SCOPE OF WORK and TECHNICAL REQUIREMENTS Laboratory Safety Equipment and Calibration Purpose: According to the regulation issued by Ministry of Health Government of Vietnam, all lab equipment must be annually calibrated, especially, the laboratories that want to obtain biosafety certificate by authorities’ agencies. These equipment (e.g. Biosafety Cabinet -BSC, autoclave, incubator pipette) must be calibrated regularly. National Institute of Hygiene and Epidemiology, Pasteur Institute in Nha Trang, Pasteur Institute in Ho Chi Minh City have received training on BSC certification but require additional equipment for performing calibration services. Accordingly, U.S. CDC will support 3 Regional Institutions in providing those basic lab equipment items for calibration. Listed below are brand names equivalent can be quoted. I. Description of Goods No. Description Specification Quantity 1 Smoke Generator 1 2 Vibration Meter 1 Features: - Frequency range 10 Hz - 1 kHz, sensitivity relative meet ISO 2954 - Professional vibration meter supply with vibration sensor & magnetic base, full set. - Metric & Imperial display unit - Acceleration, Velocity, Displacement measurement. - RMS, Peak value, Max. Hold measurement. - Wide frequency range. - Data hold button to freeze the desired reading. - Memory function to record maximum and minimum reading with recall. - Separate vibration probe with magnetic base, easy operation. - RS 232 computer interface. - Data Logger. - Optional data acquisition software. - Optional data logger ( data collection ) software. - Super large LCD display with bar graph indicator - Microcomputer circuit, high performance. - Auto shut off saves battery life. - Built-in low battery indicator - Heavy duty & compact housing case - Complete set with the hard carrying case. General Specifications: - Display: 52 mm x 38 mm, LCD display. 16 mm ( 0.63" ) digit size. With bar graph indicator.
Transcript

SCOPE OF WORK and TECHNICAL REQUIREMENTS

Laboratory Safety Equipment and Calibration

Purpose: According to the regulation issued by Ministry of Health – Government of Vietnam, all lab

equipment must be annually calibrated, especially, the laboratories that want to obtain biosafety

certificate by authorities’ agencies. These equipment (e.g. Biosafety Cabinet -BSC, autoclave,

incubator pipette) must be calibrated regularly. National Institute of Hygiene and Epidemiology,

Pasteur Institute in Nha Trang, Pasteur Institute in Ho Chi Minh City have received training on BSC

certification but require additional equipment for performing calibration services. Accordingly, U.S.

CDC will support 3 Regional Institutions in providing those basic lab equipment items for calibration.

Listed below are brand names – equivalent can be quoted.

I. Description of Goods

No. Description – Specification Quantity

1 Smoke Generator 1

2 Vibration Meter 1

Features:

- Frequency range 10 Hz - 1 kHz, sensitivity relative meet ISO 2954

- Professional vibration meter supply with vibration sensor & magnetic

base, full set.

- Metric & Imperial display unit

- Acceleration, Velocity, Displacement measurement.

- RMS, Peak value, Max. Hold measurement.

- Wide frequency range.

- Data hold button to freeze the desired reading.

- Memory function to record maximum and minimum reading with recall.

- Separate vibration probe with magnetic base, easy operation.

- RS 232 computer interface.

- Data Logger.

- Optional data acquisition software.

- Optional data logger ( data collection ) software.

- Super large LCD display with bar graph indicator

- Microcomputer circuit, high performance.

- Auto shut off saves battery life.

- Built-in low battery indicator

- Heavy duty & compact housing case

- Complete set with the hard carrying case.

General Specifications:

- Display: 52 mm x 38 mm, LCD display. 16 mm ( 0.63" ) digit size. With

bar graph indicator.

- Measurement: Velocity, Acceleration, Displacement

- Function: Main: RMS, Peak, Max. Hold.; Others: Data hold, Max. &

Min. value, Data logger.

- Frequency range: 10 Hz to 1 KHz

- Circuit: Exclusive microcomputer circuit.

- Data hold: Freeze the desired reading.

- Peak measurement: To measure the peak value.

- Max. hold measurement: To measure and update the max. peak value.

- Memory: Maximum & Minimum value

- Power off: Auto shut off, saves battery life, or manual off by push

button.

- Sampling time: Approx. 1 second.

- Sampling Time of Data Logger: 0, 1, 2, 10, 30, 60, 600, 1800, 3600 sec.

- Data Logger No.: 500 no. max.

- Data output: RS 232 serial output, isolate.

- Operating temperature: 0 to 50 (32 to 122 ).

- Operating humidity: Less than 80% RH.

- Power supply: Alkaline or heavy duty type, DC 9V battery, 006P,

MN1604 (PP3) or equivalent.

- Power consumption: Approx. DC 13 mA.

- Weight: Meter: 230 g/0.50 LB; Vibration sensor: 38 g/0.09 LB

- Dimension: Meter: 180 x 72 x 32 mm (7.1 x 2.8 x1.3 inch ).

Vibration sensor probe: Round 19 mm Dia. x 21 mm.

Electrical Specifications:

Acceleration ( RMS, Peak, Max Hold )

- Unit: m/s^2

- Range: 0.5 to 199.9 m/s^2

- Resolution: 0.1 m/s^2

- Accuracy: ± ( 5 % + 2 d ) reading; @160Hz, 80Hz, 23 ± 5oC

- Calibration Point: 50 m/s^2 (160 Hz)

- Remark: Acceleration peak range : 1.0 to 199.9 m/s^2

- Unit: G ; @ 1G = 9.8 m/s^2

- Range: 0.05 to 20.39 G

- Resolution: 0.01 G

- Accuracy: ± ( 5 % + 2 d ) reading; @160Hz, 80Hz, 23 ± 5oC

- Calibration Point: 50 m/s^2 (160 Hz)

- Remark: Acceleration peak range : 1.0 to 199.9 m/s^2

- Unit: ft/s^2

- Range: 2 to 656

- Resolution: 1 ft/s^2

- Accuracy: ± ( 5 % + 2 d ) reading; @160Hz, 80Hz, 23 ± 5oC

- Calibration Point: 50 m/s^2 (160 Hz)

- Remark: Acceleration peak range : 1.0 to 199.9 m/s^2

Velocity ( RMS, Peak, Max Hold )

- Unit: mm/s

- Range: 0.5 to 199.9 mm/s

- Resolution: 0. 1 mm/s

- Accuracy: ± ( 5 % + 2 d ) reading; @160Hz, 80Hz, 23 ± 5oC

- Calibration Point: 50 mm/s ( 160 Hz )

- Remark: Velocity peak range : 1.0 to 199.9 mm/s

- Unit: cm/s

- Range: 0.05 to 19.99 cm/s

- Resolution: 0. 01 cm/s

- Accuracy: ± ( 5 % + 2 d ) reading; @160Hz, 80Hz, 23 ± 5oC

- Calibration Point: 50 mm/s ( 160 Hz )

- Remark: Velocity peak range : 1.0 to 199.9 mm/s

- Unit: inch/s

- Range: 0.02 to 7.87 inch/s

- Resolution: 0.01 inch/s

- Accuracy: ± ( 5 % + 2 d ) reading; @160Hz, 80Hz, 23 ± 5oC

- Calibration Point: 50 mm/s ( 160 Hz )

- Remark: Velocity peak range : 1.0 to 199.9 mm/s

Displacement p-p ( RMS, Max Hold )

- Unit: mm

- Range: 1.999 mm

- Resolution: 0.001 mm

- Accuracy: ± ( 5 % + 2 d ) reading; @160Hz, 80Hz, 23 ± 5oC

- Calibration Point: 0.141 mm ( 160 Hz )

- Unit: inch

- Range: 0.078 inch

- Resolution: 0.001 inch

- Accuracy: ± ( 5 % + 2 d ) reading; @160Hz, 80Hz, 23 ± 5oC

- Calibration Point: 0.141 mm ( 160 Hz )

3 Analytical Balances 1

Features:

* Accurate Results:

- High resolution weighing cell (HRT ) with two built-in adjustment

- proFACT - fully automatic time- and temperature-controlled internal

adjustment

- Low minimum weight

- Full metal housing

* Efficient Operation:

- HCD High contrast display (18 mm high digits)

- Easy cleaning, incl. QuickLock for easy glass panel removal

- ErgoDoor left/right operation

- Telescopic draft shield doors for reduced footprint

* Quality Assurance:

- Menulock (Settings protection)

- PipetteCheck with evaporation trap (optional)

* Seamless Process:

- 2 interfaces - USB and RS232

- Bluetooth option

- PC Direct application for easy data transfer (no software needed)

Specifications:

Maximum capacity (fine range/full range): 120 g

Readability, full range: 0.01 mg

Repeatability (at nominal load): 0.04 mg

Repeatability (at low load): 0.02 mg (10g)

Settling time* (fine range/full range): 3s

Minimum sample weight* (acc. to USP): 30mg

Minimum sample weight* (U=1 %, k=2): 3mg

Supply Include:

- Main Equipment

- Standard accessories

- Manual User

4 Wireless High Temperature and Pressure Data Logger system 2

Supply included 3 items below (4.1-4.3)

4.1 *Module 1: Interface Package 2

Code: IFC400

- Interface Cable Package for the HiTemp140 and PR140 Data Loggers.

USB Connection with Docking Station Cradle.

- Can to connect with many probes: Temperature, Pressure…

4.2 *Module 2: Temperature sensor 10

Code: 140-1

- The HiTemp140 is a rugged, high precision, temperature data logger that

is built for use with autoclaves and harsh environments. This stainless steel

device is submersible, can withstand temperatures up to 140 °C (284 °F)

and has an accuracy of +/-0.1 °C (0.18 °F).

- The HiTemp140 can store up to 32,700 readings, and features a rigid

external probe capable of measuring extended temperatures, up to 260°C

(500°F). Custom probe lengths up to 7 inches are available. The device

records date and time stamped readings, and has non-volatile solid state

memory that will retain data even if the battery becomes discharged. The

HiTemp140 can be used in both wet and dry applications up to 140°C

indefinitely

Specification:

- Temperature Sensor: 100 Ω platinum RTD

- Probe Range: –200 °C to +260 °C (-328 °F to +500 °F)

- Temperature Resolution: 0.01 °C

- Temperature Accuracy:

±0.1 ºC/±0.18 ºF (20 ºC to +140 ºC/68 ºF to +284 ºF)

±0.3 °C/±0.54 ºF (-20 °C to +19.99 °C/-4 °F to +67.98 °F)

±0.4 °C/±0.72 ºF (-40 °C to -20.01 °C/-40 °F to -4.02 °F)

- Memory: 32,700 readings

- Reading Rate: 1 reading every second up to 1 reading every 24 hours

- Baud Rate: 125,000 baud

- Typical Battery Life: 1 year (1 minute reading rate at 25°C)

- Operating Environment: –40°C to +140°C, 0%RH to 100%RH

- Material: 316 Stainless Steel

- Dimensions (Body: HiTemp140-1): 1.1 in x 0.125 in dia. (0.188 in

transitional diameter)

27 mm x 3.2 mm dia. (4.8 mm transitional diameter)

- Weight: 4.2 oz (120 g)

4.3 *Module 3: Pressure sensor 2

Code: PRTemp140

- The PRTemp140 is a pressure and high temperature data logger all in

one, created specifically for use in autoclave validation, mapping studies

and pressurized processes. Featuring the familiar body style of the 140

series of data loggers and made of rugged stainless steel, this dual purpose

device is less than 2.5 inches in length, allowing it to fit discretely inside

small spaces. The PRTemp140 is able to measure and record temperatures

from -20°C to 140°C, and pressure from 0 bar to 5 bar, accommodating a

wide range of applications even in the harshest conditions.

Specification:

- Pressure Sensor: Strain Gauge

- Pressure Range: 0 bar to 5 bar

- Pressure Resolution: 0.0001 bar

- Calibrated Accuracy: ±0.03 bar

-Temperature Coefficient: 0.03 bar accuracy applies from +20 °C to +140

°C

- Temperature Sensor: Resistance Temperature Detector (RTD)

- Temperature Range: -20 °C to +140 °C (-4 °F to +284 °F)

- Calibrated Accuracy: ±0.1 °C

- Memory: 32,767 readings

- Reading Rate: 1 reading every second up to 1 reading every 24 hours

- Battery Life: 2 years typical (1 minute reading rate)

- Operating System: Compatibility XP SP3/Vista/Windows 7/Windows 8

- Operating Environment: -20 °C to +140 °C (-4 °F to +284 °F),

0 %RH to 100 %RH, 0.002 PSIA to 100 PSIA

- Material: 316 Stainless Steel

- Dimensions (Body): Flush Top: 1.98 in x 0.97 in x 0.97 in (50.2 mm x

24.6 mm x 24.6 mm) NPT Top: 2.3 in x 0.97 in x 0.97 in (58.2 mm x 24.6

mm x 24.6 mm)

- Weight: Flush Top: 90 g (3.2 oz); PT Top: 100 g (3.5 oz)

5 High Temperature Data Logger system

Supply included 5 items below (5.1-5.5) 2

5.1 *Module 1: Interface Package 2

Code: IFC200

- MadgeTech’s IFC200 interface cable communicates between stand alone

data loggers and the MadgeTech software. Through this interface, loggers

can be started, stopped or downloaded.

- The IFC200 has been redesigned, and is now “plug and play”. It can be

directly connected to a computer without installing any drivers and is

compatible with Windows XP SP3, Windows Vista (32 and 64 bit), and

Windows 7 (32 and 64 bit) operating systems. The improved IFC200 can

now operate at up to 500 Volts RMS relative to the computer’s earth

ground when attached.

- The IFC200 is equipped with communication LEDs, that provide a quick

indication of the device status. The blue light illuminates when the device

is plugged in and successfully recognized by Windows. The red light

flashes when data is sent, and the green light flashes when data is received.

5.2 *Module 2: OctRTD Data Logger 8 channel 2

Code: OctRTD

- The OctRTD is an 8 channel, battery powered, stand alone, RTD based

temperature data logger. This is an all-in-one compact, portable, easy to

use device that will measure and record data for up to 10,922

measurements per channel. The storage medium is non-volatile solid state

memory, providing maximum data security even if the battery becomes

discharged. The device can be started and stopped directly from your

computer and its small size allows it to fit almost anywhere.

Specification:

- Temperature Sensor: External: 2,3, or 4 wire 100 Ω Platinum RTD

- Temperature Range: -200 °C to +850 °C

- Temperature Resolution: 0.01 °C

- Temperature Calibrated Accuracy: ±0.1 °C

- Memory: 10,922 readings per channel

- Reading Rate: 1 reading every 2 seconds up to 1 reading every 12 hours

- Baud Rate: 2,400

- Channels: 8

- Battery Type: 9V Lithium or alkaline battery included, user replaceable

- Battery Life: 1 year

- Material: Anodized aluminum

- Dimensions: 3.5 in x 4.4 in x 1.5 in (89 mm x 112 mm x 39 mm)

- Weight: 450 g

5.3 *Module 3: Wiring the Data Logger 16

- Teflon cover, senser 100Ohm Class B DIN Platinum RTD

- Temperature Range: -50 °C to 200 °C

- Temperature Accuracy: ±0.15 °C

- Length: 180cm

5.4 *Module 4: RTD Connecter (for RTD 3 or 4 wire) 16

5.5 '*Module 5: Computer for data processing 2

- CPU: Intel Core™ i3 4005U (2* 1.7G/ 3MB Cache)

- Chipset Intel HM87

- RAM: 4GB 1600MHz DDR3

- VGA Intel HD 4400

- HDD 500GB SATA 5400rpm

- DVD-RW

- Display: 14” HP WLED

- Microsoft Windows 8

- Microsoft office 2013

6 Biological Indicator Monitoring system 1

Supply included 3 items below (6.1-6.3):

6.1 Smart-Well Incubator 1

Features:

Dual Integrated Digital Thermometers

Improved Processor and Storage

Larger, Color LCD Interface

Password Protection

Temperature Progress Mete

Specifications:

- Overall Dimensions (W x H x D): 215 x 61 x 175 mm

- Incubation Cavities:

+ Quantity: 11

+ Diameter: 1.1 cm

+ Depth: 4 cm

- Thermometer Cavity:

+ Quantity: 1

+ Diameter: 0.386 cm

+ Depth: 2.121 cm

- Integrated Thermometer Accuracy: +/- 1°C

- Electrical Ratings (for use with a certified class 2 power supply):

12VDC, 30W, 2500mAh

- Operating Temperature Range: 60 ± 2°C

- Environmental Conditions:

+ Ambient Temperature: 18°C to 37°C

+ Relative Humidity: 0% - 80% Noncondensing

- Conformance: CSA/C/US, CE, RoHS, FCC Class A, ICES-003 Class A

6.2 High speed, LCD control POS printing 1

6.3 Smart-Read EZTest 1

- Spore: stearothermophilus

- Incubation: 62oC

- Population: 106

- Units: 100

(or equivalent)

7 CO2 Meter 1

Features:

- NDIR method principal for CO2 ( Carbon dioxide ) measurement,

available for long term operation.

- High repeatability and high accuracy.

- Separate probe, easy operation and convenient for remote measurement.

- CO2 function with alarm setting.

- Large S-TN LCD, high contrast, easy readout.

- Data hold function for freezing the desired value on display.

- Records Maximum and Minimum readings with Recall.

- RS232/USB computer interface.

- Microprocessor circuit assures maximum possible accuracy, provides

special functions and features.

- Heavy duty & compact housing with hard carrying case, designed for

easy carry out & operation.

- Auto shut off is available to save battery life.

- Power supply from batteries or DC 9V adapter in.

General Specifications:

- Circuit: Custom one-chip of microprocessor LSI circuit.

- Display: LCD size: 52 mm x 38 mm dual function LCD display.

- Measurement: CO2 (Carbon dioxide), Temp.

- Unit: ppm

- Response Time: < 2 min. typically.

@ Reach the 63% reading value

@ Depend the environment air circulation.

- CO2 altitude compensation setting: 0 to 9,000 meters.

- Temperature Compensation: Automatic temp. Compensation.

- Advanced setting: CO2 altitude value setting; CO2 alarm value setting; oC/oF setting; Auto power off enable/disable setting

- Alarm setting: For CO2 measurement only.

- Data Hold: Freeze the display reading:

- Memory Recall: Maximum & Minimum value.

- Display Sampling Time: Approx. 1 second.

- Power off: Auto shut off saves battery life or manual off by push button.

- Operating Temperature: 0 to 50oC

- Operating Humidity: Main instrument : Less than 85% R.H.

C02 probe : Less than 85% R.H.

- Power Supply: DC 1.5 V battery (UM3, AA ) x 6 PCs, or equivalent.

- Power Current: Approx. DC 9.6 mA for 90% period.

Approx. DC 128 mA for 10% period.

- Weight: Main instrument: 372 g/0.82 LB.

@ Battery is included.

CO2 probe : 158 g/0.35 LB.

- Dimension: Main instrument : 173 x 68 x 42 mm

CO2 Probe : 185 x 38 x 26 mm

Electrical Specifications:

CO2 ( Carbon dioxide )

- Range: 0 to 4,000 ppm

- Resolution: 1 ppm

- Accuracy:

± 40 ppm at ≤ 1,000 ppm.

± 5% of reading at >1,000 ppm ≤ 3,000 ppm

± 250 ppm typically at >3,000 ppm, reference only

- Repeatability: ± 20 ppm at ≤ 3,000 ppm.

Temperature

- Range: 0oC to 50oC, 32oC to 122oC .

- Resolution: 0.1 degree

- Accuracy: oC: 0.8oC; oF: 1.5oF

Supply Include:

- Main Equipment

- CO2 Probe

- Hard Carrying case

- Instruction manual

(or equivalent)

8 AccuBalance capture hood 1

- Ergonomic design and ultra-light weight for easy,

one-person operation

- Automatically senses and displays supply or return

flows, saving time on the job

- Back pressure compensation ensures accurate readings

- Multiple hood sizes available for easy, cost effective

use across multiple jobs

- Detachable digital micromanometer offers flexibility to

use in multiple applicationse

Velocity Range:

Pitot probes: 25 to 15,500 ft/min (0.125 to 78 m/s)

Air flow probe: 25 to 5,000 ft/min (0.125 to 25 m/s)

Velocity matrix: 25 to 2,500 ft/min (0.125 to 12.5 m/s)

Accuracy: ±3% of reading ±7 ft/min (±0.04

m/s) at velocities >50 ft/min

(>0.25 m/s)

Units ft/min, m/s

Resolution: 1 ft/min (0.01 m/s

Pressure:

- Differential pressure: ±15 in. H2O (±3735 Pa);

150 in. H2O (37.5 kPa), maximum safe operating pressure

- Absolute pressure: 15 to 40 in. Hg (356 to 1016 mm Hg)

- Accuracy: ±2% of reading ±0.0001 in. H2O (±0.025 Pa) static and

differential; ±2% of reading absolute

- Units: in. H2O, in. Hg, Pa, hPa, kPa, mm Hg,

cm Hg, mm H2O, cm H2O,

- Resolution: 0.00001 in. H2O (0.001 Pa) static and differential; 0.01 in.

Hg (1 mm Hg) absolute

Volume:

- Range: 25 to 2,500 ft3/min (42 to 4250 m3/h) capture hood

- Accuracy: ±3% of reading ±7 ft3/min (±12 m3/h) at flows >50 ft3/min

(>85 m3/h)

- Units: ft3/min, m3/h, m3/min, l/s

- Resolution: 1 ft3/min (1 m3/h)

pH

- Range: 5 to 95% RH temperature/RH probe

- Accuracy: ±3% RH

- Resolution: 0.1% RH

Temperature

- Sensor in base: 40 to 140°F (4.4 to 60°C)

- Temperature/RH probe: 14 to 140°F (-10 to 60°C)

- Accuracy: ±0.5°F (±0.3°C)

- Units: °F, °C

- Resolution: 0.1°F (0.1°C)

Instrument Temperature Range:

Operating 40 to 140°F (4.4 to 60°C)

Storage -4 to 160°F (-20 to 71°C)

Statistics : min, max, average

Data Storage: 26,500 samples, time and date stamped

Logging Interval:User selectable

Response Time:2 to 8 seconds, differential pressure

sensor

Dimensions (manometer only):

7.4 in. x 4.5 in. x 2.3 in. (18.8 cm x 11.4 cm x 5.8 cm)

Pressure Connection:1/4 in. (6.35 mm) OD straight ports with barbed ends

for use with 3/16 in. (4.76 mm) ID flexible tubing

Weight: 3.4kg

Power requirements: Four AA-size cell or AC adapter

Supply with :

2 ft x 2 ft (610 mm x 610 mm) air capture hood/frame/base, Swirl X Flow

Conditioner manometer, 4 AA size rechargeable NiMH batteries, multi-

country AC adapter, 18 in. (46cm) Pitot probe, 2 Static Pressure probes,

16 ft (4.8m) Neoprene tubing, wheeled luggage-style carrying case, NIST-

traceable calibration certification, downloading software, USB interface

cable, and manual.

9 Temperature & Pressure Data loggers of Datatrace 1

- Controlling a manufacturing process requires accurate monitoring. And

for that, your sensor must be close to the product being monitored,

especially in food processing where one container is inside another. The

ultra-small DataTrace® Micropack III (MPIII) Pressure Data Logger has a

body measuring just 1.9 inches long.

- MPIII Pressure Data Loggers fit inside containers to get pressure

readings from 0 PSIA (0 BAR) to 72.5 PSIA (5 BAR) with exceptional

accuracy at temperatures up to 140°C (284°F). With their wide range and

high precision, MPIII Pressure Data Loggers are not only well suited to

canning and pouch processes, but an array of applications where pressure

affects products and quality

- Wireless Pressure Data Logger

- Compact (1.9″ x 0.7″) Body

- 16,000 NIST-traceable Data Points

- FDA 21 CFR Part 11 Compliant

- Intrinsically Safe

Range

Operating Temperature: +25°C to +140°C (+77°F to +284°F)

Temperature Accuracy: ± 0.25°C (± 0.45°F)

Resolution: 0.001 PSI

Operating Pressure: 0 PSIA (0 BAR) to 72.5 PSIA (5 BAR)

Pressure Accuracy: ± 0.45 PSIA (± 0.03 BAR)

Physical

Size:

Width: 0.7 in (18mm)

Length: 1.9 in (48mm)

Weight: 1.9 oz (55 grams)

Material: 316 stainless steel; Sapphire IR windows

Mechanical

Media: Compatible with air and fluids, including many corrosive liquids

Calibration: Factory calibrated (NIST traceable)

Battery Type: Lithium (field-replaceable)

Clock accuracy: 26 seconds per 24 hours

Certifications: ATEX (for intrinsic safety) - under review - to be certified

soon

Monitoring Parameters

Memory: 8,000 time/temperature observations and 8,000 time/pressure

observations

Sample rates: Operator programmable 1 sec. to 1 hr. intervals

Sensor Elements

Temperature: Thermistor

Pressure: Strain gauge bridge

10 Microplate Calibration Standards 96 well of Fireflysci 1

- FireflySci is currently offering a full range of 96 well test plate reference

standards including adapters that enable our UV/VIS calibration standards

to be used. Test plates cover absorbance, fluorescence and luminescence

- All benchmark calibration standards are certified in the ISO-9001, ISO-

17025 and ISO Guide 34-accredited NRC Canada Government Metrology

Laboratory.

Capabilities:

Enables the user to use any of our cuvette-shaped calibration standards in a

96 well plate form

You can use the following standards in this adapter: absorbance

standards, wavelength standards, stray light standards, spectral resolution

standards.

Product Specs:

Made with anodized aluminium for improved performance and resilience.

Total dimensions for 96 well plate: 85x125x12.5mm

Patent-pending design

11 Handheld Air Particle Counters (6 Channel) 1

- Particle Size Range: 0.3um - 10um in size channels per ISO 14644-1 (FS

209E) industry standard

- Number of size channels: 6

- Weight: Ultra-lightweight at 1.5 lbs for comfortable handheld use

- Flow rate: 2.83 lpm. Take ISO 14644-1 class 5 compliant 0.3um

samples in a minute

- USB support: Simple data export to memory stick or direct connect to

PC via USB cable. No software required.

- Ethernet support: Access particle data over network with common

browsers such as Windows Explorer, Safari and Firefox

- Display: Large 3.5" High resolution 320 x 240 color for super clear

viewing with intuitive icons for easy navigation

- Display configuration: User selectable size channels with super-large

font options

- Display data modes: Traditional tabular particle counts or trend graph for

easy particle profiling

- Enclosure: High impact polycarbonate ABS for exceptional durability

- Dimensions: 10.7" x 3.9" x 2.1". Contoured for convenient single

handed operation

- Battery duration: Over 10 hours with typical use model. Minimum 5.5

hours continuous sampling

- Data storage: 10,000 records for the most demanding applications

- Counting Efficiency: 50% @ 0.3 µm; 100% for particles > 0.45 µm (per

ISO 21501)

- Zero Count Level: 1 count / 5 minute (per JIS B9921)

- Concentration Limits: 10% at 4,000,000 per cubic foot (per ISO21501)

- Count Modes: Raw counts, N/CF, N/CM, N/L in Cumulative or

Differential mode

- Security: Administrator password controlled (optional)

- UI Languages: English and Japanese

- Alarms: User defined: Counts

- Battery charging: 3.5 hours

- Operating environment: 50°F to 104°F (10°C to 40°C) / < 95% non-

condensing

- Storage environment: 14°F to 122°F (-10°C to 50°C) / Up to 98% non-

condensing

12 High precision weights OIML class

Option 1: E1 Class 1mg-200g 1

- OIML class E1

- OIML Weights - Premium Line

- Weight Sets With Wooden Box

- Nominal Value: 1mg - 200g (Incl. 23 units/set)

II. Terms and conditions:

1. Price:

The price is inclusive of all fees and labor related to shipping, handling, insurance, customs

clearance, and distribution to the sites outlined below

2. Capability of vendor/supplies

(a) Having at least 2 staff trained on biosafety and biosecurity

(b) Obtaining ISO 9001-2008

(c) Having an Authorized Letter of Manufacture for goods.

3. Delivery Terms

(a) Delivery Date is to be understood as the time the Goods as listed above have to be available at

the delivery locations listed in section 6.

(b) The Supplier will make 1 (one) distribution within 4-6 weeks of Purchase Order signing to

CDC’s recipient sites (“Recipients”) listed in section 6. Both parties will confirm in writing the

quantity of goods, time of delivery, and Recipient’s name and address before the delivery.

4. Quality Assurance

(a) Supplier commits that all Goods supplied under this Purchase Order shall be in compliance with

all appropriate national regulatory authorities, requirements of the Government of Vietnam and

with relevant certification standards for the production of Goods.

(b) Supplier guarantees that the Goods are 100% new and are in accordance with the

manufacturer’s highest standards.

(c) Supplier will make reasonable accommodations for Purchaser to inspect Goods before delivery.

Purchaser also has the right to inspect Goods after delivery. Upon inspection of the Goods,

USCDC shall promptly notify the Supplier of any damages to the Goods and Supplier shall

work with USCDC to remedy these damages.

(d) Purchaser may conduct quality surveys on randomly selected items. If surveys detect testing

result discrepancies not in line with the manufacturer’s specifications, Supplier agrees to

replace the damaged or defective Goods.

5. Services

(a) Supplier agrees to provide a one-year warranty for all equipment at each site.

(b) Supplier shall conduct installation and commissioning of the Goods and the Initial Training in

for each site. The costs for installation, commissioning and the initial training are included in

the price of goods.

(c) Supplier will undertake to supply spare part, chemical for running instrument in 5 years.

6. List of designated destinations: Door to door delivery as follows:

No Name of recipient Address Person of contact Telephone number

1 Biosafety and Quality

Management Systems

Department at National

Institute of Hygiene and

Epidemiology

No. 1 Yersin Street.,

Hanoi, Vietnam

2 Biosafety Department of

Nha Trang Pasteur

Institute

No. 8 Tran Phu Street.,

Nha Trang city, Khanh

Hoa Province, Vietnam

3 Pasteur Institute in Ho

Chi Minh City

No. 167 Pasteur Street.,

District 3, Ho Chi Minh

City, Vietnam

Prefer: Vendor obtained ISO 9001-2008: The Quality Management System is applicable to consulting,

providing, high technology equipment and training documents on scientific research and production.

Distribute equipment, chemicals in the field of medical, pharmaceutical and life science. Import,

storage and distribute vaccines-biomedical reagents. Research and develop applied of biotechnology


Recommended