+ All Categories
Home > Documents > Section 2 System Components

Section 2 System Components

Date post: 02-Jun-2018
Category:
Upload: norman-morales
View: 228 times
Download: 0 times
Share this document with a friend

of 28

Transcript
  • 8/10/2019 Section 2 System Components

    1/28

    System Components

    Section Objectives

    At the conclusion of this topic, the student will be able to:

    1. Describe the function and physical characteristics of the 208 DAIU.

    2. Identify computer requirements and 208 DAIU connecting options.

    3. Perform basic tasks using ADRE for Windows software including:

    Set the computer clock

    Select a printer

    Set ADRE preferences

    Save ADRE databases

    Identify database caution screens

    Use the on-line help and glossary

  • 8/10/2019 Section 2 System Components

    2/28

    System Components Page 1

    REV C 134515

    I. The 208 DAIU

    A) Function Description

    The 208 Data Acquisition Interface Unit (DAIU) is a multi-channel

    processing unit that interfaces directly to a computer. Two DAIUs can be

    connected together for up to 16 channels of simultaneous data acquisition.The computer controls data acquisition. As data is captured by the 208

    DAIU, it is downloaded directly into the computers RAM (random access

    memory).

    Figure 1. The 208 DAIU Front Panel

    B) Rear Panel Connections

    Figure 2. The 208 DAIU Rear Panel

    BatteryCompartment

    Power Switch

    Signal Inputs

    KeyphasorInputs

    Keyphasor Outputs

    Communication Ports

    Power Supply

    InputRelay Input

  • 8/10/2019 Section 2 System Components

    3/28

    Page 2 ADRE for Windows

    REV C 134515

    Input Signals: Each 208 DAIU box accepts up to 8 channels of

    displacement, velocity, and acceleration inputs. The 208 DAIU also

    accepts process variable inputs, and input signals from most generic

    dynamic transducers.

    Keyphasor Inputs/Outputs: Up to 2 keyphasor! inputs can be assigned to

    pairs of the signal input channels, allowing data acquisition on

    machine trains with more than one shaft speed. See the discussion on

    Keyphasor Assignments in Section 3 of this course for rules on

    connecting Keyphasor signal cables to the 208 DAIUs.

    Relay Inputs: Data acquisition can be initiated via remote relay by using the

    Relay Input. (See Chapter 11 Advanced Sampling, page 15)

    Communication Ports: The 208 DAIU is connected to the computer and to

    an additional 208 DAIU through the Computer/DAIU Input and

    DAIU Output ports.

    Power Supply Input: Input from DC adapter is used to charge the internal

    battery and to run on AC power.

    C) The 208 DAIU Internal Battery

    The internal nickel-cadmium battery will last up to 4 hours on a singlecharge, and can typically handle up to 500 full charge cycles with less than

    15% capacity loss. A full drain of the battery is recommended every 1-2

    months to optimize battery performance. Charge times vary depending on

    the battery type:

    Battery Part Number Full Recharge Time

    Standard Part No. 107305-01 Up to 5 hours

    High Temperature Part No. 107306-01 Up to 15 hours

    Power Supply Charger Upgrade: A hardware upgrade is available which

    improves the performance of the internal battery charger. The

    upgrade is accomplished by replacing PWA 102387-01 with PWA

    131866-01.

  • 8/10/2019 Section 2 System Components

    4/28

    System Components Page 3

    REV C 134515

    II. Computers

    To operate ADRE!for Windows version 4.70, the target computer must have at

    least the following requirements:

    486/Pentium!processor

    Windows98, NT 4.0, or 2000

    16 megabytes (MB) RAM

    80 MB hard disk space

    Up to 28.2 MB additional hard disk space for each database stored (if no

    external medium)

    CD-ROM drive for installation

    VGA display

    Mouse

    Enhanced Parallel Port (EPP), available ISA slot, or available PCMCIA

    slot

    Several connecting options exist for operating ADRE! for Windows on your

    computer. These options depend on whether you are installing ADRE on a laptop

    or a desktop computer as well as the operating system that you are using.

  • 8/10/2019 Section 2 System Components

    5/28

    Page 4 ADRE for Windows

    REV C 134515

    A) Laptop Computers

    Connecting options for laptop computers depends upon the operating

    system used.

    Figure 3. Laptop Computer Connected to ADRE 208-P DAIU

    Windows NT 4.0/2000

    To use ADRE!for Windows with a laptop computer and Windows

    NT 4.0/2000, you must have:

    1. A 208-P DAIU

    2. A 25-80 pin EPP to DAIU connector cable

    3. An Enhanced Parallel Port (EPP) that can be set to EPP mode by

    the user or an EPP PCMCIA card

    Windows 98

    To use ADRE!for Windows with a laptop computer and Windows

    98, you must have:

    1. A 208-P DAIU

    2. A 25-80 pin EPP to DAIU connector cable3. An Enhanced Parallel Port (EPP) with EPP BIOS support, an

    Enhanced Parallel Port (EPP) that can be set to EPP mode by the

    user, or an EPP PCMCIA card

  • 8/10/2019 Section 2 System Components

    6/28

    System Components Page 5

    REV C 134515

    Laptop computers must use the 208-P DAIU instead of the 208 DAIU. This

    is because the 208 DAIU can only connect to computers using an ISA

    port/card which laptop computers do not have.

    If your laptop printer port is an Enhanced Parallel Port (EPP), you canconnect the 208-P DAIU directly to it using the 25-80 pin connector cable.

    If your laptop printer port is not an EPP, you will need to create one by

    obtaining an EPP PCMCIA card, such as the Quatech SPP-100 PCMCIA

    card, and connect the 208-P DAIU to it using the 25-80 pin connector cable.

    Figure 4. 25-80 pin Connector Cable

    B) Desktop Computers

    Connecting options for desktop computers depends upon the operatingsystem used.

    Windows 2000

    To use ADRE!for Windows with a desktop computer running

    Windows 2000 you must have:

    1. A 208-P DAIU

    2. A 25-80 pin EPP to DAIU connector cable

    3. An Enhanced Parallel Port (EPP) that can be set to EPP mode by

    the user

  • 8/10/2019 Section 2 System Components

    7/28

    Page 6 ADRE for Windows

    REV C 134515

    Windows NT 4.0

    To use ADRE!for Windows with a desktop computer running

    Windows NT 4.0 you must have:

    1. A 208-P DAIU2. A 25-80 pin EPP to DAIU connector cable

    3. An Enhanced Parallel Port (EPP) that can be set to EPP mode

    by the user or an ISA EPP card, such as the Quatech MMP-100,

    with an unused ISA slot and address.

    If your desktop printer port is an EPP, you can connect the 208-P

    DAIU directly to it using the 25-80 pin connector cable. If your

    desktop printer port is not an EPP, you will need to create one by

    obtaining an EPP ISA card and connect the 208-P DAIU to it using

    the 25-80 pin connector cable.

    Figure 5. 208-P DAIU Connected to a Desktop EPP

    Windows 98

    To use ADRE!for Windows with a desktop computer running

    Windows 98 you must have:

    1. A 208-P DAIU

    2. A 25-80 pin EPP to DAIU connector cable

    3. An Enhanced Parallel Port (EPP) with EPP BIOS support or an

    EPP port that can be set to EPP mode by the user

  • 8/10/2019 Section 2 System Components

    8/28

    System Components Page 7

    REV C 134515

    In addition to the 208-P DAIU, desktops running Windows 98 can

    also use the 208 DAIU. This is the only platform that supports the

    208 DAIU.

    To use ADRE!for Windows with a 208 DAIU, you must have:

    1. A 208 DAIU

    2. A 68-80 pin ISA to DAIU connector cable

    3. An unused ISA slot

    4. An empty ISA address space

    5. A special half length ISA interface card provided by Bently

    Nevada.

    Figure 6. Desktop Computer, ISA Communications Card, and 68-80 Pin ISA to

    DAIU Connector Cable

    Note: While collecting data through your EPP-compatible port, you cannot

    simultaneously use the port for other devices. You cannot have a printer, tape

    drive, or other device connected to the LPT port. Even having devices assigned to

    this port may cause problems with data collection. These devices will interrupt

    signals to and from the DAIU(s) unpredictably.

    Note: You cannot connect one 208-P DAIU and one 208 DAIU together. If you

    are using two DAIUs, both must be the same type.

  • 8/10/2019 Section 2 System Components

    9/28

    Page 8 ADRE for Windows

    REV C 134515

    III. ADRE!!!!for Windows Software

    A) Automated Diagnostics for Rotating Equipment (ADRE)

    Figure 7. ADRE!for Windows compact disc

    ADRE!for Windows software is the data acquisition controller. ADRE!

    for Windows directs sampling performed by the 208 DAIU, and receives

    digitized data in computer memory in return.

    Analog Signal

    208 DAIU

    A/D Conversion

    Digital Information

    Computer

    RAM

    Figure 8.The Data Path

    All data, both during and after sampling, remains in computer memory

    (RAM) until saved to the hard drive or other storage medium. Be sure to

    SAVE YOUR DATA! If you have Repeating Data Collection selected in

    the user preference (See Setting ADRE!for Windows Preferences later

    in this section), then you will have a copy of your data on the hard disk. It

    is still recommended for you to save your data with your own naming

    convention. The Database created from the Repeating Data Collection

    can then be a reliable backup copy of your raw data.

  • 8/10/2019 Section 2 System Components

    10/28

    System Components Page 9

    REV C 134515

    B) Running ADRE!!!!in a Windows Environment

    1) Setting the Computer Clock

    During sampling, ADRE!for Windows uses the computers internal

    clock to record the date and time of each sample record. Though thedate and time tags for a database can be adjusted after sampling

    through the ADRE! for Windows utility menu, insuring that the

    computer clock and calendar are set correctly prior to sampling is

    good practice.

    Figure 9. Setting the Clock and Calendar

  • 8/10/2019 Section 2 System Components

    11/28

    Page 10 ADRE for Windows

    REV C 134515

    For setting the clock on Windows 98, see Figures 9 or 10.

    Figure 9, Click: Figure 10, Click:

    1. Double Click on My Computer Icon 1. Start

    2. Double Click on Control Panel Icon 2. Settings3. Double Click on Date/Time Icon 3. Control Panel

    4. Change Date if necessary 4. Double click on Date/Time Icon

    5. Change Time if necessary 5. Change Date if necessary

    6. Click Apply 6. Change Time if necessary

    7. Click OK 7. Click Apply

    8. Click OK

    Figure 10. Setting the Clock and Calendar

  • 8/10/2019 Section 2 System Components

    12/28

    System Components Page 11

    REV C 134515

    2) Selecting a Printer Driver

    ADRE! for Windows automatically uses the default Windows

    printer driver each time the program is executed. To select a default

    Windows printer driver from Windows 98, double-click on the "My

    Computer" icon on the Desktop. Next, select Control Panel, thenPrinters. Select an installed printer by clicking on File, then Set as

    Default.

    Figure 11. Setting the Default Windows Printer

  • 8/10/2019 Section 2 System Components

    13/28

    Page 12 ADRE for Windows

    REV C 134515

    You can select any installed Windows printer from within ADRE! for

    Windows. This selection, however, does not change the default Windows

    printer. The next time ADRE!for Windows is run or a different database

    is opened (either new or existing), the default Windows printer will be

    selected again.

    To select a printer from within ADRE! for Windows, click on the File

    pull down menu, select Print, then click on the Printer Setup button for

    options.

    Figure 12. Selecting a Printer from ADRE for Windows

  • 8/10/2019 Section 2 System Components

    14/28

    System Components Page 13

    REV C 134515

    C) Setting ADRE!!!!for Windows Preferences

    Figure 13. Setting Preferences

    Database Auto Save

    Since data is stored generally in computer memory after sampling

    until manually saved to a storage medium by the user, enabling

    autosave options can provide temporary protection against the

    accidental loss of data. When a Database Auto Save option is

    selected, or checked, the current database is saved to the

    C:\ADREWIN\DATABASE\DEFAULT

    subdirectory during execution of the specified function:

    On Exit

    Upon exit of ADRE! for Windows software, the current

    database is automatically saved to the default subdirectory. No

    prompt is given.

  • 8/10/2019 Section 2 System Components

    15/28

    Page 14 ADRE for Windows

    REV C 134515

    Retrieve Database

    When a database is retrieved and loaded into ADRE! for

    Windows, the database is automatically saved to the default

    subdirectory.

    Leave Store EnableUpon the close of the Store Enable function, the newly

    collected database is saved to the default subdirectory.

    Realtime Backup

    When Realtime Backup is selected, or checked, the current database

    is saved to the backup subdirectory C:\ADREWIN\BACKUP during

    execution of data collection. If both Minutes and Samples numbers

    are non-zero, ADRE! for Windows will backup the data whenever

    the time interval expires or the sample count reaches the limit.

    Minutes

    The time interval in minutes between two consecutive backups.

    NOTE: ADRE downloads samples from the DAIU in groups of 10 vectors. The

    backup database will include only samples that have been downloaded from the

    DAIU to the computer. For example, if only 7 vector samples have been taken

    by the DAIU when the backup database is saved, there will be no samples in the

    backup database (because those samples have not yet been downloaded to the

    computer). If 15 vector samples have been taken when the backup database issaved, there will be 10 samples in the backup database.

    Samples

    The number of samples collected between two consecutive

    backups.

    Data Acquisition

    Repeat Data Collection

    When Repeating Data Collection is selected, ADRE repeatsdata collection and saves each database until the amount of free

    disk space on the disk drive is less than 28.2 MB. Data

    collection will stop and the last data collection will need to be

    saved by using the option: FILE - SAVE.

  • 8/10/2019 Section 2 System Components

    16/28

    System Components Page 15

    REV C 134515

    Each database collected while in repeat mode will be saved as:

    \ADREWIN\DATABASE\REPEAT\DDMMMYY\HOUR_MIN_SEC

    The DDMMMYY and HOUR_MIN_SEC directories will bethe date and time of the first sample in the database with a valid

    time tag. If the first sample in the database was collected with

    a time tag of 01 May 2000 12:01:01, the database would be

    saved as:

    \ADREWIN\DATABASE\REPEAT\01MAY00\12_01_01.

    If there are no samples in the database with a valid time tag, the

    database will always be saved as:

    \ADREWIN\DATABASE\REPEAT\01JAN80\00_00_00.

    Comments will be generated containing the following

    information for each database:

    Session started with Database: dd mmm yy - hr:mn:sc

    (Date and time of the first database)

    Repeating Data Collection must be selected in order to activatethe Useable Drive and Search options. The Path will also be

    dimmed unless Repeating Data Collection is checked.

    When this option is selected, a System Error message may be

    displayed if Windows cant find a disk in a Useable Drive.

    Selecting the Cancel button will remove the message from your

    screen and allow you to continue.

    Reset Trigger

    When you select Repeating Data Collection and ResetTrigger, ADRE will collect data until the database is full, save

    it, and then automatically prepare to collect data. If the

    machine is operating in the triggered region, data collection

    will trigger and ADRE will again collect data and save the

    database when it is full. This will occur repeatedly until the

    amount of free disk space on the disk drive is less than 28.2

    MB. If the machine is not operating in the triggered region, no

  • 8/10/2019 Section 2 System Components

    17/28

    Page 16 ADRE for Windows

    REV C 134515

    more data will be collected after the first database fills up and

    is saved. However, the system will again collect data if the

    machine conditions change to meet the trigger requirements. If

    free disk space runs low (below 28.2 MB), data collection will

    stop and the last database will need to be saved by using theoption: FILE - SAVE.

    If you select Repeating Data Collection, but do not select

    Reset Trigger, after the first database is saved, the system

    will automatically begin collecting data again. The system will

    collect and save data repeatedly (no matter what the machine

    conditions are) until you manually stop data collection or the

    hard drive contains less than 28.2 MB of free space.

    If you select Reset Trigger and have also configured your

    system to Trigger on Time, you will get a warning message that

    data collection will not re-trigger. This is because the time

    trigger condition can only be true once.

    Useable Drive

    The useable drive is the drive to which you will save each of

    your databases when Repeating Data Collection has been

    checked. In order for a drive to be useable, it must have morethan 28.2 MB of free disk space and it must be a writeable disk

    drive. Drives A and B are excluded from the list of useable

    drives since they cannot normally be used to store 28.2 or more

    megabytes of information. Click on the Down Arrow button to

    see a list of available drives.

  • 8/10/2019 Section 2 System Components

    18/28

    System Components Page 17

    REV C 134515

    Figure 14. Useable Drives

    Search

    Click on the search key to find all the Useable Drives (from D -

    Z) on your system. Note that only fixed and removable drives

    can be used for data storage. If a useable drive becomes

    invalid or does not show up on the list when it normally should,

    check the drive to make sure it has sufficient free space and

    that it is not write protected.

    Path

    The path or location where the databases will be stored on the

    useable drive. Each database will be stored in the

    \ADREWIN\DATABASE\REPEAT\DDMMMYY\

    subdirectory with the filename HOUR_MIN_SEC, where

    DDMMMYY is the date when the database was collected and

    HOUR_MIN_SEC is the time of the first vector sample with a

    valid time tag. For example if the first sample in the database

    was collected with a time tag of 01 May 2000 12:01:01, the

    database would be saved as:

    \ADREWIN\DATABASE\REPEAT\01MAY00\12_01_01.

    RPM Values

    ADRE does not always sample at the exact delta RPM value.

    Because it may be a couple of RPMs off, it normally rounds the actual

  • 8/10/2019 Section 2 System Components

    19/28

    Page 18 ADRE for Windows

    REV C 134515

    RPM value to the nearest delta RPM value (up to a maximum value

    of 10) to simplify plot presentation and add clarity. By default, the

    Disable RPM Rounding checkbox is left unselected to display the

    rounded RPM values in each plot displayed. If you want to display

    the exact RPM value that each sample was collected at, select theDisable RPM Rounding checkbox. The two figures below display

    the same samples (collected with delta RPM setting of 8) and the

    difference that this preference setting makes.

    Figure 15.Samples displayed with RPM Rounding Enabled

    Figure 16. Samples displayed with Disable RPM Rounding selected

  • 8/10/2019 Section 2 System Components

    20/28

    System Components Page 19

    REV C 134515

    Note: If you have a plot open and you change the RPM Values

    setting, you must close and reopen the plot to see the change.

    208 DAIU Settling

    The starting store enable and the sampling mode feature in storeenable requires the 208 DAIU to go through a settling time. If a

    bandwidth of 120 is selected, a long settling time is not required and

    no option will be given to select a 208 DAIU settling time. However,

    if a bandwidth of 12 is selected, it is recommended to check the 12

    RPM (7 seconds) box, allowing for a 7 second delay before the data is

    downloaded. If this is not done, the data collected immediately after

    a sample mode change will show the 1X, 2X, and nX charging up or

    settling.

    Note: Older versions of ADRE! for Windows (before 4.60) have a FileManager option within the Preferences dialog box. (The File Manager

    feature will be discussed further in the next subsection) The default File

    Manager for older versions is winfile.exe. To change the File Manager

    preference, type in the desired File Manager path and name (i.e.,

    c:\windows\explorer.exe). ADRE version 4.60 uses Windows Explorer

    when the File Manager feature is selected and does not offer a choice in the

    Preferences dialog box.

    D) File Manager

    ADRE allows you to access Windows Explorer without leaving the programto move, rename, and delete files. To run the File Manager from within

    ADRE, click File from the main menu bar, then select File Manager.

  • 8/10/2019 Section 2 System Components

    21/28

    Page 20 ADRE for Windows

    REV C 134515

    Figure 17. Running File Manager from ADRE for Windows

    E) Saving ADRE!!!!for Windows Databases

    ADRE! for Windows databases are stored in directories rather than

    individual files. When saving a database, the directory name is selected ortyped in. The default path is C:\ADREWIN\DATABASE\directory

    name. The File pull down menu option has two ways to save a database:

    Save and Data Range Save:

  • 8/10/2019 Section 2 System Components

    22/28

    System Components Page 21

    REV C 134515

    Figure 18. Using Save

    Save

    - Current directory name listed in entry field

    - Vector and gap reference data- Comments entry field (comment.txt)

    - Empty database (configuration only)

  • 8/10/2019 Section 2 System Components

    23/28

    Page 22 ADRE for Windows

    REV C 134515

    Data Range Save

    - Directory name entry field blanked out

    - Data range limit adjustment

    - Required disk space

    - Comments entry field (comment.txt)

    Database Comments

    To save notes about your sample run, type the notes into theComments box. These notes will be saved as part of your database in

    the comment.txt file, which is located in the database subdirectory

    Notes.

    Saving to Multiple Floppy Disks

    Databases requiring more disk space than available on a single floppy

    disk can be saved to multiple floppy disks. When the first disk is full,

    the program will prompt for an additional disk:

    Figure 20. Saving To Multiple Floppy Disks

    Figure 19. Using Data Range Save

  • 8/10/2019 Section 2 System Components

    24/28

    System Components Page 23

    REV C 134515

    The Default Subdirectory

    Figure 21. The Default Subdirectory

    Each time ADRE! for Windows is run it automatically loads the

    database currently residing in the

    C:\ADREWIN\DATABASE\DEFAULT\

    directory. The Database Autosave options, when selected, use this

    directory to store databases. Each time a database is stored to the

    C:\ADREWIN\DATABASE\DEFAULT\directory, the previous one is overwritten.

    Note: This directory should never be deleted as this will cause

    ADRE!for Windows not to start up and you would get the following

    message:

    Figure 22. Loading ADRE!for Windows Error Message

  • 8/10/2019 Section 2 System Components

    25/28

    Page 24 ADRE for Windows

    REV C 134515

    F) Caution Screens

    The following operations are the most common cause of caution screens:

    Configuration ChangesChanges to the configuration of an existing database that affect how

    the DAIU collects data will invalidate the database. These changes

    result in a configuration error because the revised configuration no

    longer matches the existing data. A caution screen gives you the

    opportunity to protect the original database.

    Figure 23. Warning Screen Following Configuration Change

    Opening a Database or Entering Store Enable

    Either of these actions, when done prior to saving newly acquired

    data, require erasing the current database from computer memory andbringing in new data. A caution screen gives you the opportunity to

    protect the original database.

    Figure 24. Caution Screen When Going to Store Enable or Opening a Database

  • 8/10/2019 Section 2 System Components

    26/28

    System Components Page 25

    REV C 134515

    Entering Store Enable in some situations would delete the vector waveform

    data in memory and in the Backup Directory. A caution screen gives you the

    opportunity to protect the original database.

    Figure 25. Caution Screen Going to "Store Enable"

    Overwriting a Previously Stored Database

    Figure 26. Caution Screen When Overwriting a Database

    After a database has been stored to disk, it is often desirable to make

    changes such as adding plot configurations, adjusting the limits of the

    database range, and changing the run type. To save these changes,

    you can either overwrite the original database (save to same name) or

    save to a database under a new name. If you choose to save to the

    same name, the caution screen above reminds you that you are about

    to overwrite the original database.

  • 8/10/2019 Section 2 System Components

    27/28

    Page 26 ADRE for Windows

    REV C 134515

    G) Using Help/Glossary

    ADRE! for Windows on-line Help system is context-sensitive and has

    flexible search capabilities. Help is available from every dialog box, taking

    you directly to the Help information you need.

    Figure 27. Accessing Help From The Sample Mode Configuration Dialog Box

  • 8/10/2019 Section 2 System Components

    28/28

    System Components Page 27

    The Glossary is available from the main tool bar pull-down Help menu. The

    glossary is in alphabetical order and contains an extensive list of definitions

    commonly used in the rotating machinery field.

    Figure 28.

    Using The Glossary


Recommended