+ All Categories
Home > Documents > User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

Date post: 06-Jul-2018
Category:
Upload: cruzrenato
View: 219 times
Download: 0 times
Share this document with a friend

of 52

Transcript
  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    1/52

    Model: 3029 and 3039

    Countertop Incubator 

    10 cu. ft capacity

    Operating and Maintenance Manual

    Manual No: 7013029

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    2/52

    Models 3029 and 3039_____________________________________________________________________________________ 

    Read This Instruction Manual.

    Failure to read, understand and follow the instructions inthis manual may result in damage to the unit, injury to operat-

    ing personnel, and poor equipment performance.

    CAUTION! All internal adjustments and maintenance must 

    be performed by qualified service personnel.

    Refer to the serial tag on the back of this manual.

    The material in this manual is for information purposes

    only. The contents and the product it describes are subject to

    change without notice. Forma Scientific, Inc. makes no repre-

    sentations or warranties with respect to this manual. In no event

    shall Forma Scientific, Inc. be held liable for any damages, direct

    or incidental, arising out of or related to the use of this manual.

    MANUAL NUMBER 7013029

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    3/52

    Models 3029 and 3039_____________________________________________________________________________________ 

    Table of Contents

    Section 1 - Inspection  . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 - 1

    1.1 Preliminary Inspection . . . . . . . . . . . . . . . . . . . . . . . .1 - 1

    1.2 Shipping Claims . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 - 1

    1.3 Responsibility for Shipping Damage . . . . . . . . . . . . .1 - 1

    Section 2 - Basic Set-Up Procedure  . . . . . . . . . . . . . . . .2 - 1

    2.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1

    2.2 Leveling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1

    2.3 Drain Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1

    2.4 Water Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1

    2.5 CO2 Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1

    2.6 Door Heater . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1

    2.7 Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1

    2.8 Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1

    2.9 Stabilization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1

    2.10 Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1

    2.11 Stabilization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1

    2.12 CO2

     . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 12.13 Door Heater . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1

    Section 3 - Specifications  . . . . . . . . . . . . . . . . . . . . . . . .3 - 1

    Section 4 - General Description . . . . . . . . . . . . . . . . . . .4 - 1

    4.1 The Model 3029/3039 CO2 Incubator . . . . . . . . . . . .4 - 1

    4.2 The CH/P Control System . . . . . . . . . . . . . . . . . . . . .4 - 1

    4.3 Standard Features of the 3029/3039 . . . . . . . . . . . . . .4 - 1

    4.4 Auxiliary Equipment . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 1

    Section 5 – Physical Installation . . . . . . . . . . . . . . . . . . .5 - 1

    5.1 Leveling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 1

    5.2 Attaching the Drain Lines . . . . . . . . . . . . . . . . . . . . . .5 - 1

    5.3 Connect to Power . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 1

    5.4 Shelf Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 1

    Section 6 - Power Module  . . . . . . . . . . . . . . . . . . . . . . . .6 - 1

    6.1 Main Power Switch and Indicating Light . . . . . . . . . .6 - 16.2 Temperature Limit Safety Thermostat . . . . . . . . . . . . .6 - 1

    6.3 Door Ajar Indicating Light . . . . . . . . . . . . . . . . . . . . .6 - 1

    6.4 Door Heaters and Control Potentiometer . . . . . . . . . .6 - 1

    Section 7 - Temperature Module  . . . . . . . . . . . . . . . . . .7 - 1

    7.1 Temperature Power Switch and Indicating Light . . . .7 - 1

    7.6 Temperature Calibration Procedure . . . . . . . . . . . . . .7 - 1

    Section 8 - Humidity Module  . . . . . . . . . . . . . . . . . . . . .8 - 1

    8.1 Humidity Power Switch and Indicating Light . . . . . .8 - 1

    8.2 CH/P Humidity Controller and Digital Display . . . . .8 - 1

    8.3 Humidity Indicating Light . . . . . . . . . . . . . . . . . . . . .8 - 1

    8.4 Setting the Humidity . . . . . . . . . . . . . . . . . . . . . . . . . .8 - 1

    8.5 Setting Up the Humidification System . . . . . . . . . . . .8 - 1

    8.6 Humidity Reservoir and Probe Check . . . . . . . . . . . .8 - 1

    8.7 Humidity Probe Placement . . . . . . . . . . . . . . . . . . . . .8 - 18.8 Humidity Calibration Check and Procedure . . . . . . . .8 - 2

    Section 9 - Carbon Dioxide  . . . . . . . . . . . . . . . . . . . . . . .9 - 1

    9.1 CO2 Power Switch and Indicating Light . . . . . . . . . .9 - 1

    9.2 CH/P CO2 Controller and Digital Display . . . . . . . . .9 - 1

    9.3 Audible CO2 Alarm and Alarm Indicating Light . . . .9 - 1

    9.4 Alarm Silence Push Button . . . . . . . . . . . . . . . . . . . . .9 - 1

    9.5 Alarm Disable Switch . . . . . . . . . . . . . . . . . . . . . . . . .9 - 1

    9.6 CO2 Inject Light . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 19.7 Sample Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 1

    9.8 Sealing the Cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 1

    9.9 CO2 Standardization Adjustment . . . . . . . . . . . . . . . .9 - 1

    9.10 Connecting the CO2 Source . . . . . . . . . . . . . . . . . . . .9 - 1

    911 Setting the CO2 Content . . . . . . . . . . . . . . . . . . . . . . .9 - 2

    9.12 CO2 Standardization Check . . . . . . . . . . . . . . . . . . .9 - 2

    9.13A CO2 Standardization Procedure . . . . . . . . . . . . . . .9 - 2

    9.13B Recommended CO2 Standardization Procedure . . .9 - 2

    9.14 CO2 Calibration Check and Procedure . . . . . . . . . . .9 - 2

    Section 10 - Preventive Maintenance  . . . . . . . . . . . . . . .10-1

    10.1 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-1

    10.2 Sterilization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-1

    Section 11 - Troubleshooting  . . . . . . . . . . . . . . . . . . . . . .11-1

    11.1 3029 Troubleshooting Guide . . . . . . . . . . . . . . . . . . .11-1

    11.2A Temperature Troubleshooting Overview . . . . . . . .11-1

    11.2A Temperature System Troubleshooting Overview . .11-111.3A Humidity System Troubleshooting Overview . . . .11-1

    11.3B Humidity System Troubleshooting Guide . . . . . . .11-2

    11.4A CO2 System Troubleshooting Overview . . . . . . . . .11-1

    11.4B CO2 System Troubleshooting Guide . . . . . . . . . . . .11-2

    Section 12 - Servicing  . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-1

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    4/52

    Section 1 - Inspection

    1.1 Preliminary Inspection

    Immediately upon receipt of the equipment, carefully

    examine the shipment for loss or damage. Make note of any

    visible damage. Unpack the shipment and inspect both interior 

    and exterior for any in-transit damage.

    1.2 Shipping Claims

    If any loss or damage is discovered, note any discrepancies

    on the delivery receipt and call the delivering carrier and

    request that their representative perform an inspection. Do not

    discard any of the packing material and do not move the ship-

    ment from the receiving area.

    1.3 Responsibility for Shipping Damage

    For products shipped F.O.B. Marietta, Ohio, the responsi-

     bility of Forma Scientific, Inc. ends when the merchandise is

    loaded onto the carrier’s vehicle.

    On F.O.B. Destination shipments, Forma Scientific and the

    carrier’s responsibility ends when your Receiving Department

     personnel sign a free and clear delivery receipt.

    Whenever possible, Forma Scientific will assist in settling

    claims for loss or in-transit damage.

    Models 3029 and 3039 ________________________________________________________________________ Inspection

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    5/52

    2.8 Temperature (Reference Section 7)

    Push in temperature set button and adjust setscrew untilthe desired temperature setpoint appears on the display.

    Remove screwdriver (knob at left of sample port). Turn

    temp power switch on.

    2.9 Stabilization

    Shut both doors and allow the incubator to run for 1

    hour.

    2.10 Humidity (Reference Section 8)

    Turn RH power switch on. Push in the humidity knob

    and use the coarse and fine adjustments to obtain the

    desired setpoint.

    2.11 Stabilization

    Shut both doors and allow the incubator to run for 6 to 8

    hours or overnight.

    2.12 CO2 (Reference Section 9)

    Turn the CO2 power switch on. Set the display to ‘00.0’

     by using the standardization adjustment, and use thecoarse and fine adjustments to obtain the desired set-

     point. Release the knob. Turn on the CO2 at the source.

    Allow the incubator to stabilize for 30 minutes. Check 

    the CO2 reading with a Fyrite. Adjust the display to

    match the Fyrite reading by using the standardization

    adjustment.

    2.13 Door Heater (Reference Section 6.4)

    If there is condensation of the inner glass door, adjust

    the door heater to 10.

    Models 3029 and 3039_________________________________________________________________Basic Set-Up Procedure

    Section 2 - Basic Set-Up Procedure

    2.1 Unpacking (Reference Section 1)

    Unpack the incubator and inspect for damage.

    Remove and discard the aluminum shipping plate from

    the front of the incubator.

    2.2 Leveling (Reference Section 5.1)

    Locate the cabinet and level, allowing 1/2” clearance

     beneath the cabinet.

    2.3 Drain Lines (Reference Section 5.2)

    Remove the valve shipped inside the cabinet and install

    in the drain line in place of the pipe plug. Attach tubing

    to thevalve

    .The overflow drain should be trapped and piped to an

    open drain and the trap filled with water.

    2.4 Water Supply (Reference Section 8.5)

    Remove the floor section from the incubator. Fill the

    floor pan with 1-1/2 gallons of distilled water.

    Fill the water bottle and allow the floor pan to fill. Thewater level should be 1/8” to 3/8” from the top of the

    overflow.

    Position the humidity probe so the tip is submerged 1/8”

    to 1/16” below the surface of the water.

    2.5 CO2 Supply (Reference Section 9.10)

    Connect a gaseous CO2 supply to the compression fittingon the back of the unit. Set the regulator to 4 to 5 PSIG.

    DO NOT turn on the CO2 supply at this point.

    2.6 Door Heater (Reference Section 6.4)

    T h d h k b ( b h b l ) h

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    6/52

    Section 3 - Specifications

    a. Capacity and Weight

    Chamber Capacity: 10 cu. ft. (285 liters)

    Shelf Capacity: 24 with 97 sq. ft. available space

    Shipping Weight: 275 lbs (117kg)

    b. Dimensions

    Exterior: 40.0”W x 38.5”H x 26.0”F-B

    (102cm x 98cm x 66cm)

    Interior: 27”W x 29”H x 22”F-B

    (69cm x 74cm x 56cm)

    c. Construction

    Exterior: 20 Gauge Stainless Steel 2-B

    Interior: 18 Gauge Stainless Steel 2-B

    Insulation: 1” (2.5cm) Urethane (Cabinet)

    2” (5.1cm) Fiberglass (Door)

    Shelves: (6) Stainless Steel

    d. Heaters

    Water: 300W, 120V, Stainless Steel

    Air: 300W, 120V, Wirewound Stainless Steel

    Door: 100W, 120V, Silicone Strip

    e. Temperature

    Control Sensitivity: ±0.01°CRange: +4°C above ambient to +60°C

    Repeatability: ±0.06°C

    Uniformity: ±0.25°C

    f. CO2

    Control Sensitivity: ±0.1%

    Range: 0% to 20%

    g. Humidity (at +21°C/50% RH ambient)

    Control: ±5%

    Performance: 50% to 96% at +60°C chamber temp

    70% to 96% at +37°C chamber temp

    Control Sensitivity: ±0.1%

    h. Electrical

    3029: 120VAC, 50/60Hz, 1 Phase, 6 FLA

    3039: 220VAC, 50/60 Hz, 1 Phase, 3.5 FLA

    i. Utility Connections

    CO2: 1/4” Compression Fitting

    Drains: 1/4” MPT Chamber Drain

    1/4” MPT Overflow Drain

    Water: Self-Contained 1 gallon (3.8 liters) Bottle -supplies chamber reservoir,

    external supply optional

    Models 3029 and 3039_________________________________________________________________________Specifications

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    7/52

    Section 4 - General Description

    4.1 The Model 3029/3039 CO2 Incubator 

    The Model 3029/3039 incubator is designed to provide

    continuous temperature, CO2, and humidity control in a

    10 cu. ft., (285 liters), forced draft environment.

    Separate solid state CH/P controllers provide direct digi-

    tal setpoint control and readout of temperature in 0.1°C

    increments, CO2 (within 0.1%), and relative humidity

    (within 1.0%). Setpoint and readout of all three parame-

    ters are presented on individual L.C.D. displays.

    The temperature range of the chamber is approximately

    4°C above ambient to +60°C. Temperature control is

    +0.1°C, and temperature uniformity is better than

    +0.25°C.

    The CO2 control system is also fully automatic, allowing

    the operator a CO2 content range of 0% to 20% with a

    control sensitivity of +0.1%.

    Rated in a +21°C/50 RH (relative humidity) ambient, a

    humidity range of 50% to 96% is achieved at +60°C

    chamber temperature, and a range of 70% to 96% is

    achieved at +37°C chamber temperature.

    4.2 The CH/P Control System

    The Forma CH/P control system for the 3029/3039 incu- bator provides independent control of temperature, CO2

     percentage, and humidity percentage. The system has

     been designed for easy troubleshooting and repair.

    The controls are easy to operate. Setpoints can be set by

    simply pushing in the set knob for the desired parameter,

    and turning to the coarse and fine adjustment to obtain

    the chosen setpoint. Individual digital displays continu-

    ally monitor all three parameters.

    An independent alarm system monitors deviations from

    CO2 setpoint and gives audible and visual warnings if 

    CO2 content deviates from setpoint by more than 1%

    (nominal). An alarm bypass switch deactivates the CO2

    alarm system and a spring-loaded reset temporarily

    silences the alarm warning signal.

    4.3 Standard Features of the 3029/3039

    Self-Diagnostic Control Design: The control systemshave been designed for easy troubleshooting by the cus-

    tomer. Not only does self-diagnosis decrease downtime,

     but it also saves on service calls.

    Positive Forced Air Convection: The air flow system

    circulates air horizontally across the shelving in a left-

    to-right direction. This design assures uniform distribu-

    tion of temperature and humidity, eliminates CO2 strati-

    fication, and assures rapid CO2 and RH recovery in the

    event of loss. Conditioned air is evenly distributed at allshelf levels under varying load conditions.

    Cabinet Construction: The interior is polished stainless

    steel and the exterior is constructed with heavy gauge

    stainless steel and finished with polyurethane enamel.

    Double wall construction is employed to maximize envi-

    ronmental stability and provide more economical opera-

    tion.

    Simplified Cleaning and Sterilization: Interior shelf sup- ports and side panels are easily removable for cleaning

    and sterilization. Side panels are bracket mounted and

    the floor panel is connected with two screws to make the

    humidity reservoir easily accessible.

    Door Construction: A full view safety glass inner door 

    seals onto a peripheral silicone gasket. The outer door is

    fitted with a peripheral magnetic gasket and seals

    against the cabinet exterior. The door heater minimizes

    the condensation on the inner door.

    Shelving Design: Solid stainless steel shelves are

    adjustable on 1" vertical center channel brackets. Six

    shelves are standard equipment, and the incubator has a

    24 shelf capacity to provide as much as 97 sq. ft. (9 sq.

    meters) of available space.

    4.4 Auxiliary Equipment

    Stacking Adapter: Required for vertical stacking of two

    cabinets, the adapter includes a frame support and an air 

     blower system. It adds 5" (12.7 cm) to the total height

    of two stacked units. (Forma Catalog Number 513028)

    Additional Shelves: Additional stainless steel, non-per-

    forated shelves may be purchased Incubator capacity is

    Models 3029 and 3039____________________________________________________________________General Description

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    8/52

    Models 3029 and 3039___________________________________________________________________ Physical Installation

    Section 5 – Physical Installation

    5.1 Leveling

    Locate the incubator on a firm, level surface in an area of 

    minimum ambient temperature fluctuation. Level the cabinet

     by using the four adjustable feet on the bottom of the unit. The

    incubator MUST be level to ensure proper operation. Be sure

    adequate clearance exists on the left side of the unit for access

    to the water supply and electrical components.

    Allow ½” minimum clearance at the bottom of the cabinet.

    If this area is blocked, the rated low limit for humidity will be

    affected.

    5.2 Attaching the Drain Lines

    There are two drain lines on the incubator; the chamber 

    drain and the overflow drain.To attach the chamber drain line:

    1) Remove the valve that was shipped inside the unit.

    2) Install the valve in place of the pipe plug in the drain on

    the back of the unit.

    3) Attach tubing to the valve and direct toward a drain.

    To attach the overflow drain line:

    1) The overflow drain should be rigid piped and trapped.2) Fill the trap with water.

    5.3 Connecting to Power 

    Connect the incubator to an adequate power source.

    3029: 120VAC, 50/60Hz, 1 Phase, 6 FLA

    3039: 220VAC, 50/60 Hz, 1 Phase, 3.5 FLA

    5.4 Shelf Mounting

    The shelves in the incubator have been designed to make

    cleaning easier.

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    9/52

    Section 6 - Power Module

    6.1 Main Power Switch and Indicating Light

    The Main Power switch controls all electrical power to the

    incubator. The switch must be ON for any of the incubator sys-

    tems to be functional.

    The main power indicating light is activated whenever the

    main power switch is in the ON position and the incubator is

    connected to power.

    6.2 Temperature Limit Safety Thermostat

    The temp limit safety thermostat is designed to be set for a

    temperature slightly higher than the operating setpoint tempera-

    ture of the incubator. In the event of an overtemp condition, the

    temp limit thermostat performs two functions:

    • Activates the audible alarm and overtemp indicator light.

    • Assumes control of power to immersion and air heaters.

    The temp limit control is not directly calibrated. The num-

     bers (0 to 10) on the dial are for reference only and do not cor-

    respond to any specific temperature.

    In the event of an overtemp condition, the alarm will

    sound, the overtemp indicating light will activate and the incu-

     bator heaters will cycle under the control of the temp limit ther-

    mostat. The temperature will be maintained at the temp limit

    setting until the source of the overtemp condition has been

    determined and corrected.Note: The temp safety control should be checked quarterly to

    ensure proper operation. To check the temp safety, turn the

    temp safety knob down toward the setpoint until the alarm and

    overtemp light are activated. Reset the temp safety after the

    test.

    a. Setting the Temp Limit Safety Thermostat

    Once chamber temperature is stable, the temp limit safety

    thermostat should be adjusted as follows:

    1. Turn the temp limit control knob slowly counterclock-

    wise until the alarm sounds and the overtemp light is

    activated.

    2 Turn the control knob clockwise no more than two units

    6.3 Door Ajar Indicating Light

    A Door Open indicating light is located on the power mod-

    ule. The red light activates when the outer door is open or hasnot been securely closed.

    When the Door Ajar light is on, power to the heaters, blow-

    ers and the CO2 injection solenoid is interrupted. This feature

    reduces gas usage and conserves energy.

    6.4 Door Heaters and Controller 

    Two heaters, installed in the outer door, can be used whenthere is condensation present on the inner glass door. The heater 

    control has to be adjusted to operate in conjunction with the

    regular temperature control system. The door heater control

    should never be set higher than necessary, or above 50. If set

    higher, it could have an adverse effect on the temperature con-

    trol of the incubator.

    The control potentiometer for the door heaters is located

    above the water bottle in the left side compartment.

    Use the door heaters only if there is excess condensationon the inner door.

    a. To Set the Door Heaters

    1. Allow the incubator to stabilize at the desired operating

    temperature.

    2. Locate the heater adjustment.

    3. Adjust the heater knob to 10. Allow the cabinet to stabi-lize overnight.

    4. If condensate is still present, adjust the door heater to

    15. Continue to make adjustments in increments of 5,

    allowing the cabinet to stabilize overnight with each

    adjustment, until there is no condensate present.

    It may be necessary to adjust the door heater after each

    change in the operating temperature setting.

    Note: Use of the door heater can slightly alter the temperaturein the chamber.

    Models 3029 and 3039_______________________________________________________________________ Power Module

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    10/52

    Section 7 - Temperature Module

    7.1 Temperature Power Switch and Indicating Light

    The temperature power switch controls electrical power to

    the temperature module. The power indicating light on the

    module is activated when both the main and temperature power 

    switches are in the ON position.

    7.2 CH/P Temp Controller and LCD Temp Display

    The LCD digital display on the temperature module contin-

    ually shows actual chamber temperature in °C, and the setpoint

    temperature when the temp set button is pushed.

    The temperature setpoint is changed by pushing the temp

    set knob in and rotating the knob to the desired temperature.

    Fine adjustments can be made to the setpoint immediately after 

    the knob is released.

    When the setpoint temperature is changed, the digital dis-

     play will show the new setpoint for 5-15 seconds before revert-ing to continuous readout of actual chamber temperature.

    Since the temperature sensor is located in the conditioning

     plenum, small changes in setpoint may result in the desired

    temperature being reached in less then 20 seconds. In such

    cases, the digital display will not indicate the progressive tem-

     perature change as actual temperature moves toward setpoint.

    Rather, only the new setpoint will be displayed.

    7.3 Heat Indicator Light

    The Heat indicator light activates when the temperature

    controller calls for more heat in the chamber.

    7.4 Setting the Temperature Control

    To set the temperature control, push in on the temp setknob and dial to the desired setpoint. Fine adjustments can be

    made immediately after releasing the knob.

    If the last digit on the display flickers after the chamber 

    temperature has stabilized, a slight readjustment of the setpoint,

    using the fine adjustment, will correct the problem.

    Models 3029 and 3039___________________________________________________________________ Temperature Module

    7.5 Temperature Calibration Check

    Note: To obtain the best results, all calibrations should be per-

    formed in the following order; temperature, humidity, CO2.The CH/P temperature controller has been calibrated prior 

    to shipment of the unit and should not require recalibration for 

    most laboratory applications. A periodic check of temperature

    control calibration, however, is recommended as a matter of 

    routine maintenance.

    A check of temperature calibration is especially advisable

    when the ambient temperature is less than 19°C or more than

    29°C.

    The calibration check should be performed with a calibrat-

    ing thermometer of known accuracy, using the following

    method.

    1. Place the calibrating thermometer inside the chamber,

    away from the door and side panels.

    2. Allow adequate time for the chamber temperature to sta-

     bilize. If the setpoint temperature has not been readjust-

    ed before the check, stabilization should be achieved

    within 15 minutes. If the temperature setpoint has been

    changed by 15°C or more before the calibration check,

    stabilization may require as long as 4 hours.

    3. Check to see if the temperature indicated by the digital

    display agrees with the temperature shown by the ther-

    mometer in the chamber.

    If the temperature controller requires recalibration, use the

     procedure in Section 7.6.

    7.6 Temperature Calibration Procedure

    Temperature calibration should be performed following the

    replacement of either the temperature module or the tempera-

    ture probe.

    A high quality, calibrating thermometer of known accuracy

    is required for this procedure. Place the thermometer near the

    center of the incubator chamber and allow the chamber temper-ature to stabilize. Refer to Section 7.5.

    1. Turn off power to the temperature module. Pull the mod-

    ule out and snap it onto the extender module. Re-install

    the module and turn the power on.

    2. Adjust switch S2 to position 7. See diagram in Section

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    11/52

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    12/52

    Section 8 - Humidity Module

    8.1 Humidity Power Switch and Indicating Light

    The humidity power switch controls electrical power to the

    humidity system. The power indicating light on the module is

    activated when both the main and humidity power switches are

    in the ON position.

    8.2 CH/P Humidity Controller and Digital Display

    The LCD digital display on the humidity module continual-

    ly shows the percent of humidity (RH) and the setpoint when

    the RH set button is pushed.

    The RH setpoint is changed by holding the RH set knob in

    and rotating the knob to the desired percentage. Fine adjust-

    ments can be made to the setpoint immediately after the knob is

    released.

    When the setpoint is changed, the digital display will showthe new setpoint for 5-15 seconds before reverting to continu-

    ous readout of the actual chamber RH.

    8.3 Humidity Indicating Light

    The humidity indicating light activates when the humidity

    controller calls for greater humidity percentage in the chamber.

    8.4 Setting the Humidity

    Put the humidity power switch in the ON position. Push in

    on the RH% knob and rotate the knob to the desired humidity

    setpoint percentage as indicated on the display.

    When running low humidity in the incubator, the excess

    water in the cabinet must be eliminated. The incubator should

     be free of all condensate in the chamber and under the floor 

     panel. The humidity reservoir should be at a normal operatinglevel.

    Excess water may be eliminated by leaving the chamber 

    door ajar for approximately one hour. Due to the door safety

    switch, the fan and heater will not operate when the door is

    open.

    Remove the floor section from the incubator. Fill the floor 

     pan with 1-1/2 gallons of water. This amount of water should

    almost fill the floor pan, but still be below the point where thefloat level is activated.

    Open the water bottle compartment on the left side of the

    cabinet and fill the bottle with distilled or demineralized water.

    Allow the chamber to fill with water until the float level is

    activated and the water ceases to flow. Let the water level sta-

     bilize overnight. Check the water level. It should be 1/8 to 3/8

    inches from the top of the overflow.

    If the water level in the floor pan is too high or too low, thefloat level control needs to be adjusted. Do this by adjusting

    the thumbscrew at the base of the float shaft and, if necessary,

    tilting the float mounting bracket.

    8.6 Humidity Reservoir and Probe Check

    The humidity reservoir water level should be checked peri-

    odically to ensure proper humidification levels in the chamber.

    The humidity reservoir is reached by removing the floor panel.

     Be sure to disconnect power before removing the floor panel.

    Make sure that the humidity probe tip is positioned so that

    the tip extends 1/8” to 1/16” below the float.

    The chamber reservoir water level should be maintained at

    1/8” to 3/8” below the overflow level as determined by the

     position of the 90° elbow located in the rear left corner of the

    reservoir.

    8.7 Humidity Probe Placement

    The humidity probe must be properly positioned for the incu-

    bator to perform optimally in registering humidity changes.

    To adjust the humidity probe, it is necessary to remove the

    chamber floor panel. The floor panel is connected by screws in

    the front corners of the chamber.

    Be sure to disconnect power before removing the floor panel

    Models 3029 and 3039_____________________________________________________________________ Humidity Module

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    13/52

    8.8 Humidity Calibration Check and Procedure

    Humidity calibration should be performed following thereplacement of the humidity module or the humidity probe.

    A hygrometer is required for this procedure. It should be

     placed inside the chamber, and the chamber should be permitted

    to stabilize for approximately 4 hours. During this time, the

    inner door should not be opened. If the hygrometer reading dif-

    fers from the digital display, recalibration is necessary.

    1. Turn off power to the humidity module. Pull the module

    out and snap it into the extender card. Turn the power 

     back on.

    3. Adjust switch S2 to position 1. See diagram in Section

    11.3 for the switch location.

    4. Turn the zero adjustment on the side of the module until

    the reading on the digital display agrees with the

    hygrometer reading.

    5. Turn switch S2 back to 0.

    6. Allow time for the chamber to stabilize. Compare the

    hygrometer reading with the reading on the digital dis-

     play. When the two readings agree, the system is cali-

     brated.

    Models 3029 and 3039____________________________________________________________________ Humidity Module

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    14/52

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    15/52

    Section 9 - Carbon Dioxide

    9.1 CO2 Power Switch and Indicating Light

    The CO2 power switch controls electrical power to the CO2

    system. The power indicating light on the module is activated

    when both the main and CO2 power switches are in the ON

     position.

    9.2 CH/P CO2 Controller and Digital Display

    The LCD digital display on the CO2 module continually

    shows the percent of CO2 and the setpoint when the CO2 set

    knob is pushed.

    The CO2 setpoint is changed by pushing the CO2 set knob

    in and rotating the knob until the desired percentage is indicat-

    ed on the display. Fine adjustments can be made to the setpoint

    immediately after the knob is released.

    When the setpoint is changed, the digital display will show

    the new setpoint for 5-15 seconds before reverting to continu-

    ous readout of actual percent CO2.

    9.3 Audible CO2 Alarm and Alarm Indicating Light

    The audible CO2 alarm and alarm indicating light activate

    when the percentage of CO2 deviates from setpoint by ±1%

    (nominal), or the CO2 set knob is pushed in..

    9.4 Alarm Silence Push Button

    The alarm silence button is located on the front of the CO2

    module. When pushed, it will silence the CO2 alarm, and de-

    energize the alarm light. The alarm will remain deactivated

    until another alarm condition causes it to sound.

    9.5 Alarm Disable Switch

    The alarm disable switch is located on the side of the CO2

    module. It is necessary to pull the CO2 module out slightly to

    allow access to the switch.

    9.6 CO2 Inject Light

    The CO2 Inject light is activated whenever there is ademand for CO2 to meet the setpoint percentage requirements.

    Since the CO2 inject light is independent of the CO2 alarm, it

    will continue to signal a need for CO2 when the CO2 alarm is

    set to either the Defeat or Silence position.

    9.7 Sample Port

    A sample port is located on the control panel for checking

    the CO2 percentage by independent means, such as with a

    Fyrite or similar instrument.

    To prevent CO2 loss, the sample port should be capped when

    not in use.

    9.8 Sealing the Cabinet

    When using CO2, the cabinet must be sealed before the unit

    is turned on. Check to ensure that the sample port is capped.

    Disconnect power to the unit and open the electrical access

     panel on the left side of the incubator. Check the Permagum

    seals on the lower right. Also check the orange plug in the

    upper left. If these openings are not sealed, CO2 will be drained

    from the cabinet and temperature and humidity will not stabi-

    lize properly.

    9.9 CO2 Standardization Adjustment

    The CO2 calibration adjustment is located to the right of 

    the CO2 set knob. This adjustment is used for standardizing the

    CO2 controller for specific ambient conditions.

    See Sections 9.12 and 9.13 for further information about

    standardizing the controller.

    9.10 Connecting the CO2 Source

    A main supply of liquid CO2, supplied from tanks without

    siphon tubes, will provide for the most economical use of CO2.

    It is also recommended that a t o stage press re reg lator ith

    Models 3029 and 3039______________________________________________________________________ Carbon Dioxide

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    16/52

    To connect the CO2 supply, insert the 1/4” O.D. metal tub-

    ing (or the 1/4” O.D. copper tubing piece supplied with the unit

    if plastic tubing is used) from the source into the CO2 compres-sion fitting, and tighten the nut one-and-a-quarter turn. As the

    nut is tightened, the ferrule (conical shaped ring) inside the nut

    exerts pressure on the tubing to hold it firmly in place. (If plas-

    tic tubing is used, the tubing should be firmly clamped to the

    copper tubing piece to ensure a leakproof seal.)

    The CO2 source must be regulated at a pressure level of 4 to 5

     PSIG. Higher pressure levels may damage CO2 filter system.

    9.11 Setting the CO2 Content

    The following conditions must be satisfied before the CO2

     percentage can be set:

    1. The temperature and humidity in the chamber must be

    allowed to stabilize for 6-8 hours or overnight, with the

    outer door closed. Since the CO2 sensor is affected bychanges in density and water content of the air, the incu-

     bator must be adjusted for site conditions of temperature

    and humidity.

    2. Check the CO2 control standardization (Section 9.12).

    Although all units are standardized at the factory before

    shipment, the CO2 control standardization may not be

    correct for the ambient conditions in which the incubator 

    is to be used. A check is therefore mandatory to ensure

     proper operation of the CO2 system. Read Section 9.12

    and 9.13 before using the incubator with CO2.

    To set the CO2 percentage, turn the CO2 power switch ON.

    Depress the CO2 set knob and rotate the knob until the desired

     percentage is indicated on the digital display.

    9.12 CO2 Standardization Check

    Check standardization with every adjustment of the tempera-

    ture setpoint and/or change of the humidity setting.

    The check procedure requires the use of a Fyrite CO2 tester 

    or similar measuring device However it is possible to stan

    To check the standardization, allow chamber temperature

    and humidity to stabilize, then sample the CO2 content using

    the measuring device and compare the test instrument readingwith the incubator display. Two samples should be taken con-

    secutively to allow for any variation in temperature and humidi-

    ty. If the readout does not match the test instrument, perform

    one of the standardization procedures in Section 9.13.

    9.13 Standardization Procedure

    a. Procedure A (CO2 Tester Not Required)

    1. Disconnect CO2 at source (if connected) and fan the

    door about 10 times to eliminate CO2 from chamber.

    2. Allow the chamber temperature and humidity to stabi-

    lize for at least 4 hours. Do not open door during stabi-

    lization.

    3. Locate the CO2 standardization adjustment located on

    the control panel.

    4. Using a jeweler’s screwdriver, turn the standardizationadjustment slowly counterclockwise until the digital dis-

     play reads 00.0%. Allow one minute for stabilization.

    The unit is now standardized and can be adjusted to the

    desired CO2 operating setpoint.

    b. Procedure B (CO2 Tester Required)

    1. Disconnect CO2 at source (if connected) and fan thedoor about 10 times to eliminate CO2 from chamber.

    2. Allow the chamber temperature and humidity to stabi-

    lize for at least 4 hours. Do not open door during stabi-

    lization.

    3. Locate the CO2 standardization adjustment located on

    the control panel.

    4. Using a jeweler’s screwdriver, turn the standardization

    adjustment slowly counterclockwise until the digital dis- play reads 00.0%. Allow one minute for stabilization.

    5. Connect the CO2 source and set the CO2 controller for 

    the desired setpoint and allow approximately 30 minutes

    for stabilization.

    6. Using the CO2 tester, read the actual CO2 percentage in

    Models 3029 and 3039_______________________________________________________________________Carbon Dioxide

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    17/52

    9.14 CO2 Calibration Check and Procedure

    CO2 calibration should be performed following the replace-ment of the CO2 module or the CO2 sensor.

    A Fyrite or similar CO2 test device is required for this pro-

    cedure. The procedure must be followed carefully to ensure

     proper calibration of the CO2 controller. Test the CO2 content

    with the test instrument. If the reading differs from the reading

    on the digital display, recalibration is necessary.

    1. Disconnect the CO2 source and eliminate the CO2 from

    chamber by fanning the door 10 times.

    2. Close the doors and allow sufficient time for the temper-

    ature and humidity in the chamber to stabilize.

    Stabilization may take up to 4 hours. Do not open the

    door during the stabilization period.

    3. Turn off power to the CO2 module. Pull the module out

    and snap it into the extender card. Re-install the module

    and turn the power back on.

    4. Adjust both the coarse and fine adjustments on the CO2

    setpot fully counterclockwise.

    5. Wait 60 seconds and change the switch S2 to position 1.

    Turn the current adjustment on the side of the module

    until the display reads 07.0. Return switch S2 to 0.

    6. Turn the calibration adjustment on the front panel until

    the display reads 00.00.

    6. Adjust the CO2 setpoint to 5%.

    7. Reconnect or turn on the CO2 supply to the chamber andallow time for the system to stabilize. In most cases, 30

    minutes should be sufficient.

    8. Using the test instrument, sample the actual CO2 level in

    the chamber. Note the reading on the digital display.

    9. To prevent additional CO2 from entering the chamber,

    turn the CO2 setpoint knob fully counterclockwise.

    10. Adjust the span pot on the CO2 module until the reading

    on the digital display agrees with the test instrumentreading.

    Note: This calibration should be suitable for most laboratory

    applications. If more precise calibration is necessary, this proce-

    dure should be repeated.

    Models 3029 and 3039_______________________________________________________________________Carbon Dioxide

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    18/52

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    19/52

    Section 10 - Preventive Maintenance

    10.1 Cleaning

    The chamber interior should be cleaned with alcohol and/or 

    soap and water. The interior panels should be disinfected with a

    general-use laboratory disinfectant. Information about disinfec-

    tants recommended by Forma Scientific is included as a supple-

    ment at the end of the manual.

    The cabinet exterior may be cleaned with soap and water,

    or any non-abrasive commercial spray cleaner.

    See Section 10.2 for a sterilization procedure.

    10.2 Sterilization

    The incubator can be easily sterilized in approximately 30

    minutes. Full strength Amphl Spray or O-Syl disinfectant in 1%

    or 3% solutions (dilute with sterile distilled water) is recom-mended.

    Remove the shelves, interior duct sheets and bottom from

    the incubator and set them aside. Spray all interior surfaces.

    Sponge down all surfaces to remove any foreign matter, and

    rinse the surfaces at least twice with distilled water.

    Be sure that the door gasket is also thoroughly cleaned. It

    may be necessary to lift up the edge of the gasket to clean it

    thoroughly. Clean the interior of the glass door with the solu-tion, and rinse it with distilled water. The drip trough should be

    wiped out and carefully rinsed.

    Run some of the solution through the drain lines.

    If desired, all surfaces can then be rubbed or sprayed with

    70% alcohol. Be sure to drain and sponge all solution from the

    incubator. All shelves, duct sheets, and removable components

    can be washed down with the same solution before they are

    returned to the incubator.

     Do not direct a hard spray of the solution at the temperature

    sensor, or rub it directly. To do so could bend the sensor wire.

     Do not use strong alkaline or caustic agents. Stainless steel is

    corrosion-resistant, not corrosion-proof.

    The use of volatiles or aromatics is not recommended for 

    cleaning cabinet interiors. Residues could cause contamina-

    tion of the cabinet environment.

    General Instructions

    Use the mildest cleaning procedure that will do the job

    effectively. To ensure maximum effectiveness and to avoid mar-

    ring the surface, always rub in the direction of the finish polish

    lines. Be sure to rinse the surface thoroughly after each clean-

    ing operation. To avoid water marks, wipe the surface dry.

    a. Passivation Procedure for Surface

    Contamination Removal

    Note: Safety requires use of rubber gloves and eye protection.

    Equipment required:

    1. Nitric acid 70 to 80 percent

    2. Non-metallic mixing container at least 2 qt. capacity

    (2) required3. Acid brush

    4. Rubber gloves

    5. Goggles

    6. Bicarbonate of soda

    1) Procedure A - Vertical surfaces

    1. Dilute acid 4 to 1 with water (add acid to water slowly).

    2. With acid brush, apply nitric acid half strength to areasshowing any signs of oxidation.

    3. Leave acid on surface for 30 to 60 minutes depending

    upon severity of blemish.

    4. Prepare a solution of sodium bicarbonate and water 

    (approximately 4 oz. per quart).

    5. With cloth dipped in sodium bicarbonate solution, rinse

    acid-treated surfaces to neutralize acid.

    6. Final rinse with plain water.

    2) Procedure B - Horizontal surfaces (tank bottoms, etc.)

    1. Dilute acid 4 to 1 with water (add acid to water slowly).

    2. Pour acid until surface is just covered.

    Models 3029 and 3039_________________________________________________________________Preventive Maintenance

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    20/52

    4. Remove acid.

    5. Prepare a solution of sodium bicarbonate and water 

    (approximately 4 oz. per quart). Pour solution until sur-

    face is just covered and let stand for 15 minutes.6. Remove this solution and finally rinse with plain water.

    Note: Rusting of 300 Series stainless steel is purely a surface

    effect and can be caused by iron carbide precipitation at joints

    heated by welding or contact with carbon steel chips, tools, etc.

    Surface treatment will remove this contamination.

    Models 3029 and 3039________________________________________________________________Preventive Maintenance

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    21/52

    1  0 - 3 

    M o d  e l   s  3  0 2  9  a n d  3  0  3  9  _ _ _ _ _ _ _ _ _ _ _ _ _

     _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

     _ _ _ _ _ _ _ _P r  e  v e n t  i   v e M a i  n

     t   e n a n c  e 

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    22/52

    1  0 -4 

    M o d  e l   s  3  0 2  9  a n d  3  0  3  9  _ _ _ _ _ _ _ _ _ _ _ _ _

     _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

     _ _ _ _ _ _ _ _P r  e  v e n t  i   v e M a i  n

     t   e n a n c  e 

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    23/52

    1  0 - 5 

    M o d  e l   s  3  0 2  9  a n d  3  0  3  9  _ _ _ _ _ _ _ _ _ _ _ _ _

     _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

     _ _ _ _ _ _ _ _P r  e  v e n t  i   v e M a i  n t   e n a n c  e 

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    24/52

     

     

     

     

     

         

     

    Models 3029 and 3039_____________________________________________________________________Troubleshooting

    Section 11 - Troubleshooting

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    25/52

    Models 3029 and 3039______________________________________________________________________Troubleshooting

     

    d l 3029 d 3039 bl h i

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    26/52

    Models 3029 and 3039_____________________________________________________________________Troubleshooting

    11.2 Temperature Troubleshooting Overview

    a. General Instructions

    Follow the instructions on the flowchart to diagnose prob-

    lems with the temperature system. For this diagnostic aid to be

    effective, all checks must be made in the sequence indicated by

    the flowchart and all data carefully recorded in the table provid-

    ed.

    b. Temperature System Troubleshooting

    Turn the temp module off and remove the temperature and

    logo modules. Transfer the extender card from the logo module

    to the module slot. Check the alignment. Add the temp module

    to the extender and check the alignment.

     Do not plug the module in upside-down. To do so will damage

    the unit.

    General Notes

    • Before removing any control module from the unit, turn

    the power to that module off.

    • Unless otherwise noted in the procedure, the module

    referred to is the temp module. The switches referred to are

    located on the temp module.

    • Switch locations for the temp module are labeled on the

    following module diagram.• When rotating set knob fully in one direction, it will be

    necessary to push the knob in at intervals to get the knob to

    rotate completely.

    • When taking readings with the set knob both in and out,

    wait 20 seconds after taking the reading with the knob in

     before taking the reading with the knob out.

    • See Drawing 3029-51-0-D for location of the plug.

    11.3 Humidity Troubleshooting Overview

    a. General Instructions

    Follow the instructions on the flowchart to diagnose prob-

    lems in the humidity system. For this diagnostic aid to be effec-

    tive, all checks must be made in the sequence indicated by the

    flowchart and all data must be carefully recorded.

    b. Humidity System Troubleshooting Procedure

    Turn the RH module off and remove the RH and logo mod-ules. Transfer the extender card from the logo module to the

    RH module slot. Snap the extender card into the connector.

    Check the alignment. Add the RH module to the extender and

    check the alignment.

     Do not plug the module in upside-down. To do so will damage

    the unit.

    General Notes

    • Before removing the control module from the unit, turn the

     power off.

    • Unless otherwise noted in the procedure, the module

    referred to is the RH module. The switches referred to are

    located on the RH module.

    • Switch locations for the RH module are labeled on the

    module diagram.• When rotating set knob fully in one direction, it will be

    necessary to push the knob in at intervals to get the knob to

    rotate completely.

    • When taking readings with the set knob both in and out,

    wait 20 seconds after taking the reading with the knob in

     before taking the reading with the knob out.

    • Recalibration of any of the controls may be necessary fol-

    lowing component replacement.• Place the thermometer through the hole on the front of the

    incubator floor. The thermometer should be lowered until it

    touches the bottom of the humidity pan.

    M d l 3029 d 3039 T bl h ti

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    27/52

    Models 3029 and 3039_______________________________________________________________________Troubleshooting

    Log RH readings on this table.

    Display Data LimitsS2 SettingSilence/Set Button Out Silence/Set Button In

    Your Results

    1 80 through 90%, ±5%  N/A

    2 64.0 through 96.0  N/A

    3 See temp setpoint  N/A

    4 40 0 h h 60 0 N/A

    Log temp readings on this table.

    Display Data LimitsS2 SettingSilence/Set Button Out Silence/Set Button In

    Your Results

    1 120.0 through 180.0  N/A

    *2 64.0 through 96.0  N/A

    3 -120.0 through –180.0  N/A

    *4 -40.0 through –60.0  N/A

    5 20.0 through 30.0  N/A

    6 20.0 through 60.0

    7 See thermometer    N/A

    8 -40.0 through –60.0 50.0 through 96.0

    *Limits are for temp or RH.

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    28/52

     

     

     

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    29/52

     

     

     

     

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    30/52

     

     

     

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    31/52

     

     

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    32/52

     

     

     

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    33/52

     

     

     

       

     

     

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    34/52

     

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    35/52

     

    Models 3029 and 3039_____________________________________________________________________Troubleshooting

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    36/52

    11.4 CO2 System Troubleshooting Overview

    a. General Instructions

    Follow the instructions on the flow chart to diagnose prob-

    lems with the CO2 controller. For this diagnostic aid to be effec-

    tive, all checks must be made in the sequence indicated by the

    flow chart and all data must be carefully recorded.

    b. Stabilizing the Cabinet

    Prior to beginning the procedure, shut the cabinet door andset the temperature and humidity at normal operating levels.

    The temperature and humidity should be stabilized by Step #12

    of the procedure.

    c. CO2 System Troubleshooting Procedure

    Turn the CO2 module off and remove the CO2 and logo

    modules. Transfer the extender card from the logo module to

    the CO2 module slot. Snap the extender card into the connector.Check the alignment. Add the CO2 module to the extender and

    check the alignment.

     Do not plug the module in upside-down. To do so will damage

    the unit.

    General Notes

    • The first eleven steps of the procedure can be completed

    with the temperature and humidity in an unstable condi-

    tion.

    • The first fifteen steps can be performed with the CO2

    turned off.

    • Before removing the control module from the unit, turn the

     power to the module off.

    • At various times in this procedure, the CO2 alarm may

    sound. Unless otherwise noted, the alarms may be silenced

     by pressing the alarm silence button.

    • Unless otherwise noted in the procedure, the module

    referred to is the CO2 module and the switches referred to

    are located on the CO2 module.

    Models 3029 and 3039_______________________________________________________________________Troubleshooting

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    37/52

    Log CO2 readings on this table.

    Display Data LimitsS2 SettingSilence/Set Button Out Silence/Set Button In

    Your Results

    1 15.0 through 25.0  N/A

    2 approximately ½ reading 3  N/A

    3 35.0 through 50.0  N/A

    4 See Fyrite™  N/A

    5 00.2 through 20.0  N/A

    6  N/A N/A

    7 96.0 through 140.0  N/A

    8 -40.0 through –60.0  N/A

       

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    38/52

     

         

     

     

     

         

     

     

       

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    39/52

     

     

      

       

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    40/52

     

     

       

     

     

       

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    41/52

     

     

     

     

     

    Models 3029 and 3039_______________________________________________________________________________Service

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    42/52

    Section 12 - Service

    12.1 Electrical Servicing

    Access to all electrical components is provided by opening

    the component compartment door on the left side of the cabinet.

    The door is secured by two 1/4 turn fasteners.

    The CO2, temperature, and humidity modules can be easily

    removed. Turn the power to the particular module OFF. Pull the

     black button out with a suitable tool to release the module, and

    carefully ease the module out of the panel.The Model 3029 incubator has been designed so that no

    internal electrical servicing should be required under conditions

    of normal use.

     If electrical servicing should become necessary, it should be

     performed by qualified service personnel only!

    For electrical information, consult the electrical schematic(Drawing #3029-50-0-D) and/or the wiring diagrams (Drawing

    #3029-51-0-D, #3029-53-0-D) included at the end of manual.

    For maximum safety, always disconnect the power before

    servicing.

    12.2 Temperature Sensor Replacement

    The temperature sensor is located behind the left side duct

    sheet, inside the incubator. To remove and replace the sensor:

    1. Turn the power switch to OFF and disconnect the unit

    from the power source.

    2. Remove the shelves from the incubator. Loosen the two

    screws that secure the floor of the incubator. Lift out the

    floor and set it aside.

    3. Carefully lift the left side duct sheet from its mounting

     brackets and set it aside.

    4. Locate the temp sensor (#245054 on drawing #3029-53-

    0-D, Detail 1). It is mounted on the top rear of the left

    interior wall.

    5. Unscrew the two screws that attach the sensor to the

    7. Set the temp sensor on the floor of the unit.

    8. Open the side component compartment and locate the

    motherboard (#285055 on drawing #3029-51-0-D).

    9. Locate the temp sensor (plug #1 on the mother board).

    10. Unplug the temp sensor, taking care not to pinch the

     plug or removal will be difficult.

    11. Carefully feed the temp plug and lead wire through the

    access port and into the chamber. It may be necessary to

    clear the Permagum from the port. Remove any tie

    wraps to permit the plug to move freely into the cham-

     ber. The Permagum should be replaced and the port

    sealed after the sensor is replaced.

    12. Remove the sensor from the chamber.

    13. Replace the sensor with part number #245054 by revers-

    ing the above procedure. Be sure to seal all screws with

    silastic.

    14. The temperature controller may need to be recalibrated

    following this procedure. Perform a temperature calibra-

    tion check (Section 7.5) to determine if recalibration is

    necessary.

    12.3 Replacing the Chamber Heater 

    The chamber heater is located under a protective plate

     beneath the floor of the incubator. To remove and replace the

    chamber heater:

    1. Turn the power switch to OFF and disconnect the unit

    from the power source.2. Remove the shelves from the incubator. Loosen the two

    screws that secure the floor of the incubator and set the

    floor aside.

    3. Locate the heater (#600210 on drawing #3029-53-0-D).

    It is mounted to the protective plate.

    4. While holding the nut beneath the each screw, remove

    the four screws from the top of the mounting plate.

    Note: The heater will drop down from the mounting plate whenthe screws are removed.

    5. Carefully pivot the heater out from under the bracket.

    6. Remove the nuts from the heater connections and

    remove the wires.

    Models 3029 and 3039_______________________________________________________________________________Service

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    43/52

    12.4 Replacing the Humidity Probe

    The humidity probe is located inside the incubator beneath

    the floor. To remove and replace the humidity probe:1. Turn the power switch to OFF and disconnect the unit

    from the power source.

    2. Remove the shelves from the incubator. Loosen the two

    screws that secure the floor of the incubator and set the

    floor aside.

    3. Locate the humidity probe (#285053 on drawing #3029-

    53-0-D, Detail 2). It is held to the float at the front of the

    water pan.4. Remove the probe by pulling it upward and out of the

    holder.

    5. Unscrew the silver ground wire from the float mounting

     bracket.

    6. Feed the humidity probe through the bushing in the float

    mounting bracket.

    7. Open the side component compartment and locate the

    motherboard (#285055 on Drawing 3029-51-0-D.)

    8. Locate the humidity probe plug (plug #2 on the mother-

     board).

    9. Unplug the humidity sensor, taking care not to pinch the

     plug as removal will become more difficult.

    10. Carefully feed the humidity plug and lead wire through

    the access port and into the incubator chamber.

    Note: It will be necessary to clear the Permagum from the port

    and clip any tie wraps to permit the plug to move freely into the

    chamber. The Permagum should be replaced and the port sealed

    after the probe is replaced.

    11. Remove the probe from the incubator chamber.

    12. Replace the probe with part #245053 by reversing the

    above procedure.

    13. When replacing the probe, be sure that the tip extends

    only about 1/16” below the surface of the water.

    14. The humidity controller may need to be recalibrated fol-

    lowing this procedure. Make a humidity calibration

    check. (See Section 8.8.)

    12.5 Replacing the Immersion Heater 

    3. Locate the immersion heater (#610005 on drawing

    #3029-53-0-D, Detail 2). It is clipped to the floor of the

    humidity reservoir.

    4. Remove the nuts from the clips securing the heater to

    the floor.

    5. Note the color of the heater lead wires.

    6. Open the component compartment and locate the heater 

    wires (labeled Wet Heater on Drawing #3029-51-0-D).

    Disconnect the heaters by removing the wire nuts. Be

    sure to note the wiring configuration.

    7. Carefully feed the heater wire leads through the access port and into the incubator chamber.

    Note: It may be necessary to clear Permagum from the port and

    remove any tie wraps that restrict passage of the heater wires.

    The Permagum should be replaced and the port sealed after the

    heater is replaced.

    8. Remove the heater from the chamber.

    9. Replace the heater with part number #610005 by revers-

    ing the above procedure.

    12.6 Replacing the CO2 Sensor 

    The CO2 sensor is located inside the incubator behind the

    left duct sheet. To remove and replace the CO2 sensor:

    1. Turn the power switch to OFF and disconnect the unit

    from the power source.

    2. Remove the shelves from the incubator. Loosen the two

    screws that secure the floor of the incubator and set the

    floor aside.

    3. Carefully lift out the left duct sheet from its mounting

     bracket and set it aside.

    4. Locate the CO2 sensor (#245057 on Drawing #3029-53-

    0-D, Detail 1). It is mounted near the center of the left

    interior wall.

    5. Remove the two screws that attach the sensor to the

    wall. Note how the washers and tie wraps are posi-

    tioned. Retain the screws and washers to mount the new

    sensor. Remove the tie wrap holding the temp sensor 

    lead wire to the CO2 sensor block.

    Note: It is very important that the tie wrap is replaced when the

    Models 3029 and 3039_______________________________________________________________________________Service

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    44/52

    10. Carefully feed the CO2 sensor plug and lead wire

    through the access port and into the incubator chamber.

    Note: It may be necessary to clear Permagum from the port and

    remove any tie wraps that restrict movement of the lead wire.

    The Permagum should be replaced and the port sealed after the

    sensor is replaced.

    11. Remove the sensor from the chamber.

    12. Replace the sensor with part number #245057 by revers-

    ing the above procedure. Be sure to seal all screws with

    silastic.

    13. The CO2 controller may need to be recalibrated follow-

    ing this procedure. See Section 9.14 for the CO2 calibra-

    tion check procedure.

    12.7 Replacing the CO2 Solenoid Valve

    The CO2 solenoid valve is located in the component com-

     partment. To remove and replace the solenoid valve:

    1. Turn the power switch to OFF and disconnect the unit

    from the power source.

    2. Open the component compartment and locate the sole-

    noid valve (#250003 on Drawing #3029-51-0-D).

    3. Follow the wire from the solenoid to the motherboard

    and disconnect plug #3 from the board.

    4. Remove the tygon tubing from both sides of the sole-

    noid.

    5. Remove the two screws securing the solenoid to the wall

    of the incubator.

    6. Remove the screws from the bottom of the solenoid

    mounting and remove the solenoid. Note the position of 

    the energy-absorbing pads.

    7. Replace the solenoid valve with part number #250003

     by reversing the above procedure. Place the energy-

    absorbing pads in their original positions.

    12.8 Replacing the Blower 

    The blower is located inside the component compartment,

    with a portion of it extending into the incubator chamber. Toremove and replace the blower:

    1. Turn the power switch to OFF and disconnect the unit

    from the power source.

    2. Remove the shelves from the incubator. Loosen the two

    screws that secure the floor of the incubator and set the

    floor aside.

    3. Carefully lift out the left duct sheet from its mounting

     bracket and set it aside.4. Locate the blower wheel (#131006 on Drawing #3029-

    53-0-D, Detail 1).

    5. Loosen the blower wheel with an Allen wrench, and lift

    the wheel from the shaft.

    6. Open the component compartment and locate the blower 

    motor (#810700 on Drawing #3029-51-0-D).

    7. Remove the four screws securing the blower and pull

    the blower out of the incubator wall.

    Note: The area around the blower shaft has been sealed with

    silastic. As a result, the blower may be difficult to remove. It

    may be necessary to use a razor knife to cut the silastic.

    Remember to seal around the new motor with silastic to prevent

    leakage from the incubator.

     Do not cut, or otherwise damage, the blower mount gasket.

     It will be necessary to reuse this gasket with the new blower motor.

    8. Remove the wiring connections to the blower motor.

     Note the positions of the wires.

    9. Replace the blower motor with part number #810700 by

    reversing the above procedure.

    Models 3029 and 3039_________________________________________________________________________Parts List

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    45/52

    Section 13 - Parts List

    Model 3029

    Qty Stock # Description

    1 285812 Plug, 2 Pole, 3 Wire, 15 Amp, Nylon

    1 950020 Cord Set, 16/3 SJT, 8 ft.

    1 260039 Switch, Toggle, DPST

    1 260042 Switch, Sealed Snap Action

    1 270010 Thermostat, Low-temp, B-10

    1 285055 Board, Circuit, Mother  

    1 285660 Knob, 1-1/4”, Blue w/ Clear Skirt

    1 290113 Light, Pilot, Red, 5/16 Neon, 115V

    8 124001 Guides, Card

    1 285057 Extender, Module Board

    1 510410 Grommet, Nylatch (for modules)

    1 510411 Plunger, Nylatch (for modules)

    1 231002 Module, CO2 Control

    1 231003 Module, Temp Control

    1 231004 Module, Humidity Control

    1 131006 Wheel, Blower  

    2 136000 Heater, Exterior Door  

    1 205025 Filter, RFI/EMI

    2 250013 Relay, DPDT, Enclosed

    1 270080 Switch, Infinite Heat

    1 285639 Switch, Dial Heat

    1 285660 1-1/4” Knob

    1 285798 Snap-In Outlet

    1 285805 Receptacle, Door  

    9.5 400296 Gasket, Feather, Inner Door 

    Model 3029 (continued)

    Qty Stock # Description

    2.8 431125 Magnet, 1”

    2.9 431124 Magnet, 1/2”

    1 431144 Gasket, Outer Door  

    1 463028 Door, Glass, Inner  

    1 600210 Heater, 300W, Air  

    1 610005 Heater, Immersion, 300W

    1 810700 Motor, Blower  

    1 950015 Cord, Electrical, Door, w/ Plug

    1 890139 Gallon Bottle

    1 60000 Clamp, Water Bottle

    1 980055 Valve, Float

    1 285020 Alarm, Sonalert, 110V

    1 245057 Sensor Assembly, CO2, w/ Plug

    1 250003 Valve Assembly, Solenoid, CO2

    1 245054 Sensor Assembly, Temperature

    1 245053 Sensor Assembly, Humidity

    1 400008 Module, Power Supply

    2 250021 Relay, Solid State, 10A, 240VAC

    4 515081 Foot, Leveling, Adjustable

    1 195105 Filter, CO2 Inlet

    1 960001 Orifice, CO2 Injection

    1 136005 Heater, Silicone Rubber, 6W, 120V

    1 475576 Door Hinge, Interior  

    2 505546 Door Hinge, Exterior  

    6 503028 Shelf  

    Models 3029 and 3039_____________________________________________________________________________Parts List

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    46/52

    Model 3039

    1 950045 Cord, 16/3 SJT, 8 ft.

    1 260039 Switch, Toggle, DPST1 260042 Switch, Sealed Snap Action

    1 270010 Thermostat, Low-temp, B-10

    1 285055 Board, Circuit, Mother  

    1 285660 Knob, 1-1/4”, Blue w/ Clear Skirt

    3 290113 Light, Pilot, Red, 5/16 Neon, 115V

    1 285057 Extender, Module Board

    1 510410 Grommet, Nylatch (for modules)

    1 510411 Plunger, Nylatch (for modules)1 231002 Module, CO2 Control

    1 231003 Module, Temp Control

    1 231004 Module, Humidity Control

    1 131006 Wheel, Blower  

    2 136001 Heater, Exterior Door  

    1 205025 Filter, RFI/EMI

    2 300065 Relay, DPDT, Enclosed

    1 270090 Switch, Infinite Heat1 285639 Switch, Dial Heat

    1 285660 1-1/4” Knob

    1 285821 Receptacle, 125V, 15A

    1 285805 Receptacle, Door  

    9.5 400296 Gasket, Feather, Inner Door 

    2.8 431125 Magnet, 1”

    2.9 431124 Magnet, 1/2”

    1 431144 Gasket, Outer Door  1 463028 Door, Glass, Inner  

    1 600211 Heater, 300W, Air  

    1 610006 Heater, Immersion, 300W

    Model 3039 (continued)

    Qty Stock # Description

    1 810701 Motor, Blower  1 950015 Cord, Electrical, Door, w/ Plug

    1 60000 Clamp, Water Bottle

    1 980055 Valve, Float

    1 285020 Alarm, Sonalert, 110V

    1 245057 Sensor Assembly, CO2, w/ Plug

    1 250003 Valve Assembly, Solenoid, CO2

    1 245054 Sensor Assembly, Temperature

    1 245053 Sensor Assembly, Humidity1 400008 Module, Power Supply

    2 250021 Relay, Solid State, 10A, 240VAC

    4 515081 Foot, Leveling, Adjustable

    1 195105 Filter, CO2 Inlet

    1 960001 Orifice, CO2 Injection

    1 136006 Heater, Silicone Rubber, 6W, 120V

    3 1210104 Resistor, 100K Ohm 1/4W 10%

    1 1211822 Resistor, 8.2K Ohm 1/2W 10%1 505522 Latch Tab

    1 505523 Latch Stud

    1 890139 Gallon Bottle

    1 60000 Clamp

    1 34014 Plug 1/2”

    1 34028 Cap Vinyl

    1 34053 Plug 1-1/4”

    1 515072 Latch, Pushbutton6 503028 Shelf  

    1 195125 Bacterial Air Vent Filter  

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    47/52

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    48/52

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    49/52

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    50/52

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    51/52

  • 8/17/2019 User Manual - Forma - 3029, 3039 - Countertop Incubator 10 Cu.ft. - 7013029[1] Copy

    52/52


Recommended