+ All Categories
Home > Documents > Monte Carlo response function simulations for the HEXITEC ... · detectors such as CdTe (Z = 50)...

Monte Carlo response function simulations for the HEXITEC ... · detectors such as CdTe (Z = 50)...

Date post: 18-Oct-2020
Category:
Upload: others
View: 0 times
Download: 0 times
Share this document with a friend
1
Monte Carlo response function simulations for the HEXITEC CdTe detector Kjell A.L Koch-Mehrin 1* , John E. Lees 1 , Sarah L. Bugby 1 , Matt D. Wilson 2 1 Space Research Centre, Michael Atiyah Building, University of Leicester, LE1 7RH, UK 2 Science and Technology Facilities Council, Rutherford Appleton Laboratory, Harwell Campus, Oxfordshire, OX11 0QX, UK 1. Introduction High atomic number semiconductor detectors such as CdTe (Z = 50) offer the possibility to detect X-ray energies above what is possible with Silicon or Germanium sensors [1]. Furthermore, due to a large band gap, CdTe detectors exhibit less thermal charge leakage and can therefore be operated at room temperature. However, phenomena such as polarization, event pile- up and charge sharing, affect CdTe detector performance and are yet to be fully understood. 2. The HEXITEC system High energy X-ray imaging technology (HEXITEC) is a family of spectroscopic, single photon counting, pixel detectors developed for high energy X-ray and -ray spectroscopy applications. The HEXITEC ASIC has previously demonstrated good spectroscopic results when coupled with a 1mm thick CdTe detector, giving an energy resolution of ~1 keV FWHM at the 59.5 keV [2] and detecting photons up to ~ 200 keV. The ASIC anode consists of an 80x80 pixel array on a 250μm pitch. the CCE by favouring charge collection from the electrons over the holes. Both weighting potentials are included in the model. the ‘tail’ cannot be modelled using the CCE. At 122 keV for 57 Co, the model for the weighted CCE gives a good fit, whereas the planar CCE significantly over-corrects. Figure 5 shows charge loss between two HEXITEC pixels – a loss of up to ~5 keV can be seen. *Email: [email protected] 1. S. Del Sordo, Progress in the development of CdTe semiconductor radiation detectors for astrophysical and medical applications, Sensors (2009) 2. M. L. Wilson et al, A 10cm10cm CdTe Spectroscopic Imaging Detector based on the HEXITEC ASIC, JINST 10 (2015) P10011 3. D.D. Duarte, Simulation of active-edge pixelated CdTe radiaton detectors, Nuc. Inst. and Meth. In Phy. Res. A. (2015) Figure 2. Weighting potential for the Hecht equation to calculate CCE for a planar detector w=L, and HEXITEC detector, w/L = 0.25, showing small pixel effect [3]. Figure 3. 57 Co spectrum from HEXITEC and model. Figure 4. Model vs HEXITEC data for a single pixel, biased at -500V and 80e - noise. 57 Co Cd & Te escape peaks: 99 keV, 95 keV Cd & Te K α XRF: 23 keV, 27 keV XRF peaks from Pb shielding 57 Co gamma emission: 122 keV, 136.5 keV Anode side Cathode side 5. Conclusions The tail to the left of the photopeak at lower energies (22 and 59 keV) may not be due to charge trapping but instead result from effects such as charge sharing or polarization. At higher energies (122 keV) the dominant source of the tail appears to be charge trapping. 3. Method A Monte Carlo model, written in Python, has been developed. X-ray attenuation by Compton scattering, photoelectric effect and Rayleigh scattering is considered. The charge cloud size due to electron ranges, cloud diffusion during drift and charge collection efficiency (CCE) due to carrier trapping is calculated. Observations using the HEXITEC were made with radioisotopes 109 Cd, 241 Am and 57 Co which have primary photopeaks at 22, 59.5 and 122 keV respectively. The HEXITEC detector benefits from the small-pixel effect; when the pixel size, w, is small compared to the thickness of the detector, L. This influences the weighting potential, shown in figure 2, which improves 4. Results and Discussion Figure 3 shows a 57 Co spectrum from the HEXITEC collected over all pixels compared with a simulated model. Both spectra are normalized to the main photopeak at 122 keV. With the exception of the Te escape peak, a good agreement for the Cd escape peak and Cd & Te XRF which are re-absorbed within the detector is seen, with a slight over-estimation from the model. Figure 5. Charge sharing between adjacent pixels for 59 keV line from 241 Am source. 241 Am 109 Cd Figure 4 shows the simulated model against HEXITEC data for a single pixel – the orange curves represent CCE for a planar detector (i.e. worst case scenario), and the green curve the CCE after correction by the small- pixel effect weighting. For the photopeaks at 22 and 59 keV for 109 Cd, 241 Am respectively, Figure 1. Left: HEXITEC, right: CZT detector [2].
Transcript
Page 1: Monte Carlo response function simulations for the HEXITEC ... · detectors such as CdTe (Z = 50) offer the possibility to detect X-ray energies above what is possible with Silicon

Monte Carlo response function simulations

for the HEXITEC CdTe detector Kjell A.L Koch-Mehrin1*, John E. Lees 1, Sarah L. Bugby 1, Matt D. Wilson 2

1 Space Research Centre, Michael Atiyah Building, University of Leicester, LE1 7RH, UK2 Science and Technology Facilities Council, Rutherford Appleton Laboratory, Harwell Campus, Oxfordshire, OX11 0QX, UK

1. IntroductionHigh atomic number semiconductordetectors such as CdTe (Z = 50) offer thepossibility to detect X-ray energies abovewhat is possible with Silicon or Germaniumsensors [1]. Furthermore, due to a large bandgap, CdTe detectors exhibit less thermalcharge leakage and can therefore beoperated at room temperature. However,phenomena such as polarization, event pile-up and charge sharing, affect CdTe detectorperformance and are yet to be fullyunderstood.

2. The HEXITEC system

High energy X-ray imaging technology(HEXITEC) is a family of spectroscopic, singlephoton counting, pixel detectors developedfor high energy X-ray and -ray spectroscopyapplications. The HEXITEC ASIC haspreviously demonstrated good spectroscopicresults when coupled with a 1mm thick CdTedetector, giving an energy resolution of ~1keV FWHM at the 59.5 keV [2] and detectingphotons up to ~ 200 keV. The ASIC anodeconsists of an 80x80 pixel array on a 250µmpitch.

the CCE by favouring charge collection fromthe electrons over the holes. Both weightingpotentials are included in the model.

the ‘tail’ cannot be modelled using the CCE.At 122 keV for 57Co, the model for theweighted CCE gives a good fit, whereas theplanar CCE significantly over-corrects. Figure5 shows charge loss between two HEXITECpixels – a loss of up to ~5 keV can be seen.

*Email: [email protected]

1. S. Del Sordo, Progress in the development of CdTe semiconductor radiation detectors for astrophysical and medical applications, Sensors (2009)

2. M. L. Wilson et al, A 10cm10cm CdTe Spectroscopic Imaging Detector based on the HEXITEC ASIC, JINST 10 (2015) P10011

3. D.D. Duarte, Simulation of active-edge pixelated CdTe radiaton detectors, Nuc. Inst. and Meth. In Phy. Res. A. (2015)

Figure 2. Weighting potential for the Hecht equation tocalculate CCE for a planar detector w=L, and HEXITECdetector, w/L = 0.25, showing small pixel effect [3].

Figure 3. 57Co spectrum from HEXITEC and model.

Figure 4. Model vs HEXITEC data for a single pixel,biased at -500V and 80e- noise.

57Co

Cd & Te escape peaks: 99 keV, 95 keV

Cd & Te Kα XRF: 23 keV, 27 keV

XRF peaks from Pb shielding

57Co gamma emission:

122 keV, 136.5 keV

Anode sideCathode side

5. Conclusions• The tail to the left of the photopeak at

lower energies (22 and 59 keV) may notbe due to charge trapping but insteadresult from effects such as charge sharingor polarization.

• At higher energies (122 keV) the dominantsource of the tail appears to be chargetrapping.

3. MethodA Monte Carlo model, written in Python, hasbeen developed. X-ray attenuation byCompton scattering, photoelectric effect andRayleigh scattering is considered. The chargecloud size due to electron ranges, clouddiffusion during drift and charge collectionefficiency (CCE) due to carrier trapping iscalculated. Observations using the HEXITECwere made with radioisotopes 109Cd, 241Amand 57Co which have primary photopeaks at22, 59.5 and 122 keV respectively.

The HEXITEC detector benefits from thesmall-pixel effect; when the pixel size, w, issmall compared to the thickness of thedetector, L. This influences the weightingpotential, shown in figure 2, which improves

4. Results and DiscussionFigure 3 shows a 57Co spectrum from theHEXITEC collected over all pixels comparedwith a simulated model. Both spectra arenormalized to the main photopeak at 122keV. With the exception of the Te escapepeak, a good agreement for the Cd escapepeak and Cd & Te XRF which are re-absorbedwithin the detector is seen, with a slightover-estimation from the model.

Figure 5. Charge sharing between adjacent pixels for 59

keV line from 241Am source.

241Am 109Cd

Figure 4 shows the simulated model againstHEXITEC data for a single pixel – the orangecurves represent CCE for a planar detector(i.e. worst case scenario), and the greencurve the CCE after correction by the small-pixel effect weighting. For the photopeaks at22 and 59 keV for 109Cd, 241Am respectively,

Figure 1. Left: HEXITEC, right: CZT detector [2].

Recommended