+ All Categories
Home > Documents > Top class high accuracy roundness cylindrical profile ...

Top class high accuracy roundness cylindrical profile ...

Date post: 18-Dec-2021
Category:
Upload: others
View: 2 times
Download: 0 times
Share this document with a friend
6
RONDCOM NEX DX RONDCOM NEX SD *Equipped off-set typed CNC detecting holder with RONDCOM NEX Rs 300 system Top class high accuracy roundness cylindrical profile measuring instrument Rotation accuracy (0.02 + 3.2 H/10000) μm Equipped with full new function and meets a need of machined parts high accuracy measurement. It is a top class high accuracy roundness cylindrical profi le measuring instrument. Opposed diameter measuring function patented Superior feature to measure inner/ outer diam- eter with high repeatability. Measure a workpiece at angles of 0 and 180 degrees on the table. The evaluation algorithm implemented as the standard to correct the errors by temperature change and generatrix line shifting, performs highly-precise diameter measurement. R-axis taper following function* The straightness of tapered surface can be measured by the function. Taper angle and straightness can be mea- sured even if it excess the range of the detector. *Taper angle may have an impact on the measurement accuracy at straightness. Contact us for details. Measure at an angle of 0 degrees Move from 0 degrees to 180 degrees Measure at an angle of 180 degrees Example of the measurement RONDCOM NEX SURFCOM 5000 36.202 ° 18.085 ° (36.170 °) Example of the measurement Comparison of the measurement results, by the high accuracy contour measuring instrument (SURFCOM 5000) and by RONDCOM NEX. Dedicated catalog is available. Roundness and Cylindrical Profile Measuring Instruments Table-Rotating Type CNC Measuring Instrument Roundness and Cylindricity Profile Measuring Instruments 20
Transcript
Page 1: Top class high accuracy roundness cylindrical profile ...

RONDCOM NEX DXRONDCOM NEX SD* Equipped off-set typed CNC detecting holder with RONDCOM NEX Rs 300 system

Top class high accuracy roundness cylindrical profile measuring instrument

Rotation accuracy (0.02 + 3.2 H/10000) µmEquipped with full new function and meets a need of machined parts high accuracy measurement.It is a top class high accuracy roundness cylindrical profi le measuring instrument.

Opposed diameter measuring function patented

Superior feature to measure inner/ outer diam-eter with high repeatability.Measure a workpiece at angles of 0 and 180 degrees on the table.The evaluation algorithm implemented as the standard to correct the errors by temperature change and generatrix line shifting, performs highly-precise diameter measurement.

R-axis taper following function*The straightness of tapered surface can be measured by the function.Taper angle and straightness can be mea-sured even if it excess the range of the detector.* Taper angle may have an impact on the measurement accuracy at straightness. Contact us for details.

Measure at an angle of0 degrees

Move from 0 degrees to 180 degrees

Measure at an angle of180 degrees

Example of the measurement

RONDCOM NEX SURFCOM 5000

36.202 ° 18.085 ° (36.170 °)Example of the measurement

Comparison of the measurement results, by the high accuracy contour measuring instrument (SURFCOM 5000) and by RONDCOM NEX.

Dedicated catalog is available.

Roundness and Cylindrical ProfileMeasuring Instruments Table-Rotating Type CNC Measuring Instrument

Roundness and Cylindricity Profile Measuring Instruments20

Page 2: Top class high accuracy roundness cylindrical profile ...

CNC upgrade availableTo meet the needs of many users, it is now possible to upgrade a manual machine to a CNC machine. The CNC conversion is amazingly easy and makes no change to the installation space. If you have installed a manual machine because only a small quantity of workpieces had to be measured, the machine can be upgraded in the same space whenever necessary.

*Excepts ROUNDCOM NEX Rs sries and sries.

Offset CNC detector holder RONDCOM NEX 300 system standard accessory

A mechanism enabling the measurement of various shapes of workpieces without interference with the R axis is provided as a standard feature. The manual type, which is mounted on the NEX100/200 systems, features an easy operation that switches between outer diameter measurement and plane measurement of the top surface only by pushing down the holder. The CNC type, which is mounted on the NEX 300 system, can control the detector posture fully automatically for the measurement of inner and outer diameters, top and bottom surfaces and tapered surface, etc., and signifi cantly improves the measurement effi ciency. Since the detectors of the manual type holder and the CNC type holder are common, if you have both, you do not need to keep a spare detector for maintenance, which results in cost reduction.

NEX 200/300NEX 100

Conventional measuring instrument

RONDCOM NEX 100 and NEX 200/300 series

Upgrade can be performed at the site.

Repurchase

RONDCOM

CNCmachine

Manualmachine

RONDCOM

RONDCOM NEX DX type

Almost maintenance-free by automatic lubrication to column.

Automatic lubrication function mounted on Z-axis column

Minimization of effect of disturbance from air-conditioner and others by functional design.

Full-covered main body and columnThe linear expansion coeffi cient of ceramic is smaller and weighs half of iron but the material is harder. Its weight is reduced but higher in rigidity and additionally it is hardly affected by changes in environmental temperature.

Weight saving and high rigidity by employing ceramics for R-axis arm

Adopting SUS for the table frees from rust.Unnecessity of oil coating, Maintenance-free.

In case of measurement using tilt adjusting function, it is not necessary to set up the measuring height.

Rust proof by adopting SUS table

Equipped optical linear scale in Z-axis column

Extend the centering stroke to ±5 mm by extending the table diameter to Φ 235 mm.

Extension of centering stroke

Innovative approach to measurement with new concept All-in-one software for measurement and analysis based on electronic form system.

ACCTeeIntegrated Analysis Software

PrinterWith drawer mechanism.

Spiral cylindricity measurement by combining table rotation with rectilinear movement. Unnecessity of Z positioning saves 30% of cylindricity measurement time compared to conventional manner.

Spiral cylindricity measuring function

Storage partfor PC

patented

Tab

le R

ota

tin

g T

ype:

CN

C

Roundness and Cylindricity Profile Measuring Instruments 21

Page 3: Top class high accuracy roundness cylindrical profile ...

Measurement of the fi rst workpiece (location)

Measurement of multiple workpieces (locations) with no changeover by moving the table in X/Y directions

Shared Option for RONDCOM NEX Series

XY-Axis Automatic Stage

Labor-saving solution achieving significantreductions in the operator's workload

When attached to RONDCOM NEX Rs α 300 DX-22

NEW

Continuous measurement of multiple workpieces/locations with no changeover Patend pendingContinuous measurement of multiple workpieces/locations saves the trouble of changeover, achieving reductions in time and labor required for measurement.Data of different evaluation items such as roundness and cylindricity as well as surface roughness, lead and twist (by attaching on RONDCOM NEX Rs/NEX Rs series) can be automatically captured at once for multiple workpieces.The supplied counterweight offsets eccentric loads during the table feeding and realizes a highly accurate measurement in a wide range of stroke.

Can be freely attached, removed,and capable of being retrofittedSince it can be attached and removed by customer, customer can choose between the standard table and the XY-axis automatic stage depending on the workpiece.As this option is available for retrofi tting, it can be attached to an existing machine*.

*By modifying part of the table of the measuring machine at the site.

Roundness and Cylindrical ProfileMeasuring Instruments Table-Rotating Type CNC Measuring Instrument

Roundness and Cylindricity Profile Measuring Instruments24

Page 4: Top class high accuracy roundness cylindrical profile ...

Item Specifications of RONDCOM NEX serieswith XY-axis automatic stage

Drive rangeCx-axis (mm) 200 (±100)

Cy-axis (mm) 100 (±50)

WorkpieceLoadable range (mm) Ф 300 from the center of XY-axis automatic stage's table

Max. loadable weight (kg) 5

Drive speed of the table (mm/s) Max. 20

Rotational Accuracy *

Radial direction (µm) (0.08 + 6H/10000)

Axial direction (µm) (0.08 + 6R/10000)

Z-axis Parallelism (µm/mm) 0.5/150

Installation dimensions and weight

Width x Depth x Height (mm) 540 x 356 x 132

Height from the upper surface of the machine's table to the upper surface of XY-axis automatic (mm)stage's table

120

Weight (kg) Approx. 20 (except for the standard equipped counterweights)

Applicable model RONDCOM NEX / NEX 200/300RONDCOM NEX Rs / NEX Rs 200/300

Storage place of counterweight Table(Workpiece loadable range)

9─M5 Depth: 7.5

* JIS B 7451-1997 compliant. H is the height of the measurement point from the upper surface of the machine's table in mm,R is the distance from the rotational center of the machine's table in mm

●The other specifi cation items conform to the Specifi cations of RONDCOM NEX series

External view

Specifications

Tab

le R

ota

tin

g T

ype:

CN

C

Roundness and Cylindricity Profile Measuring Instruments 25

Page 5: Top class high accuracy roundness cylindrical profile ...

980720 225

245

Max. measuring diameter φ300

φ235

800

1070

to 1

370

700

1095

(500

mm

col

umn)

895(

300

mm

col

umn)

330

Max

.mea

surin

g he

ight

500

(500

mm

col

umn)

Max.m

easu

ring h

eight

300

(300 m

m co

lumn)

System rack (option)E-DK-S24A

Anti-vibration table (option)E-VS-R16B

780

580

800

980720 225

245

Max. measuring diameter φ300

φ235

800

1070

to 1

370

700

1095

(500

mm

col

umn)

895(

300

mm

col

umn)

330

Max

.mea

surin

g he

ight

500

(500

mm

col

umn)

Max.m

easu

ring h

eight

300

(300 m

m co

lumn)

System rack (option)E-DK-S24A

Anti-vibration table (option)E-VS-R16B

780

580

800

SD

820

90

1330

1000

740

1000

1400

245Max. measuring diameter

φ300

1770

(500

mm

col

umn)

Max

.mea

surin

g he

ight

500

(500

mm

col

umn)

Max

.mea

surin

g he

ight

300

(300

mm

col

umn)

1570

(300

mm

col

umn)

310

210

480

φ235

820

90

1330

1000

740

1000

1400

245Max. measuring diameter

φ300

1770

(500

mm

col

umn)

Max

.mea

surin

g he

ight

500

(500

mm

col

umn)

Max

.mea

surin

g he

ight

300

(300

mm

col

umn)

1570

(300

mm

col

umn)

310

210

480

φ235

DX

RONDCOM NEX/NEX Specification

RONDCOM NEX Series External View(common with NEX Rs)

*1 NEX-11 (Max. loading mass 30 kg, 300 mm column), NEX-12 (Max. loading mass 30 kg, 500 mm column) NEX -21 (Max. loading mass 60 kg, 300 mm column), NEX -22 (Max. loading mass 60 kg, 500 mm column), NEX -23 (Max. loading mass 60 kg, 900 mm column)

*2 Please contact our sale personnel as there may be limitations due to the measurement diameter, and the combination of detector and stylus.*3 JIS B 7451-1997 compliant. H is the height of the measurement point from the upper surface of the table in mm, and R is the distance from the rotational center of the table in mm.*4 When using measurement diameter extension offset-type detector holder E-DH-RB86A (optional)

■Hardware

Z=300 size (-11, -21)Z=500 size (-12, -22)

Model

Item

RONDCOM NEX (-11, -12) RONDCOM NEX (-21, -22, -23)

100 200 300SD DX SD DX SD DX

Model *1 11 12 11 12 11 12 11 12 11 12 11 1221 22 23 21 22 21 22 23 21 22 21 22 23 21 22

Alignment Manual CNCOffset type detectror holder Manual CNC

Measuring range

Max. measuring diameter (mm) outer diameter: Ф 300 (Ф 350) *4 inner diameter: Ф 360 (Ф 410) *4

Radial feed range (R-axis) (mm) 180Up/down feed range (Z-axis) (mm) 300 500 900 300 500 300 500 900 300 500 300 500 900 300 500Max. loading diameter (mm) Ф 580Max. measuring height (mm) 300 500 900 300 500 300 500 900 300 500 300 500 900 300 500Depth of measurement (mm) 150 *2

Accuracy

Rotation accuracy *3

Radial direction (µm) (0.02 + 3.2H/10000)Axis direction (µm) (0.02 + 3.2R/10000)

Straightnessaccuracy

Up/down direction(Z-axis) (µm/mm)

0.10/100 0.20 /100 0.10/100 0.10/100 0.20

/100 0.10/100 0.10/100 0.20 /100 0.10/100

0.15 /300

0.23 /500

0.90 /900

0.15 /300

0.23 /500

0.15 /300

0.23 /500

0.90 /900

0.15 /300

0.23 /500

0.15 /300

0.23 /500

0.90 /900

0.15 /300

0.23 /500

Radial direction (R-axis) (µm/mm) 0.7/180Parallelismaccuracy Z-axis/T-axis (µm/mm) 0.7

/3001.0

/5002.0

/9000.7

/3001.0

/5000.7

/3001.0

/5002.0

/9000.7

/3001.0

/5000.7

/3001.0

/5002.0

/9000.7

/3001.0 /500

Squarenessaccuracy R-axis/T-axis (µm/mm) 1.0/150

Scale indicationaccuracy

R-axis (µm) (0.5 + L/180 + 2L⊿T/100) L: travel distance (mm) ⊿T: temperature difference between standardcondition (20℃) and environmental temperature (℃).

Speed

Measurementspeed

Rotational spped (θ-axis) (/min) 1 to 10Up/down speed (Z-axis) (mm/s) 0.5 to 10Radial direction speed (R-axis) (mm/s) 0.5 to 10

Movementspeed

Rotational speed (θ-axis) (/min) max. 20Up/down speed (Z-axis) (mm/s) 5 to 60Radial direction speed (R-axis) (mm/s) 5 to 30

Table

Table diameter (mm) Ф 235Centering range (mm) ±5Tilting range (°) ±1

Max. loadingmass

NEX (kg) 30NEX (kg) 60

Detector/Stylus

DetectorE-DT-R120B(standardly equipped)

Measuring force (mN) 30 to 100Linear range (µm) ±1000Functions Switching outer or inner diameter, Front/over travel adjustment function, Emergency stop function

StylusEM46000-S302(standardly equipped)

Stylus ball diameter (mm) Ф 1.6Stylus length (mm) 53Stylus ball material Carbide

Roundness and Cylindrical ProfileMeasuring Instruments Table-Rotating Type CNC Measuring Instrument

Roundness and Cylindricity Profile Measuring Instruments26

Page 6: Top class high accuracy roundness cylindrical profile ...

SD

■Software

■Specification

RONDCOM NEX/NEX Specification

RONDCOM NEX Series External View(common with NEX Rs)

*5 The installation dimensions and weight of NEX -23 (Max. loading mass 60 kg, 900 mm column) are the values when using the anti-vibration table E-VS-R86A (optional).

Brackets(4pc.)

Air Inputφ 8 Tube

Power Sapply

170 550

610

265

720 (225)

Maximum diameterof measurement

φ 300

Max

imum

hei

ght

of m

easu

rem

ent

900

φ 235

138 517

945

45

NE

X

: 1

600

NE

X R

s

: 162

5

(830

)

330

500

NE

X

: (2

100)

NE

X R

s

: (21

25)

When mounted on the anti-vibration table E-VS-R86A

Z=900 size (-23)

Installation dimension *5

Width (mm) 720 1074 1400 720 1074 1400 720 1074 1400Depth (mm) 580 824 820 580 824 820 580 824 820

HeightNEX (mm) 895 1095 1570 1770 895 1095 1570 1770 895 1095 1570 1770NEX (mm) 900 1100 2100 1570 1770 900 1100 2100 1570 1770 900 1100 2100 1570 1770

Weight *5

NEX Machine (kg) Approx.170

Approx.180 Approx.

330Approx.340

Approx.170

Approx.180 Approx.

330Approx.340

Approx.170

Approx.180 Approx.

330Approx.340

Computer (kg) Approx.10 Approx.10 Approx.10

NEX Machine (kg) Approx.190

Approx.200

Approx.560 Approx.

350Approx.360

Approx.190

Approx.200

Approx.560 Approx.

350Approx.360

Approx.190

Approx.200

Approx.560 Approx.

350Approx.360

Computer (kg) Approx.10 Approx.10 Approx.10

Power supply Voltage, frequency (V, Hz) AC100 to 240, 50/60 (grounding required)Power consumption (VA) Approx. 530

Air supply

Supply air pressure

NEX (MPa) 0.35 to 0.7 0.35 to 0.7 0.35 to 0.7 0.35 to 0.7NEX (MPa) 0.45 to 0.7

Working air pressure

NEX (MPa) 0.3 0.3 0.3 0.3NEX (MPa) 0.4

Air consumption volume

NEX (NL/min) 30 30 30 30NEX (NL/min) 40

Air supply connecting nipple (main unit) One-touch pipe joint for outer diameter Ф 8 mm hose

Operating environmentOperating temperature (℃) 10 to 30Guarranteed accuracy tem-erature range (℃) 20±2

Model

Item

RONDCOM NEX (-11, -12) RONDCOM NEX (-21, -22, -23)

100 200 300SD DX SD DX SD DX

Model *1 11 12 11 12 11 12 11 12 11 12 11 1221 22 23 21 22 21 22 23 21 22 21 22 23 21 22

Number of sampling (point) 14400Type of filter Digital filter Gaussian/2RC/spline/robust (spline)

Cutoff valueRotational direc-tion (θ-axis)

Low pass settable any value in range of 15, 50, 150, 500, 1500 peaks/rotation, 15 to 1500 peaks/rotation

Band pass 1 to 1500 peaks/rotationRectilinear direction (Z-axis) Low pass 0.025, 0.08, 0.25, 0.8, 2.5, 8 mm (any value in 0.0001 mm units)

Roundness evaluation of form errorMZC (min. zone circle method), LSC (least square circle method),

MIC (max. inscribed circle method), MCC (min. circumscribed circle method), N.C. (no compensation)

Measuring items

Rotational direction Roundness, flatness, flatness (compound), parallelism, concentricity, coaxiality, cylindricity, diameter deviation, squareness, thickness variation, partial circle

Rectilinear direction Straightness (Z), straightness (R), cylindricity, squareness, parallelism, diameter deviation, axis straightness

Analysis processing functions

Notch function (level, angle, cursor), combination of roundness evaluation methods, nominal value collation, cylinder 3D profile display (line drawing, shading, contour line),

real-time display, profile characteristic graph display (bearing area curve, amplitude distribution function, power spectrum), CNC automatic measuring function,

automatic centering/tilting adjustment function (except for NEX 100, NEX 100)

Display item Measuring conditions, measuring parameters, comments, printer output conditions, profile graphics (expansion plan, 3D plan), error messages, etc.

Tab

le R

ota

tin

g T

ype:

CN

C

Roundness and Cylindricity Profile Measuring Instruments 27


Recommended