PID ANALYZERS, LLC SANDWICH, MA HTTP://WWW.HNU.COM Model 115 Portable FID 1 PID Analyzers, LLC 2012
Transcript
Page 1: Model 115 portable fid presentation 812

P I D A N A L Y Z E R S L L C

S A N D W I C H M A

H T T P W W W H N U C O M

Model 115 Portable FID 1

PID Analyzers LLC 2012

Introduction

The Model 115 is a small hand held analyzer

Smallest amp Lightest FID on the market today- analyzer (2) hydride tank (22)- most are 11-12

Safe low pressure H2 storage as a metal hydride (gt 60 hours of operation)

Bluetooth enabled

H2 flow regulator is stable to +- 1-2 through the pressure range-keeping the operating T and sensitivity constant

2

PID Analyzers LLC 2012

The Model 115 is Easy to use

ThermoFoxboro- TVA 1000-FID 27 step startup procedure

PID-115- 8 steps with Calibration Turn on H2 for 30 sec by opening the H2 tank valve Turn on analyzer-Press the On button Press red button (piezo starter) to ignite Check T reading should be increasinf if the flame is litb (takes about

20 min to reach operating T) Press cal button Cal with zero gas (1 min) Cal with Cal gas (1 min) Ready to operate amp calibrated

3

PID Analyzers LLC 2012

Description Photo

The fill system consists of a customer H2 (99999) Tank with a regulator capable of setting to 250 PSI and a fluidics cable (supplied) with a quick disconnect connector at one end and a frac14rdquo connector at the other end (H2 tank)

Easy to fill Cylinder 4

PID Analyzers LLC 2012

Model 115 FID H2 Tank

Model 115 FID amp Tank 5

PID Analyzers LLC 2012

Detail of Model 115 Analyzer 6

PID Analyzers LLC 2012

FID and Belt Mounted Tank FID Belt Clip

Accessories for the 115 7

PID Analyzers LLC 2012

Shoulder strap for the 115 is also available

8

PID Analyzers LLC 2012

Tank Pressure vs H2 Flow

PID Analyzers LLC 2012

9

0

5

10

15

20

25

0 50 100 150 200 250

H2

Flo

w

Tank Pressure (PSI)

Effect of flow on Tank Presure

Linearity of Model 115

0

500

1000

1500

2000

2500

3000

3500

4000

4500

0 500 1000 1500 2000 2500 3000 3500 4000 4500

FID

Re

ad

ing

115

ppm Methane

FID Linearity

10

PID Analyzers LLC 2012

Schematic of FID

PID Analyzers LLC 2012

11

1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC

waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection

httpwwwhnucom

What can I do with a portable FID analyzer

115 Specifications

PID Analyzers LLC 2012

13

Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4

  • Model 115 Portable FID
  • Introduction
  • The Model 115 is Easy to use
  • Easy to fill Cylinder
  • Model 115 FID amp Tank
  • Detail of Model 115 Analyzer
  • Accessories for the 115
  • Slide Number 8
  • Tank Pressure vs H2 Flow
  • Linearity of Model 115
  • Schematic of FID
  • 115 Specifications
Page 2: Model 115 portable fid presentation 812

Introduction

The Model 115 is a small hand held analyzer

Smallest amp Lightest FID on the market today- analyzer (2) hydride tank (22)- most are 11-12

Safe low pressure H2 storage as a metal hydride (gt 60 hours of operation)

Bluetooth enabled

H2 flow regulator is stable to +- 1-2 through the pressure range-keeping the operating T and sensitivity constant

2

PID Analyzers LLC 2012

The Model 115 is Easy to use

ThermoFoxboro- TVA 1000-FID 27 step startup procedure

PID-115- 8 steps with Calibration Turn on H2 for 30 sec by opening the H2 tank valve Turn on analyzer-Press the On button Press red button (piezo starter) to ignite Check T reading should be increasinf if the flame is litb (takes about

20 min to reach operating T) Press cal button Cal with zero gas (1 min) Cal with Cal gas (1 min) Ready to operate amp calibrated

3

PID Analyzers LLC 2012

Description Photo

The fill system consists of a customer H2 (99999) Tank with a regulator capable of setting to 250 PSI and a fluidics cable (supplied) with a quick disconnect connector at one end and a frac14rdquo connector at the other end (H2 tank)

Easy to fill Cylinder 4

PID Analyzers LLC 2012

Model 115 FID H2 Tank

Model 115 FID amp Tank 5

PID Analyzers LLC 2012

Detail of Model 115 Analyzer 6

PID Analyzers LLC 2012

FID and Belt Mounted Tank FID Belt Clip

Accessories for the 115 7

PID Analyzers LLC 2012

Shoulder strap for the 115 is also available

8

PID Analyzers LLC 2012

Tank Pressure vs H2 Flow

PID Analyzers LLC 2012

9

0

5

10

15

20

25

0 50 100 150 200 250

H2

Flo

w

Tank Pressure (PSI)

Effect of flow on Tank Presure

Linearity of Model 115

0

500

1000

1500

2000

2500

3000

3500

4000

4500

0 500 1000 1500 2000 2500 3000 3500 4000 4500

FID

Re

ad

ing

115

ppm Methane

FID Linearity

10

PID Analyzers LLC 2012

Schematic of FID

PID Analyzers LLC 2012

11

1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC

waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection

httpwwwhnucom

What can I do with a portable FID analyzer

115 Specifications

PID Analyzers LLC 2012

13

Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4

  • Model 115 Portable FID
  • Introduction
  • The Model 115 is Easy to use
  • Easy to fill Cylinder
  • Model 115 FID amp Tank
  • Detail of Model 115 Analyzer
  • Accessories for the 115
  • Slide Number 8
  • Tank Pressure vs H2 Flow
  • Linearity of Model 115
  • Schematic of FID
  • 115 Specifications
Page 3: Model 115 portable fid presentation 812

The Model 115 is Easy to use

ThermoFoxboro- TVA 1000-FID 27 step startup procedure

PID-115- 8 steps with Calibration Turn on H2 for 30 sec by opening the H2 tank valve Turn on analyzer-Press the On button Press red button (piezo starter) to ignite Check T reading should be increasinf if the flame is litb (takes about

20 min to reach operating T) Press cal button Cal with zero gas (1 min) Cal with Cal gas (1 min) Ready to operate amp calibrated

3

PID Analyzers LLC 2012

Description Photo

The fill system consists of a customer H2 (99999) Tank with a regulator capable of setting to 250 PSI and a fluidics cable (supplied) with a quick disconnect connector at one end and a frac14rdquo connector at the other end (H2 tank)

Easy to fill Cylinder 4

PID Analyzers LLC 2012

Model 115 FID H2 Tank

Model 115 FID amp Tank 5

PID Analyzers LLC 2012

Detail of Model 115 Analyzer 6

PID Analyzers LLC 2012

FID and Belt Mounted Tank FID Belt Clip

Accessories for the 115 7

PID Analyzers LLC 2012

Shoulder strap for the 115 is also available

8

PID Analyzers LLC 2012

Tank Pressure vs H2 Flow

PID Analyzers LLC 2012

9

0

5

10

15

20

25

0 50 100 150 200 250

H2

Flo

w

Tank Pressure (PSI)

Effect of flow on Tank Presure

Linearity of Model 115

0

500

1000

1500

2000

2500

3000

3500

4000

4500

0 500 1000 1500 2000 2500 3000 3500 4000 4500

FID

Re

ad

ing

115

ppm Methane

FID Linearity

10

PID Analyzers LLC 2012

Schematic of FID

PID Analyzers LLC 2012

11

1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC

waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection

httpwwwhnucom

What can I do with a portable FID analyzer

115 Specifications

PID Analyzers LLC 2012

13

Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4

  • Model 115 Portable FID
  • Introduction
  • The Model 115 is Easy to use
  • Easy to fill Cylinder
  • Model 115 FID amp Tank
  • Detail of Model 115 Analyzer
  • Accessories for the 115
  • Slide Number 8
  • Tank Pressure vs H2 Flow
  • Linearity of Model 115
  • Schematic of FID
  • 115 Specifications
Page 4: Model 115 portable fid presentation 812

Description Photo

The fill system consists of a customer H2 (99999) Tank with a regulator capable of setting to 250 PSI and a fluidics cable (supplied) with a quick disconnect connector at one end and a frac14rdquo connector at the other end (H2 tank)

Easy to fill Cylinder 4

PID Analyzers LLC 2012

Model 115 FID H2 Tank

Model 115 FID amp Tank 5

PID Analyzers LLC 2012

Detail of Model 115 Analyzer 6

PID Analyzers LLC 2012

FID and Belt Mounted Tank FID Belt Clip

Accessories for the 115 7

PID Analyzers LLC 2012

Shoulder strap for the 115 is also available

8

PID Analyzers LLC 2012

Tank Pressure vs H2 Flow

PID Analyzers LLC 2012

9

0

5

10

15

20

25

0 50 100 150 200 250

H2

Flo

w

Tank Pressure (PSI)

Effect of flow on Tank Presure

Linearity of Model 115

0

500

1000

1500

2000

2500

3000

3500

4000

4500

0 500 1000 1500 2000 2500 3000 3500 4000 4500

FID

Re

ad

ing

115

ppm Methane

FID Linearity

10

PID Analyzers LLC 2012

Schematic of FID

PID Analyzers LLC 2012

11

1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC

waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection

httpwwwhnucom

What can I do with a portable FID analyzer

115 Specifications

PID Analyzers LLC 2012

13

Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4

  • Model 115 Portable FID
  • Introduction
  • The Model 115 is Easy to use
  • Easy to fill Cylinder
  • Model 115 FID amp Tank
  • Detail of Model 115 Analyzer
  • Accessories for the 115
  • Slide Number 8
  • Tank Pressure vs H2 Flow
  • Linearity of Model 115
  • Schematic of FID
  • 115 Specifications
Page 5: Model 115 portable fid presentation 812

Model 115 FID H2 Tank

Model 115 FID amp Tank 5

PID Analyzers LLC 2012

Detail of Model 115 Analyzer 6

PID Analyzers LLC 2012

FID and Belt Mounted Tank FID Belt Clip

Accessories for the 115 7

PID Analyzers LLC 2012

Shoulder strap for the 115 is also available

8

PID Analyzers LLC 2012

Tank Pressure vs H2 Flow

PID Analyzers LLC 2012

9

0

5

10

15

20

25

0 50 100 150 200 250

H2

Flo

w

Tank Pressure (PSI)

Effect of flow on Tank Presure

Linearity of Model 115

0

500

1000

1500

2000

2500

3000

3500

4000

4500

0 500 1000 1500 2000 2500 3000 3500 4000 4500

FID

Re

ad

ing

115

ppm Methane

FID Linearity

10

PID Analyzers LLC 2012

Schematic of FID

PID Analyzers LLC 2012

11

1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC

waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection

httpwwwhnucom

What can I do with a portable FID analyzer

115 Specifications

PID Analyzers LLC 2012

13

Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4

  • Model 115 Portable FID
  • Introduction
  • The Model 115 is Easy to use
  • Easy to fill Cylinder
  • Model 115 FID amp Tank
  • Detail of Model 115 Analyzer
  • Accessories for the 115
  • Slide Number 8
  • Tank Pressure vs H2 Flow
  • Linearity of Model 115
  • Schematic of FID
  • 115 Specifications
Page 6: Model 115 portable fid presentation 812

Detail of Model 115 Analyzer 6

PID Analyzers LLC 2012

FID and Belt Mounted Tank FID Belt Clip

Accessories for the 115 7

PID Analyzers LLC 2012

Shoulder strap for the 115 is also available

8

PID Analyzers LLC 2012

Tank Pressure vs H2 Flow

PID Analyzers LLC 2012

9

0

5

10

15

20

25

0 50 100 150 200 250

H2

Flo

w

Tank Pressure (PSI)

Effect of flow on Tank Presure

Linearity of Model 115

0

500

1000

1500

2000

2500

3000

3500

4000

4500

0 500 1000 1500 2000 2500 3000 3500 4000 4500

FID

Re

ad

ing

115

ppm Methane

FID Linearity

10

PID Analyzers LLC 2012

Schematic of FID

PID Analyzers LLC 2012

11

1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC

waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection

httpwwwhnucom

What can I do with a portable FID analyzer

115 Specifications

PID Analyzers LLC 2012

13

Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4

  • Model 115 Portable FID
  • Introduction
  • The Model 115 is Easy to use
  • Easy to fill Cylinder
  • Model 115 FID amp Tank
  • Detail of Model 115 Analyzer
  • Accessories for the 115
  • Slide Number 8
  • Tank Pressure vs H2 Flow
  • Linearity of Model 115
  • Schematic of FID
  • 115 Specifications
Page 7: Model 115 portable fid presentation 812

FID and Belt Mounted Tank FID Belt Clip

Accessories for the 115 7

PID Analyzers LLC 2012

Shoulder strap for the 115 is also available

8

PID Analyzers LLC 2012

Tank Pressure vs H2 Flow

PID Analyzers LLC 2012

9

0

5

10

15

20

25

0 50 100 150 200 250

H2

Flo

w

Tank Pressure (PSI)

Effect of flow on Tank Presure

Linearity of Model 115

0

500

1000

1500

2000

2500

3000

3500

4000

4500

0 500 1000 1500 2000 2500 3000 3500 4000 4500

FID

Re

ad

ing

115

ppm Methane

FID Linearity

10

PID Analyzers LLC 2012

Schematic of FID

PID Analyzers LLC 2012

11

1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC

waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection

httpwwwhnucom

What can I do with a portable FID analyzer

115 Specifications

PID Analyzers LLC 2012

13

Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4

  • Model 115 Portable FID
  • Introduction
  • The Model 115 is Easy to use
  • Easy to fill Cylinder
  • Model 115 FID amp Tank
  • Detail of Model 115 Analyzer
  • Accessories for the 115
  • Slide Number 8
  • Tank Pressure vs H2 Flow
  • Linearity of Model 115
  • Schematic of FID
  • 115 Specifications
Page 8: Model 115 portable fid presentation 812

8

PID Analyzers LLC 2012

Tank Pressure vs H2 Flow

PID Analyzers LLC 2012

9

0

5

10

15

20

25

0 50 100 150 200 250

H2

Flo

w

Tank Pressure (PSI)

Effect of flow on Tank Presure

Linearity of Model 115

0

500

1000

1500

2000

2500

3000

3500

4000

4500

0 500 1000 1500 2000 2500 3000 3500 4000 4500

FID

Re

ad

ing

115

ppm Methane

FID Linearity

10

PID Analyzers LLC 2012

Schematic of FID

PID Analyzers LLC 2012

11

1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC

waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection

httpwwwhnucom

What can I do with a portable FID analyzer

115 Specifications

PID Analyzers LLC 2012

13

Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4

  • Model 115 Portable FID
  • Introduction
  • The Model 115 is Easy to use
  • Easy to fill Cylinder
  • Model 115 FID amp Tank
  • Detail of Model 115 Analyzer
  • Accessories for the 115
  • Slide Number 8
  • Tank Pressure vs H2 Flow
  • Linearity of Model 115
  • Schematic of FID
  • 115 Specifications
Page 9: Model 115 portable fid presentation 812

Tank Pressure vs H2 Flow

PID Analyzers LLC 2012

9

0

5

10

15

20

25

0 50 100 150 200 250

H2

Flo

w

Tank Pressure (PSI)

Effect of flow on Tank Presure

Linearity of Model 115

0

500

1000

1500

2000

2500

3000

3500

4000

4500

0 500 1000 1500 2000 2500 3000 3500 4000 4500

FID

Re

ad

ing

115

ppm Methane

FID Linearity

10

PID Analyzers LLC 2012

Schematic of FID

PID Analyzers LLC 2012

11

1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC

waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection

httpwwwhnucom

What can I do with a portable FID analyzer

115 Specifications

PID Analyzers LLC 2012

13

Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4

  • Model 115 Portable FID
  • Introduction
  • The Model 115 is Easy to use
  • Easy to fill Cylinder
  • Model 115 FID amp Tank
  • Detail of Model 115 Analyzer
  • Accessories for the 115
  • Slide Number 8
  • Tank Pressure vs H2 Flow
  • Linearity of Model 115
  • Schematic of FID
  • 115 Specifications
Page 10: Model 115 portable fid presentation 812

Linearity of Model 115

0

500

1000

1500

2000

2500

3000

3500

4000

4500

0 500 1000 1500 2000 2500 3000 3500 4000 4500

FID

Re

ad

ing

115

ppm Methane

FID Linearity

10

PID Analyzers LLC 2012

Schematic of FID

PID Analyzers LLC 2012

11

1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC

waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection

httpwwwhnucom

What can I do with a portable FID analyzer

115 Specifications

PID Analyzers LLC 2012

13

Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4

  • Model 115 Portable FID
  • Introduction
  • The Model 115 is Easy to use
  • Easy to fill Cylinder
  • Model 115 FID amp Tank
  • Detail of Model 115 Analyzer
  • Accessories for the 115
  • Slide Number 8
  • Tank Pressure vs H2 Flow
  • Linearity of Model 115
  • Schematic of FID
  • 115 Specifications
Page 11: Model 115 portable fid presentation 812

Schematic of FID

PID Analyzers LLC 2012

11

1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC

waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection

httpwwwhnucom

What can I do with a portable FID analyzer

115 Specifications

PID Analyzers LLC 2012

13

Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4

  • Model 115 Portable FID
  • Introduction
  • The Model 115 is Easy to use
  • Easy to fill Cylinder
  • Model 115 FID amp Tank
  • Detail of Model 115 Analyzer
  • Accessories for the 115
  • Slide Number 8
  • Tank Pressure vs H2 Flow
  • Linearity of Model 115
  • Schematic of FID
  • 115 Specifications
Page 12: Model 115 portable fid presentation 812

1 Determine when to put on an SCBA amp protective clothing 2 Find buried barrels of haz waste amp check drums for VOC

waste 3 Check soil or water headspace for contamination 4 Leak Detection- liquid fuels natural gas 5 Emergency response 6 EPA Method 21 ndashleak detection

httpwwwhnucom

What can I do with a portable FID analyzer

115 Specifications

PID Analyzers LLC 2012

13

Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4

  • Model 115 Portable FID
  • Introduction
  • The Model 115 is Easy to use
  • Easy to fill Cylinder
  • Model 115 FID amp Tank
  • Detail of Model 115 Analyzer
  • Accessories for the 115
  • Slide Number 8
  • Tank Pressure vs H2 Flow
  • Linearity of Model 115
  • Schematic of FID
  • 115 Specifications
Page 13: Model 115 portable fid presentation 812

115 Specifications

PID Analyzers LLC 2012

13

Range- 0-50000 ppm 0-5000 ppm Detection limit- 01 ppm Response time (sec)- 3 sec Weight Analyzer- 2 Weight Tank- 22 H2 (liters) whenfull 70 Battery life- 12 hours Battery Capacity- 0-100 (7 V to 5 VDC) Dimensions- Analyer- 115rdquo x 3rdquo x 2375rdquo Tank- 85rdquo x 275rdquo x 25rdquo Analog output- 0-1 VDC ADC- 16 bit Precision- +- 5 Operating Range- 0-45oC Humidity range - 0-98 RH (non condensing_ Flame Temp Readout- 0-250oC Cal Gas 0-1000 or 0-10000 ppm CH4

  • Model 115 Portable FID
  • Introduction
  • The Model 115 is Easy to use
  • Easy to fill Cylinder
  • Model 115 FID amp Tank
  • Detail of Model 115 Analyzer
  • Accessories for the 115
  • Slide Number 8
  • Tank Pressure vs H2 Flow
  • Linearity of Model 115
  • Schematic of FID
  • 115 Specifications

Recommended