12 The Open Cardiovascular Medicine Journal, 2008, 2, 12-19 1874-1924/08 2008 Bentham Science Publishers Ltd. Quantitative Doppler-Echocardiographic Determination of Regurgitant Volume in Patients with Aortic Insufficiency Paul Schoenhagen* ,a , Ludwig Drude b , Hermann H. Klein c and Mario J. Garcia a a From the The Cleveland Clinic Foundation, Cleveland, Ohio, USA; b Kardiologische Gemeinschaftspraxis Marburg, Germany; c Klinikum Idar-Oberstein GmbH, Germany Abstract: Background: The severity of aortic regurgitation (AR) can be determined by invasive or echocardiographic methods. We systematically compared quantitative invasive and echocardiographic data with semiquantitative invasive grades in a prospective series of patients. Methods: Using Doppler-echocardiography we determined the cardiac output over the aortic, pulmonary and mitral valve in 27 patients (20 with, 7 without AR). Aortic regurgitant volume was calculated as the difference between the cardiac output over aortic and pulmonary valve/ mitral valve. During angiography the severity of AR was assessed semiquantita- tively by aortography and the regurgitant volume was calculated invasively as the difference between the left- and right ventricular cardiac output. Results: The echocardiographically and invasively determined regurgitant blood volume correlated closely (R 0.8). The regurgitant volume increased with higher angiographic grade but there was significant overlap between adjoining qualita- tive grades. Conclusion: In patients with AR, quantitative echocardiographic and angiographic measurements of the regurgitant vol- ume correlate closely. Keywords: Aortic Insufficiency, Doppler-Echocardiography, Cardiac Catheterization, Regurgitant Volume. INTRODUCTION The natural history of aortic regurgitation (AR) is charac- terized by an asymptomatic period of left ventricular com- pensation followed by decompensation with eventually irre- versible left ventricular dysfunction. The management of patients with AR and in particular the timely indication of valve surgery is therefore dependent on a reliable estimation of the severity and progression of regurgitation and LV func- tion [1]. The severity of AR has traditionally been estimated during cardiac catheterization by semiquantitative grading [2-4]. Quantitative echocardiographic methods have been developed to determine the hemodynamically active regurgi- tant volume noninvasively. The comparison between quanti- tative echocardiographic methods and angiographic grades has shown only modest correlation, which has been attrib- uted at large to inaccuracy of the echocardiographic meth- ods. We hypothesized that the discrepancy between echocar- diography and angiography is due to the comparison of quantitative with qualitative data. We therefore systemati- cally compared quantitative echocardiographic, quantitative invasive and semiquantitative invasive measurements in a prospective series of consecutive patients. METHODS Patients Between July ‘90 and May ‘93 we examined 136 con- secutive patients who were referred to the color imaging laboratory of the Philipps-University-Marburg for further *Address correspondence to this author at the Cleveland Clinic Foundation, Hb-6, 9500 Euclid Avenue, Cleveland, Ohio 44195, USA; Tel: (216) 444- 7579; Fax: (216) 445-1492; E-mail [email protected] evaluation of suspected valvular disease. All patients had undergone 2D and conventional Doppler studies before re- ferral. Informed consent was obtained according to institu- tional guidelines. The study was approved by the Philipps- University Marburg. Fourty-four patients were excluded from the final analysis for the following reasons: no reliable visualization (n=2), pulmonary valvular disease or combined aortic valve disease (n=6), mitral valve disease (n=24), atrial septal defect (n=9), ventricular septal defect (n=3). Of the remaining 92 patients, only those 27 patients that were ex- amined by cardiac catheterization and echocardiography within 24 hours were included in this study. Twenty of these patients had AR and 7 did not. The patients with AR had no concomitant, significant aortic stenosis (maximal systolic gradient of less than 30mmHg). The 7 patients in the control group had the following diseases: dilated cardiomyopathy (n=2), CAD (n=2), HTN (n=2), mitral valve prolapse with- out regurgitation (n=1). Cardiac catheterization: General Procedure Left Heart Catheterization Access was established using the Seldinger technique from a femoral artery. After coronary angiography a 7F pig- tail catheter was placed in the mid left ventricle. A left ven- triculogram was recorded in the 30 degree RAO projection over several cardiac cycles. The catheter was then withdrawn into the ascending aorta for the aortography in a 30 degree RAO projection. For both ventriculography and aortography, 40 ml of contrast material were injected with a flow of 14 ml/sec.
Transcript

12 The Open Cardiovascular Medicine Journal, 2008, 2, 12-19

1874-1924/08 2008 Bentham Science Publishers Ltd.

Quantitative Doppler-Echocardiographic Determination of Regurgitant Volume in Patients with Aortic Insufficiency

Paul Schoenhagen*,a, Ludwig Drude

b, Hermann H. Klein

c and Mario J. Garcia

a

aFrom the The Cleveland Clinic Foundation, Cleveland, Ohio, USA;

bKardiologische Gemeinschaftspraxis Marburg,

Germany; cKlinikum Idar-Oberstein GmbH, Germany

Abstract: Background: The severity of aortic regurgitation (AR) can be determined by invasive or echocardiographic

methods. We systematically compared quantitative invasive and echocardiographic data with semiquantitative invasive

grades in a prospective series of patients.

Methods: Using Doppler-echocardiography we determined the cardiac output over the aortic, pulmonary and mitral valve

in 27 patients (20 with, 7 without AR). Aortic regurgitant volume was calculated as the difference between the cardiac

output over aortic and pulmonary valve/ mitral valve. During angiography the severity of AR was assessed semiquantita-

tively by aortography and the regurgitant volume was calculated invasively as the difference between the left- and right

ventricular cardiac output.

Results: The echocardiographically and invasively determined regurgitant blood volume correlated closely (R 0.8). The

regurgitant volume increased with higher angiographic grade but there was significant overlap between adjoining qualita-

tive grades.

Conclusion: In patients with AR, quantitative echocardiographic and angiographic measurements of the regurgitant vol-

ume correlate closely.

Keywords: Aortic Insufficiency, Doppler-Echocardiography, Cardiac Catheterization, Regurgitant Volume.

INTRODUCTION

The natural history of aortic regurgitation (AR) is charac-terized by an asymptomatic period of left ventricular com-pensation followed by decompensation with eventually irre-versible left ventricular dysfunction. The management of patients with AR and in particular the timely indication of valve surgery is therefore dependent on a reliable estimation of the severity and progression of regurgitation and LV func-tion [1]. The severity of AR has traditionally been estimated during cardiac catheterization by semiquantitative grading [2-4]. Quantitative echocardiographic methods have been developed to determine the hemodynamically active regurgi-tant volume noninvasively. The comparison between quanti-tative echocardiographic methods and angiographic grades has shown only modest correlation, which has been attrib-uted at large to inaccuracy of the echocardiographic meth-ods. We hypothesized that the discrepancy between echocar-diography and angiography is due to the comparison of quantitative with qualitative data. We therefore systemati-cally compared quantitative echocardiographic, quantitative invasive and semiquantitative invasive measurements in a prospective series of consecutive patients.

METHODS

Patients

Between July ‘90 and May ‘93 we examined 136 con-secutive patients who were referred to the color imaging laboratory of the Philipps-University-Marburg for further

*Address correspondence to this author at the Cleveland Clinic Foundation,

Hb-6, 9500 Euclid Avenue, Cleveland, Ohio 44195, USA; Tel: (216) 444-

7579; Fax: (216) 445-1492; E-mail [email protected]

evaluation of suspected valvular disease. All patients had undergone 2D and conventional Doppler studies before re-ferral. Informed consent was obtained according to institu-tional guidelines. The study was approved by the Philipps-University Marburg. Fourty-four patients were excluded from the final analysis for the following reasons: no reliable visualization (n=2), pulmonary valvular disease or combined aortic valve disease (n=6), mitral valve disease (n=24), atrial septal defect (n=9), ventricular septal defect (n=3). Of the remaining 92 patients, only those 27 patients that were ex-amined by cardiac catheterization and echocardiography within 24 hours were included in this study. Twenty of these patients had AR and 7 did not. The patients with AR had no concomitant, significant aortic stenosis (maximal systolic gradient of less than 30mmHg). The 7 patients in the control group had the following diseases: dilated cardiomyopathy (n=2), CAD (n=2), HTN (n=2), mitral valve prolapse with-out regurgitation (n=1).

Cardiac catheterization:

General Procedure

Left Heart Catheterization

Access was established using the Seldinger technique from a femoral artery. After coronary angiography a 7F pig-tail catheter was placed in the mid left ventricle. A left ven-triculogram was recorded in the 30 degree RAO projection over several cardiac cycles. The catheter was then withdrawn into the ascending aorta for the aortography in a 30 degree RAO projection. For both ventriculography and aortography, 40 ml of contrast material were injected with a flow of 14 ml/sec.

Echo Assessment of Aortic Regurgitant Volume The Open Cardiovascular Medicine Journal, 2008, Volume 2 13

Semiquantitative Assessment of Aortic Regurgitation (Aortography)

The semiquantitative grading of AR was done according to established guidelines [2]. Grade 1 was defined as backflow without contrast of the complete LV, grade 2 as backflow with faint contrast of LV, grade 3 as dense contrast of LV same as aorta and grade 4 as contrast of LV more dense than aorta.

Ventriculography

The images of the ventriculogram were analyzed off line. End-diastolic and end-systolic frames were chosen for planimetry. The pictures were projected from a projector (Tagano 35 CX) onto a calibrated screen. The heart base, the maximal length of the ventricular cavity and the left ven-tricular circumference were manually traced on the screen. The left ventricular stroke volume (LV) was calculated ac-cording to the area-length method [5,6]. The results of three consecutive cardiac cycles in sinus rhythm were averaged.

Thermodilution

The right heart catheterization was performed via a femo-ral vein with a pulmonary artery flotation catheter. The tip of the catheter was positioned in the main stem of the right pulmonary artery. Through the proximal sidehole 10 ml 0.9% NaCl at room temperature were injected. At the ther-mistor at the distal end of the catheter the temperature were recorded.

The right ventricular (RV) cardiac output was calculated semiautomatically according to the Stewart Hamilton equa-tion [7,8].

Quantitative Assessment of Aortic Regurgitation

The invasive regurgitant volume (RVinvasive) was cal-culated as the difference between the left- and right-ventricular cardiac output (LVCO/RVCO) determined by ventriculography (LVCO) and thermodilution (RVCO).

Echocardiography

General Procedure

All examinations were performed by an experienced phy-sician sonographer, who was blinded to the results of the cardiac catheterization. The examinations were performed with a Sonotron Vingmed CFM750 machine, equipped with an imaging transducer of 3.5 or 2.75 Megahertz for the M-mode and 2D images. The echo-machine was connected on-line with a Macintosh Apple III personal computer equipped with an image processing software package. For off line analysis, the images were stored on optical disk and also printed on paper and recorded on a video recorder.

The examination was performed in a left decubitus posi-

tion. Complete 2D, conventional Doppler and color Doppler

study with evaluation of all valves and ventricular function were performed according to ASE standards. The cross sec-

tional areas (CSA) of the aortic, pulmonic and mitral valve

were determined using the formula CSA = r2 as follows: The maximal diameter of the left ventricular outflow tract

(LVOT) was measured from a 2D-parasternal long axis

view. The maximal diameter of the pulmonary artery (PA) was measured from a 2D-parasternal short axis view at the

level of the pulmonic valve. The maximal diameter of the

mitral valve (MV) was measured from a 2D apical 4-

chamber view at the level between the mid and distal third of the valve leaflets (Figs. 1-3).

The flow velocity was measured by pulsed Doppler. It was measured in a parasternal short axis view for the pulmo-

nary valve with the sample volume directly distal of the

valve. For the aortic valve it was measured from an apical 5-chamber view with the sample volume in the LVOT . For the

mitral valve flow velocity was measured from a apical 4-

chamber view with the sample volume at the level of the mitral valve, respectively (Fig. 1-3).

Quantitative Image Analysis

The Doppler flow signals of the aortic, mitral, and pul-monary valves were manually traced on the screen of the

Echo machine and the velocity time integral (VTI) was cal-

culated semiautomatically by planimetry. For each valve, 3 representative cycles that followed a sinus beat were ana-

lyzed, and the results averaged. The stroke volume was cal-

culated according to the formula [9-13].

SV= VTI x A

(SV= stroke volume, VTI= velocity time integral, A= valve area)

The cardiac output (CO) was calculated by multiplying

stroke volume (SV) and heart rate (HR):

CO= SV x HR

Calculation of Regurgitant Volume (RVecho1, RVecho2)

The regurgitant volume (RVecho1) was calculated as the

difference between the cardiac output of the aortic and pul-

monary valve [14].

RVecho1= CO LVOT – CO PV

RVecho2 was calculated as the difference between the car-diac output of the aortic and mitral valve [15].

RVecho2 = CO LVOT – CO MV

Statistical Analysis

For all measurements the mean, standard deviation and ranges are described.

Comparisons of the results were made by paired and un-paired t-test for paired and unpaired data, respectively. Sim-

ple linear regression analysis was employed to test the rela-

tion between two continuos parameters. Spearman rank cor-relation was used for the comparison of discrete angi-

ographic grades and continuos parameters.

RESULTS

Patient Population

Mean age was 49.4±12.6 (23-69) years. Eighteen

(66.7%) patients were male. Mean EF was 60.5±17.8%. LVEDd was 5.9±1.1 cm (4.1-7.7), LVESd was 4.0±1.01 cm

(2.4-6.4). All patients were in normal sinus rhythm at the

time of examination. The mean heart rate during catheteriza-tion was 73.9±14.4 bpm and during echocardiography

70.3±20.2 bpm.

14 The Open Cardiovascular Medicine Journal, 2008, Volume 2 Schoenhagen et al.

Fig. (1). Example of the measurement of the flow area and the velocity time integral over the aortic valve (LVOT). Panel A: The area of the

LVOT was calculated from a diameter in the parasternal long axis view. Panel B: The VTI in the LVOT was measured from the apical 4-

chamber view with the sample volume in the LVOT.

Fig. (2). Example of the measurement of the flow area and the velocity time integral over the pulmonary valve. Panel A: The area of the pul-

monary artery was calculated from a diameter measured at the level of the valve in the parasternal short axis. Panel B: The VTI in the pulmo-

nary artery was measured from a parasternal short axis view with the sample volume at the level of the valve.

Fig. (3). Example of the measurement of the flow area and the velocity time integral over the mitral valve. Panel A: The area of the mitral

valve was measured from a diameter measured at the level between the mid and distal third of the valve leaflets in the apical 4-chamber view.

Panel B: The VTI in the mitral valve was measured from the apical 4-chamber view.

Echo Assessment of Aortic Regurgitant Volume The Open Cardiovascular Medicine Journal, 2008, Volume 2 15

Echo

Cardiac Output Over the Aortic, Pulmonary and Mitral

Valve

In the 20 patients with AR the cardiac output over the

aortic, pulmonary, and mitral valve were: 9.4±3.2 l/min (5.5-

17.0), 4.72±1.1 l/min (2.7-7.3), and 5.1±1.1 l/min (3.3-7.4), respectively.

In the 7 patients without AR the cardiac output over the

aortic, pulmonary, and mitral valve were: 5.3±1.2 l/min (3.7-

7.1), 5.1±1.1 l/min (3.5-7.3), and 5.1±1.4 l/min (3.3-7.8), respectively.

Regurgitant Volume

The calculated regurgitant volume in patients with AR was 4.7±2.7 (1.4-10.8) for the comparison of the aortic and

pulmonary valve (RVecho1) and 4.3±2.6 (0.01-10.3) for the

comparison of the aortic and mitral valve (RVecho2).

The calculated regurgitant volume in patients without AR

was 0.2±0.8 (-0.8-1.6) for the comparison of the aortic and pulmonary valve (RVecho1) and 0.3±0.9 (-0.7-1.7) for the

comparison of the aortic and mitral valve (RVecho2).

INVASIVE

Semiquantitative Grading of Aortic Regurgitation

Semiquantitative grading of aortic regurgitation by aor-

tography showed grade 0 in 7 patients (26%), grade 1 in 1 patients (4%), grade 2 in 8 patients (30%), grade 3 in 9 pa-

tients (33%) and grade 4 in 2 patients (7%).

Cardiac Output as Determined by Thermodilution and

Ventriculography.

In the 20 patients with AR the right-and left-ventricular

cardiac output were 5.9±1.3 l/min (3.6-9.0) and 10.7±3.8 l/min (6.0-19.8), respectively.

In the 7 patients without AR the right-and left-ventricular cardiac output were 5.0±0.9 l/min (3.5-6.6) and 5.1±1.7 l/min (2.7-7.3), respectively.

Regurgitant Volume

In the 20 patients with AR the calculated regurgitant vol-ume was 4.7±3.3 l/min (0.81-12.8).

In the 7 patients without AR the calculated regurgitant volume was 0.06±1.1 l/min (-1.9-1.3).

Comparison between Echocardiographic and Invasive Method

Comparison of RVecho1 to RVecho2 (Fig. 4)

The regurgitant volume calculated from aortic and pul-monary valve (RVecho1) was compared to the regurgitant

Fig. (4). Comparison of the echocardiographic regurgitant volume

measured by comparison of the aortic and pulmonary valve (RVe-

cho1) and aortic and mitral valve (RVecho2).

Fig. (5). Comparison of the echocardiographic regurgitant volume (RVecho1 and RVecho2) with the invasive measured RV.

16 The Open Cardiovascular Medicine Journal, 2008, Volume 2 Schoenhagen et al.

volume calculated from aortic and mitral valve (RVecho2). For the entire group, the correlation was 0.98; p<0.0001; y= 0.93x-0.011; SE= 0.038.

Comparison of RVinvasive to RVecho (Fig. 5)

The echocardiographic regurgitant volume RVecho1 (aortic/pulmonary valves) and RVecho2 (aortic/mitral valves) were compared with the angiographic regurgitant volume (RVinvasive). For RVecho1 the correlation was 0.8; p<0.0001; y=0.69x+1.07; SE= 0.10.

For RVecho2 the correlation was 0.78; p<0.0001. y=0.64x+0.99; SE=0.10.

Comparison of RV to Angiographic Grades (Fig. 6 and 7)

Table 1 shows the comparison of the RV calculated by echocardiographic and invasive method and the angiographic grades. There was a significant correlation between higher grades of severity and increased RVecho 1 (p=0.004, rho=0.70), RVecho2 (p=0.005, rho=0.68), and RVinvasive (p=0.002, rho=0.73; Spearman Rank Correlation).

Fig. (7). Comparison of the invasive regurgitant volume (RVinva-

sive) with the angiographic grades.

However, because of a wide range of the regurgitant vol-ume in each grade, no significant differences of the echocar-diographic RV was found between angiographic grades 2 and 3. (RVecho1: 3.4+/-1.8 vs. 5.3+/-2.8 l/min; p=0.12, RVecho2: 3.03+/-1.7 vs. 4.97+/-2.6 l/min; p=0.10, RVinva-sive: 3.8+/-2.9 vs. 4.68+/-3.1 l/min; p=0.56)

DISCUSSION

In this study we systematically compared quantitative echocardiographic, quantitative invasive and semiquantita-tive invasive data determining the severity of aortic insuffi-ciency. We found a high correlation (R=0.98) between two quantitative echocardiographic methods quantifying the re-gurgitant volume over the aortic valve (RVecho1 and RVe-cho2). Although the correlation between angiographic and echocardiographic quantitative methods was good, the corre-lation between semiquantitative angiographic grades and RV was only moderate and semiquantitative grading could not differentiate grade 2 and 3 because of significant overlap between grades.

These results have important implications for the ap-proach to patients with AR. The angiographic semiquantita-tive classification has been the historical “gold standard” for estimating severity of AR and grade 3 has been used as a cutoff point above which to consider valve surgery. Our re-sults suggest that angiographic grades 2 and 3 do not reliably differentiate groups with different hemodynamically active regurgitant volume. This is consistent with results of previ-ous studies. Kitabatake examined 10 patients with isolated AR, 13 with additional valve disease and 10 control patients [14]. Using duplex Doppler echocardiography the aortic re-gurgitant fraction (RF) was calculated from systolic aortic and pulmonary flow. A fair correlation was found between RFEcho and semiquantitative grades (r=0.80, p<0.01), with significant overlap between grades. Rokey et al. examined 6 patients with isolated AR and 19 patients with isolated MR [15]. Mitral and aortic valve flow were obtained with 2-dimensional doppler and RF was calculated. Results were compared with measurements of RF obtained by combined LV angiography and thermodilution. No significant correla-

Fig. (6). Comparison of the echocardiographic regurgitant volume (RVecho1 and RVecho2) with the angiographic grades.

Echo Assessment of Aortic Regurgitant Volume The Open Cardiovascular Medicine Journal, 2008, Volume 2 17

tion was found between RF by either method and angi-ographic grades. Kurokawa examined 15 patients with AR, 17 patients with MR and 41 normal controls [15]. The authors measured cardiac output over the mitral and aortic valve respectively. RV and RF was calculated. The compari-son with angiographic grading showed significant overlap between adjacent grades.

These and other studies [17] describe significant overlap of the regurgitant volume between angiographic grades. This overlap is caused by the fact that the angiographic grades comprise a continuum of values of regurgitant volume with-out a sharp border between grades. Also other factors such as location of the catheter, size of the aorta and LV, and aortic blood pressure may influence the degree of LV opacification during aortography.

Previous studies generally report a good correlation be-tween echocardiographic, szintigraphic and invasive quanti-tative methods determining RV: Kitabatake calculated the regurgitant fraction (RF) invasively from right ventricular stroke volume, determined by thermodilution and left ven-tricular stroke volume determined by ventriculography [14]. A close correlation was found between the RF estimated with Doppler and catheter technique in patients with isolated AR (r=0.96, p<0.01). Rokey compared Doppler-echocardiography with measurements of RF obtained by combined LV angiography and thermodilution [16]. A sig-nificant correlation was observed between the two methods (r=0.91, SEE = 7%). Zhang examined 26 patients with AR and 23 normal controls by combined 2-dimensional and Doppler-echocardiography and calculated RF by comparing transmitral volume and left-ventricular stroke volume [17]. The RF was also calculated by radionuclide ventriculogra-phy. The comparison between echocardiographic and inva-sively calculated RF was good. Kurokawa calculated RF by radionuclide angiography [15]. The RF calculated by echo-cardiographic and invasive measurements correlated well (r = 0.79 p < 0.01; n = 11).

In calculating the cardiac output over individual valves by multiplying the velocity time integral (VTI) and the out-flow tract area several technical limitations have to be con-sidered: The calculation of the outflow tract area from one measured diameter assumes a circular shape during systole. Several studies have shown a nearly circular anatomy of the aortic and pulmonary valves and a constant valve area during systole [18-20]. On the other hand the mitral valve has been shown to have an oval anatomy with variable valve area dur-

ing systole secondary to the biphasic flow pattern in diastole [21,22]. Therefore the mitral valve orifice has often been calculated by oval formulas or determined by planimetry [23-25]. However, several studies have reported good results with calculation based on circular formula [26,27]. We found a high correlation between the two echocardiographic meth-ods to calculate the regurgitant volume confirming the use-fulness of a circular formula to calculate the cross-sectional area of the mitral valve.

The goal of our study was to compare noninvasive, echo-cardiographic quantification of aortic insufficiency versus quantitative and qualitative invasive methods. We did not perform a comparison of the described echocardiographic quantitative method to a comprehensive echocardiographic evaluation as performed typically in clinical practice. This typically includes flow mapping to determines the spatial extension of the regurgitant jet [28-31], the determination of the jet area at the valve level (vena contracta) [32], direc-tional flow in the ascending /descending aorta [33], pressure-half-time [34,35] and the proximal isovelocity surface area (PISA) method [36,37].

However, as quantification of flow events has more re-cently been described using three dimensional echocardi-ography [38,39] and magnetic resonance imaging [40-42], routine quantification of the regurgitant volume in patients with aortic insufficiency may become possible and the de-scribed method can serve as a reference.

Limitations

The small number of patients is a major limitation of our study. Also, the regurgitant volume is dependent on hemo-dynamic conditions. Therefore the time delay between echo-cardiographic and invasive examination may have affected the correlation between echocardiography and angiographic grades. In order to minimize that influence we included only patients who had the two examinations within 24 hours. The results of the RV calculated in patients without AR show a relative large range with both echocardiographic and inva-sive calculation. A careful comparison of qualitative results obtained with 2D Echo and quantitative echocardiographic results and clinical findings is therefore necessary.

CONCLUSION

The results of our study show a good correlation between echocardiographic and invasive quantitative calculations of the regurgitant volume in patients with aortic insufficiency.

Table 1. Comparison of Echocardiographic and Invasive determined Regurgitant Volume

RVecho1 RVecho2 RVinvasive

Grade 0 (n=7) 0.03+/-0.15 0.04+/-0.15 -0.06+/-0.3

Grade 1 (n=1) 4.97 4.62 3.2

Grade 2 (n=8) 3.4+/-1.8 3.03+/-1.7 3.8+/-2.9

Grade 3 (n=9) 5.30+/-2.8 4.97+/-2.6 4.68+/-3.1

Grade 4 (n=2) 6.76 6.22 9.5

RVecho1 = echocardographically determined regurgitant volume (comparison aortic and pulmonary valve). RVecho2 = echocardographically determined regurgitant volume (comparison aortic and mitral valve). Rvinvasive = invasively determined regurgitant volume.

18 The Open Cardiovascular Medicine Journal, 2008, Volume 2 Schoenhagen et al.

The comparison to the semiquantitative angiographic grades show that grades 2 and 3 do not reliably differentiate groups with significant different regurgitant volume. Although an-giographic grading of severity is considered the traditional “gold standard”, management decisions in patients with aor-tic regurgitation should therefore not rely on angiographic grades alone but incorporate a quantitative assessment of the hemodynamically active regurgitant volume.

ABBREVIATIONS

AR = Aortic Regurgitation

CAD = Coronary Artery Disease

CSA = Cross Sectional Area

REFERENCES

[1] Bonow RO, Lakatos E, Maron BJ, et al. Serial long-term assess-

ment of the natural history of asymptomatic patients with chronic

aortic regurgitation and normal left ventricular function. Circula-

tion 1991; 84,4: 1625-1635.

[2] Sellers RD, Levy MJ, Amplatz L, et al. Left retrograde cardioangi-

ography in acquired cardiac disease. Am J Cardiol 1964; 14,4: 437-

447.

[3] Baron MG. Angiographic evaluation of valvular insufficiency.

Circulation 1971; XLIII: 599-605.

[4] Croft CH, Lipscomb K, Mathis K, et al. Limitations of qualitative

angiographic grading in aortic and mitral regurgitation. Am J Car-

diol 1984; 53: 1593-1598.

[5] Dodge HT, Sandler H, Baxley WA, et al. Usefulness and limita-

tions of radiographic methods for determining left ventricular vol-

ume. Am J Cardiol 1966; 18: 10-23.

[6] Sandler H, Dodge HT. The use of single plane angiocardiograms

for the calculation of left ventricular volume in man. Am Heart J

1968; 75,3: 325-334.

[7] Ganz W, Swan HJC. Measurement of blood flow by thermodilu-

tion. Am J Cardiol 1972; 29: 241-246.

[8] Weisel RD, Berger RL, Hechtmann HB. Measurement of cardiac

output by thermodilution. N Engl J Med 1975; 27: 682-684.

[9] Huntsman LL, Stewart DK, Barnes SR, et al. Noninvasive doppler

determination of cardiac output in man. Clinical validation. Circu-

lation 1983; 67,3: 593-602.

[10] Ihlen H, Amlie JP, Dale J, et al. Determination of cardiac output by

doppler echocardiography. Br Heart J 1984; 51: 54-60.

[11] Loeppky JA, Hoekenga DE, Greene R,et al. Comparison of nonin-

vasive pulsed doppler and fick measurements of stroke volume in

cardiac patients. Am Heart J 1984; 107,2: 339-346.

[12] Magnin PA, Stewart JA, Myers S, et al .Combined doppler and

phased-array echocardiographic estimation of cardiac output. Cir-

culation 1981; 63,2: 388-392.

[13] Schuster AH, Nanda NC. Doppler echocardiographic measurement

of cardiac output: Comparison with a non-golden standard. Am J

Cardiol 1984; 53: 257-259.

[14] Kitabatake A, Ito H, Inoue M, et al. A new approach to noninva-

sive evaluation of aortic regurgitant fraction by two-dimensional

doppler echocardiography. Circulation 1985; 72,3: 523-529.

[15] Kurokawa S, Takahashi M, Sugiyama T, et al. Noninvasive evalua-

tion of the magnitude of aortic and mitral regurgitation by means of

doppler two-dimensional echocardiography. Am Heart J 1990; 120,

3: 638-648.

[16] Rokey R, Sterling LL, Zoghbi WA, et al. Determination of regurgi-

tant fraction in isolated mitral or aortic regurgitation by pulsed

doppler two-dimensional echocardiography. J Am Coll Cardiol

1986; 7: 1273-8.

[17] Zhang Y, Nitter-Hauge S, Ihlen H, et al. Measurement of aortic

regurgitation by doppler echocardiography. Br Heart J 1986; 55:

32-38.

[18] Francis GS, Hagan AD, Oury J, et al. Accuracy of echocardiogra-

phy for assessing aortic root diameter. Br Heart J 1975; 37: 376-

378.

[19] Loeber CP, Goldberg SJ, Marx GR, et al. How much does aortic

and pulmonary artery area vary during the cardiac cycle. Am Heart

J 1987; 113,1: 95-10041.

[20] Seed WA, Wood NB. Velocity patterns in the aorta. Cardiovascular

Research 1971; 5: 319-330.

[21] Ormiston JA, Shah PM, Tei C, et al. Size and motion of the mitral

valve annulus in man. I. A two-dimensional echocardiographic

method and findings in normal subjects. Circulation 1981; 64,1:

113-120.

[22] Rusted IE, Scheifley CH, Edwards JE. Studies of the mitral valve.

I. Anatomic features of the normal mitral valve and associated

structures. Circulation 1952; 6: 825-831.

[23] Goldberg SJ, Dickinson DF, Wilson N. Evaluation of a elliptical

area technique for calculating mitral blood flow by doppler echo-

cardiography. Br Heart J 1985; 54: 68-75.

[24] Fisher DC, Sahn DJ, Friedman MJ, et al. The mitral valve orifice

method for non-invasive two dimensional echo-doppler determina-

tions of cardiac output. Circulation 1983; 67,4: 872-877.

[25] Zhang Y, NitterHauge S, Ihlen H, et al. Doppler echocardiographic

measurement of cardiac output using the mitral orifice method. Br

Heart J 1985; 53: 130-136.

[26] Tribouilloy C, Slama M, Shen WF, et al. Determination of left

ventricular inflow by pulsed doppler echocardiography: Influence

of mitral orifice area and blood velocity measurements. European

Heart J 1991; 12: 39-43.

[27] Lewis JF, Kuo LC, Nelson JG, et al. Pulsed doppler echocardio-

graphic determination of stroke volume and cardiac output: clinical

validation of two new methods using the apical window. Circula-

tion 1984; 70,3: 425-431.

[28] Ciobanu M, Abbasi AS, Allen M, et al. Pulsed doppler echocardi-

ography in the diagnosis and estimation of severity of aortic insuf-

ficiency. Am J Cardiol 1982; 49: 339-343.

[29] Veyrat C, Ameur A, Gourtchiglouian C, et al .Calculation of pulsed

doppler left ventricular outflow tract regurgitant index for grading

the severity of aortic regurgitation. Am Heart J 1984; 108,3: 507-

515.

[30] Miyatake K, Okamoto M, Kinoshita N, et al. Clinical applications

of a new type of real time two dimensional doppler flow imaging

system. Am J Cardiol 1984; 54: 857-868.

[31] Perry GJ, Helmcke F, Nanda NC, et al. Evaluation of aortic insuf-

ficiency by doppler color flow mapping. J Am Coll Cardiol 1987;

9,4: 952-959.

[32] Veyrat C, Lessana A, Abitbol G, et al. New indexes for assessing

aortic regurgitation with two-dimensional doppler echocardio-

graphic measurement of the regurgitant aortic valvular area. Circu-

lation 1983; 68,5: 998-1005.

[33] Boughner DR. Assessment of aortic insufficiency by transcutane-

ous doppler ultrasound. Circulation 1975; 52: 874-879.

[34] Gozzelino G, Aletto C, Curti MT,et al. The effect of heart rate on

the slope and pressure half-time of the doppler regurgitant velocity

in aortic insufficiency. J Am Soc Echocardiogr 1996; 9: 516-26.

[35] Ishii M, Jones M, Shiota T, et al. What is the validity of continuous

wave doppler grading of aortic regurgitation severity? A chronic

animal model study. J Am Soc Echocardiogr 1998; 11: 332-7.

[36] Tribouilloy CM, Enriquez-Sarano M, Fett SL, et al. Application of

the proximal flow convergence method to calculate the effective

regurgitant orifice area in aortic regurgitation. J Am Coll Cardiol

1998; 32: 1032-9.

[37] Sato Y, Kawazoe K, Nasu M, et al. Clinical usefulness of the

proximal isovelocity surface area method using echocardiography

in patients with eccentric aortic regurgitation. The Journal of Heart

Valve Disease 1999; 8: 104-111.

[38] Irvine T, Li XN, Rusk R, et al . Three dimensional colour Doppler

echocardiography for the characterisation and quantification of

cardiac flow events. Heart 2000; 84(suppl): ii2-ii6.

Echo Assessment of Aortic Regurgitant Volume The Open Cardiovascular Medicine Journal, 2008, Volume 2 19

[39] Berg S, Torp H, Haugen BO, et al. Volumetric blood flow meas-

urement with the use of dynamic 3-dimensionl ultrasound color

flow imaging. J Am Soc Echocardiogr 2000; 13: 393-402.

[40] Chatzimavroudis GP, Oshinski JN, Franch RH, et al. Quantifica-

tion of the aortic regurgitant volume with magnetic resonance

phase velocity mapping: a clinical investigation of the importance

of imaging slice location. J Heart Valve Dis 1998; 7: 94-101.

[41] Dulce MC, Mostbeck GH, O'Sullivan M, et al.Severity of aortic

regurgitation: interstudy reproducibility of measurements with ve-

locity-encoded cine MR imaging. Radiology 1992; 185: 235-40.

[42] Sechtem U, Pflugfelder PW, Cassidy MM, et al. Mitral or aortic

regurgitation: quantification of regurgitant volumes with cine MR

imaging. Radiology 1988; 167: 425-30.