USEPA Methods 8270 and 8260 on a Single GCMS Without Changing Columns Richard Whitney, Laura Chambers, Clifford Taylor Shimadzu Scientific Instruments, Inc. Columbia, MD Spotlight on Method 8270 Instrumentation
Transcript
Page 1: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

USEPA Methods 8270 and 8260 on a Single

GCMS Without Changing Columns

Richard Whitney, Laura Chambers, Clifford Taylor

Shimadzu Scientific Instruments, Inc.

Columbia, MD

Spotlight on Method

8270 Instrumentation

Page 2: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

USEPA Methods 8270 and 8260 on a Single

GCMS Without Change Columns

Richard Whitney, Laura Chambers, Clifford Taylor

Shimadzu Scientific Instruments, Inc.

Columbia, MD

Spotlight on Method

8270 Instrumentation

Venting the MS to . . .

Page 3: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

• Small laboratory

• Limited budget

• Multiple applications

• No time

Problem

3

The Big Bang Theory – CBS.com

Page 4: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

• 8260 and 8270

• Single GCMS

• Two inlets, two columns

• No venting

Solution

4

The Big Bang Theory – CBS.com

Page 5: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

What We Did

Instrument configuration

Optimize method parameters

Analytical results (aka data)

5

Page 6: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Instrument Configuration

GCMS-QP2010 SE

Production Workhorse 6

Page 7: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Instrument Configuration

GCMS-QP2010 SE

Production Workhorse 7

Page 8: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Instrument Configuration (cont.)

8

GCMS Twin Line Kit

Page 9: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Instrument Configuration (cont.)

9

Page 10: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Instrument Configuration (cont.)

10

Page 11: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Instrument Configuration (cont.)

58 L/sec (He)

Pfeiffer TMP (Rotary pump 30 L/min)

11

Page 12: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Instrument Configuration (cont.)

12

Page 13: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Instrument Configuration (cont.)

13

Ecology Mode saves ~26%

Page 14: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Instrument Configuration (cont.)

Parameter EPA Method 8260 EPA Method 8270

Sample introduction Purge-and-trap Liquid syringe

Liner Shimadzu multi-purpose

split liner, no glass wool

Shimadzu multi-purpose

split liner, no glass wool

Injection conditions Split 75:1 Split 10:1

Column Rxi-624Sil MS

20 m x 0.18 mm x 1.4 µm

Rxi-5Sil MS

20 m x 0.18 mm x 0.18 µm

Maximum column

temperature Program = 320 °C Program = 350 °C

Carrier gas Helium, constant linear velocity mode

MS pump Single stage TMP, max flow 4 mL/min (He)

MS interface 225 °C

Ion source EI, 200 °C

14

Page 15: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Instrument Configuration (cont.)

Born in Kyoto

Made in the USA

15

Page 16: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

• Primary objectives

1. Keep total column flow rate

below 4 mL/minute

2. Keep maximum oven

temperature below 320 °C

3. Keep GC run time as short

as practical

Method Optimization

16

The Big Bang Theory – CBS.com

Page 17: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Method Optimization (cont.)

Parameter Method 8260

Primary

Method 8270

Inactive

Column constant

linear velocity

45 cm/second

~0.8 mL/min

30 cm/second

~0.5 mL/min

Maximum total

column flow ~1.3 mL/minute

Oven program

35 °C (5 min)

20 °C/min to 220 °C

(hold 2.5 min)

NA

GC run time 12.5 minutes NA

17

Page 18: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Method Optimization (cont.)

Parameter Method 8260

Inactive

Method 8270

Primary

Column constant

linear velocity

30 cm/second

~0.5 mL/min

40 cm/second

~0.7 mL/min

Maximum total

column flow ~1.2 mL/minute

Oven program NA

45 °C (0.5 min)

25 °C/min to 315 °C

(hold 4.2 min)

GC run time NA 16 minutes

18

Page 19: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Why Constant Linear Velocity?

Van Deemter Plot

Linear Velocity – Isothermal Oven

Optimal LV for

each carrier

Gas

19

Page 20: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Why Constant Linear Velocity? (cont.)

Head Pressure (kPa)

Flow (ml/min)

Linear Velocity (cm/sec)

Oven Temp

With temperature programming:

1. Viscosity of gas increases

2. Gas expands

3. Flow rate and LV drop

20

Page 21: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Why Constant Linear Velocity? (cont.)

Linear Velocity

Linear Velocity

Oven Temp

Head Pressure

Flow

Linear Velocity

Constant

Pressure

21

Page 22: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Why Constant Linear Velocity? (cont.)

Linear

Velocity

Linear

Velocity

Head Pressure

Flow

Linear Velocity

Oven Temp

Constant

Flow

22

Page 23: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Why Constant Linear Velocity? (cont.)

Constant

Linear Velocity

Head Pressure

Linear Velocity

Flow

Oven Temp

Linear Velocity

23

Page 24: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Why Constant Linear Velocity? (cont.)

Constant Flow

Linear velocity = 20 cm/sec

Constant Pressure

Linear velocity = 15 cm/sec

Constant Linear Velocity

Linear velocity = 30 cm/sec

24

Page 25: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Analytical Results

BFB Spectrum

EPA Method 8260

BFB Tune Criteria

Passed

25

Page 26: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Analytical Results (cont.)

EPA Method 8260

10 µg/L calibration standard

TIC – 66 compounds resolved

26

Page 27: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Analytical Results (cont.)

DFTPP Spectrum

EPA Method 8270

DFTPP Tune Criteria

Passed

27

Page 28: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Analytical Results (cont.)

DDT Breakdown

DDT (expanded)

DDE DDD

Benzidine and Pentachlorophenol Tailing

Benzidine

Pentachlorophenol

28

Page 29: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Analytical Results (cont.)

DDT Breakdown

DDT (expanded)

DDE DDD

Benzidine and Pentachlorophenol Tailing

Benzidine

Pentachlorophenol

29

Page 30: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Analytical Results (cont.)

EPA Method 8270

50 µg/mL calibration standard

TIC – 66 compounds resolved

30

Page 31: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Analytical Results (cont.)

EPA Method 8270 Calibration

Range 8 points, 0.4 – 160 µg/mL

6 Internal Standards Acenaphthene-d10, Cyrysene-d12, 1,4-Dichlorobenzene-d4,

Naphthalene-d8, Perylene-d12, Phenanthrene-d10

40 µg/mL each

Response Factor 54 compounds < 15% RSD

Min. RF = 0.132, Pentachlorophenol

Max. RF = 3.966, bis(2-chloroisopropyl)ether

Linear Curve 12 compounds > 15% RSD

Min r = 0.997, 2,4-dinitrophenol

Max r = 0.999, 8 compounds

31

Page 32: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Analytical Results (cont.)

2-Nitrophenol (2 µg/mL)

r = 0.999

32

Page 33: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Benzo(b&k)fluoranthenes (1 µg/mL)

Analytical Results (cont.)

RSDb = 6.1%

RSDk = 12.0%

33

Page 34: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Analytical Results (cont.)

0.0

2.0

4.0

6.0

8.0

10.0

12.0

%R

SD

% Relative Standard Deviation (0.4 µg/mL*, n = 8)

62/66 RSD < 8%

39/66 RSD < 4%

34

Page 35: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Analytical Results (cont.)

0.0

20.0

40.0

60.0

80.0

100.0

120.0

Ave

rag

e %

Re

co

ve

ry

Average % Recovery (0.4 µg/mL*, n = 8)

57/66 Recovery 80 – 120%

35

Page 36: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Analytical Results (cont.)

0.00

0.10

0.20

0.30

0.40

0.50

0.60

0.70

0.80

IDL (

µg

/mL

)

Statistical Instrument Detection Limit (0.4 µg/mL, n = 8)

60/66 IDL < 0.2 µg/mL

36

Page 37: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Analytical Results (cont.)

DCM carryover < 1.0 ppb

37

Page 38: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Analytical Results (cont.)

Acetone carryover ~ 1.0 ppb

38

Page 39: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

• Two methods

• Two intro techniques

• Two injection ports

• Two columns

• One MS

• All method criteria passed

• One happy customer

• Shimadzu application

note GCMS-1302

Summary

39

The Big Bang Theory – CBS.com

Page 40: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing

Acknowledgements

Richard R. Whitney, Ph.D.

Configuration insight and

all the hard work!

Julie Kowalski

Advice re column and conditions

Columns and standards

40

Page 41: USEPA Methods 8270 and 8260 on a Single GCMS Without ...nemc.us/docs/2013/presentations/Tue-Spotlight on Method 8270... · USEPA Methods 8270 and 8260 on a Single GCMS Without Changing