+ All Categories
Home > Documents > QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique •...

QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique •...

Date post: 04-Jun-2018
Category:
Upload: phungdan
View: 221 times
Download: 0 times
Share this document with a friend
25
QSIT QUALITY SYSTEM INSPECTION TECHNIQUE Robert L. Turocy May 6, 2002
Transcript
Page 1: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

QSIT• QUALITY SYSTEM

INSPECTIONTECHNIQUE

• Robert L. Turocy May 6, 2002

Page 2: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

WHAT SPECIFICALLY DOYOU WANT OR NEED TO

KNOW ABOUT QSIT??�MANAGEMENT�QUALITY & REGULATORY�ENGINEERING�PRODUCTION, - - - , SERVICE

Page 3: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

WHAT IS QSIT?

�QSIT IS AN OPTIONAL FDAINSPECTION PROCESS

�QUALITY SYSTEM ORIENTED�TOP DOWN VERSUS BOTTOM UP�PRE-INSPECTION ACTIVITIES�SAMPLING�FOCUS ON MANAGEMENT

Page 4: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

QSIT PILOT INSPECTIONSFDA 483s

0

10

20

30

40

50

60

M C P D O

• M = MGMT. 57• C = CAPA 50• P = PAPC 45• D = DESIGN 26• O = OTHER 22

Page 5: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

WHY DOES THE FDA USEQSIT?

� QSIT IS FOCUSED, HARMONIZED, EFFICIENT,INCREASES COMPLIANCE, & MOSTIMPORTANT, QSIT ASSISTS IN THEPROTECTING THE PUBLIC FROM UNSAFEMEDICAL DEVICES

� USED TO DETERMINE IF A MANUFACTURER’SQUALITY SYSTEM IS CONFORMING WITHREGULATIONS

Page 6: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

WHO DEVELOPED QSIT?

�FDA REENGINEERING EFFORT

�ASSISTED BY INDUSTRY, TRADEASSOCIATIONS, ANDCONSULTANTS

�FDLI FACILITATED MEETINGS,ETC.

Page 7: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

HOW IS QSITIMPLEMENTED?

�LEVEL 1, (Abbreviated)�CAPA + 1 OTHER SUBSYSTEM

�LEVEL 2, (Baseline)�MGMT CONTROLS, DESIGN CONTROLS,

CAPA, PAPC, AND RETURN TO MGMTCONTROLS

�LEVEL 3, (Compliance Follow-up toOFFICIAL ACTION INDICATED)

Page 8: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

WHEN IS QSIT USED?

�MANUFACTURER’S COMPLIANCEHISTORY IS THE MAJOR FACTOR

�RISK OF DEVICE

Page 9: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

DOES QSIT WORK?�YES!!! PILOT INSPECTION RESULTS

�QSIT IS HARMONIZED WITH THEEU PROCESS OF INSPECTIONS

�QSIT REDUCED IN-PLANT TIMEFROM APPROXIMATELY 67 HOURSTO 28 IN PLANT HOURS

Page 10: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

BREAK DOWN OF 28HOURS

0

5

10

15

20

25

30

M D C P T

� M = MGMT 4.2� D = DESIGN 5.2� C = CAPA 10.7� P = PAPC 8.1� T = TOTAL 28.2

6 IN-PLANT HOURSEQUALS 1 DAY

Page 11: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

QSIT DOES

�REVIEW THE QUALITY SYSTEM

�VALIDATE “ESTABLISHED”

�REVIEW MANAGEMENT

Page 12: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

QSIT DOES NOT

�ELIMINATE “FOR CAUSE”INSPECTIONS

�FIND AN INFINITE NUMBER OFPRODUCT PROBLEMS

Page 13: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

MANAGEMENT CONTROLS� QUALITY POLICY� MANAGEMENT REVIEW� INTERNAL QUALITY AUDIT� QUALITY PLAN� QUALITY SYSTEM PROCEDURES� MANAGEMENT REPRESENTATIVE� SUITABLITY & EFFECTIVENESS� ORGANIZATIONAL STRUCTURE,

RESPONSIBILITY, AUTHORITY, &RESOURCES

Page 14: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

DESIGN CONTROLS

�WHEN ARE DESIGN CONTROLSREVIEWED?

� SUBMIT PDP?� SUBMIT IRB?� SUBMIT IDE?� SUBMIT 510(K)?� SUBMIT PMA?� MARKET DEVICE?

Page 15: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

DESIGN CONTROLS cont.

� DC PROCEDURES� DESIGN PLAN� DESIGN INPUTS� ACCEPTANCE

CRETERIA� DESIGN OUTPUTS� DESIGN

VERIFICATION� DESIGN VALIDATION

� SW VALIDATION� RISK ANALYSIS� PRODUCTION UNIT

VALIDATED� DESIGN CHANGE

CONTROL� DESIGN REVIEWS� DESIGN TRANSFER� DESIGN HISTORY

FILE

Page 16: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

CORRECTIVE & PREVENTATIVEACTION (CAPA)

� CAPA PROCEDURES� INFORMATION SOURCES IDENTIFIED� INFORMATION ANALYZED� COMPLETE, ACCURATE, & TIMELY

INFORMATION� STATISTICAL METHODS� FAILURE ANALYSIS VERSUS RISK� ROOT CAUSE ANALYSIS� APPROPRIATE CAPA TAKEN & DOCUMENTED� SHARE INFORMATION – MANAGEMENT

REVIEW

Page 17: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

PRODUCTION & PROCESSCONTROL (PAPC)

� PAPC PROCEDURES� CONTROLS & MONITORS� DEVICE HISTORY RECORDS� NON-CONFORMING ACTIONS� EQUIPMENT ADJUSTMENT,

CALIBRATION, & MAINTENANCE� VALIDATION OF PROCESSES� SW VALIDATION� PERSONNEL QUALIFICATIONS

Page 18: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

LINKAGES TO OTHERSUBSYSTEMS

�MATERIAL CONTROLS

�RECORDS/DOCUMENTS/CHANGECONTROLS

�FACILITY & EQUIPMENTCONTROLS

Page 19: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

MEDICAL DEVICE REPORTING(MDR) A SATELITE TO CAPA

�MDR PROCEDURES�MDR FILES ESTABLISHED�MDR INFORMATION COMPLETE�DEATHS, SI & SI, AND

MALFUNCTIONS

Page 20: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

CORRECTIONS &REMOVALS (CAR) ASATELITE TO CAPA

�CAR PROCEDURES�CARs SUBMITTED�CARs COMPLETED�CAR FILE ESTABLISHED

Page 21: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

MEDICAL DEVICETRACKING A SATELITE TO

CAPA

• FAILURE CAUSES ADVERSE HEALTHCONSEQUENCES

• OBLIGATION FOR TRACEABILITY

Page 22: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

SAMPLING CONFIDENCE LIMIT OF (.99) MEANS THAT WE ACCEPT A 99%PROBABILITY THAT NO MORE THAN 10% OF THE REMAINING

CASES DO NOT MEET OUR EXPECTATION. THIS IS BASED ON THEFACT THAT WE FIND “O” BAD CASES OUT OF 51 SAMPLES.

TABLE 2 BINOMIAL SAMPLING

190161107.05 UCLF

907351.10 UCLE21 OUT OF1 OUT OFO OUT OF

Page 23: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

STERILIZATION PROCESSCONTROLS A SATELITE TO

PAPC� STERILIZATION PROCEDURES� PROCESS VALIDATED� PROCESS CONTROLLED & MONITORED� APPROPRIATE HANDLING OF NON-

CONFORMANCES� EQUIPMENT ADJUSTMENT, CALIBRATION, &

MAINTENANCE� SW VALIDATION� PERSONNEL QUALIFIED & TRAINED

Page 24: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

QSIT REFERENCES

�21 CFR 820, Preamble & Regulation�***QSIT HANDBOOK GUIDE�CD ROM COMPUTER BASED

TRAINING�http://www.fda.gov/cdrh/dsma/cgmphome.h

tml�INSP. DEVICE MFGRS. CP 7382.845

Page 25: QUALITY SYSTEM INSPECTION TECHNIQUE - IMDRF · qsit • quality system inspection technique • robert l. turocy may 6, 2002. what specifically do ... qsit pilot inspections fda 483s

XQSIT


Recommended