+ All Categories
Home > Documents > Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE...

Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE...

Date post: 06-Nov-2020
Category:
Upload: others
View: 3 times
Download: 0 times
Share this document with a friend
82
IN DEGREE PROJECT MECHANICAL ENGINEERING, SECOND CYCLE, 30 CREDITS , STOCKHOLM SWEDEN 2018 Standardization of Eddy Current Testing Calibration for Valve Spring Wire ANNICK INGABIRE ROBIN OLSSON KTH ROYAL INSTITUTE OF TECHNOLOGY SCHOOL OF INDUSTRIAL ENGINEERING AND MANAGEMENT
Transcript
Page 1: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

IN DEGREE PROJECT MECHANICAL ENGINEERING,SECOND CYCLE, 30 CREDITS

, STOCKHOLM SWEDEN 2018

Standardization of Eddy Current Testing Calibration for Valve Spring Wire

ANNICK INGABIRE

ROBIN OLSSON

KTH ROYAL INSTITUTE OF TECHNOLOGYSCHOOL OF INDUSTRIAL ENGINEERING AND MANAGEMENT

Page 2: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

KTH Industrial Engineering and Management

Standardization of Eddy Current Testing

Calibration for Valve Spring Wire

Thesis supervisor MSc students

Company: Hanna Ahl, MSc Robin Olsson

KTH: Robert Tomkowski, PhD Annick Ingabire

Examiner: Andreas Archenti, Professor

Master of Science Thesis MG213X

KTH Royal Institute of Technology

ITM / Dep Production Engineering

TPRMM-16

SE-100 44 STOCKHOLM

Page 3: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

1

Sammanfattning

Elektromagnetisk provning (ECT) har blivit en av de mest använda metoderna för att

kvalitetssäkra produkter där icke-destruktiv testning av material krävs. Vid provning av ståltråd

interagerar utrustningen med det testade materialet och upptäcker ytdefekter och, i viss

omfattning, om strukturen skiljer sig från det kalibrerade provet. Om produkten befinner sig

utanför specifikationen, skrotas den eller ombearbetas. Denna Mastersuppsats undersöker

kalibreringsförfarandet för elektromagnetisk provning som utförs av Suzuki Garphyttan, som är en

av de största tillverkarna i världen av ventil- och transmissionsfjädrar för bilindustrin. Genom de

slutsatser som framgår av denna rapport, som baseras på undersökningar gjorda i vetenskapliga

artiklar och genom att analysera den data som inhämtats från produktionen, presenteras en

standardisering av kalibreringsförfarandet. Detta är nödvändigt för att säkerställa såväl testernas

tillförlitlighet, såväl som minimering av risken för att skrota ut material på grund av felaktiga

inställningar, till exempel på grund av otillräckligt signal-brus (S/N)-förhållande. Fokus ligger på

sond-baserad, roterande testning, i denna avhandling kallad circografen, eftersom den är manuellt

kalibrerad.

Några av de konstaterade resultaten i rapporten är:

Standard Operating Procedures (SOP)-baserade instruktioner implementeras i företagets

kvalitetssystem. Detta för att minska variationer i kvalitet mellan olika operatörer och

maskiner.

Ett förslag på intervaller för värden (fasvinkel, förstärkning, korrigering av filter och så

vidare) presenteras. Detta är baserat på insamlad unik produktionsdata från operatörer

och utförda test.

Fasvinklarna som används varierar inom specifika intervaller och bestäms av materialval i

allmänhet och frekvensval i synnerhet.

Konduktivitets- och permeabilitetsvärdena för oljehärdad tråd, liksom penetrationsdjupet

för tre olika frekvenser presenteras.

Härdningsfel kan inte detekteras i roterande provning

Ökat kolinnehåll minskar den elektriska ledningsförmågan och ger ökad resistivitet, vilket

gör att fasen flyttar sig och resulterar i ett minskande gap mellan brussignal och

spricksignal.

Nyckelord: NDT, ECT, ståltråd, konstgjorda defekter, sprickor, kalibrering

Page 4: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

2

Abstract

Eddy current testing (ECT) has become one of the most extensively used methods to secure the

products and constructions when non-destructive testing is required. In typical cases of steel wire

testing, the equipment interacts with the tested material and discovers surface defects and, to a

limited extent, if the inner structure is differing from the calibration sample. If the product is

found to be outside specification, it is either scrapped or reworked. This master thesis investigates

the Eddy current testing calibration procedures performed by steel wire manufacturer Suzuki

Garphyttan, which is one of the largest producers in the world of valve and transmission spring

wire for the automotive industry. By the research shown in this thesis, based on the investigation

made in scientific papers and by analyzing data extracted from production, a standardization of

the calibration procedure is being presented. This is to secure both the testing reliability, and

minimizing the risk of scrapping material due to inaccurate settings, for example due to

insufficient signal to noise (S/N) ratio. The focus is on probe-based, rotating testing, in this thesis

called the circograph, since it is manually calibrated.

Some of the findings established in the report:

Standard Operating Procedures (SOP) based instructions is being implemented in the

company's Quality system. This is to decrease the process variations between different

operators and machines.

Suggestions of intervals for values (Phase angle, gain, filter correction and so forth) are

presented. These values are based on collected unique production data from operators and

machines, as well as performed tests.

The phase angles used are ranging between specific value intervals, and set by material

choice in general and choice of frequency in particular.

The conductivity and permeability values for oil-tempered wire, as well as penetration

depth for three different frequencies, are presented.

Hardening error cannot be detected in the circograph.

Increased carbon content is decreasing conductivity and increasing resistivity, causing the

phase to move slightly and decreasing the gap between noise signal and crack signal.

Keywords: NDT, ECT, Steel wire, artificial defect, crack, Calibration.

Page 5: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

3

Page 6: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

4

Foreword

This thesis hereby acknowledges that it is not the result of one’s single effort but rather a

product of combined efforts from different people and organizations. Our gratitude is

hence extended for the advice, guidance, expertise and encouragements shared with us to

everyone who helped put together this piece of work.

Special appreciation is hereby expressed to the following professionals for their remarkable

contribution to the project:

KTH Thesis Supervisor PhD. Robert Tomkowski for his immense support and

enlightenment throughout the project.

Suzuki Garphyttan Thesis supervisor Mrs. Hanna Ahl, MSc, who helped us

through the project in every aspect of the standardization of their ECT

procedures.

KTH Professor Andreas Archenti, as the examiner of this thesis work.

Suzuki Garphyttan as a company, and their professional staff including Stefan

Kandelid (R&D engineer), Erik Elfversson (process manager), Mika Hynninen

(Eddy current technician and expert), Thomas Wrenninge (operations manager),

who has been of great value sharing their knowledge and discussing results and

theories.

Some of the operators have been of great value sharing the knowledge of

calibration, and all operators have been collecting the information of today’s best

praxis, making it possible to compile it into a full instruction. We cannot mention

all here, but we show gratitude to Therese Magnusson, Fredrik Nilsson and Per

Rickardsson and others for taking time to share their knowledge and competence.

All our KTH colleagues in the School of Industrial Technology and Management

are hereby thanked very much for their everyday support during the project period,

as well as for the 2 years of the Master. Among many, Ove Bayard, PhD, has

been of great help and support.

Our Final Thanks go to our families and friends for their constant encouragements

and support they have demonstrated throughout the duration of studies at KTH,

especially when the project was ongoing.

Robin Olsson and Annick Ingabire

Stockholm, June 2018

Page 7: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

5

NOMENCLATURE

Abbreviations

Abbreviation Significate

SG Suzuki Garphyttan

NDT Non-destructive Testing

NDE Non-destructive evaluation

EC Eddy Currents

ECT Eddy Current Testing

OT Oil-Tempered

ISO International Standards Organization

AISI American Iron and Steel Institute

ASTM American Society for testing and Materials

SIS Swedish Standards Institute

SS Svensk Standard

TS Technical Committee

SI International System of Units

SOP Standard Operating Procedure

CM Color marks

UHT Ultra High Tensile Steel

EDM Electrical Discharge machining

S/N Signal to noise

AC Alternative current

Page 8: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

6

Notations

Latin Symbols Unit Description

A [mm2] Area

D [mm] Diameter

f [KHz]or [kg/m3] Frequency

G [dB] Gain

R [Ohms] or [Ω] Resistance

L [m] Length

H [A/m] Magnetic field strength

B [T or Wb/m2] Magnetic flux density

Greek symbols Unit Description

µ [H/m] or [Wb/(A m)] Magnetic Permeability

σ [S/m] Electrical conductivity

ρ [Ohm meter] or [Ωm] Resistivity

ω [Rad/s] Angular frequency

δ [μm] Standard Penetration depth

π 3.14159 Pi

Page 9: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

7

Contents

1. INTRODUCTION .................................................................................................................... 9

1.1 TECHNOLOGY OVERVIEW ....................................................................................................................9 1.2 THESIS AIMS .................................................................................................................................... 10 1.3 PROJECT SCOPE .............................................................................................................................. 10 1.4 ABOUT SUZUKI GARPHYTTAN ........................................................................................................... 11

1.4.1 Wire production ...................................................................................................................... 11 1.4.2 ECT and the reason for inline testing. ..................................................................................... 14

1.5 THE CHALLENGE OF THIS WORK ........................................................................................................ 15 1.6 THESIS LAYOUT ............................................................................................................................... 16

2. FRAME OF REFERENCE ...................................................................................................... 17

2.1 EXISTING TECHNOLOGY ALTERNATIVES .............................................................................................. 17 2.2 EDDY CURRENT TESTING .................................................................................................................. 19

2.2.1 History .................................................................................................................................... 19 2.2.2 Generation of Eddy currents .................................................................................................... 19 2.2.3 Factors affecting Eddy Current Testing ................................................................................... 20

2.3 EC TESTING REFERENCE STANDARDS ................................................................................................. 31 2.3.1 Standards for EC testing equipment ........................................................................................ 31 2.3.2 Standards for calibration ......................................................................................................... 32 2.3.3 Standard Operating Procedure (SOP) for calibration ............................................................... 32

3. METHODOLOGY ...................................................................................................................35

3.1 QUANTITATIVE AND QUALITATIVE EVALUATION OF THE CALIBRATION PROCEDURE ................................. 35 3.1.1 Quantitative analysis ............................................................................................................... 35 3.1.2 Qualitative analysis ................................................................................................................. 35

3.2 EQUIPMENT USED ............................................................................................................................ 36 3.2.1 Circograph and the procedure of calibration ............................................................................ 36 3.2.2 Defectomat ............................................................................................................................. 44 3.2.3 Contracer and other measuring equipment ............................................................................... 44 3.2.4 Calibration samples ................................................................................................................. 45 3.2.5 Microscope and mounted samples ........................................................................................... 46 3.2.6 Generally about the methods for this thesis paper .................................................................... 46

4. RESULTS .............................................................................................................................. 47

4.1 RESULTS ANALYSIS ........................................................................................................................... 47 4.1.1 Phase angle determination ...................................................................................................... 48 4.1.2 Gain ........................................................................................................................................ 51 4.1.3 Filter correction ...................................................................................................................... 54 4.1.4 Conductivity and permeability measurement ............................................................................ 57 4.1.5 Frequency and phase ............................................................................................................... 60

4.2 OTHER OBSERVATIONS ..................................................................................................................... 62 4.2.1 How the amount of color marks seemed to trend with increased alloy content ......................... 63 4.2.2. Defect of different conductivity (lead, Pb) .............................................................................. 64 4.2.3 Artificial defect reliability ........................................................................................................ 65 4.2.4. Calibration with lead (Pb) disturbance or directly from lead ................................................... 67 4.2.5. Dirt, oxide remains and oil covering the artificial defect .......................................................... 69

4.3 DEFINITION OF IMPROVEMENT POSSIBILITIES ...................................................................................... 70 4.4 RESULT OF SOP INSTRUCTION FOR OPERATOR ................................................................................... 71

5. CONCLUSIONS ..................................................................................................................... 73

5.1 CONCLUSIONS .................................................................................................................................. 73

Page 10: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

8

5.2 DISCUSSIONS ................................................................................................................................... 74 5.3 FUTURE WORK ................................................................................................................................. 75

LIST OF TABLES ...................................................................................................................... 76

LIST OF FIGURES ..................................................................................................................... 76

REFERENCES ............................................................................................................................ 78

Page 11: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

9

1. Introduction

This Introduction section is divided into six parts: 1) Technology overview, 2) thesis aims,

3) project scope, 4) a summary about Suzuki Garphyttan where the work has been

performed, 5) the challenges of this thesis and 6) the thesis layout. The aim is to give an

overview and introduction of the thesis work.

1.1 Technology overview

It is a common phenomenon that valve spring wires, which are used in engines for the

automotive industry, can have mechanical defects or cracks causing a spring failure,

which eventually lead to a failure in the complete engine of the car or the truck. The

defects causing the breakages can be present in the raw material, as it is delivered as wire

rod from the steel mill, or introduced later on during the various stages of wire production

to its final dimension and final material properties. Depending on the context, such

defects must be clearly identified and removed before reaching the customer, since even

the smallest crack or defect can cause a breakage of the spring, and causing a failure in

the engine. This means both the wire and the spring needs to be tested, to ensure no

critical defects exist.

There are several established methods for non-destructive testing, including but not

limited to: X-rays, ultrasonic testing and eddy current testing (ECT) [2].

During valve spring wire production, every precaution is taken to avoid a damaged wire

surface. Material is removed in an initial process called shaving, where the outer oxide

layer of the wire rod from the steel mill is removed, including potential cracks. After that,

the wire is carefully protected throughout the wire production process, to avoid sharp

edges as well as protection from metal to metal contact in the different process steps.

Still, a final, non-destructive testing procedure needs to be done to ensure the quality,

and this testing consists of using ECT. ECT is suitable for testing of surfaces on

conductive materials like steel. By using ECT, surface imperfections deeper than 40µm

[3], can be detected, marked with color and subsequently removed in the spring coiling

operation.

The expected result from this project is to update instructions used by the operators

during calibration of ECT, and to establish value intervals of important parameters used

during ECT calibration. By having a standardized calibration, the possibility to minimize

signal to noise level and to avoid false alarm color marked defects is enhanced. It is also a

securement that no defects will be missed in the future due to lack of competence or

machine failures.

Page 12: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

10

There are a lot of properties, both in settings of the equipment and of the material, that

influence the result, including conductivity, phase angle, gain, steel properties, artificial

defect depth, natural defect depth and so forth. The key to a reliable result is to set those

parameters right. The importance and explanations of the terms will be further examined

in the Literature review part, and connected in Result section.

1.2 Thesis aims

The main purpose of this master thesis is to;

Update current official instructions used by the operators

Investigate how different calibration settings affect the testing reliability and try to

find the best praxis of current knowledge, to set a standard for the future.

Fill the knowledge gap of used intervals for different parameters as well as material

properties and its impact on the testing.

Today there is a short instruction on calibration settings, but the description is vague and

referring mainly to operator competence, and in the increasing requirement of modern

production, a framework resulting in an instruction should be implemented, based on the

operators’ knowledge, previous production data and a scientific background.

1.3 Project Scope

The thesis aim is to update the instructions on what settings to use when calibrating the

circograph for ECT of valve steel wire produced by a specific company. The company is

one of the market leaders in the world producing valve spring wire. The instruction

includes recommended intervals of gain, phase angles and other parameters that the

operator or management use when the calibration is being executed.

Limitations for the project are:

Only oil-tempered wire between 2.0-6.55 mm used at Suzuki Garphyttan will be

investigated.

The only non-destructive testing, NDT, method examined is eddy current testing,

ECT. This is due to the fact that ECT is the only NDT-based evaluation

performed today in the company. ECT is suitable for detecting surface defects and

cracks in conductive materials, which is a main failure cause for spring breakages

in automotive engines. Other NDT-methods are presented shortly, but the main

focus is to standardize the existing technology, not to compare different

Page 13: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

11

techniques. The company executes other tests to detect inclusions and other

material failures, which is outside the scope of this thesis.

Only circograph (rotating probes) will be investigated, not defectomat (encircling

coil). The reason for this choice is that a daily calibration is being done in the

circograph, but the defectomat already have standardized values.

The results are only valid for Foerster GmbH equipment, a German producer

considered as a leading manufacturing of ECT equipment. Foerster GmbH has also

given the names “circograph” and “defectomat”, which are used in this report.

The results and the values are only valid for the specific environment, product, site

and process which the tests has been performed. The results are however

comparable to other environments, and other examinations, articles and reports

has served as a guide on how to set the experiments and values presented in this

report. The aim is to extend this field of knowledge with the material type used

for valve spring wire.

1.4 About Suzuki Garphyttan

Suzuki Garphyttan has produced wire since 1906. Since 1927 the main product is oil

tempered valve spring wire for the automotive industry. During the years, other special

wires such as stainless round wire, flat rolled and shaped wire in carbon, low allowed and

stainless material have been added to the production portfolio.

The company’s focus is spring wire for combustion engines, mainly valve and transmission

springs, piston rings, compression rings, fuel injection springs used in cars and trucks all

over the world.

Eddy current testing is carried out on material with high demands on surface quality. It is

performed with both rotating (R) and stationary (D) probe test equipment.

1.4.1 Wire production

Basically, oil tempered wire for the automotive industry has the following procedure:

The production starts with wire rod from external suppliers, commonly called “coils” in

the industry (not to be confused with eddy current coils, which in this thesis is referred to

mainly as probes). The weight of a wire-coil is around 2 tons per coil, and is delivered in

a few different standardized dimensions, like 8.00mm or 9.00mm, and then processed to

the demands of every specific customer. The goal of the wire production is as follows:

Wire with correct dimension, like 3.95 mm or egg-shape like 3.00x2.39mm2.

The surface should be defect free

The black oxide layer should be stable and cover the wire

Page 14: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

12

It has to be hardened with the right properties (tensile strength, yield strength,

fine martensitic structure and so forth).

To achieve this, the wire is processed in 5 steps as described shortly below. The steps are:

1) Shaving

2) Recrystallization

3) Drawing

4) Hardening

5) ECT

Common procedures and production flow in wire production is shown in a more

visualizing way on the following figure 1, where also the steel grain structure and the

color of the surface is visualized.

Figure 1: The production steps and a schematic simplified view of structure, grain size, oxide layer and dimension

Now follows a more detailed description of the process steps.

(i) Shaving

Shaving is the first step performed and means that the wire is machined through a

shaving die, simply removing the top layer of the wire rod (oxide scale mainly, but also

cracks and decarburization), making it looking very shiny (see figure 2). The advantage of

this procedure is that removes cracks and defects originated from the wire rod mill or

from transportation. The disadvantage of this is that a significant part of the weight is

being removed, which is costly. Other alternatives to shaving are chemical material

Page 15: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

13

removal (pickling) and shot blasting, which is mainly removing the oxide but not deep

defects or wire rod cracks. Therefore, for the toughest applications, material removal is

necessary.

Figure 2:How the wire rod looks after shaving

(ii) Recrystallization

When the shaving process is done, the surface layer of the wire becomes very smooth, but

the upper surface layer is also becoming martensitic or deformed perlite. Since the next

production step is heavy reduction in the drawing machine, the shaved layer will crack up

when drawing reduction is performed due to the tension between the martensitic top layer

and the ductile core. This would cause heavy damage to the wire. To avoid it, the wire is

being recrystallized in a furnace. The temperature is lower than in hardening furnaces, but

enough for the surface grain structure to relocate and become more ductile.

(iii) Drawing

Drawing is the heart of wire production. Here the final dimension will be set, which the

customer has required. The drawing machines consists of several drawing capstans,

typically between 3-10 capstans, which will drag the wire through drawing dies, each die

size smaller and smaller, forcing the wire to reduce in size. Since the force is high, the

drawing process is very sensitive for defects, typically scratches and mechanical defects,

which can appear when the drawing die get damaged, the cooling in the machine is

insufficient, or there is other contact between metal equipment and the wire. The

principle of drawing is shown in figure 3.

Page 16: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

14

Figure 3: wiredrawing die

After drawing it is time to set the final material properties in hardening process.

(iv) Hardening

Hardening is a process where the wire is transformed through high temperature to

austenite, then cooled quickly and quenched in oil creating a martensitic structure, which

makes the wire hard but also brittle. Then the wire is being annealed in lower

temperature, to keep the main part of the strength but gaining more flexibility and less

brittleness.

Figure 4: Hardening procedure.

The heating process is done by using a combination of high frequency induction heater

and an indirect heating furnace. The pipes usually hold inert atmosphere, to avoid

decarburization. The furnace keeps the wire’s temperature constant until immediately

before the oil-quenching, and during the time in the furnace a surface oxide film grows,

which also plays an important role as lubrication of the wire during the spring

manufacturing. The heated wires are then oil-quenched, creating a fine martensitic

structure, and finally annealed in molten lead bath to set the tensile strength and other

mechanical properties. After tempering, water cooling bath is used for cooling the wire

and rust protection is applied before enter WIP area in front of ECT. The flow is

visualized in figure 4.

1.4.2 ECT and the reason for inline testing.

ECT as a procedure will be further analyzed in the coming sections. However, the reason

for the wire to be checked inline is to discover and sort out surface defects. There are

Page 17: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

15

many types of defects, but mainly wire rod defects and so called button type defects are

the most serious. Also mechanical marks and scratches are common. The example below

(figure 5) shows how a wire rod defect can look like on the wire, as well as its

characteristic appearance, the crack is around 270µm, and the limit is 40µm. The crack is

partly covered by material. The defects are usually difficult to see by the human eye,

unless it is specified exactly what part and position of the wire that is damaged. In ECT,

the defects, also partly covered by material, are quickly detected, which is one of the

reasons for the methods’ popularity in wide field areas like nuclear plants, airplanes, pipes

and so forth.

Figure 5: Wire Rod crack, surface picture and its corresponding crack appearance in a cut

When a defect is detected it is marked by a white color, and later sorted out by the

spring producer. Therefore, no defects appear in the engines it is intended for. In the

Results part, color marks (CM) are mentioned a few times in an evaluation purpose. CM

means the number of times a fault is discovered and color marked, and can be measured

in a total amount or for comparison purposes by CM/100kg or CM/1000m or other

preferred units. The color marking is done even if the equipment is (calibrated) right or

wrong, or disturbed in another way causing false alarms. The intention is to only mark

flaws that are outside specification, for valve spring wire material 40µm or above.

1.5 The challenge of this work

The challenge in this project is consisting of mainly the following points:

1) There is a limited description within the company on how to do a calibration,

since it is considered a craftwork and something the operator learn by experience

Page 18: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

16

and peer-to-peer learning. However, in modern production, a calibration system

shall be established to ensure that measuring and test equipment are maintained

by periodic calibration.

2) It is thought within the company, that the safety margins for defects are too high,

or that the equipment, by itself or by the human factor of insufficient calibration,

for unknown reason are too sensitive and giving signal to defects that in fact are

small, so called false alarms. This is costing the company a lot, as well as giving

extra work for their customers to remove those defected material parts. Figure 6

shows a defect that according to the equipment measurement is over 40µm, when

in fact, the depth is only around 20µm.

Figure 6: Depth of the defect

Compared to the wire rod crack shown on figure 5, this defect (figure 6) is only 20µm

and should not have been sorted away, since the limit is 40µm. This is causing over-

scrapping of material.

1.6 Thesis layout

The project thesis is arranged in five sections as listed below:

Introduction

The Introduction consists of an overview of the project including the technology

overview, thesis aims, the project scope, and a summary about Suzuki Garphyttan,

the challenge of this work and thesis layout.

Frame of reference

Page 19: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

17

In this chapter, an overview of the existing technology alternatives will be

described, as well as an in-depth description of ECT and is influencing factors.

Previous research papers will be referred to, to provide the knowledge and

information about ECT. Finally, ECT reference of standards used in the project

will be described, and a description of the SOP-based instruction that will be the

base for future calibration in the company.

Methodology

In this section, the calibration procedure of the circograph is described, as well as

other equipment used in this project, their methods and purpose which help to

reach the goals for the project.

Results

The results that are obtained with the methods described in the Methodology

section are compiled and analyzed as well as the results of the updated SOP for

calibration of the circograph.

Conclusions

Finally, a discussion of the results, conclusions and future work is presented in

Conclusions. The conclusions are based from the analysis with the intention to

answer the thesis questions in 1.2 Thesis Aims.

2. Frame of reference

The aim of this section is to provide a summary of the existing knowledge and previously

performed research in pursuing the project objectives. Existing technology alternatives in

the field of nondestructive testing is presented in this chapter, some of the factors

affecting eddy current testing as well as reference standards used in ECT.

2.1 Existing technology alternatives

Non-destructive testing, NDT, techniques are used in the metal industries to evaluate the

properties of a wide variety of materials without causing damage to the specimens. Some

of the most common NDT techniques utilized for inspection of conductive materials such

as copper, aluminum or steel are electromagnetic eddy current based, ultrasonic or liquid

penetrant testing. An overview of different techniques is described in table 1.

Page 20: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

18

Table 1: Different NDT method used in industry [1]

Method Principles Application Advantages Limitation

Ultrasonic

examination

High frequency sound

pulses from a transducer

propagate through the

test material, refracting

at interfaces

Most materials can be

examined if sound

transmission and

surface finish are good

and the shape is not

complex

Provide precise, high

sensitivity results quickly.

Thickness information,

depth and type of flaw

can be obtained from one

side of component

No permanent record

(usually). Material

attenuation, surface finish

and contour. Required

couplant

Radiographic

testing

Radiographic film is

exposed when radiation

passes through the test

object. Discontinuities

affect exposure

Most materials, shape

and structures.

Examples include welds,

castings, composites

etc. As manufactured

or in service

Provides permanent

record and high

sensitivity. Most widely

used and accepted

volumetric examination

Limited thickness based

on material. Density,

orientation of planar

discontinuities is critical.

Radiation hazard

Magnetic

particle

testing

Uses reflected or

transmitted light from

test object that is

received by a light

sensing device

Applied in industries

ranging from raw

material to finished

products and in-service

inspection

Can be inexpensive and

simple with minimum

training required. Broad

scope of uses and benefits

Only surface conditions

can be evaluated.

Effective source of

illumination required.

Penetrant

liquid

A liquid containing

visible or fluorescent dye

is applied to the surface

and enters

discontinuities by

capillary action

Virtually any solid non-

absorbent material

having uncoated surface

that are not

contaminated

Relatively easy and

materials are inexpensive.

Extremely sensitive, very

versatile. Minimal training

Discontinuities open to

the surface only. Surface

condition must be

relatively smooth and free

of contaminants

Eddy current

testing

Its technique is based on

the interaction between

a magnetic field source

and the test material

which induces Eddy

currents in the test

piece

Detect the presence of

very small cracks by

monitoring changes in

the Eddy current flow

ECT is high speed

testing, it can be

automated.

Discontinuities at or near

surface can be reliably

directed. Accurate

measuring of conductivity

and coating thickness

measurements. No

physical contact required,

low cost and portable

Limited penetration into

test material, several

variable simultaneously

affect output indication,

discontinuities are

qualitative not

quantitative indications,

material must be

conductive, requires skills

when many variables are

involves, false indications

can result from edge

effects and parts geometry

The focus of this thesis is to improve and standardize ECT calibration procedures in

Suzuki’s valve spring wires. Nevertheless, knowledge about other testing techniques can

improve the understanding of how Eddy current testing works and its advantages and

limitations. The thesis does not rank the different techniques. The objective is to

understand the present testing process based on eddy currents, and to improve the

stability in the process, not to compare different techniques. In general, ECT is suitable

since the wire is conductive, which is the base to use ECT. Eddy current based evaluation

is also easy to run in line with high productivity. ECT is surface based mainly, and is very

sensitive to changes in the surface made by cracks, scratches or similar.

Page 21: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

19

2.2 Eddy current testing

Eddy current testing, ECT, is the specific technology for this project and the objective of

this subsection is to describe its history and principles, and a brief discussion is made

about some of the factors that affect ECT.

2.2.1 History

The principle of ECT has its origin based on Michael Faraday's discovery of

electromagnetic induction in 1831. Later in the 1900th century scientists noticed changes

in the properties of a conductive coil when placed in contact with metals of different

conductivity and permeability. During Second World War these effects were put into

practical use in testing materials. Much work was done in the 1950’s and 60’s, in the

aircraft and nuclear industries. Friedrich Foerster, founder of Foerster GmbH, and Robert

McMaster are seen as the main pioneers within the field [28]. Today, Eddy current testing

is an accurate and well-understood inspection technique [4].

2.2.2 Generation of Eddy currents

Eddy currents are created through a process called electromagnetic induction, and to

generate Eddy currents for an inspection, a probe is used. Inside the probe, there is a

length of electrical conductive material which is formed into a coil. Alternating current

flows in the coil at a frequency chosen based on the product to be tested. A dynamic

expanding and collapsing magnetic field forms in and around the coil, as the alternating

current flows through it. When an electrically conductive test material is placed in the

coil’s dynamic magnetic field, electromagnetic induction occurs, and Eddy currents are

induced in the material [5]. Eddy currents flowing in the material will generate their own

secondary magnetic field which opposes the coil’s primary magnetic field as shown on

figure 7. This entire electromagnetic induction process to produce eddy currents may

occur from several hundred to several million times each second depending upon the

chosen frequency [21]. In this work the main study has been done on settings with 300

kHz, with complimentary test of the frequencies 1000 kHz and 100 kHz. When a flaw is

introduced to the conductive material or test piece, the Eddy currents are disrupted, and

the signal is transferred to a screen where the crack has a different phase compared to

the normal undisrupted noise based on the calibration sample.

Page 22: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

20

Figure 7: Generating Eddy current on the test piece [25]

The phase is built up on the analysis of the impedance, which is based on the resistance,

inductance and the reactance of the material. A change within the tested material, by

mechanical defects or by a change of material type and/or structure, will result in a

change of the impedance angle, and the signal will change, which can be detected.

There are different types of ECT-equipment; the two main types are the defectomat and

the circograph, as they are named in this thesis. The names are taken, as previously

mentioned in the Introduction section, from the manufacturing company Foerster GmbH.

The circograph, which is rotating with two or more identical probes around the test

material, is mainly for longer defects, like scratches. The defectomat, with an encircling,

non-rotating coil, has better ability to detect shorter defects. In this report, the focus is

on the circograph, since it requires adjustment every time a new dimension is run, and

several recommended values are missing. For a more in-depth description of the

calibration of the circograph, see Methodology section. Both types are however affected

by the same factors which will be presented in section 2.2.3.

2.2.3 Factors affecting Eddy Current Testing

Apart from a defect, there is a variety of factors that affect the response of an Eddy

current probe. Those factors include electrical conductivity and magnetic permeability of

the test material, the operational frequency of the eddy current probe, the skin effect and

penetration depth, the proximity of the probe to the surface and geometric properties of

the test material. Also, a factor which is partly setting based, is the so called Noise to

Signal (S/N) ratio, which is described after the equipment and material factors. A

summary of those factors is given in this subsection.

Page 23: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

21

(i) Conductivity

Conductivity of a material is a main factor to be considered when using eddy current

testing inspection. The greater the conductivity of the material, the better flow of eddy

currents is on the surface. The conductivity is inversed to the resistivity. There are

different factors affecting conductivity of a material such as material composition, heat

treatment, hardness, temperature of a material, internal residual stresses and conductive

coatings [22].

Concerning the material properties, grain size and structure type, when a material (metal

or alloy) is subjected to heat treatment, the metal properties are changed. This change in

is causing an internal change in the material which also affects the conductivity of the

material [5]. The change in conductivity can also be detected by ECT methods, but

according to the analysis in this thesis, the defectomat (encircling testing) is more likely

to notice the change in structure, caused by for example a hardening error. The encircling

coil discovers even a slight change in conductivity (like a hardening error), while a

rotating unit mainly discovers a complete change of material, for example lead residues on

the steel surface. This is visualized in the figure 8 below, which shows how the signal

appears when the material has been unsatisfactory hardened and the material is almost

perlite. The calibration sample, which the machine refers to, is hardened and annealed

martensitic. The signal does not reach the trigger level in the rotating circograph but

gives a large signal in the encircling defectomat. Since this thesis is focusing on the

circograph and not defectomat, hardening errors are only shortly discussed here, but more

focus is put on complete changes of conductivity, like when the ECT is disturbed by

another type of material, like lead or by a change of material. The behavior of the

different steel types investigated is described in the Results section.

Figure 8: Hardening error (perlite structure). A small detection signal can be seen in the circograph (rotating probe

testing), but at large signal can be seen in the defectomat

Page 24: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

22

Figure 8 shows hardening error (perlite structure instead of hardened martensitic

structure). Not detected, meaning it is not reaching the trigger level, by circograph (left)

but by defectomat (right), since the conductivity difference for the rotating probe is too

small. Examples of lead interference are further explained in the Results section.

In general, it is difficult to measure conductivity on wire with small diameter range, since

the probes usually used in common conductivity devices are too big for the thin wire

dimension used in this research. According to the data on typical annealed steels, the

conductivity decreases by increasing carbon content. Table 2 show different AISI steels,

and the figure shows that increasing levels of alloys in general, and carbon specifically,

decreases the conductivity. This also implies that the resistivity is increasing with

increasing carbon content.

Table 2: Different AISI-steel and their alloy content and Conductivity values [11]

Figure 9: The conductivity decreases with increasing carbon content for some random chosen annealed AISI-steels

Page 25: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

23

Since the impedance, which is the signal picked up in ECT, is consisting of the

hypotenuse of the resistance and the reactance/inductance, a change in Resistance will

impact the impedance [24].

In this project, actual values (complete year 2017’s data) of different steel grade of wire

used at Suzuki Garphyttan are being used to analyze the steel composition. Around 20

000 tons produced were analyzed. By analyzing those values, it can be shown somewhat

different values as shown in table 3:

Table 3. Average alloy values of different grade of Suzuki Garphyttan wire

In general, the alloy content is increasing according to the end number of the steel,

OTXXX. OT101 has higher alloy content than OT70 based on this logic. An exception

from this logic is 75KD, which is used for other products than valve springs and also has

different testing requirements. As a result from this analysis, figure 10 shows that the

composition of alloys is increasing, and this is in Results section correlated to the phase

angle and the conductivity.

Figure 10: Chemical composition analysis regarding average of carbon (%)

Page 26: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

24

Although it is difficult to get conductivity values for this thin wire with the devices found

on the market, another way of receiving conductivity values is to base the conductivity on

the resistivity values. One way is to recalculate conductivity by using Poulliet’s law [26]:

Knowing this, it is also established that conductivity is the inverse of the resistivity:

By combining these two and replacing the resistivity, the following equation defines the

conductivity:

Where:

σ is Electrical conductivity [S/m]

ρ is Resistivity [Ohm meter] or [Ωm]

A is Area [m2]

R is Resistance [Ohms] or [Ω]

L is Length [m]

This equation is used to calculate the conductivity values presented in 4.1.4 (Results).

(ii) Permeability

The permeability of a material is described as the ability with which a material can be

magnetized. If the magnetic field strength at various locations varies even slightly, these

small variations have a large effect on the impedance of the coil. These changes in the

impedance of the coil are often so large that they mask all other changes such as changes

in conductivity or dimensions [8].

High magnetic permeability makes the standard penetration depth decrease. To explore

the material internally, ferromagnetic materials are inspected at lower frequencies that

non-ferromagnetic materials [5]. In the testing environment of this thesis, the material is

being demagnetized before entering the circograph, as well as the residual magnetism is

measured once every quarter for every ECT-line. The main focus for the de-magnetization

is to eliminate the artificially created magnetic increase before the defectomat testing,

where the material is magnetized according to the hysteresis curve, to give a more reliable

testing in the defectomat. The material is then put through the demagnetization unit

which gives the material is original properties.

Page 27: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

25

Magnetic fields consist of two closely related fields, and the relation between them defines

the permeability [27]. The two fields are called B and H. B represents the magnetic flux

density, and is measured in Tesla (T) or Weber per square meter (Wb/m2). H represents

the magnetic field intensity and is measured in Ampere per meter (A/m).

The values of the magnetic field intensity (H) is measured quarterly within the company

to control that the de-magnetization works and to secure that the magnetic properties are

not too intense. If the values are too intense the wire and the springs will attract metal

particles during the spring manufacturing, shot-blasting, annealing as well as in the motor

of the end customer. To avoid this, the machines do not start if it the demagnetization

does not work. It is in general difficult to measure the magnetic intensity. The field

intensity can differ largely on the same material and possible also how the cut is made

(by saw or by other cutting tools). On the surface, which has an oxide layer on the

outside, almost no magnetization can be noticed. The biggest value found, when the

probe is moved back and forward, is the one noticed in the protocol.

The permeability is given by [27]:

Where:

H is Magnetic field strength [A/m]

B is Magnetic flux density [T or Wb/m2]

µ is Magnetic Permeability [H/m] or [Wb/(A m)]

This equation is used to calculate the permeability values presented in 4.1.4 (Results).

(iii) Frequency

The operating frequency chosen for the inspection also have a significant effect on the

eddy current probe response. Frequency selection affects both the phase relationship and

the relative signal strength of the response from different flaws. Choosing the proper

frequency is critical to acquiring optimal resolution between flaw signals and noise

contributions from the material under test [21]. The difference between noise and signal

should be as large as possible.

Frequency can be an operator-controlled variable, but it is standardized and locked in the

testing environment which was investigated in this thesis. The main use of the frequency

is controlling the depth penetration, the density and phase of induced Eddy currents. In

general terms, higher frequencies are used to detect surface breaking discontinuities, and

lower frequencies for subsurface testing [8]. Phase difference between noise and defect is

also affected by the frequency level. Depending on the type of material this can be

Page 28: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

26

significant differently. An example is given in figure 11 showing the phase difference

between the signal of a defect and the sample’s natural noise as a function of frequency,

the difference between noise and signal is 30°, and is reached between 250-300 kHz in in

this particular case[9]. This phase angle difference is individual according to the material

tested. As can be seen in the Results section the difference is larger for the material used

for this thesis research. Still, figure 11 and the curve seen is an example of what to be

expected in the environment this research is performed in.

Figure 11: Difference between noise and defect signal [9]

For the oil tempered wires that were investigated, almost all defects are parted around

50°-90°, which gives a more reliable testing condition with better S/N-ratio. On the other

hand, a frequency of 300 kHz gives lower penetration depth, meaning that mainly the

surface is being evaluated, increasing the risk for false alarms due to rough surface or

metal particles [9].

Page 29: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

27

Figure 12: Example of the difference between the noise and the crack for 300 kHz, around 70°-90° in this case

(iv) Skin effect and penetration depth

As mentioned above, the skin effect and the penetration depth, are affecting the

sensitivity of the testing. The penetration depth is set by a combination of parameters

consisting of frequency, permeability and conductivity. The following equation number 5

shows how to calculate the penetration depth. Standard penetration depth depends on

electrical conductivity, magnetic permeability of the test material and on Eddy current

frequency (see figure 13 for a schematic view).

or simplified:

Where:

f =frequency [kHz]or [kg/m3]

μ = permeability [H/m] or [Wb/(Am)]

σ =conductivity [S/m]

ω =angular frequency [Rad/s]

ρ = resistivity [Ωm]

Page 30: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

28

The later equation (6) is useful when conductivity is not given and the penetration depth

is calculated on metals or materials. Standard penetration depth decreases as

conductivity, permeability or inspection frequency increase as shown on figure 13.

Equation (5) shows how frequency influences the penetration depth. When using the low

frequency, permeability or conductivity, the penetration depth increases.

Skin effect is when Eddy current flowing in the test object at any depth produce magnetic

field which oppose the primary field thus reducing net magnetic flux and causing a

decrease in current flow as depth increases. Alternatively, Eddy currents near the surface

can be viewed as shielding the coil’s magnetic field thereby weakening the magnetic field

at greater depths and reducing induced currents [8]. Skin effect is also a limiting factor of

increasing frequency as desired.

Figure 13:Relative Effect of frequency, conductivity and permeability on the depth of penetration for a typically single-

coil ECT probe

(v) Proximity

Lift-off and fill factor are the terms used to describe the space that occurs between the

specimen under test and the inspection coil or probe. Both have a similar effect on the

eddy current response. Lift-off and fill factor are essentially the same thing, lift-off is

applied when rotating probes are used, and fill factor is used for encircling testing. Lift-off

is the impedance change that occurs when there is a variation in the distance between the

inspection coil probe and the test piece. Those small variations in spacing can mask many

indications and create noises. Fill factor is applied in an encircling or an internal coil;

basically it is a measurement of how well the conductor (test specimen) fits the coil [5].

The rotating surface probes should be presented to the test surface at a constant angle to

the surface with constant lift-off and pressure if possible [8]. Phase angle and wire speed

Page 31: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

29

on every eddy current testing line is taken into consideration to analyze the changes to

the signal from the rotating surface probe.

(vi) Geometry

Geometry of the test material such as thickness and edge or end-effect can also influence

the Eddy current response. Material thickness influences the measurements, especially

when material thickness is less that the effective depth of penetration. Edge effect or

End-effect is described as a phenomenon that occurs when inspecting coil is at the end of

test piece, it occurs when Eddy current testing probe approaches the edge of a specimen,

where current has no place to flow, this may result in false indication [22]. The distortion

results in a false indication that is known as edge effect [8].

(vii) Signal to noise ratio

Signal to noise (S/N) ratio is an important word in the NDT vocabulary and should be

introduced early. It is not a characteristic of the material or equipment, but is important

to get a meaningful testing in that sense that a high ratio gives a good reliability, so the

ECT equipment sorts out real defects and not general noise. Signal to noise ratio is a

parameter that defines and relates the height of the signal to the height of the noise that

comes natural from the material itself. The response of a crack should be considerably

greater than the signal amplitude of the background noise [5]. The appearance of a crack

in ECT will be further explained in Methodology part, but a schematic picture on how to

calculate the ratio can be seen in figure 14 and equation 7.

Figure 14: Relative Effect of frequency, conductivity and permeability on the depth of penetration for a typically single-

coil ECT probe

The S/N ratio is simply calculated by:

Page 32: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

30

A ratio less than 3:1 or 300 % should be rejected, since the risk for false alarms or

missing real defects are too big. In figure 14, the testing should not be performed

according the calculation using equation (7), the signal noise ratio is equal to:

2,33 =35,00

15,00

In this report there are two ways of measuring S/N, one conventional definition of noise

and one unique of this thesis. The conventional way, from articles and from the

equipment supplier Foerster GmbH, is to simply measure noise and signal, for example by

the Vector mode, see figure 15. There the difference between the top (in the middle) and

the level on the sides (the noise) is measured. This is illustrated in figure 15, and

correlates to the figure 16. It is called longitudinal noise or just “noise”.

Figure 15: S/N between then top (crack signal) and the sides (surface roughness noise)

A possible unfavorable part of the longitudinal noise evaluation is that it does not

consider the stability, neither the length, especially if the noise is laid down as it usually is

when ECT is performed in Garphyttan. See figure 16 and 17. The noise height is then

close to 0 (figure 17), but the signal width could be very unstable. In this thesis, for some

dimensions also the latitudinal length is measured. Often, however, the noise values in

both longitudinal and latitudinal directions correlates. High longitudinal noise also means

high latitudinal noise.

Figure 16: Normal noise (here called longitudinal) and latitudinal noise

Page 33: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

31

To further visualize this phenomenon, figure 17 is showing the concept with latitudinal

noise, showing two different scenarios. The height of both the crack and the height of the

noise is the same in both figures; therefore the conventional S/N ratio is the same. But

the width may not be equivalent in different scenarios. Therefore, also the width and

stability of the signal should be evaluated.

Figure 17: Further developing the latitudinal noise in yellow (short) and purple (long)

A calibration resulting in a high latitudinal value, causing a planar but wide signal, is

more likely to create false alarm compared to a shorter noise when running in motion,

even if both have a really good S/N ratio during non-motion calibration. This is due to

actual production, which will always include minor vibrations, disturb from dirt etcetera.

Therefore more noise appears during running compared to when the calibration was done.

2.3 EC testing reference standards

Suzuki Garphyttan has a quality management system certified according to the

international standards for quality ISO TS 16949. The work is ongoing to also achieve the

upcoming new standard in the automotive industry, the IATF 16949. Main suppliers of

wire rod have QMS certified by a third party according to ISO 9001 as a minimum and

are also regularly audited by Suzuki Garphyttan [17].

To inspect wire by ECT, different standards are considered to provide uniformity of the

testing procedure. In this subsection different standards will be used as a reference and

the SOP instruction is based on this information.

2.3.1 Standards for EC testing equipment

In this thesis, statistical studies of data, provided by operators from different lines in the

ECT department, have been conducted to analyze the variation in the results of all

measuring equipment as specified in ISO TS 16949 section 7.6.1.

Page 34: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

32

When using ECT equipment, its electrical characteristics shall be evaluated and used

under stable conditions. The characteristics and values are only valid for the unique

testing conditions where the instruction is applied as recommended in SS-EN ISO 15548-

1:2013 section 4.2. Therefore, as previously stated, the Results and settings stated are

unique to the product and environment it has been designed and tested for.

2.3.2 Standards for calibration

As specified in ASTM E-1212 section 9.2.2 Measuring and test equipment are calibrated

and controlled to specified requirements. A calibration system is established to ensure

that measuring and test equipment are maintained by periodic calibration against certified

equipment traceable to nationally or locally recognized standards and serviced so that

equipment function properly and are within their prescribed limits.

In this project, calibration is done using reference standard calibration samples with an

artificial EDM-made defect with a known depth. If a profile shaped wire is being

calibrated, two artificial defects are used, one on each side. An inspection system has to

be stated and it is important to make functional checks and verify the capability of the

inspection system and to provide calibrations curves as specified in SS-EN ISO

15549:2011 8.4.

It is necessary that all operators must be trained to handle ECT equipment during testing

and to perform the instructed calibration instructions.

There is a possibility to avoid calibration by using a “library”, with previous and “good”

calibrations, to reuse the settings. However, it has to be done carefully and “machine by

machine”, since no equipment is exactly the same [23].

2.3.3 Standard Operating Procedure (SOP) for calibration

SOPs are a set of instructions to complete tasks that are created for the equipment

operation, calibration and maintenance, to name a few. SOPs give a visual overview of

the calibration and how it should be performed correctly at the specified company or

under certain circumstances.

The focus for this research and project is to update the existing simple instruction for

calibration of ECT equipment and convert into a SOP-instruction, which describes the

process and give instructions to follow when performing calibration activities and what

values to use.

Page 35: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

33

In making changes to an existing SOP, any documents referenced or affected by SOP

should be considered including forms, templates, or other documents that are part of the

process related to the SOP being updated [18].

It is preferable to have a process map (or flowchart) of the tasks which gives an overview

of linked activities that must be managed to achieve the predetermined output. Focus is

to define its boundaries and make sure that the start and final points are clear, and it

should be easy to see if the action corresponding to the existing calibration is missing

[19].

The key steps of creating or updating SOPs through process mapping are described in the

following figure 18.

Figure 18: SOP development using process mapping

(i) Standard Operating Procedure (SOP) Contents

There are common SOP sections to be considered during SOP development. A typical

SOP should include at least the following sections [20]:

Purpose: a brief of statement of the intent of the SOP.

Scope: describes the boundaries for the applicability of the SOP, certain

circumstances, conditions, organizational units, and etcetera. It is important to set

the scope within the real control of the SOPs.

Policies: contains key rules and/or constraints that apply to the process.

Responsibilities: key responsibilities of the participants in the process.

Procedures: describe all activities to be done.

Not limited to the above sections, the following sections in the SOPs is helpful [20]:

Equipment and Materials

References

Definitions

Tools

Attachments/Appendices

Document History

Page 36: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

34

(ii) Standard Operating Procedure (SOP) Template

The SOP is standardized based on the needs of each organization. SOP template should

be easy to use. Consistent sentence flow does have benefits compared to the over-

standardizing sentences which may make them unmanageable and hard to understand.

Also, using some pictures and drawings is recommended.

Aspects of the SOP layout that need to be standardized are [20]:

SOP numbering

headers and footers

pagination

section headings

section numbering

test appearance (fonts, margins, etc.)

Page 37: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

35

3. Methodology

In this section the process methodology for the calibration of the circograph is described,

the used method and its purpose which helps to reach the goals for the project.

3.1 Quantitative and qualitative evaluation of the

calibration procedure

The present procedure of how calibration is being performed is a combination of

quantitative and qualitative evaluation. Both have to be considered important, but as can

be seen in the Results section, it is preferable to move from Qualitative to Quantitative

evaluation where there is a possibility and reason to do so. To have as much values as

possible is also important both for education of new employees, for revisions and audits

and for the possibility to standardize the production and for product safety.

3.1.1 Quantitative analysis

Example of quantitative evaluation

The depth of the artificial defect that the operator calibrates.

The height of the signal and the noise (S/N).

All values put in the system.

3.1.2 Qualitative analysis

Example of qualitative evaluation or semi-qualitative (meaning that today no values are

being extracted even if there might be a possibility to do so) evaluation:

How “stable” the signal is, e.g. if it has a lot of small peaks vibrating.

The shape of the curve or of the noise, see Figure 19. In the figure it is very

obvious that the left curve is distorted, since it has several peaks. But for an

unexperienced eye it can be difficult to determine which of the peaks the signal is,

and what peak(s) are noise peaks. To the right the signal is very stable and clean,

which is correct.

The difference between the legs holding up the crack peak.

The quality and life-length of the calibration sample, if it has a lot of dirt inside, if

the oxide layer is removed and so forth.

As well as many of the guidelines for calibration, which is mouth-to-mouth based

or experience-based.

Page 38: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

36

Figure 19: Example of qualitative evaluation, based on the judgement of the appearance of the curve, Insufficient shape

of curve (left, red)

3.2 Equipment used

A complete ECT line consists of the following order of equipment:

1. Incoming material WIP storage area

2. Pay-off (rotating, so the wire is winded of its load carrier)

3. Defectomat

4. Demagnetization

5. Circograph

6. Automatic color station for marking defected wire parts, above the specified limit

(40µm)

7. Cameras for checking the color marks

8. Rust protection apply

9. Pay-on (rotating, so the wire is winded on its delivery carrier)

10. Packing

Describing the circograph is the focus and will be presented below in next section.

3.2.1 Circograph and the procedure of calibration

Circograph sensor system Ro 20 P shown on figure 20, is the one of nondestructive EC

testing equipment manufactured by Foester GmbH. Ro 20 P scans the surface of round

materials (wires and profiles, typically egg-shape and elliptic), with scanning probes which

allow the maximum flaw resolution for surface-exposed, rotating systems that are used to

Page 39: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

37

detect longitudinal surface defects. Circograph sensor system Ro 20 P has a capacity to

cover the diameter range from 2mm to 20mm. The flaw detectability for bright material

surface is from flaw depth of approximately 30μm or more. The testing is today

performed from 40μm and above. The risk of testing for even less deep defects is that the

S/N-level becomes too low. Ro 20 P equipment uses testing speed of up to 3m/s and

continuous testing [12].

Figure 20: Circograph sensor system Ro 20 P with wire going through

The purpose of the calibration, which is being done for each shift of dimension, is to

adjust the settings to the specific properties of the wire. Simplified, the calibration

equipment is based on the following parts (see figure 21):

Page 40: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

38

Head for probes.

Probes rotating with over 14000 rpm.

Guides on both sides of the rotating head to keep the wire straight.

A calibration sample with an artificial EDM created mark of around 80µm depth.

The samples come from wire that the company has produced previously, but is not

the same wire running in the current machine and in many cases the steel grade is

unknown. The artificial mark of 80µm depth is made externally in a dual-source

arrangement from two different companies.

Demagnetization box.

Computer and screen.

Figure 21: Schematic simplified view of the calibration parts changed during calibration

The calibration is made on different views in the computer system, and the two most

important are the Y mode and the XY mode, see figure 22. There is also another mode

called the V mode, Vector mode, which can be used, see figure 44. The main

requirement, according to the knowhow of the observed company and their procedure (it

is not stated yet, before this thesis was made, in any official instruction) is that the crack

after the calibration should reach the trigger level of 80µm and the phase angle should

not be pointing directly straight up, but in the same time not lean more the +/- 10° from

the Y-axis on the positive side, see figure 23 and 24. Some operators refer to the clock

+/- 10 minutes, but that gives quite a big difference compared to degrees.

Figure 22: The screens for XY-mode (left) and Y-mode (right)

Page 41: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

39

All conductive materials have a so called natural noise, affected by previous mentioned

properties as conductivity, permeability but also surface roughness. The noise can if

handled incorrectly disturb the equipment and the difference between actual crack and

normal surface can be difficult to distinguish. This is in general referred to as the signal

to noise (S/N) ratio. To avoid this, the natural noise of the wire inserted should be planar

or close to planar, which also can be seen in the figures below (figure 23). In the

experience perceived in this thesis, everything that is differing of the natural noise, it

might either be a change in conductivity or a defect, the signal is pointing up, not

necessarily in the range +/- 10° from the Y-axis, but still up, which is a setting of the

system.

Figure 23: Crack point up and noise down, which gives the best S/N-ratio

To the help to reach this result, the operator has quite a lot of parameters to adjust:

Phase angle

Gain

Filter correction

Gain for second-probe

Bandwidth

Different thresholds or trigger levels

Depending on machine type, in some cases there are many more options. And then there

are “hidden” system views, where it is possible to change more in depth-settings, like

Page 42: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

40

rotating speed, testing frequency and other parameters. Usually, the operator should not

and does not change these settings, but the settings have been used in this thesis to test

different settings.

Figure 24: Visualizing of adjustment based by crack, noise, gain, and filter correction. The crack signal points up

between +/- 10°

Phase angle

Adjusting the direction of both noise and crack is crucial. The phase angle is the single

most important setting to get an optimal S/N-ratio, since it is possible to adjust the

direction of the noise. An increased phase angle value moves the signal and noise to the

left, and a decreased phase value moves the signal to the right, see figure 25.

Figure 25: Schematic adjustment of noise or crack with phase

Page 43: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

41

Gain

Since the difference of the noise direction and the signal direction is not 90° (which has

been optimal, meaning the noise is already planar and the crack is pointing up without

the need of additional gain), the focus in calibration is to adjust the noise to be planar by

adjusting the phase angle value. The crack will then follow after the phase adjustment

and pointing obliquely. By increasing or decreasing the gain the crack will, although it is

oblique, having the value of 80% which is 80 µm, which in this case is the desired

calibration value. Figure 26 shows how the noise and crack follow each other according

to the adjustment of the phase. In the example it is likely that the gain needs to be

lowered to make sure the signal does not get values over the trigger level. The trigger

level during calibration is the depth of the artificial crack. During production the trigger

level is set to the depth the product wire specification it should be tested against. There

is a possibility to have several trigger levels, as can be seen in several of the figures; an

example of this is figure 26.

Figure 26: The adjustment of phase and gain to get an approved result. Dashed lines are trigger levels.

Filter correction

Filtering of frequencies, where a negative value moves the filter towards lower frequencies,

meaning more low frequencies are passing through. The filter correction is moving both

the direction of the crack, as well as the shape of the legs. A crack closer to the Y-axis

has less need for gain to compensate for the oblique appearance. Positive values of filter

correction compress the graph and move the signal towards left. Negative values widens

the graph and moves the signal towards right, see figure 27.

Page 44: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

42

Figure 27: Filter correction. Positive values compress the graph and move the signal towards left. Negative values

widens the graph and moves the signal towards right

Before starting with calibration, it is important to change the precision guide according to

the wire dimension, for example to calibrate a wire of 2.9 mm a guide of 2.95 mm is

used. After fixing the guides, then the calibration of the wire can start. To get a correct

clearance/proximity, the probes are moved in direct contact with the wire, and then

removed a certain distance, which the computer then is compensating for. During this

process, it is also important to have clean equipment, a correctly made artificial defect,

probes that are not worn out and other parameters. The schematic view of the

changeable system is shown in figure 28.

The most difficult part for the operator is to find a suitable compensation between the

crack, the noise and the compensation the computer has to do between the highest and

the lowest point, see figure 28. The crack has to be of the same depth wherever on wire

it is found, even though there are different distances from probe to crack depending

which probe is closest to the crack as seen in figure 28. To achieve the same depth from

0-360°, as compensation is being made by system. The algorithm behind this

compensation is not known and not investigated in this research.

Figure 28: Different signal height depending on the distance from probes (0-360°)

Page 45: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

43

To eliminate this distance difference, the operator twists the wire around and tries to find

the biggest signal, reports that and then tries to find the smallest signal and report that.

The computer then defines the compensation. Example of how to find the highest point

(strongest signal or closest to the probe) is given in the figure 29. In this example, the

phase angle is around 216°. Correctly done, for round wire, the lowest point should be

opposite (180°) of the highest point, in this case 216°-180°=36°. The lowest point can be

seen already when pointing out the highest point.

Figure 29: Example of highest point and lowest point of the signal during calibration. The highest point is at 216° in

this example.

If the calibration is done correctly the defect should be compensated so that the height of

the defect in Y-mode is at the same level from 0 to 360°, or +/- 10μm since it can be

hard to achieve exactly the same height.

Figure 30: Confirm calibration when the values of the defect is the same during rotation from 0-360°

Page 46: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

44

As seen on the figure 30, the height is within +/- 10μm when the calibration sample is

rotating in different positions. When all the parameters are set, the operator can Click

OK, and line in the material in the machine and start running. The procedure of

calibration takes around 45min in average.

3.2.2 Defectomat

The encircling defectomat manufactured by Foester GmbH is one of nondestructive EC

testing types on the market, and is used for detecting transverse flaws and short flaws by

using an encircling coil. The test piece is then moved through or passed by with a

synchronous signal display. The systems are fully automated and defects can be marked,

classified and automatically discarded in exactly the same way as the circograph.

The defectomat has a test speed up to 150m/s [13] and continuous testing. The

maximum speed is never used in reality since the circograph is in the same line with a

maximum speed of 3m/s.

3.2.3 Contracer and other measuring equipment

A contour measuring machine called a contracer, manufactured by Mitutoyo, was initially

used in this project to measure the depth of artificial defect. During the measurement of

depth of the artificial defect, the wire to be examined should be well placed and it is

important to adjust the artificial defect of the wire according to the probe of the

contracer. Measured contour data can be compared with design data and shapes in terms

of actual data and shapes rather than just analysis of individual dimensions [15]. In this

project, the measured result of the artificial defect contour was not correlating to the

values certified by the supplier. A discussion with the company and indirectly with

Mitutoyo was held, since either the Mitutoyo equipment or the calibration samples were

incorrectly made. When measuring the crack in microscope by cutting it and placing it in

a mounted sample (see 3.2.5 Microscope and mounted samples), the values perceived of

the artificial defect depth were very close to the certificate values given by the

manufacturers of the calibration samples. Mitutoyo explained that this is probably due to

the measuring speed, according to the technician who held the contact. The lowest speed

is not enough to catch enough data point for such a small crack that is almost not visible

for the human eye. Due to time limit of this thesis and the cost correlated this could not

be arranged in this project. Still, the contracer works well for shape measurement

purposes. The contracer is not used for the calibration purpose in normal production.

Page 47: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

45

Figure 31: Contracer [16]

There are other devices used in this project. Figure 32 shows one device used to measure

resistance. For permeability measurement a device to measure magnetic field intensity (H)

was used as well as magnetic flux density (B) measured in mT units, see figure 33.

Figure 32: Resistance measurement in mΩ Figure 33: B-field hall-effect Meter instrument used to

measure magnetic flux density (B)

3.2.4 Calibration samples

Before inspecting the quality of the wire, calibration of the ECT equipment is performed

using calibration samples. The calibration samples are used as reference standards or

standardized samples to teach the machine what defects to look for. Calibration samples

used at the company observed are made by two external companies, and the delivery of

the sample is within 2 weeks from the date an order is placed. The wire is made by the

company, but as previously mentioned, it is not necessarily the same steel grade even

though the dimension is the same as the material to be tested. It is preferable the test

material and the calibration sample have the same material, alloy, shape (round or

profile), temperature, diameter etcetera. Location, size and depth of the artificial defects

made to the calibration sample must be well known before calibration.

Page 48: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

46

After adjusting the ECT equipment for a specified response based on the calibration

sample, a signal of approximately the same amplitude and phase should be obtained from

the condition of the wire with ECT equipment.

3.2.5 Microscope and mounted samples

Due to the size of defect, a microscope can be used to investigate the location and depth

of the defect. In order to view the depth of the defect in optical microscope, a sample or

part with a defect on must be dipped in Bakelite, and should be completely covered with

Bakelite and then mounted. After heating the Bakelite, the mounted sample is cooled.

Then the polishing machine using different polishing papers to make the surface more and

more smooth. Finally the sample is etched as shown on figure 34, and the depth can be

measured using optical microscope.

Figure 34: mounted sample

3.2.6 Generally about the methods for this thesis paper

The evaluation was mainly done through different methods:

Previous data found in the production system. This includes the steel composition

values and how many color marks per employee/machine type etcetera.

Unique production data was collected. This meant the operators wrote down their

production values on a special sheet created for this purpose. The total amount of

units (steel coils) with production data was over 1800 coils or over 2200 tons of

material. The data was then analyzed with pivot tables in Excel and Minitab.

Tests made by the thesis authors were made with a lot of calibrations done in the

circograph. Especially 2.95mm and 5.00mm were used, since they are “big

runners”. Also several other dimensions were used.

Regarding devices they are already described well. This includes measuring

magnetic flux density, resistivity, magnetic intensity, conductivity, contracer and

so forth.

The equipment hardware and software is made by Foerster GmbH and operating

on Windows platform.

Page 49: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

47

4. Results

In this chapter, the results will be presented that were obtained with methods described

Methodology and that are compiled, analyzed and compared with the existing knowledge

and theory presented in the Frame of reference.

A general SOP has been implemented and signed by the process manager and the quality

manager. The main purpose was to determine what parameters that influence the final

result, and to find out what parameters are used today by the operators, and what should

be recommended in the future. A summary of the majority of the parameters can be seen

in table 4.

Table 4: Values for calibration to be used in the future.

In total, unique values from the operators were collected from over 1800 coils (production

units), representing around 2200 tons of material. Added to that comes the tests made by

undersigned of the thesis report. The values are chosen to give stable and reliable values

for ECT in this specific environment and for valve spring wire.

4.1 Results analysis

The results have been shown, and in the following section several of the different values

will be further explained.

Page 50: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

48

4.1.1 Phase angle determination

At start, the operators thought the phases values were more or less random. By analyzing

the actual values of the operators used when they calibrated, typical phase angle values

could be determined. When analyzing the values, it was discovered that in fact two

machines were having phase angles 180° differently from the others. This is related to

signal output. The operators claim that the phase angle is consistent even if the

circograph or the probes were changed, but if the computer was changed, the phase could

shift. A change of probes was tested during the project and the phase angles were the

same as previously also after new probes were installed. However, the fact that the

different circographs have operated in such different values, the general opinion has been

among the operators that “everyone is doing different”. This is not entirely correct. This

thesis is showing that it’s is not random values, but there are two ECT-lines where the

output signal is opposite 180°, which is being visualized in the figure 35. There are still

large value intervals that are never used, and they should not, since the tests performed

show the S/N-ratio will be too low outside the phase value intervals suggested in this

thesis.

After evaluation, the interval is set to:

Option interval 1: 330°-+20° (most common).

Option interval 2: 150°-200°.

For Option interval 2, the curve of the crack would have been upside down if the phase

value interval from Option 1 was used. It is not examined what happens with the

reliability of the testing if an upside-down signal is used, but the supplier thinks the

testing reliability probably is unchanged. A decision was made to run production with two

intervals as presented above, so the curve appearance is unchanged.

Figure 35: Alternating current and pickup signal pickup 180° from each other, resulting in a upside down curve

Page 51: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

49

In fact, the 8 lines investigated have the follow spectra:

Line 1 Option interval 1: (330°-20°)

Line 2 Option interval 1: (330°-20°)

Line 3 Option interval 2: (150°-200°)

Line 4 Option interval 1: (330°-20°)

Line 5 Option interval 1: (330°-20°)

Line 6 Option interval 1: (330°-20°)

Line 7 Option interval 2: (150°-200°)

Line 8 Option interval 1: (330°-20°)

This phase value interval is covering the absolute majority of all values observed, but the

interval gives an opportunity to discover values that might mean that something is wrong

with the equipment, the calibration sample or the knowledge of the operator. An example

of this could be that an incorrect frequency is used, which would completely change the

value intervals. When the values now are standardized, the intervals can probably be

tightened more in the future, but the risk is that there will be production downtime when

production cannot run because one or more values are outside the tolerated intervals.

There are obviously many parameters to use which influence each other. That might

make it more difficult to fit the received values into a tight specification of useable,

approved values.

There is a correlation between the alloy content in the material and the phase angle as

shown in figure 35 and was discussed in “Frame of reference”. The ultra-high tensile steels

(OT90/91/101), the UHT steels, have decreased conductivity (see figure 46), increased

resistivity and a changed impedance point causing a changed impedance phase angle (see

figure 37). The difference is still quite small, with around 8°, which does not have a

significant different effect on the calibration result, see figure 36. To compensate for the

reduction of impedance, the operators compensate by changing the gain or lifting up the

curve with the filter correction. The recommendation is to stop using a random steel

grade during the calibration. Instead UHT material should be calibrated with UHT

calibration samples. Figure 36 is also including the standard deviation, showing the

difference for at least ultra-high tensile steels like OT101, OT91 and OT90 compared to

the steel grades with lower alloy (carbon) content.

Page 52: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

50

Figure 36: Phase comparison with standard deviation, values over 360° means that the value in machine is slightly over

0°. Values come from collected production data.

Figure 37: The impedance angle, the difference between noise and crack, is slightly smaller for the UHT-steels. The

angle for OT70 (α) is larger than for OT91 (β)

The defectomat is not investigated in this report, but since it is reacting heavily on

hardening structure errors, see figure 8, it is likely that the change in conductivity has a

larger impact in the defectomat than in the circograph, which is also seen in the figure

52, where the color marks are increasing with increasing alloy content.

Page 53: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

51

4.1.2 Gain

Gain is an important parameter, and is used to decrease or increase a signal. The

operator has two-three gain parameters to calibrate, but the focus has been on the main

gain, which is giving the highest values, and is based on probe number one. The second

gain values are for probe number two, and is usually following probe number one but

have lower values.

Figure 38: Gain and Phase angle comparison for 5,00 mm OT70 wire. Filter correction is kept at 0

All dimensions, grades and machines have an optimal value range. In figure 38, the

optimal gain and phase is shown for 5.00 mm. Optimal in this case means as low gain as

possible, since the gain is not only increasing the crack but also the noise. For the given

example, as well as the majority of the wire, a phase angle around 340 to 370° (+10) also

gives the lowest values of gain, but it can be observed that the curve is quite flat on the

bottom. This is correlated to that the gain is still minimized if the phase is moved and

adjusted on the other side of the Y-axis, which can be seen in figure 39.

Page 54: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

52

Figure 39: Different crack curves. A curve with a phase angle far from the trigger level has to use more gain, since it

has a longer distance from the trigger calibration level.

It is always possible to lift up a crack signal by increasing the gain. It is hard to define

when the limit for the gain is. The system investigated has a gain upper limit of 63.4dB.

This means that theoretically it is possible to lay the crack phase down as oblique as it is

still possible to reach the trigger level with a total gain of 63.4dB. “Possible” does not

mean “recommended”. Using over 60dB will give a much distorted signal.

The operators tell, regarding the direction of the crack, that a crack direction from -10 to

+ 10° from the Y-axis is acceptable. Some say “ten minutes to or over 12 o’clock”. None

of them is completely wrong or right. However, since the signal vs noise often is around

70°-90°from each other, it is quite natural that the crack is standing up when the noise is

lying down. This also means that less gain needed, and to use as less gain as possible to

reach the trigger line is preferred. Figure 40 is showing when the “clock”-rule is used. The

gain is in this case increased, the phase moved 45° and the noise is almost -45° from the

Y-axis line. Meaning: 10 to or 10 over, if referring to the clock. This setting makes it

more difficult for the machine to get the right settings, since the noise is standing up and

consequently the S/N-ratio will be insufficient.

Page 55: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

53

Figure 40: Phase is changed so it is laid down 45°. This is lifting up the noise direction and gives a disturbed signal.

It is recommended that degrees should be used, and not the clock. This gives a

requirement that the phase should differ from +10° to -10° from the Y axis. It is not

extremely important to keep this interval values, but using the clock as some operators

refer to leads to a more unstable curve so too much gain is used.

Figure 41: Showing correct interpretation of a spectrum giving the lowest gain and the lowest noise

A factor that has high correlation with the gain is the lift off, or clearance as it is called

by the program made by Foerster GmbH. A larger distance from the probe to the surface

requires more gain to compensate for a weakening signal, see figure 42. This is also

increasing the noise level. However, it is hard to know what the optimal distance is. The

risk is that if the wire and the probe are too close, so called kinks (knots in the wire)

touches the probe giving a high signal or in worst case break the probes completely,

which is creating a high cost when replacing. But a high gain value can indicate that the

probes are not mounted correctly so the lift off distance is larger than suitable. A possible

recommendation is that the thinnest dimension, from 2-3 mm, could have a decreased

Page 56: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

54

clearance, especially since the operators quite often complain the finer sizes are harder to

calibrate than the thicker dimensions.

Figure 42: Increase of gain necessary when lift off or clearance is increased

Worth mentioning, the clearance was not affecting the phase angle according to the tests

performed in this thesis. Regarding gain for different frequencies, this is introduced in the

next section.

4.1.3 Filter correction

Filter correction, to tune in and out different frequencies, is mainly used to correct the

legs during calibration and to make them “straight”. Filter correction also has the ability

to lift the crack signal, see Methodology part and especially figure 27. The main part of

all operators use the filter correction -1, which means the filter accepts more signals with

less frequency. See figure 43. Using -1, and negative values in general, stabilizes the curve

which is usually good for the testing. According to our experience the value of 0 is usually

as good as -1 and could be used more often. Positive values increase the signal to noise

ratio. A value of +3, which has never been used by the operators during this work, is

actually giving the best S/N-values, when used in calibration vector mode. But, the signal

is then very unstable, which can be seen on all the peaks seen in the vector mode (see

figure 44). It is almost impossible to run on +3 in real production.

Also, noise in S/N calculation is based on the vector mode. But considering the

latitudinal noise, which was introduced in Methodology, it is clearly seen in the figure that

Page 57: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

55

the latitudinal noise level increases (which is negative for testing) when using a filter

correction of +2 and above.

Figure 43: Pie chart of the values used in production. -1 is the most common value used.

Using -1 or -2 is giving the possibility to decrease the gain and get straight legs. However,

at the same time the latitudinal noise is increasing. The majority of the employees put

the phase down and does not care about the signal being very unstable. By using value of

0 or +1, it can be estimated that the total S/N-level reaches the lowest point, and this

thesis claims that seeking the optimal point is decreasing the risk for unwanted false

alarms. -2 could increase the straightness of the legs, but the S/N ratio should according

to this thesis be more important than the signal appearance. In fact, the collected data

from the operators also shows the operators have the same view.

Figure 44: Filter correction +3 has the lowest S/N-ratio, but has a high latitudinal noise and the very unstable. The

two small figures to the right show filter correction value 0 and +3. +3 has best S/N ratio has, but the noise is full of

sharp peaks.

Page 58: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

56

A lot of tests have been made on how different frequencies are correlated to filter

correction settings. No big changes, at least not significant, have been observed. Some

observations can be stated:

The higher frequency used, the filter correction tends to be more negative, since

the higher frequency means a more unstable signal. A value of 1000 kHz requires

more stabilization than 100 kHz, therefore negative values are more relevant to

use. This can also be seen in the figure 45 below. For 1000 kHz, negative filtering

values are recommended. If positive filtering values are used, the gain rises rapidly.

The gain for 300 kHz and 1000 kHz is following each other on the negative

filtering values. However, when using positive filtering values, the green curve

representing the 1000 kHz needs more and more gain compared to 300 kHz and

100 kHz.

The gain for 100 kHz has consequently higher gain than the other two frequencies

when using negative filtering values, meaning that it is not optimal to use for this

material. Even though the penetration depth is larger, which could stabilize the

testing, but increased gain is not preferable.

Finally, it is important to conclude that the gain is depending on the filter

correction, see figure 45. This is true on all frequencies and the most obvious

observation.

Figure 45: Gain is correlating heavily on the filter correction. Values average based on dimension 2,6 mm, 2,95 mm,

2,90 mm, 3,75 mm, 3,95 mm, 5,00 mm and 5,50 mm.

Page 59: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

57

4.1.4 Conductivity and permeability measurement

In this subsection, analyses of different factors that can affect conductivity and

permeability have been made.

(i) Values in production

The equation (3) that was used to calculate the conductivity was:

The results of these calculations are being shown in figure 46 and table 5. The results

follow well the values received in the “Frame of Reference” - section when annealed AISI-

steels were compared regarding conductivity. When the carbon content increases, the

conductivity decreases. See figure 46, which is based on over 30 measured values. Since

resistivity is the reverse of conductivity as given in the equation (2),

resistivity increases with UHT-steel.

Figure 46: The different conductivity values for selected materials. Carbon content is increasing from the left to the

right on steel grade

The average values of conductivity of different steel grade are also shown in the following

table:

Page 60: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

58

Table 5: Steel grades and their conductivity value in oil tempered structure

Permeability values are more difficult to achieve, especially since magnetic properties are

variating a lot depending on where (and how) the wire is cut and where the measuring

probe is placed. As explained in Frame of Reference, it is necessary to calculate the

magnetic flux density (B) and the magnetic intensity (H) of the wire, to receive the

permeability. This has been done and the values are as follows:

Figure 47 shows the average calculating of the values got from measurement of

Magnetic flux density (B) = 1.1mT

And figure 48 shows the average calculation of values got from measurement of

the Magnetic intensity (H) = 6.40 A/cm.

Therefore, Magnetic permeability = 1.7*10-6Wb/Am (valid for all steel grades of oil-

tempered wire produced in Garphyttan). The values are not split among the different

steel grades, since the unsatisfactory variation of the values may lead to unwanted

conclusions. Totally around 40 samples measured in two ends, and the highest value of

each sample represents the wire result. An enhanced study of more values and during

longer time period may lead to a differentiation of the steel grades based on the

permeability. Instead, an average is being presented, which can be seen in the figures

showing the standard deviation for B and H. The permeability value, 1.7*10-6Wb/Am, is

used for the penetration depth calculations.

Page 61: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

59

Figure 47: Magnetic flux density on oil tempered hardened samples (mixed steel grades)

Figure 48: Magnetic density on oil tempered hardened samples (mixed steel grades)

The penetration depth for different steel grades is presented in Table 6. The equation

used is:

Page 62: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

60

where the permeability is kept constant (1.7*10-6 Wb/Am). In reality, this is not correct,

but could serve as an understanding of the penetration depth. The conclusion is:

The penetration depth is around 900μm for 100 kHz, 500μm for 300 kHz and

300μm for 1000 kHz.

UHT-steel should in general have a bigger penetration depth.

Table 6: The penetration depth measured in μm for different steel grades and frequencies. The permeability is

considered constant for all grades

Wire rod cracks have been found the latest years in production with a maximum level of

around 500μm. The frequency 300 kHz is in general sufficient for this material. To go

deeper in the material will make the ECT less sensitive for false alarms, but the phase

angle difference between noise and crack will shrink and the S/N ratio will decrease,

causing more color marks. The tradeoff for less sensitive testing is a higher S/N ratio,

and this is not worth the change (for this type of material). 300 kHz is probably not the

most efficient frequency, but it is close. To get the perfect frequency, it is necessary to do

the evaluation made in the figure presented in Frame of Reference, see figure 11, and to

do it for every steel grade.

4.1.5 Frequency and phase

Throughout the thesis, the frequency has been a central part of the research. It has been

concluded:

The penetration depth is affected by the frequency.

Trends have been noticed that higher frequency tends to increase the need for a

negative filter. Also, gain is higher for 100 kHz for negative filter values.

Page 63: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

61

The largest difference in daily work for the operator is the correlation between phase

angles and the frequency, if the frequency can be changed. All other recommended values

presented would be within the current intervals if the frequency is changed, but not the

phase interval. This can be visualized in figure 49 and figure 50. Figure 49 show typical

values for a random diameter examined which a values around:

300°for 100 kHz

335° for 300 kHz

60° (360+60=420° adjusted value)

This is a typical example; the real interval needs to be set through a lot of collected

values. Some dimensions have been investigated, which can be seen in figure 50. The

general conclusion is that the larger the frequency from 100-1000 kHz is, the more

increased phase angle values are obtained. In figure 49 the adjusted value is shown, in

reality it is 60° (360+60=420° adjusted value), the adjusted value makes it to easier show

the trend within a diagram and not in a unit circle.

Figure 49: Phase correlation with frequency for 100 kHz, 300 kHz and 1000 kHz.

Discussion has been held about the impact on dimensions regarding the phase. As can be

seen in figure 50, it has been difficult to prove or disprove. No significant difference can

be seen regarding this when changing the dimension on the same frequency.

Page 64: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

62

Figure 50: Phase angle for planar noise of different dimensions of wire used for frequency 100 kHz, 300 kHz and 1000

kHz

The conclusion is that all phase angle values presented in the SOP instruction have to be

revised if the phase frequency changes. If increased to 1000 kHz, the range will be around

+60° according to observations.

4.2 Other observations

During the work several notes has been taken about other parameters affecting the

reliability of the testing. Some are explained, some of the notes are in the SOP, and some

needs more evaluation. The subjects presented here are:

How the amount of color marks seemed to increase with the steel grade and an

increased alloy amount.

“Defects” consisting of residues of a material with a completely different

conductivity, in the case lead (Pb).

False calibration on lead mark.

Dirt, oxide remains and oil covering the artificial defect

Page 65: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

63

4.2.1 How the amount of color marks seemed to trend with increased

alloy content

Initial information achieved pointed in a direction that an increasing alloy content created

more color marks (more defects), see figure 51. A possible reason for this trend is that

UHT materials are more difficult to process, and therefore more damages appear and

more defects are found. But when evaluating the real defects, there could not be

confirmed that the defects found in for example an UHT high-alloyed steel like OT90,

was worse (deeper depth and more rough surface) than for defects found in a steel grade

like low-alloyed OT70. Still, the color marks per weight or length increased, both in

defectomat (encircled testing) and circograph (rotating testing), by increased alloy

content. This can be seen in figure 51.

Figure 51: CM (color marks) is increasing with increasing alloy content, by meter units.

By analyzing the figures, the increasing trend due to increasing alloy content is only

significantly relevant for the defectomat. The increased amount of defects was increased

for all dimensions and shapes in the defectomat. The circograph, which is the focus in

this report, is not affected by the steel grade to the same extent. This is due to that the

phase is laid down as planar as possible by the operator, which is being described

previously in this report. Therefore, the increase of amount of color marks in circograph is

not related to the grade, but the fact that an increased part of the UHT material is based

on profile shape. This can be very well seen in figure 52, the left figure. The values for

different steel grades are consistent regarding color marks, but are separated by shape, if

it is round or profile. Profile wire has almost twice as many color marks compared to

Page 66: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

64

round, indicating the compensation of the dimension ratio is not working optimal. It is

not known if and how much this can be improved.

Figure 52: Round and profile, color marked defects in circograph and defectomat.

Figure 52 shows that by separating the round and profile, there is no longer an increase in

CM in circograph by steel grade. As results from this analysis, the increased amount of

CM in circograph is because of UHT-steels (OT90,91,101) mainly consists of profiles,

which is more difficult to calibrate than round. This is correlating with the results shown

in this report, that there is a slight change in conductivity, affecting the defectomat more

than the circograph.

4.2.2. Defect of different conductivity (lead, Pb)

An issue in ECT for wire that has been annealed in lead (see figure 53 and 54), is the

residues that can be found on the wire occasionally, often resulting in a reaction in the

circograph, misinterpreting the lead for a defect, since the magnetic field is changed.

Since it is possible to sort out certain frequencies and/or phases, it is tempting to sort

out the signal from lead, if it differs from other mechanical defects or scratches. This is

possible to do in the equipment. However, as can be seen on the figure 53, the signal that

comes from lead marks is actually in the similar phase as normal defects, except that the

form is straighter with no clear “legs” visible.

Page 67: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

65

Figure 53: Lead signal in XY-mode noticed in production. Note the similarity between normal crack’s appearance

Figure 54 is showing how a lead (Pb) mark can look like. This is enough for the

circograph to give a false alarm.

Figure 54: A lead mark sticks to the steel wire

4.2.3 Artificial defect reliability

Tests have been made with different types of artificially made defects, both made by saw

and by EDM. An example of this comparison can be seen in figure 55. In general, the

defect types look very much the same. An issue was that the Contracer, as described in

3.2.3., did not show reliable values, and due to cost limitations, the testing of this did not

continue. The standard in the SOP is still the EDM marked artificial defects.

Page 68: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

66

Figure 55 EDM artificial defect compared to saw manufactured.

(i) Having the right defect on the right place

According to Foerster GmbH, the supplier of the testing equipment, the defect should be

placed on the longest and the shortest distance from the probes. For round wire the

clearance or proximity is the same wherever the artificial defect is placed. For profile

shaped wires, which have two calibration marks instead of one, the two defects should be

set in a certain angle, to make sure the longest and the shortest distance from the probes

are achieved.

Figure 56: Placement of artificial defects on Oval and Profiles wires. Picture from Foerster and from Garphyttan wire

(right)

However, in this thesis, the angle is always 90° between the two defects for profiles,

which is some cases, might not be correct. But the egg shaped is usually designed to have

the shortest and longest position very close to the nominal values, see Figure 56. It has

not been proved any bigger difference with this wrong placement of the defect. It is easier

for the manufacturer of the calibration samples to place the defect 90° from each other,

as shown in figure 57.

Page 69: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

67

Figure 57 Egg- shaped wire an incorrect placement of an artificial crack.

A more correct placement would be as in the figure 56 (to the right), where the angle

made in green would represent where the defect for this specific profile would be placed

on the flat side. It could not be stated through research how much this disorientation of

the defect actually influenced noise levels and compensation.

4.2.4. Calibration with lead (Pb) - disturbance or directly from lead

This thesis work shows that sometimes there are lead residues also on top of the

calibration samples, both in direct connection to the artificial defect, or on the other side,

or nearby the crack. This increases heavily the risk that the operator calibrates with lead

disturbance or completely on a lead defect. This is something the thesis work suggests

the company to examine before using the sample in production, and this is instructed in

the SOP.

Page 70: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

68

Figure 58: Lead is present on a calibration sample; it is direct connection to the artificial crack (to the right)

Figure 58 shows how it can look like when calibrating on lead and not on the defect.

Typically, the values suggested by the machine are strange, for example a gain of 57dB

(that is why it is so important with value intervals to detect easily when the calibration

probably is not correctly executed). However, still, it can be quite similar in shape

compared to a calibration on an artificial defect, see figure 59. The stability of noise is

very poor when calibrating based on a Pb mark.

Figure 59: A calibration made on the calibration sample with a lead (upper). When putting in the real defect (bottom),

supposed to be 80µm, it shows much higher values.

Page 71: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

69

As mentioned, although the lead-based calibration is very unstable, the signal is quite

similar to a calibration from a conventional artificial crack. However, when calibrating

first on lead, then putting in the real crack, the picture and noise level is extremely high,

as seen in the figure. This will be the same in actual production, the number of “defects”

will be very high, since the testing will be very sensitive, see figure 59 for an explanation.

4.2.5. Dirt, oxide remains and oil covering the artificial defect

In several cases, when the calibration samples were investigated, remains of oil, dirt or

oxide were covering the artificial defect, see figure 60. It is of high importance for the

operator to make sure no lead, dirt or anything else is covering the defect or the probes.

Figure 60: Two examples when the depth of the calibration sample is covered by dirt

Before calibration process begins, it is necessary to clean the calibration sample especially

in the artificial defect or in the calibration mark. Figure 61 shows one calibration sample

with the calibration mark from figure 60 when it has been cleaned.

Figure 61: Cleaned artificial defect on the calibration sample

Page 72: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

70

4.3 Definition of improvement possibilities

By taking out defects that have been previously considered over 40µm by the ECT and

measure their real values, a potential of improvement can be estimated. An example of

this is a defect seen in figure 61. When the defect is cut into two pieces and checked in a

microscope, the defect can be measured. In the example in figure 61, at least 53% of the

defect depth is actually not a real depth. The example is calculated as follows:

Depth of the defect was 18.49µm.

Nominal value is 40.00µm (at least)

“Potential” = 40.00-18.49 = 21.11 µm.

21.11/40.00= 53%

In many cases, when color marks of so called defects were examined, no defects could be

found at all on the surface. This means it is likely some of the defects are only false alarm

defects.

Figure 62: Depth of the defect was 18.49µm. Potential= 40µm-18.49=21.11 µm

Since cutting defects and check the real depth in a microscope is a very time consuming

measurement method, and no 3D microscope was available, another way to evaluate the

variations in the process is by checking the amount of color marks per employee over a

long time. Or in other words, how many times the equipment judges the defects to be

over the trigger level 40µm. The result shows a big spread in values from the “best”

operator (or the machine she/he works on) with the lowest amount of color marks, where

the best has less than half of the color marks compared to the operator with the “worst”

Page 73: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

71

number has, see table 7. Still, the almost the same wires are being processed in the

machines, so the outcome should be similar.

Table 7: Big spread of the amount of color marks (defects) detected by different operators during 1 year and 4 months

in the circograph. This indicates larges variance and an unstable process.

The potential with the standardization is that these the color marks per meter per

employee should show less variance in the future, resulting in a more correctly performed

ECT.

4.4 Result of SOP instruction for operator

The main thesis aim was to create an instruction which is easy for all operators to read,

understand and apply. The instruction is a summary of all the parameters investigated in

this report, which is important to apply to get an accurate result. The instruction gives an

overview of calibration activities to follow when calibrating the ECT equipment

circograph. This SOP instruction is designed according to ISO standards and best known

praxis, as shown in the first page of this SOP instructions. Previous research on the

SOPs for calibration has been used, as shown in the chapter “Frame of reference”. The

following figure, see figure 63, shows the first page of the SOP instructions done in this

project. The rest is belonging to the company and is a part of SG know-how.

Page 74: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

72

Figure 63: Part of the confidential SOP instruction for operators working with ECT.

Page 75: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

73

5. Conclusions

A discussion of the results, conclusions and future work is presented in this chapter. The

conclusions are based on the analysis with the intention to answer the formulation of the

thesis questions that were presented in the first chapter Introduction.

5.1 Conclusions

Several findings have been established in this thesis work. Suitable settings of phase, gain

and filter correction has been established. The values were obtained from production

data, from around 1800 “coils” or production units, whereby the operators filled in new,

unique information about these values, which had never been collected before. The values

have then been confirmed by tests and analysis done, and regarding the big amount of

data, the conclusions could be made with larger certainty.

Also, scientific articles have confirmed the general trends and conclusions, even if there is

a slight difference in the materials from the wire studies in this thesis work compared to

the research articles. ISO-standards of calibration have been referred to, to make sure the

instructions on how to make standard operating calibration by the operators is correctly

done.

Some of the conclusions confirmed through this thesis are:

There are certain phase angles intervals that are suitable for testing. Since rotating

testing is made by alternate currents, the signal can sometimes be picked up 180°

from the normal interval range, according to the sinus curve appearance. This

creates a primary and secondary phase angle interval, with the difference of 180°.

The gain is adjusted according to how much the signal must be lifted up. To use

low gain is preferable.

Filter correction is filtering the frequency, and is increasing or decreasing the need

to use gain. Filter correction is also increasing or decreasing the S/N ratio, and

make the curve more or less stable. Negative filter values give a more stable curve

but decreased S/N ratio.

Frequency is the main contributor to set the phase angle interval. Three different

frequencies have been examined, where the middle alternative of 300 kHz shows

the best properties.

The penetration depth has been calculated.

Conductivity and permeability values have been obtained.

Page 76: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

74

Steel with higher carbon content has higher resistivity, lower conductivity and a

change of the impedance point occurs. This means the phase angles increase for

UHT steels.

It is important to check so no lead (Pb) or dirt is present during testing.

S/N ratio has been discussed and how to set phase, gain and filter correction to

get the best values.

Putting the noise phase planar is of major importance to get a sufficient S/N ratio

(3:1). Therefore, it is important to find the biggest angle difference between the

noise and the crack, so the crack stands up and the noise is longitudinal. This can

be done by trying more frequencies.

Instructions are important to avoid running on values that differs from normal

production, and an SOP makes it easier to follow for engineers, operators and

technicians. By having value intervals, abnormalities will easier be detected and

eliminated.

The purpose of this thesis work has been fulfilled:

1) Updated instructions used by the operators

2) It has been explained how different calibration settings affect the testing reliability

3) Filled the knowledge gap of used intervals for different parameters as well as

material properties and its impact on the testing.

5.2 Discussions

There are several notes to be made during this thesis work. Certain dimension of high

runner has been chosen. However, there are a lot of individual parameters to set, and the

equipment (circograph) has its own algorithms to calculate compensation, as well as the

incoming material has slightly different properties and the calibration artificial defects

may also have a variation in their shape. Therefore, the results may differ slightly when

repeated, but the general trends are the same.

The SOPs instruction for operator consists of how to calibrate ECT equipment

(circograph), and will not only help the operators in daily basis but will be used for new

operators as introduction training, which will save time and cost for new operators.

By implementing this SOPs instruction, calibration setting parameters will be used by the

employees. The objective is to decrease the variation between machines and operators,

which hopefully will decrease the over-scrapping of material. It will also ensure customers

that the same testing is performed every time.

Page 77: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

75

From practical tests done during this project and also by using several research papers

and ISO standards, the contribution of this project on the overall global knowledge of

ECT consists of gathering together all required quantitative values used in calibration of

ECT equipment used in this specific environment. It is based on a combination of real

production data from inline ECT, as well as tests performed when the machine is not

moving, which makes the results reliable.

5.3 Future work

It was not easy to get the permeability values throughout this thesis. As said in the frame

of reference part, the values of magnetic field intensity (H) measured in A/m, have been

measured quarterly within the company to control that the de-magnetization works and

to secure the magnetic properties are not too strong.

To get values of magnetic flux density (B) measured in Tesla (T) or Weber per square

meter (Wb/m2) because of limited time and unavailability of instrument, an instrument

called B-field hall-effect Meter was used. Since the variation was quite big, an average of

all steel grades was applied. If more values were obtained, individual values of

permeability for each steel grade could be obtained. A future study could increase the

sample size and the reliability of the equipment to find differences between steel grades

and how magnetic permeability of test material affects Eddy current testing.

A key for better reliability in testing is to find the optimal testing frequency, where the

angle between signal and noise is as close to 90° as possible. It was only possible to

measure three frequencies in this report due to machine settings. This study can be

extended in the future.

Also, the finding that the profile shape is heavily affecting the output for defects could be

of further interest, as well as the placement of a artificial defect.

Finally, the expectation is that the SOP instruction will be a living document where

technicians, operators and engineers together find the best praxis continuously. Although

eddy current is well understood and evaluated concept, every situation is unique, and

small changes may lead to a revise in the standards.

Page 78: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

76

List of Tables TABLE 1: DIFFERENT NDT METHOD USED IN INDUSTRY [1] ................................................................................... 18 TABLE 2: DIFFERENT AISI-STEEL AND THEIR ALLOY CONTENT AND CONDUCTIVITY VALUES [11] ................................ 22 TABLE 3. AVERAGE ALLOY VALUES OF DIFFERENT GRADE OF SUZUKI GARPHYTTAN WIRE .......................................... 23 TABLE 4: VALUES FOR CALIBRATION TO BE USED IN THE FUTURE. ........................................................................... 47 TABLE 5: STEEL GRADES AND THEIR CONDUCTIVITY VALUE IN OIL TEMPERED STRUCTURE ......................................... 58 TABLE 6: THE PENETRATION DEPTH MEASURED IN ΜM FOR DIFFERENT STEEL GRADES AND FREQUENCIES. THE

PERMEABILITY IS CONSIDERED CONSTANT FOR ALL GRADES ........................................................................... 60 TABLE 7: BIG SPREAD OF THE AMOUNT OF COLOR MARKS (DEFECTS) DETECTED BY DIFFERENT OPERATORS DURING 1

YEAR AND 4 MONTHS IN THE CIRCOGRAPH. THIS INDICATES LARGES VARIANCE AND AN UNSTABLE PROCESS. ....... 71

List of Figures FIGURE 1: THE PRODUCTION STEPS AND A SCHEMATIC SIMPLIFIED VIEW OF STRUCTURE, GRAIN SIZE, OXIDE LAYER AND

DIMENSION .............................................................................................................................................. 12 FIGURE 2:HOW THE WIRE ROD LOOKS AFTER SHAVING ........................................................................................... 13 FIGURE 3: WIREDRAWING DIE .............................................................................................................................. 14 FIGURE 4: HARDENING PROCEDURE. .................................................................................................................... 14 FIGURE 5: WIRE ROD CRACK, SURFACE PICTURE AND ITS CORRESPONDING CRACK APPEARANCE IN A CUT .................... 15 FIGURE 6: DEPTH OF THE DEFECT ....................................................................................................................... 16 FIGURE 7: GENERATING EDDY CURRENT ON THE TEST PIECE [25] ........................................................................... 20 FIGURE 8: HARDENING ERROR (PERLITE STRUCTURE). A SMALL DETECTION SIGNAL CAN BE SEEN IN THE CIRCOGRAPH

(ROTATING PROBE TESTING), BUT AT LARGE SIGNAL CAN BE SEEN IN THE DEFECTOMAT ................................... 21 FIGURE 9: THE CONDUCTIVITY DECREASES WITH INCREASING CARBON CONTENT FOR SOME RANDOM CHOSEN ANNEALED

AISI-STEELS ............................................................................................................................................ 22 FIGURE 10: CHEMICAL COMPOSITION ANALYSIS REGARDING AVERAGE OF CARBON (%) .............................................. 23 FIGURE 11: DIFFERENCE BETWEEN NOISE AND DEFECT SIGNAL [9] ........................................................................... 26 FIGURE 12: EXAMPLE OF THE DIFFERENCE BETWEEN THE NOISE AND THE CRACK FOR 300 KHZ, AROUND 70°-90° IN THIS

CASE ....................................................................................................................................................... 27 FIGURE 13:RELATIVE EFFECT OF FREQUENCY, CONDUCTIVITY AND PERMEABILITY ON THE DEPTH OF PENETRATION FOR A

TYPICALLY SINGLE-COIL ECT PROBE........................................................................................................... 28 FIGURE 14: RELATIVE EFFECT OF FREQUENCY, CONDUCTIVITY AND PERMEABILITY ON THE DEPTH OF PENETRATION FOR

A TYPICALLY SINGLE-COIL ECT PROBE ........................................................................................................ 29 FIGURE 15: S/N BETWEEN THEN TOP (CRACK SIGNAL) AND THE SIDES (SURFACE ROUGHNESS NOISE) ......................... 30 FIGURE 16: NORMAL NOISE (HERE CALLED LONGITUDINAL) AND LATITUDINAL NOISE ................................................. 30 FIGURE 17: FURTHER DEVELOPING THE LATITUDINAL NOISE IN YELLOW (SHORT) AND PURPLE (LONG) ........................ 31 FIGURE 18: SOP DEVELOPMENT USING PROCESS MAPPING ..................................................................................... 33 FIGURE 19: EXAMPLE OF QUALITATIVE EVALUATION, BASED ON THE JUDGEMENT OF THE APPEARANCE OF THE CURVE,

INSUFFICIENT SHAPE OF CURVE (LEFT, RED) ................................................................................................ 36 FIGURE 20: CIRCOGRAPH SENSOR SYSTEM RO 20 P WITH WIRE GOING THROUGH ...................................................... 37 FIGURE 21: SCHEMATIC SIMPLIFIED VIEW OF THE CALIBRATION PARTS CHANGED DURING CALIBRATION ........................ 38 FIGURE 22: THE SCREENS FOR XY-MODE (LEFT) AND Y-MODE (RIGHT) .................................................................. 38 FIGURE 23: CRACK POINT UP AND NOISE DOWN, WHICH GIVES THE BEST S/N-RATIO ................................................ 39 FIGURE 24: VISUALIZING OF ADJUSTMENT BASED BY CRACK, NOISE, GAIN, AND FILTER CORRECTION. THE CRACK SIGNAL

POINTS UP BETWEEN +/- 10° .................................................................................................................... 40 FIGURE 25: SCHEMATIC ADJUSTMENT OF NOISE OR CRACK WITH PHASE ................................................................... 40 FIGURE 26: THE ADJUSTMENT OF PHASE AND GAIN TO GET AN APPROVED RESULT. DASHED LINES ARE TRIGGER LEVELS.

.............................................................................................................................................................. 41 FIGURE 27: FILTER CORRECTION. POSITIVE VALUES COMPRESS THE GRAPH AND MOVE THE SIGNAL TOWARDS LEFT.

NEGATIVE VALUES WIDENS THE GRAPH AND MOVES THE SIGNAL TOWARDS RIGHT ............................................ 42 FIGURE 28: DIFFERENT SIGNAL HEIGHT DEPENDING ON THE DISTANCE FROM PROBES (0-360°) ................................... 42

Page 79: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

77

FIGURE 29: EXAMPLE OF HIGHEST POINT AND LOWEST POINT OF THE SIGNAL DURING CALIBRATION. THE HIGHEST POINT

IS AT 216° IN THIS EXAMPLE. ..................................................................................................................... 43 FIGURE 30: CONFIRM CALIBRATION WHEN THE VALUES OF THE DEFECT IS THE SAME DURING ROTATION FROM 0-360° .. 43 FIGURE 31: CONTRACER [16] .............................................................................................................................. 45 FIGURE 32: RESISTANCE MEASUREMENT IN MΩ FIGURE 33: B-FIELD HALL-EFFECT METER INSTRUMENT USED TO

MEASURE MAGNETIC FLUX DENSITY (B) 45 FIGURE 34: MOUNTED SAMPLE ............................................................................................................................ 46 FIGURE 35: ALTERNATING CURRENT AND PICKUP SIGNAL PICKUP 180° FROM EACH OTHER, RESULTING IN A UPSIDE DOWN

CURVE .................................................................................................................................................... 48 FIGURE 36: PHASE COMPARISON WITH STANDARD DEVIATION, VALUES OVER 360° MEANS THAT THE VALUE IN MACHINE IS

SLIGHTLY OVER 0°. VALUES COME FROM COLLECTED PRODUCTION DATA. ....................................................... 50 FIGURE 37: THE IMPEDANCE ANGLE, THE DIFFERENCE BETWEEN NOISE AND CRACK, IS SLIGHTLY SMALLER FOR THE UHT-

STEELS. THE ANGLE FOR OT70 (Α) IS LARGER THAN FOR OT91 (Β) .............................................................. 50 FIGURE 38: GAIN AND PHASE ANGLE COMPARISON FOR 5,00 MM OT70 WIRE. FILTER CORRECTION IS KEPT AT 0 ......... 51 FIGURE 39: DIFFERENT CRACK CURVES. A CURVE WITH A PHASE ANGLE FAR FROM THE TRIGGER LEVEL HAS TO USE MORE

GAIN, SINCE IT HAS A LONGER DISTANCE FROM THE TRIGGER CALIBRATION LEVEL. ........................................... 52 FIGURE 40: PHASE IS CHANGED SO IT IS LAID DOWN 45°. THIS IS LIFTING UP THE NOISE DIRECTION AND GIVES A

DISTURBED SIGNAL. .................................................................................................................................. 53 FIGURE 41: SHOWING CORRECT INTERPRETATION OF A SPECTRUM GIVING THE LOWEST GAIN AND THE LOWEST NOISE .. 53 FIGURE 42: INCREASE OF GAIN NECESSARY WHEN LIFT OFF OR CLEARANCE IS INCREASED ........................................... 54 FIGURE 43: PIE CHART OF THE VALUES USED IN PRODUCTION. -1 IS THE MOST COMMON VALUE USED. ........................ 55 FIGURE 44: FILTER CORRECTION +3 HAS THE LOWEST S/N-RATIO, BUT HAS A HIGH LATITUDINAL NOISE AND THE VERY

UNSTABLE. THE TWO SMALL FIGURES TO THE RIGHT SHOW FILTER CORRECTION VALUE 0 AND +3. +3 HAS BEST

S/N RATIO HAS, BUT THE NOISE IS FULL OF SHARP PEAKS. ........................................................................... 55 FIGURE 45: GAIN IS CORRELATING HEAVILY ON THE FILTER CORRECTION. VALUES AVERAGE BASED ON DIMENSION 2,6 MM,

2,95 MM, 2,90 MM, 3,75 MM, 3,95 MM, 5,00 MM AND 5,50 MM. ................................................................... 56 FIGURE 46: THE DIFFERENT CONDUCTIVITY VALUES FOR SELECTED MATERIALS. CARBON CONTENT IS INCREASING FROM

THE LEFT TO THE RIGHT ON STEEL GRADE ................................................................................................... 57 FIGURE 47: MAGNETIC FLUX DENSITY ON OIL TEMPERED HARDENED SAMPLES (MIXED STEEL GRADES) ........................ 59 FIGURE 48: MAGNETIC DENSITY ON OIL TEMPERED HARDENED SAMPLES (MIXED STEEL GRADES) ................................ 59 FIGURE 49: PHASE CORRELATION WITH FREQUENCY FOR 100 KHZ, 300 KHZ AND 1000 KHZ. .................................... 61 FIGURE 50: PHASE ANGLE FOR PLANAR NOISE OF DIFFERENT DIMENSIONS OF WIRE USED FOR FREQUENCY 100 KHZ, 300

KHZ AND 1000 KHZ ................................................................................................................................. 62 FIGURE 51: CM (COLOR MARKS) IS INCREASING WITH INCREASING ALLOY CONTENT, BY METER UNITS. ........................ 63 FIGURE 52: ROUND AND PROFILE, COLOR MARKED DEFECTS IN CIRCOGRAPH AND DEFECTOMAT. ................................ 64 FIGURE 53: LEAD SIGNAL IN XY-MODE NOTICED IN PRODUCTION. NOTE THE SIMILARITY BETWEEN NORMAL CRACK’S

APPEARANCE ........................................................................................................................................... 65 FIGURE 54: A LEAD MARK STICKS TO THE STEEL WIRE ........................................................................................... 65 FIGURE 55 EDM ARTIFICIAL DEFECT COMPARED TO SAW MANUFACTURED. .............................................................. 66 FIGURE 56: PLACEMENT OF ARTIFICIAL DEFECTS ON OVAL AND PROFILES WIRES. PICTURE FROM FOERSTER AND FROM

GARPHYTTAN WIRE (RIGHT) ...................................................................................................................... 66 FIGURE 57 EGG- SHAPED WIRE AN INCORRECT PLACEMENT OF AN ARTIFICIAL CRACK. ................................................ 67 FIGURE 58: LEAD IS PRESENT ON A CALIBRATION SAMPLE; IT IS DIRECT CONNECTION TO THE ARTIFICIAL CRACK (TO THE

RIGHT) .................................................................................................................................................... 68 FIGURE 59: A CALIBRATION MADE ON THE CALIBRATION SAMPLE WITH A LEAD (UPPER). WHEN PUTTING IN THE REAL

DEFECT (BOTTOM), SUPPOSED TO BE 80µM, IT SHOWS MUCH HIGHER VALUES. ................................................ 68 FIGURE 60: TWO EXAMPLES WHEN THE DEPTH OF THE CALIBRATION SAMPLE IS COVERED BY DIRT ............................. 69 FIGURE 61: CLEANED ARTIFICIAL DEFECT ON THE CALIBRATION SAMPLE .................................................................. 69 FIGURE 62: DEPTH OF THE DEFECT WAS 18.49µM. POTENTIAL= 40µM-18.49=21.11 µM .......................................... 70 FIGURE 63: PART OF THE CONFIDENTIAL SOP INSTRUCTION FOR OPERATORS WORKING WITH ECT. .......................... 72

Page 80: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

78

References

1. Damhuji Rifai, Ahmed N. Addalla, Kharudin Ali and Ramdan Razali. (2016).

Giant magnetoresistance Sensors: A review on Structures and Non-Destructive

Eddy Current Testing Applications.

2. B.P.C Rao. (2007). Practical Eddy Current Testing. ISBN: 978-1842652992.

Indian Society for Non Destructive Testing.

3. Institut Dr. Foerster GmbH & Co. KG. (February 2011): Statement regarding

Eddy Current Testing of Valve Spring Wire.

4. Joseph M. Buckley, formerly of HOCKING NDT. An introduction to Eddy Current

Testing theory and technology

5. Javier Garcia-Martin, Jaime Gomez-Gil and Ernesto Vazquez-Sanchez. (2011).

Non-Destructive techniques based on Eddy Current Testing. ISSN 1424-8220

6. Placko, D. and Dufour, I. (October 1992). Eddy current sensors for nondestructive

inspection of graphite composite materials. In Proceedings of the IEEE Conference

of the industry Applications Society, Houston, TX, USA; PP 1676-1682

7. Xu, P., Huang, S. and Zhao, W. ( February 2010). Differential eddy current

testing sensor composed of double gradient winding coils for crack detection. In

Proceedings of the IEEE sensors application symposium , Limerick, Ireland; pp. 59-

63

8. International Atomic Energy Agency. (2011). Eddy current Testing at Level 2:

Manual for the Syllabi Contained in IAEA-TECDOC-628.Rev. 2 “Training

Guidelines for Non-Destructive Testing Techniques.Vienna

9. D. Salathé, R.K. Maix. (1988). Quality Control of RutherFord-Type Cables by the

Eddy Current Test Method. BBC Brown Boveri Lth, Dept. ISM CH-8050 Zurich

Switzerland (Available IEEE, DOI: 10.1109/20.11431)

10. Conductivity and resistivity values for Iron & Alloys tables URL:https://www.nde-

ed.org/GeneralResources/MaterialProperties/ET/Conductivity_Iron.pdf , read

20180517

11. AISI 1008 Carbon Steel (UNS G10080) standard

URL:https://www.azom.com/article.aspx?ArticleID=6538 , read 20180517

12. Institut Dr. Foerster GmbH & Co. KG. (2008). Circograph sensor system Ro 20 P

6.460, Operating Instructions

13. Defectomat DS 2.815 Foerster. Defectomat DS 2.815.

14. J. Schöder, B. Lutz, M, Halter. Electromagnetic test equipment fully integrated in

quality assurance by networks. Institute Dr Förster.

15. Mitutoyo. Quick Guide to Precision Measuring instrument. E4329. URL:

https://www.mitutoyo.co.jp/eng/pdf/E4329_QuickGuide.pdf

Page 81: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

79

16. Mark Izumi. New CV-2100N4 and M4 Contracers from Mitutoyo. URL:

https://www.mitutoyo.com/wp-content/uploads/2014/06/CV-2100M.jpg

17. Suzuki Garphyttan.(August 2014). Oil tempered valve spring wire

18. J. Gough, D Nettleton. (2010). Standard Operating Procedures (SOPs)-

Companies Applying What Regulations Say To Their Specific Operations

19. I.Colligon and M.Rosa .(2006). GLP SOPs for Equipment Calibration and

Maintenance Part 3: Process mapping for SOP Development

20. I.Colligon and M.Rosa .(2007). GLP SOPs for Equipment Calibration and

Maintenance Part 5: SOP Templates and SOP on SOPs

21. J.Blitz and T.S. Peat. (1981). The application of multi-frequency Eddy Currents

to Testing Ferromagnetic Metals” NDT international, Vol 14, part 1, PP 15-17

22. A.A. Diaz, R.A. Mathews, J.Hixon and S.R.Doctor. (2007). Assessment of Eddy

Current Testing for the Detection of Cracks in Cast Stainless Steel Reactor Piping

Components. Pacific Northwest National Laboratory NUREG/CR-6929 PNNL-

16253

23. J. Blitz, D.J.A. Williams and J. Tilson. (1981) Calibration of eddy-current test

equipment.

24. M. Kashefi, M.S. Amiri. (2010) Investigation of Variables Affecting Impedance

Plane in Eddy Current Testing of Carburized Steels

25. Ho-Young Mun, Chang Eob Kim. (January 2014) Comparison of ECT Probes in

Diagnosis of Defects

26. W. Lowrie. (2007) Fundamentals of Geophysics. Cambridge University Press. pp.

254

27. W. Eccles. (2011) Pragmatic Electrical Engineering: Fundamentals - Synthesis

Lectures on Digital Circuits and Systems, pp 166

28. P. E. Mix. (2005) Introduction to nondestructive testing - A training guide,

ISBN-13 978-0-471-42029-3

Page 82: Standardization of Eddy Current Testing Calibration for ...1236071/FULLTEXT01.pdf5 NOMENCLATURE Abbreviations Abbreviation Significate SG Suzuki Garphyttan NDT Non-destructive Testing

TRITA ITM-EX 2018:545

www.kth.se


Recommended