Sterilization – In General What’s the process? Decontamination – removes > 80% of microbes Sterilization – Kills/inactivates the remaining microbes What are the Sterilants? Saturated Steam Ethylene Oxide – Ozone Hydrogen Peroxide High Level Disinfectants (Glutaraldehyde, OPA, PAA)
Transcript

Sterilization – In General

• What’s the process?– Decontamination – removes > 80% of microbes– Sterilization – Kills/inactivates the remaining microbes

• What are the Sterilants?– Saturated Steam– Ethylene Oxide– Ozone – Hydrogen Peroxide– High Level Disinfectants (Glutaraldehyde, OPA, PAA)

Sterilization – In General

• What’s the process?– Decontamination – removes > 80% of microbes– Sterilization – Kills/inactivates the remaining microbes

• What are the Sterilants?

–Saturated Steam– Ethylene Oxide– Ozone – Hydrogen Peroxide– High Level Disinfectants (Glutaraldehyde, OPA, PAA)

Steam Sterilization

• “Critical” Variables– Time.– Temperature.– Saturated Steam. (Depends on Temperature and Pressure)

• Other Variables (Incomplete List)– Loading– Sterilant contact with items to be sterilized– Outside weather changes– Boiler Chemicals– Wrapping Materials– Etc, etc, etc…….

Saturated Steam

0

10

20

30

40

50

60

250 254 258 262 266 270 276 280

Temperature

Pressure (PSIA)

Wet Steam/ Liquid Water

Superheated Steam

Saturated Steam – What is it?

• It is a colourless gas, containing < 3% liquid water

• Depends on temperature and pressure• It will condense if it cools down a fraction

of a degree• It releases latent or potential heat energy

as it condenses• It is the latent or potential heat transferred

to microbes that kills/inactivates them

Latent Heat = Potential Energy

16Tons

16 Ton weight is sitting on the floorIt has zero Potential energy

Researcher knows the16 ton weight has zeroPotential energy.She is not afraid.

Latent Heat = Potential Energy

16Tons

2 meters

16 ton weight is now 2 metersoff the ground. It has quite a lotof potential energy with respect to the researcher

The researcher is nowvery afraid of the 16 tonweight’s potential energy

Latent Heat = Potential Energy

16Tons

2 meters

Latent or Potential energy from the 16 ton weight is now realized. Especially by the lateResearcher.

Latent Heat = Potential Energy

16Tons

16Tons

Steam (Gas)

Water (Liquid)

The “Experiment”Generating Saturated Steam at 1

Atmosphere

Temp

Time

100ºC

All the energy is now going in To creating “latent” heat in the steamWith no increase in temperature

It takes 150 KJoules to heat 1 lb of water from 20ºC to 100ºCand 1023 Kjoules to convert 1 lb of water from liquid to gas at 100ºC

Sterilization Monitoring

• We monitor the Process not sterility of items

• Because the process is validated (by the manufacturer of the sterilizer) we assume that, if the process functions correctly, the goods in the process will be sterile.

Process Monitors for Steam Sterilization

• 3 Major types used:• Biological Indicators:

– Monitor all variables in the process

• Chemical Indicators:– Monitor one or more of the critical variables

• Mechanical Indicators:– Monitor one of the critical variables

Process Monitors for Steam Sterilization

• Individually they provide information, but only together do they tell the whole story!

• A failure in any one of the monitors implies the process is incomplete in one or more necessary variables

Biological Indicators

• Large # (> 105 or 106) of resistant spores

• Usually “Self-contained” type

• Placed in a normally loaded sterilizer

• At least one per day per cycle (CSA)

• Final Readout made at 1, 3, 24 or 48 hours

• Test the process to ensure it is capable of killing microorganisms

VialVial

SporeSporeStripStrip

Ampoule + Ampoule + Growth mediaGrowth media& Indicators & Indicators

FilterFilter

CapCap

Self-contained Biological Indicator

How do they Work?

• Colour change indication:– If spore survives, it is incubated– Spore consumes nutrients, excretes acid– Spore creates subsequent generations that

also consume nutrients and excrete acids– Purple (or other) dye is acid sensitive– If enough acid is produced, dye will change

from purple to yellow (or other colour change)– Process can take up to 48 hours to get

reliable prediction of kill

How do they Work?

• Rapid Read BIs– Spore survives the sterilization cycle and is

incubated– Nutrients are “tagged” with a U.V. fluorescent

dye– As spore consumes nutrients the dye is

released and begins to fluoresce– Fluorescence is detected electronically– Process takes up 3 hours to get reliable

prediction of kill

What bout these “Extended Cycles”?

• G.stearothermophilus will still be the test spore.

• Requires a new Process Challenge Device (test pack) not a new BI

• Extended cycle required because bigger. Heavier more complex sets protect microbes better

• PCD must provide the same protection to the BI

BIs and CIs

• All current sterilization standards (AAMI, CSA and ISO) require the use of Biological indicators.

• Chemical Indicators cannot replace Biological indicators because:– Biologics are far more complex– Biologics have the potential to recover after serious damage– Chemistries only approximate the response of Biologics– Chemical kinetics are far simpler than biologic kinetics– Chemical indicators only measure the critical variables of the

process. Biologics integrate all the variables of microbial death

Understanding Chemical Indicators

• To fully understand Chemical indicators one must understand:– BIER/CIER Vessels/Test methods used by

manufacturers– Stated Values (SVs)– ISO/AAMI Classifications of Chemical

Indicators– How it all relates to YOUR sterilization

process– What a pass/fail result really means

Test Methods for Testing Chemical Indicators

• Must be reproducible, controlled and consistent• Based on standard test “Sterilizers” called BIER

or CIER vessels• BIER = Biological Indicator Experimental

Resistometer• CIER = Chemical Indicator Experimental

Resistometer• BIER/CIER vessels performance specified by

ISO 18472• BIER/CIER Vessels behave very differently from

a typical health care sterilizer

BIER/CIER Vessels

• Very Small Chamber• Very tight controls • Highly reproducible

results• Basis for all Biological and

Chemical indicator testing• All CI “Stated Values” are

based on this vessel

Some Requirements of CIER Vessels

• Come-up time < 10 seconds• Come-down time < 10 seconds• Vacuum air removal < 2 minutes• No steam allowed during air removal• Set Temperature within 0.5ºC• Set Time within 1 second• Pressure within 3.5 Torr• One data point recorded per second

Contrast CIER Vessel to Health Care Sterilizer

• Come-up time 5 to 15 minutes• Come-down time 5 to 15 minutes• Vacuum air removal 5 to 15 minutes• Steam used during air removal• Set Temperature within 2ºC (?)• Set Time within ??• Pressure within ??• One data point recorded per 30 seconds

Test Methods for Chemical Indicators

• All testing done in a CIER Vessel

• Must meet requirements of ISO 11140-1

• ISO 11140 is written for Manufacturers not end users

• Specifications are broken in to 6 categories

• Only need to show the ability to reach the stated values.

Stated Values: What are they?

• For any given chemical indicator to be compliant with ISO 11140-1, it must have “Stated Values” for the parameters it measures.

• Example: For a Class 3 Chemical Indicator, it must have a stated value for temperature. This is the temperature at which we can expect to see the change in the CIER vessel.

Stated Values and ISO 11140-1

• The standard uses the “Stated Values” to define how the indicator will perform.

• Example: A Class 4 Indicator• Stated Values:

– 134ºC– 4 Minutes

• ISO allows a range:• Colour change (Pass result) must be seen at 134ºC,

after 4 minutes exposure in the CIER vessel. A fail (no colour change) result must be seen at 132ºC after only 3 minutes exposure. No additional testing is required at points in between.

Stated Values and ISO 11140-1

• A Class 6 example:• Stated Values:

– 134ºC, 4 Minutes (Saturated steam)– A pass result must be seen ( in the CIER

vessel) after 4 minutes exposure at 134ºC– A Fail result must be seen at 133ºC after 3

min 45 second exposure in a CIER vessel.– No additional testing of points in between is

required.

Stated Values and ISO 11140-1

• The BIG QUESTION

• So, if a chemical indicator shows a “pass” result in your sterilizer does it mean that the Stated Values given for that particular chemical indicator have been achieved?

Chemical Indicators

• Changes (reactions that cause colour change) start at temperatures well below the stated values for the CI

• Temperature at which the reaction starts is the initiation temperature

• Changes (reactions) will continue once initiated and the temperature is above the initiation temperature

Rates of Reaction and Initiation Energies

In our house my wife allows me thehonour of pressing my own shirts.

As part of that process I have learned a lot about the chemistry of burningfabrics.

Rates of Reaction and Initiation Energies

For example, I discovered that if I place an unplugged (cool) iron on a shirt,I can leave it there indefinitely andit will not scorch the shirt.

That’s because the energy imparted tothe shirt from the cool iron is well belowthe initiation energy required to start the burning reaction of the fabric.

It also doesn’t remove wrinkles

Rates of Reaction and Initiation Energies

• I’ve also learned that I can leave an iron on its lowest setting on the shirt indefinitely and it will not scorch the shirt either. I guess the low setting still doesn’t deliver enough energy to start the burning reaction

Rates of Reaction and Initiation Energies

• Of course, an iron on medium setting will scorch the shirt, but it takes quite a few minutes to do so,

• On high setting, the wrinkles go, but if I leave the iron on the shirt for even a short time, it will scorch.

The Scorched Shirt Example

0

10

20

30

40

50

60

70

80

90

20 80 140 200 260 320 380

Iron Temperature

Time to Scorch

Chemical Indicators

• Example (Theoretical) CI has stated values of 134ºC, 4 minutes: Initiation temperature may be 100ºC and reaction (change) will start to occur at that temperature and continue while the temperature remains over 100ºC.

• The rate of change will vary with temperature once initiated. Usually faster as temperature increases

BIER/CIER Vessel Testing

Same CI in a Health Care Sterilizer

Next Big Question(s)

• If chemical indicators start to react below the stated values:– How fast do they react at lower temperatures?– Will they change if they spend sufficient time

at lower temperatures?– What does that mean for monitoring my

sterilizer?

Extended Cycle Time Temperature Plot: Set Points 132ºC, 18 min)

0

20

40

60

80

100

120

140

Time (min)

Tem

per

atu

re (

ºC)

Sterilizer Chamber

Inside the Set

Conclusions so far…..

• Stated Values cannot be duplicated in Health Care Sterilizers

• Stated values can be used only for comparative purposes between Chemical Indicators of the same class

• Stated Values are used by Manufacturers of Chemical Indicators to make CSA/ISO or AAMI Classification claims.

CSA/ISO/AAMI Classification of Chemical Indicators

• All follow the ISO 11140, 6 tier classification structure

• All state the class has no hierarchical significance. (i.e. a Class 2 is not “better” than a Class 3 or a Class 6 is not “better” than a Class 5.)

• Classification structure is to provide specifications and targets for manufacturers of chemical indicators

• It may be used by users for comparing CIs within the same class

Definitions

Critical Variable– For steam sterilization Time, Temperature and

Saturated Steam– For Ethylene Oxide: EtO concentration, humidity,

time, temperature

• Critical Parameter– A set value of a critical variable that must be attained

• Stated Values– The critical parameters measured by a Chemical

Indicator in a CIER vessel

CSA/ISO/AAMI Classification

• Class 1 – Process Indicators– Autoclave tapes, built-in CIs on peel pouches– Simply identify processed from unprocessed

packages

• Class 2 – Special Indicators– Bowie-Dick Type Tests

• Class 3, 4, 5, and 6 are all Internal Indicators

Classes 3 - 6

• Class 3 Indicator:– Measures only one of the critical variables of

the process

• Class 4 Indicator:– Measures 2 or more of the critical variables of

the process– E.g. Time and temperature, or time saturated

steam.

Classes 3 - 6

• Class 5 Integrating Indicator– Measures all critical variables of the process, and

correlates with the test spore for that process across a range of sterilization conditions

– For steam sterilization, the Class 5 must correlate with G.stearothermophilus at 121ºC, 128ºC* and 135ºC

– Has a known response over a range of sterilization temperatures

* May be other temperatures within the range of 121 - 135ºC

Class 5 Integrating Indicator Response

0

2

46

8

10

1214

16

18

Temperature (C)

Tim

e (M

inu

tes)

Test Spore

Class 5 Integrator

Class 3 - 6

• Class 6 Emulating Indicator– Measures all critical variables of the process– Tightest specifications of all indicators– Stated Values are for a specific time and

temperature for steam sterilization

Class 6 Emulating Indicator Response

0

2

4

6

8

10

12

14

16

121

123

125

127

129

131

133

135

Temperature (C)

Tim

e (M

inu

tes) Test Spore

Class 6 StatedValue

Class 6 Emulating Indicator Response

0

2

4

6

8

10

12

14

16

121

123

125

127

129

131

133

135

Temperature (C)

Tim

e (M

inu

tes) Test Spore

Class 6 StatedValue

Possible Responsebelow Temp SV

Conclusions:

• Saturated Steam is critical to the sterilization process

• Non-condensable gases (air, CO2, N2) are poor heat transfer media and poor sterilants

• It is the latent heat of saturated steam, not the temperature that kills the microbes

Conclusions

• All process monitors must agree that the cycle was successful

• A failure in any one of the 3 types of monitor (Biological, chemical or mechanical) means the process was insufficient and processed goods are suspect

Conclusions

• Biological Indicators are the only indicator to integrate all the variables of he sterilization process

• They are the only direct measure of the lethality of the process

• The BI response can be approximated by chemical indicators, but CIs cannot replace BIs

Conclusions

• The basis for classification of Chemical Indicators in Canada is CANISO 11140

• This standard is written for manufacturers of chemical indicators.

• The standard has no hierarchical significance

• Stated values are only realizable in the test vessel (CIER vessel)

Conclusions

• Selection of biological indicators and chemical indicators must be made considering patient safety, the information needed from the indicators, the standard requirements, best practices and economic considerations. No single indicator will likely accommodate every circumstance