Instruction IN/RandC/007–EN SM500F Paperless recorder Battery replacement Measurement made easy 1 Introduction This publication details replacement of the battery (B12454) fitted to the processor board of the SM500F Paperless recorder. The procedure must be carried out by a trained technician. 2 Safety 3 Tools required — T8 Torx driver Important note dispose of the old components as detailed in the Instruction manual (IM/SM500F).
Transcript
Page 1: SM500F Paperless recorder Battery replacement IN/RandC/007–EN SM500F Paperless recorder Battery replacement Measurement made easy 1Introduction This publication details replacement

Instruction IN/RandC/007–EN

SM500FPaperless recorder Battery replacement

Measurement made easy

1 Introduction

This publication details replacement of the battery (B12454) fitted to the processor board of the SM500F Paperless recorder. The procedure must be carried out by a trained technician.

2 Safety

3 Tools required

— T8 Torx driver

Important notedispose of the old components as detailed in the Instruction manual (IM/SM500F).

Page 2: SM500F Paperless recorder Battery replacement IN/RandC/007–EN SM500F Paperless recorder Battery replacement Measurement made easy 1Introduction This publication details replacement

2 IN/RandC/007–EN| SM500F | Papeless recorder | Battery replacement

4 Replacing the battery

1. Isolate the recorder from the power supply.

Referring to Fig. 4.1: 2. Unlock the door with the key (if required) A, press release

catch B and open the door.

Referring to Fig. 4.2: 3. Use a T8 Torx driver to remove 6 Torx screws C securing

the door assembly rear cover and remove the rear cover.

Referring to Fig. 4.3: 4. Disconnect processor board flexi-cable D, display

flexi-cable E keypad membrane cable F and inverter cable (if fitted) G from the processor board. Lift out and discard the processor board.

5. Turn the processor board over.

Referring to Fig. 4.4: 6. Remove battery H and insert replacement.

Referring to Fig. 4.3: 7. Position the processor board in the door housing and

reconnect inverter cable G (color recorders only – if fitted) keypad membrane cable F, display flexi-cable E and processor board flexi-cable D.

Referring to Fig. 4.2: 8. Refit the door assembly rear cover and use a T8 Torx driver

to secure 6 Torx screws C.

9. Close and lock the door.

10. Restore the recorder's power supply.

CAUTION – Property damageThe recorder is vulnerable to electrostatic damage. Wear an anti-static strap or place the instrument on an anti-static workbench.

Fig. 4.1 Opening the door

Fig. 4.2 Removing the rear cover

A

B

C

Fig. 4.3 Disconnecting / removing the processor board

Fig. 4.4 Replacing the battery

D

E

F

H

Page 3: SM500F Paperless recorder Battery replacement IN/RandC/007–EN SM500F Paperless recorder Battery replacement Measurement made easy 1Introduction This publication details replacement

SM500F | Papeless recorder | Battery replacement | IN/RandC/007–EN 3

5 Setting the time and date

1. Enter the Configuration level of the SM500F.

2. Enter password (if required) and select Edit current configuration.

3. Select the Common configuration parameter:The Common Configuration menu is displayed:

4. Use the arrows to highlight the Time tab:

5. Press the key to select the Date and timeparameter. A warning dialog is displayed:

6. Press . The Date and time menu is displayed:

7. Set the required time and date.8. Once finished, press and exit and save the

configuration as normal.

Important noteOnce the battery has been removed, the battery failed icon is displayed on the front panel of the unit. To restore normal operation, the correct time and date must set.

Important noteChanging the time will potentially cause loss of data. Ensure that all data archiving is up to date before proceeding.

Page 4: SM500F Paperless recorder Battery replacement IN/RandC/007–EN SM500F Paperless recorder Battery replacement Measurement made easy 1Introduction This publication details replacement

Contact us

IN/R

andC

/007

–EN

01.2

017ABB Limited

Industrial AutomationHoward Road, St. NeotsCambridgeshire, PE19 8EUUKTel: +44 (0)870 600 6122Fax: +44 (0)1480 213 339Mail: [email protected]

ABB Inc.Industrial Automation125 E. County Line RoadWarminster, PA 18974USATel: +1 215 674 6000Fax: +1 215 674 7183

www.abb.com

NoteWe reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document.

We reserve all rights in this document and in the subject matter and illustrations contained therein. Any reproduction, disclosure to third parties or utilization of its contents in whole or in parts – is forbidden without prior written consent of ABB.

Copyright© 2017 ABBAll rights reserved


Recommended