1 PZ PZ PZ PZ LASER LASER LASER LASER IPL IPL IPL IPL PZ101 PZ101 PZ101 PZ101 Operation Operation Operation Operation Manual Manual Manual Manual Zhengzhou Zhengzhou Zhengzhou Zhengzhou PZ PZ PZ PZ Laser Laser Laser Laser Slim Slim Slim Slim Technology Technology Technology Technology Co., Co., Co., Co., Ltd Ltd Ltd Ltd
Transcript

1

PZPZPZPZ LASERLASERLASERLASER IPLIPLIPLIPL

PZ101PZ101PZ101PZ101

OperationOperationOperationOperation ManualManualManualManual

ZhengzhouZhengzhouZhengzhouZhengzhou PZPZPZPZ LaserLaserLaserLaser SlimSlimSlimSlimTechnologyTechnologyTechnologyTechnology Co.,Co.,Co.,Co., LtdLtdLtdLtd

2

TTTTableableableable ofofofof ContentsContentsContentsContents

1.1.1.1. Safety Instruction………………………………………………………………….4

1.1 Electric Safety ………………………………………………………………………………… 4

1.2 Movement and Transportation safety ………………………………………………………… 4

1.3 Optical Safety………………………………………………………………..………………… 5

1.4 Fire Caution of the Treatment Room…………………………………………………5

1.5 Safety Equipment…………………………………………………………….………5

1.6 Safety Signs and Marks………………………………………………………………………..6

2. Products Descriptions……………………………………………..………………… 7

2.1 Equipment Specification…………………………………………………………………...…7

2.2 Operation Specification…………………………………………………………………….....8

2.3 Treatment Parameter……………………………………………………………………………9

2.4 System Configuration……………………………………………………………………….9

3. Principle and Features for Treatment…………………………………………10

3.1 Treatment Principle …………………………………………………………………………10

3.2 Treatment Features……………………………………………………………………………11

4. Technique Parameter…………………………………………………………….…11

4.1 Parameter…………………………………………………………………………….………..…11

4.2 Relative Requirement……………………………………………………………………………12

5. Installation and Adjustment…………………………………………………………12

5.1 Installation………………………………………………………………………………………..12

6. Operation…………………………………………………………………………16

6.1 Start ing ………………… …………………………………………………………16

3

6.2 Operation Steps……………………………………………………………………………….16

6.3 Treatment steps………………………………………………………………………………. 21

6.4Skin Test……………………………………………………………………………………..22

6.5 Cautions…………………………………………………………………………………...….23

6.6 Turning off the System………………………………………………………………………..23

7. Indications and Contraindication…………………………………………….…24

7.1 Indications…………………………………………………………………………………….24

7.2 Contraindication………………………………………………………………………………24

8. After-treatment Care………………………………………………………………..24

9. Maintenance and Troubleshooting……………………………………………25

9.1 Cleaning the Unit……………………………………………………………………….25

9.2 Cleaning the Light Guide……………………………………………………………………..25

9.3 Trouble eliminating…………………………………………………………………………25

9.4 Malfunction Information…………………………………………………………………….26

10. After Service Instructions…………………………………………...………….30

11. Contact Us…………………………………………………………………………….31

4

CCCChapterhapterhapterhapter 1111 SSSSafetyafetyafetyafetyThe chapter describes safety issues regarding the use of the IPL system. National

criterion was conformed earnestly during the design and production of the system, to

minimize the hazard for both people and the surroundings.

Caution:Caution:Caution:Caution:

Personnel operating or maintaining this equipment must read this manual

thoroughly before attempting to operate the IPL HR&SR System.

This Product carries out standard number: YZB/京 0442-2003 And GB-97061.11.11.11.1 ElectricElectricElectricElectric SafetySafetySafetySafety

Safety style: class I type B

The system requires 230 V/110V, unidirectional power source, the maximum

output Current is less than 7A.

2000W Voltage Stabilizer is required in order to reduce the interference of the

fluctuation of power net waves to the system and extend the lifetime of the system.

Cautions:Cautions:Cautions:Cautions:

After turning off the system, dangerous voltage still left. Removing the covers

creates a safety hazard. Because remaining high voltage presents inside the console

during course of the treatment.

Note:

The system requires international standard three-phase power plug, which owns

a grounded pin in the power cable. This protective grounding is essential for safe

operation.

1.21.21.21.2 MovementMovementMovementMovement andandandand TransportationTransportationTransportationTransportation SafetySafetySafetySafety

The unit consists of delicate components and long-distance transportation after

first installation and adjustment is not allowed. If long transportation is a must, please

contact with us. During the movement, the shock-reducing package originated from

the manufacturer must be used. Otherwise it might cause adverse effect to the Optical

system, influencing the energy output.

5

1.31.31.31.3 OpticalOpticalOpticalOptical SafetySafetySafetySafety

Cautions:

Intense pulse light with wavelength of 640-1200nm and 530-1200nm sourced

from the system, can create intensive energy. All personnel including clients in the

treatment room must wear protective glasses whenever the unit is in use

Cautions:

Intense pulse light emission may cause an eye hazard. Please follow warning and

instructions where the system is being used.

Cautions:

No reflective objects such as jewelry, mirrors and watches to reflect the intense

pulse light in treatment room.

Cautions:

Never look directly at the intense pulse light emitted from the treatment handle,

even when wearing the protective glasses.

Never direct the intense pulse light at any skin except the test patch and the

target area.

1.41.41.41.4 TreatmentTreatmentTreatmentTreatment RoomRoomRoomRoom andandandand FireFireFireFire SafetySafetySafetySafety

1>.Make sure that no explosive and flammable anesthesia drugs, liquid, gas

(such as alcohol, ether, oxygen, and the mixed gas of nitrous oxide) present in and

around the treatment area, in addition the protection measures should be well

prepared.

2>Never direct the intense pulse light at any other part except the target area

3>Make sure the caution signs are posted in a prominent place of the treatment

room.

1.51.51.51.5 SystemSystemSystemSystem SafetySafetySafetySafety EquipmentEquipmentEquipmentEquipment

This product strictly carries out the Registration Standard of the Medical Devices

to make sure the safety of using,standard number is : YZB/京 0442-2003

This product is designed with the following safety protection devices:

6

1 Circuit-- The separated safety circuit system will cut off the output of the intense

light when necessary thus to avoid the system working in abnormal condition

2 Keys of the System Console――This fitting is the chief switch of the power

supply for the system. The key should be kept by the responsible man after

turning off the system in case of the illegal operation and incorrectly operation.

3 Emergency Shut-off Knob――Shut down the system on emergency

Caution:

The emergency shut-off knob should be used carefully, the devices software will

be damaged once frequently used.

1.61.61.61.6 CautionCautionCautionCaution SignsSignsSignsSigns andandandand MarksMarksMarksMarks

ST-B IPL System adopts the following signs, please pay attention to them:

OFF ON STOP

① KeyKeyKeyKey switchswitchswitchswitch ② TTTThehehehe emergencyemergencyemergencyemergency shut-offshut-offshut-offshut-off knobknobknobknob

4 IIIIntensentensentensentense safetysafetysafetysafety markmarkmarkmark ④ voltagevoltagevoltagevoltage dangerdangerdangerdanger markmarkmarkmark

AC230V 10A 50HZ

⑤ GroundGroundGroundGround protectionprotectionprotectionprotection ⑥ PowerPowerPowerPower connectorconnectorconnectorconnector

⑦ LLLLightightightight emittingemittingemittingemitting ⑧ CounterCounterCounterCounter

O U T F A L L

IntenseIntenseIntenseIntense lightlightlightlight emittingemittingemittingemitting

windowwindowwindowwindow

countercountercountercounterHHHHandandandand switchswitchswitchswitch

7

⑨OutfallOutfallOutfallOutfall

● Key Switch is set on the above part of the machine;OFF is the chief switch for

power off ,ON controls the operation starting.

● The Emergency Switch is set in the front part of the machine; when necessary,

press it to cut off the power supply.

● The Intense Emitting Window is set in the front of the handle.

● Dangerous voltage presents in the connector of the power board inside the

machine.190v-365v voltage appears when the machine operating.

● Ground protection set in the back of the power plugs.

● Power connector set in the back of the system, right side, below.

● Handle switch in the lifting handle of the treatment handle to countering the

radiation times.

● The key of the counter is on the holder of the handle, It is used for countering

the accumulating times of the flashes.

● Outfall at the below part of the machine, when water inside the machine need

to be changed , press the blue circle, then the black stopper can be pulled out.

CCCChapterhapterhapterhapter 2222 SystemSystemSystemSystem DescriptionsDescriptionsDescriptionsDescriptions

2.12.12.12.1 DeviceDeviceDeviceDevice DescriptionDescriptionDescriptionDescription

①The IPL Skin-care system can generate intense pulse light to take effect on the

target issue for removal of pigment, blood streaks, unexpected hairs and skin

rejuvenating.

②The system utilizes the selective absorption elements of the light.

③The system is controlled by computer system which consists of following major

components: CPU, controlling console, intense light treatment head

④ The system console contains running module and computer system. The

embedded microprocessor monitors and controls the system continuously.

8

⑤Operating Interface consists of 8 inch color LCD touch screen, Key switch and

emergency shut off knob. The screen shows the system work program, treatment

parameters and number of light emitting. The treatment handle houses the

mechanisms that generate the intense pulse light and the light guide, designed for

operating by hand. (See Chapter 6)

2.22.22.22.2 IntroductionIntroductionIntroductionIntroduction totototo thethethethe OperationOperationOperationOperation

The main functions of the system are as followings:

Main effect: removing unexpected hair on armpits, body, face lips, hairline, bikini

and limbs. Improve the rough skin, enhance skin elasticity, repair

acne

Additional effect: Pathology of Pigments, Pathology of Blood Vessels, Ageing of

the Skin Black eye socket and other abnormal pigments.

Before and after

(hair removal)

9

(Skin rejuvenation)

2.32.32.32.3 TreatmentTreatmentTreatmentTreatment ParametersParametersParametersParametersThe following parameters are allowed to adjust by the operator and 20-50J/cm2

Adjustable,:Maximum energy frequency: Program 3 T1 T2 T3 T4 T5

9.9 05 9.9 05 9.9Minimum energy frequency: 20J/cm squ±2J/cm squThe maximum error between the screen energy and the test energy ≤±20%

Pulse type:Three sub-pulses, adjustable pulse duration and pulse delay. (Pleasesee chapter 6)

2.42.42.42.4 SystemSystemSystemSystem StructureStructureStructureStructure

Key Switch

Emergency Switch

Handle Piece

LCD touch screen

10

2.4.12.4.12.4.12.4.1 ListListListList ofofofof SystemSystemSystemSystem ComponentsComponentsComponentsComponentsLCD touch screenEmergency shut-off knobKey switchMain consoleTreatment handle support2 IPL treatment handle

2.4.22.4.22.4.22.4.2 SystemSystemSystemSystem ComponentsComponentsComponentsComponents DescriptionDescriptionDescriptionDescriptionLCD touch screen: Display the work status of the systemEmergency shut-off knob: Shut down the system on emergencyKey switch: Turn on and shut off the power supply of the systemMain console: House the cooling system, power supply system, operating system

and microprocessor.Treatment handle: Emit the intense pulse light and guide the light to the target

area.

Standard accessories:

CCCChapterhapterhapterhapter 3333 PrinciplePrinciplePrinciplePrinciple andandandand FeaturesFeaturesFeaturesFeatures forforforfor TreatmentTreatmentTreatmentTreatment

3.13.13.13.1 TreatmentTreatmentTreatmentTreatment PrinciplePrinciplePrinciplePrinciple

The ST-B IPL HR & SR system emits varied wavelength, intense pulse, and

broad-spectrum light. It can permeate the cuticle to the derma and take effect on the

abnormal pigment to break the abnormal pigment. Highly controlled flashes of light

are selectively absorbed by hair follicles and target tissue. The absorbed light heats

FunnelGlass

Eyepatch

Powersupply line

Cooling gel

Silicagel circles

Wateroutlet forStopcock

11

the hair follicle and the target tissue, which damages or destroys the growing potential

of the follicle, stimulates the proliferation of collagen and improving the

rearrangement of elastic fibers, finally achieving the purpose of hair removal and skin

rejuvenation.

3.23.23.23.2 FeaturesFeaturesFeaturesFeatures ofofofof thethethethe SystemSystemSystemSystem1 Mature products with Good mini design

2Error self-detecting and error correcting system

3triple light filter systems: Effectively filtering the harmful light

4 Silver catoptric system: remarkable optical effect

5Three cooling systems: Built-in Circle water-cooling system

Air cooling system

Semiconductor cooling device

6 Multiplex counter system: insuring the quality

7Water-electricity separator inside: insuring the system safety

8 Magnetic drive pump + metal water box controlled water circulation: lower noise,

longer lifetime

9.3 adjusting programs, every program with a submenu aim at different skin types.

CCCChapterhapterhapterhapter 4444 TechnicalTechnicalTechnicalTechnical SpecificationsSpecificationsSpecificationsSpecifications

4.14.14.14.1 TechnicalTechnicalTechnicalTechnical SpecificationsSpecificationsSpecificationsSpecifications

The system emits the intense pulse light with three pulses and varied wavelength,

its system specifications is as following:

Light source Intense pulsed light

Light guide Crystal light guide

Wavelength 640-1200nm, 530-1200nm (420-1200nm)

Fluency 20-50J/cm2

Pulse duration 11-149.7ms

Frequency ≤1P/3S

Spot size 8×34, 16×57mm(error≤0.2mm)

12

Cooling system Semi-conductor cooling system&air

cooling & embedded circle water cooled

Size 400 x 350 x 1030 mm

Weight (net) 37 kg

Fuse specification T10AL250V Φ5×20/ T16AL250V

Φ5×20

Input power Standby:450VA

Transient: 1500VA

Power requirement AC220V±22V, 50HZ±1HZ

4.2 RelevantRelevantRelevantRelevant requirementsrequirementsrequirementsrequirements

①Environment requirement: Operating Temperature: 10-30ºC;

Operating Relative humidity ≤80%

2 Transportation and storage temperature: -10℃~+40℃

3 Relative humidity: ≤80%

4 Atmospheric pressure: 500hPa~1060hP

CCCChapterhapterhapterhapter 5555 InstallationInstallationInstallationInstallation andandandandAdjustmentAdjustmentAdjustmentAdjustment

5.15.15.15.1 InstallationInstallationInstallationInstallation

5.1.15.1.15.1.15.1.1 InstallationInstallationInstallationInstallation

The IPL Skin-rejuvenating system is used in hospitals, clinics and beauty salons.

Only a little pre-installation preparation is needed. The installation procedures are to

be performed by technicians authorized by our company.

The installation procedures include the following content:

List of components and accessories

ComponentsComponentsComponentsComponents QQQQuantityuantityuantityuantity NotesNotesNotesNotes AAAAccessoriesccessoriesccessoriesccessories QuantityQuantityQuantityQuantity NotesNotesNotesNotes

Main console 1 IPL coupling

gel

2

Intense pulse

light treatment

2 Available: 640 -

1200nm, 530 -

Operator’s

manual

1

13

head 1200nm (optional

420-1200nm )

Switch key 2 Power supply

line

1

goggle 1 silicone ring

against water

leakage

10

Protective eye

guard

1 Service

guarantee

card

1

Treatment

handle support

1 Draining

stopper

1

Shock-absorbing

case

1

5555.1..1..1..1.2222 InstallationInstallationInstallationInstallation RRRRequirementequirementequirementequirementssss

14

Examine the treatment area before removing the package as the requirements

described in this chapter.

The system should keep enough distance from blocks such as radiators, walls

and desks.

The electrical and mechanical requirements of the system are as following:

① Single-phase 220V±10%, 10A, 50Hz (standard outlet according to country

and area).

② The system is grounded through the grounding conductor in the power cable.

This protective grounding is essential for safe operation.

③Unstable voltage cause bad influence to the system, the voltage-transformer

is necessary to the system.

④ For operation safety, the temperature requirement of the treatment room

should be about 20ºC, and the relative humidity should be below 80%.

During the operation of the system, the thermal power is about 100W. Air –

conditioning device is necessary to the system.

5.1.3 Installation

① Remove the package of the system and place the unit at expected place.

② Take out the treatment handle support and install it at according place. (see

picture 1)

Picture 1

③Fill the tank with distilled water or pure water, which is at the back of the

15

unit. Never fill the tank with mineral water and piped water. keep filling the water

until it running out of the tank, then turn the key of power supply to start the cooling

system, allowing it operating for 5min. The operation of filling water is finished. (see

picture 2 )

Picture 2

Caution:

The cooling water need to be changed one time per month.

5 Insert the power supply line. (see picture 3)

Picture 3

16

ChapterChapterChapterChapter 6666 OperationOperationOperationOperation

This Chapter contains operating instructions for the system.

After the system is installed well, the operator shall select suitable treatment

parameters according to the client’s skin property. The specific steps are as follows:

6.16.16.16.1 StartStartStartStart thethethethe systemsystemsystemsystem

Make sure the emergency shut-off knob loosed. Turn right the key clockwise tostarting the cooling system, waiting for seconds to igniting the treatment handle andLCD screen, the system start to work. (you can hear the sound of water circulating ifnormal )

6.26.26.26.2 OperationOperationOperationOperation stepsstepsstepssteps

6.2.1 Start the machine and it shows the LOGO interface, after 3 seconds it comes to the main

operating interface ( see picture 4 )

(picture4)

6.2.2 First of all is the language selection interface(see picture5):

17

(picture 5)

6.2.3 Select the flag icon corresponding to language, at the same time to enter the

function selection interface, it shows HR(hair removal), SR(freckle removal

and skin rejuvenation), VR(removing resaca)(see picture 6):

(picture 6)

6.2.4 Press HR or SR or VR to choose the different function and meantime enter the

normal mode main operating interface ( see picture 7 )

18

( picture 7 )

6.2.5 The main interface function of general mode

6.2.5.1 Interpreting for noted number on the interface

1111 standards for the treatment type as chosen before (HR for undesired hair removal, SR for skinrejuvenation)

2222 energy changing button .I can be changed from 20J/ cm²to 50J/cm² with down and us switchesAt the same time the parameter has a related change shown on the columnar stick (energychanging scope 20J/ cm²--50J/ cm²)

1111

8888

7777

6666

5555

4444

3333

2222

19

3333 reset the shots number of treatment ..When the hand-piece has one shots ,it will show one on thepulse counter button .If more ,it shows the related number that the hand-piece has.Press the resetbutton ,it will catch the original state .

4444 Capacity release mode, includes program 1program 2,program 3 program Program 1"representing a discharge of two pulses, one pulse interval, "Program 2" "Program 3" representing adischarge of three pulses, two pulse interval, and their discharge time and space is different

5555 it is a cooling control key (it is designed always on with machine working state )

6666 back to the former state (when necessary )

7777 pre-ignition key .It is used to controlling hand-piece lighting . “Star/stop” button is in the fourthrow of the main interface as well, there is represents pre-ignition key handle, press the buttonwhen working the lamp inside the handle be lightened. Press it again, it is stopped.8888 pulse width adjustment key. It will enter the following interface as pressed the button

6.2.5.2

The “pulse type”, press it then enter into the following interface (picture 8), the interface

shows that the selected features, the selected mode and discharge time (T1, T3, T5) and interval

time (T2, T4) of the specific parameters, in the “program 1” state, the pulse parameters are three

sets of adjustable parameters, in the “program 2, program 3” state, the pulse parameters are five

sets of parameters adjustable, press “+ -” for increase or decrease. When it is set well press “OK”

to return to the main operation interface (the adjustment of parameters will be restored to its

original state after shutdown, for the parameter adjustment range and default parameters, please

see “ table 1”.

( Picture 8)

20

Time parameter adjustment range table:

For Asiantreatment

handle T1 T2 T3 T4 T5energydensity

freckle SR 4.0-4.5 25-35 5.0-5.5 30-40 4.5-5.0 20-30

Age-spot SR 3.5-4.0 30-45 4.5-5.0 25-40 4.0-4.5 20-26

Red-bloodvessel

VR 2.5-3.0 25-35 4.5-5.0 20-30 5.6 20-28

Black eyesocket

SR 3.0 40-60 6.5 35-55 6.3 20-25

Wrinkleremoval

SR 2.8-3.5 25-35 5.0-6.6 25-60 5.0-6.3 20-40

Skinrejuvenation

SR 3.5 35-50 5.0-5.5 30-45 4.5-5.0 20-27

Hair removal HR 7.0 25-45 7.5-8.5 25-45 7.5-8.0 22-40

Hairremoval(accrding to color of

skin)

handle T1 T2 T3 T4 T5energydensity

white hairremoval

HR 7.0-7.5 20-30 7.5-8.0 20-30 7.5-8.5 P3/25-35

Skinrejuvenation

SR 3.5-4.0 25-30 4.0-5.0 25-30 4.5-5.5 P2/23-30

Red blood vesselremoval

VR 2.5-3.5 21-26 4.0-5.5 20 4.5-6.0 P2/22-30

Yellow hairremoval

HR 5.0-7.0 20-30 6.0-7.0 25-40 7.0-8.5 P3/22-35

Skinrejuvenation

SR 3.0-4.0 35-45 4.5-5.5 30-40 4.0-5.0 P2/20-30

Red blood vesselremoval

VR 2.0-3.0 25-30 4.5-5.5 20-30 4.5-5.5 P1/23-28

black hairremoval

HR 2.0-3.0 30-40 4.0-5.0 P1/20-25

TTTTableableableable 1111

6.36.36.36.3 TreatmentTreatmentTreatmentTreatment StepsStepsStepsSteps

21

Start the system and wait for the screen into the main menu.

A) Select suitable program by pressing “program 1”,“program 2” or“program 3”.

B) It is recommended to always use low setting in the first treatment. Use lower

frequency for skin.

C) Testing the client’s skin for selecting suitable energy level. Skin test is necessary

before the treatment for the client’s safety.

D) Apply a 1-2mm layer of cold coupling gel to the treatment site. Light pulses

should be triggered on cold gel, so spread gel only on immediate treatment area

and treat right away, then spread gel on next area.

E) Direct the light guide at the treatment site, apply minimal pressure and slightly

touch the skin, or keep the distance of 1-2mm from the skin. The light guide must

be in complete contact with the coupling gel.

F) Press down the emitting button that is on the treatment handle and trigger a shot.

Every shot accompanied buzzing sound; both counters on the treatment handle

and screen count one time.

G) Wipe off the gel, wait a few minutes and examine the treatment site.

H) After finishing the treatment, record and file the treatment parameters for every

client.

6.46.46.46.4 SkinSkinSkinSkin TestTestTestTest

Skin test is essential before the treatment for the safety operation. The details are

follows:

1) Skin test should be performed before the treatment for the client when the first

time. The test patch should be under the ears or other skins not influencing the

facial beauty.

22

2) For hair-removal with the Hair-removal treatment handle (640-1200nm

wavelength):

First, cut off the hair and leave only 1-2 mm. clean the test patch with pure water,

and apply a layer of cold coupling gel at the test patch. Take the treatment handle

from the supporter, and direct the light guide to the test patch. Trigger 2-3 shots (1/3

area of the spot size overlapping the treatment site is permitted.), and wait for 2-3

minutes to examine the test patch:

� If swelling and/or reddening around the hair follicles appear within a

few minutes, continue the treatment with the same settings.

� If no visible changes appear around the follicles, continue the test with

2j/cm² higher energy level.

If the skin shows adverse effect such as excessive reddening or swelling, lower the

energy level.

For skin rejuvenating (530-1200nm wavelength)

Clean the test patch with pure water and wash- milk; apply a layer of cold coupling

gel at the test patch. Take the treatment handle from the supporter, and direct

the light guide to the test patch. Trigger 2-3 shots (1/3 area of the spot size

overlapping the treatment site is permitted.), and wait for 2-3 minutes to

examine the test patch:

If lightly swelling and/ or reddening at the test patch appears several minutes later,

continue the treatment with the same settings of energy level and pulse type.

If no visible changes at the patch appear, continue the test with 2j/cm² higher energy

level. Re-perform the test for several times until the above situation happens.

If the skin show adverse effect or excessive reddening and/or swelling at the test patch

just after shot, continue the test with lower energy level.

6.56.56.56.5 CautionsCautionsCautionsCautions

All personnel including clientele must wear protective glasses when the system is in

use

If do skin-rejuvenating for all face, start the treatment from the ear-root, and spread to

all face

One time overlapping to the pathology site is allowed.

Place the treatment handle perpendicular to the skin, apply minimal pressure and

23

slightly touch the skin. The light guide should be in complete contact with the

gel. Leave no blank site and overlapping 1/3 of the spot size is allowed.

Better not shoot the light at upper eyelid

Use higher Settings of T2, T4 for darker skin

Use lower settings of T2, T4 for light skin

Warning:Warning:Warning:Warning:

ItItItIt isisisis recommendedrecommendedrecommendedrecommended totototo alwaysalwaysalwaysalways useuseuseuse thethethethe lowlowlowlow settingssettingssettingssettings inininin thethethethe firstfirstfirstfirst treatment.treatment.treatment.treatment.6.66.66.66.6 TurningTurningTurningTurning offoffoffoff thethethethe systemsystemsystemsystem

1>Loose the pedal switch or the emitting button

2>Place the treatment handle on the supporter

3>Turn the power supply key contra clockwise to first grade. Wait for some second

and turn to “Off”.

4>Clean the light guide with paper or cotton ball

Chapter 7 Indications and Contraindication

7.17.17.17.1 IndicationsIndicationsIndicationsIndications

Hair removal: Unexpected hairs on limbs, body, face and lips (most effective)

Pathology of the pigment: Speckles, age spot, sun-induced freckles and all kinds of

pigments sourced from cosmetics (use 560-1200nm curing handle better).

Pathology of the blood vessel: Red blood streak, bottlenose, erythematic spot and so

on

Aged skin: wrinkles, acne scar, low skin elasticity, skin slack, skin dark, black eye

circle, black spot and so on

7777.2.2.2.2 ContraContraContraContraindicationindicationindicationindication

� Pregnant woman

� Using photosensitive medicine the 560-1200nmlight

� Recently tanned skin (within the last month)

� History of keloid scarring

� Diabetes, unless under control

� Cancer, particular skin cancer

� Heart disease, epilepsy

Chapter 8 Post-treatment Care

24

● NoNoNoNo exposureexposureexposureexposure totototo sunlightsunlightsunlightsunlight

Clients should use SPF15 factor sunscreen and sunscreen cream for one month

after treatment. when stay out, clients should take beach umbrella or sunbonnet

against sunlight. Tanning after treatment sessions may enhance melanin regeneration,

which may result in hyper-pigmentation

● FreeFreeFreeFree fromfromfromfrom alwaysalwaysalwaysalways touchingtouchingtouchingtouching andandandand pressingpressingpressingpressing

Clients must do not press and scratch the treated area against infection.

● MakeupMakeupMakeupMakeup

Immediate makeup after the treatment is not allowed for better curative-effect,

advising no use of makeup for four to five days following treatment, for makeup ,no

using chemical cosmetic, the biological is better,. Remove the makeup against

infection and pigmentation when swelling and reddening appear at the treated site,

and inform the doctor. At the same time, clients take some medicine such as VitC,

VitE.

According to the normal physiology, client should take the next treatment

21dayslater,.one treatment course contains 3-5 curing times. After one treatment, the

wrinkles, spots, and so on, incurred part will have a obvious effect.

● FoodFoodFoodFood

Advising not eat the hot and spicy food because the skin is sensitive during this

period.

● OtherOtherOtherOther

Advise that improve other causes of disease, such as adjusting the endocrine situation

by taking Chinese herb of medicine, taking few medicines that include VitC and VitE.

Caution:Caution:Caution:Caution:

According to the normal physiological process, clients should take the next treatment

21 days later. The whole treatment needs 3-5 times.

Due to physical difference between individuals, a few of clients may have temporary

discomfort such as blister, deposit of pigment, swelling and so on. Advise using ice

bag to cover the treated area for 15-20 minutes. Use make up 7 days later after

treatment.

Chapter 9 Maintenance and Troubleshooting

9999.1.1.1.1 CleaningCleaningCleaningCleaning thethethethe UUUUnitnitnitnit

Clean the system exterior at least once a week. Wipe all surfaces with soft, damp,

25

non-abrasive cloth. Mild and neutral cleaning solutions may be used. Not spill liquids

in the unit.

Change the purified or distilled water once every month in the water pathway.9999.2.2.2.2 CleaningCleaningCleaningCleaning thethethethe LLLLightightightight GGGGuideuideuideuide

The light guide should be kept clean at all the time and should be cleaned after

each treatment.

To clean the light guide, first dry it with a lint-free cloth. Then clean it with a

cloth moistened with water-free alcohol and let it dry thoroughly.9999.3.3.3.3 TroubleshootingTroubleshootingTroubleshootingTroubleshooting

Warning

There are some high voltage sets in the machine. Incorrect operation may cause

serious result. As a routine, if you maintain the machine, you should cut off electrical

source first.

9.49.49.49.4 MMMMalfunctionalfunctionalfunctionalfunction InformationInformationInformationInformation

AnalysisAnalysisAnalysisAnalysis onononon simplesimplesimplesimple malfunctionmalfunctionmalfunctionmalfunctionssss

The chapter below shows some possible trouble and solution accordingly.

SymptomSymptomSymptomSymptom MethodsMethodsMethodsMethods ofofofof examinationexaminationexaminationexamination SolutionSolutionSolutionSolution

No power Power disconnected Check the electrical socket or

outlet.

Ineffective switch key Connect the two terminals of

key directly.

If it was bad, change it please.

Ineffective crash stop button Connect the two terminals of

crash stop button directly.

If it was bad, change it please.

Main board relay failed to

work

Check if there is 230V in

output.

If there is no 230V in the

output, change it please.

The touch screen is not in

normal:

No sound of the button or no

picture of the screen or no

data is displayed or

the button is not in power

Check if there is AC 230V in

the two wires of LCD and

main board

If the pin and wires were bad,

change it please.

Check the LCD screen

electricity supplier and data

conducting lines of display

device

No useful when a new data

conducting line was changed,

then change a new LCD please

26

Bad connection of data

conducting lines of the board

and LCD

No useful when a new data

conducting line was changed,

then change a new LCD

control board please

Check the emergency

shut-off button

Loose the emergency shut-off

knob by turning it clockwise

Abnormal symbols appears Check if there is some thing

wrong with chips

Change a same kind of chip, if

problem still persists, change

the LCD.

The system can’t irradiate

after pressing the handle

button

Check whether the system

has entered into the ready

mode.

Whether on the operation

mode.

Whether the handle button is

in normal.

Make sure the mode and the

operation is correct and the

handle button is in normal

The system failed to work Check the power cable and

fuse

Make sure the power cable and

fuse is ok

The output light turns to be

weaker.

Check whether the voltage of

the power supply is too low.

Whether the light head is

clean.

Whether the energy is too

low

Sufficient power supply and

keep the light head clean.

Water pathway is abnormal Check if there is AC 230V in

the wires of the pump

If not, make sure the normal

pressure. If it is ,change the

same kind of pump

The water tank is leaking. Dry the water, and check the

position leaking water

Drain the water, and then paste

the gap by 704704704704 glue sealing the

slight gap. If the gap were

large then need to change a

water tank

27

The connection position of

the light head and pedestal

is leaking

Check the connection place Change a O-ring seal

9.4.19.4.19.4.19.4.1 NoNoNoNo powerpowerpowerpower

☆SolutionSolutionSolutionSolution::::① Check the electrical socket or outlet.

② check the key switch and the emergency switch(the emergency switchshould not be pressed down).

③ Check the fuse is damaged or not

28

9.4.2.9.4.2.9.4.2.9.4.2.WaterWaterWaterWater cancancancan notnotnotnot fillfillfillfill inininin

9.4.3.9.4.3.9.4.3.9.4.3.WhenWhenWhenWhen presspresspresspress ““““readyreadyreadyready”””” thethethethe lamplamplamplamp cancancancan notnotnotnot bebebebe lightenedlightenedlightenedlightened,,,, andandandand theretheretherethere isisisis ““““buzzbuzzbuzzbuzz”””” sound.sound.sound.sound.

9.4.4.9.4.4.9.4.4.9.4.4. TheTheTheThe machinemachinemachinemachine workworkworkwork properproperproperproper butbutbutbut thethethethe triggertriggertriggertrigger isisisis notnotnotnot workingworkingworkingworking

☆SolutionSolutionSolutionSolution::::

29

9.4.59.4.59.4.59.4.5.WaterWaterWaterWater leakageleakageleakageleakage overoveroverover thethethethe handlehandlehandlehandle connectorconnectorconnectorconnector

☆SolutionSolutionSolutionSolution::::

9.4.69.4.69.4.69.4.6.MachineMachineMachineMachine workworkworkwork proper,proper,proper,proper, butbutbutbut cancancancan notnotnotnot displaydisplaydisplaydisplay onononon screen.screen.screen.screen.

☆SolutionSolutionSolutionSolution::::

Fix the Data wire

30

Note: If the system malfunctions during start-up or normal operation, turn the

system off immediately and contact with our service department. Only the technician

authorized by our company can service the system. Otherwise, we bear no

responsibility.

Chapter 10 After-sale service instructions

We earnestly state:� Strict examination presents with all products before sale. ONE-years

guarantee of service without payment from the date of purchasing is

authorized to clients. During the guarantee period, a malfunction due to the

quality of machine shall be responsible by manufacturer. (the main power

system is gived 1 years guarantee ,the treatment handle used less than 20,000

spots in one year can be maintained freely )

� If there is some problem during the warranty, we provide the unconditionally

maintain. The machines are provided technical support for life time, And if

there is some spare parts need to be changed we only charge for the cost fees.

Malfunctions caused by following reasons are out of guarantee service:

A. Malfunctions caused by incorrect operation

B. Malfunctions caused by unauthorized transferring, dismantling, collision

or other human errors.

C. Malfunctions caused by force majeure.

D. Energy reduction caused by bad environment of moisture and dust.

E. Malfunctions caused by unsound spare parts and accessories

� The system is maintained for lifetime, when the components need to be

changed, we only charge the cost fee and traffic fee accordingly.

31