Revital-Ox™ RESERT™ R60 Solution Test Strips
Document # M4208EN.2015-09, Rev. A ©2015 STERIS Corporation. All Rights Reserved.
For further information, please contact STERIS in the
United States at 800-548-4873, in Canada at 800-661-3937
or visit our website at www.steris.com.
Quality Test for Revital-Ox RESERT Solution Test StripFrequency: Upon receipt of each shipment, test each lot in the shipment1. Prepare test solutions.
PASS Solution Dispense 30 ml of Revital-Ox RESERT solution from a newly opened bottle. Label the container.
FAIL Solution Mix 15 ml of tap water and 15 ml of Revital-Ox RESERT solution from a newly opened bottle.
Label the container.
2. Confirm that each solution’s temperature is ≥ 20˚C - 68˚F.
3. Remove three strips from the bottle. Test each strip per steps 4-6 above. Record the results.
4. Remove three additional strips from the bottle. Test each strip per steps 4-6 above. Record the results.
5. Interpretation:
a. Passing Testi. All three indicator strips pass in PASS Solutionii. All three indicator strips fail in FAIL Solution
b. Failing test – Do Not Usei. One or more strips fail in PASS Solutionii. One or more strips pass in FAIL Solution
Disclaimer: This is general information to assist users. Read and follow all product labeling and instructions for use.Revital-Ox® RESERT® solutions Include:• Revital-Ox® RESERT® High Level Disinfectant• Revital-Ox® RESERT® High Level Disinfectant-Chemosterilant• RESERT® XL HLD High Level Disinfectant
Unexposed / Fail FailPass
Frequency: Prior to each use of Revital-Ox® RESERT® solutions to confirm MRC (Minimum Recommended Concentration)
Materials: Required for MRC testing Solution Test Strips Timer Paper Towels Container Thermometer
Confirmation of MRC for Revital-Ox RESERT Solutions using Revital-Ox RESERT R60 Solution Test StripsNOTE: Refer to the instruction for use of the Revital-Ox RESERT solutions and Automated Endoscope Reprocessor (AER) for detailed high level disinfection instructions.
1. Wear appropriate PPE for working with Revital-Ox RESERT solution.
2. Confirm expiration date of Revital-Ox RESERT solution and test strips. Do not use Revital-Ox RESERT solution beyond expiration regardless of test strip result.
Not all products available in all markets.
3. Check solution’s temperature.
a. AER – Dispense a 30 ml specimen of solution from AER (per AER instructions). Confirm temperature is ≥ 20˚C/68˚F before proceeding.
b. Manual soaking – Confirm temperature of solution in the basin is ≥ 20˚C/68˚F before proceeding.
4. Remove one test strip and dip (Either in 30ml specimen or soak basin) for two seconds. Remove from the solution and start a 60 second timer.
5. Remove excess solution from strip pad by touching the short edge to a paper towel. Do not blot test strip.
6. At 60 seconds, interpret strip color through the viewing window.