+ All Categories
Home > Documents > Safety Aspects of Magnetic Resonance Imaging455

Safety Aspects of Magnetic Resonance Imaging455

Date post: 05-Apr-2018
Category:
Upload: xis182
View: 218 times
Download: 0 times
Share this document with a friend

of 41

Transcript
  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    1/41

    HOW I DO CMR SCANNING SAFELY

    Elisabeth Burman

    Research Sister

    Royal Brompton Hospital, London UK

    This presentation is posted for members of scmr as an

    educational guideit represents the views and practices of

    the author, and not necessarily those of SCMR.

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    2/41

    Magnetic ResonanceOne of the safest imaging modalities but there

    are still safety issuesAdvantages

    Non-invasive

    No ionizing radiation

    Repeat studies present no problem

    Fewer ethical problems imaging volunteers

    Why are there safety issues? Potential for accidents and even fatalities

    Damage to equipment / cost

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    3/41

    Static magnetic field biological effectsmechanical effects

    Gradient Field induced currents (PNS)auditory damage

    RF Field RF power depositionBurns

    Otherissues contrast agents

    critically ill patients/resuscitationmonitoring

    quench

    claustrophobia

    Safety issues centre around the 3 electromagnetic fields

    and issues indirectly related

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    4/41

    Static magnetic field

    1.5 T is 30,000 x the strength of earths magnetic field

    Measured in Gauss or Tesla (10,000G equivalent to 1T)

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    5/41

    Static Magnetic Fields

    1. Biological effects (potential risk)

    - exposure to static magnetic fields of up to 4T are not thoughtto be harmful

    Biological effects relevant to clinical imaging- distorted ECG (magnetohydrodynamic effect)

    - consider prudency with pregnancy

    2. Mechanical effects (very real risk)

    - translational or attractiveforces on metallic objects whenbrought into the field

    A superconducting magnet is always switched on!

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    6/41

    Static Magnetic Field

    Biological effects

    Magnetohydrodynamic effectaugmented T wave

    Outside field

    0.5 Tesla

    1.5T

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    7/41

    Magnetohydrodynamic effect

    seen as augmentation of T-wave Caused by the effect of the static magnetic field on

    moving blood (systole) as a conducting fluid.

    The gradient and RF fields also affect the configurationof the ECG

    Morphological ECG changes are therefore difficult todetect and diagnose, but rhythm is usually recognised

    Any concern regarding rhythm, remove patient fromscanner and perform 12 lead ECG

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    8/41

    Static Magnetic Field

    Pregnancy

    Patients 1st trimesterprudent to avoid CMR where possible

    2nd and 3rd trimesterdecision made on a risk versusbenefit determination. For example if it avoids the patient

    being subjected to x-rays.

    Health Care Workers

    May enter MR scanning room regardless of trimester

    Should not remain in the room when scanner is operational,avoiding exposure to gradient and radiofrequency fields

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    9/41

    Static Magnetic Field

    Mechanical effects

    Projectile or missile effect- the attractive forces exerted by the static magneticfield present the greatest potential for patient injury

    - objects will be pulled out of hands, pockets etc, and fly intomagnet which has caused injury and death.

    Effect on ferromagnetic implants

    - electro-mechanical eg pacemakers

    - biomedical eg valves, stents

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    10/41

    What is typically ferromagnetic?

    EQUIPMENT PERSONAL ITEMS (leave outside) Oxygen cylinders Keys, pens

    Wheelchairs Bleeps

    Trolleys Mobile phones

    IV stands Coins Monitoring equipment Stethoscopes

    Ventilators Scissors

    It is easy to forget objects, particularly when

    responding to an emergency!

    Remember alsomagnetic strips will be wiped (credit cards)

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    11/41

    Warnings to minimise accidents?

    OUTSIDE THE SCANNER ROOM

    Signage

    Barrier cords to scanner room

    Restricted access - swipe card / coded entry system

    INSIDE THE SCANNER ROOM (fringe field5 Gauss line)Older systems - passively shielded

    - extensive fringe field gave a warning

    Newer systems - actively shielded

    - small fringe field greater hazard because there is no gradual warning

    field strength increases hundreds of gauss in a few inches

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    12/41

    Never be complacentaccidents do happen

    Oxygen cylinder Infusion pump

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    13/41

    Floor buffer

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    14/41

    Cerebral aneurysm clips

    Metallic foreign body in the eye Shrapnel, bullets (in critical area)

    Ocular implants (containing metal)

    Swan-Ganz

    Contraindications to CMR*

    Implants & metal Electromechanical implan

    Pacemakers /ICDs

    Pacing wires

    Cochlear implants

    Neurostimulators

    Hydrocephalus shunts

    Any device electrically or mechanically activated* In some circumstances MR has been performed despite contraindications, especially in specialist centres.

    Work is in progress to make some devices safe. For example, some cerebral aneurysm clips may now be scanned.

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    15/41

    Implants

    Cerebral clips modern clips are considered safe (titanium, elgiloy)

    older clips(martensic steel is highly ferrous)

    obtain operation notes with serial number of clip

    radiologist to consent patient seek advice from specialist centre

    Foreign bodies

    Maybe situated near vascular or nervous tissue If in doubtx-ray

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    16/41

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    17/41

    Potential effects of MRI on pacemakers

    excessive heating / induced currents in the

    pacemaker lead

    temporary or permanent modification of

    pacemaker function

    inappropriate sensing or triggering of the device

    movement of the pulse generator or lead

    pacing the heart

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    18/41

    What can be safely scanned at 1.5T?Lists of approved implants are available

    CARDIAC

    sternal wires heart valves & annuloplasty rings

    coronary stents epicardial wires (cut short and taped)

    othervascular stents, coils and filters

    If unsure, best to check atcheck: www.mrisafety.com

    occlusion devices

    catheters (without guidewire)

    BIOMEDICAL IMPLANTSOTHER

    orthopaedic implants contraceptive devices

    dental implants

    Effects of implants on imaging -signal void / local distortion

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    19/41

    Safety Checklistcomprehensive but concise

    Removal of accessories - watch, jewellery (except

    wedding rings), body piercing rings, hearing aids, glasses,false teeth, artificial limbs and prostheses

    Removal of clothes containing metal eg zips, bras

    Important to know

    Previous heart surgery?

    Diabetic or epileptic?

    Asthmatic or allergies?

    (in relation to contrast)

    Tattoos or permanent eye liner(iron oxide)?

    Essential to know

    Cardiac pacemaker? Previous neurosurgery?

    Implants or metal in the body?

    Pregnant (prudent approach)?

    Drug patch with foil backing?

    Essential preparation

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    20/41

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    21/41

    Gradient fields - Induced Currents

    Gradient fields induce an electric field and thus

    a current in the patient, potentially this can be

    of sufficient intensity in modern systems to

    produce a physiological response

    - peripheral nerve stimulation (PNS)

    - cardiac stimulation is not considered possible

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    22/41

    Hearing protection mandatory above 90dB

    time averaged for:

    patients staff remaining in the scanner room

    relatives accompanying children or patients

    Gradient fieldsAuditory damage

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    23/41

    Radiofrequency (RF) fields

    Thermogenic effects - health & safety concerPhysiological tissue heating response

    most of the transmitted RF power is transferred into heat

    within the patients tissue all MR systems have safety thresholds to avoid

    dangerous levels

    Patients with compromised thermoregularory systems are

    at greatest risk

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    24/41

    Specific Absorption Rate (SAR)

    SAR is the RF power absorbed per unit mass of tissue

    (expressed in W/kg)

    complex function of numerous variables

    calculated by software from the average forwardpower passing into the RF transmitter coil and thebody mass situated in the RF transmitting fieldTherefore an accurate patient weight is vital

    SAR increases 9 fold from 0.5T to 1.5T

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    25/41

    Radiofrequency (RF) fields

    Potential for burns

    1, 2, 3 burns have occurred in the past in

    patients undergoing MRI

    This is a result of excessive heat developing in the

    devices or objects

    ECG system is often the culprit Interventional MRI poses greater risk

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    26/41

    Prevention of burns

    Electrodes - carbon fibre studs

    - placed close together

    ECG leads - carbon fibre

    - fibre optic

    - high impedance

    - short as possible (plaited if necessary)

    All conductive leads should be placed in a linear fashioncoming out of bore of scanner

    Avoid crossed limbs where possible

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    27/41

    RF Burn from non-Carbon Electrode

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    28/41

    Other Issues

    Contrast agents

    Critically ill patients / resuscitation

    Monitoring and equipment related hazards

    Quench of superconducting magnets

    Claustrophobia

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    29/41

    Contrast agents

    all gadolinium chelates

    overall patient tolerance high incidence of adverse reactions very low

    Possible reactions:- headache, nausea and vomiting, pain if extravasation occurs

    anaphylaxis 1:100,000

    Contraindications pregnancy (risk versus benefit ratio)

    breast feeding mothers infants < 2 years

    Caution asthmatics, history of previous reactions

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    30/41

    Critically ill patients

    Assess patient the day before

    Lengthen and prime IV lines on ward

    Replace electrodes - carbon studs Lengthen chest drains

    Check intubation circuits for metal

    Educate nursing staff in preparing patient

    Ensure status of patient has not changed since referral Screen all accompanying personnel

    AICU and High Dependency Patients

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    31/41

    Sedated patients

    Departmental sedation protocol with anaestheticapproval

    Nil by mouth for 6 hrs (food & milk) 2 hrs water

    Medical examination

    Informed consent

    Tipping trolley, piped oxygen, full resuscitation

    equipment

    Responsible adult to collect from department and remainwith for 24 hours

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    32/41

    Resuscitation

    Alert cardiac arrest team Call for help

    BLS only can be performed in scanner room

    Keep MR Compatible tipping trolley adjacent to

    scanner Patient must be moved to designated resuscitation area asquickly as possible

    Appoint non-clinical person to prevent access to scannerroom

    Practice regular cardiac arrest scenarios to identifypotential problems

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    33/41

    Monitoring

    MR Safe

    Device when used in the MR environment, has beendemonstrated to present no additional risk to the patient

    BUT may affect the quality of diagnostic information

    MR Compatible

    Device has been demonstrated to neither significantly

    affect the quality of the diagnostic information or to beoperationally affected by the scanner in use

    Equipment must be MR Safe or MR Compatible

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    34/41

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    35/41

    Elective Quenching

    The magnet should only be quenched in twosituations:-

    If someone is trapped to the scanner and isinjured and/or distressed

    If there is a fire in the immediate vicinity onorder to reduce risk to the Fire Brigade

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    36/41

    Action to be taken in the event of

    a Quench Evacuate the room as quickly as possible Ensure the door is kept open during evacuation

    Close door after evacuation If trapped in room stay close to floor level Seek the advice of a senior physicist immediately Call scanner engineer

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    37/41

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    38/41

    Factors to reduce anxiety

    education and explanation trial visit to the department

    maintaining physical and verbal contact

    presence of a relative or friend

    use of a mirror, prism glasses good communication system

    alarm

    good light and ventilation

    music pleasant thoughts

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    39/41

    Ensuring Safe Practice

    Safety Checklistcomprehensive but concise

    Restricted access to magnetic field area

    MR Compatible equipmentmonitors, wheelchairs,syringe pumps, stethoscopes

    Awareness of medication common to patient areacontrast agents, stress agents

    Staff Induction

    Education and training of staff of all disciplines

    Risk Assessments

    Policies and Procedurescardiac arrest, quench

    This co ld be o !

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    40/41

    This could be you!

    References

  • 7/31/2019 Safety Aspects of Magnetic Resonance Imaging455

    41/41

    References

    Shellock F.G. and Kanal E. (2001)

    - Magnetic Resonance Bioeffects, Safety, and PatientManagement

    Health Protection Agency (HPA)

    (was National Radiological Protection Board -NRPB)

    International Electrical Committee (IEC) Medical and Healthcare products Regulatory Agency

    (MRHA) (was Medical Devices Agency - MDA)

    http://www.mrisafety.com (Shellock and Kanal)

    http://www.imrser.org/ http://www.magneticresonancesafetytesting.com/


Recommended