA LIMPIEZA, UN

    CEPILLO PLSTICO PARA LADESEMPREGNACIN DE

    ELEMENTOS MS RGIDOS,QUE PUEDAN FIJARSE EN L.

    1 --DESCONECTAR EL EQUIPO DE LA RED

    ELCTRICA

    2 --VERIFICAR SI EL DISRUPTOR SE

    ENCUENTRA EN TEMPERATURA ELEVADA.

    3 -- SI LOS PROCEDIMIENTOS 2 Y 3

    FUERON REALIZADOS CON XITO,

    BASTA PUJAR EL FILTRO COMOMUESTRA LA FIGURA ARRIBA

  • 7/25/2019 Manual de Operaciones Autoclaves

    39/47

    39

    MANTENIMIENTO PREVENTIVO

    CAMBIE EL AGUA DEL ALAJIBE SEMANALMENTE

    TORNILLO DE DRENAJEDEL ALJIBE

    A

    SERCON

    S PPARTIDASELEO

    AMPLIAO DO FILTRO

    CMARA INTERNA

    RESISTNCIA

    COLOQUE UN RECIPIENTE EN LA BASE DELTORNILLO DEL DREN.

    RETIRE EL TORNILLO Y AGUARDE TODO ELDESAGE DEL AGUA DEL ALJIBE.

    COLOQUE NUEVAMENTE EL TORNILLO YLIMPIE EL ALJIBE CON LCOHOL 70 %.

  • 7/25/2019 Manual de Operaciones Autoclaves

    40/47

    40

    MANTENIMIENTO PREVENTIVO

    SISTEMA DE CIERRE, TRABAMIENTO Y ESTANCACIN DE LA PUERTA (AHMC

    13)SON DE EXTREMA IMPORTANCIA PARA GARANTIZAR LAS CONDICIONES DE SEGURIDAD EN

    EL PROCESO. MISMA CON EL CICLO REALIZADO ABAJO PRESIN CONSIDERADA BAJA, LA

    FUERZA APLICADA EN LA EXTENSIN DEL REA DE LA TAPA ES MUY ALTA.

    PARA TANTO ES IMPRESCINDIBLE QUE EL CIERRE Y LA APERTURA DE LAS TAPAS OCURRAM

    SIEMPRE EN CONDICONES CONFORME LAS DE PROYECTO

    LUBRICACIN

    LA LUBRICACIN DE LAS PARTES QUE COMPONEN EL SISTEMA DE CIERRE DE LASPUERTAS ES DE EXTREMA IMPORTANCIA, PUES LOS DEPSITOS DE SUCIEDAD Y DEGRASA DESCOMPUESTA PUEDEN FCILMENTE DAAR EL ACOPLAMIENTO ENTRE LASPARTES MVILES Y DE PRECISIN

    EJE CENTRAL

    PRA EFECTUARSE LA LUBRICACIN DEL EJE CENTRAL EN LA PUERTA (EJE SIN FIN) SE DEBE

    REMOVER TODOS LOS RESIDUOS DE SUCIEDAD Y DE GRASA YA UTILIZADA A FIN DE QUE SE

    MANTENGA LAS PARTES LIMPIAS Y LA UTILIZACIN DE LA GRASA MOLIKOTE ML G1-2 PARA

    UNA NUEVA LUBRICACIN

    BISAGRAS

    PARA EFECTUARSE LA LUBRICACIN DE LAS BISAGRAS, SE DEBE REMOVER TODOS LOSRESIDUOS DE SUCIEDAD Y DE GRASA YA UTILIZADA A FIN DE QUE SE MANTENGA LASPARTES LIMPIAS Y LA UTILIZACIN DE LA GRASA MOLIKOTE ML G1-2 PARA UNA NUEVALUBRICACIN..

    ASTAS DE CIERRE

    LAS ASTAS DE CIERRE TIENEN LA FUNCIN DE EJERCER UNA FUERZA DE APRETADURA DE LA

    TAPA EN LA GUARNICIN A FIN DE QUE IMPIDA EL PASAJE DE AIRE DE LA CMARAINTERNA AL AMBIENTE EXTERNO.o EFECTE LA REAPRETADURA DE LA ESFERA DE MOVIMIENTO DE LAS

    ASTAS DE CIERRE.o VERIFIQUE LA HOMOGENEIDAD DEL CIERRE DE LAS ASTAS A FIN DE

    QUE SE OBTENGA UNA HOMOGENEIDAD EN LA ESTANCACIN.EFECTE LA REAPRETADURA DE LOS TORNILLOS DE FIJACIN DE LOS CALZOS DE AJUSTEDE LAS ASTAS

  • 7/25/2019 Manual de Operaciones Autoclaves

    41/47

    41

    ANTES DE LLAMAR LA ASISTENCIA TCNICA

    ALARMA 0

    PUEDE SER OCASIONADO POR MAL TRABAMIENTO DE LA PUERTA

    CIERRE LA PUERTA DEL EQUIPO CORRECTAMENTE E INTENTE NUEVAMENTE. EN LA PERSISTENCIA DEL PROBLEMA LLAME LA ASISTENCIA TCNICA

    ALARMA 1

    PUEDE SER OCASIONADO POR FALTA DE AGUA EN EL ALJIBE

    VERIFIQUE EL NIVEL DE AGUA DEL ALJIBE E INTENTE NUEVAMENTE

    EN LA PERSISTENCIA DEL PROBLEMA LLAME LA ASISTENCIA TCNICAALARMA 2

    LLAME LA ASISTENCIA TCNICA

    ALARMA 3

    LLAME LA ASISTENCIA TCNICA

    ALARMA 4PUEDE SER OCASIONADO POR SUCIEDAD EN LA GUARNICN DE LA PUERTA

    EFECTE LA LIMPIEZA DE LA GUARNICIN DE LA PUERTA E INTENTENUEVAMENTE.

    EN LA PERSISTENCIA DEL PROBLEMA LLAME LA ASISTENCIA TCNICA

    ALARMA 5

    PUEDE SER OCASIONADO POR SUCIEDAD EN LA GUARNICN DE LA PUERTA

    EFECTE LA LIMPIEZA DE LA GUARNICIN DE LA PUERTA E INTENTENUEVAMENTE.

    EN LA PERSISTENCIA DEL PROBLEMA LLAME LA ASISTENCIA TCNICA

  • 7/25/2019 Manual de Operaciones Autoclaves

    42/47

    42

    ALARMA 6

    LLAME LA ASISTENCIA TCNICA

    ALARMA 7

    PUEDE HABER SIDO OCASIONADO POR OSCILACIONES EN LA ALIMENTACIN ELCTRICA,DANDO CIERTA INSTABILIDAD AL EQUIPO, SE DEBE CANCELAR EL CICLO. SI EL PROBLEMAPERSISTIR ES ACONSEJABLE LA INSTALACIN DE UN ESTABILIZADOR DE TENSIN(RECUERDE QUE LA FUENTE DE ALIMENTACIN DEBE SER PRIORITARIA, O SEA, NO DEBESER COMPARTIDA CON LADRN O CADENA DE APARATOS).EN CASO DE HABERSE INTERRUMPIDO UN CICLO Y EL EQUIPO EST CON AGUA DENTRO DELA CMARA INTERNA, SE DEBE SELECCIONAR NUEVAMENTE EL CICLO QUE ESTABA

    ANTERIORMENTE Y DEJAR QUE EL MISMO ALCANCE APROXIMADAMENTE 0,70 Kgf/cm DEPRESIN, CANCELNDOLO (PULSE SIMULTNEAMENTE Y , LUEGO SELECCIONE H20 (DESCARGA) PARA QUE EL AGUA VUELVA AL ALJIBE

    EN LA PERSISTENCIA DEL PROBLEMA LLAME LA ASISTENCIA TCNICA

    ALARMA 8

    PUEDE SER OCASIONADO POR SUCIEDAD EN EL FILTRO DE DRENAJE EFECTE LA LIMPIEZA DEL FILTRO E INTENTE NUEVAMENTE EN LA PERSISTENCIA DEL PROBLEMA LLAME LA ASISTENCIA TCNICA

    ALARMA 9

    PUEDE HABER SIDO OCASIONADO POR HABERSE DADO LA PARTIDA EN EL CICLO CON LATEMPERATURA DE LA CMARA INTERNA ARRIBA DE 70C

    AGUARDE EL ENFRIAMIENTO DEL EQUIPO E INTENTE NUEVAMENTE EN LA PERSISTENCIA DEL PROBLEMA LLAME LA ASISTENCIA TCNICA

    ALARMA 10

    LLAME LA ASISTENCIA TCNICA

    ALARMA 11

    LLAME LA ASISTENCIA TCNICA

  • 7/25/2019 Manual de Operaciones Autoclaves

    43/47

    43

    ALARMAS

    ALARMAS

    POSIBLESCAUSAS

    C

    ERRAMENTOD

    ELAPUERT

    M

    ICROD

    ALAPUERTA

    S

    ENSOR

    DETEMPERATURA

    S

    ENSOR

    DEPRESIN

    S

    ALIDAVALV.AR

    S

    ALIDAVALV.

    GUA

    S

    ALIDAREL

    O

    BSTRUCCIN

    DELDRENO

    O

    BSTRUCCIN

    DELFILTRO

    N

    VELDELRESERVATRIO

    R

    EL

    T

    ENSIN

    INCOMPATIBLE

    LLAVEGENERALROTA

    R

    ESISTENCIAQUEMADA

    E

    MPAQUETADURA

    V

    ALV./FUSIBLESDESEGUR

    T

    EMPERATURADEPARTIDA

    V

    ALV.ENTRADAGUA

    V

    ALV.SALIDADEAR

    P

    LUGC

    ON

    MALOC

    ONTACT

    ALARMA 00FALHANOSENSORDAPORTA V V

    ALARMA 01 FALTADEGUAOUFALHANOSENSOR V V V V V V V V V V

    ALARMA 02FALTADEAQUECIMENTOOUSENSOR V - - -

    ALARMA 03SOBREPRESSO V V V V V V V V

    ALARMA 04SOBPRESSO V V V V V V -

    ALARMA 05SOBRETEMPERATURA V V - V -

    ALARMA 06SOBTEMPERATURA V V V

    ALARMA 07 TEMPODEAQUECIMENTOULTRAPASSADO V V V V V - V V V V

    ALARMA 08TEMPODEDESCARGAULTRAPASSADO V V V V V

    ALARMA 09FALHANORESFRIAMENTO V V V V

    ALARMA 10 SENSORDEPRESSO -

    ALARMA 11 SENSORDETEMPERATURA V

    LIGALALLHAVEGENERALYNADAACURRE V V

    LIGALALLHAVEYDISPLAYPARPODEANDO V V VPRESENADELVAPORSALINDOPOR GABINETE V

    AGUA/ VAPORESCAPANDOPORLAPUERTA - - V

    RESTODEGUAENELFONDODELACAMARA V - - V -

  • 7/25/2019 Manual de Operaciones Autoclaves

    44/47

    44

    ESPECIFICACIONES TCNICAS

    MODELO PESOKg CAMARAC

    MXTENSO

    VCAPACIDADE

    LPOTENCIA

    WCONSUMO

    kWhDESJUNTOR

    ACABLE

    mm

    AHM 35 ALUMINIO 127 110 21 1700 1,5 20 4,0

    AHM 35 INOX127ou134

    110 21 1700 1,5 20 4,0

    AHM 35 ALUMINIO 127 220 21 1700 1,5 10 2,5

    AHM 35 INOX127ou

    134

    220 21 1700 1,5 10 2,5

    AHM (BV) 35 INOX127ou134

    220 21 1700 1,5 12 4,0

    AHM 45 INOX127ou134

    220 25 1700 1,5 10 2,5

    AHM 80 INOX127ou134

    110 42 2300 2,0 25 4,0

    AHM 80 INOX127ou134

    220 42 2300 2,0 13 4,0

    AHM (BV) 80 INOX127ou134

    220 42 2300 2,0 15 4,0

    AHM (BV) 100 INOX127ou134

    220 54 4500 4,0 25 10,0

  • 7/25/2019 Manual de Operaciones Autoclaves

    45/47

    45

    DIMENCIONES

    IDENTIFICACIN DEL PRODUCTO

    CAPACIDADE ALTURA EXT. LARGURA EXT. PROF. EXT. DIAM. INT. PROF. INT.

    21 LITROS 465 mm 540 mm 500 mm 300 mm 300 mm

    25 LITROS 465 mm 540 mm 500 mm 300 mm 370 mm

    42 LITROS 465 mm 550 mm 800 mm 300 mm 600 mm

    54 LITROS 465 mm 610 mm 1000 mm - 600 mm

    FECHA DE FABRICACIN

    NMERO DE FABRICACIN

    MXIMA PRESIN DE TRABAJO ADMISIBILE

    MXIMA TEMPERATURA ADMISIBILE

    POTENCIA

    VOLTAGEM

    PRODUCIN DE VAPOR

  • 7/25/2019 Manual de Operaciones Autoclaves

    46/47

    46

    TRMINO DE GARANTALa Sercon Indstria e Comercio de Aparelhos Mdicos e Hospitalares Ltda, concede la garanta de este equipo, por ellafabricado y comercializado; por el perodo de 270 das, calculados a partir del fin del plazo legal de 90 das de la fechade emisin de factura al consumidor, completando un total de 360 das, desde que el mismos haya sido instalado y

    utilizado acorde con las orientaciones contenidas en el manual de instrucciones.Durante el perodo establecido, la garanta cubre totalmente la mano de obra y piezas en el reparo de defectosdebidamente constatados como siendo de fabricacin. Solamente un tcnico autorizado por el fabricante esthabilitado a reparar defectos cubiertos por la garanta, por medio de presentacin de la factura original.

    El consumidor tiene el plazo de 90 das para reclamar de irregularidades (vicios) aparentes, de fcil e inmediataobservacin en el producto, como los tem que constituyen la parte externa y cualquier otra accesible al usuario.Piezas expuestas al desgaste natural, piezas plsticas y accesorios en general tienen garanta restricta al plazo legal de90 das

    LA GARANTA CUBRE:

    o Eventuales defectos del material empleado en la fabricacin del equipoo Eventuales defectos de funcionamiento y performance mecnica originados de la fabricacin.

    LA GARANTA NO CUBRE:

    o La instalacin o utilizacin del producto en desacuerdo a las recomendaciones del manual del propietario.o Averas causadas por la falta correcta de la conservacin del aparato acorde con las orientaciones del

    manual (por lo tanto lalo atentamente antes de usar el equipo)o Averas resultantes de la impericia del operador, uso indebido, cadas, agente de naturaleza y accidentes de

    cualquier naturaleza.o Componientes expuestos a desgaste natural por el uso (guarniciones, retenedores, etco Daos causados por la falta de mantenimeinto, limpieza y conservacino Desempeo insatisfactorio del producto debido a instalaciones o red elctrica inadecuadaso Motores elctricos, llaves elctricas y dems componientes de la cadena elctrica.o Transporte y remocin del producto para reparo.

    LA GARANTA PIERDE SU EFECTO:

    o Asistencia tcnica prestada por personas no autorizadaso La no observacin de las prescripciones constantes del manual de instrucciones, funcionamiento,

    instalaciones y mantenimiento preventivo.o Introduccin de modificacin y usos de accesorios impropios y no originales.o Por la suspensin o atraso en los pagos de los plazos vencidoso Fin del plazo de validez de la garanta.

    GENERALIDADES:

    o La garanta tendr validez por el plazo arriba especificado, calculado a partir de la fecha de adquisicin porel primer comprador/consumidor, mismo que la propiedad del producto sea transferida.

    o Las piezas sustituidas sin deuda, en decurso de la garanta, son de propiedad del fabricante.o Est a criterio del fabricante, la eleccin del sitio de ejecucin del trabajo, siendo de responsabilidad del

    cliente los gastos de traslado hasta la asistencia tcnica.o El fabricante no se responsabiliza por perjuicios en decurso de la paralizacin de la mquina o accidente.Son de responsabilidad del reclamante los gastos en decurso de atendimiento para llamadas de asistencia tcnica

    cuando sean juzgadas improcedentes

  • 7/25/2019 Manual de Operaciones Autoclaves

    47/47

    La SERCON se reserva el derecho de alterardatos tcnicos de este manual en funciones deadecuaciones para la necesidad de mercado.

    Sercon Indstria e Comrcio de Aparelhos Mdico Hospitalares Ltda.08770-040 R. Ten. Onofre Rodrigues de Aguiar, 1201 Vila Industrial Mogi das

    Cruzes - SPTelefone: 55.11.4721.1733 Fax: 55.11.4721.1733 r. 240