Home > Documents > FlexiMag Plus EN - Ventiladores Pulmonares · Upper LED for distance viewing of alarms Maximum...

FlexiMag Plus EN - Ventiladores Pulmonares · Upper LED for distance viewing of alarms Maximum...

Date post: 30-Jul-2018
Category:
Upload: vancong
View: 213 times
Download: 0 times
Share this document with a friend
2
C M Y CM MY CY CMY K FlexiMag Plus EN.pdf 1 27/04/16 13:43
Transcript

C

M

Y

CM

MY

CY

CMY

K

FlexiMag Plus EN.pdf 1 27/04/16 13:43

Optional Accessories

Ventilation Modes

VCV (can be assisted); PCV (can be assisted); PRVC; PLV (neonatal); V-SIMV + PS; P-SIMV + PS; DualPAP; APRV; CPAP/PS; MMV and VS

NIV (No Invasive Ventilation) available in all ventilation modes

Internal Li-ion battery 11,8 VDC; with intelligent charger and autonomy up to 210 minutes (at standard operating conditions)

Features and Resources

High resolution 15-inch color LCD screen

Touchscreen or "turn and confirm" knob operation

Standard serial interface RS-232C

Keys for quick access: Standby mode (standby); Silence alarms (Mute); Timer for direct oxygen delivery (50% to 100%) for 2 minutes; Nebulisation syncronized with inspiratory phase (also adjusting this volume with tidal volume automatically); Tracheal gas insufflation (TGI) synchronized with expiratory phase; Inspiratory cycle manual trigger; Inspiratory and expiratory pauses with adjustable time; Freeze graphs with cursor for browsing; Manual trigger for extra breath.

Measurements of Regulated input gas pressures, O2 concentration, Airway Pressure Pmax, PEEP, Plateau and Pmean, as well as flow in the circuit for automatic compensation

Auxiliary pressure measurement; Adjustable sign in volume modes

Flow pattern adjustable in volume modes

Leakage compensation automatically; Circuit compliance compensation

Compensation for ET tube resistance (ATC)

Magnamed Remote Assistance

Monitoring and calculation of tidal volume according condition (ATPD/BTPS)

External input 100-240 VAC - 50-60 Hz - with automatic commutation

Galvanic cell or paramagnetic system (optional) for FIO2 monitoring

Distal or proximal external flow sensor (optional)

Automatic switch over to handle ventilation with single available gas in case the other fails

Log of trend and events occurred until 72 hours and records over 100,000 events

Tidal Volume: 5 to 3000 mL; Respiratory rate: 0 to 180 rpm

Rise time: 0 to 2.0s; Pause: 0 to 70% or 0 to 2.0s

Inspiratory pressure: 0 to 120 cmH2O; Support pressure (delta PS) 0 to 120 cmH2O

PEEP: 0 to 50 cmH20

Pressure Trigger: 0.0 to -20 cmH2O; Flow Trigger: 0.0 to 30 L/min

Flow cycling in pressure support: 5 to 80%

O2 concentration: 21 to 100%; Inspiratory time: 0.05 to 30s

Waveform of inspiratory flow: Square, Descending or Decelerated, Ascending or Accelerated, Sinusoidal or Senoid

CPAP: 1 to 50 cmH2O

High pressure: 1 to 60 cmH2O; Low pressure: 0 to 40 cmH2O

High time: 0.20 to 59.8s; Low time: 0.20 to 59.8s

I:E ratio:1599 to 10.1; Respiratory flow: 1 to 180 L/min

Backup option for spontaneous mode; Saving of last set parameters

Height setting for obtaining patient ideal weight (IBW)

Ventilation Parameter Settings

P0.1; Slow Vital Capacity; P/V Flex with possibility of graphic analysis;Pimáx (NIF); Trapped Volume

Special features

Curves of Pressure x Time; Flow x Time; Volume x Time; CO2 x Time (optional capnograph); SpO2 x Time (optional oxymetry)

Loops of Pressure x Volume; Volume x Flow; Pressure x Flow with saving up to 3 loops;

Bargraph with numerical monitoring of the instant pressure, peak pressure or plateau pressure; Presentation of 4 curves simultaneously

Ventilation Monitor

Maximum pressure, mean pressure and plateau pressure (numerical)

Auxiliary and estimated tracheal pressure (in ATC)

PEEP and intrinsic PEEP; Tidal volume expired and inspired

Volume minute; I:E Ratio; Total and spontaneous respiratory rate

Inspiratory and expiratory time; Inspiratory and expiratory airway resistance

FiO2; Static and dynamic compliance

Elastance; Time constant; Rapid shallow breathing index Tobin (RSBI)

Work imposed (WOBi); T Insp / T Total; Leakage % and flow

Oxymetry (optional): SpO2, pulse rate, perfusion index (PI), variability index (PVI)

Capnography (optional): EtCO2 and CO2 Insp.

Power On Self Test Checks for integrity of sensors and valves;

Optional diagnosis for checking the integrity of sensors and valves

Calibration options in main menu

Leak measurement and complacency of respiratory circuit

Release safety valve; ACTIVE overpressure valve; Safety relief valve 100 cmH2O

Upper LED for distance viewing of alarms

Maximum Pressure - High/Low Alarm OFF: 0 to 120 cmH2O

PEEP - High/Low Alarm OFF: 0 to 80 cmH2O

Total tidal volume - High/Low Alarm OFF: 0 to 3000 mL

Respiratory rate - High/Low Alarm OFF: 0 to 200 rpm

Volume Minute - High/Low Alarm OFF: 0 to 99 L

FiO2 - High/Low Alarm OFF: 18 to 100%; EtCO2 - High/Low Alarm OFF: 0 to 80 mmHg (1); CO2 Insp - High/Low Alarm OFF: 0 to 80 mmHg(1)

SpO2 - Low Alarm: OFF; 0 to 100%(1)

Respiratory rate - High/Low Alarm: OFF; 0 to 240 bpm(1)

Automatic adjustment of alarm limits: OFF; 10%, 20% and 30% of mean values; Time of apnea: OFF; 0 to 60s

Low battery; Heart frequency High/Low off – 150 bpm

Low O2/Air pressure; Low and High Air and O2 Inlet Pressure

Respiratory circuit disconnection or obstruction

Lack of electrical energy; Inoperative(1) Optionally available

Alarm Systems and Safety

Physical and Environmental Specifications

Connection to oxygen source: Oxygen inlet - DISS Standard

Gas supply pressure: 36 to 150 psi (250 to 1034 kPa)

Hose and dimensions: compliance with ISO 5359:2000

Dimensions (basic unit): H 1335 mm x W 453 mm x D 542 mm (with handle)

Weight: 18 kg

Operation: temperature of -10 to 50°C, barometric pressure 500-1200 cmH2O and relative humidity between 5-95%

Trolley with castors and locks; Articulated arm with support

Distal flow sensors; Adult patient respiratory circuit; Humidifier support

Extension for O2 and extension for Air

Standard Accessories

O2 concentration measurement cell (paramagnetic)

Heated humidifier with adult chamber and mounting support

Pediatric patient respiratory circuit; Neonatal patient respiratory circuit

Proximal flow sensors and lines; Sensor for oxymetry

Sensor for Capnograph; Compressor