With gold standard disinfection in 3 minutes –
Thermo Scientific Microm HM550 cryostat
Safety Ergonomics Quality Precision
Part of Thermo Fisher Scientific
Avoid the risk of occupational infections – with the new gold standard cold disinfection in cryostats
Every year, tens of thousands of health-care workers suffer from occupational infections, such as HBV, HCV and HIV. Their risk of catching these infections is significantly higher than in the average population. Working with infectious material is inherent to work in pathology laboratories, which is why it is our goal to minimize any health hazards. With enhanced safety and efficiency in mind, the new patented cold disinfection solu-tion is being introduced as standard with the Thermo Scientific Microm HM550. Cold disinfection from Thermo Scientific offers customers instant disinfection, significantly reducing risks to the occup-ational health of laboratory staff.
Outstanding advantages• Ready for use in just three minutes
– no workflow interruption• Immediately available for use after
disinfection cycle• Utilizes Sanosil®, a powerful non-
hazardous disinfectant – complete-ly safe for lab technicians
The keys to a new gold standard• Instant disinfection of the entire
cold cryo-chamber – 100 times faster than existing methods
• Fumigation of disinfection by ultra-sound – instrument is immediately available for use, resulting in signif-icantly higher throughput
• Micro-droplets uniformly moisten the entire cryo-chamber
Sanosil® – no need to be toxic to be effective• A disinfectant and virucide that
harnesses the natural, antimicrobial powers of Hydrogen Peroxide and Silver (Ag+)
• The combination of these two elements provides pure, 99.99% disinfecting power
• Contains NO chlorine, aldehydes, phenolics, ammonium compounds, sodium hydroxide or any other harmful toxins or carcinogens
• Clear and odorless
Perfect results quickly and safely – whilst meeting the highest standards for hygiene.Ready for safe operation in just 3 minutes. Create and maintain a safe and germ-free working environment. Thermo Scientific Microm HM550 is now equipped with the new cold disinfection feature. Hygiene and cleanliness of your cryo-chamber is now assured – with no compromise. Enjoy a fully flexible system created with safety, precision and flexibility in mind.
Snap freezing via active deep freezing device• Integrated Peltier: fast freezing of the tissue at a temperature
down to -60°C• Offers four positions for chucks with a diameter up to 30 mm• The 9 position fast freezing rail provides temperature control
down to -40°C
Perfect specimen orientation• “0-positioning“ device – automatically locks the specimen
orientation into place• Fast automatic approach of the tissue towards the knife edge (ACA)
Ergonomics, safety and design• Fully adjustable electronic handwheel brake in any
specimen position• Designed to be operated while sitting or standing• Keyboard and display intuitively guide through all operations• Reflex-free chamber illumination• Large storage surfaces• Drain system with containers for waste fluids
Option: Integrated Vacutome• Stretching sections and disposal of section debris via a suction device• Suction power can be selected by the user• Pre-cooled coarse filter of the dual filter system collects section
debris and purifies the air
Option: Specimen cooling with “quick release“ mechanism• Specimen cooling is completely integrated into the microtome• Chamber volume and storage shelves are completely maintained• “Quick release” mechanism for clamping chucks allows for fast
changing of specimen
Option: Motorized cutting drive• Three different cutting modes• Double click philosophy provides protection against unintended
motion
NEWOvercoming obstacles of conventional disinfectionNo interruption of your workflow• Current disinfection after warming up the cryo-chamber takes
up to 3 hours – with cold disinfection you are ready in just 3 minutes
Complete disinfection of the entire cryo-chamber • Not all current disinfections are fully effective, for example, with
UV light only disinfects the areas reached by light – cold disinfection reaches the entire cryo-chamber
No health risks caused by the disinfectant reagent• Current disinfection, for example, by vaporization of formalin or
others, is potentially toxic – cold disinfection with Sanosil is non-toxic with proven effectivness
See the benefits of the new cold disinfection system at: www.cold-disinfection.com
Uncompromised safety – with a vast variety of additional application options
Thermo Scientific Microm HM550 cryostat with cold disinfection specifications
Dimensions (W, D, H) 640 (w/o handwheel) x 760 x 1200 (w/o stage) mm
Weight 125 – 155 kg (depending on model)
Certificates CE; c-CSA-us
Easy-to-read displays show important data such as set/actual temperature, section sum, remaining travel, number of sections, actual time, trim and fine sectioning thickness, retraction function
Automatic approach system (ACA) for quick, precise and safe movement of the specimen towards the knife and/or blade
Electronical handwheel brake – can be locked in any position
Continuous temperature control of the cooling chamber from 0°C to -35°C
Fast freezing station for the simultaneous initial freezing of 12 specimens (30 mm chucks)
Automatic defrosting function for immediate on-demand defrosting
Heated sliding window
Energy saving modes: standby and sleep-mode
High performance rotary microtome
Section thickness range from 1 μm – 100 μm
FINE section thickness range from 1 μm – 10 μm in 1 μm steps
from 10 μm – 20 μm in 2 μm steps
from 20 μm – 100 μm in 5 μm steps
TRIMMING section thickness range from 5 μm – 500 μm
from 10 μm – 100 μm in 10 μm steps
from 100 μm – 200 μm in 20 μm steps
from 200 μm – 500 μm in 50 μm steps
Max. specimen size 70 x 55 mm
Total horizontal feed 28000 μm
Vertical cutting stroke 60 mm
Sample orientation 8° along the X-Y-axis, rotatable 360°
Sample retraction during return stroke 40 μm, can be deactivated
Intuitive 0-positioning of the object orientation
Accelerated object holder return movement for fast operation
Cutting speed 0 – 250 mm/s, can be continuously adjusted (motor option)
Cutting window selectable (motor option)
Continuous specimen temperature control from +10°C to -50°C (object cooling option)
Coarse feed and reset three different sped levels (motor option)
Three operating modes interval, single and continuous stroke (motor option)
Motorized sectioning controlled via foot pedal or button
Motor brake and emergency stop switch for safe operation (motor option)
Integrated Vacutome system for stretching cryo sections as well as debris removal (vacutome option)
Integrated fast freezing device with separate cooling element (Peltier) down to max. -60°C for four specimen (30 mm chucks) (option)
Power supply 100...240 V 1.6 A +/-10% 50...60 Hz
4481 Campus DriveKalamazoo, MI 49008USA+1 (800) 522-7270
www.thermo.com/pathology
© 2010 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Anatomical Pathology
ROW
MAM
I/100
1/00
43GB
Robert-Bosch-Strasse 4969190 WalldorfGERMANY+ 49 (0) 6227-8360