RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY © Corporation Scientifique Claisse ® , 2013. All Rights Reserved. Anne-Catherine Breton, Mathieu Bouchard, John Anzelmo and Sébastien Rivard Corporation Scientifique Claisse, Québec, Canada
Transcript
Page 1: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

RAW MATERIALS XRF APPLICATION USING AUNIVERSAL BORATE FUSION METHODOLOGY

© Corporation Scientifique Claisse®, 2013. All Rights Reserved.

Anne-Catherine Breton, Mathieu Bouchard, John Anzelmo and Sébastien Rivard

Corporation Scientifique Claisse, Québec, Canada

Page 2: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

21ST CENTURY INDUSTRY REALITY

• Raw Materials:– Natural ores and rocks– Use of industrial by-products– Not necessarily obtain from well

known sources

• Fuels: – Less expensive fuel and waste

2

Page 3: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

LAB MANAGERS REALITY

• Analyze a very wide range of sample types(raw materials, in process samples, final products)

• Meet requirements of standard test methods

• Laboratory operators and/or technicians with limited scientific background

• Limit cost of analysis– Need safe, simple and efficient analytical methods– Need to limit the sources of manipulation errors– Need to limit the use of reference materials and/or pure

oxides

3

Page 4: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

4PRESSED POWDERS FOR XRF

Pressed pellets can generate good accuracy only when:

• A distinct calibration curve is made for every type of matrix

• Standards that match exactly the unknown samples for particle size and mineralogy are available

• Grinding and pressing conditions are optimized and rigorously repeated

• The standards and the samples can be ground enough to reduce the particle size effect significantly

• The pellets are stable enough (use of binder) to avoid causing cross-contamination

Page 5: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

5

FUSION FOR XRF

• No particle effect

• All elements transformed into simple oxides

• No mineralogical effect

• All standard and unknown specimens have the same matrix which is a borate glass

• Calibration is much simplerand flexible

Page 6: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

MATERIAL TYPES

• Cement• Blended cement• Cement with additions• Aluminate cement• Clinker• Kiln feed• Raw Mix• Limestone

• Gypsum• Sand• Clay• Bauxite• Silica Fume• Slag• Fly ash• Iron Ore

6

.

Page 7: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

DEVELOPMENT OF THE GLOBAL FUSION METHOD

• 200 different samples from 14 material types

• Both ignited and non ignited samples

• 3 different sample:flux ratios

• We tried different dry oxidation methods

7

Page 8: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

THE EQUIPMENT USED

Muffle Furnace:

- Fisher ScientificIsotemp® Programmable- Max. Temp.: 1125°C- Power: 50/60 Hz 208/240 VAC

Spectrometer:

- Bruker S4 Explorer- WDXRF- Tube: 1000 Watts

8

Automatic Fluxer:

- Claisse M4 or TheOxFluxer

Page 9: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

THE EQUIPMENT USED - FLUXER

Automatic Electric Fluxer:Automatic Gas Fluxer:

- Claisse® M4TM Fluxer- Propane/Air Heating- 3 Positions

- Claisse® TheOxTM Fluxer- Electrical Heating- 6 Positions

11

Page 10: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

CEMENT AND RAW MATERIALS FUSION METHOD

• Ignition: Necessary for All Cement related and Raw Materials

950oC for 1 hour

• Crucible: Claisse Optimix crucible, Pt-Au alloy (95%-5%)

• Mold: 32mm Pt-Au alloy (95%-5%)

• Flux: Li2B4O7 / LiBO2 49.75% / 49.75% flux with integrated

LiBr non-wetting agent 0.50%

• Ratio: 1:10 sample to flux

• Mixing: 15 seconds on a vortex mixer

• Fusion time: 18 min. (including cooling) for M4 Fluxer

22 min. (including cooling) for TheOx

10

Page 11: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

CALIBRATION STANDARDS

• The calibration of the WDXRF raw materials application was executed using a wide variety of RMs

• Bureau of Analysed Samples Ltd. (BAS): British Chemical Standard Certified Reference Materials

• Domtar® Inc. Research Center: Canadian Certified Reference Materials• China National Analysis Center for Iron and Steel: NCS DC Reference Material• European Committee for Iron and Steel Standardization • European Coal and Steel Community (ECSC): Euro-Standard• Geological Institute for Chemical Minerals: GBW Reference Material• Institut de Recherches de la Sidérurgie (IRSID) : Échantillon-Type• Instituto de Pesquisas Tecnológicas (IPT) : Reference Material• Japan Cement Association (JCA): Reference Materials for X-ray Fluorescence

Analysis • National Institute of Standards & Technology (NIST): Standard Reference

Material®• Slovak Institute of Metrology (SMU): Slovak Reference Material• South Africa Bureau of Standards: SARM Certified Reference Material

12

Page 12: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

CALIBRATION STANDARDS12

RM element concentration as an oxide equivalent

Page 13: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

MATERIAL TYPES

• Not included in this analytical process1) Copper Slags2) Other samples with copper at higher level than traces3) Some Iron Ores

13

• Easy to fuse

Page 14: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

14CALIBRATION

Inter-element corrections and squarred correlation coefficients for raw materials application

Page 15: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

SENSITIVITY, PRECISION AND ACCURACY RESULTS15

Raw Materials application results

Page 16: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

CONCLUSION

• Borate fusion allows to developed a robust and universalsample preparation method which is able to fuse the largerange of raw materials and cements

• It is possible to obtain a precise and accurate universalcalibration, able to cover a wide range of raw materials

• Precise and accurate results can be obtained using bothtype of fluxers: TheOx or M4.

16

Page 17: Raw Materials XRF application using a universal borate ... · RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY ... – Need to limit the sources of manipulation

QUESTIONS AND COMMENTS

If you have questions, please feel free to ask me now or later at HRT Labortechnik’s booth.

Thank you for your attention!