1 2011/1/20 1 New Work Item Proposal of Automated Soluble Salt Field Measurement Standard JSTRA H. Hattori 2011/1/20 2 Contents 1. Salinity Measurement Methods Ion detection tube method Bresle Method Automated Methods Comparison 2. Example of Automated Method (SSM-21P) 3. Summary 4. Proposal
Transcript
Page 1: New Work Item Proposal of Automated Soluble Salt Field ...1 2011/1/20 1 New Work Item Proposal of Automated Soluble Salt Field Measurement Standard JSTRA H. Hattori 2011/1/20 2 Contents

1

2011/1/20 1

New Work Item Proposal of

Automated Soluble Salt Field

Measurement Standard

JSTRA

H. Hattori

2011/1/20 2

Contents

1. Salinity Measurement Methods

Ion detection tube method

Bresle Method

Automated Methods

Comparison

2. Example of Automated Method (SSM-21P)

3. Summary

4. Proposal

Page 2: New Work Item Proposal of Automated Soluble Salt Field ...1 2011/1/20 1 New Work Item Proposal of Automated Soluble Salt Field Measurement Standard JSTRA H. Hattori 2011/1/20 2 Contents

2

2011/1/20 3

Salinity Measurement Methods

ISO 8502-5 (1998)

Ion detection tube

method (wiping)

ISO 8502-6/9

(1998)

Bresle method

Automated method

(2010)

Kitagawa Electric Conductivity

“Chloride Ion” “All Soluble Salt Ions”

4

Ion Detection Tube Method

Kitagawa Chloride Detector Tube

Page 3: New Work Item Proposal of Automated Soluble Salt Field ...1 2011/1/20 1 New Work Item Proposal of Automated Soluble Salt Field Measurement Standard JSTRA H. Hattori 2011/1/20 2 Contents

3

5

ISO8502-6/9 “Bresle Method”

6

Page 4: New Work Item Proposal of Automated Soluble Salt Field ...1 2011/1/20 1 New Work Item Proposal of Automated Soluble Salt Field Measurement Standard JSTRA H. Hattori 2011/1/20 2 Contents

4

2011/1/20 7

Automated Methods for ISO 8502-6/9

From ARP Website From IMCS 2008

NST presentation

HEDON/ARP :

Soluble Salt Meter

NST Center :

Salt Smart

DKK-TOA : Surface

Salinity Meter SSM-

21P

2011/1/20 8

Comparison of Measurement MethodsIon detection tube

Method Bresle method

Automated method

SSM-21P

Sampling Wipe with gauzeBresle Patch and

injector

Fix to plate with

powered stirring

Detection

Method

Silver dichromate

method

Electric Conductivity

Method

Electric Conductivity

Method

Work Time 20 minutes 10 minutes 3 minutes

Pure Water

Consumption150ml 10ml 10ml

Sampling Area 2500cm2 12.5cm2 12.5cm2

Measurement

Calculation

・Cl - measurement

using sensor tube

・Calculate to NaCl

concentration.

・Conductivity

Measurement

・Calculate to NaCl

concentration

・Conductivity

Measurement

・Convert to NaCl

concentration by

software

Consumables

( Cost /times)

Tube,Gauze,Tapes,

Water

(25)

Patch, Water

(90)

Water

(1)

Page 5: New Work Item Proposal of Automated Soluble Salt Field ...1 2011/1/20 1 New Work Item Proposal of Automated Soluble Salt Field Measurement Standard JSTRA H. Hattori 2011/1/20 2 Contents

5

9

Less human error

Results are Reliable

Less consumables

Fast measurement

Measured readings stored

Advantages

10

Use of magnet means no glue remains

on metal surface

(unlike Bresle Method)

No glue after using automated method

Glue remains after removing Bresle patch

Page 6: New Work Item Proposal of Automated Soluble Salt Field ...1 2011/1/20 1 New Work Item Proposal of Automated Soluble Salt Field Measurement Standard JSTRA H. Hattori 2011/1/20 2 Contents

6

11

Advantages Less human error

Results are Reliable

Less running cost

Fast measurement

Measured readings stored

No Glue will remain

Disadvantage

•Initial cost is high

2011/1/20 12

Example of Automated Method (SSM-21P)

撹拌スイッチ

Cell

本体

測定セル

Pure water inlet

Main Unit

DetectorSwitch

Electric Stirrer

Permanent

Magnet

O-ring

Page 7: New Work Item Proposal of Automated Soluble Salt Field ...1 2011/1/20 1 New Work Item Proposal of Automated Soluble Salt Field Measurement Standard JSTRA H. Hattori 2011/1/20 2 Contents

7

2011/1/20 13

Mounting & Measurement

1. Mount Detector

2. Inject Pure Water

3. Activate StirrerPure water

Air

Magnet

SteelPure water

ElectrodeTemp. Sensor

Electric

Stirrer

4. Result is Displayed

2011/1/20 14

Measured Surface Can be Any Orientation

Downward-

horizontal surface

Upward-horizontal

surface

Vertical surface

Downward

inclined surface

Upward inclined

surface

Page 8: New Work Item Proposal of Automated Soluble Salt Field ...1 2011/1/20 1 New Work Item Proposal of Automated Soluble Salt Field Measurement Standard JSTRA H. Hattori 2011/1/20 2 Contents

8

2011/1/20 15

Inject Pure Water into Cell

2011/1/20 16

Extraction – Stirring

Commences Start Key on the main unit & Stirrer SW on the detector

Page 9: New Work Item Proposal of Automated Soluble Salt Field ...1 2011/1/20 1 New Work Item Proposal of Automated Soluble Salt Field Measurement Standard JSTRA H. Hattori 2011/1/20 2 Contents

9

2011/1/20 17

Measurement at block stage

SSM-21P

● Majority of Japanese shipyards are already

using Model SSM-14P & SSM-21P extensively.

2011/1/20 18

Summary

Automated Soluble Salt Measurement Method has the following advantages:

Higher accuracy Higher productivity Higher reproducibility Eco-friendliness

Page 10: New Work Item Proposal of Automated Soluble Salt Field ...1 2011/1/20 1 New Work Item Proposal of Automated Soluble Salt Field Measurement Standard JSTRA H. Hattori 2011/1/20 2 Contents

10

2011/1/20 19

Proposal

We would like to propose to standardize

automated soluble salt measurement for

reducing environmental load and

development of industry with

implementation of advanced technology.

2011/1/20 20

Thank you for your attention