    User manual

    JOFRA DLC-155/158/250/700Copyright 2010 AMETEK Denmark A/S

  • 7/30/2019 User Manual Dlc

    2/17

  • 7/30/2019 User Manual Dlc

    3/17

    127414 01 20-12-2010 3

    Table of contents: Page

    1.0 General information ......................42.0 Introduction....................................5

    3.0 Functionality ..................................6

    3.1 Functional description..................6

    3.2 Connections ................................6

    3.3 Serial number..............................8

    4.0 Operation........................................9

    4.1 Operation area ............................9

    5.0 Maintenance.................................10

    5.1 Calibration .................................11

    6.0 Technical specifications .............15

  • 7/30/2019 User Manual Dlc

    4/17

    4 20-12-2010 127414 01

    1.0 General information

    This manual is only effective for the following

    products:

    JOFRA DLC-155JOFRA DLC-158JOFRA DLC-250JOFRA DLC-700

    The products are manufactured by:

    AMETEK Denmark A/S

    Gydevang 32-343450 Allerd - Denmark

    TEL: +45 48 16 80 00FAX: +45 48 16 80 80

  • 7/30/2019 User Manual Dlc

    5/17

    127414 01 20-12-2010 5

    2.0 Introduction

    The JOFRA DLC-155, DLC-158, DLC-250 and DLC-700 probes are designed for fast and traceablecalibration and temperature measuring with JOFRARTC-156/157 B/C, JOFRA RTC-158 B/C, JOFRARTC-250 B/C and JOFRA RTC-700 B/C systemsand are ready for use.

    Please read this manual carefully before use, toobtain maximum value of your calibration system.

    Warning

    Read this manual before use.

    Do not use in hazardous area.

    Handle carefully.

    Never exceed temperature range

  • 7/30/2019 User Manual Dlc

    6/17

    6 20-12-2010 127414 01

    3.0 Functionality

    3.1 Functional description

    The probes can be used for temperatures in therange :

    DLC-155 : -60C to 155C / -76F to 311F

    DLC-158 : -60C to 155C / -76F to 311F

    DLC-250 : 0C to 250C / 32F to 482F

    DLC-700 : 0C to 700C / 32F to 1292F

    The JOFRA DLC-155, DLC-158, DLC-250 and DLC-700 probes may be supplied with a certificate for alimited temperature range.

    3.2 Connections

    The probes are delivered with a connecting cable.The pin-layout is as follows:

    Pin 1 : TC V-Pin 2 : TC V+Pin 3 : Memory GND

    Pin 4 : Memory I/O

  • 7/30/2019 User Manual Dlc

    7/17

    127414 01 20-12-2010 7

    The figure below is shown from the connector sideof the probe connector.

    1

    23

    4

  • 7/30/2019 User Manual Dlc

    8/17

    8 20-12-2010 127414 01

    3.3 Serial number

    The serial number is placed on the probe as shownon the figure below:

    Serial number

  • 7/30/2019 User Manual Dlc

    9/17

    127414 01 20-12-2010 9

    4.0 Operation

    4.1 Operation area

    All the probes are intended for use in areas whichmeet the following:

    Ambient temperaturerange : -20C to 70C (-4F to 158F)

    Humidity : 0% to 90%

    Protection class : IP 50

    WarningDo not use in hazardous areas.

  • 7/30/2019 User Manual Dlc

    10/17

    10 20-12-2010 127414 01

    5.0 Maintenance

    The probe does not require specific maintenancebefore or after use. The user may carry out thefollowing procedure himself:

    Cleaning the probe : Use alcohol or water

    and a soft cloth.

    Caution

    The probe must always be protectedagainst any mechanical damage.

    The probe must never be exposed to

    mechanical shock effects. Avoid thermal shock

    Any bending of the probe may causepermanent damage

  • 7/30/2019 User Manual Dlc

    11/17

    127414 01 20-12-2010 11

    5.1 Calibration

    Required equipment:

    DLC-probe to be calibrated. RTC-calibrator with DLC-input.

    An insert with correct size holes for the DLC-probe and the temperature probe.

    For calibrating DLC-155, DLC-158 and DLC-250:an AMETEK 127587 temperature probe(recommended) or similar with short measuringzone is required.

    For calibrating DLC-700: a STS-200 A 970temperature probe (recommended) or similar isrequired.

    Procedure:

    1. Place the DLC-probe in the insert and connectthe probe to the DLC-input of the calibrator.

    2. Go to the Sensor setup menu.

    3. Set Use load compensation to YES. If thereference probe is connected to the REF input,

    set Sensor type to External and SET followsTRUE to No. If no reference probe isconnected, set Sensor type to Internal.

    4. Enter the DLC-menu.

    5. Read the A, B and C-coefficients of the probeand enter the values in the EXCEL spreadsheet

    (127588 DLC-155.XLS, 127884 DLC-158.XLS,127885 DLC-250.XLS or 127589 DLC-700.XLS).

  • 7/30/2019 User Manual Dlc

    12/17

    12 20-12-2010 127414 01

    The calibration temperatures are:Sensor DLC-155 DLC-155 DLC-700

    Calibrator RTC-156 RTC-157 RTC-700

    -30C/-22F -45C/-49F 33C/91.4F-20C/-4F -20C/-4F 200C/392F

    0C/32F 0C/32F 350C/662F

    50C/122F 50C/122F 500C/932F

    100C/212F 100C/212F 660C/1220F

    *Recommendedcalibrationtemperatures

    155C/311F 155C/311F 700C/1292F

    Lower

    measuring point

    Bottom of

    insert

    Bottom of

    insert

    Bottom of

    insertUppermeasuring point

    60 mm /2.4 frombottom

    60 mm /2.4 frombottom

    80 mm / 3.1from bottom

    Curve fitwarning limit

    0.01C/0.018F

    0.01C /0.018F

    0.05C /0.09F

    * Note : The temperatures must be entered in Celcius.

  • 7/30/2019 User Manual Dlc

    13/17

    127414 01 20-12-2010 13

    Sensor DLC-158 DLC-250

    Calibrator RTC-158 RTC-250

    -22C/-72F 28C/82F

    -10C/-4F 50C/122F0C/32F 100C/212F

    50C/122F 150C/302F

    100C/212F 200C/392F

    *Recommendedcalibrationtemperatures

    155C/311F 250C/482F

    Lowermeasuring point

    Bottom ofinsert

    Bottom ofinsert

    Uppermeasuring point

    60 mm /2.4 frombottom

    60 mm /2.4 frombottom

    Curve fitwarning limit

    0.01C/0.018F

    0.01C /0.018F

    * Note : The temperatures must be entered in Celcius.

    The following procedure is repeated for eachtemperature:

    1. Set the SET temperature.

    2. Place the reference temperature probe in thebottom of the insert.Wait until the temperature have stabilized and theDLC-reading is 0.00C. If the DLC-readingcontinuously shows a positive value at -30C(RTC-156) / -45C (RTC-157) / -22C (RTC-158),increase the SET-temperature in steps of 1C

    until the DLC-reading displays 0.00C.

  • 7/30/2019 User Manual Dlc

    14/17

    14 20-12-2010 127414 01

    3. Record the temperature of the reference probe inthe excel spreadsheet.

    4. Move the reference probe to the upper position

    60 mm (RTC-156/157/158/250) or 80 mm (RTC-700).

    5. Wait until the temperature have stabilized and theDLC-reading is 0.00C.

    6. Record the temperature of the reference probeand the DLC-probe in the excel spreadsheet.

    7. Go to next SET-temperature.

    After the last STEP is completed:

    8. Check that all values in the calculated curve fitare less than the values specified in the table. Ifthis is OK, the new coefficients can be

    downloaded using the CON050 software.

  • 7/30/2019 User Manual Dlc

    15/17

    127414 01 20-12-2010 15

    6.0 Technical specifications

    Probe specifications DLC-155:

    Sensor type : Differential TC type T

    Probe length : 177 mm (6.96 in)

    Temperature range : -60C to 155C / -76F to311F

    Diameter : OD3 mmDepth : 140 mm (5.51 in)

    Media compatibility : AISI316

    Connection : Redel plug with build inmemory is standard

    Probe specifications DLC-158:

    Sensor type : Differential TC type T

    Probe length : 192 mm (7.56 in)

    Temperature range : -60C to 155C / -76F to

    311F

    Diameter : OD3 mm

    Depth : 140 mm (5.51 in)

    Media compatibility : AISI316

    Connection : Redel plug with build inmemory is standard

  • 7/30/2019 User Manual Dlc

    16/17

    16 20-12-2010 127414 01

    Probe specifications DLC-250:

    Sensor type : Differential TC type T

    Probe length : 192 mm (7.56 in)

    Temperature range : 0C to 250C / 32F to482F

    Diameter : OD3 mm

    Depth : 140 mm (5.51 in)

    Media compatibility : AISI316

    Connection : Redel plug with build inmemory is standard

    Probe specifications DLC-700:

    Sensor type : Differential TC type N

    Probe length : 177 mm (6.96 in)

    Temperature range : 0C to 700C / 32F to1292F

    Diameter : OD4 mm

    Depth : 200 mm (7.9 in)

    Media compatibility : INCONEL 600

    Connection : Redel plug with build inmemory is standard

  • 7/30/2019 User Manual Dlc

    17/17

    127414 01 20-12-2010 17

    Certificate:

    If the DLC probes are supplied with a certificate, thecalibration is carried out as recommended below

    according to the ITS 90 temperature scale.

    The DLC-155/158 probe is as standard calibrated inthe range :

    45C to 155C (-49F to 311F).

    The DLC-250 probe is as standard calibrated in therange :

    28C to 250C (82F to 482F).

    The DLC-700 probe is as standard calibrated in therange :

    33C to 700C (91.4F to 1292F).


Recommended