Temperature Measurement
Gerald Recktenwald, Evan ThomasPortland State University
Department of Mechanical Engineering
Temperature Measurement
• Liquid bulb thermometers
• Gas bulb thermometers
• bimetal indicators
• RTD: resistance temperature detectors (Platinum wire)
• thermocouples
• thermistors
• IC sensors
• Optical sensors
Pyrometers⊲ ⊲ Infrared detectors/cameras liquid crystals⊲
IC Temperature Sensors
• Semiconductor-based temperature sensors or thermocouple reference-junctioncompensation
• Packaged suitable for inclusion in a circuit board
• Variety of outputs: analog (voltage or current) and digital
• More useful for a manufactured product or as part of a control system than aslaboratory instrumentation.
Thermistors
A thermistor is an electrical resistor used tomeasure temperature. A thermistor designed suchthat its resistance varies with temperature in arepeatable way.A simple model for the relationship betweentemperature and resistance is
ΔT = kΔR
A thermistor with k > 0 is said to have a positivetemperature coefficient (PTC). A thermistor withk < 0 is said to have a negative temperaturecoefficient (NTC).
Photo from YSI web site:www.ysitemperature.com
Thermistors
Advantages
• Output is directly related to absolute temperature - no reference junction needed.
• Relatively easy to measure resistance
• Sensors are interchangeable (±0.5◦C)
Disadvantages
• Possible self-heating error
⊲ Each measurement applies current to resistor from precision current source ⊲ Measure voltage drop ΔV , then compute resistance from known current and ΔV . Repeated measurements in rapid succession can cause thermistor to heat up⊲
• Can be more expensive than thermocouples for comparable accuracy: $10 to $20/eachversus $1/each per junction. Thermistors costing less than $1 each are available fromelectronic component sellers, e.g. Digikey or Newark.
• More difficult to apply for rapid transients: slow response and self-heating
Thermistors
50Data
Calibration uses the Steinhart-Hart equation
Nominal resistance is controllable bymanufacturing.
Typical resistances at 21◦C:10 kΩ, 20 kΩ, . . . 100 kΩ.
45 Curve Fit
40
35
30
25
20
15
10
5
05 10 15 20 25 30
Resistance (kΩ)
Thermistors
Two-wire resistance measurement:
Resistance in the lead wires can lead to inaccurate temperature measurement.
Ohmmeter
ThermistorI
RT V
Thermistors
Four-wire resistance measurement eliminates the lead resistance1
Ohmmeter
Rlead
Thermistor
RTRlead I
Rlead
Rlead
V
1Sketch adapted from Hints for Making Better Digital Multimeter Measurements, Agilent Technologies Corporation,www.agilent.com.
Thermocouples
• Principle of operation
• Wire types: B, E, J, K, N, R, S, T
• Formats: prefab, homemade, fast response, slow response
• Circuit diagrams: reference junction compensation
• Good practice
Heat
Metal A
Metal BVAB
+
-
Type K • Chromel {90% nickel and 10% chromium} • Alumel {95% nickel, 2% manganese, 2% aluminum and 1% silicon}• 41 µV/°C • −200 °C to +1350 °C / -330 °F to +2460 °F range.
Seebeck Effect
Temperature gradient in a conductor induces a voltage potential
T2
E12
Voltmeter
T1
E12 = σ(T2 − T1) (1)
where σ is the average Seebeck coefficient for the range T1 ≤ T ≤ T2.
Seebeck Effect
Perturb T2 while holding T1 fixed:
(2)
Subtract Equation (1) from Equation (2) to get
(3)
Rearrange
(4)
σ is an intrinsic property of the material, so
(5)
Seebeck Effect
Applying the limit yields the derivative:
so
(6)
Equation (6) is the definition of the Seebeck Coefficient
Nominal values of Seebeck Coefficient
σ values are small, so the voltage output from thermocouples is small, typically on theorder of 10−3 V.
Reference Junction
Thermocouples are only capable ofmeasuring temperature differences.
To measure the temperature of anobject, we need a known referencetemperature. The thermocouple isused to measure the temperaturedifference between the object andthe known reference temperature.
Physical Circuit:x
Tt Tr
1 Cu2
Voltmeter 5 Cu 4
Conceptual T(x) Plot:
Tj
Tt 1
Cu
Tr 2
material P Tj
3material N
3
5
Cu
4
x
Ejr
emf relativeto referencejunction
Etr
0
Ice Point Reference Junction
Ice-point reference junction used in the thermal lab:
Glass tube, partiallyfilled with oil
Ice/watermixture
Insulated thermos
Calibration Curves
Working with calibration data
• Integrals are never evaluated.• Data is tabulated and curve fit
• Standard polynomial curve fitsand coefficients are available forcommon thermocouple types
Calibration data for T-Typethermocouples:
E (mV) T (◦C)
0.0000 0
2.0357 50
4.2785 100
6.7041 150
9.2881 200
12.0134 250
14.8619 300
17.8187 350
20.8720 400
Temperature Data Acquisition and Analysis
Evan ThomasPortland State University
Department of Mechanical Engineering
Resolution – the number of bits involved in an analog to digital conversion. Conversion time – the time it takes for an ADC to convert an analog voltage sample to a digital code.Quantization error – The rounding error between the true analog value and the recorded digital value.
Analog to Digital Conversion
Voltage 2-Bit Digital Representation0 to 2.5 002.5 to 5 015 to 7.5 107.5 to 10 11
Digital Data• Hexadecimal numbers
– Save space
• Signing bit– Represent negative numbers– Known as twos complement
Decimal Hexadecimal0 01 12 23 34 45 56 67 78 89 910 A11 B12 C13 D14 E15 F
Bits Unsigned value 2's complement value0111 1111 127 127 0111 1110 126 126 0000 0010 2 2 0000 0001 1 1 0000 0000 0 0 1111 1111 255 −1 1111 1110 254 −2 1000 0010 130 −126 1000 0001 129 −127 1000 0000 128 −128