Recent Advances on CdTe/CdZnTe detectors Tad Takahashi Institute of Space and Astronautical Science (ISAS) Japan Aerospace eXploration Agency (JAXA) and Univ. of Tokyo Acknowledgement : Z. He(Michigan), P.Luke (LBNL), F. Harrison(Caltech), O. Limousin(CEA), C. Szeles (eV), R. Ohno(ACRORAD), J. Matteson (UCSD) For High Energy PHOTON NDIP08 T.Takahashi
Transcript
Page 1: Advances e detectors

Recent Advanceson CdTe/CdZnTe detectors

Tad TakahashiInstitute of Space and Astronautical Science (ISAS)

Japan Aerospace eXploration Agency (JAXA)and

Univ. of Tokyo

Acknowledgement : Z. He(Michigan), P.Luke (LBNL), F. Harrison(Caltech), O. Limousin(CEA), C. Szeles (eV), R. Ohno(ACRORAD), J. Matteson (UCSD)

For High Energy PHOTON

NDIP08

T.Takahashi

Page 2: Advances e detectors

Outline

• Demand and CdTe/CdZnTe

• Recent Progress on technology• Crystal and ASIC

• X-ray Imager• Pixel & Strips

• Gamma-ray Detector• Coded Mask /PET

• Compton Camera

• Summary/Future Prospects

NDIP08

T.Takahashi

Page 3: Advances e detectors

Demands

3

In the field of Medical ApplicationHomeland SecurityAstronomical Observation

Need Detector Material which can be used as an alternative to Si (in terms of Efficiency) an alternative to Ge (in terms of Operating Temperature) Good Energy Resolution similar to Ge (0.2%@662keV)

High Efficiency, above 10 keV upto 1 MeVPosition resolution a few hundred micronA detector can be operated at room temperature

Seems to be very difficult and would need another 10 to 20 years for the final answer

NDIP08

T.Takahashi

Page 4: Advances e detectors

All next generation telescopes need a Hard-X camera above 10 keV, where Si becomes transparent

4

NuSTAR Small Explorer

Two hard X-ray (6 - 79 keV) focusing telescopes

Launch August 2011

17h12m17h14m17h16m

-39d30m

-40d00m

NeXT Mission(JAPAN)

Launch 2013 Launch 2014

Simbol-X(ASI/CNES)

To take a photo ofSuperNova explosion

in hard X-ray

NDIP08

T.Takahashi

Page 5: Advances e detectors

With Newly developedHard X-ray Mirror

Mirror Detector10-20 m

Nagoya U.

NDIP08

T.Takahashi

Page 6: Advances e detectors

• High Z semiconductor (ZCd = 48, ZTe = 52), ! =5.9 g/cm3

• Room Temperature Operation or Cool Environment

P. Luke (2006)

100 keV

511 keV

Takahashi and Watanabe (2000)

CdTe/CdZnTe seem to be the only candidateat least, at this moment

NDIP08

T.Takahashi

Page 7: Advances e detectors

Slow mobility/ short lifetime of carriers are a bit problem for CdTe/CdZnTe

!"#$%&'

"(%&'

eh

)*+,'-$'./$#-&'0'(&,-%(-#$'-.(#'1"!#.%(-&'0#$

!

2

3'+'!#1%(-%(+4

$%+'-%(+4

cathode

anode

eh

"

2mm

cathode

anode

eh

"

We now know how to handle this Charge Collection Issue (Coplanar Grid/Cross Strip etc.)

For the case of thin detector, and with high bias voltage (400 V/0.5mm), can collect full charge.

Thick vs Thin approaches, I’ll explain later

Signal is Depth Dependent

Takahashi and Watanabe (2000)

NDIP08

T.Takahashi

Page 8: Advances e detectors

CdTe/CdZnTe seem to be the only candidateat least, at this moment

137Cs

FWHM2.1 keV0.3 %

662keV

CdZnTe

epoxy

gold stud

ASIC

CdTe/ACRORAD

NuSTAR/Caltech

Takahashi et al. 2005

Thin CdTe diode at -20 deg

260 eV at 6.4 keV

NDIP08

T.Takahashi

Because, they’ve already shown good performance

Page 9: Advances e detectors

CdTe/CdZnTe Commercial ProductsAmptek AXION

aguila (US)

Radiation-Threat-DetectorX-ray Spectrometer

CdTe dental panoramic digital imaging system

NDIP08

T.Takahashi

Page 10: Advances e detectors

Courtesy of AJAT (Finland)

CCDScintillator

CMOS

CdTeCMOS

NDIP08

T.Takahashi

Page 11: Advances e detectors

Hard X-ray ImagerIntegrated type

Commercial Products

AJAT, Finland

1 dim Imager (30cm)

100 micron pitch

NDIP08

T.Takahashi

Page 12: Advances e detectors

Technologies

Crystal&

ASIC

NDIP08

T.Takahashi

Page 13: Advances e detectors

Who makes CdTe/CdZnTe Crystal?

From a review talk by P.J. Sellin (2005)

NDIP08

T.Takahashi

Page 14: Advances e detectors

Recent Advances on TechnologiesACRORAD (JAPAN)

Quartz ampoule

CdTe poly crystal

Te-rich Cl-doped Cd-Te solvent

Heater unit

Cl-doped CdTe grown single crystal

CdTe single crystal seed

Travel Heater Method (THM)

Careful treatment of post heating

Very uniform wafer

Large Single Crystal

THM

1st Large Scale CdTe Camera in Space (INTEGRAL)

NDIP08

T.Takahashi

Page 15: Advances e detectors

Recent Advances on Technologies

2m

1mCdZnTe 4x4x2mm

eV Products (High Pressure Bridgman and improved method)

Large Crystal

SWIFT !-ray satellite (2004-)

NDIP08

T.Takahashi

Page 16: Advances e detectors

Recent Advances on TechnologiesREDLEN

succeeded to make CZT by THMLarge Single Crystal

Chen et al. JAP, 2008

662 keV 1. 18% (7.8 keV, FWHM)without additional signal correction)

2 x 2 x 1.05 cm3

monolithic pixel pixel size 2.46 mm

IR imagevery small Te

precipitate/inclusion

NDIP08

T.Takahashi

Page 17: Advances e detectors

We need ASIC : If you need CdTe/CdZnTe imaging detectors

Level-sensitive

Discriminator

Semigaussian“fast” shaper

Monostable(fixed width)

Semigaussian“slow” shaper

ChargeIntegrator(preamp.)

S/H Vss

Vdd

TriggerOut

TAVA

Multiplexer

analog out

ADCSpectrum

Photon Counting

Need 50 e- at 0pF for both cases

NDIP08

T.Takahashi

Page 18: Advances e detectors

Hard X-ray Camera

For photonsabove 10 keVbelow 100 keV

NDIP08

T.Takahashi

Page 19: Advances e detectors

CdTe Photon-counting imager using XPAD chip

Basalo et al. , NIMA 2008

0.7 mm thick CdTe20 kpixels (130!m x 130!m) 2 ms/frame 106 photons/pixel

NDIP08

T.Takahashi

Page 20: Advances e detectors

16 capacitors

CSA

toADC

for NuStar Satellite

32 x 32 array, 0.6 mm pitch 2 mm thick CdZnTe

CdZnTe Imager with Spectroscopic Capability

non-uniformity of the image comes from CZT, not from ASIC

Prototype (for HEFT)

Flat Image

0.5mm pitch, 2mm thickTwo hybrids: 24 x 48

NDIP08

T.Takahashi

Page 21: Advances e detectors

for NuStar Satellite

32 x 32 array, 0.6 mm pitch 2 mm thick CdZnTe

CdZnTe Imager with Spectroscopic Capability

non-uniformity of the image comes from CZT, not from ASIC

Prototype (for HEFT)

Flat Image

0.5mm pitch, 2mm thickTwo hybrids: 24 x 48

E

coun

ts

5 mm CdZnTe, 800 V, 1C Single pixel 155Eu 900 eV FWHM @ 86 keV

F. Harrison, 2008

NDIP08

T.Takahashi

Page 22: Advances e detectors

! HED: mosaic of 64 independent CdTe (Al/CdTe/Pt) cameras! Caliste 64: first prototype of detection unit

! See A. Meuris’ talk in this session

241Am spectrum at -10°C, 500V with the single events of the 64 pixels

0.66 keV fwhm @ 13.94 keV0.84 keV fwhm @ 59.54 keV

for Simbol-X SatelliteCdTe Imager with Spectroscopic Capability NDIP08

Page 23: Advances e detectors

5 cm

ISAS/JAXA

Large Area 1024 pixel CdTe Array(pixel size 1.4 x 1.4 mm2)

Large Area Hard X-ray Imager

0.5 mm thick

NDIP08

Page 24: Advances e detectors

X-ray Imager

201Tl99mTc

69-80 keV140 keV

Allow us to study where in the leaf absorb which kind of metal

(multi-tracer imaging)

NDIP08

Page 25: Advances e detectors

Fine Pitch CdTe StripsPreviously 1) Difficult to make strips on the barrier electrode (In) for CdTe diode.2) Wire-bond does not work on CdTe. With new electrode materialon CdTe (Al as anode/Pt as cathode)we have succeeded to make fine pitch Double Sided Cross Strip detector

CdTe

ASIC

ASIC

ceramic board with through holesIn/Au stud bump

wire-bonding

wire-bondingstrip electrode

strip pitch 100-400 micron

NDIP08

Page 26: Advances e detectors

x [ch]0 10 20 30 40 50 60

y [c

h]

0

10

20

30

40

50

60

0

20

40

60

80

100

120

140

31keV

x [ch]0 10 20 30 40 50 60

y [c

h]

0

10

20

30

40

50

60

0

20

40

60

80

100

120

14060keV

x [ch]0 10 20 30 40 50 60

y [c

h]

0

10

20

30

40

50

60

0

20

40

60

80

10081keV

x [ch]0 10 20 30 40 50 60

y [c

h]

0

10

20

30

40

50

60

0

20

40

60

80

100122keV

Results(imaging)Shadow Image

nut(M3)

nut(M2)washer(M3)

solder

(0.6mm)

!x~400!m Imaging Spectroscopy !!

various RIs(241Am, 133Ba, 57Co)

–20ºC, 500V bias

spectra

241Am133Ba57Co

NDIP08

Page 27: Advances e detectors

Gamma-ray

above 100 keV

NDIP08

T.Takahashi

Page 28: Advances e detectors

CdTe & CdZnTe detectors for gamma-rayThick Approach

Energyat room temperature P.Luke (LBNL)

NDIP08

T.Takahashi

Page 29: Advances e detectors

Gamma-ray Detector

Cross Strip CdZnTe

511keV3.1 % FWHM

California Breast Cancer Research Program, CBCRP Grant Number 12IB-0092

For a Large NIH Program 1 mm spatial resolution, ~2% energy resolution at 511 keV

by C. Levin of Stanford,and J. Matteson et al. 2008

NIH Grant #R01CA120474

Edge on geometryEff. of 86 % for 511 keVby 4cm CZT

Depth Correctionby using Cathode/Anode

Info.

NDIP08

T.Takahashi

Page 30: Advances e detectors

Gamma-ray Detector

Mini Coded Mask (cross strip CZT)

J. Matteson &Aguila Tech., 2008

supported by HSRAPA

Redlen CZT

DGAS specs:!- Image a 5 mCi source at >5 m in less than 10 sec, and localize it to <10 degrees!- Energy band of 40 - 250 keV!- Better than 10% energy resolution at 122 keV

NDIP08

T.Takahashi

Page 31: Advances e detectors

CdTe & CdZnTe detectors for gamma-rayThin Approach

40 layer = 20 mm thick CdTe

!Concept

Full Charge Collection (No Tail)-> Stack thin CdTe layersto get high efficiency.

Area 5 x 5 mmThickness 2.25 mm(three layers)

AmptekXR-100T-CdTe-STACK

Watanabe, TT et al. 2002

NDIP08

T.Takahashi

Page 32: Advances e detectors

Gamma-ray Detector

CdTe PET (1st Result) Achieved Spatial Resolution< 1mmArray of 1mmX1mmX5mm

5120 CdTe BAR

Rat Mouse

Brain

Tohoku Univ.

Press Release by K. Ishii, Tohoku Univ., Japan

Edge on geometry

NDIP08

T.Takahashi

Page 33: Advances e detectors

Gamma-rayCompton Camerafor High Sensitivity

Imaging

NDIP08

T.Takahashi

Page 34: Advances e detectors

Z. He et al. Michigan U. (2008)

• Cathode/Anode Ratio (Depth Info)• Timing Measurement (Drift Time) ( Multiple Interaction/ Depth Info)

Gamma-ray Detector

CdZnTe Compton Camera (Thick)

2!2!1.5 cm3 = 6 cm3 CZT(eV Products)

ASIC front-end(Gamma-Medica-Ideas AS)

NDIP08

T.Takahashi

Page 35: Advances e detectors

Room-temperature handheld CdZnTe !-ray imaging spectrometers with energy resolution ~ 1% FWHM at 662 keV

NDIP08

T.Takahashi

Page 36: Advances e detectors

Systematicerror shouldremain constant

Demonstration of !-Ray Imaging using a Single 1.5"1.5"1 cm3 CZT Detector (FOV=4pi)

angular resolution ~10 degrees at 667 keV

AccidentallyDetected a 10 !C 137Cs source next door

90deg

NDIP08

T.Takahashi

Page 37: Advances e detectors

Gamma-ray Detector

Si/CdTe Compton Camera (Thin)!"#"$%&%'()*+,-.'/012%345'6"44%7'"&'!"#$'

89%:'%;#413"<1$'=-3"5'*%4%&61#%>?)"@%44A@%'

:A44'B&%'A@'013'@C%'&%$&A<D%'2%"&B3%2%$@'

0312'EFF'G%H'@1'IFF'G%HJ

%&'&('&)"*+')",&-"'*('.%-*/"0"012'&A$6%'A@'A&'41:'K'2"@%3A"4'"$7'212%$@B2'10'%4%6@31$'

"31B$7'@C%'$B64%A'A&'&2"44'-L'M%&&'"N%6@%7'O5'P1##4%3'

Q31"7%$A$R

Concept (Takahashi et al. SPIE 2003).Narrow FOV (field-of-view)

Compton camera.Compton kinematics to

suppress backgrounds

34''+*5"0('67'89:';;'

<=&->'%&'?&@"+

A'+*5"0('67'89B:';;'

<=&->'C)$"'?&@"+

D'EF&<

NDIP08

T.Takahashi

Page 38: Advances e detectors

Gamma-ray Detector

Si/CdTe Compton Camera!"#$#$%&'

Angular resolution ~3 degrees at 511 keV 0 100 200 300 400 500 6000

50

100

150

200

250

energy

Z

DSSDs

CdTe Bottom

CdTe Side47.4

14.4

X

47.4

source direction

Compton cone

!"#$%&%'()$*+

,-%&%'()$*+

.//%0$1%23-45%+36*7$%89:.%"$;*$$+<

=>?%"$;%@A1

.//%0$1

NDIP08

()*+',*-'$-*./-0111-23-45567 T.Takahashi

Page 39: Advances e detectors

Good performance for both multiple and diffuse sources. Large FOV and good energy resolution

!"#$%

!"&$%

Si/CdTe Compton Camera NDIP08

'()*+,)-+.-)/0-1222-34-56678

ISAS, 2008

T.Takahashi

Page 40: Advances e detectors

Summary/Future Prospects1. CdTe and CdZnTe are now in the phase ofreal application.2. Large and Thin CdTe wafers are widely used for the commercial products of hard X-ray camera3. Photo Counting Detector (109-1010 photons/pixel /s)would be the next step.4. For gamma-ray detection, Thick approach and Thin approach both work.5. Homeland Security and Medical Imaging boost the development6. Space missions are always one step ahead in terms of technological requirements, which is good.

NDIP08

T.Takahashi