3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional Director Medical Physics Texas Oncology March 8, 2015 Disclaimer No financial relationships to disclose with any of the companies discussed or presented in this talk
Transcript
Page 1: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

1

Vendor Tools and Uses for TG-142

Bret H. Heintz, PhD, DABR

Regional Director Medical Physics

Texas Oncology

March 8, 2015

Disclaimer

No financial relationships to disclose with any of the companies discussed or presented in this talk

Page 2: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

2

Do you currently use a commercial TG-142 software program?

20%

20%

20%

20%

20% 1. Yes

2. No

10

Do you have baselines setup in your TG-142 program?

20%

20%

20%

20%

20% 1. Yes

2. No

10

Page 3: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

3

Do you know where to find the baseline values for your TG-142 program?

20%

20%

20%

20%

20% 1. Yes

2. No

10

Do you know who you could reach out to for help setting up your TG-142 program?

20%

20%

20%

20%

20% 1. Yes

2. No

10

Page 4: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

4

Goal of a QA program

To assure that the machine characteristics do not deviate significantly from their baseline values acquired at the time of acceptance and commissioning

Managing QA Results

• In-house• Paper

• Electronic

• Commercial

Page 5: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

5

Quote of the Day!

As of now, all Physics data (monthly, annuals, post-repair reports, post software upgrade reports, instrument calibration reports etc) are currently paper-based.

TG-142 Project

• Categories

• Business Plan• Pros and Cons

• Outline of software requirements

• RFP

• Evaluation

• Implementation

Page 6: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

6

TG-142 Project - Categories

• Major• Mechanical

• Safety

• Imaging

• Dosimetry

• Minor• Respiratory Gating

PROS and CONS of commercial TG-142 software

• Pros• Setup Instructions within software

• Implement baseline values

• Trending

• Frequency of Tests• Daily

• Monthly

• Annually

• Standardization

• Centralized location for all data

• Cons• Upfront cost

• Maintenance

• Conformity

• Time

Page 7: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

7

TG-142 Project – Software Requirements

• Software architecture

• Networking

• Centralization

• Connectivity

• Upgrades

• Customer support

• Functionality

• Features

TG-142 Project – Software Requirements

• Ease of Use

• Trending

• Reports

• Multi-functional

• Analysis tools

• Adaptable

Page 8: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

8

TG-142 Project

• RFP• Software package #1

• Imaging

• Software package #2• Mechanical

• Dosimetry

• Safety

• Evaluation• Team assembled to evaluate each product

• Evaluate each module within software package

• Provide feed back to vendor on pros and cons of product

• Implementation

Vendors

Page 9: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

9

Evaluation of TG-142 Imaging software

• User Interface

• Functionality

• Features

• Modules• CBCT

• MV Imaging

• kV Imaging

• Star Shot

• Winston Lutz

• Light Field/Radiation Field

• MLC

Evaluation of TG-142 Imaging software

• Functionality• Load Images

• Analysis

• Display results

• Generate Report

• Trending

Page 10: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

10

Features - Software

• Imaging Toolbar• Window/Level

• Ruler

• Angle

• Rotation

• ROI

• Image Type• TIFF

• JPG

• DICOM

•Film Scanner• Color

• Gray Scale

•Scale

•Line Profile

•Coordinate System

•Graticule

•Approve

Features - Report

• Site • Logo

• Name

• Machine• SN

• Image• Acquisition Date

• Evaluation Date

• Baseline

• Tolerances

•Evaluator

•Approver

•Approval Date

•Customized•Full

•Condensed

Page 11: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

11

User Interface

User Interface

Page 12: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

12

User Interface

Imaging Tests – CBCT

Page 13: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

13

Imaging Tests – CBCT

Imaging Tests – CBCT Report

Page 14: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

14

Imaging Tests – CBCT Report

Imaging Tests – MV/kV tests

Page 15: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

15

Imaging Tests – Star shot

Imaging Tests – Star shot

Page 16: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

16

Imaging Tests – Star shot report

Imaging Tests – Winston Lutz

Page 17: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

17

Imaging Tests – Winston Lutz

Imaging Tests – Light Field / Radiation Field

Page 18: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

18

Imaging Tests – Light Field Radiation Field

Imaging Tests – Light Field / Radiation Field

Page 19: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

19

Imaging Tests - MLC

Approval

Page 20: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

20

Implementation

• Software Package #2• Categories

• Mechanical

• Safety

• Dosimetry

• Frequency• Daily

• Monthly

• Annually

• Customization/Development• Machine types

• HDR

• CT

• Error Log

• Documents

• X-ray and RAM

Daily QA

Page 21: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

21

Daily QA - Safety

Daily QA - Mechanicals

Page 22: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

22

Daily QA - Dosimetry

Tolerances and Trending

Page 23: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

23

Review and Approval

Review and Approval

Page 24: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

24

Monthly Mechanical

Monthly output

Page 25: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

25

Error Log Documentation

HDR-QA

Page 26: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

26

CT QA

Recommendations - Vendors

• Simplify

• Images

• Generic imaging toolbar for all modalities• Window\Level

• Ruler

• Angle

• ROI

• Rotation

• Line Profile

• Uniform Graphical Interface (GUI)

• Approval

• Reports

Page 27: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

27

Recommendations - Customers

• Develop • How will the site use the software

• Must have features

• Implementation plan

• Baseline data• Acquire or restore

• Tolerances• Develop and implement

• Validate the algorithms before implementing module clinically

• Support• Reach out to vendors

• Reach out to fellow AAPM members

Summary

•RFP has been a very challenging and time consuming project for several physicists

•TG-142 software programs•Daily, Monthly, Annually

•Dosimetry, Mechanical, Safety, and Imaging

•PROS and CONS

•Functionality

•Features

•Product Enhancements

Page 28: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

28

Recognition

•Texas Oncology•Eric Stauffer, MS

•Marc Kleiman, MS

•Ed Scarbrough, MS

•Manish Goyal, MS

•Colin Delaney, MS

•Kyle Antes, MS

•Several other team members

•Vendors

What are the four main categories tested in TG-142?

20%

20%

20%

20%

20% 1. Dosimetry, Mechanical, Safety, MLC

2. Dosimetry, Mechanical, Safety, Training

3. Mechanical, Safety, Training, CBCT

4. Safety, Imaging, Training, Driving

5. Dosimetry, Mechanical, Safety, Imaging

10

Page 29: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

29

Answer

Dosimetry, Mechanical, Safety, Imaging• Quality Assurance or medical accelerators: Task Group Report – 142, Medical Physics 36 (9), September 2009

What are some PROS of a TG-142 software program?

20%

20%

20%

20%

20% 1. Trending, Standardization

2. Dosimetry

3. Mechanical

4. Training

5. Gating

10

Page 30: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

30

Answer

Trending and Standardization• Quality Assurance or medical accelerators: Task Group Report – 142, Medical Physics 36 (9), September 2009

The software program should be tested for ______ and _______ prior to clinical use?

20%

20%

20%

20%

20% 1. Accuracy

2. Consistency

3. Accuracy and Consistency

4. Safety and Accuracy

5. Dosimetry and Mechanical

10

Page 31: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

31

Answer 3

Accuracy and Consistency

• Quality Assurance or medical accelerators: Task Group Report –142, Medical Physics 36 (9), September 2009

The baseline data should come from which of the following?

20%

20%

20%

20%

20% 1. Customer Acceptance Testing

2. Similar machine

3. After annual evaluation

4. TG-142 software vendor

5. Friend

10

Page 32: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

32

Answer 4

Customer Acceptance Testing

• AAPM code of practice for radiotherapy accelerators: Report of AAPM Radiation Therapy Task Group No. 45

The Pass/Fail criteria should be based on:

20%

20%

20%

20%

20% 1. TG-142 software vendor

2. TG-43

3. Baseline data collected from specific machine over given period of time

4. Similar machine

5. Survey monkey results

10

Page 33: Vendor Tools and Uses for TG-142amos3.aapm.org/abstracts/pdf/97-25839-352470-109885-19623690… · 3/17/2015 1 Vendor Tools and Uses for TG-142 Bret H. Heintz, PhD, DABR Regional

3/17/2015

33

Answer 5

Baseline data collected from specific machine over given period of time

• Quality Assurance or medical accelerators: Task Group Report –142, Medical Physics 36 (9), September 2009


Recommended