+ All Categories
Home > Documents > Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes...

Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes...

Date post: 26-Sep-2018
Category:
Upload: doantuong
View: 223 times
Download: 0 times
Share this document with a friend
40
Reamer/Irrigator/Aspirator (RIA). For intramedullary reaming and bone harvesting. Technique Guide
Transcript
Page 1: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Reamer/Irrigator/Aspirator (RIA).For intramedullary reaming and boneharvesting.

Technique Guide

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite Cvr1

Page 2: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite Cvr2

Page 3: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Introduction

Surgical Technique

Product Information

Bibliography

Overview 2

Indications 4

Preparation 5

Assembly 10

Reaming 19

Bone harvesting 21

Disassembly 24

Single-patient-use instruments 25

Other material 27

General instruments 28

Optional instruments 30

Other accessory instruments 31

Power Equipment 32

Cleaning instruments 33

Set List 33

34

Table of contents

Synthes 1

Image intensifier control

WarningThis description is not sufficient for immediate application of the instrumentation. Instruction by a surgeon experiencedin handling this instrumentation is highly recommended.

Reprocessing, Care and Maintenance of Synthes InstrumentsFor general guidelines, function control and dismantling of multipart instruments please refer to: www.synthes.com/reprocessing

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 1

Page 4: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

– Lower intramedullary pressure – Lower fat embolisation – Lower heat generated – Removal of infected tissue

Reaming with lowered risk of systemic complication Fast, efficient reaming

Deep flutes to prevent cloggingSharp edges for

optimized cutting

Removal of marrow andbone through aspirationholes creates negative intramedullary pressure

Constant cooling with irrigation fluid, also emulsifies reamings for easy evacuation

– Time saving, one-step procedure – Easier handling– Sharp reamer heads for optimized cutting

Tube assemblyReamer head

Flow of irrigating fluid

Bone marrow, morselized bone

Overview

2 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 2

Page 5: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Locking clip

Drive shaft

Aspiration of bone marrow and finely morselized bone

Mesh sized to capturemorselized bone and bone marrow

Aspiration tube connected to vacuumsource

Gravity fed irrigationinto cannulation ofdrive shaft

Drive shaft seal

Synthes 3

Autograft harvesting

– Low morbidity– Possible range: 30 to 80cc per procedure – Easy, well-known technique– Osteoinductive and osteogenic

Graft filter 100cc capacity

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 3

Page 6: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

Indications

Indications– To clear the medullary canal of bone marrow and reaming

debris – To clear the medullary canal of infected bone tissue– To effectively size the medullary canal for the acceptance

of an intramedullary implant or prosthesis – To harvest finely morselized autogenous bone and bone

marrow for any surgical procedures that require bone graftin order to facilitate fusion and/or fill bone defects. Theseprocedures include spinal fusion, joint arthrodesis, totaljoint replacement, fracture repair, nonunion, maxillofacialreconstruction, and tumor removal

Note: When deciding on bone harvesting, the patient’s history, bone quality, physiological condition and compliancemust be taken into account.

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 4

Page 7: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Synthes 5

1Select appropriate length of tube assembly

Intruments

314.745S RIA Tube Assembly, for RIA Drive Shaft oflength 360mm, for No. 314.742, sterile

314.746S RIA Tube Assembly, for RIA Drive Shaft of length 520mm, for No. 314.743, sterile

Select the appropriate length of the RIA tube assembly basedon the length of the canal.

Length of the canal Tube assembly

Less than 350 mm 360 mm

Greater than 350 mm 520 mm

Preparation

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 5

Page 8: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Preparation

6 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

2Confirm reaming diameter

Intrument

351.717 Depth Gauge for Medullary Nails

To estimate the canal diameter, position image intensifier foran AP view of the limb at the level of the isthmus. Hold theradiographic depth gauge perpendicular to the limb andoverlay the diameter tabs over the isthmus. Read the diame-ter on the tab that fills the canal. Repeat with a ML view.

a For reaming only, choose the implant diameter. Select areamer head 1.0 mm to 1.5 mm larger than the chosenimplant diameter.

b For bone harvesting only, select a reamer head 1.0 mm to1.5 mm larger than the canal diameter at the isthmus.

Note: Measure both in AP and LM view.

Caution: The distance of the depth gauge from the bone and the position of the receiver affect the diameter measure-ment. Always place the depth gauge on the side of the limb closestto the receiver. Estimate the real width as follows:Distance between depth gauge and bone~ 25 mm = 1 mm larger reading~ 50 mm = 2 mm larger reading~100 mm = 3 mm larger reading

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 6

Page 9: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Synthes 7

3Access canal and place reaming rod

Instruments

393.100 Universal Chuck with T-Handle

352.032 SynReam Reaming Rod � 2.5 mm, short,length 950 mm

Gain access to the intramedullary canal using standard open-ing procedures and instrumentation for the chosen implantor entry point. Reduce the fracture, if present.

Attach the universal chuck with T-handle to the reaming rod.Insert the reaming rod into the canal to the physeal scar. Ver-ify its position with the image intensifier.

Note: Ensure the reaming rod is centered in the canal inboth AP and LM view.

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 7

Page 10: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Preparation

8 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

4Confirm nail length (for reaming only)

Instruments

351.717 Depth Gauge for Medullary Nails

351.719 Elongation Tube for Reaming Rods oflength 950 mm

Assemble the extension tube to the depth gauge.

Place the depth gauge assembly over the reaming rod.

Note: The tip of the depth gauge assembly should be locatedat the entry site.

Caution: Use the depth gauge only with the 950 mm reaming rod.

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 8

Page 11: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Synthes 9

Determine the length of the canal. Choose the length of theimplant accordingly.

Note: Consider dynamization when choosing the length ofthe implant.

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 9

Page 12: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

1Attach reamer head

Instruments

352.250S – 352.259S RIA Medullary Reamer Heads 12.0 mm – 16.5 mm, sterile

352.261S – 352.265S RIA Medullary Reamer Heads 17.0 mm – 19.0 mm, sterile

314.745S RIA Tube Assembly, for RIA DriveShaft of length 360 mm, for No. 314.742, sterile

314.746S RIA Tube Assembly, for RIA DriveShaft of length 520 mm, forNo. 314.743, sterile

Select the appropriate reamer head. Insert it into the tube assembly.

Note: The reamer head is attached correctly when it can spinfreely and is retained by the tube assembly.

Caution: Reamer heads are sharp. Use care when attachingthe reamer head to the tube assembly.

10 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

Assembly

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 10

Page 13: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

2Attach drive shaft to tube assembly and reamer head

Instruments

314.742 RIA Drive Shaft, length 360 mm

314.743 RIA Drive Shaft, length 520 mm

Select the respective length of the drive shaft. Guide the tipof the drive shaft through the RIA tube assembly.

Synthes 11

Mate the hexagonal end of the RIA drive shaft with thehexagonal recess in the reamer head.

Note: Visually check mating through the aspiration holes.

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 11

Page 14: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Retaining Groove

12 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

Note: The drive shaft is properly attached when the hex flatsare not visible. The helix is also visible when properly attached.

Note: The Drive Shaft hex is properly engaged in the reamerhead when the retaining groove is hidden.

Assembly

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 12

Page 15: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Synthes 13

Push the outer sleeve of the RIA drive shaft forward to engagethe body of the RIA tube assembly.

Note: The drive shaft is properly attached when the tan/grayplastic coupling shaft is no longer visible.

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 13

Page 16: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

14 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

3Attach locking clip

Instrument

352.260S Locking Clip for RIA, sterile

Slide the locking clip onto the assembled RIA drive shaft andtube assembly.

Note: The small ridge on the inner surface of the locking clipmatches the groove between the drive shaft and the tube assembly.

Assembly

Buttons on tube assembly

Rotate the locking clip until the circular notches on the lock-ing clip align with the corresponding buttons on the tube assembly.

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 14

Page 17: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Synthes 15

4Attach drive shaft seal

Instrument

351.718.02S RIA Drive Shaft Seal

Attach the drive shaft seal to the proximal end of the driveshaft.

Note: The drive shaft seal is important to retain irrigation inthe cannulation of the drive shaft.

5Attach drive unit with large quick coupling

Instruments

530.605 Battery Reamer/Drill

530.760 Quick Coupling for DHS/DCS Triple Reamers, for Battery Power Line

Select a cannulated drive unit that will deliver 3.5 to 4.5 Nmof torque and 700 to 900 rpm (standard drill speed). Use thecannulated quick coupling to attach the drive shaft to thedrive unit.

Caution: Do not use a reduction drive. Drills with a torquegreater than 6 Nm must not be used. Power equipment de-signed for reaming must not be used.

Drive shaft seal

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 15

Page 18: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

16 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

Alternative with chuck with key

Alternative instrument

530.730 Jacob’s Chuck with Key

Use the chuck with key to attach the drive shaft to the driveunit.

Note: Insert key into one hole of chuck and tighten. Movekey to all subsequent holes on the chuck and tighten again.

6Connect irrigation

Instrument

Irrigation Tube

Other material

1– 2 liter container of saline solution (e.g. isotonic 0.9% NaCl)

Suspend a 1 or 2 liter container of saline solution or irrigat-ing fluid approximately one meter above the level of the RIAassembly.

Attach the end with spike on the irrigation tube to the irriga-tion container.

Note: The clamp on the irrigation set should be closed untilreaming begins.

Assembly

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 16

Page 19: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Synthes 17

Connect the opposite end of the irrigation tube to the irriga-tion port on the tube assembly.

Note: The irrigation port is the smaller of the two ports andis indicated by an “I”.

7Connect aspiration

Instrument

Aspiration Tube

Other material

2 – 5 liter Suction canister

Vacuum source

Connect the aspiration tube (non-flared end) to the aspira-tion port on the tube assembly.

Note: The aspiration port is the larger of the two ports andis indicated by an “A”.

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:48 Uhr Seite 17

Page 20: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

18 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

Alternative for bone harvesting

Instrument

352.229S Graft Filter for RIA, sterile

Connect the graft filter to the aspiration port on the tubeassembly.Connect the distal end of the graft filter to the aspirationtube.

Assembly

Connect the other end of the aspiration tube (flared end) tothe suction canister. Connect the suction canister directlyto the vacuum source.

Note: Connect the irrigation/aspiration tube directly to thesuction canister to avoid a reduction in suction. Never con-nect several suction canisters.

Start vacuum source, if necessary.

Note: Set adjustment to maximum vacuum.

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:49 Uhr Seite 18

Page 21: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Synthes 19

1Introduce RIA assembly into medullary canal

Guide the reamer head over the reaming rod.Open the irrigation clamp to start flow. Turn on the vacuumsource.

Note: Prior to insertion into patient’s limb, visually verify flowof irrigation fluid at the tip of the reamer head.

Advance the RIA assembly over the reaming rod until the aspiration holes are fully immersed into the bone. Check position on the image intensifier. A flow of blood and bonemarrow becomes visible in the aspiration tube.

Note: Flow of aspiration begins when aspiration holes arefully immersed in the bone.

Caution: Never ream when there is no irrigation/aspiration.The irrigation/aspiration cools the reamer heads and removesbone marrow and morselized bone from the medullary cav-ity. It is crucial for the properties of the system.

Reaming

Aspirationholes

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:49 Uhr Seite 19

Page 22: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

20 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

2Ream

Commence reaming using a gradual advancing/retractingtechnique. Slowly advance 20 –30 mm and then retract50– 80 mm allowing the irrigation to flow in front of reamerhead. Then advance until resistance is felt and repeat.

Note: An insufficient irrigation volume may lead to clogging.Never advance too fast to avoid clogging.

Caution: Periodically check that the reaming aspirate is flow-ing back via the aspiration port and tube into the suctioncanister. Stop reaming if there is no flow.

After the reamer head reaches the desired depth, withdrawthe RIA assembly while maintaining rotation with the drill.

Note: Reverse the drill if advancement of the reamer becomes difficult.

Reaming

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:49 Uhr Seite 20

Page 23: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Synthes 21

1Turn off irrigation and aspiration

Instrument

352.229S Graft Filter for RIA, sterile

Stop irrigation after withdrawing the RIA assembly from thepatient. Turn off the suction or clamp the suction tubing.Disconnect the filter canister from the lid and move it to theback table.

2Prepare graft

Hold graft filter vertically. Compress the bone graft by gentlypushing down on the plunger. Estimate the volume of bonegraft with the scale on the outer canister.

Bone Harvesting

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:49 Uhr Seite 21

Page 24: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

22 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

3Remove graft from filter

Invert filter vertically with the plunger in a downward posi-tion. Remove the inner filter from the canister.

Caution: When handling, do not hold the opening down-wards as there is a risk of dropping the graft.

Bone Harvesting

Invert the inner filter vertically. Push the bone graft out of thefilter with the plunger.

Hold the inner filter over an appropriate container.

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:49 Uhr Seite 22

Page 25: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Synthes 23

4Augment with a bone substitute (optional)

Optional implants

710.019S chronOS Granules, 1.4–2.8 mm, 10.0 cc, porosity 60%*

710.021S chronOS Granules, 1.4–2.8 mm, 20.0 cc, porosity 60%*

710.026S chronOS Granules, 2.8–5.6 mm, 10.0 cc, porosity 60%*

710.027S chronOS Granules, 2.8–5.6 mm, 20.0 cc, porosity 60%*

To increase the volume of harvested bone graft and to pro-vide it with structure, augment it with chronOS Granules.

Note: chronOS is a synthetic , fully resorbable and phasepure �-Tricalcium Phosphate bone substitute. The inter -connected porosity allows for a complete remodeling to newbone within 6 to 18 months.

*availability dependent on regulatory conditions

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:49 Uhr Seite 23

Page 26: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

24 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

1Remove drive shaft

Remove and dispose of the irrigation/aspiration tube andlocking clip. Push the outer sleeve of the drive shaft towardthe body of the RIA tube assembly.

Disassembly

Keep the outer sleeve in a forward position while withdrawingthe drive shaft from the body of the tube assembly.

Remove the drive shaft from the RIA tube assembly.

Note: Dispose of the tube assembly and reamer head. Remove and dispose of the drive shaft seal.

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:49 Uhr Seite 24

Page 27: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Synthes 25

Single-Patient-Use instruments

Reamer Heads – Conical tip helps centralize reamer– Deep flutes prevent clogging– Sharp cutting edges for every case– Diameters: 12 mm –19 mm in 0.5 mm increments– Stainless-steel– Sterile packed– Attaches to RIA Tube Assemblies

Article No. Diameter (mm)

352.250S 12.0

352.251S 12.5

352.252S 13.0

352.253S 13.5

352.254S 14.0

352.255S 14.5

352.256S 15.0

352.257S 15.5

352.258S 16.0

352.259S 16.5

352.261S 17.0

352.262S 17.5

352.263S 18.0

352.264S 18.5

352.265S 19.0

Product Information

Tube Assemblies – Two lengths available (attach to corresponding drive shafts) – Sterile packed

Article No. Length (mm)

314.745S 360

314.746S 520

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:49 Uhr Seite 25

Page 28: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

26 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

351.718.02S RIA Drive Shaft Seal, sterile, pack of 2 units– Elastomeric seal– Prevents flow of irrigation fluid into drive unit– Sterile packed

352.229S Graft Filter for RIA, sterile– Collects finely morselized bone chips and marrow– Capacity: 100 cc– Graduations for estimating volume– Plunger for easy removal of graft– Sterile packed

352.260S Locking Clip– Secures attachment between drive shaft and tube assembly– Sterile packed

Irrigation/Aspiration Tubes– Spike on irrigation tube to connect with irrigation source– Clamp on irrigation tube to control flow of irrigation fluid– Aspiration tube to connect to suction canister

Product Information

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:49 Uhr Seite 26

Page 29: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Synthes 27

Other Material

Irrigation Fluid Bag– 1 – 2 Liters for reaming– 3 Liters minimum for harvesting

Suction Canister– 2 Liters minimum for reaming– 3 Liters minimum for harvesting

Vacuum Source – Wall suction or portable suction

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:49 Uhr Seite 27

Page 30: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

28 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

General Instruments

Article No. Length (mm)

314.742 RIA Drive Shaft, length min 360 mm

314.743 RIA Drive Shaft, length min 520 mm

– Made of superelastic, Nitinol, an alloy of nickel and titanium

– Two lengths available – Attaches to RIA tube assemblies

351.050 Tissue Protector

351.717 Depth Gauge for Medullary Nails

351.719 Extension Tube for Reaming Rods;assembles to Depth Gauge for MedullaryNails, for No. 351.717

Product Information

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:50 Uhr Seite 28

Page 31: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Synthes 29

355.010 Medullary Tube, synthetic

393.100 Universal Chuck with T-Handle

352.032* SynReam Reaming Rod � 2.5 mm, short,length 950 mm

*Available non-sterile or sterile packed. Add “S” to the catalogue number to ordersterile products.

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:50 Uhr Seite 29

Page 32: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

30 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

Optional Instruments

Opening instruments

351.020 Reverse Awl, small

03.010.04 Awl � 12.0 mm, cannulated

03.010.115 Guide Wire � 3.2 mm, 290 mm

03.010.036 Drill Bit � 12.0 mm, cannulated, length300 mm, for No. 532.015

351.240 Cutter for UTN/CTN and UniversalMedullary Nail, � 11.0 mm, length 350 mm

Product Information

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:50 Uhr Seite 30

Page 33: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Synthes 31

Other accessory instruments

351.782 Holding Forceps for SynReam Reaming Rod� 2.5 mm

351.060 Centering Pin � 4.0 mm, length 400 mm,for No. 351.240

351.920 Hand-Reamer for Medullary Canal � 6 mm

351.930 Hand-Reamer for Medullary Canal � 7 mm

351.940 Hand-Reamer for Medullary Canal � 8 mm

03.010.093 Rod Pusher for Reaming Rod withHexagonal Screwdriver � 8.0 mm

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:50 Uhr Seite 31

Page 34: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

32 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

Power equipment

530.605 Battery Reamer/Drill

530.730 Chuck with Key, for Compact

530.760 Quick Coupling for DHS/DCS TripleReamers, for Battery Power Line

Product Information

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:50 Uhr Seite 32

Page 35: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Synthes 33

Cleaning instruments

352.041 Cleaning Brush � 3.6 mm, length 600 mm,for flexible shafts

351.800 Air Jet, for cleaning Instruments withCompressed Air. Do not sterilize

351.810 Air Tube � 2.0 mm, for No. 351.800

RIA-Instruments Set (105.309)

Instruments

314.742 RIA Drive Shaft, length 360 mm

314.743 RIA Drive Shaft, length 520 mm

351.050 Tissue Protector

351.717 Depth Gauge for Medullary Nails

351.719 Extension Tube for Reaming Rods; assembles to Depth Gauge for MedullaryNails, for No. 351.717

355.010 Medullary Tube, synthetic

393.100 Universal Chuck with T-Handle

690.308 Graphic Case for RIA Instrument Set,without Contents

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:50 Uhr Seite 33

Page 36: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

34 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide

Fat embolism and reamed/unreamed nailing

Danckwardt-Lilliestrom G, Lorenzi GL (1970) Intramedullarynailing after reaming – an investigation on the healingprocess in osteotomized rabbit tibias. Acta Ortho Scand S134

Sturmer KM (1993) Measurement of intramedullary pressurein an animal experiment and propositions to reduce the pres-sure increase. Injury 24(3):7-21

Robinson CM, Ludlam CA, Ray DC, Swann DG, Christie J(2001) The coagulative and cardiorespiratory responses toreamed intramedullary nailing of isolated fractures. J BoneJoint Surg Br 83(7):963-73

Joist A, Schult M, Ortmann C, Frerichmann U, Frebel T,Spiegel HU, Kropfl A, Redl H (2004) Rinsing-suction reamerattenuates intramedullary pressure increase and fat intravasa-tion in a sheep model. J Trauma 57(1):146-51

Pape HC, Zelle BA, Hildebrand F, Giannoudis PV, Krettek C,Van Griensven M (2005) Reamed femoral nailing in sheep:does irrigation and aspiration of intramedullary contents alterthe systemic response? J Bone Joint Surg Am. 87(11):2515-22

Schult M, Kuchle R, Hofmann A, Schmidt-Brakling T, Ort-mann C, Wassermann E, Schmidhammer R, Redl H, Joist A(2006) Pathophysiological advantages of rinsing-suction-reaming (RSR) in a pig model for intramedullary nailing. J Or-thop Res 24(6):1186-92

Muller CA, Green J, Sudkamp NP (2006) Physical and techni-cal aspects of intramedullary reaming. Injury 37(4):39-49

Husebye EE, Lyberg T, Madsen JE, Eriksen M, Roise O (2006)The influence of a one-step reamer-irrigator-aspirator tech-nique on the intramedullary pressure in the pig femur. Injury37(10):935-40

Pape HC, Giannoudis P (2007) The biological and physiologi-cal effects of intramedullary reaming. J Bone Joint Surg Br89(11):1421-6

Van Gorp CC, Falk JV, Kmiec SJ Jr, Siston RA. (2008) TheReamer/Irrigator/Aspirator Reduces Femoral Canal Pressure inSimulated TKA. Clin Orthop Relat Res Apr 19 [Epub ahead ofprint]

Bibliography

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:50 Uhr Seite 34

Page 37: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

Synthes 35

Osteomyelitis

Bellapianta J, Gerdeman A, Sharan A, Lozman J (2007) Useof the Reamer Irrigator Aspirator for the treatment of a 20-Year recurrent osteomyelitis of a healed femur fracture.J Orthop Trauma 21(5):343-6

Zalavras CG, Singh A, Patzakis MJ (2007) Novel technique formedullary canal débridement in tibia and femur osteo -myelitis. Clin Orthop Relat Res 461:31-4

Bone harvesting

Wenisch S, Trinkaus K, Hilda A, Hose D, Herde K, Heiss C,Kilian O, Alt V, Schnettler R (2005) Human reaming debris: asource of multipotent stem cells. Bone 36:74-83

Schmidmaier G, Herrmann S, Green J, Weber T, Scharfen-berger A, Haas NP, Wildemann B (2006) Quantitative assess-ment of growth factors in reaming aspirate, iliac crest, andplatelet preparation. Bone 39(5):1156-63

Stafford PR, Norris BL (2007) Reamer-Irrigator-Aspirator as aBone Graft Harvester. Techniques in Foot & Ankle Surgery6(2):100-107

Hammer TO, Wieling R, Green JM, Südkamp NP, Schneider E,Müller CA (2007) Effect of re-implanted particles from in-tramedullary reaming on mechanical properties and callusformation: A laboratory study. J Bone Joint Surg Br89(11):1534-8

Newman JT, Stahel PF, Smith WR, Resende GV, Hak DJ, Morgan SJ (2008) A new minimally invasive technique forlarge volume bone graft harvest for treatment of fracturenonunions. Orthopedics (3):257-61

Kobbe P, Tarkin IS, Frink M, Pape HC (2008) Voluminousbone graft harvesting of the femoral marrow cavity for autol-ogous transplantation: An indication for the “Reamer-Irriga-tor-Aspirator-” (RIA-)technique. Unfallchirurg 111(6):469-72

Nichols TA, Sagi HC, Weber TG, Guiot BH (2008) An alterna-tive source of autograft bone for spinal fusion: the femur:technical case report. Neurosurgery 62(3 Sup 1):E179

Kobbe P, Tarkin IS, Pape HC (2008) Use of the “reamer irriga-tor aspirator” system for non-infected tibial non-union afterfailed iliac crest grafting. Injury 39(7):796-800

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:50 Uhr Seite 35

Page 38: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:50 Uhr Seite 36

Page 39: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:50 Uhr Seite Cvr3

Page 40: Technique Guide - OSTEOSYNTESE.dkosteosyntese.dk/wp-content/uploads/2014/11/RIA.pdf · 4 Synthes Reamer/Irrigator/Aspirator (RIA) Technique Guide Indications Indications – To clear

0123Presented by: 036.

000.

553

SE_1

5736

6 A

A

3007

0093

©

07/

2008

Syn

thes

, Inc

. or

its a

ffili

ates

A

ll rig

hts

rese

rved

Sy

nthe

s is

a t

rade

mar

k of

Syn

thes

, Inc

. or

its a

ffili

ates

0x6.000.553_AA.qxp:0x6.000.553 26.08.2008 14:50 Uhr Seite Cvr4


Recommended