Date post: | 21-Dec-2016 |
Category: |
Documents |
Upload: | tanveer-ahmad |
View: | 219 times |
Download: | 2 times |
ORIGINAL PAPER
New UPLC–MS/MS method for simultaneous determinationof irbesartan and hydrochlorthiazide in human plasma
Seema Zargar • Tanveer Ahmad Wani
Received: 27 July 2013 / Accepted: 7 February 2014
� Iranian Chemical Society 2014
Abstract Ultra-performance liquid chromatography–tan-
dem mass spectrometry (UPLC–MS/MS) is a preeminent
analytical tool for rapid biomedical analysis with the
objective of reducing analysis time and maintaining good
efficiency. In this study a simple, rapid, sensitive and
specific ultra-performance liquid chromatography–tandem
mass spectrometry method was developed and validated
for quantification of the angiotensin II receptor antagonist,
irbesartan and hydrochlorthiazide in human plasma. After a
simple protein precipitation using methanol and acetoni-
trile, irbesartan, hydrochlorthiazide and internal standard
(IS) telmisartan were separated on Acquity UPLC BEHTM
C18 column (50 9 2.1 mm, i.d. 1.7 lm, Waters, USA)
using a mobile phase consisting of acetonitrile:10 mM
ammonium acetate:formic acid (85:15:0.1 % v/v/v)
pumped at a flow rate of 0.3 mL/min and detected by
tandem mass spectrometry with negative ion mode. The ion
transitions recorded in multiple reaction monitoring mode
were m/z 427.2 ? 193.08 for irbesartan, m/z 295.93 ?268.90 for hydrochlorthiazide and m/z 513.2 ? 287.14 for
IS. The assay exhibited a linear dynamic range of
30–500 ng/mL for irbesartan and 1–500 ng/mL in human
plasma with good correlation coefficient of (0.996) and
(0.997) and with a limit of quantitation of 30 and 1 ng/mL
for irbesartan and hydrochlorthiazide, respectively. The
intra- and inter-assay precisions were satisfactory; the
relative standard deviations did not exceed 10.13 % for
irbesartan and 11.14 % for hydrochlorthiazide. The pro-
posed UPLC–MS/MS method is simple, rapid and highly
sensitive, and hence it could be reliable for pharmaco-
kinetic and toxicokinetic study in both animals and
humans.
Keywords Irbesartan � Hydrochlorthiazide � Ultra-
performance liquid chromatography � Tandem mass
spectrometry � Pharmacokinetic � Toxicokinetic � High-
throughput analysis
Introduction
Most cardiovascular events are attributed to high blood
pressure. Hence, antihypertensive therapy is to reduce
considerably the risk of developing cardiovascular com-
plications that cause a high mortality rate in the patients
with hypertension. Hydrochlorothiazide (HCT) 6-chloro-
3,4-dihydro-2H-1,2,4-benzothiadiazine-7-sulfonamide 1,1-
dioxide, is one of the oldest thiazide diuretics, often pre-
scribed in combination with other antihypertensive drugs
such as beta blockers, angiotensin-converting enzyme
inhibitors, or angiotensin II receptor blockers [1–3]. Irbe-
sartan (IBS), 2-butyl-3-[[20-(tetrazol-5-yl)biphenyl-4-yl]-
methyl]-1,3-diazaspiro[4.4]non-1-en-4-one, is a potent and
selective angiotensin II subtype 1 receptor antagonist
indicated for use in patients with hypertension, in addition
to those with type 2 diabetes mellitus and nephropathy [4].
Angiotensin II is regarded as the main effector of AT1
receptor in renin–angiotensin system. It causes vasocon-
striction, tachycardia, increase of aldosterone secretion
from the adrenal cortex and retention of sodium and body
fluid [4, 5]. The combination of IBS and HCT dosage form
S. Zargar
Department of Biochemistry, College of Science, King Saud
University, P.O. Box 22452, Riyadh 11211, Saudi Arabia
T. A. Wani (&)
Department of Pharmaceutical Chemistry, College of Pharmacy,
King Saud University, P.O. Box 2457, Riyadh 11451,
Saudi Arabia
e-mail: [email protected]
123
J IRAN CHEM SOC
DOI 10.1007/s13738-014-0429-3
has been indicated for treatment of edema and hypertension
and clinical studies conducted in this regard have shown
that the combination is clinically effective with a good
safety profile [6, 7]. IBS/HCT provided consistent blood
pressure lowering and tolerability regardless of age, obes-
ity, and prevalence type 2 diabetes and greater efficacy in
patients with high cardiovascular risk [8].
A number of methods have been employed for the
analysis of hydrochlorothiazide concentration alone or in
combination with other drugs in biological samples by
high-performance liquid chromatography (HPLC) with
ultraviolet or electrochemical detection [9–14], LC–MS
[15, 16] or with tandem LC–MS/MS [17–21]. However,
most of them were time consuming, insufficiently sensitive
and employed tedious liquid–liquid extraction. The liquid–
liquid extraction methods involved endogenous several
steps yielding poor separation from the plasma interfer-
ences and gave highly variable and relatively low recov-
eries. Several HPLC methods have also been reported for
the determination of IBS in human plasma which include;
HPLC coupled with UV [22], fluorescence [23, 24].
Although the simultaneous determination of IBS and HCT
has previously been suggested by other workers, who
employed HPLC coupled with UV detection [25, 26], or
with tandem LC–MS/MS detection [27] such methods
suffered from lack of sensitivity, demonstrated by the
lower limits of quantitation.
UPLC is a new category of separation science which
builds upon well-established principles of liquid chroma-
tography, using sub-2 lm porous particles. These particles
operate at elevated mobile phase linear velocities to pro-
duce significant reductions in separation time and solvent
consumption. Literature indicates that a UPLC system
allows approximately ninefold decreases in analysis time
as compared to the conventional high-performance (HP)
LC system using 5 lm particle size analytical columns, and
approximately threefold decrease in analysis time in
comparison with 3 lm particle size analytical columns
without compromise on overall separation [28–33]. Ac-
quity UPLC columns contain hybrid X-Terra sorbent,
which utilizes bridged ethyl siloxane/silica hybrid (BEH)
structure, ensures the column stability under the high
pressure and wide pH range (1–12) [33]. In all documented
references, no UPLC–MS/MS method has been used to
determinate IBS and HCT presence and concentration in
human plasma until now.
The present study describes the development and vali-
dation of a UPLC method coupled with tandem mass
spectrometry (UPLC–MS/MS) for the determination of
IBS and HCT in human plasma. The proposed method used
is more sensitive and relatively simple extraction procedure
using methanol and acetonitrile to directly precipitate
protein in combination with UPLC–MS/MS detection.
Experimental
Materials and methods
Hydrochlorthiazide was obtained from Sigma Chemical
Co. (St. Louis, USA). Telmisartan and irbesartan (IS) were
obtained from AK Scientific Inc. (CA, USA). Human
plasma was obtained from normal healthy volunteers at
King Khalid University Hospital (Riyadh, Saudi Arabia),
and they were kept frozen at -20 �C until analysis. HPLC-
grade acetonitrile, methanol and ammonium acetate were
obtained from Winlab Laboratory, UK. Formic acid was
obtained from BDH Laboratory, UK. All other reagents
were of analytical grade unless stated otherwise. All
aqueous solutions was prepared using water that was
purified using Milli-QR Gradient A10R (Millipore, Mosc-
heim Cedex, France) having pore size 0.22 lm.
Apparatus and operating condition
Liquid chromatography
The UPLC system included quaternary solvent manager, a
binary pump, degasser, autosampler with an injection loop
of 10 lL and a column heater-cooler. The separation was
performed on Acquity UPLC BEHTM C18 column
(50 9 2.1 mm, i.d., 1.7 lm, Waters, USA) maintained at
40 �C. The mobile phase was composed of acetoni-
trile:10 mM ammonium acetate:formic acid (85:15:0.1 %
v/v/v)) pumped at a flow rate of 0.3 mL/min The injection
volume was 5 lL in partial loop mode and the temperature
of the autosampler was kept at 4 �C.
Mass spectrometric conditions
Waters Acquity liquid chromatography system coupled
with a Waters TQD triple quadrupole mass spectrometer
was used (Waters, Milford, USA). Mass spectrometric
detection was carried out using an electrospray interface
(ESI) operated in the negative ionization mode with
multiple reaction monitoring (MRM) for all IBS, HCT and
IS. Nitrogen was used as a desolvating gas at a flow rate
of 550 L/h. The desolvating temperature was set at 350 �C
and the source temperature was set at 150 �C. The colli-
sion gas (argon) flow was set at 0.1 mL/min. The capillary
voltage was set at 3.2 kV. Decrease in capillary voltage
had minimal effect on peak intensity, however, decreasing
the capillary voltage reduced the peak intensity drasti-
cally. The MS analyzer parameters were as follows: LM1
and HM1 resolution 15.0 and 15.0; ion energy 1, 0.8 V;
LM2 and HM2 resolution 12.0 and 14.0, respectively, ion
energy 2, 0.1 V, dwell time, 0.146 s. The cone voltage
and collision energy were optimized in case of each
J IRAN CHEM SOC
123
analyte so as to maximize the signal corresponding to the
major transition observed in the MS/MS spectra, follow-
ing the fragmentation of the [M?H]? ions corresponding
to the selected compounds. The Mass Lynx software
(Version 4.1, SCN 805) was used to control the UPLC–
MS/MS system as well as for data acquisition and
processing.
Calibration standards and quality control samples
A standard stock solution of IBS, HCT and telmisartan (IS)
were prepared by dissolving the compounds in methanol, to
give a final concentrations of 1 mg/mL. The 1 mg/mL
stock solution of IBS and HCT was serially diluted to
prepare working solutions in the required concentration
range with diluent methanol–water (50:50, v/v). The cali-
bration standards and quality control (QC) samples were
prepared by spiking with working solutions yielding eight
standard solutions ranging from 30 to 500 ng/mL for IBS
and 1–500 ng/mL for HCT. QC stock solutions were pre-
pared separately in methanol–water (50:50, v/v). QC
samples at three different concentrations levels: 60, 250,
and 400 ng/mL for IBS and 3, 250 and 400 ng/mL for
HCT. Spiked plasma calibration standards and quality
control samples were kept at -80 �C until assayed or used
for validating the assay procedures. The IS working solu-
tion (1 lg/mL) for routine use was prepared by diluting the
telmisartan stock solution in methanol and kept in refrig-
erator for storage.
Plasma blank: 190 lL of plasma was spiked with 10 lL
of methanol–water (50:50, v/v).
Plasma blank with internal standard: 190 lL of plasma
was spiked with 10 lL of 1 lg/mL IS.
Sample preparation
A simple protein precipitation method was used to extract
IBS, HCT and IS. Plasma samples stored at around -80 �C
were thawed, left for 1 h and vortexed for 30 s on room
temperature before extraction to ensure homogeneity. To
190 lL of plasma sample, 10 lL (0.6 lg/mL) of IS was
added. The samples were vortex mixed for about 30 s and
then 100 lL of methanol was added to it and vortex mixed
again for another 30 s. After vortex mixing, further 500 lL
of acetonitrile was added to the sample. The samples were
again vortex mixed gently for 1.0 min and the supernatant
was separated after centrifugation at 15,000g for 10 min
and evaporated to dryness under a gentle stream of nitrogen
at 40 �C. The residue was reconstituted with 190 lL of
methanol–water (50:50, v/v) and transferred to UPLC vials.
5 lL volumes (in partial loop with needle over fill mode)
of the sample were subjected to the analysis by UPLC–MS/
MS.
Bioanalytical method validation
A full method validation was performed according to
guidelines set by the United States Food and Drug
Administration (US-FDA) and European Medicines
Agency (EMEA) guidelines. [34, 35] The validation of this
procedure was performed in human plasma in order to
evaluate the method in terms of selectivity, linearity of
response, accuracy, precision, recovery, dilution integrity
and stability of analytes during both short-term sample
processing and long-term storage. Selectivity, linearity,
accuracy and precision exercise was also performed in
human plasma.
Selectivity and specificity
The selectivity of the method towards endogenous plasma
matrix components, metabolites and component medica-
tions was assessed in human blank plasma. Among the
analyzed plasma batches, plasma batch showing no or
minimal interference at the retention time of analytes and
internal standards was selected. They were processed and
analyzed using the proposed extraction protocol spiked
with standard IBS and HCT at lower limit of quantification
(LLOQ) level (30 and 1 ng/mL, respectively, for IBS and
HCT) and IS 30 ng/mL.
Carry over
Carryover effect was evaluated to ensure that the rinsing
solution used to clean the injection needle and port was
able to avoid any carry-forward of injected sample in
subsequent runs. The design of the experiment comprised
blank plasma, LLOQ and upper limit of quantitation
(ULOQ) followed by blank plasma to check for any pos-
sible interference due to carryover.
Linearity and standard curve
The linearity of the method was determined by analysis of
standard plots associated with eight-point standard cali-
bration curve recorded individually for IBS and HCT. Each
was plotted against corresponding concentration level in
the ranges 30–500 ng/mL for IBS, and 1–500 ng/mL for
HCT. Calibration curves from accepted three precision and
accuracy batches were used to establish linearity. Curves
were best fitted using a least-square linear regression model
y = mx ? b, weighted by 1/x2, in which y is the peak area
ratio, m is slope of the calibration curve, b is the y axis
intercept of the calibration curve and x is the analyte (HCT
or IBS) concentration. Back-calculations were made from
these curves to determine the concentration of IBS and
HCT in each calibration standards and the resulting
J IRAN CHEM SOC
123
calculated parameters were used to determine concentra-
tions of analyte in quality control samples. The determi-
nation coefficient r2 [ 0.98 was desirable for all the
calibration curves. The lowest standard on the calibration
curve was to be accepted as the lower limit of quantifica-
tion (LLOQ), if the analyte response was at least ten times
more than that of drug-free (blank) extracted plasma. In
addition, the analyte peak of LLOQ sample should be
identifiable, discrete, and reproducible with accuracy
within ±20 % and a precision B20 %. The deviation of
standards other than LLOQ from the nominal concentration
should not be more than ±15.0 %.
Precision and accuracy
Intra- and inter-day accuracies expressed as a percentage of
deviation from the respective nominal value. The precision
of the assay was measured by the percent coefficient of
variation (% CV) at four concentrations in human plasma.
Intra-day precision and accuracy were assessed by analyzing
six replicates of the quality control samples at three levels
(quality control) during a single analytical run. The inter-day
precision and accuracy were assessed by analyzing 18 rep-
licates of the quality control samples at each level through
three precision and accuracy batches runs on three consec-
utive validation days. The deviation at each concentration
level from the nominal concentration was expected to be
within ±15.0 % except LLOQ, for which it should not be
more than 20.0 %. Similarly, the mean accuracy should not
deviate by ±15.0 % except for the LLOQ where it can be
±20.0 % of the nominal concentration.
Extraction recovery and matrix effect
To investigate extraction recovery, a set of samples (n = 6
at each low, medium, and high concentration levels in
unique lots of plasma) was prepared by spiking IBS and
HCT into plasma at 60, 250, and 400 ng/mL for IBS and 3,
250 and 400 ng/mL for HCT, respectively. Each of the
samples were processed as per the procedure described
previously. A second set of plasma samples was processed
and spiked post-extraction with the same concentrations of
IBS, HCT and IS that actually existed in the pre-extraction
spiked samples. Extraction recovery for each analyte was
determined by calculating the ratios of the raw peak areas
of the pre-extraction spiked samples to those of the samples
spiked after extraction. The matrix effect was evaluated by
analyzing MQC sample.
Stability and dilution integrity evaluation
Stability of IBS and HCT in plasma was assessed by
analyzing six replicates of QC samples at low and high
concentrations under a variety of storage and processing
conditions. Six aliquots of each low and high concentration
quality control samples were taken to evaluate the bench-
top stability (short-term stability), freeze–thaw stability,
auto sampler storage stability and long-term stability.
Bench-top stability was assessed after exposure of the
plasma samples to room temperature for *6 h, which
exceeds the residence time of the sample processing pro-
cedures. The freeze–thaw stability was evaluated after
undergoing three freeze (at around -80 �C)–thaw (room
temperature) cycles. The autosampler storage stability was
determined by storing the reconstituted QC samples for
*48 h under autosampler condition (maintained at 8 �C)
before being analyzed. Long-term stability was assessed
after storage of the test samples at around -80 �C for
60 days. The working solutions and stock solutions of IBS,
HCT and IS were also evaluated for stability at room
temperature for 24 h and at refrigerator temperature (below
10 �C) for 30 days. All stability exercises were performed
against freshly spiked calibration standards. The samples
were considered stable in plasma at each concentration if
the deviation from the mean calculated concentration of
stability quality control samples was within ±15 %.
The dilution integrity experiment was intended to vali-
date the dilution test to be carried out on higher analyte
concentrations (above ULOQ), which may be encountered
during real subject samples analysis. It was performed at
1.6 times the ULOQ concentration. Six replicates samples
of half and quarter concentration were prepared and their
concentrations were calculated by applying the dilution
factor of 2 and 4, respectively, against the freshly prepared
calibration curve. The integrity of the samples was con-
sidered to be maintained if % nominal is within ±15 % of
nominal values and % CVs B 15 % at both diluted levels.
Result and discussion
Optimization and validation of assay
Optimization of chromatographic condition
Initial feasibility experiments of various mixture(s) of
organic solvents such as acetonitrile and methanol along
with Millipore water; both having 0.1 % formic acid, also
these organic solvents along with different concentration of
ammonium acetate (2–15 mM) with altered flow-rates (in
the range of 0.20–0.50 mL/min) were performed to optimize
an effective chromatographic conditions of IBS, HCT and IS
(chemical structures given in Fig. 1). The best conditions
were achieved with mobile phase comprising acetonitrile:
10 mM ammonium acetate:formic acid (85:15:0.1 % v/v/v))
pumped at a flow rate of 0.3 mL/min, on Acquity UPLC
J IRAN CHEM SOC
123
BEH� C18 column (50 9 2.1 mm, i.d. 1.7 lm. The selected
conditions were found to be suitable for the determination of
electrospray response for IBS, HCT and IS.
UPLC–MS/MS operation parameters were carefully
optimized for the determination of IBS. Analytes were
detected by tandem mass spectrometry using MRM of
precursor–product ion transitions with 0.146 s dwell time,
at m/z 427.2 ? 193.08 for IBS, 295.93 ? 268.90 for HTZ
and m/z 513.2 ? 287.14 for IS. A standard solution
(100 ng/mL) of IBS, HCT and the IS were directly infused
along with the mobile phase into the mass spectrometer
with ESI as the ionization source. The mass spectrometer
was tuned initially in the negative ionization modes for
IBS, HCT and IS. Parameters, such as capillary and cone
voltage, desolvation temperature, ESI source temperature
and flow rate of desolvation gas and cone gas, were opti-
mized to obtain the optimum intensity of protonated mol-
ecules of IBS, HCT and IS for quantification. Among the
parameters, capillary and cone voltage, especially cone
voltage, were important parameters. The cone voltage was
optimized using cone ramp (2–100) V. The precursor ion
intensities increased significantly when cone voltage was
raised gradually. Lastly, analytes produced the strongest
ion signals when cone voltage was set up at 42 V for IBS
and at 46 V for HCT. Decrease in the cone voltage reduced
the ion signal and increase in the cone voltage had minimal
effect on the ion signal. The collision energy was investi-
gated from 2 to 80 eV to optimize the response of product
ion, and the best values were found to be 28 eV for the
product ions m/z 193.08 for IBS and 18 eV for the product
ions m/z 268.9 for HCT. For IS, m/z 287.14 spectra were
produced at cone voltage of 48 eV optimum collision
energy of 34 eV.
Optimization of sample processing
Protein precipitation was used for sample preparation in this
study. Protein precipitation can be helpful in producing a clean
sample and avoiding endogenous substances in plasma with
the analytes and IS onto the column and MS system. Clean
samples are essential for minimizing ion suppression and
matrix effect in UPLC–MS/MS analysis. Two organic
Fig. 1 Chemical structure of
IBS (a), HCT (b) and
telmisartan (IS) (c)
Fig. 2 a–c Represent chromatograms obtained from blank plasma
showing no interference at the retention time of IBS, HCT and IS,
respectively. d–f Represent chromatogram of LLOQ for IBS, HCT
and IS, respectively. g–i Representative chromatogram HQC for IBS,
HCT and IS, respectively
J IRAN CHEM SOC
123
solvents, acetonitrile and methanol, were used for precipita-
tion of these proteins. Finally a combination of methanol and
acetonitrile was found to be optimal, which can produce a
clean chromatogram for a blank plasma sample and yield the
highest recovery for the analytes from the plasma.
Selectivity
Selectivity of the method was assessed by comparing the
chromatogram of blank plasma with the corresponding
spiked LLOQ sample. Six different batches of blank human
plasma were tested to identify the peaks due to the possible
biogenic plasma components. Thus the method looks to be
selective enough for determination of IBS, HCT and IS in
plasma. Representative chromatograms obtained from
blank plasma showing no interference at the retention time
of IBS, HCT and IS are shown in Fig. 2a–c, respectively.
Representative chromatogram of LLOQ for IBS, HCT and
IS is shown in Fig. 2d–f, respectively; whereas, represen-
tative chromatogram HQC for IBS, HCT and IS are shown
in Fig. 2g–i, respectively.
Linearity and sensitivity
The linearity of the method was determined by a weighted
least-square regression analysis of standard plot associated
with an eight-point standard curve for both IBS and HCT.
The calibration curves were generated by plotting area ratio
(IBS/IS) as a function of IBS concentration and was found to
be linear from 30 to 500 ng/mL for IBS in human plasma and
(HCT/IS) as a function of HCT concentration and was found
to be linear from 1 to 500 ng/mL for HCT in human plasma.
The determination coefficients (r2) were consistently greater
than 0.995 during the course of validation. The lower limit of
quantification for this assay was 30 and 1 ng/mL for IBS and
HCT in plasma, respectively. Representative LLOQ is sen-
sitive enough to investigate the pharmacokinetic behavior of
IBS and HCT in human plasma.
Precision and accuracy
Tables 1 and 2 summarize the inter- and intra-day preci-
sion and accuracy values for QC samples. The coefficient
of variation values of both intra- and inter-day results of
plasma were 2.54–10.13 and 1.14–2.29 %, respectively, for
IBS and intra- and inter-day results of plasma were
2.21–11.14 and 0.62–1.06 %, respectively, for HCT. These
results indicate that the method has good precision and
accuracy and are within the acceptance limit of\15 % and
±\15 % for precision and accuracy, respectively.
Recovery
At three QC concentration levels 60, 250, and 400 ng/mL
for IBS and 3, 250 and 400 ng/mL for HCT the percent
extraction recoveries (mean ± SD) of IBS and HCT
obtained are given in Table 3. The mean extraction
recovery for IBS was 82.04 ± 1.80 % and for HCT was
85.85 ± 1.43 %. The mean recovery for the IS telmisartan
at the concentration employed was 88.62 ± 9.68 %. This
result indicates that the extraction efficiency for IBS and
HCT using protein precipitation method was satisfactory,
consistent and concentration independent.
Table 1 Intra- and inter-day precision and accuracy of IBS in human
plasma
Spiked
conc.
(ng/mL)
Run Measured conc.
(ng/mL ± SD)
Precision
(CV, %)
Accuracy
(recovery, %)
Intra-day variation (six replicates at each concentration)
60 1 58.88 ± 5.96 10.13 98.13
2 61.51 ± 5.12 8.32 102.53
3 59.49 ± 5.23 8.80 99.16
250 1 248.61 ± 15.33 6.16 99.44
2 256.11 ± 12.73 4.96 102.68
3 248.19 ± 17.04 6.86 99.27
400 1 402.38 ± 17.04 4.23 100.59
2 406.37 ± 10.33 2.54 101.60
3 404.80 ± 15.50 3.83 101.20
Inter-day variation (18 replicates at each concentration)
60 59.96 ± 1.37 2.29 96.38
250 251.17 ± 4.80 1.91 100.91
400 406.19 ± 4.66 1.14 101.54
Table 2 Intra- and inter-day precision and accuracy of HCT in
human plasma
Spiked
conc.
(ng/mL)
Run Measured conc.
(ng/mL ± SD)
Precision
(CV, %)
Accuracy
(recovery, %)
Intra-day variation (six replicates at each concentration)
3 1 3.03 ± 0.32 10.87 101.11
2 3.05 ± 0.34 11.14 101.72
3 3.04 ± 0.23 7.60 101.61
250 1 250.28 ± 10.74 4.29 100.11
2 248.38 ± 14.16 5.70 99.35
3 247.19 ± 18.35 7.42 98.87
400 1 393.83 ± 8.71 2.21 98.45
2 400.61 ± 11.81 2.94 100.15
3 401.63 ± 15.65 3.89 100.40
Inter-day variation (18 replicates at each concentration)
3 3.08 ± 0.03 1.07 96.38
250 248.61 ± 1.55 0.62 100.91
400 398.69 ± 4.23 1.06 101.54
J IRAN CHEM SOC
123
Matrix effect and other recoveries
In this study, the matrix effect was evaluated by analyzing
MQC sample. The matrix effect was calculated via the
formula:
Matrix effect (%) = X2/X1 9 100 (%)
where X1 = response of neat concentrations and X2 is
response of post-spiked concentrations
From the calculations, it was observed that IBS and
HCT showed an average (n = 6) matrix factor of 101.98 %
with a CV of 3.72 % for IBS and 101.72 % with a CV of
5.21 % for HCT at MQC level
Stability and dilution integrity
The stabilities of IBS and HCT were investigated at two
concentrations of QC samples (low and high concentra-
tions) to cover expected conditions during analysis, storage
and processing of all samples, which include the stability
data from various stability exercises like in-injector, bench
top, freeze/thaw and long-term stability tests. The stability
results summarized in Table 4 showed that IBS and HCT
spiked into human plasma was stable for at least 6 h at
room temperature, for at least 48 h in final extract at 8 �C
under autosampler storage condition, for 30 days at around
-80 �C, and during three freeze–thaw cycles when stored
at around -80 �C and thawed to room temperature. The
stock solutions and working standard of IBS, HCT and IS
were stable for 30 days at refrigerator temperature (below
10 �C) and at least for 24 h at room temperature.
In dilution integrity study, the % accuracy of two and
four times diluted sample are indicated in Table 4 and
showed %CV of less than 15 %. These results conclude
that the dilution of the concentrated plasma sample up to
four times maintains legibility and integrity of IBS and
HCT concentration.
Table 3 Recovery data of irbesartan, hydrochlorthiazide (three QC
samples) and telmisartan in human plasma
Compound Spiked conc. (ng/mL) Recovery (% ± SD)
IBS (analyte) 60 80.30 ± 7.41
250 81.91 ± 5.67
400 83.91 ± 4.11
Mean ± SD 82.04 ± 1.80
HCT (analyte) 3 84.38 ± 6.43
250 85.91 ± 6.34
400 87.25 ± 5.79
Mean ± SD 85.85 ± 1.43
Telmisartan (IS) 30 88.62 ± 9.68
Table 4 Stability and dilution integrity data of irbesartan and hydrochlorthiazide in human plasma
Stability Drug Spiked conc.
(ng/mL)
Measured conc.
(ng/mL ± SD)
Precision
(CV, %)
Accuracy
(recovery, %)
Bench top (6 h) Irbesartan 60 59.75 ± 2.75 4.61 99.58
400 399.24 ± 16.17 4.05 98.88
Hydrochlorthiazide 3 3.01 ± 0.34 11.28 100.5
400 396.00 ± 14.70 3.71 99.00
Freeze thaw (3 cycle) Irbesartan 60 61.41 ± 4.26 6.94 102.36
400 402.57 ± 11.80 2.93 100.64
Hydrochlorthiazide 3 3.15 ± 0.31 10.52 100.50
400 402.67 ± 15.77 3.91 100.66
Auto sampler (48 h) Irbesartan 60 59.41 ± 5.98 10.07 99.03
400 397.52 ± 19.40 4.88 99.38
Hydrochlorthiazide 3 2.94 ± 0.28 9.50 98.27
400 399.33 ± 13.03 3.26 99.83
30 days at -80 �C Irbesartan 60 61.25 ± 6.65 10.85 102.08
400 400.91 ± 16.32 4.07 100.22
Hydrochlorthiazide 3 2.94 ± 0.28 9.50 98.27
400 401.00 ± 11.36 2.83 100.25
Dilution integrity Irbesartan 160 157.95 ± 9.29 5.88 98.72
320 316.44 ± 8.61 2.72 98.88
Hydrochlorthiazide 160 159.62 ± 11.04 6.92 99.76
320 324.61 ± 13.16 4.05 101.44
J IRAN CHEM SOC
123
Advantages of the proposed method over the reported
methods
This study represents the first report describing the deter-
mination of IBS and HCT in human plasma by UPLC–MS/
MS method. The proposed method is superior to the pre-
viously reported LC–MS methods in terms of the sensi-
tivity, simplicity as the method described herein is based
on simple one-step protein precipitation for sample prep-
aration. The run time was only 2 min which is suitable for
high-throughput analysis and reduction in the use of
organic solvents as flow rate of 0.3 mL/min was used for
just 2 min for each sample run.
Conclusions
A novel simple, economical high-throughput and highly
sensitive UPLC–MS/MS method was successfully devel-
oped and validated for the determination of IBS and HCT
in human plasma. The method involved simple one-step
protein precipitation method for plasma sample preparation
for analysis and short runtime (2.0 min). The proposed
method could be practical and reliable for pharmacokinetic
and toxicokinetic study for IBS and HCT in humans.
Acknowledgements This research project was supported by a grant
from the ‘Research Center of the Center for Female Scientific and
Medical Colleges’, Deanship of Scientific Research, King Saud
University.
References
1. A. Lant, Drugs 29, 162 (1985)
2. A.S. Manalan, H.R. Besch Jr, A.M. Watanabe, Circ. Res. 49, 326
(1981)
3. K. Wellington, D.M. Faulds, Drugs 62, 1983 (2002)
4. K.F. Croom, M.P. Curran, K.L. Goa, C.M. Perry, Drugs 64, 999
(2004)
5. J.K. Kirk, Am. Fam. Physician 59, 3140 (1999)
6. J.M. Neutel, E. Saunders, G.L. Bakris, W.C. Cushman, K.C.
Ferdinand, E.O. Ofili, J.R. Sowers, M.A. Weber, J. Clin. Hy-
pertens. 7, 578 (2005)
7. J.M. Neutel, S.S. Franklin, P. Lapuerta, A. Bhaumik, A. Ptas-
zynska, J. Hum. Hypertens. 22, 266 (2008)
8. M.R. Weir, J.M. Neutel, A. Bhaumik, M.E.D. Obaldia, P.
Lapuerta, J. Clin. Hypertens. 9(S 12), 23 (2007)
9. R.B. Miller, C. Amestoy, J. Pharm. Biomed. Anal. 10, 541 (1992)
10. I. Niopas, A. Daftsios, J. Liq. Chromatogr. Relat. Technol. 25,
487 (2002)
11. D. Zendelovska, T. Stafilov, P. Milosevski, Biomed. Chromatogr.
18, 71 (2004)
12. A. Medvedovici, C. Mircioiu, V. David, D.S. Miron, Eur. J. Drug
Metab. Pharmacokinet. 25, 91 (2000)
13. K. Richter, R. Oertel, W. Kirch, J. Chromatogr. A 729, 293
(1996)
14. B.S. Kuo, A. Mandagere, D.R. Osborne, K.K. Hwang, Pharm.
Res. 7, 1257 (1990)
15. L. Li, J. Sun, P. Yang, Z. He, Anal. Lett. 39, 2797 (2006)
16. A. Vonaparti, M. Kazanis, I. Panderi, J. Mass Spectrom. 41, 593
(2006)
17. M.F. Tutunji, H.M. Ibrahim, M.H. Khabbas, L.F. Tutunji, J
Chromatogr. B 877, 1689 (2009)
18. M. Song, T. Hang, H. Zhao, L. Wang, P. Ge, P. Ma, Rapid
Commun. Mass Spectrom. 21, 3427 (2007)
19. H. Li, Y. Wang, Y. Jiang, Y. Tang, J. Wang, L. Zhao, J. Gu, J.
Chromatogr. B 852, 436–442 (2007)
20. N.V.S. Ramakrishna, K.N. Vishwottam, S. Manoj, M. Koteshw-
ara, S. Wishu, D.P. Varma, Biomed. Chromatogr. 19, 751 (2005)
21. W. Fang, W. Xie, J.Y.K. Hsieh, B.K. Matuszewski, J. Liq.
Chromatogr. Relat. Technol. 28, 2681 (2005)
22. R.T. Sane, M. Francis, S. Pawar, Indian Drugs 40, 104 (2003)
23. A.K. Shakya, Y.M. Al-Hiari, O.M.O. Alhamami, J. Chromatogr.
B 848, 245 (2007)
24. W.L. Wang, Q. Liu, Zhongguo Yaofang 18, 2512 (2007)
25. N. Erk, J. Chromatogr. B 784, 195 (2003)
26. S. Najma, A.M. Saeed, A.S. Shahid, S. Shahnawaz, Chin.
J. Chromatogr. 26, 544 (2008)
27. L.F. Tutunji, M.F. Tutunji, M.I. Alzoubi, M.H. Khabbas, A.I.
Arida, J. Pharm. Biomed. Anal. 51, 985 (2010)
28. R. Russo, D. Guillarme, D.T.T. Nguyen, C. Bicchi, S. Rudaz,
J.L. Veuthey, J. Chromatogr. Sci. 46, 199 (2008)
29. D.T. Nguyen, D. Guillarme, S. Rudaz, J.L. Veuthey, J. Sep. Sci.
29, 1836 (2006)
30. J.R. Mazzeo, U.V. Neue, M. Kele, R.S. Plumb, Anal. Chem. 77,
460 (2005)
31. A. de Villiers, F. Lestremau, R. Szucs, S. Gelebart, F. David, P.
Sandra, J. Chromatogr. A 1127, 60 (2006)
32. S.A. Wren, P. Tchelitcheff, J. Chromatogr. A 1119, 140 (2006)
33. L. Novakova, L. Matysova, P. Solich, Talanta 68, 908 (2006)
34. US Food and Drug Administration, Center for Drug Evaluation
and Research (CDER). Guidance for industry, bioanalytical
method validation. 2001. http://www.fda.gov/downloads/Drugs/%
e2%80%a6/Guidances/ucm070107.pdf
35. European Medicines Agency. Guideline on bioanalytical method
validation. 2011. http://www.ema.europa.eu/docs/en_GB/docu
ment_library/Scientific_guideline/2011/08/WC500109686.pdf
J IRAN CHEM SOC
123