5975 inert MSD
Chris SandyEuropean GC-MS Product SpecialistAgilent Technologies
Presenter: Chris Sandy
Welcome to our e-Seminar:
Introduction of the new 5975 inert MSD for expanding your application capabilities
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 3Chairperson: John Vis, Audio: +44 207 162 0125
What Has Changed?
What’s the same?
• Reliability
• Quality
• Inertness
But NOT Performance – it’s better!
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 4Chairperson: John Vis, Audio: +44 207 162 0125
What’s Different?
• Functional industrial design
• More efficient vacuum system
• Extended mass range
• AutoCI
• New sensitivity specifications
• Synchronous SIM/Scan
• EI with CI ion source
• 6850 GC control
• QuickSwap MSD interface
• Review of differentiators
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 5Chairperson: John Vis, Audio: +44 207 162 0125
Quick Remove Top Cover
Easy analyzer access
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 6Chairperson: John Vis, Audio: +44 207 162 0125
Functional Front Window
• Wire connections
• Filament operation
• Column insertion
• CI Source identification
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 7Chairperson: John Vis, Audio: +44 207 162 0125
Modified MSD mainframe
for quick/easy turbo pump
maintenance
Modified Mainframe
New turbo and mechanical vacuum pumps are more efficient for Helium
and Hydrogen
• New 70 L/sec standard pump
• New 262 L/sec performance pump
• New 2.5 m3/hour mechanical pump
(67% more capacity)
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 8Chairperson: John Vis, Audio: +44 207 162 0125
G3397A – optional gauge/controller
Integrated Ion Gauge/Controller
• Greater vacuumprecision
• Excellent for trouble shooting small leaks –every system should have one!
Software displayLocal display
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 9Chairperson: John Vis, Audio: +44 207 162 0125
New Sensitivity Specifications –EI Scan
182:1 s/n100:160:1EI1 pg OFN
59755973
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 10Chairperson: John Vis, Audio: +44 207 162 0125
New Sensitivity Specifications –CI Scan
125:175:1PCI100 pg BZP
300:1 (100 fg)
500:1(1 pg)
NCI1 pg/100 fg
OFN
59755973
704:1 s/n (100 fg OFN under NCI)
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 11Chairperson: John Vis, Audio: +44 207 162 0125
Extended Mass Range
• 1050 m/z for 5975 inert MSD
• Quadrupole modified
• New electronics
• 800 m/z for all 5973 MSDs
(Software upgrade does not enable 1050 m/z)
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 12Chairperson: John Vis, Audio: +44 207 162 0125
Extended Mass Range – PBDE 209
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 13Chairperson: John Vis, Audio: +44 207 162 0125
959
957
961
955
963
953
965
960 962
958956
953
954
955
956
957
958960
961
962
963
964
965
959
953 954 955 956 957 958 959 960 961 962 963 964 965
5975 inert MSD Theoretical
Extended Mass Range –PBDE 209
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 14Chairperson: John Vis, Audio: +44 207 162 0125
AutoCI
• AutoCI – Fully automated setup/tuning
• No external manual control
• CI made as easy as EI!
The new 5975
CI interface
The old 5973 manual CI interface
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 15Chairperson: John Vis, Audio: +44 207 162 0125
Automated EI with CI Source
• March 2004: 5973 MSD
• Application note on EI with CI ion source
5989-0595EN
• Now: 5975 MSD
• Automated switching between EI and CI in a
sequence
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 16Chairperson: John Vis, Audio: +44 207 162 0125
Automated EI with CI Source
Setup steps:
1. Install CI Source
2. Tune instrument in PCI – create PCICH4.u
3. Auto create EI tune file – create ei-cisource.u
4. Tune instrument in NCI – NCICH4.u
5. Assign tune file with method and run in sequence
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 17Chairperson: John Vis, Audio: +44 207 162 0125
6850 GC Control via ChemStation
• 5973N MSD (G2570A)
• 5973 inert MSDs
• 5975 inert MSDs
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 18Chairperson: John Vis, Audio: +44 207 162 0125
Agilent MSD Unique Features
• Inert Ion source
• Hyperbolic quad
• Quad temp
• CI system
• Analyzer modularity
Synchronous SIM/Scan
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 20Chairperson: John Vis, Audio: +44 207 162 0125
Set Up Synchronous SIM/Scan
• SIM/Scan data
analysis requiresmixed mode dataview.
• SIM/Scan data issaved in the samedata file directory.
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 21Chairperson: John Vis, Audio: +44 207 162 0125
Comparison of SIM/scan Implementations
Implementation (1) – ‘coarse granularity’
Agilent Implementation– ‘fine granularity’ for whole acquisition time
Implementation (2) – ‘fine granularity’ at specific times
11.20 11.40 11.60 11.80 12.00 12.20 12.40 12.60 12.80 13.00 13.20 13.40
Scan SIM
TIME
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 22Chairperson: John Vis, Audio: +44 207 162 0125
• AUTOSIM is a standard feature of G1701DA Software
• AUTOSIM provides rapid and automatic conversion of a Scan method to SIM/Scan
• AUTOSIM creates SIM method based on Calibration table
• The operator has fine control over SIM parameter settings
• The SIM ions/groups/times are saved automatically to the method
Synchronous SIM/Scan –AutoSIM
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 23Chairperson: John Vis, Audio: +44 207 162 0125
Inject Standard
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 24Chairperson: John Vis, Audio: +44 207 162 0125
AutoSIM
Application note: 5988-4188EN
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 25Chairperson: John Vis, Audio: +44 207 162 0125
Edit SIM Method
AutoSIM automatically creates SIM groups and populates ions in each group
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 26Chairperson: John Vis, Audio: +44 207 162 0125
Synchronous SIM/Scan
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 27Chairperson: John Vis, Audio: +44 207 162 0125
1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.0011.0012.0013.00
10000
30000
50000
70000
90000
110000
130000TIC: 1 ppb 60 VOCs.D\DATA.MS
1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.0011.0012.0013.00
10000
30000
50000
70000
90000
TIC: 1 ppb 60 VOCs.D\DATASIM.MS
Scan and SIM Chromatograms from SIM/Scan Analysis of 60 VOCs at 1 ppb Each
SIM
Scan
• Velocity XPT/6890N/5975 P&T/GC/MS system• 20-m X 0.18 mm X 1.0 µmDB-VRX column• Scan sampling rate = 2; Dwell =20 ms for SIM
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 28Chairperson: John Vis, Audio: +44 207 162 0125
Synchronous SIM/Scan Analysis of VOC Target Gases at 1 ppb each
1.20 1.30 1.40 1.50 1.60 1.70 1.80 1.90 2.000
1000
2000
3000
4000
5000
6000
7000
8000
9000
10000
1.151.201.251.301.351.401.451.501.551.601.651.701.751.801.851.901.950
1000
2000
3000
4000
5000
6000
7000
8000
9000
10000
Scan
SIM
1
2 3
4 5
6
1
2 3
4 5
6
VOC Gases (1 ppb each)
1. Dichlorodifluoromethane
2. Chloromethane
3. Vinyl chloride
4. Bromomethane
5. Ethyl chloride
6. Trichlorofluoromethane
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 29Chairperson: John Vis, Audio: +44 207 162 0125
SIM/Scan and Scan-Only Sensitivity Comparison: 5-mL Purge of 1 ppb Ethyl Chloride (a Low Responding Gas)
2075Scan(SIM/Scan)
1773Scan(Scan only)
254749SIM(SIM/Scan)
Peak/Peak S/N
RMS S/N
• SIM as part of SIM/Scan run gives 10X greater sensitivity
• Still get ≥≥≥≥ 8 scans across peak using same scan rate (n = 2)
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 30Chairperson: John Vis, Audio: +44 207 162 0125
Why Our SIM/Scan is So Good
22
Performance Electronics
Nearly no fall off with increasing scan speed
21100
50
Phenanthrene
20 ~ 6250 amu/sec
23 ~ 781 amu/sec
Nearly no fall off with increasing scan speed
Standard Electronics
Fall off with increasing scan speed
21
22
100
50
Phenanthrene
23 ~ 781 amu/sec
Fall off with increasing scan speed
20 ~ 6250 amu/sec
Moving from 23 to 22
doubles scans
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 31Chairperson: John Vis, Audio: +44 207 162 0125
For a 0.25 mm column, a typical peak width (at base) is 6 seconds. Switch from scan at 2^2 mode to SIM/scan@2^1, there will still be the same 15 cycles over the peak, but gets the 4-ion SIM group data for FREE.
Scan ���� SIM/scan45 to 555 => 510 amu
Scan
45 to 555 amu, Scan only @ 2^2
2.87 scan/sec1563 amu/s
SIM
4 ions, 40 ms
1 sec 2 sec
2^06250 amu/s
3.76 cycle/s
2^13125 amu/s
2.88 cycle/s
2^21563 amu/s
1.94 cycle/s
Same width (SIM)
a cycle
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 32Chairperson: John Vis, Audio: +44 207 162 0125
Synchronous SIM/Scan –Key Features
• Performance Electronics really make the difference
• SIM Improvements – 100 groups and 60 ions per group
• Minimum dwell time is reduced by half to 5 ms
• Even more data points per SIM peak
• Synchronous SIM/Scan available on 5973 MSDs with performance electronics (req. Smart Card 3+)
• Upgrades: G1088A and G1088B AND SW upgrade
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 33Chairperson: John Vis, Audio: +44 207 162 0125
Synchronous SIM/ScanApplication examples
•Allergens analysis: Proposed 5975 method, 26 Targets
SIM/Scan mode
- SIM table built with AUTOSIM
- Quant with SIM target ion
- Confirmation by DRS with Scan data
•Pesticides analysis with microfluidics technology for flexible GC and GC/MS analysis
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 34Chairperson: John Vis, Audio: +44 207 162 0125
Allergens analysis in SIM/Scan
After shave1% v/v 1ul Split 20 : 1SCAN
SIM
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 35Chairperson: John Vis, Audio: +44 207 162 0125
Allergens Overlaid SIM/Scan
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 36Chairperson: John Vis, Audio: +44 207 162 0125
Allergens SIM/Scan analysis and deconvolution with DRS
With courtesy of Chris Sandy and J-F Garnier
Quant based on SIM target ion
Confirmation based on deconvoluted full mass spectrum
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 37Chairperson: John Vis, Audio: +44 207 162 0125
Synchronous SIM/Scan –Summary
• Great for target analysis and identifying unknowns
• SIM data for quantitation
• Scan data for confirmation of unknowns
• Our SIM/Scan delivers real performance based on superior electronics
Overview of MicrofluidicTechnologies
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 39Chairperson: John Vis, Audio: +44 207 162 0125
Connections to Microfluidic Splitter
•Deactivated, very low internal volume to prevent peak tailing orband broadening
•A single, special design metal ferrule
• More inert that graphite/vespel, does not leak at high oven temperatures
To Detector 1To Detector 2
Plate
Metal
Ferrule
Nut
Column
Channel
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 40Chairperson: John Vis, Audio: +44 207 162 0125
Microfluidic Splitter hardware installation
Deactivated microfluidicsplitter operates to 350 C.
Uses metal ferrules to eliminate leaks and
retightening
To MSD
To Det 1
Column in
Makeup from Aux EPC
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 41Chairperson: John Vis, Audio: +44 207 162 0125
AUX EPC
MSD
Column in
Blocked
Microfluidic Splitter operating modes * (1) Single Column
* Aux EPC Flow enables column change-over without venting MSD
* Aux EPC Flow enables inlet / septum maintenance without ingressof air into column.
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 42Chairperson: John Vis, Audio: +44 207 162 0125
Microfluidic Splitter operating modes (2) Splitter
AUX EPCColumn in
MSD
Detector 1
* Aux EPC Flow enables column change-over without venting MSD
* Aux EPC Flow enables inlet / septum maintenance without ingressof air into column.
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 43Chairperson: John Vis, Audio: +44 207 162 0125
Splitting to 2 Detectors
Effluent Splitter
with Makeup
6890N GC
Column
AUX EPC
3.8 psig
0.814 m X 0.18 mm id
1.1 m X 0.18 mm id
15 m X 0.25 mm id X 0.25 um HP-5MS
Configuration for 2:1 Split to MSD
mECD
Auto-sampler
5975 InertMSD
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 44Chairperson: John Vis, Audio: +44 207 162 0125
Microfluidic Splitter – ‘Chromatographically Transparent’
MSD
ECD
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 45Chairperson: John Vis, Audio: +44 207 162 0125
100 ppb VOAs, 100 ppm Gasoline in Water, Ambient Headspace
0 2 4 6 8 10 12
0 2 4 6 8 10 12
0 2 4 6 8 10 12
Scan
ECD
SIM
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 46Chairperson: John Vis, Audio: +44 207 162 0125
AUX EPCColumn 1 / Inlet 1
MSD
Column 2 / Inlet 2
Microfluidic Splitter operating modes :(3) Dual Column Inlet
* Aux EPC Flow enables column change-over without venting MSD
* Aux EPC Flow enables inlet / septum maintenance without ingressof air into column.
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 47Chairperson: John Vis, Audio: +44 207 162 0125
Dual Column Configuration*
Back : 30m x 0.25mm ID x 0.25um HP5-MS
Front : 12.4m x 0.25mm ID x 0.25um HP5-MS
25 mins
6 mins
*Requires performance turbo pump
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 48Chairperson: John Vis, Audio: +44 207 162 0125
AUX EPCColumn 1 / Inlet 1
MSD
Column 2 / Inlet 2
Microfluidic Splitter operating modes :(4) Dual Column Inlet + Olfactory Port + MSD
ODO II* Aux EPC Flow enables column change-over without venting MSD
* Aux EPC Flow enables inlet / septum maintenance without ingressof air into column.
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 49Chairperson: John Vis, Audio: +44 207 162 0125
AUX EPC3.8 psig Column inlet
23.73 psig
MSD
Detector 1
Splitter mode – acquiring data
AUX EPC25.0 psig Column inlet
1.0 psig
MSD
Backflush mode
Detector 1
Microfluidic Splitter operating modes *
(5) Splitter mode with Backflush
* Aux EPC Flow enables column
change-over without venting MSD* Aux EPC Flow enables inlet / septum
maintenance without ingress of air into column
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 50Chairperson: John Vis, Audio: +44 207 162 0125
6890N/5975 inert Pesticide Analysiswith 3-Way Splitter
3-way Effluent Splitter with Makeup
Auto-sampler
6890N GC
Column
Phosphorus FPD
30 m X 0.25 mm id X 0.25 um HP-5MS
1X method with 10:10:1 split FPD:MSD:ECD
5975InertMSD
uECDAUX EPC
3.8 psig
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 51Chairperson: John Vis, Audio: +44 207 162 0125
1-injection-4-signal System
• 5975, RTLocked, 30m x 0.25mm x 0.25um HP-5MS
• Samples: pesticide mix at 10, 1, and 0.1 ppm
• Injection: 10:1 split
• At the end of the column, eluent flow is split 3 ways: MSD:FPD:µECD = 10:10:1
10:1 split
SIM/Scan
µECD
FPD
~480 pg
~480 pg
~48 pg
Injection 10 ppm (10000 pg)
1000 pg
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 52Chairperson: John Vis, Audio: +44 207 162 0125
5975 SIM/scan + 3-way splitterUltimate Pesticide System
Scan: ~ 480 pg
SIM: ~ 480 pg
FPD: ~ 480 pg7 peaks
5.00 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00
µECD: ~ 48 pg
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 53Chairperson: John Vis, Audio: +44 207 162 0125
QuickSwap MSD interfaceAvailable October 1st, 2005
• Replacement for NoVent
• Remove column w/o venting
• Backflush mode
• Inlet maintenance reverse flow
• Removes heavies from column
• Maintain constant flow to MSD
MS
Transferline
Column
EPC inlet
Microfluidic technology
Requires performance turbo for flow rates > 2 cc/min
Introduction to eMethods and the new MSD Chemstation
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 55Chairperson: John Vis, Audio: +44 207 162 0125
What are eMethods?
• eMethods is short for Electronic Methods.
• Can now package existingGC/MSD methods
• Export them to other 5973 MSDs
• Export them to 5975 inert MSDs
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 56Chairperson: John Vis, Audio: +44 207 162 0125
eMethod Benefits
• Fast setup of new 5975 inertMSD
• Method sharing between5973 and 5975 MSDs
• Download apps from websiteand import entire method intoMSD ChemStation
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 57Chairperson: John Vis, Audio: +44 207 162 0125
New 5975 inert eMethods
• PBDE Flame Retardants
• SIM/Scan – Pesticides in Water and Food
• VOCs via Headspace – European method
• Residual Solvents via Headspace
• NIDA Drugs
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 58Chairperson: John Vis, Audio: +44 207 162 0125
MSD ChemStation Software
• Two GC/MS perChemStation
• GC and GC/MS control
• Retention time locking
• Custom reports
• eMethods
• Four configurations
• Enhanced mode
• EnviroQuant
• DrugQuant
• Reformulated Gas
Most popular GC/MS software in the world
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 59Chairperson: John Vis, Audio: +44 207 162 0125
Instrument Control Enhancements
• Combined Top/Instrument Control Screen
• New icons for quick access to tune, EMF and print
• Configurable sequence table
• More right mouse click functions in editing sequences
• Printing selected sections of method and sequence
• Import of GC ChemStation method parameters(excluding 5890 GC)
• Display of IP address(s) of GC and MSD instruments
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 60Chairperson: John Vis, Audio: +44 207 162 0125
Data Analysis Enhancements
• Explorer-style navigation tool
• Interactive editing of Integration Parameters
• Switchable right mouse click functions
• Spreadsheet-like SIM Parameters Table
• Icons for custom tools, DA options, variable watch
• Toolbar on/off capabiity
• Selectable DA color themes
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 61Chairperson: John Vis, Audio: +44 207 162 0125
Data Analysis Enhancements
•QEdit functionality is improved:
• Separate window for Baseline Display
• Generating reports directly from QEdit
• Custom configuring X-axis
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 62Chairperson: John Vis, Audio: +44 207 162 0125
Explorer Navigation
Explorer navigation
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 63Chairperson: John Vis, Audio: +44 207 162 0125
Interactive Edit Integration
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 64Chairperson: John Vis, Audio: +44 207 162 0125
Right Click Features
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 65Chairperson: John Vis, Audio: +44 207 162 0125
SIM Parameter Table
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 66Chairperson: John Vis, Audio: +44 207 162 0125
QEdit
Reports
Baseline Display
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 67Chairperson: John Vis, Audio: +44 207 162 0125
Library Options
• NIST05
• Wiley 7th ed
• Stan pesticide
• Pfleger/Maurer/Weber MS Drug Library, 3rd ed
• NIST Chemical Structures Library
NEW
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 68Chairperson: John Vis, Audio: +44 207 162 0125
Other Software Options
• Deconvolution Reporting Software
- New Revison A.02 released Aug 1, 2005
• Security ChemStation
• Headspace software
• RTL Pesticide Database : FAMES, Flavors, Volatiles Organics, Organotin derivatives, PCB congeners, Forensic toxicology
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 69Chairperson: John Vis, Audio: +44 207 162 0125
Expanding the portfolio of RTL / DRS Databases
• Updated RTL Pesticides Database, Nov 2005
Adding ~ 200 new compounds
• TIC Toxic Industrial Chemicals Database, Nov 2005
~ 800 compounds (VOC, SVOC, Dioxins, PCBs andPesticides)
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 70Chairperson: John Vis, Audio: +44 207 162 0125
GC/MS Configurations
• 5973N EI with diffusion pump
• 5975 inert EI with standard turbo pump
• 5975 inert EI with performance turbo pump
• 5975 XL EI/PCI/NCI with performance turbo pump
All systems available with choice of either 6850 or 6890 GCs
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 71Chairperson: John Vis, Audio: +44 207 162 0125
Summary
Agilent 5975Inert MSD
• Better reliability and quality
• Enhanced performance
• eMethods
• AutoSIM and SIM/scan
• Higher mass range
• AutoCI
• Sensitivity
• Software improvements
5975
Che
mSta
tion
eMet
hod
Sum
mar
y
Slide 72Chairperson: John Vis, Audio: +44 207 162 0125
Thank you for attending today’s Agilent e-Seminar.
Our e-Seminar schedule is expanding regularly.
Please check our web site frequently at:
www.agilent.com/chem
If you would like to receive regular updates please
update your account information under
Stay Current with e-Notes
Wrap-up E-Seminar Questions