D A K O T A U L T R A S O N I C S S O U N D S O L U T I O N S The Best of Both Worlds... The Best of Both Worlds... The DFX-7 Series The DFX-7 Series Flaw Detector & Thickness Gauge Flaw Detector & Thickness Gauge Flaw Detector: Sizing Toolkits: DAC, AWS, TCG, DGS. Exceptional visibility in sunlight (AMOLED) color VGA display (320x240 pixels). P.R.F. - 8 to 333 Hz, adjustable. Screen Refresh Rate: Adjustable 60 & 120 Hz. Detection: Z-Cross, Flank & Peak. Thickness Gauge: Automatic: probe zero, probe recognition, and Temperature compensation. Measurement: Variety of modes to address a number of applications. Large data storage with multiple formats: Alpha numeric grid and sequential w/auto identifier. Up to 12 hours of battery life using 3 AA cells. 4Windows PC interface software included.
Transcript

D A K O T A U L T R A S O N I C S

S O U N D S O L U T I O N S

The Best of Both Worlds...The Best of Both Worlds...The DFX-7 SeriesThe DFX-7 SeriesFlaw Detector & Thickness GaugeFlaw Detector & Thickness Gauge

Flaw Detector:▶ Sizing Toolkits: DAC, AWS, TCG, DGS.▶ Exceptional visibility in sunlight (AMOLED)

color VGA display (320x240 pixels).▶ P.R.F. - 8 to 333 Hz, adjustable.▶ Screen Refresh Rate: Adjustable 60 & 120 Hz.▶ Detection: Z-Cross, Flank & Peak.

Thickness Gauge:▶ Automatic: probe zero, probe recognition, and

Temperature compensation.▶ Measurement: Variety of modes to address a

number of applications.▶ Large data storage with multiple formats: Alpha

numeric grid and sequential w/auto identifier.▶ Up to 12 hours of battery life using 3 AA cells.4Windows PC interface software included.

D F X - 7 S E R I E S S P E C I F I C AT I O N SGeneralSize: 2.5W x 6.5H x 1.24D in (63.5 x 165 x 1.5mm).

Weight: 14 ounces (.397kgs), with batteries.

Case: Extruded aluminum body with nickel plated aluminum end caps (gasket sealed).

Display: 1/4 VGA AMOLED color display (320 x 240 pixels). Viewable area 1.7 x 2.27 in (43.2 x 57.6 mm). 16 color pallete, multiple color op-tions, and variable brightness.

Screen Refresh Rate: Selectable 60 or 120Hz.

Display Views: Flaw Detector: Full wave, +/- Rec-tified, or RF. Thickness Gauge: Digits, +/- Recti-fied, RF, or B-Scan.

Timing: Precision 25MHz TCXO with single shot 100 MHz 8 bit ultra low power digitizer.

Measurement Gates: Two independent gates (Flaw), and three gates (thickness). Start & width adjustable over full range. Amplitude 5-95%, 1% steps. Positive or negative triggering for each gate with audible and visual alarms.

Operating Temperature: 14 to 140F (-10C to 60C).

Environmental: Meets IP65 requirements.

CalibrationAutomatic Calibration: Longitudinal (straight), or Shear (angle).

Probe Types: Single Contact, Dual, Delay, and Angle .

Units: English (in), Metric (mm), or Time (µs).

Velocity: 0.0100 to .6300 in/µs (256–16,000 m/s).

Test Range: 0 to 0.280in (7.11mm) minimum, to 1200in (30,480mm) maximum at steel velocity. Continuously variable.

Zero Offset (Probe Zero): 0–999.999 µs.

Material Velocity Table: Contains longitudinal and shear velocities for a variety of material types.

PulserPulser Type: Two adjustable square wave pulsers and receivers.

P.R.F.: 8 to 333Hz in selectable steps (8, 16, 32, 66, 125, 250, 333Hz).

Pulser Voltage: 200 volt peak amplitude, rise/fall time < 10ns into 50ohm.

Pulse Width: 40 to 400 ns. Selectable step op-tions 40, 80 & 400 ns (labeled spike, thin & wide).& Thickness Gauge

ReceiverGain: 0 to 110dB with 0.2dB resolution. Manual and AGC control.

Damping: 50, 75, 100, 300, 600, & 1500 ohms.

Frequency Bands: DFX-7 & 7+: Broadband 1.8 - 19 MHz (-3dB). DFX-7+: Three narrow bands at 2MHz, 5MHz, 10MHz.

Horizontal Linearity: +/- 0.4% FSW.

Vertical Linearity: +/- 1% FSH.

Amplifier Linearity: +/- 1 dB.

Amplitude Measurement: 0 to 100% FSH, with 1% resolution.

Delay: 0 - 999in (25,375mm) at steel velocity.

Flaw Detector FeaturesTRIG: Trigonometric display of beam path, depth, surface distance, and curved surface cor-rection. Used with angle beam transducers.

DAC: Up to 8 points may be entered and used to digitally draw a DAC curve. Reference -2, -6, -10, (-6/-12), (-6/-14), (-2/-6/-10) dB. Amplitude displayed in %DAC, dB, or %FSH.

AWS: Automatic defect sizing in accordance with AWS D1.1 structural welding code.

AVG/DGS: Automatic defect sizing using probe data. Stores up to 64 custom setups.

TCG: Time corrected gain. 50 dB dynamic range, 20 dB per microsecond, up to 8 points for curve definition.

Measurement Mode: Pulse-Echo (P-E) measures from 0.025 in to 100 ft. (0.63mm to 3048 cm).

Auto-Cal: Provides automatic calibration with two reference points.

Detection Modes: Zero Crossing, Flank and Peak.

Display Freeze: Hold current waveform on screen.

Peak Memory: Captures peak signal amplitude.

Thickness Gauge FeaturesMeasurement Modes (Dual Element):

Pulse-Echo Mode (P-E) - (Pit & Flaw Detection) measures from 0.025 in to 100 ft. (0.63mm to 3048 cm).

Pulse-Echo Coating Mode (PECT) - (Material, Coating, Pit & Flaw Detection): Material: 0.025 in to 100 ft. (0.63mm to 3048 cm). Coating: 0.001 to 0.100 inches (0.01 to 2.54 millimeters).

Pulse-Echo Temp Comp Mode (PETP) - (Pit & Flaw Detection) Auto temperature compensation -measures from 0.025 in to 100 ft. (0.63 mm to 3048 cm).

Thickness Gauge (Features) - (Con’t)Echo-Echo Mode (E-E) - (Thru Paint & Coatings) measures from 0.050 to 4.0 inches (1.27 to 102 millimeters). Will vary based on coating.

Echo-Echo Verify (E-EV) - (Thru Paint & Coatings) measures from 0.050 to 1.0 inches (1.27 to 25.4 millimeters). Will vary based on coating.

Coating Only Mode (CT) - (Coating Thickness) Measures from 0.0005 to 0.100 inches (0.0127 to 2.54 millimeters). Range will vary +/- depending on the coating.

One and two point calibration option for mate-rial & coating, or selection of basic material types.

Auto probe zero, recognition and temperature compensation.

High speed scan up to 50 readings per second.

Audible alarm with hi/lo limits.

Built-in differential mode for QC inspections.

64 custom setup configurations.

MemoryLog Formats: Grid (Alpha Numeric), or Sequen-tial (Auto Identifier).

Graphics On: 8,000 readings, A/B Scan image, & all gauge settings for every reading.

Graphics Off: 210,000 readings (coating, mate-rial, min & max. (Thickness gauge only).

Custom Setups: 64 user configurations.

Power SourceBattery: Three 1.5V alkaline, 1.2V AA Nicad cells, 1.2V AA NI-MH, or other other equivalent power source. Battery life (continuous use): Alkaline (12 hrs), Nicad (5hrs), and NI-MH (12hrs), with default settings.

ConnectionsOutput: RS232 serial interface. PC software & USB converter cable included.

Transducer Connectors: Two LEMO 00 connec-tors.

CertificationThickness Gauge: Factory calibration traceable to NIST & MIL-STD-45662A.

Flaw Detector: EN12668-1 compliant.

Warranty2 year limited

M A D E I N T H E U S ADistributed by:

DAKOTA ULTRASONICS1500 Green Hills Road, #107Scotts Valley, CA 95066Ph: (831) 431-9722Fax: (831) 431-9723Website: www.dakotaultrasonics.comEmail: [email protected]

D A K O T A U L T R A S O N I C S

S O U N D S O L U T I O N S

One on hand equalstwo on the job!

One on hand equalstwo on the job!

The DFX-8 SeriesThe DFX-8 SeriesFlaw Detector & Thickness GaugeFlaw Detector & Thickness Gauge

Flaw Detector:▶ Sizing Toolkits: DAC, AWS, TCG, DGS.▶ Exceptional visibility in sunlight (Blanview) transmissive color QVGA display (320x240 pixels).▶ P.R.F. - 8 Hz to 2 kHz, adjustable.▶ Screen Refresh Rate: 60 Hz.▶ Detection: Z-Cross, Flank & Peak.

Thickness Gauge:▶ Automatic: probe zero, probe recognition, and Temperature compensation.

▶ Measurement: Variety of modes to address a number of applications.

▶ Large data storage with multiple formats: Alpha numeric grid

and sequential w/auto identifier.

▶ Li-Ion pack up to 18 hrs, emergency backup 6

AA cells up to 12 hrs.▶ Windows PC

interface software included.

▶ 2 year warranty.

D F X - 8 S E R I E S S P E C I F I C AT I O N SGeneralSize: 8.5W x 6.5H x 2.5D in (216 x 165 x 70mm).

Weight: 4.5 lbs (2.04 kgs), with batteries.

Case: Extruded aluminum body with nickel plated aluminum end caps (gasket sealed).

Display: Blanview sunlight readable QVGA TFT color display (320 x 240 pixels). Viewable area 4.54 x 3.40 in (115.2 x 86.4 mm), or 5.7 in (144.78 mm) diagonal. 16 color pallete, multiple color options and variable brightness.

Screen Refresh Rate: 60Hz.

Display Views: Flaw Detector: Full wave, +/- Rectified, or RF. Thickness Gauge: Digits, +/- Rectified, RF, or B-Scan.

Timing: Precision TCXO timing with single shot 100 MHz 8 bit ultra low power digitizer.

Measurement Gates: Two independent gates (Flaw), and three gates (thickness). Start & width adjustable over full range. Amplitude 5-95%, 1% steps. Positive or negative triggering for each gate with audible and visual alarms.

Operating Temperature: 14 to 140F (-10C to 60C).

Environmental: Meets IP65 requirements.

PulserPulser Type: DFX-8 Two adjustable square wave puls-ers. DFX-8+ Two tone burst pulsers.

P.R.F.: 8 to 2000Hz in selectable steps (8, 16, 32, 66, 125, 250, 333, 1000, 2000 Hz).

Pulser Voltage: DFX-8 100 - 200 volt peak amplitude, rise/fall time < 10ns into 50ohm. DFX-8+ 100 - 400v.

Pulse Width: 40 to 400 ns. Selectable step options 40, 80 & 400 ns (labeled spike, thin & wide).

ReceiversGain: 0 to 110dB with 0.2dB resolution. Manual and AGC control.

Damping: 50, 75, 100, 300, 600, & 1500 ohms.

Frequency Bands: DFX-8 & DFX-8+ Broadband 1.8 - 19 MHz (-3dB). Four narrow bands at 1MHz, 2MHz, 5MHz, 10MHz. DFX-8+ Additional narrow bands at .5MHz, 15MHz.

Horizontal Linearity: +/- 0.4% FSW.

Vertical Linearity: +/- 1% FSH.

Amplifier Linearity: +/- 1 dB.

Amplitude Measurement: 0 to 100% FSH, with 1% resolution.

Delay: 0 - 999in (25,375mm) at steel velocity.

Power SourceLithium Ion Pack: 10.8v, 2 amp hrs, typical operation 18hrs.

Battery Backup: Emergency battery backup. Six 1.5V alkaline, 1.2V AA Nicad cells, 1.2V AA NI-MH, or other other equivalent power source. Battery life (continuous use): Alkaline (12 hrs), Nicad (5hrs), and NI-MH (12hrs), with default settings.

ConnectionsUSB: Direct USB 1.1 PC connectivity.

Power Connector: 12v @ 2amps, adapter 100-240 VAC, .7 Amps, 50-60 Hz.

5 Pin Lemo (includes):

RS232Output: RS232 PC serial interface. DFX-8+ For use with B-Scan encoders (crawlers).

Alarm Outputs: Two independent alarm outputs trig-gered by the gates.

Analog Out: DFX-8+ Proportional outputs (amplitude or distance), 0-10 volts.

Transducer Connectors: Two LEMO 00 connectors.

CalibrationAutomatic Calibration: Longitudinal (straight), or Shear (angle).

Probe Types: Single Contact, Dual, Delay, and Angle .

Units: English (in), Metric (mm), or Time (µs).

Velocity: 0.0100 to .6300 in/µs (256–16,000 m/s).

Test Range: 0 to 0.076in (1.93mm) minimum, to 1200in (30,480mm) maximum at steel velocity. Continuously variable.

Zero Offset (Probe Zero): 0–999.999 µs.

Material Velocity Table: Contains longitudinal and shear velocities for a variety of material types.

MemoryLog Formats: Grid (Alpha Numeric), or Sequential (Auto Identifier).

Capacity: 4 Gb internal & up to 64 Gb External SD slot.

Screen Capture: bitmap graphic capture for quick documentation.

Custom Setups: 64 user configurations.

VideoRemote Commander: Java PC software allows remote display and control for training/presentation purposes, and custom system integration.

Flaw Detector FeaturesTRIG: Trigonometric display of beam path, depth, surface distance, and curved surface correction. Used with angle beam transducers.

DAC: Up to 8 points may be entered and used to digi-tally draw a DAC curve. Reference -2, -6, -10, (-6/-12), (-6/-14), (-2/-6/-10) dB. Amplitude displayed in %DAC, dB, or %FSH.

AWS: Automatic defect sizing in accordance with AWS D1.1 structural welding code.

AVG/DGS: Automatic defect sizing using probe data. Stores up to 64 custom setups.

TCG: Time corrected gain. 50 dB dynamic range, 20 dB per microsecond, up to 8 points for curve definition.

Measurement Mode: Pulse-Echo (P-E) measures from 0.025 in to 100 ft. (0.63mm to 3048 cm).

Auto-Cal: Provides automatic calibration with two reference points.

Detection Modes: Zero Crossing, Flank and Peak.

Flaw Detector Features (Cont’d)Display Freeze: Holds current waveform on screen.

Peak Memory: Captures peak signal amplitude.

Auto Interface Gate - DFX-8+ Automatic adjustment of interface gate for immersion testing (water path).

Thickness Gauge FeaturesMeasurement Modes (Dual Element):Pulse-Echo Mode (P-E) - (Pit & Flaw Detection) mea-sures from 0.025 in to 100 ft. (0.63mm to 3048 cm).

Pulse-Echo Coating Mode (PECT) - (Material, Coat-ing, Pit & Flaw Detection): Material: 0.025 in to 100 ft. (0.63mm to 3048 cm). Coating: 0.001 to 0.100 inches (0.01 to 2.54 millimeters).

Pulse-Echo Temp Comp Mode (PETP) - (Pit & Flaw Detection) Auto temperature compensation -measures from 0.025 in to 100 ft. (0.63 mm to 3048 cm).

Echo-Echo Mode (E-E) - (Thru Paint & Coatings) mea-sures from 0.050 to 4.0 inches (1.27 to 102 millimeters). Will vary based on coating.

Echo-Echo Verify (E-EV) - (Thru Paint & Coatings) mea-sures from 0.050 to 1.0 inches (1.27 to 25.4 millimeters). Will vary based on coating.

Coating Only Mode (CT) - (Coating Thickness) Mea-sures from 0.0005 to 0.100 inches (0.0127 to 2.54 mil-limeters). Range will vary +/- depending on the coating.

One and two point calibration option for material & coating, or selection of basic material types.

Auto probe zero, recognition and temperature com-pensation.

High speed scan up to 50 readings per second.

Audible alarm with hi/lo limits.

Built-in differential mode for QC inspections.

64 custom setup configurations.

TransducersDelay line: High Frequency single element delay line style for precision testing of thin materials.

Pencil: High Frequency single element delay line style for testing of materials in tight access areas and difficult geometries.

Contact: Single element contact style for general pur-pose longitudinal & Shear wave flaw detection.

Dual: Pitch/Catch dual element style for longitudinal & Shear wave corrosion inspections..

CertificationThickness Gauge: Factory calibration traceable to NIST & MIL-STD-45662A.

Flaw Detector: EN12668-1 compliant.

Warranty2 year limited

M A D E I N T H E U S ADistributed by:

DAKOTA ULTRASONICS1500 Green Hills Road, #107Scotts Valley, CA 95066Ph: (831) 431-9722Fax: (831) 431-9723Website: www.dakotaultrasonics.comEmail: [email protected]

D A K O T A U L T R A S O N I C S

S O U N D S O L U T I O N S

DFX Flaw Detector SeriesDFX Flaw Detector SeriesIsn’t performance what you need?

The DFX series ultrasonic flaw detectors are fullfeatured digital scopes designed to handle the simplest to the most difficult applications. Whether the application requires high penetra-tion for materials that are difficult to measure, or high resolution for precision aerospace parts, the DFX series flaw detectors are equipped with the necessary features to do the job.

▶ Transflective Color TFT Display.

▶ Broad & Narrow Band Amplifier.

▶ Square Wave Pulser (Active EdgeTM).

▶ Extruded Aluminum Case(IP65).

▶ USB / RS232 / Video / Analog Outputs.

▶ Trig Functions.

▶ Curved Surface Correction.

▶ DAC/TCG/AWS/DGS (Defect Sizing).

▶ Automatic Calibration.

▶ Long Battery Life (Li-Ion Pack).

▶ Water Path Correction (638 only).

▶ Fast Setup Screen Facility.

▶ Alphanumeric Text.

▶ Help Facility.

DFX series can go where you go, to do the work you

do, saving you time and money, making your

inspections fast and accurate.

D F X 6 1 5 / 6 2 5 / 6 3 5 / 6 3 8 S P E C I F I C AT I O N SPhysicalSize: 9.25W x 7.13H x 4.0D in (235W x 181H x 102D mm).

Weight: 7.5 lbs. (3,4 Kgs) with Li-Ion cells.

Case: Extruded Aluminum.

Display: Transflective TFT Color.

Display area: 4.39in x 3.29in (111.4mm x 83.5mm) 320 x 240 pixels.

A-Scan area: 255 x 200 pixels (315 x 200 expand-ed), 8 color options and variable brightness.

Temperature:Operating –10° to +55°C or 14° to 131°F –20° to +70°C or –4° to 158°F (survivable).Storage –40° to +75°C or –40° to 167°F

Environmental: Meets IP65 requirements.

SpecificationsUnits: English (in), Metric (mm), or Time (μs).

Probe Zero: 0–999.999μs.

Velocity: 615 & 625 0.0393–0.3937in/μs (1000–9999m/s). 635 & 638 0.0100in/μs (256–16000m/s).

Test Range: 615 & 625 0–0.2in (5mm) up to 0–400in (10,000mm) at steel velocity. Variable in sequence, 0.4 in or 0.04 in (10mm or 1mm). 635 & 638 0–0.05 in (1 mm) up to 0–800 in (20,000mm) at steel velocity. Variable in 1, 2, 5 sequence or continuously in 0.05 in (1mm) incre-ments. Also from 1 to 5000 (μs).

Test Modes: Pulse-echo and transmit/ receive.

Gates: 615 & 625 Start & Width adjustable over full range. Amplitude 0–100%, 0.5% steps. Visual & audible alarms. Single gate positive trigger, and two gates positive and negative triggering. Gate 2 has selectable 0.6 second delay on alarm. 635 & 638 Two fully independent gates with positive and negative triggering for each gate.

Gate Expansion: 635 & 638 Expands range to width of Gate 1.

Gate Monitor Delay: 635 & 638 Selectable0.6 sec delay on gate 2 negative monitor.

Measurement Modes:

Signal Monitor - 635 & 638. Depth - Depth and amplitude of signal in gate.

Echo Echo - Echo-Echo distance, automatic gate 2 position.

Gate to Gate - 625 Echo-Echo distance, manual gate 2 position. 635 & 638 Independent gates.

Trig - Trigonometric display of beam path, depth and surface distance. Calculation of skip depth and curve surface compensation, X-offset for transducer.

T-Min - Holds minimum thickness in depth mode.

System Linearity: Vertical = 1% Full Screen Height (FSH) Amplifier Accuracy ±0.1%dB. Hori-zontal ±0.4% Full Screen Width (FSW).

Pulser Voltage: 615 & 625 200 volt peak ampli-tude, rise/fall time < 10ns into 50ohm. 635 & 638 100V–350V (450V 638 ) square wave pulser. Pulse width from spike to 2000ns duration—rise/fall times < 5ns into 50 ohms.

Pulser Width: 615 fixed at 100ns. 625 30-250ns linked to filter band. 635 & 638 Adjustable in 2% of nominal width, minimum 1ns maximum 40ns.

ActiveEdge: Unique active pulse control for enhanced near surface resolution and signal response. Replaces traditional damping control.

P.R.F.: 615 & 625 Selectable 35 to 1000Hz(635 & 638 5000Hz), 5Hz steps.

Screen Update Rate: 50 or 60Hz.

Rectification: Full wave, positive or negative half wave and unrectified RF.

Delay: 615 & 625 0–400 in (10,000mm) 635 & 638 0–800in (20,000mm) at steel velocity in 0.02 steps (0.05mm).

Gain: 0 to 110dB. Adjustable in 0.5, 2, 6, 14 and 20dB steps.

Frequency Bands: 615 4Broadband 1–10 MHz (-6dB). 625 4 Narrow Bands centered at 1MHz, 2MHz, 5MHz, & broadband 1.5–15MHz. 635 & 638 6 narrow bands centered at 0.5, 1, 2.25, 5, 10 & 15 MHz. Broad band at 2 to 22 MHz (-6dB) and 1 to 35 MHz (-20dB).

Vertical Linearity: 1% full screen height. ampli-fier Linearity: +/- 0.1 dB. Horizontal Linearity: +/- 0.4% full screen width (FSW).

Reject: 615 & 625 50% suppressive reject. LED warning when selected. 635 & 638 80% linear reject.

MemoryThickness Logging: Storage for 8000 readings stored in Block/Location/Number coding or alpha-numeric pre-programmed work sheets. Transferable to Excel using optional PC software.

Panel Memory: 100 storage locations for calibra-tion settings.

A-Scan Memory: 800 waveforms.

FeaturesWave Smoothing: Produces a smoothsignal envelope.

AGC: Automatic Gain Control sets selected echo to a user defined level (10–90%).

DAC: Up to 10 points may be entered and used to digitally draw a DAC curve. Reference -2, -6, -10, -12, -14 dB curves can be selected for JIS, ASME and EN1714 codes.

AWS: Automatic defect sizing in accordance with AWS D1.1 Structural Welding Code.

API: Automatic defect sizing in accordance with API 5UE.

AVG/DGS: Automatic defect sizing using probe data. 10 probe data sets can be stored.

TCG: 625, 635 & 638 Time corrected gain. 40 dB dynamic range, 30 dB per microsecond, up to 10 points for curve definition.

Auto-Cal: Provides Automatic calibration with two echoes.

Reference Waveform: Recalled waveform can be shown in a different color to live waveform for direct comparision.

Display Freeze: Hold current waveform on screen.

Peak Memory: For echodynamic pattern deter-mination.

Online Help: Instant operator guidance on operation accessed from direct key.

Language Support: Six user selectable lan-guages from: English, German, French, Spanish, Dutch, Italian, Russian, Polish, Czech, Finnish & Hungarian. Others available on request.

Power SourceBattery: Lithium Ion battery pack 14.4V,5.0 Amp hrs. Minimum 11 hours use, typical15 hours, indication of battery charge. Recharge time 4 hours.

Charger: 100–240 VAC, 50–60Hz.

ConnectionsUSB: For connection to PC, keyboard and printer.

Outputs: Serial Interface, composite video (NTSC & PAL), analog output for amplitude and distance updated at PRF rate. Transmitter sync output.

Transducer Connectors: available with BNC or LEMO 1 connectors (factory option).

Additional 638 FeaturesInterface trigger: Interface gate locks to surface echo and eliminates water path variation.

High Power Pulser: 450 volt boost.

Warranty & Certification2 year limited / EN12668-1 compliant.

M A D E I N T H E U S ADistributed by:

DAKOTA ULTRASONICS1500 Green Hills Road, #107Scotts Valley, CA 95066Ph: (831) 431-9722Fax: (831) 431-9723Website: www.dakotaultrasonics.comEmail: [email protected]

D A K O T A U L T R A S O N I C S

S O U N D S O L U T I O N S

MVXA/B Scan Ultrasonic

Thickness Gauge

▶ Powered by a 20MHz platform with ultra low power 10 bit digitizer.▶ 1/8” VGA grayscale display (240 x 160 pixels). Screen Refresh rate of 25 Hz.▶ Manual or AGC gain option, depending on measure mode selected, 40 dB gain control.▶ Display veiws: RF, +/- Rectified, B-Scan (cross section), or Large Digits.▶ Two independent gates with shared threshold.▶ Measure modes: (P-E) pulse-echo (flaws & pits), (P-E GT) pulse-echo (flaws & pits w/blanking gate & treshold), and (E-E) echo-echo (thru-paint).▶ Data Storage: 12,000 page capacity (each page contains waveform, measurement, and all gauge settings).▶ Windows® PC interface software.▶ Multiple language support.▶ 2 year limited warranty.

MVXA/B Scan Ultrasonic

Thickness Gauge

M V X S P E C I F I C AT I O N S

PhysicalWeight:13.5 ounces (with batteries).

Size:2.5 W x 6.5 H x 1.24 D inches(63.5 W x 165 H x 31.5 D mm).

Operating Temperature:-14˚ to 140˚F (-10˚ to 60˚C).

Keyboard:Membrane switch with twelve tactile keys.

Case:Extruded aluminum body with nickel-plat-ed aluminum end caps (gasket sealed).

Data Output:Bi-directional RS232 serial port; Windows® PC interface software.

Display:1/8in VGA grayscale display (240 x 160 pix-els); viewable area 2.4 x 1.8in (62 x 5.7mm); EL backlit (on/off/auto invert).

Ultrasonic SpecificationsMeasurement Modes:Pulse-Echo (flaws, pits).Echo-Echo (thru-paint).

Pulser:Square wave pulser with adjustable pulse width (spike, thin, wide).

Receiver:Manual or AGC gain control with 40dB range, depending on mode selected.

Timing:20 MHz with ultra low power 10 bit digi-tizer.

DisplayDisplay Views:

A-Scan: Rectified +/- (flaw view) RF (full waveform view).

B-Scan: Cross section view; display speed of 15 secs per screen.

Large Digits: Standard thickness view; Digit Height: 0.400 in (10mm).

Scan Bar: Thickness 6 readings per second; Viewable in B-Scan and Large Digit views.

Bar Graph: Indicates stability of measure-ment.& Thickness Gauge

Power SourceThree 1.5V alkaline or 1.2V NiCad AA cells.

Typically operates for 35 hours on alkaline and 10 hours on NiCad (charger not included).Auto power off if idle 5 minutes.

Battery status icon.

MeasuringRange:

Pulse-Echo Mode: (Pit & Flaw Detection) measures from 0.025 9.999 inches (0.63 to 254 millimeters).

Echo-Echo Mode: (Thru Paint & Coatings) measures from 0.050 to 4.0 inches (1.27 to 102 millimeters). Range will vary +/- de-pending on the thickness of coating.

Flaw Prove Up: Basic mode using angle beam transducers.

Resolution: +/- .001 inches (0.01 mm).

Velocity Range:.0492 to .3936 inches/ms1250 to 9999 meters/sec

Single and Two point calibration option, or selection of basic material types.

Units: English & Metric

TransducerTransducer Types:

Dual Element (1 to 10 MHz).

Single Element (1 to 10 MHz) - converter cable required.

Locking quick disconnect LEMO “00” con-nectors.

Standard 4 foot cable.

Custom transducers and cable lengths available for special applications.

MemoryData Structure:

Grid (alpha numeric)

Cell contents:

12,000 readings , A or B Scan image, & gauge settings for every reading.

OBSTRUCT to indicate inaccessible loca-tions.

Memory:

16 megabit non-volatile ram.

Features:Setups:

64 custom user-definable setups; Factory setups can also be edited by the user.

Gates:

Single gate in pulse-echo mode, or single gate with holdoff in echo-echo mode; Adjustable threshold.

Selectable Transducers:

Selectable transducer types with built-in dual path error correction for improved linearity.

Alarm Mode:

Set Hi and Lo tolerances with audible beeper and visual LEDs.

Fast-Scan Mode:

Takes 32 readings per second and displays the minimum reading found when the transducer is removed.

CertificationFactory calibration traceable to NIST & MIL-STD-45662A.

Warranty2 year limited

M A D E I N T H E U S ADistributed By:

DAKOTA ULTRASONICS1500 Green Hills Road, #107Scotts Valley, CA 95066Ph: (831) 431-9722Fax: (831) 431-9723Website: www.dakotaultrasonics.comEmail: [email protected]

FixedVelocity

ThicknessGauges

D A K O T A U L T R A S O N I C S

S O U N D S O L U T I O N S

MXSeries

MXSeries

FixedVelocity

ThicknessGauges

The MX-2 is ideal for theperson that tests a few differentmaterials regularly. The MX-2 con-tains 8 velocities of commonmaterials and 2 velocities thatthe user can set using a simplesoftware program.

The MX-1 and MX-2 arepackaged in all metal cases,sealed with gaskets to protectthem from harsh working envi-ronments. These gauges can gowhere you go, to do the work you do, saving you time and money, making fast accurate measure-ments.

Dakota Ultrasonics offers quality gauges that are reliable, innova-tive and competitively priced - all backed by the longest warranty in the business—5 full years.

The MX-1 and MX-2 are fixedvelocity gauges making themaffordable precision tools thatrequire no special training tooperate.

The MX-1 will hold a singlevelocity, making it the perfect choice where one material is being tested day after day. With the preset velocity in the gauge, you simply turn on the gauge and you are ready to work. The velocity is set in the MX-1 with a simple software program that is suppliedwith the unit.

M X - 1 & M X - 2 T H I C K N E S S G A U G E S

PhysicalWeight:10 ounces (with batteries).

Size:2.5 W x 4.5 H x 1.24 D inches(63.5 W x 114.3 H x 31.5 D mm).

Operating Temperature:-20 to 120F (-30 to 50C).

Case:Extruded aluminum bodywith nickel-plated aluminumend caps (gasket sealed).

KeypadSealed membrane that isresistant to both water andpetroleum products.

Four or five tactile-feedbackkeys.

TransducerDual-element (transmit andreceive).

1 to 10 MHz frequency range.

Locking quick disconnectLEMO connectors.

4 foot cable.

Custom transducers availablefor special applications.& Thickness Gauge

Power SourceTwo 1.5V alkaline or1.2V NiCad AA cells.

Typically operates for80 hours on alkaline and20 hours on NiCad.

Display flashes when battery islow. Unit turns off automaticallywhen battery is too low tooperate reliably.

DisplayMulti-function 4.5 digit liquidcrystal display with 0.500 inchnumerals, backlit for use inpoor light conditions.

Backlight is selectable on/off/auto(illuminates only when taking ameasurement).

Bar graph indicates stabilityof reading.

Computer InterfaceRS232 serial interface. PC software & USB converter cable included.

CertificationFactory calibration traceable to NIST & MIL-STD-45662A.

Warranty5 year limited.

MeasuringRange:Measures from 0.025 to 19.999inches (0.63 to 500 millimeters).Range dependent on materialand transducer type.

Units:English & Metric

Resolution:0.001 inches (0.01 millimeters)

Velocity Range:0.0492 to .5511 in/μs.(1250 to 14,000 m/sec)

Four readings per second forsingle point measurements.

Velocity Settings:The MX-1 has 1 customvelocity setting.

The MX-2 has 8 factory presetsand 2 programmable customvelocity options. The presetvelocities are as follows:

1. Steel 4340 .233 2. Stainless Steel 303 .223 3. Aluminum 2024 .251 4. Cast Iron .179 5. Plexiglass .106 6. PVC .094 7. Polystyrene .092 8. Polyurethane .070 9. Custom10. Custom

M A D E I N T H E U S ADistributed by:

DAKOTA ULTRASONICS1500 Green Hills Road, #107Scotts Valley, CA 95066Ph: (831) 431-9722Fax: (831) 431-9723Website: www.dakotaultrasonics.comEmail: [email protected]

DAKOTA ULTRASONICS rugged thickness gauge family has been designed and built to satisfy the roughest industry conditions. The features offered in the MX Series allow the user to select a quality tool that will meet or exceed their specific application needs. Our 5 year limited warranty indicates how we feel about the reliability and durability of the MX Series.

S P E C I F I C A T I O N S

TheMX SeriesfromDAKOTAULTRASONICS

D A K O T A U L T R A S O N I C S

S O U N D S O L U T I O N S

TheStandardHas Been

Set!

TheStandardHas Been

Set!

PACKAGING All Metal Case.The units are sealed with gaskets to protectthem from harsh working environments.

SIZE 2.5”W x 4.5”H x 1.2”D. The units are small andportable and will operate on 2 AA batteries for 80 hours.Very compact and convenient to transport to any job.

FEATURES The MX Series offers the features you haverequested; Differential Mode, High Speed Scan Mode,Alarm Mode and On-Board Data Logging.

WARRANTY All gauges are covered by a 5 yearlimited warranty.

SUPPORT Dakota excels in in this area. OurEngineering and technical staff, combined with a knowledgable and friendly sales staff, are always here to help out.

AS YOU CAN SEE, the standard has been set withthe MX SERIES. Please give us a call to discuss yourapplications.

TheMX SeriesfromDAKOTAULTRASONICS

M X S E R I E S T H I C K N E S S G A U G E S

PhysicalWeight:10 ounces (with batteries).

Size:2.5 W x 4.5 H x 1.24 D inches(63.5 W x 114.3 H x 31.5 D mm).

Operating Temperature:-20 to 120F (-30 to 50C).

Case:Extruded aluminum bodywith nickel-plated aluminumend caps (gasket sealed).

KeypadSealed membrane that isresistant to both water andpetroleum products.

Six or nine tactile-feedback keys.

TransducerDual-element (transmit andreceive).

1 to 10 MHz frequency range.

Locking quick disconnectLEMO connectors.

4 foot cable.

Custom transducers availablefor special applications.& Thickness Gauge

Power SourceTwo 1.5V alkaline or1.2V NiCad AA cells.

Typically operates for80 hours on alkaline and20 hours on NiCad.

Display flashes when battery islow. Unit turns off automaticallywhen battery is too low tooperate reliably.

DisplayMulti-function 4.5 digit liquidcrystal display with 0.500 inchnumerals, backlit for use inpoor light conditions.

Backlight is selectable on/off/auto(illuminates only when taking ameasurement).

Bar graph indicates stabilityof reading.

SoftwareComes complete with cable andPC software (MX-5 DL only).

MemoryNumeric data storage, 10 files of 100 readings per file, for 1000 read-ings (MX-5 DL only).

CertificationFactory calibration traceable to NIST & MIL-STD-45662A.

Warranty5 year limited.

MeasuringRange:Measures from 0.025 to 19.999inches (0.63 to 500 millimeters).Range dependent on materialand transducer type.

Units:English & Metric

Resolution:0.001 inches (0.01 millimeters)

Velocity Range:0.0492 to .5511 in/μs.(1250 to 14,000 m/sec)

Four readings per second forsingle point measurements, or16 per second in Scan Mode -captures the minimum thickness.

Features (MX-5, MX-5-DL)Differential Mode:Enter acceptable thickness value,unit will display+/- the differencefrom the value entered.

Alarm Mode:Enter a minimum acceptablethickness value. If measurementfalls below minimum, red LED willilluminate and sound beeper. Ifmeasurement is above minmumvalue, green LED will illuminate.

Data Output:RS232 serial interface. PC software & USB converter cable included.

M A D E I N T H E U S ADistributed by:

DAKOTA ULTRASONICS1500 Green Hills Road, #107Scotts Valley, CA 95066Ph: (831) 431-9722Fax: (831) 431-9723Website: www.dakotaultrasonics.comEmail: [email protected]

DAKOTA ULTRASONICS rugged thickness gauge family has been designed and built to satisfy the roughest industry conditions. The features offered in the MX Series allow the user to select a quality tool that will meet or exceed their specific application needs. Our 5 year limited warranty indicates how we feel about the reliability and durability of the MX Series.

S P E C I F I C A T I O N S

Versatility & Convenience

D A K O T A U L T R A S O N I C S

S O U N D S O L U T I O N S

The MMX Series has arrived! These gauges are the answer toboth versatility and convenience in every way! The MMX series

defines versatility by combining features like: Fast Scan, Alarm(audible/visual), and data logging, along with multi-mode for

reading through and eliminating the thicknessof painted or coated materials. The MMX demonstrates

convenience by making all these features available to theuser with just the touch of a button.

Conveniently toggle between pulse-echo mode(detect flaws & pits), and echo-echo mode

(eliminate paint & coating thickness), to cover allyour inspection requirements. No need

to remove or allow for errors due to paintand coatings. The MMX gauges use a highlydamped dual element style transducer for both

modes, eliminating the need to switch betweentransducer types.

The MMX gauges are packaged in an all alu-minum sealed case, making them rugged

and resistant to the working environ-ment—Dakota’s trademark.

5 Year Limited Warranty

TheMMX-6 SeriesVersatility & Convenience

TheMMX-6 Series

M M X S E R I E S T H I C K N E S S G A U G E S

PhysicalWeight:

10 ounces (with batteries).

Size:

2.5 W x 4.5 H x 1.24 D inches(63.5 W x 114.3 H x 31.5 D mm).

Operating Temperature:

-20 to 120 F (-30 to 50 C).

Case:

Extruded aluminum body with nickel plated aluminum end caps (gasket sealed).

KeypadSealed membrane that is resistant to both water and petroleum products.

Nine tactile-feedback keys.

TransducerDual-element (transmit and receive).

1 to 10 MHz frequency range.

Special high-damped transducersavailable for Through Paint & Coating mea-surements.

Locking quick disconnect LEMO connec-tors.

4 foot cable.

Custom transducers available for special applications.

Power SourceTwo 1.5V alkaline or1.2V NiCad AA cells.

Typically operates for 60 hours on alkaline and 15 hours on NiCad.

Display flashes when battery is low. Unit turns off automatically when battery is too low to operate reliably. ckness Gauge

DisplayMulti-function 4.5 digit liquid crystal dis-play with 0.500 inch numerals, backlit for use in poor light conditions.

Backlight is selectable on/off/auto (illumi-nates only when taking a measurement).

Measurements displayed in inches, inches/microsecond, millimeters, and meters/second.

Bar graph indicates stability of reading.

Internal Data Logger(MMX-6 DL only)

Numeric Data Logging (10 files of 100 read-ings per file, total 1000 readings.

OBST indicates no reading.

FeaturesMulti-Mode:

Toggle between Pit & Flaw Detection(Pulse Echo Mode), and Through Paint& Coatings (Echo-Echo Mode) with the single press of a button.

Alarm Mode:

Enter a minimum acceptable thickness value. If measurement falls below mini - mum, red LED will illuminate and sound beeper. If measurement is above minimum value, green LED will illuminate.

Data Output:

RS232 serial interface. PC software & USB converter cable included.

MeasuringRange:

Pulse-Echo Mode (Pit & Flaw Detection): measures from 0.025 to 19.999 inches (0.63 to 500 millimeters).

Echo-Echo Mode (Through Paint & Coat-ings): measures from 0.1 to 1.0 inches (2.54 to 25.4 mm).

Extended range (ER) (Increases Echo-Echo Range): measures from 0.125 to 4.0 inches (3 to 100 mm).

Switch to select English or Metric units.

Resolution:

0.001 inch (0.01 millimeter).

Velocity Range:

0.0492 to .3937 inches./µs.( 1250 to 10,000 meters/second).

Built in:

Stainless steel reference disk for probe zeroing.

Speed:

Standard - 4 per second.Scan - 16 per second (min. capture).

Single and two point calibration option included.

CertificationFactory calibration traceable to NIST & MIL-STD-45662A.

Warranty5 year limited.

M A D E I N T H E U S ADistributed by:

DAKOTA ULTRASONICS1500 Green Hills Road, #107Scotts Valley, CA 95066Ph: (831) 431-9722Fax: (831) 431-9723Website: www.dakotaultrasonics.comEmail: [email protected]

DAKOTA ULTRASONICS’ has succeeded in adding the final touches to the rugged MX line with the multi-mode MMX series. The MMX-6 and MMX-6DL have not only been combined with all the durability and features of the MX series gauges, but have also been equipped with the ability to measure through painted or coated materials and eliminate the thickness of the paint or coat-ing. Our 5 year limited warranty indicates how we feel about the reliability and durability of the MMX Series.

APPLICATIONS: Corrosion & Pitting • Tube & Pipe • Tanks • Boilers • Glass • Variety of Applications

S P E C I F I C A T I O N S

D A K O T A U L T R A S O N I C S

S O U N D S O L U T I O N S

MMX-7Multi-Mode Thickness

Gauge with B-Scan▶ Powered by a 20MHz platform with ultra low power 10 bit digitizer.▶ 1/8” VGA grayscale display (240 x 160 pixels). Screen Refresh rate of 25 Hz.▶ Measure modes: (P-E) pulse-echo (flaws & pits) and (E-E) echo-echo (thru-paint).▶ Adjustable gain option; high, med, or low in (P-E) mode, or automatic gain control (AGC) in (E-E) mode.▶ Display veiws: Large Digits or B-Scan (cross section).▶ Store up to 64 custom setups. Selection of factory setups included.▶ High speed scan: 32 readings per second.▶ Alarm feature: Hi and Lo limit options.▶ Data Storage: 12,000 page capacity.▶ Windows® PC interface software.▶ Multiple language support.▶ 2 year limited warranty.

MMX-7Multi-Mode Thickness

Gauge with B-Scan

M M X - 7 S P E C I F I C AT I O N S

PhysicalWeight:13.5 ounces (with batteries).

Size:2.5 W x 6.5 H x 1.24 D inches(63.5 W x 165 H x 31.5 D mm).

Operating Temperature:-14˚ to 140˚F (-10˚ to 60˚C).

Keyboard:Membrane switch with twelve tactile keys.

Case:Extruded aluminum body with nickel-plat-ed aluminum end caps (gasket sealed).

Data Output:Bi-directional RS232 serial port; Windows® PC interface software.

Display:1/8in VGA grayscale display (240 x 160 pix-els); viewable area 2.4 x 1.8in (62 x 5.7mm); EL backlit (on/off/auto invert).

Ultrasonic SpecificationsMeasurement Modes:Pulse-Echo (flaws, pits).Echo-Echo (thru-paint).

Pulser:Square wave pulser with adjustable pulse width (spike, thin, wide).

Receiver:Manual or AGC gain control with 40dB range, depending on mode selected.

Timing:20 MHz with ultra low power 10 bit digi-tizer.

DisplayDisplay Views:

Large Digits: Standard thickness view; Digit Height: 0.400 in (10mm).

B-Scan: Cross section view; display speed of 15 secs per screen.

Scan Bar: Thickness 6 readings per second; Viewable in B-Scan and Large Digit views.

Bar Graph: Indicates stability of measure-ment.& Thickness Gauge

Power SourceThree 1.5V alkaline or 1.2V NiCad AA cells.

Typically operates for 35 hours on alkaline and 10 hours on NiCad (charger not included).Auto power off if idle 5 minutes.

Battery status icon.

MeasuringRange:

Pulse-Echo Mode: (Pit & Flaw Detection) measures from 0.025 9.999 inches (0.63 to 254 millimeters).

Echo-Echo Mode: (Thru Paint & Coatings) measures from 0.050 to 4.0 inches (1.27 to 102 millimeters). Range will vary +/- de-pending on the thickness of coating.

Flaw Prove Up: Basic mode using angle beam transducers.

Resolution: +/- .001 inches (0.01 mm).

Velocity Range:.0492 to .3936 inches/ms1250 to 9999 meters/sec

Single and Two point calibration option, or selection of basic material types.

Units: English & Metric

TransducerTransducer Types:

Dual Element (1 to 10 MHz).

Single Element (1 to 10 MHz) - converter cable required.

Locking quick disconnect LEMO “00” con-nectors.

Standard 4 foot cable.

Custom transducers and cable lengths available for special applications.

MemoryData Structure:

Grid (alpha numeric)

Cell contents:

12,000 readings, B Scan image & gauge set-tings for every reading.

OBSTRUCT to indicate inaccessible loca-tions.

Memory: 16 megabit non-volatile ram.

Features:Setups:

64 custom user-definable setups; Factory setups can also be edited by the user.

Gates:

Single gate in pulse-echo mode, or single gate with holdoff in echo-echo mode; Adjustable threshold.

Selectable Transducers:

Selectable transducer types with built-in dual path error correction for improved linearity.

Alarm Mode:

Set Hi and Lo tolerances with audible beeper and visual LEDs.

Fast-Scan Mode:

Takes 32 readings per second and displays the minimum reading found when the transducer is removed.

CertificationFactory calibration traceable to NIST & MIL-STD-45662A.

Warranty2 year limited

Digits View B-Scan View

M A D E I N T H E U S ADistributed By:

DAKOTA ULTRASONICS1500 Green Hills Road, #107Scotts Valley, CA 95066Ph: (831) 431-9722Fax: (831) 431-9723Website: www.dakotaultrasonics.comEmail: [email protected]


Recommended