mlts / mlcs Fast test of medical luminaires The medical lighting test (mlts) and calibration (mlcs) systems are designed for fast and easy testing and cali- bration of medical luminaires or luminaire components during their production process. Based on a luminance measurement system the mlts / mlcs is able to measure illuminance distributions with high resolution. The pho- tometric and geometric measurement of the light field is done within seconds and tested against tolerances. Approval against values requested by regulations like DIN EN 60601-2-41 medical lighting or DIN EN ISO 9680 dentistry lighting can be verified. In case the luminaires need to be adjusted or calibrated additionally, the mlcs offers a wide range of possibilities to adjust different work items as for example light field diameter or illumi- nance. Both systems, the mlts and the mlcs, are easily expandable by a spectrometer element (-spec), which enables the systems to test and calibrate all relevant colorimetric parameters such as color temperature, chromaticity coordinates and color rendering indices. www.opsira.com/mlts
Transcript
Page 1: mlts / mlcs - opsira · mlts / mlcs Fast test of medical luminaires The medical lighting test (mlts) and calibration (mlcs) systems are designed for fast and easy testing and cali-bration

mlts / mlcsFast test of medical luminaires

The medical lighting test (mlts) and calibration (mlcs)

systems are designed for fast and easy testing and cali-

bration of medical luminaires or luminaire components

during their production process. Based on a luminance

measurement system the mlts / mlcs is able to measure

illuminance distributions with high resolution. The pho-

tometric and geometric measurement of the light field

is done within seconds and tested against tolerances.

Approval against values requested by regulations like

DIN EN 60601-2-41 medical lighting or DIN EN ISO 9680

dentistry lighting can be verified. In case the luminaires

need to be adjusted or calibrated additionally, the mlcs

offers a wide range of possibilities to adjust different

work items as for example light field diameter or illumi-

nance. Both systems, the mlts and the mlcs, are easily

expandable by a spectrometer element (-spec), which

enables the systems to test and calibrate all relevant

colorimetric parameters such as color temperature,

chromaticity coordinates and color rendering indices.

www.opsira.com/mlts

Page 2: mlts / mlcs - opsira · mlts / mlcs Fast test of medical luminaires The medical lighting test (mlts) and calibration (mlcs) systems are designed for fast and easy testing and cali-bration

Typical values of standard configuration Changes are possible depending on the system configuration. Variations to the technical data may occur due to the permanent improvement and development of our measurement systems. We do not assume any juristic responsibility or liability whatsoever for such variations or misprints. The General Terms and Conditions of Trade of the opsira GmbH are valid. mlts I mlcs· E · V00103014 · © opsira GmbH · www.opsira.com/mlts. · www.koellekunter.de

Measuring quantitiesMaximum illuminance EcPosition of the max. illuminance in Cartesian and / or polar coordinatesD50 illuminance and diameterD10 illuminance and diameterRatio D50 to D10Color rendering indices Ra, R1 to R14Correlated color temperatureColor coordinates xy, u'v'Irradiance EeUV-irradiance EuvRatio Ee / Ec

DimensionsL 1800 x W 1500 x H 2100 mmWeight 250 kg

AccessoriesClient-Softwareluca Softwarespec'3 softwareSample holdersSample heatingSample cooling