+ All Categories
Home > Documents > vbSeries® Portable Instrument Hazardous Area Portable Instrument Hazardous Area Certification...

vbSeries® Portable Instrument Hazardous Area Portable Instrument Hazardous Area Certification...

Date post: 11-Mar-2018
Category:
Upload: ngotu
View: 216 times
Download: 1 times
Share this document with a friend
10
vbSeries® Portable Instrument Hazardous Area Certification Class I, Division 2 1 GE Measurement & Control vbSeries® Portable Instrument Hazardous Area Certification Class I, Division 2 Class I, Division 2, Groups A, B, C, D certification has been granted by the Canadian Standards Association for CSA labeled variants of the following instrument models: vb5™ Data Collector vb6™ Data Collector vb7™ Vibration Analyzer vb8™ Vibration Analyzer vbBalancer™ Imbalance Correction Instrument vbBalancer+™ Imbalance Correction Instrument See attached certificate for details. Notes 1. ONLY instruments bearing the CSA logo and Hazardous Area information are certified. vbSeries instruments without the CSA logo are NOT CERTIFIED. 2. The use of noncertified sensors and/or accessories WILL VOID instrument Class I, Division 2 CSA certification.
Transcript
Page 1: vbSeries® Portable Instrument Hazardous Area Portable Instrument Hazardous Area Certification Class I, Division 2 5 GE Measurement & Control vbX /S COUT INSTALLATION CONTROL DRAWING

vbSeries® Portable Instrument Hazardous Area Certification Class I, Division 2 1

GE Measurement & Control

vbSeries® Portable Instrument Hazardous Area

Certification Class I, Division 2

Class I, Division 2, Groups A, B, C, D certification has been granted by the Canadian Standards Association for CSA labeled variants of the following instrument models:

• vb5™ Data Collector • vb6™ Data Collector • vb7™ Vibration Analyzer • vb8™ Vibration Analyzer • vbBalancer™ Imbalance Correction Instrument • vbBalancer+™ Imbalance Correction Instrument

See attached certificate for details.

Notes

1. ONLY instruments bearing the CSA logo and Hazardous Area information are certified. vbSeries instruments without the CSA logo are NOT CERTIFIED.

2. The use of noncertified sensors and/or accessories WILL VOID instrument Class I, Division 2 CSA certification.

Page 2: vbSeries® Portable Instrument Hazardous Area Portable Instrument Hazardous Area Certification Class I, Division 2 5 GE Measurement & Control vbX /S COUT INSTALLATION CONTROL DRAWING

vbSeries® Portable Instrument Hazardous Area Certification Class I, Division 2 2

GE Measurement & Control

Page 3: vbSeries® Portable Instrument Hazardous Area Portable Instrument Hazardous Area Certification Class I, Division 2 5 GE Measurement & Control vbX /S COUT INSTALLATION CONTROL DRAWING

vbSeries® Portable Instrument Hazardous Area Certification Class I, Division 2 3

GE Measurement & Control

Page 4: vbSeries® Portable Instrument Hazardous Area Portable Instrument Hazardous Area Certification Class I, Division 2 5 GE Measurement & Control vbX /S COUT INSTALLATION CONTROL DRAWING

vbSeries® Portable Instrument Hazardous Area Certification Class I, Division 2 4

GE Measurement & Control

Page 5: vbSeries® Portable Instrument Hazardous Area Portable Instrument Hazardous Area Certification Class I, Division 2 5 GE Measurement & Control vbX /S COUT INSTALLATION CONTROL DRAWING

vbSeries® Portable Instrument Hazardous Area Certification Class I, Division 2 5

GE Measurement & Control

vbX/SCOUT INSTALLATION CONTROL DRAWING

FOR HAZARDOUS LOCATIONS

Drawing Number: CIL2100VBX Entire drawing is for Hazardous Location (Class I, Division 2, Groups A, B, C, D)

Warnings: 1. Recharge or change battery only in a non-hazardous area. 2. Use USB or Ethernet only in a non-hazardous area.

3. Substitution of components may impair suitability for Class I, Division 2.

Notes: 1. No revisions of this drawing are permitted without CSA approval. 2. If the sensor model is not specified, any sensor certified for the location may be used provided that the senor entity parameters meet the

above conditions and the final installation meets also the individual installation requirements for each specified sensor. 3. CSA Master Contract: 150368

PO Box 9297, Christchurch, New Zealand. Tel: (64-3) 9430-700 Fax: (64-3) 9430-727

Title:

vbX/SCOUT INSTALLATION CONTROL DRAWING

FOR HAZARDOUS LOCATIONS

Drawing No:

CIL2100VBX

Date:

3 Jun 2013

Approved by:

NL (3 Jun, rev5) Revision:

5

Ethernet USB Charger

CH1-4 TACH

Battery pack BATT0206 7.4 V 4500 mA Hr Li-Ion Do not recharge or change in hazardous locations

vbX/SCOUT

Instrument

Not to be used in hazardous locations

Any loop-powered sensor (IEPE/ICP®)

with certified input entity parameters: Vmax: ≥ 26.4 V

lmax: ≥ 47 mA Pmax: ≥ 310 mW

Li: ≤ 28 mH - Lcable Ci: ≤ 360 nF – Ccable

Any externally-powered sensor with certified output entity parameters:

Voc: ≤ 30 V Isc: ≤ 25 mA

La: ≥ Lcable Ca: ≥ 2 nF + CCable

Commtest/Bently sensor AM1-100-T2 or

A32-100-S4 or A33-100-TM, with cable: Length: ≤ 10 m Capacitance: ≤ 200 pF/m

Inductance: ≤ 1 µH/m

Any sensor powered by the vbX or Scout with certified input entity parameters:

Vmax: ≥ 5.2 V lmax: ≥ 53.3 mA

Pmax: ≥ 277 mW Li: ≤ 22.8 mH - LCable Ci: ≤ 830 µF - CCable Any externally-powered sensor

with certified output entity parameters: Voc: ≤ 30 V

lsc: ≤ 100 mA La: ≥ Lcable

Ca: ≥ 3 nF + CCable

Page 6: vbSeries® Portable Instrument Hazardous Area Portable Instrument Hazardous Area Certification Class I, Division 2 5 GE Measurement & Control vbX /S COUT INSTALLATION CONTROL DRAWING

vbSeries® Portable Instrument IP65 Rating

The following instruments have been independently verified as complying with the

relevant provisions of AS/NZS 60529:2004 for an IP65 rating:

• vb5™ Data Collector

• vb6™ Data Collector

• vb7™ Data Collector and Analyzer

• vb8™ Data Collector and Analyzer

• vbBalancer™ Imbalance Correction Instrument

• vbBalancer+™ Imbalance Correction Instrument

See attached document for details.

Notes 1. The communications area cover seal must be securely fitted to maintain the

instrument’s IP65 rating.

2. Only instruments with serial numbers 40500 and higher are IP65 certified.

3. AS/NZS 60529:2004 is reproduced from IEC 60529:2001 with minor modifications

and editorial corrections. Similarly, ANSI 60529:2004 is the US adoption of IEC

60529:2001.

Page 7: vbSeries® Portable Instrument Hazardous Area Portable Instrument Hazardous Area Certification Class I, Division 2 5 GE Measurement & Control vbX /S COUT INSTALLATION CONTROL DRAWING

APPROVED BY: TESTED BY:

Christopher Bennetts Christopher Olds IANZ Signatory Engineer

10 VANADIUM PLACE, PO BOX 9194, ADDINGTON 8243, CHRISTCHURCH, NEW ZEALAND Page 1 of 11

TEL: +64 3 339 1670 FAX: +64 3 339 1671 EMAIL: [email protected] INTERNET: www.nz.ul.com All tests reported herein have been performed in accordance with the laboratory’s IANZ approved scope of accreditation.

363 International Accreditation New Zealand (IANZ) has a Mutual Recognition Arrangement (MRA) with the National Association of Testing Authorities (NATA), Australia, such that both organisations recognise accreditations by IANZ and NATA as being equivalent. Users of test reports are recommended to accept test reports in

the name of either accrediting body.

Laboratory Test Report Report No.: 13594-00 This report may not be reproduced except in full. Date Issued: 17 January 2007 It contains no corrections or erasures.

SUBJECT:

Commtest Instruments vb7

REQUESTED BY: Commtest Instruments Ltd 28b Moorhouse Avenue Christchurch NEW ZEALAND

INSTRUCTIONS: Test for compliance with AS 60529:2004 “Degrees of protection provided by enclosures (IP code)” for an IP65 Rating.

CONTENTS: General Test Specification Date of test Description Results: AS 60529:2004

SUMMARY: All test results in this report in relation to the Commtest Instruments vb7 confirmed that the specimen complied with the relevant provisions of AS 60529:2004 for an IP65 Rating.

Page 8: vbSeries® Portable Instrument Hazardous Area Portable Instrument Hazardous Area Certification Class I, Division 2 5 GE Measurement & Control vbX /S COUT INSTALLATION CONTROL DRAWING
Page 9: vbSeries® Portable Instrument Hazardous Area Portable Instrument Hazardous Area Certification Class I, Division 2 5 GE Measurement & Control vbX /S COUT INSTALLATION CONTROL DRAWING

vbSeries® Portable Instrument Drop Test

Test Parameters

Test performed according to MIL-STD-810F-516.5-IV, 4.5.5 Procedure IV - Transit

Drop & Table 516.5-VI

Drop Height: 1.22m (nominal 1.22 meters/48 inches)

Instrument S/N: 40388

Tester Name: AJS

Date: 27 November 2007

Instrument Configuration

o Memory Module and Processing Unit secured with nylon standoffs

o LCD protected with Inseal gasket

o Keypad cables secured with Inseal foam pads

o Battery cover is glass-filled nylon

o Analog: Test signal = 100Hz 1Vrms 12Vdc

o Before testing: CH1 = CH2 = 141.9 adB US

Results

Strike Orientation Powers Up After? Comments/Observations

Face: Front √

Face: Back √

Face: Left √

Face: Right √

Face: Top √

Face: Bottom √

Edge: Front/Top √

Edge: Front/Right √

Edge: Front/Left √

Edge: Front/Bottom √

Edge: Back/Top √

Edge: Back/Right √

Edge: Back/Left √

Edge: Back/Bottom √

Page 10: vbSeries® Portable Instrument Hazardous Area Portable Instrument Hazardous Area Certification Class I, Division 2 5 GE Measurement & Control vbX /S COUT INSTALLATION CONTROL DRAWING

Edge: Right/Top √

Edge: Right/Bottom √

Edge: Left/Top √

Edge: Left/Bottom √

Corner: Front/Top/Right √

Corner: Front/Btm/Right √

Corner: Front/Bottom/Left √

Corner: Front/Top/Left √

Corner: Back/Top/Right √

Corner: Back/Btm/Right √

Corner: Back/Bottom/Left √

Corner: Back/Top/Left √

Analog Cal-Check OK? √

Inspection After Test Completion

o Memory Module, Processing Unit, Speaker, Keypad Cables all remained secure

o LCD intact and operational. Battery Cover intact. CH2 BNC slightly dented but

connector cable will fit and still operational

o Analog response after testing: CH1 = CH2 = 141.9 adB US

Conclusion

PASS

Signed:

November 27th 2007


Recommended